Oscilloscope

An oscilloscope is an electronic instrument that graphically displays voltage signals as waveforms, illustrating how electrical signals vary over time to enable visualization, measurement, and analysis in fields such as electronics, engineering, and physics. It operates by converting input signals into a visual trace on a screen, typically using a cathode-ray tube (CRT) in analog models or digital processing in modern variants, with key parameters including bandwidth (e.g., ≥100 MHz), sampling rate (e.g., ≥100 MS/s), and resolution (e.g., 8-bit ADC). Essential for debugging circuits, verifying signal integrity, and characterizing phenomena like transients or repetitive pulses, oscilloscopes support measurements such as risetime (≤3.5 ns) and deflection accuracy (±2% full scale). The invention of the oscilloscope traces back to 1897, when German physicist Karl Ferdinand Braun developed the cathode-ray oscilloscope, utilizing an electron beam deflected by electric fields to trace waveforms on a fluorescent screen, laying the foundation for modern signal visualization tools. Early analog oscilloscopes, prominent from the 1930s, relied on CRT technology for direct, continuous signal display, while triggered models introduced in 1946 stabilized repetitive waveforms for precise observation. The first digital storage oscilloscope was introduced by Nicolet in the early 1970s; by the late 20th century, digital storage oscilloscopes (DSOs) had become widespread, employing analog-to-digital converters (ADCs) to sample signals at rates ≥100 MS/s, store data in memory, and enable features like averaging, pre-trigger viewing, and equivalent-time sampling for capturing infrequent events. At its core, an oscilloscope functions through interconnected systems: the vertical axis amplifies and scales the input voltage (e.g., 5 mV/div to 5 V/div via BNC connectors), the horizontal axis generates a time-based sweep (e.g., via sawtooth waveform), and the trigger system synchronizes the display to signal edges or levels for stable traces. Probes—passive (e.g., 500 MHz bandwidth) or active—interface with circuits without significant loading, while modern digital models incorporate raster displays, cursors for measurements, and bandwidth limits (e.g., 20 MHz at -3 dB) to filter noise. Standards from organizations like NIST ensure performance, including spot size ≤0.55 mm and storage duration of at least 15 minutes for captured waveforms.

Fundamentals

Definition and Purpose

An oscilloscope is a testing instrument that graphically displays voltage signals as a function of time or other variables, converting electrical inputs into visual waveforms on a screen. This visualization allows users to observe the shape, amplitude, and timing characteristics of electrical phenomena that would otherwise be invisible. Originally developed in the late 19th century to observe varying electrical magnitudes, the oscilloscope has evolved into an essential tool for troubleshooting, design verification, and education in electronics. Invented by Karl Ferdinand Braun in 1897 using a cathode ray tube, it enabled the real-time monitoring of signal behaviors, a capability that has remained central to its purpose in modern applications. The core functions of an oscilloscope include measuring key signal parameters such as amplitude, frequency, phase, rise time, and distortion, providing insights into both analog and digital signal behaviors. For instance, it can quantify phase differences between waveforms by analyzing time delays relative to their period. These measurements support the analysis of waveform integrity, distinguishing subtle variations like noise or harmonic content that affect performance. In various fields, the oscilloscope's importance lies in its ability to enable real-time observation of transient events, such as voltage spikes or rapid signal changes, which are undetectable by static instruments like multimeters. This dynamic visualization is critical for diagnosing issues in electronic circuits, verifying sensor outputs, and ensuring compliance in manufacturing processes.

Basic Principles

Analog oscilloscopes operate on the principle of electron beam deflection within a cathode-ray tube (CRT). An electron gun generates a focused beam of electrons accelerated toward a phosphor-coated screen. Electrostatic deflection plates, positioned vertically and horizontally, apply electric fields to steer the beam: the vertical plates control the amplitude of the input signal, while the horizontal plates manage the time axis. When the beam strikes the phosphor, it produces a visible trace that persists briefly due to phosphorescence, allowing the waveform to be observed as the beam sweeps across the screen. Signal amplification is essential for processing input signals to drive the deflection plates effectively. The vertical amplifier boosts the input voltage to produce proportional beam deflection, with sensitivity typically expressed as the deflection factor, calculated as volts per division (V/div), where the beam displacement per division on the graticule corresponds to a specific input voltage. Similarly, the horizontal amplifier processes the sweep signal to ensure linear positioning along the time axis. These amplifiers maintain signal integrity across a wide bandwidth, preventing distortion while scaling the signal to match the CRT's deflection range. The time base generates a linear horizontal sweep to create a uniform time axis for the waveform display. This is achieved using a ramp voltage waveform produced by a sweep generator, which increases linearly over time to drive the horizontal deflection at constant speed, followed by a rapid flyback to reset. The slope of the ramp determines the sweep speed, ensuring that time intervals are accurately represented on the screen. In digital oscilloscopes, the input signal is digitized through analog-to-digital conversion, requiring adherence to the sampling theorem to faithfully reconstruct the waveform. The Nyquist criterion states that the sampling rate f_s must exceed twice the maximum signal frequency f_{\max} to prevent aliasing, expressed as f_s > 2 f_{\max}. This ensures that the discrete samples capture sufficient detail for accurate waveform reproduction without frequency folding artifacts. To minimize loading on the measured circuit, oscilloscopes employ high input impedance, standardized at 1 MΩ in parallel with a small capacitance (typically 10–20 pF). This value balances signal fidelity with probe compatibility, drawing negligible current from low-power sources while allowing passive probes to form a 10:1 attenuation network without significantly altering circuit behavior.

Key Components

The core of an oscilloscope consists of several integrated hardware elements that process and visualize electrical signals. The display serves as the output interface, where in analog models, a cathode-ray tube (CRT) features a phosphor-coated screen that emits light upon electron beam impact, enabling visual persistence for waveform observation. This phosphor glow allows traces to remain visible briefly after the beam passes, facilitating the viewing of repetitive signals up to frequencies around 1 GHz. In digital oscilloscopes, the display shifts to LCD or LED panels, which render digitized waveforms with vertical resolutions typically ranging from 8 to 12 bits, providing 256 to 4096 discrete voltage levels for precise signal representation. The vertical amplifier processes the input signal for Y-axis deflection, amplifying weak voltages from DC levels up to high frequencies while maintaining signal integrity. It employs attenuators and gain stages to scale the signal amplitude, supporting sensitivity settings from as low as 1 mV per division to handle a broad dynamic range without distortion. This component ensures the vertical deflection plates of the CRT (or digital equivalent) receive a proportional voltage, directly influencing the waveform's height on the display. For horizontal positioning, the horizontal amplifier works in tandem with the time-base generator to create a linear X-axis sweep. The time-base generator produces a precise sawtooth waveform, which ramps steadily to move the electron beam (or digital cursor) across the screen at controlled speeds, typically adjustable from nanoseconds to seconds per division. This setup allows time-domain visualization, where the slope of the sawtooth determines the horizontal scale, enabling accurate timing measurements. The power supply unit delivers the necessary voltages to all subsystems, including high-voltage outputs of 5 to 15 kV for accelerating electrons in the CRT anode to strike the phosphor screen effectively. It also provides stable low-voltage rails, such as +5 V to +15 V for logic circuits and amplifiers, ensuring reliable operation across the instrument. Proper regulation in this supply is critical to minimize noise and maintain deflection accuracy. The trigger circuit detects specific signal characteristics, such as rising or falling edges, to synchronize the horizontal sweep and stabilize the display for repetitive or single events. It uses comparators to monitor voltage thresholds, initiating the time-base only when the input meets predefined criteria like slope or level. In digital oscilloscopes, the analog-to-digital (A/D) converter quantizes the amplified input signal into discrete digital values for storage and processing. This component samples the analog waveform at rates often exceeding 1 GS/s, converting continuous voltages into binary data with resolutions matching the display's bit depth, typically 8-12 bits vertically, to preserve signal fidelity before rendering. Probe connections interface directly with the vertical amplifier's inputs to deliver the signal for this conversion process.

Operation

Vertical Deflection

The vertical deflection system in an oscilloscope processes the input signal to control its amplitude representation on the display, enabling accurate measurement of voltage variations along the Y-axis. This system amplifies, attenuates, and filters the signal before applying it to the vertical deflection plates of the cathode-ray tube (CRT) or the equivalent in digital displays, ensuring the waveform's height corresponds to its voltage levels. Input stages of the vertical deflection system include coupling selectors that determine how the signal is passed to the amplifier. DC coupling transmits both alternating current (AC) and direct current (DC) components of the input signal, allowing measurement of absolute voltage levels including any DC offset. In contrast, AC coupling employs a high-pass filter to block the DC component, centering the waveform around zero volts on the display and isolating AC variations, which is useful for signals with large DC offsets that might otherwise push the trace off-screen. Ground reference mode disconnects the input signal entirely, setting the trace to the zero-volt line to establish a baseline or verify calibration. Sensitivity control adjusts the vertical gain of the system, typically in a 1-2-5 sequence from 1 mV/div to 10 V/div or higher, scaling the signal to fit the screen's vertical divisions. This setting determines how much voltage produces one division of deflection; for a standard 8-division graticule, the full-scale deflection voltage V_{fs} is given by V_{fs} = sensitivity \times 8. For example, at 5 V/div, a 40 V peak-to-peak signal fills the screen. Position control shifts the amplified signal vertically on the display without altering its amplitude, allowing users to center the waveform for optimal viewing or to align multiple traces. This is achieved by adding a DC offset to the vertical amplifier output. Polarity inversion reverses the deflection direction for a selected channel, flipping the waveform upside down on the screen; this is particularly useful for comparing phase relationships between signals without rewiring probes. Bandwidth limits in the vertical system apply low-pass filtering to restrict high-frequency components, typically reducing the response to 20 MHz to minimize noise and unwanted transients while preserving lower-frequency signal details. The limit is specified in MHz and can be toggled on or off, with full bandwidth restoring the system's native frequency range.

Horizontal Deflection and Sweep

The time-base generator is a core component of the oscilloscope's horizontal deflection system, responsible for producing a linear ramp voltage that drives the electron beam or equivalent in digital displays across the screen at a constant speed, thereby establishing the time axis for waveform visualization. This ramp voltage, often in the form of a sawtooth waveform, increases steadily during the sweep period and resets abruptly, enabling precise timing representation. Sweep speeds are adjustable, typically ranging from 1 µs/div to several seconds/div, to accommodate signals spanning high-frequency pulses to slow-varying phenomena. Sweep modes determine how the time-base generator operates relative to the input signal. In free-run mode, also known as auto mode, the sweep occurs continuously without synchronization to a trigger, ensuring a persistent display even in the absence of a signal by relying on an internal timer. Triggered mode, or normal mode, initiates the sweep only upon detection of a specific signal event, providing stable, synchronized waveform views; this mode incorporates holdoff, an adjustable delay period following a trigger during which subsequent triggers are ignored, preventing premature retrace and stabilizing displays of complex or repetitive patterns like burst signals. Delayed sweep enhances resolution by introducing a programmable time offset, allowing the main time base to capture an overview while a secondary, faster time base zooms into a specific portion of the signal for detailed examination; this dual time-base approach uses the main sweep to trigger the delayed one after a set interval. In X-Y mode, the time-base generator is bypassed entirely, with an external voltage applied directly to the horizontal deflection input to plot one signal against another, producing patterns such as Lissajous figures for phase and frequency analysis. Horizontal sensitivity control adjusts the gain of the horizontal amplifier, scaling the deflection for non-time-based applications like X-Y plotting where precise voltage-to-division ratios are needed.

Triggering Mechanisms

Triggering in an oscilloscope synchronizes the horizontal sweep to specific events in the input signal, ensuring a stable and repeatable display of the waveform by initiating the acquisition at a consistent point. This mechanism prevents the trace from drifting across the screen, allowing users to observe repetitive signals clearly and capture transient events accurately. The trigger system monitors the signal for predefined conditions, such as voltage thresholds or timing anomalies, and starts the sweep generator once those conditions are met. Oscilloscopes support various trigger types to isolate specific signal characteristics. The edge trigger, the most fundamental type, activates when the signal crosses a selected voltage threshold on either the rising or falling edge, providing synchronization for standard periodic waveforms. Pulse width triggering detects pulses based on their duration relative to a set threshold, useful for identifying anomalies like short or long pulses in digital signals. Video triggering is designed for television or video signals, synchronizing to horizontal, vertical, or field components to display frame lines or fields stably. Glitch triggering captures narrow, aberrant pulses or irregularities that might otherwise be missed, such as dropouts in logic signals. The trigger source determines which signal the oscilloscope uses for synchronization. Internal triggering selects one of the input channels (e.g., CH1 or CH2) as the reference, ideal for self-contained measurements on a single waveform. External triggering uses a separate input connector for an unrelated synchronization signal, such as a clock line in a circuit. Line triggering synchronizes to the AC power line frequency (typically 50 or 60 Hz), helpful for observing hum or noise related to mains supply. Level and slope controls fine-tune the trigger point for precision. The trigger level sets the voltage threshold at which the event must occur, adjustable via a knob that positions a marker on the display, ensuring the sweep starts at the desired amplitude point within the signal range. Slope selection specifies the direction of the edge—positive (rising) or negative (falling)—to match the signal's transition, preventing false triggers on unwanted edges. Holdoff introduces a variable time delay after each sweep, during which the trigger circuit ignores subsequent events, stabilizing complex or bursty waveforms that might otherwise produce multiple overlapping traces. This control, adjustable from microseconds to seconds, allows the sweep to complete fully before re-arming, particularly useful for signals with varying repetition rates. Trigger modes dictate how the oscilloscope responds to signal absence or irregularity. In normal mode, the sweep occurs only upon a valid trigger event, resulting in a blank screen if no trigger is detected, which is essential for precise synchronization. Auto mode forces periodic sweeps using an internal timer even without a trigger, displaying a baseline or noise if the signal is absent, facilitating initial setup or observation of DC levels. These modes integrate with the sweep generator to maintain display continuity while prioritizing trigger stability.

Display Modes

Oscilloscopes primarily operate in single-trace mode, displaying a single waveform in the conventional Y-T format, where the vertical deflection represents the amplitude of the input signal (typically voltage) and the horizontal deflection represents time from an internal sweep generator. This mode provides a straightforward visualization of signal behavior over time, essential for basic waveform analysis in analog and digital oscilloscopes alike. For comparing multiple signals, dual-trace or multiple-trace modes enable the simultaneous display of waveforms from two or more input channels. In alternate mode, the oscilloscope switches between channels after each horizontal sweep, rapidly refreshing to create the illusion of simultaneous traces; this is suitable for higher-frequency signals where sweep times are short. In chopped mode, the beam alternates rapidly between channels (typically at a fixed rate like 500 kHz) during a single sweep, interlacing the traces for a segmented appearance; this works better for low-frequency signals to avoid flicker. Math functions extend these modes by allowing operations such as addition (e.g., channel A + channel B for differential measurements) or subtraction, displayed as a derived trace. X-Y mode decouples the display from the time base, using one channel (usually channel 1) for horizontal (X) deflection and another (channel 2) for vertical (Y) deflection, plotting one varying signal against another. This configuration generates Lissajous patterns, such as ellipses whose shape and orientation reveal phase differences between sinusoidal signals—for instance, a 90-degree phase shift appears as a circle. Frequency ratios between the two signals can also be determined from the pattern's complexity, like the number of loops indicating integer multiples. X-Y mode is particularly valuable for phase measurements in AC circuits or servo systems. Z-axis modulation introduces an additional input to control beam intensity, varying the brightness of the trace rather than position. In analog oscilloscopes, this directly modulates the cathode-ray tube's electron beam current; in digital models, it is emulated through pixel intensity adjustments. Applications include blanking the trace during retrace in X-Y mode to clean up the display, enhancing cursors or markers for precise measurements, or demodulating frequency-modulated (FM) signals by using the modulation envelope to intensity-modulate a carrier trace. In digital storage oscilloscopes, persistence mode retains multiple waveform acquisitions on the display, overlaying them with decaying or infinite intensity to build a composite image over time. This allows observation of transient events, noise distributions, or rare glitches that might not appear in single sweeps. A key application is generating eye diagrams for digital signal integrity analysis, where repeated bit transitions are superimposed to assess jitter, amplitude margins, and bit error rates in high-speed communications.

Controls and Interfaces

Front Panel Controls

The front panel of an oscilloscope features a set of physical controls dedicated to managing the display and operational state of the instrument, enabling users to adjust trace alignment, configure basic modes, and optimize visualization without delving into signal processing details. These controls, typically arranged around the display area, ensure precise waveform representation. In analog oscilloscopes, they directly influence the cathode-ray tube (CRT)'s deflection systems, while digital models incorporate knobs, buttons, and touchscreens for compatibility with modern workflows and enhanced usability. In analog oscilloscopes, beam controls regulate the electron beam's characteristics to produce a clear trace on the CRT screen. The focus knob sharpens the beam by adjusting the voltage on focusing electrodes, minimizing spot size for high-resolution details. Intensity controls the trace brightness by varying beam current, balancing visibility and phosphor longevity. Astigmatism adjustment fine-tunes beam circularity to reduce blurring. The beam finder button intensifies and expands the beam to locate off-screen traces. These features are absent in digital oscilloscopes, which use digital signal processing and flat-panel displays (e.g., LCD or OLED) with automatic focus and brightness adjustment via software. Modern digital front panels typically include multifunction rotary knobs for scaling and positioning, dedicated buttons for triggering and acquisition modes, and touchscreen interfaces for menu navigation and measurements. The graticule provides a standardized measurement scale, typically an 8-by-10 division grid where vertical divisions represent voltage and horizontal ones time intervals. In analog scopes, internal graticules are inscribed on the CRT's inner surface for parallax-free viewing, while external ones use a transparent sheet. Digital oscilloscopes display graticules on-screen, often with customizable overlays and illumination always active for readability in various lighting conditions. Illumination controls in analog models backlight graticule lines for low-light use. Position controls consist of vertical and horizontal knobs that shift the trace for optimal alignment. The vertical position knob centers the waveform relative to the graticule's centerline, while the horizontal one adjusts the timing reference. These are standard on both analog and digital oscilloscopes, integrating with sensitivity settings for accurate scaling. Power and mode switches manage operational states. The main power switch toggles AC connection, often with fuse and filter protection. Operate/standby modes allow low-power idle. Sweep mode selectors include continuous for repetitive signals and single-shot for transients, available in both analog (via triggered sweep) and digital (via acquisition modes) scopes. In analog oscilloscopes, the trace rotation control aligns the trace with graticule axes, compensating for CRT misalignments. This is unnecessary in digital models, where display orientation is software-controlled.

Input and Probe Interfaces

Oscilloscopes typically feature BNC (Bayonet Neill-Concelman) connectors as the primary input interfaces, which provide a secure, low-reflection connection for signal inputs. These connectors support both 50 Ω impedance for high-frequency applications, such as RF signals, and 1 MΩ impedance for general-purpose measurements, with the outer shell serving as the ground reference to minimize noise and ensure safe signal transmission. Channel selectors on oscilloscopes allow users to configure the input channels for single or dual-trace operation, commonly labeled as CH1 and CH2. In dual-trace mode, both channels can be displayed simultaneously using alternating (ALT) or chopped (CHOP) switching to interleave waveforms, while single-channel selection isolates CH1 or CH2 for focused measurements; coupling options, such as AC or DC, further condition the input signals to filter DC offsets or pass full-spectrum content. Attenuation compensation is essential for passive probes, particularly 10:1 attenuation probes, which divide the input voltage by 10 to extend the dynamic range and reduce loading effects on the circuit under test. These probes include an adjustable compensation capacitor that must be tuned—typically using the oscilloscope's built-in calibration signal, such as a 1 kHz square wave—to achieve a flat frequency response across the probe's bandwidth, preventing distortions like overshoot or undershoot in the displayed waveform. Overload protection mechanisms safeguard the oscilloscope's input circuitry from high-voltage transients or accidental overvoltages, often employing clamping diodes connected to the supply rails or chassis ground to limit peak voltages, alongside series resistors that dissipate excess energy. For 1 MΩ inputs, protection typically handles up to ±400 V peak, while 50 Ω inputs are limited to levels like 5 Vrms (approximately ±7 V peak for sinusoidal signals) or ±5 V peak depending on the model, to prevent damage to sensitive front-end amplifiers. Differential inputs, common in mixed-signal oscilloscopes, enable floating measurements by amplifying the voltage difference between two input signals without reference to ground, which is crucial for isolating signals in non-grounded circuits like power electronics or battery systems to avoid ground loops and common-mode noise. These inputs often pair with dedicated differential probes that provide high common-mode rejection ratios, allowing safe and accurate acquisition of signals up to several kilovolts.

Bandwidth and Sensitivity Settings

The bandwidth of an oscilloscope is defined as the frequency at which the amplitude response drops to -3 dB (approximately 70.7%) of its low-frequency value, representing the upper limit of accurate signal reproduction without significant attenuation. For instance, a 100 MHz oscilloscope maintains flat response up to around 100 MHz before the -3 dB roll-off begins, ensuring fidelity for signals within that range. This specification is crucial for selecting an instrument that matches the highest frequency components of the signal under test, as exceeding the bandwidth leads to distorted waveforms and underestimated peak amplitudes. Bandwidth also correlates directly with the oscilloscope's rise time, the duration for the output to transition from 10% to 90% of its final value in response to a step input. The approximate relationship is given by the formula t_r \approx \frac{0.35}{\text{BW}}, where t_r is the rise time in seconds and BW is the bandwidth in hertz; this holds for Gaussian-response systems common in many oscilloscopes. For a 100 MHz bandwidth, the corresponding rise time is about 3.5 ns, allowing precise capture of fast edges in digital signals. This relation guides users in verifying system performance, as the combined rise time of the oscilloscope and probe must be considered for overall measurement accuracy. Sensitivity settings determine the vertical scale for displaying signal amplitudes, with coarse adjustments providing discrete steps (e.g., 1 mV/div to 10 V/div in a 1-2-5 sequence) for general use and fine adjustments enabling continuous variation within those steps for precise scaling of low-level signals. These ranges accommodate weak signals down to 500 µV/div in high-sensitivity models, optimizing dynamic range while minimizing noise floor impact. Fine control is particularly useful for amplifying subtle phenomena, such as small voltage ripples, without overloading the input. Oscilloscopes often incorporate high-pass and low-pass filters to enhance signal fidelity by isolating specific frequency bands or reducing interference. Low-pass filters, such as a 20 MHz bandwidth limit, attenuate high-frequency noise while preserving the fundamental waveform, effectively acting as an analog filter to clean noisy traces. High-pass filters remove DC offsets and low-frequency components, aiding analysis of AC-coupled signals like audio or RF modulations. These settings allow targeted noise reduction without altering core vertical deflection processing. Probe compensation ensures accurate signal transmission by tuning the probe's RC compensation network to match the oscilloscope's input capacitance, preventing overshoot or undershoot in square waves. The procedure involves connecting the probe to the oscilloscope's built-in calibration output (typically a 1 kHz square wave), adjusting the probe's trimmer capacitor until the waveform edges are flat and sharp, and verifying across the probe's bandwidth. This low-frequency adjustment (around 1 kHz) aligns the probe's attenuation ratio, maintaining fidelity for high-frequency measurements. In multi-channel configurations, effective bandwidth may experience derating when multiple channels are active simultaneously, as shared sampling resources limit the per-channel sample rate, potentially constraining high-frequency capture to maintain Nyquist compliance. For example, a scope rated at 40 GS/s on one channel might derate to 10 GS/s across four channels, reducing the effective upper frequency limit for complex signals. This requires users to evaluate total channel loading to avoid bandwidth limitations in simultaneous acquisitions.

Types

Analog Oscilloscopes

Analog oscilloscopes represent the traditional form of oscilloscope technology, relying on direct analog signal processing to produce real-time visual representations of electrical waveforms. These instruments convert input voltages into electron beam deflections within a cathode ray tube (CRT), where the beam traces the signal's variations across a phosphor-coated screen without any intermediate digitization. The core components include a vertical amplifier to scale the input signal for vertical deflection, a horizontal sweep generator for time-based horizontal movement, and deflection plates or coils that steer the electron beam accordingly. This design enables immediate, continuous display of dynamic signals, making analog scopes particularly suited for observing live phenomena such as transient events in real time. A key variant of analog oscilloscopes is the dual-beam model, which employs two independent electron guns within a single CRT to generate separate beams for displaying two unrelated signals simultaneously. Unlike dual-trace scopes that alternate or chop between channels on a single beam, dual-beam designs allow each beam to operate with its own vertical and horizontal deflection systems, facilitating direct comparison of non-synchronized or independent waveforms, such as phase differences in AC signals or multiple simultaneous events. This capability is achieved by sharing the CRT's phosphor screen but maintaining distinct beam paths, often requiring specialized CRTs with dual post-accelerator structures for high performance. Examples include the Tektronix 556, introduced in 1966, which offered over 50 triggering and display modes in a compact form. To capture transient signals that occur too quickly for persistent observation, analog oscilloscopes incorporate storage mechanisms using specialized CRT phosphors or mesh-grid structures. Phosphor storage relies on the afterglow properties of materials like P31, where a high-energy write beam excites the phosphor to create a visible trace that persists for seconds to minutes, decaying exponentially to 1/e of its initial intensity; bistable variants use scattered phosphor particles to maintain a stable charge pattern without halftones, enabling screen splitting for waveform comparison. Mesh-grid storage, in contrast, employs a fine conductive mesh coated with a dielectric layer positioned near the phosphor screen; during writing, the beam charges selected areas positively, and a low-energy flood gun then selectively illuminates the phosphor by passing electrons through uncharged regions, retaining the pattern until erased. These methods allow temporary retention of single-shot events but do not permit permanent waveform saving or digital export. Despite their strengths in real-time visualization, analog oscilloscopes have notable limitations, including the inability to store waveforms for later analysis or measurement, which restricts their use for complex post-processing tasks. The CRT phosphor is susceptible to burn-in from prolonged static displays at high brightness, potentially causing permanent screen damage through uneven wear or charge migration. Bandwidth in high-end models reaches up to 1 GHz, as exemplified by the Tektronix 7104 from the 1980s, but this is constrained by analog amplifier and CRT deflection limits, beyond which signal distortion occurs. Maintenance challenges arise from aging vacuum tubes or solid-state components, with replacement parts becoming scarce. In contrast to digital storage oscilloscopes, which sample and retain data indefinitely, analog models prioritize immediate analog tracing but lack equivalent persistence for non-repetitive signals. Historically, the Tektronix 465, introduced in November 1972, exemplifies a landmark portable analog oscilloscope from the 1970s, featuring 100 MHz bandwidth, dual-trace capability, and solid-state construction except for the CRT, all within a compact, battery-optional design weighing about 11.3 kg (25 lb) for field use. Priced at $1,725 in 1973 (equivalent to roughly $12,600 in 2025 dollars), it became a standard in electronics labs for its reliability across environmental conditions, including options for EMI shielding and external DC operation.

Digital Storage Oscilloscopes

Digital storage oscilloscopes (DSOs) represent an evolution from analog oscilloscopes by digitizing input signals through an analog-to-digital converter (ADC) and storing the resulting data in digital memory for subsequent display and analysis. This digitization enables the capture of waveforms as discrete data points, allowing for flexible post-acquisition processing without the limitations of phosphor persistence in analog displays. Unlike analog scopes, which provide a direct real-time visualization limited by the CRT's response time, DSOs store multiple acquisitions, facilitating features like waveform averaging and mathematical operations on captured data. DSOs employ various acquisition modes to handle different signal types. In real-time sampling mode, the oscilloscope captures the entire waveform in a single acquisition sweep at a high sample rate, typically exceeding twice the signal's highest frequency to prevent aliasing, making it suitable for non-repetitive or transient events. For repetitive signals, equivalent-time sampling mode accumulates data over multiple sweeps, either randomly or sequentially, to achieve higher effective resolution and bandwidth beyond the real-time sample rate limitations; this mode is particularly effective for high-frequency periodic signals where full waveform capture in one pass is impractical. The memory depth, or record length, in a DSO determines the number of samples stored per acquisition, directly influencing the observable time span and resolution. For instance, a buffer of 1 million points at a 1 GS/s sample rate allows capturing a 1 ms record length while maintaining high temporal detail, enabling analysis of both long-duration events and fine waveform features without compromising sample density. Deeper memory enhances the scope's versatility for complex signals, such as those in communications or power electronics, by supporting extended observation windows. Processing capabilities in DSOs extend beyond basic display through built-in digital signal processing. The fast Fourier transform (FFT) function converts time-domain data to the frequency domain, revealing spectral content for applications like harmonic analysis in audio or RF signals. Additional math functions, such as integration to compute area under the curve or differentiation to derive rate of change, allow users to manipulate waveforms mathematically—e.g., subtracting reference signals to isolate noise—enhancing diagnostic precision without external hardware. Advanced triggering in DSOs leverages digital processing for precise event isolation. Digital pattern triggering detects specific bit sequences across multiple channels, useful for logic validation in embedded systems, while runt pulse triggering captures aberrant pulses that fail to reach full amplitude within a defined duration, aiding glitch detection in digital circuits. Serial bus decoding supports protocols like I²C and SPI by overlaying decoded data on the analog waveform, simplifying protocol compliance testing and error identification in microcontroller-based designs. Key advantages of DSOs include infinite persistence displays, where multiple acquisitions overlay on the screen with intensity grading to highlight anomalies over time, and automated measurements that compute parameters like peak-to-peak voltage or rise time via software algorithms, reducing manual effort and improving measurement repeatability. These features enable comprehensive signal analysis, such as statistical trending over thousands of cycles, far surpassing the capabilities of analog instruments for modern debugging tasks.

Specialized Variants

Specialized variants of oscilloscopes extend core functionality to address niche requirements in complex signal environments, field deployments, or cost-sensitive applications, often by integrating complementary tools or optimizing for portability and performance. These designs build on digital storage principles but incorporate hybrid capabilities for enhanced versatility in debugging, analysis, and measurement tasks. Mixed-signal oscilloscopes (MSOs) combine analog waveform display with integrated logic analyzer channels, enabling engineers to capture and correlate analog and digital signals from mixed-signal circuits in a single instrument. They typically feature 4 to 8 analog channels alongside 16 digital channels that interpret signals as binary logic states, facilitating threshold-based viewing and bus analysis without separate equipment. This integration supports serial protocol decoding, such as I2C or SPI, and time-synchronized measurements essential for embedded systems debugging. For instance, Tektronix MSO series models provide up to 2 GHz bandwidth, 10 GS/s sample rates, and advanced triggering for digital buses. Mixed-domain oscilloscopes (MDOs) merge time-domain oscilloscope functions with spectrum analysis, allowing simultaneous viewing of signals in both domains for efficient RF and mixed-signal troubleshooting. They incorporate an RF input alongside analog and digital channels, providing correlated time and frequency traces to identify interactions between baseband and RF components, such as in wireless designs. Key features include built-in spectrum analyzers with up to 6 GHz frequency range, arbitrary waveform generation, and protocol analysis in a unified interface. Tektronix MDO4000 series exemplifies this with integrated 6-in-1 capabilities, including logic analysis and power measurements, tailored for IoT and EMI testing. Handheld oscilloscopes offer compact, battery-powered portability for on-site diagnostics in industrial or field settings, where benchtop units are impractical. These devices feature rugged enclosures rated for harsh environments, such as CAT III/IV safety standards, with built-in displays and interfaces for standalone operation or USB connectivity to PCs. They support essential functions like automatic measurements, waveform storage, and FFT analysis, often combining oscilloscope and multimeter capabilities. Models like the Fluke 190-204/S provide 200 MHz bandwidth, 2.5 GS/s sampling, and up to 4 hours of battery life, while Rohde & Schwarz Scope Rider delivers lab-grade performance with 500 MHz bandwidth in a battery-operated format. PC-based oscilloscopes utilize external hardware modules connected via USB to a personal computer, leveraging the host's processing power, storage, and display for cost-effective, high-performance signal acquisition. The hardware includes signal conditioning and analog-to-digital converters (ADCs), with software handling visualization, analysis, and data export, often supporting deep memory buffers up to 2 GS. This setup enables low-cost alternatives for hobbyists and labs, with features like multi-channel inputs and protocol decoding, though limited by PC interface speeds compared to standalone units. Pico Technology's PicoScope series, for example, offers up to 500 MHz bandwidth and 5 GS/s sampling in USB-powered modules, suitable for general-purpose testing. High-speed oscilloscopes target ultra-fast signals in telecommunications and optical systems, offering bandwidths exceeding 10 GHz to capture transient behaviors in high-data-rate environments. They incorporate specialized inputs, such as optical modules for wavelengths from 1100 to 1650 nm, supporting multirate testing of 10 Gb/s and higher datacom/telecom standards. Advanced models emphasize low noise and high sample rates for precise eye diagram analysis and jitter measurements. Keysight's Infiniium UXR series achieves up to 110 GHz bandwidth with exceptional signal integrity, while Tektronix's TDS8000B provides 30 GHz optical bandwidth for component evaluation.

Applications and Uses

General Measurement Techniques

Oscilloscopes enable the analysis of electrical signals through standardized measurement techniques that quantify voltage, time, frequency, phase, and signal quality parameters. These methods rely on the instrument's display graticule, on-screen cursors, or automated functions to provide accurate readouts, often requiring proper trigger setup to stabilize the waveform for reliable assessment. Voltage measurements form a foundational aspect of oscilloscope use, capturing the amplitude characteristics of signals. Peak-to-peak voltage is determined by measuring the vertical distance from the waveform's maximum positive peak to its minimum negative peak, typically using horizontal cursors positioned at these extrema or via automatic peak detection functions on digital oscilloscopes. RMS (root mean square) voltage, which represents the effective value of an AC signal, is calculated automatically by modern instruments through integration over multiple waveform cycles, ensuring precision for power-related analyses. DC offset, the average voltage level of the signal, is measured using vertical cursors or auto-functions with DC coupling enabled to include the constant component, allowing differentiation from pure AC content. Time and frequency measurements assess the temporal behavior of signals, essential for characterizing periodic phenomena. The period, defined as the time duration of one complete waveform cycle, is quantified by placing vertical cursors at corresponding points (e.g., from one rising edge to the next) or through automated period measurement, with frequency then derived as its reciprocal (1/period in hertz). Rise time, the interval for the signal to transition from 10% to 90% of its amplitude, and fall time, from 90% to 10%, are evaluated using cursors aligned to these threshold levels or auto-functions, particularly important for evaluating signal speed in digital circuits. Duty cycle, the percentage of the period during which the signal is high (or low), is computed by measuring the pulse width relative to the period via cursors or automated tools, providing insight into waveform symmetry. Phase difference between two signals is measured to evaluate relative timing, often using dual-channel setups. In XY display mode, the oscilloscope plots one channel's voltage against the other's, forming Lissajous patterns where the phase angle is determined by the pattern's geometry—such as a circular figure indicating 90° shift—quantified via cursors or graticule comparison. Alternatively, in time-domain view, phase is calculated from the time delay between corresponding points on overlaid waveforms divided by the period, multiplied by 360°. Signal integrity assessments identify distortions that affect performance, using visual inspection and quantitative tools. Jitter, the variation in signal edge timing, is analyzed through persistence or eye diagram modes on digital oscilloscopes, where multiple acquisitions overlay to reveal timing deviations, measurable with automated jitter analysis functions. Overshoot, the excessive voltage excursion beyond the nominal level following a transition, is detected by cursor placement on waveform peaks and quantified as a percentage of the step amplitude, helping diagnose reflections or bandwidth limitations. A representative example involves probing a sine wave to verify its amplitude and frequency. With the signal connected to channel one and triggering enabled, the user adjusts vertical scaling to center the waveform (e.g., 1 V/div for a 2 V peak-to-peak signal) and horizontal scaling for several cycles (e.g., 1 ms/div for a 1 kHz tone). Cursors or auto-functions then yield peak-to-peak voltage (2 V), period (1 ms), and frequency (1 kHz), confirming the signal's specifications.

Industry-Specific Applications

In electronics design, oscilloscopes are essential for debugging circuits by visualizing voltage waveforms at various points, enabling engineers to identify anomalies such as noise, glitches, or timing errors in digital and analog systems. For instance, they facilitate the verification of operational amplifier (op-amp) responses by measuring frequency response, slew rate, and distortion under load conditions, ensuring compliance with design specifications. In the automotive industry, oscilloscopes support CAN bus analysis through protocol decoding and triggering on specific identifiers or error frames, helping diagnose issues like signal integrity degradation, crosstalk, or termination faults in vehicle networks. They also enable precise measurement of ignition waveforms, including dwell angle—the duration the ignition coil is energized—by capturing primary and secondary voltage traces to assess timing accuracy and detect mechanical wear in distributor systems. Telecommunications engineers employ oscilloscopes to generate and analyze eye diagrams, which overlay multiple bit transitions to evaluate signal quality in high-speed serial links such as optical or electrical interfaces. These diagrams quantify jitter and noise impacts at decision points, correlating eye opening dimensions to bit error rates (BER), often targeting levels below 10^{-12} to ensure reliable data transmission without excessive retransmissions. In power electronics, oscilloscopes measure switching noise in inverters by probing voltage and current across power semiconductors, revealing dv/dt and di/dt rates that indicate electromagnetic interference or insulation stress in motor drives. They also assess harmonic distortion through FFT-based analysis of output waveforms, calculating total harmonic distortion (THD) to verify compliance with standards like IEEE 519 and mitigate effects on grid stability. Although dedicated electrocardiogram (ECG) monitors are primary for clinical use, oscilloscopes serve as secondary tools in medical device development for capturing and analyzing low-amplitude ECG signals, often below 1 mV, to validate amplifier performance and noise rejection in prototypes. High waveform update rates, such as 1 million per second, aid in resolving subtle QRS complexes and artifacts during bench testing.

Integration with Software and Systems

Modern oscilloscopes increasingly integrate with computers and networks through USB and Ethernet interfaces, enabling remote control, data acquisition, and export functionalities. These interfaces support the Standard Commands for Programmable Instruments (SCPI) protocol, a standardized set of ASCII-based commands that allow programmatic interaction across vendors. For instance, USB typically operates in USB-TMC (Test and Measurement Class) mode for direct device communication, while Ethernet provides LAN connectivity for higher-speed data transfer and multi-instrument setups. This integration facilitates exporting waveform data in formats like CSV or binary for offline analysis, reducing the need for manual screen captures. Dedicated software packages enhance oscilloscope capabilities by providing advanced analysis tools directly on connected PCs. Vendor-specific solutions, such as Tektronix's TekScope, allow users to load and analyze captured waveforms, perform measurements, and decode protocols like I2C, SPI, and USB without the instrument's hardware limitations. Open-source alternatives like Sigrok offer flexible protocol decoding for a wide range of serial buses, supporting hardware from multiple manufacturers and enabling custom scripting for decoding tasks. These packages often include graphical interfaces for visualizing decoded data, making complex signal analysis more accessible. For automated testing environments, oscilloscopes integrate seamlessly with platforms like National Instruments' LabVIEW and Python scripting libraries. LabVIEW uses instrument drivers to automate measurement sequences, such as voltage sweeps or timing verifications, streamlining production testing workflows. Python, via libraries like PyVISA, enables scripted control over SCPI commands for custom automation, including data logging and error checking in real-time applications. This approach is particularly valuable in research and development, where repeatable measurements accelerate validation processes. Post-2020 advancements have introduced cloud-based solutions for remote oscilloscope access and collaborative analysis. Tektronix's TekDrive platform provides a cloud workspace for uploading scope data, enabling team-based debugging and sharing of waveforms across locations. Similarly, Keysight's Infiniium EXR-Series supports hosted remote control via web interfaces, allowing collaborative sessions for signal integrity checks in distributed teams. These systems enhance productivity by integrating oscilloscope data with cloud storage and collaboration tools. Firmware updates for oscilloscopes are delivered via USB or Ethernet connections, incorporating enhancements for emerging protocols such as 5G NR decoding. Manufacturers like Tektronix release periodic updates that add compliance testing features for 5G signals, ensuring instruments remain compatible with evolving standards without hardware changes. By 2025, these updates have enabled real-time decoding of 5G physical layer signals, supporting applications in telecommunications R&D.

Accessories and Enhancements

Probes and Accessories

Passive probes are the most common type used with oscilloscopes for general voltage measurements, featuring selectable attenuation ratios such as 1x and 10x to accommodate different signal amplitudes and bandwidth requirements. The 1x setting provides unity gain for low-voltage signals but limits bandwidth typically to under 35 MHz due to higher loading effects, while the 10x setting reduces signal loading by increasing input impedance and extends bandwidth up to several hundred MHz, making it suitable for most applications. These probes include an adjustable compensation capacitor, typically with a range of 10 to 35 pF, to match the oscilloscope's input capacitance and ensure accurate waveform reproduction by compensating for high-frequency attenuation. High-voltage variants of passive probes are designed for signals exceeding 1000 V, often rated up to 1500 V peak or more, with reinforced insulation to prevent hazards in power electronics testing. Active probes incorporate amplification circuitry at the probe tip to minimize signal distortion, particularly for high-impedance or high-frequency measurements, with FET-input designs providing input capacitances as low as 1 pF and resistances around 1 MΩ to reduce loading on sensitive circuits. These probes are powered by the oscilloscope or an external supply and offer bandwidths from hundreds of MHz to several GHz, enabling precise capture of fast edges in digital and RF signals. Current probes, a specialized active probe category, clamp around conductors to measure AC and DC currents without circuit interruption, utilizing Hall effect sensors for DC detection and transformers for AC components, with sensitivities down to milliamperes and bandwidths up to 100 MHz or more for power analysis. Accessories extend oscilloscope functionality beyond basic voltage probing; logic probes connect multiple digital channels to mixed-signal oscilloscopes, capturing up to 16 or more signals with threshold adjustments for protocols like I2C or SPI, while demodulator probes extract baseband information from modulated carriers, such as AM or FM signals in RF testing, by filtering and detecting envelope variations. Proper grounding is essential to minimize noise pickup in probe measurements; spring clips provide a short, low-inductance ground path directly at the measurement point, reducing loop area and inductive noise compared to longer alligator clip leads that can form ground loops and introduce artifacts like ringing or offset errors. When selecting probes, bandwidth matching is critical: the probe's bandwidth should equal or exceed the oscilloscope's to achieve the full system performance, often following a guideline where the overall bandwidth approximates the minimum of the two but can be enhanced by ensuring the probe rating is at least three to five times the highest signal frequency of interest.

Calibration and Maintenance

Calibration of oscilloscopes ensures the accuracy of voltage, time, and frequency measurements by verifying and adjusting key parameters such as gain, offset, and frequency response. Standard procedures begin with warming up the instrument for at least 20 minutes to achieve thermal stability, followed by connecting a probe to the built-in calibration terminal, which typically outputs a 1 kHz square wave signal. This reference signal is used to adjust probe compensation for a flat-top response, confirming square edges without overshoot or undershoot, and to calibrate vertical sensitivity by measuring the waveform's amplitude against known values. Further steps involve applying precise voltage references from external calibrators to check linearity and bandwidth, ensuring the oscilloscope's response remains flat across its specified range. To maintain traceability, these calibrations must link directly to national metrology institutes like the National Institute of Standards and Technology (NIST) through an unbroken chain of comparisons using primary or secondary standards, such as voltage references and metrology-grade instruments, under controlled environmental conditions. This traceability, guided by NIST Special Publication 250 and ISO/IEC 17025, documents measurement uncertainties and supports compliance in regulated fields like electronics testing and medical device validation. Manufacturers recommend performing full calibrations at intervals of approximately 12 months, depending on usage intensity and environmental factors. In digital storage oscilloscopes, self-calibration routines provide an internal method to optimize the signal path for accuracy, particularly by correcting for analog-to-digital (A/D) converter linearity errors through automated adjustments to offset, gain, and sampling paths. These routines, accessible via the instrument's menu, use built-in reference signals to compensate for internal drifts without external equipment, though they should be followed by external verification for critical applications. Routine maintenance addresses common issues that can degrade performance, such as vertical drift caused by temperature fluctuations or component aging, which manifests as baseline shifts in displayed waveforms, and trigger instability from noise interference or faulty trigger circuits, leading to jittery or absent traces. To mitigate these, users should perform periodic visual inspections and run diagnostic self-tests; cleaning dust from cooling fans and vents with compressed air prevents thermal buildup and related drifts, while avoiding liquid cleaners on internal components. For legacy analog oscilloscopes, gentle cleaning of the phosphor screen with approved solutions may be required to restore brightness, but digital models primarily need external connector maintenance using isopropyl alcohol and lint-free cloths. Safety during calibration and maintenance is paramount, given the potential for high voltages in test signals. Operators must observe warnings against defeating ground connections or measuring ungrounded circuits exceeding 30 V RMS without proper isolation, as oscilloscope inputs are typically rated for 300-600 V RMS to ground to prevent shock hazards. Probes used in these procedures should match the voltage ratings of the signals being measured, with common examples supporting up to 1,000 V RMS tip-to-common, and the oscilloscope chassis grounded at all times to avoid floating potentials. Laboratories performing oscilloscope calibrations adhere to ISO/IEC 17025:2017, the international standard for testing and calibration competence, which mandates validated methods, traceable equipment, and evaluation of measurement uncertainties to ensure reliable results. For high-precision laboratory scopes, annual recertification is standard to maintain errors below 1% in key parameters like gain and time base, supporting applications in research and quality control where accuracy directly impacts outcomes. Bandwidth verification, often integrated into these processes, uses fast step or pulse signals to confirm the -3 dB point aligns with specifications.

References

  1. [1]
    [PDF] Oscilloscope-Basics.pdf - Instrumentation LAB
    Oscilloscope Basics. Primer. Abstract: The oscilloscope is arguably one of the most useful general purpose tools ever created for use by electronic engineers. ...Missing: principles | Show results with:principles
  2. [2]
    [PDF] Electrical performance tests for storage oscilloscopes
    2. Oscilloscope Operating Principles. 2.1 Conventional (analog) oscilloscopes ... defined as the input value which causes 50% of the output codes to be ...
  3. [3]
    [PDF] Basic Operation of an Oscilloscope The Screen - NJIT
    An oscilloscope displays a voltage waveform versus time and has the following components: 1) a screen to display a waveform,. 2) input jacks for connecting the ...
  4. [4]
    What is an oscilloscope? - Tektronix
    May 11, 2021 · Engineers use an oscilloscope to measure electrical phenomena and solve measurement challenges quickly and accurately to verify their designs.
  5. [5]
    Understanding basic oscilloscope operation | Rohde & Schwarz
    What is an oscilloscope and how is it used? The primary purpose of an oscilloscope is to measure and display voltage versus time. They are widely used for ...Basic Oscilloscope Operation... · Waveform Display · Trigger System And Trigger...<|control11|><|separator|>
  6. [6]
    Oscilloscope – 1897 - Magnet Academy - National MagLab
    A mechanic may use an oscilloscope to measure engine function, while a medical researcher uses it to monitor heart activity.
  7. [7]
    Calculating phase difference with an oscilloscope - EDN Network
    Aug 13, 2018 · Phase difference can be measured on an oscilloscope by finding the time delay between two waveforms and their period.
  8. [8]
  9. [9]
    What is an Oscilloscope: An In-Depth Look - Used Keysight Equipment
    Think of it as a graphing tool that plots electrical voltage on the vertical axis against time on the horizontal axis. This visualization allows you to see the ...
  10. [10]
  11. [11]
    Deflection Plate - an overview | ScienceDirect Topics
    Deflection plates (DP) are components in an oscilloscope that create vertical movement of the electron beam by applying a voltage, causing the beam to be ...<|control11|><|separator|>
  12. [12]
    How Does an Oscilloscope Work: Detailed Insights for Engineers
    Amplification: A pre-amplifier boosts the incoming signal. The vertical controls adjust the level of amplification. Attenuation: Optionally, you can attenuate ...
  13. [13]
    Vertical & Horizontal Amplifiers in Cathode ray oscilloscope(CRO)
    Nov 18, 2016 · The gain of the vertical amplifier determines the smallest signal that the oscilloscope can satisfactorily reproduce on the CRT screen. The ...<|control11|><|separator|>
  14. [14]
    What is Vertical Sensitivity in Oscilloscopes? - Keysight
    This is generally expressed in volts per division (V/div), indicating how many volts correspond to each horizontal line on the oscilloscope's grid. The more ...
  15. [15]
    Oscilloscope Basics: Waveforms, Graph, & Measurement Reading
    What is an oscilloscope? An oscilloscope (informally scope, oscope, or O-scope) is a diagnostic instrument that draws a graph of an electrical signal.
  16. [16]
    What is a Time Base Generator? - Definition & Circuit Diagram
    The velocity operates with the deflecting voltage increases linearly with the time. The voltage having such characteristic is called ramp voltage.
  17. [17]
    [PDF] Oscilloscope - NYU Physics department
    Every scope has a signal generator or time base oscillator. This oscillator produces a voltage that is a linear ramp as a function of time. Fig. 2 shows the ...<|control11|><|separator|>
  18. [18]
    Evaluating Oscilloscope Bandwidth, Sample Rate, and ... - Tektronix
    Rise time describes the useful frequency range of an oscilloscope. Rise time measurements are critical in the digital world. Rise time may be a more appropriate ...
  19. [19]
  20. [20]
    [PDF] Oscilloscope Sample Rates versus Sampling Fidelity
    To satisfy Nyquist criteria, you need a scope that samples at least three to five times higher than the scope's bandwidth specification, depending on the scope ...
  21. [21]
    What is Input Voltage in Oscilloscopes? - Keysight
    Most oscilloscopes have an input impedance of 1 MΩ. Matching source impedance with oscilloscope input impedance ensures accurate measurements. Keysight ...
  22. [22]
    An Introduction to Oscilloscope Probes - Technical Articles
    Jun 29, 2017 · Most modern scopes allow you to choose either 50 Ω or 1 MΩ input resistance. For general-purpose testing, a 1 MΩ input is commonly used. A 50 Ω ...
  23. [23]
    50-Ω and 1-MΩ oscilloscope terminations - EDN Network
    Jun 26, 2025 · The 1-MΩ termination provides a high input impedance, which reduces circuit loading. It offers the highest voltage and offset ranges, but its ...Input Termination · 50-Ω Vs 1-MΩ Inputs · Dc Or Ac Input Coupling
  24. [24]
    [PDF] Tektronix: Primer > XYZs of Oscilloscopes
    Introduction to Logic Analysis: A Hardware Debug Tutorial. The XYZs of Logic Analyzers. UMTS Protocols and Protocol Testing. GPRS Protocol Testing in the ...Missing: explanation | Show results with:explanation
  25. [25]
    What is Vertical Resolution in Oscilloscopes? - Keysight
    16-bit ADC: At the top-tier, with 216 or 65536 vertical levels, the 16-bit ADC is for those scenarios where every minute detail counts. It provides unparalleled ...
  26. [26]
    Oscilloscope Systems and Controls: Functions & Triggering Explained
    The three main systems are vertical (signal amplification), horizontal (time base), and trigger (stabilizing repeating signals or triggering on single events).
  27. [27]
    [PDF] Oscilloscopes
    (12 kV) at the screen and zero at the other end. Thus, the electrons ... CRT Oscilloscope Block Diagram. 24. Page 25. A basic Cathode Ray Oscilloscope ...
  28. [28]
    What is oscilloscope coupling - Keysight
    Oscilloscope coupling is the process of connecting an oscilloscope to a signal source to measure the signal's waveform. There are three main types – AC, DC, ...Selected Deals · What Are Waveform Intervals... · Keysight OscilloscopesMissing: system | Show results with:system
  29. [29]
    [PDF] Iwatsu SS-5702 Oscilloscope
    The SS-5702 is designed for dual-trace measurement respec- tively. The vertical deflection factor is from 5 mV/div to 10 V/div in a 1-2-5 sequence.
  30. [30]
    [PDF] PORTABLE OSCILLOSCOPE
    SCOPE BW Button-Reduces the bandwidth of the vertical deflec- tion system and trigger system to 20 MHz when the button is lit. The full vertical deflection ...
  31. [31]
    Time-Base Generator : Circuit Diagram with Working Principle
    Such a voltage is called Ramp voltage. When this voltage rapidly tends to ... The oscilloscope uses several time-base generators to display various Time-base ...
  32. [32]
    Trigger Tips: The Holdoff Control | Tektronix
    Mar 20, 2013 · Trigger Holdoff adds a delay to re-arming the trigger circuit, controlling how often the oscilloscope can be triggered, and how long to ignore ...Missing: modes | Show results with:modes
  33. [33]
    How does an oscilloscope work? - Tektronix
    The trigger system looks for voltage thresholds, pulse widths, logic combinations (on multiple inputs), and many other conditions to qualify an acquisition.
  34. [34]
    None
    ### Oscilloscope Triggering Summary
  35. [35]
    [PDF] 6.007 Oscilloscope Tutorial - MIT OpenCourseWare
    3 Triggering. Now that our waveform is centered, we need to make sure the oscilloscope can lock to the signal. Figure 7: The “LEVEL” knob sets the trigger level ...
  36. [36]
  37. [37]
    [PDF] Agilent InfiniiVision 5000/6000/7000 Series Oscilloscopes User's ...
    Connect the oscilloscope to the device under test using passive or active probes or 50-ohm BNC cables. 4 Triggering. Trigger modes, trigger settings, and ...
  38. [38]
    [PDF] Agilent 54621A/22A/24A Oscilloscopes and Agilent 54621D/22D ...
    The oscilloscopes display analog and digital signals, have digital channel controls, and offer waveform math, measurement, and autoscale features.
  39. [39]
    [PDF] Agilent 6000 Series Oscilloscopes User's Guide
    Z-Axis Input in XY Display Mode (Blanking). When you select the XY display ... Z-axis turns the trace on and off (analog oscilloscopes called this Z ...
  40. [40]
    [PDF] Agilent 54621A/22A/24A/41A/42A Oscilloscopes ... - UNL EE Shop
    XY Display Mode Operating Hint (Z-axis input on 54620-series only). When you select the XY display mode, the time base is turned off. Channel 1 is the. X-axis ...<|control11|><|separator|>
  41. [41]
    [PDF] Programmer Manual TDS3000, TDS3000B, and TDS3000C Series ...
    ... Z channel source. DISPLAY:XY:ZCHANNEL CH2 sets oscilloscope channel 2 as the Z channel source in gated XYZ mode. DISplay:XY:ZLEVel. Specifies the Z channel ...
  42. [42]
  43. [43]
    Lab 3: PRBS generation, Noisy Channel Model and Eye Diagrams
    The persistence of the screen of a general purpose oscilloscope would be too short to show a steady display, so it may blink. Alternately, the oscilloscope ...
  44. [44]
    Oscilloscope Controls - HyperPhysics
    Rotate this button until the trace is sharp. Beam finder. If you do not find a trace, push this button. The screen will display what quadrant the trace is in.
  45. [45]
  46. [46]
    [PDF] GOS-6xxG Family Dual Trace Oscilloscope - GW Instek
    Each of these oscilloscopes employs a 6-inch rectangular type cathode-ray tube with red internal graticule. 623G, 653G and 658G each has a sweep ...
  47. [47]
    ABCs of Probes Primer - Tektronix
    Learn about the fundamentals of probes and discover how to choose the right probe for your oscilloscope and get reliable measurements every time.
  48. [48]
    How to tune (compensate) x10 oscilloscope probes - Pico Technology
    Jul 23, 2025 · The easiest way to compensate the LFC part of a probe is to input a square wave with a relatively slow edge but, importantly, no overshoot.
  49. [49]
    A Guide to Protecting Oscilloscope Inputs from Over-Voltage - DigiKey
    Apr 4, 2022 · A 1 MΩ input can be protected using surge protectors such as spark gaps, gas discharge tubes, or diodes that break down to discharge electrical surges to ...Missing: overload | Show results with:overload
  50. [50]
    Test Happens - Teledyne LeCroy Blog: Back to Basics: Probes (Part II)
    Nov 20, 2013 · Low-impedance (Low-Z) passive probes generally provide a 10:1 attenuation factor into the oscilloscope's 50-Ω input termination. Where the high ...<|control11|><|separator|>
  51. [51]
    6 Series B MSO - Tektronix
    For example, an oscilloscope channel operates as a low pass filter where its 3 dB down point is referred to as its bandwidth. Given a waveform of any shape, a ...
  52. [52]
    How is bandwidth related to rise time for oscilloscopes? - Tektronix
    Historically, oscilloscope frequency response tended to approximately follow the rule: Bandwidth x risetime = 0.35. This corresponds to a 1- or 2-pole ...Missing: t_r ≈
  53. [53]
    2 Series MSO Specification and Performance Verification Manual
    Sensitivity range (coarse), 1 MΩ: 1 mV/Div to 10 V/Div in a 1-2-5 sequence; Sensitivity range (fine): Continuous adjustment from 1 mV/DIV to 10 V/DIV, 1 MΩ 1x ...
  54. [54]
    How to reduce signal noise on a basic oscilloscope - Tektronix
    Jun 9, 2016 · Using the Bandwidth Limit Filter is one of the simplest ways to reduce noise in your signal. Oscilloscopes typically offer a limited set of ...
  55. [55]
    User Defined Filter Tool | Tektronix
    Low-Pass filters (LPF) are used to remove background and high-frequency noise. High-Pass filters (HPF) can be used to remove DC and low-frequency components.Missing: reduction | Show results with:reduction
  56. [56]
    [PDF] Keysight Technologies Eight Hints for Better Scope Probing
    Probe compensation is the process of adjusting the RC divider so the probe maintains its attenuation ratio over the probe's rated bandwidth. If your scope can ...
  57. [57]
    Digital Phosphor Oscilloscopes - Tektronix
    3.5 GHz, 2.5 GHz, 1 GHz, and 500 MHz bandwidth models · Up to 40 GS/s real-time sample rate on one channel, up to 20 GS/s on two channels, and up to 10 GS/s on ...
  58. [58]
    What is an Analog Oscilloscope? - Tektronix
    Dec 6, 2023 · An analog oscilloscope graphically represents electrical signals, converting voltage into a visual format displayed on a screen.
  59. [59]
    Analogue Oscilloscope: Cathode Ray Scope - Electronics Notes
    Looking at the operation of the analogue scope in more detail, it uses the cathode ray tube to display signals in both X (horizontal) and Y (vertical) axes.
  60. [60]
    Dual Beam Oscilloscopes - vintageTEK
    Dual trace scopes are a single beam / single gun design which either alternate or chop the vertical signal. Alternating draws one trace then the other.Missing: analog | Show results with:analog
  61. [61]
  62. [62]
    The Storage Oscilloscope: Storage Principles – Blog - Apex Waves
    Sep 12, 2018 · Positively charged areas of the storage target allow electrons from the flood guns to pass through the mesh towards the standard phosphor, ...
  63. [63]
  64. [64]
    Tektronix 7104 – 1 GHz analog oscilloscope – physik.co-i60.com
    Apr 14, 2023 · The Microchannel Plate CRT display is still in a good shape – no burn-ins except the typical wear-out areas (horizontal trace, annotation areas) ...<|control11|><|separator|>
  65. [65]
  66. [66]
    465 - TekWiki
    The Tektronix 465 is a portable 100 MHz dual-trace oscilloscope with dual time-bases, introduced in November 1972. It is all solid-state except for the CRT.Missing: 1970s | Show results with:1970s
  67. [67]
    Real-Time Versus Equivalent-Time Sampling | Tektronix
    When a DSO uses equivalent-time sampling, it can acquire any signal up to the analog bandwidth of the scope regardless of the sample rate. In this mode, the ...
  68. [68]
    Real-Time vs. Sampling Oscilloscopes: What Are the Differences?
    May 11, 2022 · Real-time scopes capture signals in real-time, while sampling scopes use multiple passes. Real-time scopes are better for non-repeating signals ...
  69. [69]
    Digital Storage Oscilloscope - Tektronix
    Record length is selectable, from 2000 samples up to 20 million samples for capturing long time periods. The exceptionally long record length will help you find ...
  70. [70]
    [PDF] Digital Storage Oscilloscope - TBS2000B Series Datasheet - Tektronix
    The TBS2000B Series of Oscilloscopes with a 9-inch WVGA display,. 5 million point record length and 2 GS/s sample rate, capture and.
  71. [71]
    Digital Storage Oscilloscope - Tektronix
    You can understand the frequency content of your signals with the FFT function by pressing the dedicated front-panel FFT button. Display only the FFT, or turn ...
  72. [72]
    [PDF] Digital Storage Oscilloscopes - Tektronix
    Their Fast Fourier Transform (FFT) math function allows the user to analyze, characterize, and troubleshoot circuits by viewing frequency and signal ...
  73. [73]
    How to Choose the Best Oscilloscope to Buy - Tektronix
    May 11, 2021 · Digital storage oscilloscopes (DSO) employ a serial processing architecture to capture from 10 to 5,000 wfms/s. Some DSOs provide a special mode ...
  74. [74]
    How Does an Oscilloscope Work: Detailed Insights for Engineers
    An oscilloscope comprises various components, each serving a unique function to deliver that intricate view of electrical signals.
  75. [75]
    Digital Storage Oscilloscopes - Tektronix
    Digital storage technology supports features not available on most low-cost oscilloscopes, including automatic measurements, peak detect, storage of two ...
  76. [76]
    What is a Mixed-Signal Oscilloscope? - Keysight
    A mixed-signal oscilloscope (MSO) provides insight into both analog and digital aspects of a circuit, combining DSO and logic analyzer functionality.
  77. [77]
    Mixed Signal Oscilloscopes Compared - Tektronix
    A mixed signal oscilloscope (MSO) is a type of digital storage oscilloscope designed to display and compare both analog signals and digital signals.Missing: definition | Show results with:definition
  78. [78]
    Why Would you Want a Mixed Signal Oscilloscope?
    The mixed signal oscilloscope (MSO). An MSO adds up to 16 digital channels to your 4 analog channel oscilloscope. This greatly expands the types of analysis ...Missing: definition | Show results with:definition
  79. [79]
    How to Use a Mixed Signal Oscilloscope to Test Digital Circuits
    A mixed signal oscilloscope's digital channels view a digital signal as either a logic high or logic low, just like a digital circuit views the signal. This ...
  80. [80]
    Mixed Signal Oscilloscopes | Tektronix
    Tektronix MSO/DPO5000B Series oscilloscopes provide exceptional signal fidelity, with 2 GHz bandwidth and 10 GS/s sample rate, along with advanced analysis and ...
  81. [81]
    Fundamentals of the MDO4000 Series Mixed Domain Oscilloscope
    A mixed signal oscilloscope is an oscilloscope with additional logic or digital channels that are optimized for digital signals and serial bus decode / ...
  82. [82]
    Mixed Domain Oscilloscope - Tektronix
    Mixed Domain Oscilloscopes let you see your signals from an analog, digital and RF perspective with one, highly-integrated instrument.
  83. [83]
    Oscilloscope Types - Tektronix
    Learn about the differences between digital storage oscilloscopes, digital phosphor oscilloscopes, mixed signal oscilloscopes and mixed domain oscilloscopes ...Missing: grid | Show results with:grid
  84. [84]
    Mixed Domain Oscilloscopes - Tektronix
    It is the ultimate 6-in-1 integrated oscilloscope that includes an integrated spectrum analyzer, arbitrary function generator, logic analyzer, protocol ...
  85. [85]
  86. [86]
    Handheld Oscilloscopes - Tektronix
    With 21 automated measurements, you can make common measurements quickly and accurately. ... persistence, Infinite persistence. Display format: YT and XY ...Ths3000 Series Datasheet · Replacement Products · Fast Analysis Of Your Device
  87. [87]
  88. [88]
    R&S®Scope Rider handheld oscilloscope | Application
    The R&S®Scope Rider is a modern versatile handheld oscilloscope featuring lab oscilloscope performance in a portable format with the ruggedness of a battery ...<|separator|>
  89. [89]
    Why your next scope should be a PC-based USB oscilloscope
    Jan 22, 2025 · Compact, cost-effective , and high-performing, they seamlessly combine the reliability of traditional scopes with the power and versatility of ...Decode Without Limits · Harness the Power of Your PC · Your Interface, Your Way
  90. [90]
    USB Oscilloscope - PC Based Scope - Electronics Notes
    PC & USB oscilloscopes provide high performance test instruments in a small space and for a comparatively low cost by using the user interface and some ...Pc Usb Scope Basics · 2) Fpga Based Usb Scope · Pc Usb Scope Advantages /...
  91. [91]
    Oscilloscope - an overview | ScienceDirect Topics
    Definition. The oscilloscope is an instrument that displays waveforms, allowing measurements of wave time and amplitude to be taken. Older oscilloscopes are ...
  92. [92]
    Oscilloscope | Pico Technology
    The highest performance PC-based USB-powered oscilloscopes and MSOs available. Up to 500 MHz bandwidth, 5 GS/s sampling and 2 GS memory. All models have a ...Missing: sound card external
  93. [93]
    TDS8000B - Digital Sampling Oscilloscope - Tektronix
    With its high optical bandwidth of up to 30 GHz (typical) it is well-suited for general purpose high-performance 10 Gb/s optical component testing. The ...
  94. [94]
    Infiniium UXR Series Oscilloscopes | Superior Signal Integrity
    Keysight Infiniium UXR Series Oscilloscopes offer 110 GHz bandwidth, industry-leading low noise, and exceptional signal integrity, ensuring unmatched ...
  95. [95]
    80C00 Optical Modules for DSA8300 Sampling Oscilloscope ...
    The 80C11B module is a long wavelength (1100 to 1650 nm) multirate optical sampling module optimized for testing 10 Gb/s datacom and telecom standard rates at ...
  96. [96]
    8 Series Sampling Oscilloscope (Discontinued) - Tektronix
    Use TSOVu with the TSO820 Sampling Oscilloscope Mainframe to acquire multiple channels simultaneously and analyze remotely over Ethernet or Wi-fi. Wi-fi ...
  97. [97]
    [PDF] XYZs of Oscilloscopes –– - Tektronix
    Oscilloscopes have systems and controls, measure waveforms, and come in types like digital storage, digital phosphor, mixed domain, mixed signal, and digital ...
  98. [98]
    [PDF] Measuring with oscilloscopes - Educational note
    This educational note covers the theory and practice oscilloscopes based on concrete measurement examples that address authentic everyday tasks.
  99. [99]
    How to Measure Frequency with an Oscilloscope - Tektronix
    Apr 6, 2021 · Oscilloscopes measure the voltage of an electrical signal, which is then graphically represented by waveforms. However, an oscilloscope can ...Missing: general | Show results with:general
  100. [100]
    CAN Bus Troubleshooting and Decoding with Oscilloscopes
    Discover how Tektronix CAN bus decoders simplify CAN bus troubleshooting with advanced decoding and triggering features, ensuring reliable system ...Missing: dwell angle
  101. [101]
    Dwell angle: Understanding ignition events with PicoScope (MIC#7)
    Sep 8, 2025 · Dive deep into automotive ignition theory with PicoScope. This guide explains how to use math channels to accurately calculate dwell angle, ...
  102. [102]
    Anatomy of an Eye Diagram - How to Construct and Trigger - Tektronix
    For sampling oscilloscopes, this can be 105 samples per second at a 10 Gb/s (1010 bits/second) rate. This means that most eye diagrams are composed of shallow ...
  103. [103]
    Bridging the Gap Between BER and Eye Diagrams - Tektronix
    BER and Eye Diagrams: Why Neither is a Complete Depiction. Most communication links are ultimately judged on their Bit Error Rate (BER) per-formance – how many ...Missing: telecommunications | Show results with:telecommunications
  104. [104]
    Making Measurements on 3-Phase Motor Drives with an Oscilloscope
    This primer describes methods for making measurements using inverter, motor and drive analysis software on oscilloscopes to provide stable, accurate electrical ...
  105. [105]
    Power Supply Measurement and Analysis with 4/5/6-PWR ...
    When currents with harmonics flow through the impedance of the electrical distribution system, voltage distortion can result. Heat can build-up in cabling and ...
  106. [106]
    ECG Small Signal Generation and Analysis | Keysight
    Keysight 4000 X-Series Oscilloscope with fast 1000000 wfms/sec update rate is a powerful tool to overcome the challenge of capturing low amplitude ECG ...
  107. [107]
    Capturing small ECG signals in medical applications
    Capturing and analyzing these signals directly with oscilloscopes is difficult due to the small amplitude. However, the R&S®RTE digital oscilloscope offers key ...
  108. [108]
    Getting Started with Oscilloscope Automation and Python - Tektronix
    The general syntax of SCPI commands is very similar among various manufacturers and instruments, making it easier to communicate with different instruments.What Is A Programmatic... · Downloading And Running An... · Key Python Scpi Commands For...
  109. [109]
    Import and export of data and settings via USB cable
    The live data directory includes a SETUP.scp file for users who want to determine current oscilloscope settings and associated SCPI commands. Open this ...Missing: Ethernet | Show results with:Ethernet
  110. [110]
    Oscilloscope Software: Discover Advanced Analysis Solutions
    With TekScope PC Analysis Software, you can analyze waveforms anytime from anywhere by turning your PC into an oscilloscope.
  111. [111]
    Protocol decoders - sigrok.org
    May 14, 2024 · This is a list of supported protocol decoders (PDs) already supported by libsigrokdecode and also a list of work in progress or yet to be ...
  112. [112]
  113. [113]
    Keysight Technologies Infiniium EXR-Series Offers Advanced ...
    Nov 10, 2020 · Both platforms offer advanced software applications and features that simplify debugging, power measurements and remote collaboration. Designs ...Missing: based post-
  114. [114]
    Oscilloscopes Add 5G NR Analysis Software - Microwaves & RF
    Mar 3, 2022 · Tektronix released SignalVu 5G NR analysis software for their 6 Series B MSO oscilloscopes to perform critical 5G NR measurements per 3GPP ...
  115. [115]
    [PDF] 5/6 Series MSO Firmware Update Read This First - Tektronix
    Open up a Web browser and go to www.tek.com/software. 2. Enter the model number of your oscilloscope in the search box. 3. Scan the list of instrument firmware.
  116. [116]
    Understanding passive oscilloscope probes | Rohde & Schwarz
    Two most common attenuation factors are 1x and 10x; 1x probes are good for small signals, but have limited bandwidth; 10x probes are good for larger signals ...Missing: range | Show results with:range
  117. [117]
    Passive 1x/10x Voltage Probe - Tektronix
    The P2220 and P2221 are compact passive probes with selectable 1X/10X attenuation. ... Compensation range, All oscilloscopes with 1 MΩ input. P2220, 15 to 25 pF.
  118. [118]
    [PDF] P2220 200 MHz 1X/10X Passive Probe Instructions
    The P2220 200 MHz 1X/10X Passive Probe is a compact passive probe with selectable 1X/10X attenuation. ... Compensation Range. 15 pF--25 pF. —. System Input ...
  119. [119]
    Active FET Probe | Tektronix
    The P6205 is designed with FET devices for its inputs, which allows very high input resistance values and low input capacitances.
  120. [120]
    Active Probes | Tektronix
    Tektronix active probes offer high-speed performance, low input capacitance, and connect to oscilloscopes, providing truer signal reproduction for high- ...
  121. [121]
    Types of oscilloscope probes | Rohde & Schwarz
    A standout advantage of active probes is minimal loading over a wide frequency range - thanks to the low input capacitance that translates into a high input ...
  122. [122]
    How to measure current with an oscilloscope | Tektronix
    Apr 6, 2021 · AC/DC current probes use a transformer to measure AC currents and a Hall-Effect device to measure DC current. Since they include active ...
  123. [123]
    N2780B AC / DC Current Probe, 2 MHz, 500 Arms - Keysight
    Free deliveryThe N2780B current probe is an accurate and reliable solution for measuring AC and DC currents. Using a hybrid technology that includes a Hall effect sensor ...
  124. [124]
    Current Probes Datasheet - Tektronix
    This "long nose" style clamp-on probe uses a Hall Effect current sensor to provide a voltage output to oscilloscopes. It has a BNC connector and can be used ...
  125. [125]
    Mixed Signal and Digital Phosphor Oscilloscopes - Tektronix
    P6780 Differential Logic probes provide high-bandwidth connections for up to 16 digital signals. ... The DPO7OE Series Optical probes can be used as an Optical ...
  126. [126]
    [PDF] ABCs of Probes - Tektronix
    Physically attaching the probe to the test point also establishes an electrical connection between the probe tip and the oscilloscope input. For useable ...
  127. [127]
    [PDF] Power Supply Control Loop Response (Bode Plot) Measurements
    A short spring clip ground adapter (usually shipped as an accessory with the probe and shown in the inset image of Figure 3), or better yet, a soldered-in.
  128. [128]
    Making Single-ended Measurements with a Differential Probe
    This paper will examine the use of differential probes in making single-ended measurements on differential signals and consider measurement issues that arise.
  129. [129]
    How Oscilloscope Probes Affect Your Measurement - Tektronix
    This application note describes how an oscilloscope probe changes the signal you're measuring at the test point and what probing specifications to look for to ...Missing: RC | Show results with:RC
  130. [130]
    Oscilloscope buyer's guide | Rohde & Schwarz
    One guideline for digital or other complex signals is to pick a bandwidth that is 3-5 times faster than the fastest clock or data signal. For example, if ...<|control11|><|separator|>
  131. [131]
    How To Calibrate an Oscilloscope: Step-by-Step | Fluke
    ### Step-by-Step Oscilloscope Calibration Procedure
  132. [132]
    NIST Traceable Calibration - Tektronix
    Traceability in Calibrations: A NIST traceable calibration involves maintaining a comprehensive record of the measurement process, connecting a series of ...
  133. [133]
    [PDF] Agilent 1000 Series Oscilloscopes User's Guide - Purdue Engineering
    ... Oscilloscope Utility Settings. 138. Agilent 1000 Series Oscilloscopes User's Guide. Running Self-Calibration. The automatic calibration routine adjusts the ...<|separator|>
  134. [134]
    Floating Oscilloscope Measurements ... And Operator Protection
    “Floating” a ground referenced oscilloscope is the technique of defeating the oscilloscope's protective grounding system – disconnecting “signal common” from ...Missing: spring | Show results with:spring
  135. [135]
    A Guide to ISO 17025 Calibration Standards and Compliance
    This set of standards shows that a lab complies with strict industry regulations and can deliver accurate, reliable results for you.