Monitor unit

A monitor unit (MU) is a standardized measure of radiation output from a linear in , representing the machine's delivery of calibrated to 1 centigray (cGy) at a depth of maximum dose for a 10 cm × 10 cm field at a source-to-surface of 100 cm under conditions. This unit is quantified by the charge collected in a built-in within the accelerator's treatment head, serving as a for the number of particles emitted during beam-on time. In clinical practice, monitor units are essential for treatment planning and delivery, where they are calculated by radiotherapy treatment planning systems (TPS) to ensure the prescribed dose reaches the target volume while minimizing exposure to surrounding tissues. The calculation accounts for factors such as beam energy, field size, depth, and patient-specific geometry, often verified through independent methods like Monte Carlo simulations for accuracy in heterogeneous tissues. Calibration of linear accelerators to this MU system follows international standards, such as those from the International Atomic Energy Agency (IAEA), to maintain dosimetric consistency across institutions. Variations in MU calculations can arise from differences in TPS algorithms or beam modeling, underscoring the importance of commissioning and quality assurance protocols.

Definition and Fundamentals

Definition

A monitor unit (MU) is a machine-specific measure of radiation output from a linear accelerator used in , defined as the quantity of beam delivered when the built-in registers one unit of charge. The linear accelerator is calibrated such that 1 MU corresponds to 1 cGy (centigray) of to under specified reference conditions, providing a standardized metric for . This calibration ensures consistent quantification of beam intensity across treatments. The primary purpose of the monitor unit is to enable reproducible and precise dose delivery in clinical radiotherapy by normalizing the machine's output to a verifiable . By relating the machine's internal monitoring system to , MUs facilitate the planning and verification of patient-specific treatments, where the prescribed dose is achieved through a calculated number of MUs adjusted for geometric and dosimetric factors. Monitor units translate to actual patient absorbed dose through calibration factors that incorporate beam quality, such as photon energy, and account for deviations from reference conditions using output factors and depth-dose data. For photon beams, the standard reference conditions unique to MU definition include a source-to-surface distance (SSD) of 100 cm, a collimator field size of 10 cm × 10 cm at the phantom surface, and measurement at the depth of maximum dose (dmax) specific to the nominal beam energy (typically 1.5 cm for 6 MV photons). These conditions establish the baseline for all subsequent MU-based calculations in treatment planning.

Historical Development

The monitor unit (MU) emerged in the 1970s as linear accelerators became standard in radiotherapy, providing a standardized measure of beam output calibrated at a fixed point, typically 1 cGy per MU under reference conditions, to address inconsistencies in exposure measurements for variable source-to-surface distances in isocentric setups. Prior to this, dosimetry relied on the roentgen unit for exposure, which was inadequate for precise dose delivery in megavoltage beams produced by linacs introduced clinically in the 1950s. A key milestone came with the 1976 ICRU Report 24, which formalized procedures for determining in patients irradiated by X- or gamma-ray beams, emphasizing tissue dose specification at defined points and influencing the integration of into clinical protocols. In the 1980s, the AAPM TG-21 protocol advanced MU calibration by adopting an absorbed-dose-based approach using ionization chambers, replacing earlier air-kerma methods and improving accuracy for photon and beams. This evolution aligned with the broader shift from exposure () to standards, initiated by ICRU recommendations in the 1950s for the unit and culminating in the 1975 adoption of the gray (SI unit), with IAEA TRS-277 (1987) and ESTRO guidelines further standardizing MU calculations on to water. The 1999 AAPM TG-51 protocol marked a significant refinement, simplifying dosimetry for high-energy beams with direct absorbed-dose-to-water calibrations, reducing uncertainties in MU determinations to under 1% for most clinical scenarios and becoming widely adopted globally. In parallel, IAEA TRS-398 (2000, revised 2006) provided an international harmonizing MU calibration with TG-51 principles, incorporating beam quality indices like TPR_{20,10} for consistent output referencing. Modern developments in the 2000s incorporated MUs into advanced techniques like intensity-modulated radiotherapy (IMRT), where complex fluence patterns increased MU demands, and flattening filter-free (FFF) beams, first commercialized around 2010, which required updated output factors due to higher dose rates (up to 2400 MU/min) and forward-peaked profiles, as detailed in the AAPM Therapy Emerging Technology Assessment Work Group report (2015). These updates ensured MU systems adapted to FFF and IMRT without compromising dosimetric precision, reflecting ongoing standardization by IAEA and ESTRO. Subsequent advancements include the 2022 AAPM Report 374 providing practical guidance for TG-51 reference dosimetry and the 2024 Report 385 addendum addressing high-energy electron beams. As of 2025, efforts are underway to update IAEA TRS-398 further, with proposals to harmonize it with TG-51 for unified global practices.

Calibration Principles

Primary Calibration Process

The primary calibration process for a linear accelerator establishes the relationship between monitor units (MU) and to , typically defining 1 MU as delivering 1 cGy under reference conditions. This calibration ensures accurate for clinical radiotherapy beams, focusing on high-energy photons and electrons. The process relies on absolute measurements in a to account for beam-specific characteristics. Essential equipment includes Farmer-type ionization chambers, which are cylindrical air-filled detectors with a 0.6 cm³ sensitive volume suitable for dosimetry in megavoltage beams. These chambers are paired with high-precision electrometers to measure collected charge and water phantoms that simulate tissue-equivalent conditions for depth-dose measurements. The setup also requires a reference field size of 10 cm × 10 cm and source-to-surface distance (SSD) of 100 cm to standardize conditions. The step-by-step procedure begins with positioning the at the reference depth in the water phantom—10 cm for photons in TG-51 or dref = 0.6 R50 - 0.1 cm (typically 0.7–7 cm) for electrons, where R50 is the depth at which the dose falls to 50% of maximum. A known number of (e.g., 200 ) is delivered from the linac, and the records the charge collected by the chamber. Corrections are applied for , , polarity, ion recombination, and electrometer linearity to obtain the corrected charge. The to water is then calculated using the chamber's coefficient in a beam and a beam quality correction factor, yielding the machine-specific output factor that defines the MU-to-dose calibration (e.g., 1 MU = 1 cGy at 10 cm depth for a 10 MV photon beam). Protocols such as the AAPM TG-51 (including the 2024 addendum for ) and IAEA TRS-398 (revised 2024) serve as primary standards for this calibration, emphasizing absorbed dose to as the reference quantity. In TG-51, the beam quality index Q is determined using the percentage depth dose at 10 cm depth (%dd(10)x), measured with a scanning system. TRS-398 employs the ratio (TPR20,10) as the beam quality index, calculated from readings at 20 cm and 10 cm depths in a fixed SSD setup, which minimizes electron contamination effects. Both protocols provide tabulated correction factors for various chamber types and beam energies, ensuring traceability to national standards laboratories. Calibrations are recommended annually by AAPM guidelines to maintain dosimetric accuracy within 1% tolerance, though more frequent checks may be needed after maintenance or beam adjustments. Factors influencing the process include linac-specific beam energies, such as 6-18 for photons and 4-20 MeV for electrons, which require energy-dependent indices and chamber positioning. Variations in filter or beam steering can necessitate re-verification of the output factor.

Dose Quantities and Metrics

In radiation dosimetry for monitor unit calibration, the primary quantity of interest is the absorbed dose, which quantifies the energy deposited by ionizing radiation in a medium. Absorbed dose D is defined as the quotient of the mean energy \mathrm{d}\varepsilon imparted by ionizing radiation to matter of mass \mathrm{d}m traversing a given volume, expressed as D = \frac{\mathrm{d}\varepsilon}{\mathrm{d}m}. Its unit is the gray (Gy), equivalent to 1 joule per kilogram (J/kg), though centigray (cGy), where 1 Gy = 100 cGy, remains common in clinical settings for practical dose reporting. In monitor unit contexts, reference dosimetry focuses on absorbed dose to water D_{w,Q}, measured under standardized conditions for a beam of quality Q, as it approximates dose in tissue-equivalent media. Kerma, or kinetic energy released per unit , represents the kinetic energy transferred from indirectly (e.g., photons) to charged particles per unit , defined as K = \frac{\mathrm{d}E_\mathrm{tr}}{\mathrm{d}m}, where \mathrm{d}E_\mathrm{tr} is the of kinetic energies of those particles. Also measured in , kerma is useful for describing interactions before secondary effects, particularly in air kerma for standards. Historically, served as a precursor quantity, measuring the ionization produced in air by photons, defined as X = \frac{\mathrm{d}Q}{\mathrm{d}m}, where \mathrm{d}Q is the total charge of ions in air \mathrm{d}m; its unit, the (R), equated to approximately 0.0087 in air but has been largely supplanted by metrics due to limitations in accounting for energy absorption. Several metrics are essential for interpreting monitor units in terms of clinical dose delivery, particularly for and beams in radiotherapy. Percentage depth dose (PDD) quantifies beam attenuation in a , defined as the ratio of at a specified depth to the dose at the depth of maximum dose d_m, multiplied by 100, and depends on field size, source-to-surface distance (SSD), and beam energy. Tissue phantom ratio (TPR) provides a similar measure but is SSD-independent, defined as the ratio of doses at two depths in a for the same field size defined at the deeper point, often normalized at 10 cm depth for . These depth-dose metrics enable scaling of reference doses to treatment depths. Scatter factors account for radiation scattering effects that influence output in monitor unit calibrations. The collimator scatter factor S_c (also called in-air output ratio) is the ratio of output in air for a given collimator field size to that for a reference 10 × 10 cm² field, capturing head scatter from the linac collimators. The phantom scatter factor S_p is the ratio of dose per monitor unit at a fixed depth in phantom for a given field to the reference field, isolating phantom-induced scatter; the total scatter factor S_{cp} = S_c \times S_p combines both. Output factors generalize these for non-reference conditions, such as varying applicators in electron beams, where they represent the ratio of dose per monitor unit at d_m to reference conditions. Beam quality specifiers ensure accurate across energies. The %dd(10)_x is the percentage depth dose at 10 cm depth in for a 10 × 10 cm² field under specified SSD and blocking conditions, serving as the primary beam quality index in modern protocols; it determines the beam quality conversion factor k_Q for adjusting calibrations to linac beams. These link monitor units (MUs) to clinical absorbed dose through established formalisms that scale outputs. For example, in SSD setups, the dose D at a point is related to MUs via D = \mathrm{MU} \times S_c \times S_p \times \mathrm{[OAR](/page/Oar)} \times \frac{\mathrm{PDD}}{100}, where OAR is the off-axis , incorporating depth and scatter effects to predict delivered dose from calibrated MU settings. In isocentric setups, TPR replaces PDD for equivalent scaling. Such relations ensure MUs correspond precisely to prescribed doses under conditions. Standardization of units has evolved under the International Commission on Radiation Units and Measurements (ICRU), which defines reference dosimetry conditions and promotes SI units for consistency. The transition to the gray (Gy) from the older rad (1 Gy = 100 rad) occurred in the 1970s, formalized by ICRU recommendations to align with the International System of Units (SI), facilitating global interoperability in absorbed dose measurements for monitor unit calibration. ICRU Report 64, for instance, specifies dosimetry protocols based on absorbed dose to water under reference conditions like 10 cm depth and 10 × 10 cm² field.

Calculation Methods

Basic Monitor Unit Calculation

The basic monitor unit (MU) calculation determines the number of monitor units required to deliver a prescribed radiation dose to a specified point in a for standard beam treatments using source-to-surface distance (SSD) setups. This involves accounting for beam output under reference conditions, depth dose variations, field size effects, and geometric factors. The calculation assumes a linear accelerator calibrated to deliver 1 cGy per MU at the depth of maximum dose (d_max) for a 10 × 10 cm² field at an SSD of 100 cm, as established in primary protocols. The core formula for MU in an SSD setup is: MU = \frac{D_p}{\text{OF}} \times \frac{100}{\text{PDD}} where D_p is the prescribed dose at the point of calculation (in cGy), OF is the output factor (dimensionless, relative to the reference 10 × 10 cm² field), and is the percentage depth dose at the calculation depth, field size, and energy (expressed as a percentage). This formula derives from the machine's calibrated output, adjusted for the relative dose falloff with depth (via PDD) and field size (via OF). The term \frac{100}{\text{PDD}} corrects for the reduced dose at depth relative to d_{\max}, incorporating attenuation, scatter buildup, and geometric divergence under fixed SSD conditions. The output factor (OF) is the product of the collimator scatter factor (S_c) and the phantom scatter factor (S_p), i.e., \text{OF} = S_c \times S_p. S_c accounts for scatter from the collimator jaws, depending on the field size defined at the (typically 100 cm from the source), and is close to 1.0 for the reference 10 × 10 cm² field. S_p represents scatter within the (patient), varying with field size at the depth of calculation and increasing with larger fields due to more scattered photons reaching the point. For non-reference fields, these factors are measured or tabulated and multiply to adjust the reference output (1 cGy/). For isocentric setups, where the source-to-axis distance (SAD) is fixed at 100 cm and the calculation point is at the isocenter, tissue maximum ratio (TMR) or tissue ratio (TPR) replaces PDD, but the SSD formula applies directly for surface-referenced treatments. Assumptions include homogeneous tissue, on-axis calculation points, no beam modifiers (e.g., wedges), and SSD near 100 cm to minimize Mayneord F-factor adjustments for PDD variation with distance. If SSD differs significantly from 100 cm, the PDD must be corrected using the Mayneord factor: F = \left[ \frac{\text{SSD} + d_{\max}}{100 + d_{\max}} \right]^2 \left[ \frac{100 + d}{\text{SSD} + d} \right]^2, and the adjusted PDD used in the formula. Example for open fields: Consider delivering 100 cGy at 10 depth using a 6 MV photon beam, 10 × 10 ² field size, SSD = 100 , and d_{\max} = 1.5 . For this reference field, OF = S_c \times S_p = 1.0 \times 1.0 = 1.0. The PDD at 10 depth is 67% for these conditions. Substituting into the : MU = \frac{100}{1.0} \times \frac{100}{67} \approx 149.3 Step-by-step: Compute the depth dose correction: $100 / 67 \approx 1.4925, yielding 149.25 MU at d_{\max} equivalent, adjusted by the output factor (1.0). This delivers the prescribed dose at the point, assuming primary at 100 SSD. The PDD value accounts for all geometric and dosimetric effects from d_{\max} to the calculation depth.

Secondary and Advanced Calculations

Secondary calculations for () verification in radiotherapy often employ the Clarkson sector integration algorithm to account for irregular field shapes, where the dose is computed by dividing the field into sectors and integrating () or tissue maximum ratios (TMR) across them. This method improves accuracy over equivalent square approximations for highly irregular fields by explicitly modeling primary and scatter contributions from each sector, typically achieving errors below 2% compared to measured doses in clinical setups. Superposition and methods serve as secondary verification tools by convolving a kernel representing energy deposition with the incident fluence, enabling MU checks for heterogeneous media or non-standard geometries without relying on treatment planning system outputs alone. These algorithms, such as the collapsed cone superposition, provide independent dose estimates at prescription points, with typical agreement within 3% for beams in validations. Advanced calculations for intensity-modulated radiotherapy (IMRT) and volumetric modulated arc therapy (VMAT) verify total MU using independent methods that compute the required output from planned fluence maps or control point segments, ensuring agreement with the treatment planning system within 2% for cases like and head-and-neck. Guidelines recommend independent MU verification for these modalities using simplified models that sum segmental doses from beamlets, often achieving sub-2% discrepancies relative to planning systems for and head-and-neck cases. For beams, MU determinations in advanced scenarios utilize the en bloc method for fields with custom cutouts, where output factors are derived from equivalent squares based on projections, or extended source-to-surface (SSD) techniques that apply effective SSD shifts or air-gap corrections to adjust percent depth doses. These approaches maintain accuracy within 2-4% for extended SSDs up to 120 cm, as validated in AAPM protocols for irregular electron applicators. Formula extensions for MU calculations integrate beam modifiers into the core equation, expressed as: \text{MU} = \frac{\text{Dose}}{\text{Output} \times \text{WF} \times \text{TF} \times \text{OAR}} where Output is the calibrated dose per at reference conditions, WF is the wedge factor accounting for transmission (typically 0.5-0.9 for 15°-60° wedges), TF is the tray factor for blocking accessories (often 0.95-1.00), and OAR is the off-axis ratio for points displaced from the central axis (near 1.00 on-axis, decreasing laterally). These factors are measured at the calculation depth and field size, with WF and TF exhibiting depth dependence due to scatter buildup. Software-independent hand calculations for wedged fields involve multiplying the open-field MU by the WF at the wedge's effective depth, adjusted for field obliquity via isocentric corrections if incidence exceeds 10°. For example, in a 6 MV photon beam treating a 10 × 10 cm field at 100 cm SAD with a 45° wedge to deliver 200 cGy at 10 cm depth, the MU is computed as 200 / (1.00 × 0.72 × 1.00 × 1.00) ≈ 278 , where 0.72 is the measured WF; oblique incidence adds a cosine correction to the effective depth for surfaces angled up to 30°. Such manual verifications ensure robustness against planning errors, with tolerances of ±5% for complex modifiers in .

Clinical Applications and Verification

Integration in Treatment Planning

In radiotherapy treatment planning systems () such as Varian's Eclipse and Philips' Pinnacle, monitor unit () calculations are integrated as an essential input for optimizing dose distributions and ensuring accurate delivery. These systems employ advanced algorithms, including convolution-superposition or Monte Carlo-based models, to compute MUs during plan optimization, balancing target coverage with organ-at-risk sparing while incorporating beam data from linac commissioning. The role of MUs extends to iterative refinement, where planners adjust parameters like beam angles and weights to meet clinical objectives before final dose normalization. The workflow for MU integration commences with CT simulation to acquire patient-specific geometry, including immobilization and anatomical contours, which informs subsequent beam modeling and heterogeneity corrections. Contouring of targets and organs precedes beam placement, after which the TPS performs forward or inverse to generate initial dose maps; MUs are then normalized, commonly to 1 cGy per MU at the isocenter under reference conditions, accounting for depth, size, and tissue interfaces. This normalization ensures the prescribed dose aligns with linac , with the TPS iteratively recalculating MUs to reflect patient geometry variations, such as density or bone scattering, for precise output scaling. For specific modalities, MU scaling in three-dimensional conformal radiotherapy (3D-CRT) involves per-beam adjustments based on equivalent square field sizes and effective depths, enabling straightforward delivery with static positions. In volumetric modulated arc therapy (VMAT), MUs are optimized across numerous control points—typically 177 per arc in —allowing dynamic modulation of speed, motion, and dose rate for efficient conformal dosing. Hybrid approaches, often used in gynecological cancers, integrate external beam MUs with intracavitary-interstitial dwell times within the TPS to achieve cumulative equivalent dose in 2 Gy fractions (EQD2) objectives, scaling EBRT contributions to complement high-dose-rate boosts. Upon plan approval, the TPS outputs MU values in a digital imaging and communications in medicine (DICOM) record and treatment (DCM RT) file, which interfaces directly with the linear accelerator for execution. This file specifies total MUs per beam or arc, triggering the linac's monitor chamber—a transmission ionization chamber in the treatment head—to regulate beam-on time and terminate delivery once the integrated charge matches the prescribed MUs, ensuring dosimetric fidelity. The monitor chamber's calibration, verified against TPS models, maintains output constancy at 1 cGy/MU under standard conditions, with the system halting irradiation if deviations exceed tolerances.

Quality Assurance and Verification

Quality assurance protocols for monitor units (MUs) in radiotherapy emphasize routine checks to maintain output constancy and ensure precise dose delivery. Daily output verification is conducted using ionization chambers or diodes in a phantom setup to confirm photon beam output within a ±3% tolerance, helping detect immediate deviations in linac performance. Monthly dosimetry verifications involve thermoluminescent dosimeters (TLDs) or diodes to assess output constancy at ±2%, alongside checks for beam profiles and flatness to validate MU calculations against baseline values. Verification methods focus on confirming delivered during patient treatment. In-vivo dosimetry employs metal-oxide-semiconductor field-effect transistors (MOSFETs) for real-time dose measurements at the skin entrance, enabling direct confirmation by comparing measured doses to planned values with tolerances typically within ±5%. For intensity-modulated (IMRT), electronic portal imaging device (EPID)-based captures the transmitted beam to validate accuracy by reconstructing 2D dose distributions and identifying discrepancies in modulated fields. Common error sources include linac drift from component aging or wear, and beam asymmetry due to or misalignment, both of which can lead to MU inaccuracies exceeding 2%. The AAPM Task Group 142 guidelines recommend a ±2% tolerance for beam MU delivery to minimize risks, with annual tests ensuring set MUs match delivered doses within ±1 MU or ±2%, whichever is greater. Incidents of MU errors have resulted in significant under- or over-dosing; for instance, an incorrect exposure time calculation in a unit led to a 166% overdose for 115 patients over several months, prompting the adoption of independent double-checks and audits as corrective measures. In another case, miscalibration caused a 25% overdose in 207 patients, highlighting the need for regular intercomparison exercises to prevent recurrence.

References

  1. [1]
    [PDF] Monitor Unit (MU) Calculation - AAPM
    A monitor unit (MU) is a a measure of machine output by a built-in monitor ionization chamber of a proton machine. – 1MU equal to dose delivered by a proton ...
  2. [2]
    Monitor units (MU)? - RT Medical Systems
    Apr 24, 2023 · In the realm of radiation oncology, the amount of radiation delivered to a patient is quantified in units known as monitor units (MU). The MU ...<|control11|><|separator|>
  3. [3]
    Comparison of monitor units calculated by radiotherapy treatment ...
    In radiation therapy, the monitor units (MU) needed to deliver a treatment plan are calculated by treatment planning systems (TPS). The essential part of ...
  4. [4]
    Calculating Monitor Unit (MU) in Radiation Therapy with Monte ...
    Apr 24, 2023 · Monitor Unit (MU) is an essential parameter in radiation therapy that determines the amount of radiation to be delivered to the patient ...
  5. [5]
    Comparison of monitor units calculated by radiotherapy treatment ...
    In radiation therapy, the monitor units (MU) needed to deliver a treatment plan are calculated by treatment planning systems (TPS).
  6. [6]
    AAPM's TG‐51 protocol for clinical reference dosimetry of high ...
    Sep 7, 1999 · A protocol is prescribed for clinical reference dosimetry of external beam radiation therapy using photon beams with nominal energies ...
  7. [7]
    [PDF] Absorbed Dose Determination in External Beam Radiotherapy
    Radiation treatment unit and reference conditions for Dw,Q determination. Accelerator: Nominal Acc. potential: MV. Nominal dose rate: MU/min Beam quality, Q ...
  8. [8]
    [PDF] the history and development of the clinical linear accelerator - SPRMN
    Jun 20, 2006 · The first medical linac treated its first patient in 1953. This review covers the history and development of clinical linacs, with a brief look ...
  9. [9]
    [PDF] Monitor Unit Calculations for External Photon and Electron Beams
    Jul 25, 2001 · The accurate determination of dose per monitor unit (MU) at a single calculation point is an essential part of this process. There are many ...
  10. [10]
    [PDF] TRS 430 - Scientific, technical publications in the nuclear field | IAEA
    Monitor unit/time check . ... radiotherapy physics”, Radiotherapy Physics in Practice (WILLIAMS, J.,. THWAITES, D.I., Eds), Oxford University Press, Oxford ...
  11. [11]
    [PDF] AAPM's TG-51 protocol for clinical reference dosimetry of high ...
    A protocol is prescribed for clinical reference dosimetry of external beam radiation therapy using photon beams with nominal energies between 60Co and 50 MV ...
  12. [12]
    AAPM Practice Guideline 8.b: Linear Accelerator Performance Tests
    Oct 4, 2023 · The recommended frequency and action limit are listed with each test. Implementation notes on each test follow the tables. These tests were ...
  13. [13]
    None
    ### Extracted Definitions and Historical Context
  14. [14]
    None
    Summary of each segment:
  15. [15]
    [PDF] Addendum to the AAPM's TG-51 protocol for clinical reference ...
    Mar 12, 2014 · change in reading per monitor unit from beam-on to stabi- lization and that equilibration should be achieved in less than. 5 min. Leakage ...
  16. [16]
    ICRU Report 64, Dosimetry of High-Energy Photon Beams Based on ...
    This Report examines the methods by which absorbed dose to water can be determined for photon radiations with maximum energies from approximately to 1 MeV to ...
  17. [17]
    Monitor Unit Hand Calculations | Oncology Medical Physics
    Percent Depth Dose (PDD) is the ratio of dose at a given depth to maximum dose. PDD is a function of depth, source to surface distance, and field size.
  18. [18]
    Monitor unit calculations for external photon and electron beams
    A protocol is presented for the calculation of monitor units (MU) for photon and electron beams, delivered with and without beam modifiers.
  19. [19]
    (PDF) Clarkson's Sector Integration Method for Radiation Fields ...
    Jan 23, 2019 · Clarkson's sector integration method is a dose calculation method which offers far better accuracy in dose calculation of irregular fields.
  20. [20]
    A new convolution/superposition dose calculation method for ...
    Jun 4, 1998 · This thesis was dedicated to developing a convolution/superposition dose calculation method for computing patient dose distributions as well ...
  21. [21]
    Experimental verification of convolution/superposition photon dose ...
    This work describes an experimental verification of the two-photon dose calculation engines available on the Helax-TMS (version 6.1) commercial radiotherapy ...
  22. [22]
    Report of AAPM Task Group 219 on independent calculation‐based ...
    Jul 2, 2021 · This report provides guidelines on calculation-based dose/MU verification for intensity modulated radiation therapy (IMRT) or volumetric modulated arc therapy ...Missing: S_c × S_p
  23. [23]
    Independent calculation of monitor units for VMAT and SPORT - PMC
    Dose and monitor units (MUs) represent two important facets of a radiation therapy treatment. In current practice, verification of a treatment plan is ...
  24. [24]
    [PDF] Verification of monitor unit calculations for non-IMRT clinical ...
    Dec 30, 2010 · The requirement of an independent verification of the monitor units (MU) or time calculated to deliver the prescribed dose to a patient has ...
  25. [25]
    Monitor unit calculations for external photon and electron beams ...
    Feb 26, 2014 · For electron beams, the formalism provides for calculations at the standard and extended SSDs using either an effective SSD or an air-gap ...
  26. [26]
    [PDF] Monitor Unit Calculations for Photons and Electrons: Report of TG-71
    For rectangular fields, use Square Root. Method: Se(ra,LxW) = [Se(ra,LxL) · Se(ra,WxW)]1/2. Many irregular ...
  27. [27]
    Eclipse | Varian
    Eclipse is a treatment planning system that leverages advances in imaging, treatment planning, and delivery, with features like RapidArc Dynamic and RapidPlan.
  28. [28]
    Beam complexity and monitor unit efficiency comparison in two ...
    Mar 10, 2021 · The Pinnacle 3 Auto-Planning system can optimize BC and NPC plans to achieve the same plan quality using both the Trilogy and Synergy systems.
  29. [29]
    [PDF] Eclipse Photon and Electron Algorithms Reference Guide
    Configuring the AAA or Acuros XB Using Varian Eclipse Beam Data. Varian Eclipse Beam Data is provided by Varian at the installation of the calculation.
  30. [30]
    [PDF] Commissioning of Radiotherapy Treatment Planning Systems
    (2). Calculate with RTPS MU/time needed to deliver 2 Gy to the reference point #5. (3). Perform manual MU calculation for each field and compare the results ...
  31. [31]
    Radiation Therapy Treatment Process | OncoLink
    Jun 26, 2025 · Step by step look at the process of radiation therapy, from CT Simulation to treatment.
  32. [32]
    Evaluation of Inhomogeneity Correction Performed by Radiotherapy ...
    This machine operates with a dose rate of 350 MU/min (which is the pulse rate used in clinical practice) was calibrated to deliver 1 cGy/MU at iso-center for a ...
  33. [33]
    Assessment of the Monitor Unit Objective tool for VMAT in the ... - NIH
    Feb 25, 2018 · This work aims to achieve the highest possible monitor units (MU) reduction using the MU Objective tool included in the Eclipse treatment planning system.
  34. [34]
    Advances in External Beam Radiation Therapy and Brachytherapy ...
    A hybrid intracavitary/interstitial technique can allow for improved dosimetry for larger tumours, irregular tumour shape, parametrial extension, vaginal ...
  35. [35]
    The American Brachytherapy Society (ABS) consensus guidance for ...
    May 29, 2025 · Hybrid brachytherapy treatment planning. Following applicator and needle insertion, the patient is imaged for treatment planning with one or ...
  36. [36]
    The lifetime of a linac monitor unit ion chamber - PMC - NIH
    Nov 11, 2021 · This study is the first to report the clinical lifetime of Varian Kapton sealed ion chambers as a retrospective review.