Köhler illumination

Köhler illumination is a fundamental technique in optical that provides uniform and even illumination of the specimen by imaging the light source onto the aperture diaphragm of the rather than directly onto the sample, ensuring optimal contrast and resolution without artifacts from the light source's structure. Developed in 1893 by August Köhler, a microscopist at in , , the method was initially motivated by the need to improve photomicroscopy with early long-exposure photographic plates, addressing inconsistencies from gas lamps and other uneven light sources prevalent at the time. The optical principles of Köhler illumination rely on a two-stage process: first, the source is sharply focused in the of the condenser's to produce parallel bundles of rays that uniformly fill the objective's ; second, an image of the field is focused onto the specimen , controlling the illuminated area without projecting the source's irregularities. This setup, applicable to both transmitted and epi-illuminated (reflected) microscopy, allows precise control of the cone's angle via the condenser , which directly influences the microscope's effective and thus its —typically set to 65-80% of the objective's pupil diameter for balanced . Among its key advantages, Köhler illumination delivers grainless, homogeneous lighting across the field of view, minimizing , dust effects, and unevenness that could degrade image quality, making it indispensable for high-fidelity observation and of thin specimens (up to about 8 micrometers thick). It remains a standard in modern setups, foundational for advanced techniques like phase contrast, differential interference contrast (), and confocal imaging, and requires recalibration whenever the objective or light source is changed to maintain performance.

Historical Context

Invention and Inventor

August Köhler, born on March 4, 1866, in , , was a physicist and microscopist who studied a range of scientific disciplines including , , , physics, and at institutions such as the Technical University of , the University of , and the University of Giessen, from which he received his doctorate in 1893. Early in his career, Köhler worked as a teacher while pursuing research in and photomicrography, fields that were rapidly evolving in during the late 19th century. His association with the University of Jena came later, through his long-term role at in that city, where he joined as a staff member in 1900 and eventually became head of the microscopy, microphotography, and projection department in 1938; in 1922, the university appointed him professor of microphotometry and projection, recognizing his contributions to optical sciences. In 1893, at the age of 27, Köhler invented a novel illumination technique specifically designed to enhance the quality of photomicrographs by providing uniform, glare-free lighting without imaging the light source directly onto the specimen. This method, now known as Köhler illumination, addressed key limitations in existing practices and was first detailed in his seminal titled "Ein neues Beleuchtungsverfahren für mikrophotographische Zwecke" (A New Illumination Method for Photomicrographic Purposes), published in the German journal Zeitschrift für wissenschaftliche Mikroskopie und für Mikroskopische Technik, volume 10, issue 4, pages 433–440. An English of the paper appeared the following year in the Journal of the Royal Microscopical Society, broadening its impact among international researchers; in the original work, Köhler emphasized the technique's ability to separate the illumination of the object plane from the aperture plane, ensuring even field illumination critical for high-fidelity imaging. The invention occurred amid significant advancements in during the late , a period when Abbe's theoretical work at had improved designs and , yet illumination remained a persistent challenge for accurate specimen visualization and . Prior methods, such as critical illumination—which projected the light source directly onto the sample—often resulted in uneven brightness, visible dust particles, and glare that distorted images, particularly in photomicrography used for scientific documentation. Köhler's approach, developed independently during his doctoral research, effectively superseded critical illumination by introducing a collector system that focused uniformly, marking a foundational shift toward modern optical standards.

Motivation and Development

Prior to the development of Köhler illumination, critical illumination was the predominant method in microscopy, but it suffered from significant drawbacks that compromised image quality. In critical illumination, the light source is imaged directly onto the specimen plane, leading to uneven lighting across the field of view due to irregularities in the light source, such as the filament structure in gas lamps or early electric bulbs. This resulted in glare from visible images of the light source, which interfered with clear specimen imaging and introduced artifacts, particularly problematic in photomicrography where long exposure times amplified these inconsistencies. August Köhler sought to address these issues by devising a system that provided uniform, artifact-free illumination through the separation of the light source from the specimen . His approach ensured that the specimen was illuminated by a homogeneous field rather than a direct projection of the source's imperfections, thereby minimizing glare and achieving even brightness without wasting light through diffusers. This innovation was particularly motivated by the needs of microphotography, where consistent illumination was essential to avoid objectionable backgrounds and overheating during exposures. Following Köhler's 1893 publication, the technique was rapidly adopted by for integration into their designs, marking a shift toward standardized optimal illumination in scientific . Early refinements in the late 1890s and early 1900s focused on adapting the method to evolving light sources, such as from gas to electric lamps, and improving alignment for broader applicability, solidifying its role as a foundational technique by the early .

Optical Principles

Conjugate Planes and Image Formation

In Köhler illumination, conjugate planes refer to pairs of optical planes that are in common focus along the light path, enabling the separation of illumination and imaging functions. There are two primary sets of conjugate planes: the aperture planes, which control the light source and angular distribution of illumination, and the field planes, which handle the spatial distribution and imaging of the specimen. The aperture conjugate planes include the lamp filament (or light source), the condenser aperture diaphragm (located at the front focal plane of the condenser), the back focal plane of the objective, and the eyepoint (Ramsden disk) of the eyepiece. Similarly, the field conjugate planes consist of the field diaphragm, the specimen plane, the intermediate image plane (at the fixed diaphragm of the eyepiece), and the final image plane (such as the retina, film, or sensor). This configuration ensures that the light source is imaged solely within the aperture planes, preventing its structure from interfering with the specimen image. The key to uniform illumination lies in defocusing the light image at the specimen . In this setup, the collector lens projects an enlarged image of the lamp onto the condenser aperture diaphragm, where it is precisely ed as part of the aperture conjugate set. From there, the condenser lens system transforms the diverging rays from the into parallel bundles that uniformly flood the specimen, but the itself remains out of focus at the field , avoiding the projection of details—such as filaments, , or irregularities—onto the specimen. This defocused state at the specimen ensures even brightness across the field of view without introducing glare or unevenness from the source's inherent non-uniformity. Regarding ray paths, parallel rays originating from different points on the light source are directed by the to converge at the back focal plane of , forming an image of the (and indirectly the source) there. These rays pass through the specimen in a collimated manner, maintaining uniformity, and are then focused by to form a sharp of the specimen at the intermediate plane, independent of the illumination details. This separation of ray paths—illuminating rays focused in planes and rays in planes—optimizes and by confining source artifacts to the non-imaging optical train.

Key Components and Ray Paths

Köhler illumination employs several key optical components to achieve uniform specimen illumination while preventing the light source's image from appearing in the final view. The primary elements include the collector lens, field diaphragm, condenser aperture diaphragm, and condenser lens. The collector lens, typically positioned within the lamp housing, gathers divergent rays from the light source—such as a tungsten-halogen —and projects an enlarged, focused image of the filament onto the plane of the condenser aperture diaphragm. The field diaphragm, located between the collector lens and the condenser, controls the size of the illuminated area by limiting the beam's width and serves as the illumination system's virtual light source. The condenser aperture diaphragm, situated at the front focal plane of the condenser lens, regulates the angle of the entering the condenser, thereby controlling the of the illumination and influencing and . Finally, the condenser lens system images the field diaphragm onto the specimen plane, focusing the diverging light from it to ensure even distribution without projecting the filament's structure. In the ray paths of Köhler illumination, originates from the lamp and is collected by the collector , which images the precisely at the condenser aperture ; this placement ensures that the source's irregularities do not affect the specimen's illumination uniformity. From there, the condenser redirects the , forming an of the field directly at the specimen , which defines the exact area illuminated and eliminates beyond the field of view. These paths exploit conjugate planes in the illumination train, where the field is conjugate to the specimen, and the is conjugate to the objective's back focal . The setup for transmitted and reflected illumination shares core components but differs in configuration and ray tracing to accommodate direction. In transmitted , the is mounted below the specimen stage, directing upward illumination through the sample, with the field diaphragm imaged at the specimen plane and the at the aperture; adjustment of the height focuses these images accurately. For reflected (epi-)illumination, serves a as both and , eliminating the need for a substage ; from the source passes through to the specimen and reflects back along the same path, with the aperture diaphragm positioned closer to the source and the field diaphragm conjugate to the specimen plane, simplifying alignment but requiring the objective's to match the illumination cone.

Advantages and Limitations

Primary Benefits

Köhler illumination ensures uniform illumination across the entire field of view by imaging the light source in the plane of the condenser aperture diaphragm, thereby eliminating hot spots, glare, and artifacts from imperfections or dust in the light source. This results in a grainless, extended light field at the specimen plane, providing consistent intensity without the uneven brightness or filament shadows common in critical illumination setups. The technique enhances contrast and resolution through independent control of illumination parameters using the and field diaphragms, allowing precise adjustment of and without affecting the path. By setting the to illuminate 65-90% of pupil, microscopists can optimize the balance between direct and scattered , maximizing while minimizing fringes or excessive glare, particularly for specimens with varying transparency. Furthermore, Köhler illumination is highly compatible with advanced microscopic techniques such as phase contrast, differential interference contrast (), and fluorescence microscopy, as it maintains uniform excitation without introducing wavelength-specific interference or that could degrade image quality. This setup supports high-fidelity imaging in these modalities by ensuring the delivers a clean, adjustable that aligns with the requirements of specialized .

Drawbacks and Constraints

Köhler illumination demands precise alignment of the light source, , and field diaphragm to achieve optimal performance, a process that involves multiple steps including centering the lamp filament, focusing the , and adjusting diaphragms, often requiring significant operator skill and time, particularly for users unfamiliar with the . This alignment sensitivity can lead to issues such as or uneven illumination if the is even slightly off-center, necessitating frequent recalibration when changing objectives or specimens. The system's reliance on multiple optical elements, including collector lenses and the , results in reduced light efficiency, as it does not utilize the full surface area or angular distribution of the light source, leading to losses that can limit its effectiveness in low-light applications such as certain fluorescence microscopy setups where maximizing excitation light is critical. For instance, with non-planar sources like arc lamps, only light from the central plane is properly focused, contributing to further inefficiencies and potential unevenness in illumination intensity. In setups involving very high condensers, establishing Köhler illumination presents additional challenges, as the field may not close sufficiently small to be imaged clearly in the , or the may fail to produce a visible image, often requiring adjustments like switching to lower objectives or specialized condensers to mitigate these issues. Furthermore, in ultramicroscopy techniques that demand uniform illumination over large fields of view, traditional Köhler configurations often struggle to maintain homogeneous intensity, potentially compromising image quality in extended imaging volumes.

Practical Implementation

Equipment Requirements

Köhler illumination in transmitted microscopy necessitates a equipped with an adjustable substage that includes both an iris for controlling the angle of illumination and a field iris for defining the illuminated area, along with a collector to focus the . The must be vertically focusable and horizontally centerable to enable precise positioning relative to the specimen plane. A stable , typically a 12V or a modern LED equivalent, is required to provide even, adjustable intensity without introducing artifacts from the structure. For enhanced alignment accuracy, optional tools such as a phase telescope or can be inserted into the tube to visualize the 's back focal plane, aiding in centering. A pinhole test slide may also be used to verify illumination uniformity by projecting a sharp pinhole image onto the intermediate . In reflected light configurations, including epi-illumination systems commonly used in metallurgical or , the substage is omitted in favor of a vertical illuminator module that houses the and field diaphragms, a collector , and a beamsplitter—such as a half-silvered mirror—to direct light downward through the , which serves dual roles as and . The light source remains similar, often a tungsten-halogen or housed externally, ensuring the filament image forms in the 's rear focal plane for uniform specimen illumination.

Alignment and Setup Procedure

The alignment and setup procedure for Köhler illumination ensures uniform, artifact-free lighting across the field of view by properly configuring the light source, , and diaphragms. This process typically begins after mounting a specimen and selecting a low-magnification , such as 4x or 10x, to facilitate easier adjustments. The controls the size of the illuminated area, while the regulates the angle of light entering the to optimize contrast and resolution. Follow these sequential steps for setup on a standard brightfield :
  1. Place a thin, translucent specimen on the stage and sharply on it using the coarse and fine knobs, with the positioned approximately 0.5 cm below the slide for an initial starting point. This establishes the and allows visualization of the field.
  2. Center the light source by adjusting the lamp housing or collector lens to ensure the or light emitter is aligned with the ; for lamps, project the image onto a white card placed at the front and use centering screws to position it centrally, while LED sources often require less adjustment due to their diffuse output.
  3. Close the field diaphragm fully and remove one to view the back of ; adjust the condenser height () until the edges of the field diaphragm appear sharp and in focus, then center the diaphragm image using the condenser's lateral adjustment knobs if it is off-axis. This step images the uniformly at the condenser plane.
  4. Replace the eyepiece and gradually open the field diaphragm until its edges just fill the visible field of view without encroaching on the image; this ensures the specimen is evenly illuminated without stray light artifacts.
  5. Adjust the condenser focus further if needed to achieve completely even illumination across the field, and set the aperture diaphragm to illuminate approximately 70-80% of the objective's back focal plane (visible by removing the eyepiece again) for balanced resolution and contrast.
For upright microscopes, condenser adjustments are made from below the stage, whereas inverted models require accessing controls from the side or top due to the stage's position above the condenser; consult the instrument manual for exact knob locations in both cases. LED light sources simplify filament centering but may necessitate intensity calibration to match levels for consistent results. To prevent specimen damage, particularly for light-sensitive biological samples, monitor and limit during setup to avoid over-illumination, starting at a low voltage (e.g., 6-7V for lamps) and increasing only as needed. Always ensure the does not contact the slide to avoid displacing the specimen.

Testing and Troubleshooting

To verify a proper Köhler illumination setup, microscopists employ several targeted testing techniques that confirm the alignment of conjugate planes and uniform light distribution. One primary method involves inserting a phase telescope or Bertrand lens into the imaging path and defocusing the specimen slightly; this allows observation of the light source (or ) in the back focal plane of , where it should appear as a uniformly illuminated without discernible structure or hotspots, indicating that the source image is correctly conjugated to the aperture . Another verification step requires closing the field diaphragm until its edges form a sharp polygon within the field of view, then focusing and centering these edges using the condenser focus and adjustment screws; the edges should appear crisp and centered, after which the diaphragm is opened until just vanishing from view, ensuring even coverage without spillover. Finally, placing a flat, uniform test slide (such as a blank coverslip) on the stage and observing the field under low magnification reveals even illumination if the background appears homogeneously bright without gradients or shadows. Common problems in Köhler illumination often stem from subtle misalignments or adjustments, each with straightforward diagnostic and corrective actions. Uneven lighting across the field, manifesting as brighter or dimmer regions, typically results from a decentered condenser; this is resolved by recentering the condenser using its adjustment screws while monitoring the field diaphragm edges or light source image through the phase telescope. Glare or excessive brightness, which reduces contrast, arises from an overly open condenser aperture diaphragm; closing it to fill approximately 70-90% of the objective's back focal plane (viewed via phase telescope) minimizes stray light while preserving resolution. Vignetting, characterized by darkened edges in the field of view, occurs when the field diaphragm is insufficiently open or the condenser is misfocused; opening the diaphragm until its edges are just outside the view and refocusing the condenser eliminates this issue. Key indicators of misalignment include the visibility of source structure—such as filament details or lamp irregularities—directly in the , which violates Köhler's of separating source and specimen ; this signals improper conjugation and requires rechecking the collector lens and condenser . Similarly, blurred or off-center field diaphragm edges during testing denote errors in the condenser, prompting adjustment of the condenser height knob. or fibers appearing sharply in multiple conjugate planes (e.g., at the specimen and ) further indicate issues, necessitating cleaning of accessible components like the condenser front before retesting. These diagnostics ensure reliable performance, particularly after objective changes or environmental shifts like temperature variations that can alter alignments.

Applications

In Conventional Microscopy

Köhler illumination serves as the foundational technique for achieving even and uniform specimen illumination in standard , particularly in routine imaging applications within biological and material sciences. By focusing an image of the light source in the plane while projecting a defocused image onto the specimen, this method ensures that the sample is bathed in homogeneous light without artifacts from the lamp filament, enabling clear visualization of cellular structures in tissue sections or microstructural features in metals and polymers. In transmitted microscopes equipped for conventional setups, Köhler illumination integrates seamlessly with achromatic objectives and Abbe condensers to optimize light delivery. Achromatic objectives, which correct for across visible wavelengths, benefit from the even illumination to maintain and color fidelity, while the collects and directs parallel light rays through the specimen plane, adjustable via iris diaphragms for control. This configuration is in upright and inverted microscopes used for transmitted observations, supporting magnifications from 4x to 100x without introducing uneven brightness or glare. Since its introduction in 1893 by August Köhler at , this illumination method has been a staple in educational and environments, with widespread adoption in the early as transitioned to standardized optical systems. It remains integral to teaching basic histological techniques in biology curricula and routine quality control in materials labs, where consistent illumination is essential for reproducible results across diverse samples. The uniformity provided by Köhler illumination enhances contrast in unstained specimens, facilitating subtle detail detection without additional filters.

In Advanced and Modern Techniques

Köhler illumination is highly compatible with , where it ensures uniform light distribution to translate phase shifts in unstained specimens into detectable intensity differences, enhancing contrast for transparent samples like live cells up to 10 µm thick. This setup requires a specialized with a phase annulus aligned precisely with the objective's phase ring, minimizing halo artifacts and providing even illumination across the field. In , Köhler illumination delivers collimated, polarized light essential for shearing wavefronts and exaggerating gradients, yielding high-resolution, three-dimensional-like images without halos in thick specimens. Proper alignment of the aperture and field diaphragms in the vertical illuminator maintains phase relationships between orthogonal beams, optimizing contrast in both transmitted and reflected configurations. For polarization microscopy, Köhler illumination forms the foundational step in alignment, providing even polarized light to maximize visibility and extinction in anisotropic materials, thereby improving contrast through crossed polarizers. Modern adaptations of Köhler illumination incorporate LED-based sources, which offer energy-efficient, low-heat illumination compared to traditional lamps, enabling longer imaging sessions with reduced . LED arrays can replace conventional condensers to achieve programmable illumination patterns, enhancing and in computational methods like Fourier ptychography without multiple exposures. Post-2015 developments extend its use to through concurrent transmitted light detectors, where optimized Köhler alignment with non-absorbed wavelengths (e.g., 561 nm) registers real-color images with for pigmented samples like cells. In super-resolution techniques such as structured illumination microscopy (SIM), Köhler illumination is modified by overlaying sinusoidal patterns to shift high-frequency information into the observable Fourier domain, doubling lateral to approximately 100 nm. These advancements support applications in live-cell imaging, where Köhler-enhanced phase contrast reveals dynamic cellular structures with minimal disturbance, as seen in observations of unstained biological specimens. In nanotechnology, Köhler illumination integrates with microsphere-assisted to boost effective , enabling super-resolution nano-imaging of phase gratings and reducing via optimized diaphragms and . Epi-illumination variants of Köhler provide glare-free reflected light for opaque , using the objective as a to uniformly project the source onto the rear focal plane, ideal for semiconductors and thin films in DIC or setups.

References

  1. [1]
    Köhler Illumination - Thorlabs
    Jun 18, 2021 · Köhler illumination avoids imaging the source to the sample or sensor planes and provides uniform illumination to the sample plane.
  2. [2]
    Microscopy Basics | Kohler Illumination - Zeiss Campus
    Illumination of the specimen is the most important variable in achieving high-quality images in microscopy and critical photomicrography or digital imaging.
  3. [3]
    Koehler Illumination: A Brief History and a Practical Set Up in Five ...
    Aug 17, 2017 · The technique of Koehler Illumination is one of the most important and fundamental techniques in achieving optimum imaging in any given light microscope set-up.
  4. [4]
    August Köhler Age, Death, Wife, Family, Biography - StarsUnfolded
    He was appointed as a professor of microphotometry at the University of Jena and also became an honorary professor of the Medical Faculty in 1922. He filed a ...
  5. [5]
    K?hler, August Karl Johann Valentin - Encyclopedia.com
    In 1900 Köhler joined the Zeiss firm. He spent the rest of his life in Jena. In 1922 the university made him professor of microphotometry and projection, a post ...
  6. [6]
    A Microscopy Journey Through Time - ZEISS
    Illumination device with separate control of the luminous field and condenser aperture: August Köhler (1866-1948).Missing: biography | Show results with:biography
  7. [7]
    Proper Alignment of the Microscope - ScienceDirect.com
    His technique was originally published in 1893 in the Zeitschrift für wissenschaftliche Mikroskopie, to be followed in English in the Journal of the Royal ...Missing: invention | Show results with:invention<|control11|><|separator|>
  8. [8]
    Microscope Illumination - Molecular Expressions
    August Köhler (1866-1948) - August Köhler, a German scientist and photomicrographer born in 1866, is best known for his development of a superior microscope ...
  9. [9]
    [PDF] Köhler illumination: A simple interpretation - Microscopist.co.uk
    Köhler's initial difficulties were caused especially by the irregularities of the light sources available to ... Köhler illumination in the modern microscope.
  10. [10]
    Clear and bright images with Koehler Illumination - ZEISS
    Sep 11, 2023 · Koehler Illumination achieves a homogeneously illuminated object plane imaged with optimal contrast and resolution.<|control11|><|separator|>
  11. [11]
    Microscopy Basics | Illumination and the Optical Train - Zeiss Campus
    Conjugate planes are in common focus and are critical in achieving proper Köhler illumination. A second set of planes, known as the image-forming conjugate ...
  12. [12]
    Köhler Illumination - Evident Scientific
    Köhler illumination was first introduced in 1893 by August Köhler of the Carl Zeiss corporation as a method of providing the optimum specimen illumination.
  13. [13]
  14. [14]
  15. [15]
    Understanding the microscope. 3. Illumination. By Jeremy Sanderson
    Devised about 120 years ago, Köhler's illumination was based on sound optical principles, and has therefore become universally accepted. It is by no means the ...
  16. [16]
  17. [17]
  18. [18]
    Microscope Alignment for Köhler Illumination - Nikon's MicroscopyU
    This interactive tutorial reviews both the filament and condenser alignment procedures necessary to achieve Köhler illumination.
  19. [19]
    A 6 step guide to Koehler Illumination | Scientifica
    Jan 22, 2018 · Koehler Illumination is a process that provides optimum contrast and resolution by focusing and centring the light path and spreading it evenly over the field ...
  20. [20]
    Koehler Illumination - SPOT Imaging
    Required Microscope Parts. In order to set up a microscope for Koehler illumination, the microscope has to have a vertically adjustable, centerable condenser ...
  21. [21]
    The Automatic Microscope | Nikon's MicroscopyU
    Although the function of Köhler illumination systems is to ensure uniform illumination and to control its coherence, it is also possible to achieve ...
  22. [22]
    [PDF] Chapter 7 Lenses
    Using this test slide, align the microscope for Köhler illumination using the part of the slide with just one coverglass. Bring the field iris into view and ...
  23. [23]
    Reflected Light Microscopy - Zeiss Campus
    The image-forming or field set of conjugate planes in reflected light Köhler illumination consists of the field diaphragm, the specimen surface, and the ...
  24. [24]
    How to Align Köhler Illumination in 6 Simple Steps | Olympus LS
    Feb 3, 2022 · 1. Change to a low magnification objective. Changing to a low magnification objective on your microscope, such as 10X, gives you a wide field of ...Missing: precise | Show results with:precise
  25. [25]
    Microscope Alignment for Köhler Illumination - Zeiss Campus
    The illuminated area of the specimen plane must be at least as large as the field of view for any given objective.Missing: precise | Show results with:precise
  26. [26]
    How to Set Up Koehler Illumination on Your Upright Light Microscope
    Optimize your microscope's performance with our step-by-step guide to setting up Koehler Illumination on upright models.
  27. [27]
    Kohler Illumination Procedure - Light Microscopy Core Facility
    If you want to accurately observe or capture transmitted light images, you will need to establish Kohler illumination each time you change the objective.
  28. [28]
    Education in Microscopy and Digital Imaging - Zeiss Campus
    In Köhler illumination, light emanating from each point of the source should uniformly illuminate the field diaphragm to produce a similarly uniform viewfield.
  29. [29]
    Microscopy Basics | Enhancing Contrast in Transmitted Light
    Simple brightfield imaging, with the microscope properly adjusted for Köhler illumination, provides a limited degree of information about the cell outline, ...Missing: compatibility | Show results with:compatibility
  30. [30]
    Reflected Light DIC Microscopy | Nikon's MicroscopyU
    Differential interference contrast is particularly dependent upon Köhler illumination to ensure that the waves traversing the Nomarski prism are collimated and ...Anatomy of the Reflected Light... · Reflected Light DIC... · Orientation Effects on...
  31. [31]
    Polarized Light Microscopy - Microscope Alignment
    The first step in polarized light microscope alignment is to establish proper Köhler illumination with both the polarizer and analyzer removed from the optical ...
  32. [32]
    Contrast‐enhanced, single‐shot LED array microscopy based on ...
    Aug 22, 2023 · It has been demonstrated that LED array microscopes outperform Köhler illumination microscopes in some applications. ... efficient light sources ...
  33. [33]
    Automated alignment method for coherence-controlled holographic ...
    Oct 28, 2015 · A coherence-controlled holographic microscope (CCHM) was developed particularly for quantitative phase imaging and measurement of live cell ...
  34. [34]
    Optimisation approaches for concurrent transmitted light imaging ...
    Aug 21, 2015 · Various forms of transmitted light microscopy can be achieved in association with confocal microscopy ... The importance of Köhler illumination.Missing: post- | Show results with:post-
  35. [35]
    Superresolution Structured Illumination Microscopy - Zeiss Campus
    Superresolution structured illumination microscopy was the first SIM technique to utilize transversely structured excitation illumination to sharpen lateral ...
  36. [36]
    Optical Microsphere Nano-Imaging: Progress and Challenges
    ... Kohler illumination is used for the surface imaging. Generally speaking, mechanical holders tend to affect the illumination in microsphere nano-imaging ...<|separator|>