Microscopy

Microscopy is the science and technical field of using microscopes to visualize and study objects and specimens that are too small to be seen with the unaided , typically extending observation capabilities to scales ranging from micrometers (10⁻⁶ meters) to nanometers (10⁻⁹ meters). These instruments achieve this by magnifying images through lenses or beams, revealing intricate details in biological cells, microorganisms, materials, and subatomic structures that have profoundly advanced fields like , , and . The origins of microscopy trace back to the , when early compound microscopes were developed in , with significant contributions from scientists such as , who published in 1665 describing observations of cork cells and other microstructures, and Antony van Leeuwenhoek, known as the "father of " for his single-lens microscopes that achieved up to 270× magnification and enabled the first sightings of , , and blood cells in the 1670s and 1680s. Over the subsequent centuries, microscope design evolved dramatically: the 18th and 19th centuries saw improvements in lens quality, including achromatic objectives invented by Chester Moor Hall in 1733 to reduce , and Ernst Abbe's formulation of the resolution limit in 1873—defined as approximately λ / (2 n sin α), where λ is the of , n is the , and α is the half-angle of the —which set the theoretical boundary for at about 0.2 micrometers using visible . Key 19th- and 20th-century advancements included apochromatic lenses by and in 1886 for superior color correction, August Köhler's even illumination system in 1893, and the introduction of electron microscopy in the 1930s, which uses electron beams to achieve resolutions down to 0.2 nanometers, far surpassing light-based limits. Microscopes are broadly classified into light (optical) microscopes, which employ visible or ultraviolet light and lenses for magnification up to about 1,500×, and electron microscopes, which utilize electron beams and electromagnetic lenses for much higher magnifications exceeding 1,000,000× but require vacuum conditions and typically non-living samples. Common types of light microscopy include brightfield, the simplest and most widely used for stained specimens where the sample appears dark against a bright background; darkfield, which illuminates the specimen obliquely to highlight live, unstained microbes as bright objects on a dark field; phase-contrast, developed by Frits Zernike in the 1930s to enhance contrast in transparent, unstained cells by converting phase shifts in light into amplitude differences; and fluorescence microscopy, which excites fluorescent dyes or proteins with specific wavelengths to visualize targeted structures like proteins or DNA in living cells. Electron microscopy variants encompass transmission electron microscopy (TEM) for internal ultrastructure imaging and scanning electron microscopy (SEM) for detailed surface topography. More advanced techniques, such as confocal laser scanning and superresolution methods, have pushed optical limits to around 10 nanometers by exploiting phenomena like stimulated emission depletion. In applications, microscopy is indispensable for diagnosing diseases through cytology and , such as identifying tumors via endomicroscopy; exploring cellular processes in , including cell viability and organelle function; and analyzing material properties in , from semiconductor defects to nanomaterial compositions. It has also facilitated breakthroughs like the discovery of viruses and subcellular components, underscoring its role in foundational scientific progress.

Fundamentals

Principles of Image Formation

Microscopy is the technical field of using microscopes to view objects and details too small to be seen with the naked eye, primarily by achieving magnification to enlarge the image and enhancing contrast to distinguish features within the specimen. This process relies on optical or electronic instruments that collect and focus radiation—such as visible light or electron beams—from the specimen to form a visible image, enabling observation of structures at scales from micrometers down to nanometers. The core goal is to resolve fine details that would otherwise remain invisible, with applications spanning biology, materials science, and medicine. In ray optics, the foundational model for image formation in compound microscopes treats light as rays that are bent by lenses to converge on focal points, creating a real, inverted image of the specimen. The objective lens, positioned close to the specimen, collects divergent rays from the object and forms an intermediate real image within the microscope tube by converging them through refraction, typically providing high magnification (e.g., 4x to 100x) and resolving power. The eyepiece lens, or ocular, acts as a simple magnifier placed at a comfortable viewing distance, further enlarging the intermediate image into a virtual image observed by the eye, resulting in total magnification as the product of the objective and eyepiece powers. This tandem lens system, separated by a fixed tube length (often 160 mm), ensures the final image appears erect and significantly enlarged compared to the specimen. The wave nature of light and electrons introduces diffraction as a fundamental limit to image formation, where wavefronts from the specimen interfere, blurring fine details beyond a certain scale. For light microscopy, Ernst Abbe established in 1873 that resolution is constrained by diffraction, quantified by the equation d = \frac{\lambda}{2 \mathrm{NA}}, where d is the minimum resolvable distance, \lambda is the wavelength of light, and NA is the numerical aperture of the objective lens, reflecting its light-gathering ability (NA = n sin θ, with n as the refractive index of the medium and θ the half-angle of the maximum cone of light). Similarly, electrons exhibit wave-particle duality with a de Broglie wavelength \lambda = \frac{h}{p} (h as Planck's constant and p as momentum), much shorter than visible light (e.g., ~0.005 nm at 100 keV), allowing electron microscopes to achieve superior resolution while still subject to analogous diffraction limits. Contrast in microscopic images arises from differences in how specimens interact with the illuminating beam, primarily through or changes. contrast occurs when the specimen absorbs or scatters (or electrons), reducing in certain areas, as seen in stained biological samples where darker regions indicate higher . contrast, more subtle and common in transparent specimens, results from variations in causing shifts in the phase of the wave without significant change; these shifts must be converted to detectable differences for . Specimen preparation is essential to optimize by enhancing and minimizing artifacts, involving mounting the sample between a and coverslip, often in an immersion medium like (refractive index ~1.518) to match the objective's design and increase for better . with dyes (e.g., hematoxylin for nuclei) selectively binds to cellular components, amplifying by or , while careful fixation preserves without . These steps ensure the specimen is thin, stable, and refractive index-matched to the medium, reducing scattering and enabling clear ray convergence.

Resolution and Magnification Limits

In microscopy, magnification enlarges the apparent size of a specimen, but its effectiveness depends on distinguishing fine details. Optical magnification is the ratio of the image size to the object size, while total magnification M is calculated as the product of the objective lens magnification m_o and the eyepiece magnification m_e, expressed as M = m_o \times m_e. This formula applies to compound microscopes, where the objective forms an intermediate image and the eyepiece magnifies it further. However, magnification must be paired with sufficient ; exceeding the useful range leads to empty magnification, where the image appears larger but blurry, as no additional structural details are revealed due to unresolved diffraction patterns. The useful magnification range is generally 500 to 1000 times the objective's numerical aperture (NA), beyond which empty magnification degrades image quality by introducing artifacts like halos without enhancing . Resolution defines the smallest distance between two points in a specimen that can be distinguished as separate, primarily limited by the of waves passing through the microscope's . The quantifies this limit, specifying the \theta as \theta = \frac{1.22 \lambda}{D}, where \lambda is the of and D is the of the ; in microscopes, this angular measure adapts to linear d by relating D to the , yielding d \approx 0.61 \lambda / \mathrm{NA} for the minimum resolvable distance under optimal conditions. This arises from the overlap of Airy patterns, where two points are just resolvable when the central maximum of one coincides with the first minimum of the other. Several factors influence resolution: the wavelength \lambda, where shorter wavelengths enable finer detail separation; the numerical aperture \mathrm{NA} = n \sin \theta, which captures the light-gathering capacity with n as the medium's refractive index and \theta as the half-angle of the light cone; and the refractive index n, which immersion oils (e.g., n \approx 1.52) increase to exceed air's n = 1.0, thereby boosting NA and resolution. These align with Abbe's diffraction limit, derived in his 1873 seminal work, which established the theoretical minimum resolvable distance as approximately \lambda / (2 \mathrm{NA}). Empty magnification exacerbates resolution limits, as amplifying an already diffraction-blurred image merely enlarges indistinct features, reducing practical utility in specimen analysis. Across modalities, light microscopy resolution is typically around 200 nm, constrained by visible light wavelengths of 400–700 nm. In contrast, electron microscopy achieves resolutions of about 0.1 nm, enabled by the de Broglie wavelength of accelerated electrons (e.g., 0.0037 nm at 100 keV), which is orders of magnitude shorter than light, allowing atomic-scale imaging despite similar diffraction principles.

History

Early Developments

The earliest known attempts at magnification date back to ancient civilizations, where simple devices were employed to enlarge images. In ancient Greece and Rome, glass globes filled with water served as rudimentary magnifying tools, exploiting the refractive properties of water to produce enlarged views of small objects. These devices, while limited in precision, represented an initial understanding of optical principles for visual enhancement. By the medieval period, advancements in lens-making led to the development of reading stones around the 11th century. Monks in Europe crafted these from segments of polished rock crystal or quartz, forming plano-convex lenses that could magnify text on manuscripts by about 2 to 3 times when placed directly on the page. This innovation aided presbyopic scribes in illuminating and copying religious texts, marking a practical application of magnification in daily scholarly work. The invention of the compound microscope in the late transformed magnification capabilities. In the 1590s, Dutch spectacle makers Hans and Zacharias in Middelburg, , assembled the first compound microscope by arranging multiple lenses in a tube, achieving magnifications of around 3 to 9 times through the combination of and lenses. This instrument laid the groundwork for higher-resolution observation by superimposing images from successive lenses. Early 17th-century refinements further propelled microscopy forward. In 1609, Galileo Galilei developed the occhiolino, an improved compound microscope using a convex objective lens and a concave eyepiece, which provided magnifications up to 20 times and allowed for detailed examination of minute structures. Around 1665, Robert Hooke published Micrographia, the first extensively illustrated treatise on microscopic observations, where he examined thin slices of cork and coined the term "cell" to describe the box-like compartments he observed, revealing the cellular structure of plant material. Hooke's work, conducted with a compound microscope of his design achieving up to 50 times magnification, popularized microscopy and demonstrated its potential for biological discovery. A pivotal contribution came from Antony van Leeuwenhoek in the 1670s, who crafted superior simple microscopes using a single high-quality ground from rock crystal. These instruments achieved magnifications of up to 270 times, far surpassing contemporary compound designs, and enabled Leeuwenhoek to observe and describe previously unseen microorganisms, including in 1674 and in lake water samples by 1676. His discoveries, communicated through letters to the Royal Society, established as a field and highlighted the microscope's role in unveiling the microbial world. Despite these innovations, early microscopes suffered significant optical limitations. Compound instruments were plagued by , where light rays passing through different parts of a focused at varying points, causing image blurring, and , which produced color fringing due to the unequal of light wavelengths by glass lenses. These flaws restricted and clarity, often making observations at higher magnifications impractical until later lens improvements.

Key Innovations in the 19th and 20th Centuries

In the 1830s, significant progress in optical microscopy was made through the development of achromatic lenses, which corrected for chromatic and spherical aberrations that had previously limited image clarity. Joseph Jackson Lister, a British wine merchant and amateur optician, demonstrated that combining multiple weak lenses at specific distances could minimize these distortions, publishing his findings in 1830 and collaborating with instrument maker Andrew Ross to produce practical achromatic objectives. This innovation enabled sharper, color-fringe-free images, marking a foundational step toward modern compound microscopes. The 1870s brought theoretical and practical advancements led by , a at , who established the diffraction-based limits of microscopic resolution and collaborated with glassmaker to develop specialized optical glasses for superior lenses. Abbe's 1873 theory quantified how wavelength and determine resolvability, guiding design, while his 1878 invention of homogeneous systems—initially water-based, soon followed by oil—dramatically improved light collection. By the , oil objectives, refined with Schott's flint and crown glasses, achieved magnifications up to approximately 1000× and resolutions around 0.2 μm, approaching the practical limit for visible light microscopy. These lenses filled the space between the and specimen with oil of matching , reducing light scattering and enabling visualization of fine cellular details previously indistinct. The early 20th century saw innovations addressing contrast challenges in transparent specimens, culminating in the 1930s with Frits Zernike's invention of . Zernike, a physicist, developed this technique to exploit phase shifts in light waves passing through unstained, living cells, converting them into amplitude differences for enhanced visibility without dyes that could alter biological structures. His work, first demonstrated in 1934, revolutionized biological imaging and earned him the in 1953. A transformative leap occurred in the 1940s with the advent of electron microscopy, pioneered by , who recognized that electron beams, with wavelengths far shorter than visible light, could surpass optical resolution limits. In 1931, Ruska and Max built the first prototype transmission electron microscope (TEM), refined to a functional model by 1933 using magnetic coils as lenses; the first commercial instrument followed in 1939 from . This earned Ruska the in 1986, shared for his foundational contributions. From the 1950s to 1970s, TEM underwent key refinements in instrumentation and sample preparation, enhancing usability for . Advances included higher-voltage accelerators for better penetration and resolution down to sub-nanometer scales, alongside techniques like metal shadowing in the 1950s for surface relief imaging and in the 1960s–1970s, which used heavy-metal salts to outline macromolecular structures without embedding. Concurrently, emerged from the University Engineering Department under Charles Oatley, with Dennis McMullan constructing the first practical SEM prototype in 1951; by 1965, the group’s work led to the commercial Stereoscan, enabling three-dimensional topographic imaging at resolutions of 50 nm or better. These developments solidified electron microscopy as indispensable for atomic-scale investigations.

Modern and Recent Advances

The development of confocal laser scanning microscopy (CLSM) in the 1980s and 1990s marked a pivotal advancement in optical , building on Marvin Minsky's 1957 patent for the confocal principle. This technique, which uses a pinhole to eliminate out-of-focus light, enabled true three-dimensional (3D) of thick specimens by optical sectioning, revolutionizing fields like . Commercial CLSM systems became available in the mid-1980s, with widespread adoption by the 1990s following improvements in laser technology and detectors. In the 2000s, super-resolution techniques shattered the diffraction limit of light microscopy, achieving resolutions below 200 nm. , invented by and first demonstrated in 2000, employs a doughnut-shaped depletion beam to sharpen the excitation spot, allowing nanoscale visualization of live cells. shared the 2014 for this breakthrough, alongside developments in single-molecule localization methods. Photoactivated localization microscopy (), introduced by Eric Betzig in 2006, and stochastic optical reconstruction microscopy (), developed by in the same year, rely on precise localization of individual fluorophores activated in sparse subsets, enabling resolutions around 20-30 nm. The 2010s witnessed the cryo-electron microscopy (cryo-EM) revolution, transforming by enabling atomic-level imaging of biomolecules in near-native states. Advances in direct electron detectors and computational processing achieved resolutions below 3 Å, allowing visualization of protein structures without crystals. This "resolution revolution" was recognized with the 2017 awarded to , , and Richard Henderson for developing cryo-EM methodologies. From 2020 to 2025, innovations further expanded microscopy's capabilities for complex biological imaging. In 2025, researchers at MIT introduced enhancements to expansion microscopy, incorporating lipid-optimized probes to achieve high-resolution 3D mapping of lipid membranes and protein distributions in cells, expanding physical samples up to 20-fold for nanoscale detail. Yale University's 2024 FLASH-PAINT technique advanced multiplexed super-resolution by using transient DNA adapters and eraser strands for rapid, unlimited cycling of probes, enabling simultaneous visualization of dozens of molecular targets at ~10 nm resolution without photobleaching limitations. In cryo-EM, the Thermo Fisher Scientific Krios G4, installed at UCLA in early 2025, incorporates automated sample loading and improved stability for faster data collection and higher-resolution structures, reducing acquisition times by up to 50% compared to prior models. Artificial intelligence (AI) integration has become a cornerstone of recent microscopy advances, particularly for image reconstruction and . Deep learning algorithms enhance super-resolution by predicting high-fidelity images from low-resolution inputs, improving signal-to-noise ratios in techniques like STED and . In electron microscopy, self-supervised models such as SHINE (2025) denoise raw cryo-EM data in real-time, accelerating high-throughput analysis while preserving structural details at near-atomic scales. These AI-driven methods have automated feature extraction, enabling scalable processing of large datasets from diverse microscopy modalities. The global microscopy market has grown significantly, reaching approximately $8.4 billion in 2025, fueled by automation and AI enhancements that streamline workflows and expand applications in research and industry.

Optical Microscopy

Bright-field and Dark-field Techniques

Bright-field microscopy is the simplest and most common form of optical microscopy, employing transmitted white light to illuminate the specimen directly from below, allowing light to pass through the sample and form an image based on amplitude differences caused by absorption or scattering. This technique produces a bright background with the specimen appearing darker due to variations in density and thickness, making it particularly suitable for observing stained biological samples such as bacteria or tissue sections in histology. Key components include a condenser lens that focuses and evenly distributes light onto the specimen plane, and an iris diaphragm that adjusts the aperture to control illumination intensity and contrast. However, bright-field imaging suffers from low inherent contrast when viewing unstained or transparent specimens, as these materials minimally absorb or scatter light, often rendering fine details invisible without additional preparation. Additionally, basic setups are prone to chromatic aberration, where different wavelengths of light focus at varying points, leading to color fringing and reduced image clarity, though this can be mitigated with achromatic objectives. In contrast, enhances visibility of low-contrast specimens by using oblique illumination to block the direct central beam, resulting in a dark background with the specimen appearing bright against it due to scattered or diffracted entering the objective. This method excels at highlighting edges and structures in live, unstained cells, nanoparticles, and colloidal particles by capitalizing on deflection at boundaries. Specialized setups employ or cardioid condensers to generate a hollow cone of high-angle ; the uses a reflective surface for numerical apertures up to 1.40, while the cardioid relies on mirrored internals for aberration-free illumination up to 1.30, often requiring immersion oil for optimal performance. Applications include the detection of spirochetes, such as in diagnosis, where the technique reveals motile bacteria invisible in bright-field, as well as imaging submicron particles in suspensions. The technique of was developed in the early to observe unstained microbes and overcome the contrast limitations of early transmitted systems.

Phase-contrast and Differential Interference Contrast

Phase-contrast microscopy, developed by Frits Zernike in , enhances the visibility of transparent specimens by converting phase shifts in light caused by variations in into detectable differences. When light passes through a specimen such as a live , regions with higher retard the light waves by a small phase amount, typically on the order of π/4 or less, which remains invisible in standard bright-field imaging. Zernike's method introduces a deliberate phase shift of π/2 (90 degrees) to the undiffracted direct light relative to the diffracted light from the specimen, amplifying these subtle phase differences into intensity variations through . This technique particularly highlights gradients, making structures like membranes and organelles appear as brighter or darker regions against the background. The setup for phase-contrast microscopy involves an annular diaphragm in the condenser that produces a hollow cone of illumination, ensuring the direct light passes through a transparent phase ring in the objective's rear focal plane. This phase ring, typically made of a dielectric material like magnesium fluoride, either retards (positive contrast) or advances (negative contrast) the direct light by π/2 while also slightly attenuating its intensity to balance amplitudes for optimal interference. In positive mode, phase-advanced regions (e.g., protein-rich areas in cells) appear darker, mimicking absorption, whereas negative mode reverses this to make them brighter, which can be useful for visualizing dense structures like sperm heads. A key limitation is the appearance of halos—bright or dark fringes—around phase objects due to incomplete phase cancellation, which can obscure fine details in thicker specimens. Applications of are prominent in for observing unstained, living cells in , where it reveals dynamic processes such as and organelle movement without the artifacts of fixation or staining. For instance, it enables clear visualization of nuclei, mitochondria, and cytoskeletal elements in mammalian cells, providing insights into cellular and in real time. Differential interference contrast (DIC) microscopy, pioneered by Georges Nomarski in the , builds on interferometric principles to produce pseudo-three-dimensional images of transparent specimens by exploiting local gradients. Nomarski's approach splits a polarized into two orthogonally polarized, closely sheared wavefronts using birefringent prisms; as these beams pass through the specimen at slightly offset positions (shear distance typically 0.1–1.5 µm), they acquire differential shifts proportional to the gradient, which are then recombined to generate patterns manifesting as directional shadows and highlights. This creates a relief-like, 3D appearance that emphasizes edges and surface , particularly effective for revealing subtle structures in live cells and tissues. The core components of a DIC system include a linear before the , a Wollaston or modified Nomarski in the to generate the sheared beams, a matching in , and an analyzer oriented at 90 degrees to the for recombination. Nomarski , which position the interference plane outside the body, allow greater flexibility in placement and reduce spatial constraints compared to traditional Wollaston . DIC achieves higher lateral and axial resolution than phase-contrast by utilizing the full of without restricting illumination, avoiding halo artifacts and providing sharper images of fine details like in unstained cells. However, its setup demands precise alignment of components and , and it is highly sensitive to vibrations and specimen thickness variations, which can introduce directional bias or pseudorelief effects.

Fluorescence and Confocal Methods

Fluorescence microscopy relies on the excitation of fluorophores—molecules that absorb light at short wavelengths and emit it at longer wavelengths—enabling the visualization of specific cellular components with high contrast. This process exploits the , the energy difference between absorption and emission spectra, which allows emitted light to be separated from excitation light using optical filters. The technique is particularly valuable for label-based imaging, where fluorophores such as fluorescein or are conjugated to antibodies or other probes to target molecules of interest. A common implementation is wide-field epifluorescence microscopy, where excitation light is directed through the objective lens onto the sample, and emitted fluorescence is collected from the entire . Key components include dichroic mirrors, which reflect shorter-wavelength excitation light while transmitting longer-wavelength emission, and bandpass filters that selectively pass excitation and emission wavelengths to minimize background noise. This setup is widely used in applications like , where fluorophore-labeled antibodies bind to specific proteins, allowing researchers to map their localization in fixed cells or tissues with subcellular precision. Confocal microscopy advances fluorescence imaging by incorporating a pinhole in the detection path to block out-of-focus light, producing sharp optical sections suitable for three-dimensional reconstruction via z-stacks—series of images taken at incremental focal depths. systems direct a focused beam across the sample using (galvo) mirrors for precise raster scanning, while (PMT) detectors amplify the weak signals for high-sensitivity detection. For faster imaging, tandem scanning confocal microscopes employ a spinning with thousands of pinholes to illuminate and detect multiple points simultaneously, enabling video-rate acquisition of dynamic processes. This optical sectioning improves axial resolution to approximately 0.5–1 μm, surpassing wide-field methods. Despite these advantages, fluorescence microscopy faces limitations such as , where prolonged irreversibly degrades fluorophores, reducing signal over time, and potential from dyes that can harm living cells. Recent advancements in (LED) sources have addressed some challenges by providing stable, narrow-band at lower costs and reduced heat output compared to traditional mercury lamps, facilitating wider accessibility for routine imaging. A variant, multiphoton microscopy, uses lasers to simultaneously deliver two or more photons for , minimizing and enabling deeper —up to several millimeters—while confining to the focal to limit and .

Super-resolution Techniques

Super-resolution techniques in optical microscopy overcome the diffraction limit of approximately 200 nm by exploiting the nonlinear properties of fluorescence emission or precise localization of individual fluorophores, enabling imaging at scales of 10-100 nm. These methods primarily manipulate fluorescent labels to achieve resolutions far beyond conventional wide-field or confocal approaches, revealing subcellular structures such as synaptic proteins or membrane dynamics. Developed in the early , they have transformed biological imaging by providing molecular-scale insights without the need for electron microscopy. Stimulated Emission Depletion (STED) microscopy employs a doughnut-shaped depletion beam to suppress fluorescence emission around the excitation focus, effectively shrinking the point spread function and achieving resolutions of 20-50 nm in living cells. Introduced by and Jan Wichmann in , STED uses a high-intensity depletion tuned to the fluorophore's emission wavelength to de-excite molecules via , confining emission to a central region smaller than the limit. This technique has been widely adopted for imaging dynamic processes, such as release in neurons, due to its compatibility with standard confocal setups. Localization methods, including Photoactivated Localization Microscopy (PALM) and Stochastic Optical Reconstruction Microscopy (STORM), attain resolutions of 10-20 nm by sequentially activating and localizing sparse subsets of photoswitchable fluorophores, followed by computational reconstruction of their positions. , developed by Betzig and colleagues in 2006, relies on photoactivatable fluorescent proteins that are stochastically turned on, imaged, and photobleached, allowing precise fitting of their point spread functions to determine sub-pixel locations. Similarly, , introduced by Michael Rust, Mark Bates, and in 2006, uses organic dyes that can be switched on and off repeatedly, enabling dense labeling and high-fidelity reconstructions of structures like clathrin-coated pits. These approaches excel in fixed samples but require thousands of frames for full images. Structured Illumination Microscopy () enhances to about 100 nm by illuminating the sample with a patterned grid, which interferes with high-frequency sample to extend the observable Fourier space, followed by computational . Pioneered by in 2005, linear SIM doubles the conventional by capturing shifted illumination patterns and reconstructing the image via Fourier , while nonlinear variants can theoretically achieve unlimited through saturation effects. SIM variants, such as 3D-SIM, are particularly useful for live-cell imaging of cytoskeletal elements due to their relatively low requirements compared to other super-resolution methods. Advancements in STED implementation continue, with confocal core facilities like that at the University of Maryland School of Medicine offering state-of-the-art STED systems in 2025 for routine subcellular imaging, such as membrane proteins, expanding access to super-resolution for biomedical researchers. Despite their power, super-resolution techniques face challenges including high light doses that induce and in live samples, as well as substantial computational demands for image reconstruction in localization and methods. These limitations often necessitate optimized fluorophores or adaptive illumination to balance with sample viability.

Electron Microscopy

Transmission electron microscopy (TEM) is a that utilizes a beam of electrons transmitted through an ultrathin specimen to produce high-resolution images of internal structures. The principle relies on the interaction of electrons with the sample via elastic and , where unscattered or minimally scattered electrons form the image, providing contrast based on mass-thickness differences and . Electrons accelerated to high voltages, typically 100–300 , exhibit a de Broglie of approximately 0.0037 at 100 , enabling resolutions far superior to optical microscopy due to this short ./08%3A_Structure_at_the_Nano_Scale/8.02%3A_Transmission_Electron_Microscopy) The TEM setup consists of an , usually a thermionic or field emission source, that generates the beam, followed by a series of electromagnetic lenses to condense, focus, and project the beam through the specimen. The transmitted s are then magnified by objective and projector lenses and detected on a fluorescent screen, , or modern () camera for . Specimens must be prepared as ultrathin sections less than 100 nm thick to allow sufficient without excessive ; this is achieved through , where embedded samples are sectioned using a diamond knife. Contrast is enhanced by heavy metal staining, such as for during fixation or and lead citrate post-sectioning, which scatter s differentially based on . TEM achieves resolutions down to 0.1 , allowing visualization of details in crystalline materials, with magnifications ranging from 10^5 to 10^6 times. In bright-field mode, the image forms from directly transmitted electrons, highlighting mass-thickness contrast for overall and density variations. High-resolution TEM (HRTEM), or phase-contrast mode, enables imaging by interfering diffracted and transmitted beams to reveal arrangements and defects in periodic structures. Despite its capabilities, TEM has limitations including the need for high-vacuum environments to prevent by air molecules, which restricts live imaging. Radiation damage from the electron beam can alter or destroy sensitive specimens, particularly biological or materials, limiting dose to avoid structural changes. Additionally, TEM produces 2D projections of the specimen, necessitating techniques like to reconstruct three-dimensional structures.

Scanning Electron Microscopy

Scanning electron microscopy () employs a focused beam of that is raster-scanned across the surface of a specimen to generate high-resolution images revealing surface and . The principle relies on the of the incident electron beam with the sample, producing , backscattered electrons, or characteristic X-rays that are detected to form contrast in the image. This technique provides three-dimensional-like visualization due to its large , which is approximately 100 to 300 times greater than that of optical microscopy, enabling detailed examination of surface features at magnifications from 10x to over 300,000x. The SEM setup consists of an electron column housing an electron source, such as a thermionic tungsten filament, (LaB6) gun, or (FEG) for higher resolution, along with electromagnetic lenses to focus the beam to a spot size of 1-10 nm, deflection coils for scanning, and a sample chamber maintained at high (typically 10^{-5} to 10^{-7} ) to prevent . Modern SEMs often incorporate variable pressure modes, allowing operation at up to 200 for certain applications. Specimen is crucial for optimal ; non-conductive samples are coated with a thin layer of , , or carbon (5-20 nm thick) via to enhance conductivity and prevent charging under the beam. For biological or hydrated specimens, preparation involves chemical fixation, through a graded series, and critical point drying using supercritical CO2 to avoid surface tension-induced collapse during solvent evaporation. Resolution in SEM typically ranges from 1 to 10 nm, depending on the type and accelerating voltage (0.5-30 kV), with FEG-SEM achieving sub-nanometer performance for surface details. modes include secondary (SE) detection for topographic contrast, as these low-energy s (50 eV to 50 keV) escape from near the surface; backscattered () imaging for compositional information, where yield correlates with ; and () for elemental mapping by analyzing emitted X-rays. Environmental SEM (ESEM), developed in the , modifies the system with differential pumping and gaseous secondary detectors to image hydrated or wet samples at low (up to 2000 Pa water vapor pressure), preserving natural states without extensive .

Cryo-electron Microscopy

Cryo-electron microscopy (cryo-EM) preserves the native structure of biological specimens by rapidly freezing them in vitreous ice, avoiding the formation of damaging ice crystals, and imaging at cryogenic temperatures around -180°C to minimize beam-induced damage.31325-9) This technique, pioneered by Dubochet's development of methods in the , traps biomolecules in a non-crystalline, amorphous state that mimics their hydrated environment in solution. This work, along with contributions from and Richard Henderson, earned them the 2017 for developing cryo-EM for the determination of high-resolution structures of biomolecules in solution. By maintaining samples in a frozen-hydrated state without stains or fixatives, cryo-EM enables visualization of macromolecular complexes at near-native conditions, distinguishing it from traditional electron microscopy approaches.01078-9) The primary methods in cryo-EM include cryo-transmission electron microscopy (cryo-TEM) for single-particle analysis and cryo-electron (cryo-ET) for three-dimensional cellular imaging. In single-particle cryo-TEM, developed through Joachim Frank's algorithmic advancements in the 1970s and 1980s, thousands of two-dimensional projections of individual macromolecules are computationally aligned and reconstructed into high-resolution three-dimensional structures. Cryo-ET, extending these principles to thicker specimens, acquires tilt series of images from cellular sections to generate tomograms that reveal the spatial organization of proteins within their native cellular context. Both methods rely on low-dose imaging to prevent specimen degradation, leveraging the phase contrast from unstained samples embedded in thin ice layers.01078-9.pdf) The standard workflow begins with sample preparation via plunge-freezing, where a small of purified protein solution (typically 3-5 μL) is applied to a holey carbon , blotted to form a thin film, and rapidly vitrified by plunging into cooled to approximately -180°C, achieving cooling rates exceeding 10^5 K/s to form vitreous . The frozen is then transferred under to a cryo-TEM instrument equipped with direct electron detectors, such as the or K3 models, which capture high-frame-rate movies to correct for beam-induced motion and enable for improved signal quality. Data collection involves automated acquisition of micrograph movies at low electron doses (20-50 e/Ų), followed by pipelines that include motion correction, contrast transfer function estimation, and particle picking. Resolutions in cryo-EM typically range from 1-4 Å for well-behaved proteins, allowing atomic model building and visualization of side-chain details, as demonstrated in structures like the ribosome at 2 Å. Recent advances, including the Thermo Fisher Scientific Titan Krios G4 microscope installed at institutions like UCLA in 2025, have enhanced automation and stability, enabling significantly faster data collection and sub-2 Å resolutions for challenging samples. These developments, combined with direct detector technologies, have reduced data acquisition times from days to hours, broadening accessibility for structural biology. Cryo-EM has revolutionized applications in determining structures of viruses, such as spike proteins at near-atomic resolution, and membrane proteins embedded in nanodiscs, revealing conformational dynamics inaccessible to crystallography.00331-3) Software like RELION, an open-source pipeline for single-particle reconstruction, facilitates Bayesian classification and refinement, enabling the sorting of heterogeneous states from datasets exceeding 1 million particles. These capabilities have accelerated , including inhibitor design for viral entry mechanisms and functions. Major challenges in cryo-EM include the inherently low (SNR) in micrographs, arising from the weak of electrons by light atoms in biological samples and the need for low-dose to avoid , often requiring averaging over tens of thousands to millions of particle images for sufficient contrast. Preferred orientations and conformational heterogeneity further complicate , necessitating advanced algorithms to discard suboptimal particles and resolve subtle structural variations. Ongoing efforts focus on improving sample uniformity and detector sensitivity to mitigate these issues.

Scanning Probe Microscopy

Atomic Force Microscopy

Atomic force microscopy (AFM) is a scanning probe technique that images and manipulates surfaces at the nanoscale by measuring forces between a sharp probe tip and the sample. Invented in 1986, it extends the principles of the to non-conductive materials by detecting mechanical interactions rather than electrical currents. The core component is a microfabricated with a sharp tip, typically made of silicon or , that scans over the sample surface while a laser beam reflects off the cantilever's back onto a position-sensitive to detect deflections caused by tip-sample forces, such as van der Waals or electrostatic interactions. The experimental setup includes a piezoelectric scanner that precisely positions the sample or tip in three dimensions using voltage-controlled expansion of piezoelectric materials, enabling raster scanning with sub-nanometer precision. A feedback loop maintains constant interaction parameters—such as force, amplitude, or frequency—by adjusting the scanner's z-position in real time, generating topographic maps from the deflection data. AFM operates in multiple modes to suit different samples: in contact mode, the tip maintains constant contact with the surface under a set force (typically a few nanonewtons), suitable for rigid materials but potentially damaging to soft ones; tapping mode oscillates the cantilever near its resonance frequency, allowing intermittent contact that reduces lateral forces and preserves delicate structures like biomolecules; non-contact mode keeps the tip above the surface, detecting attractive forces through frequency shifts for ultra-high vacuum or atomic-scale imaging. AFM achieves lateral resolution below 1 nm and vertical resolution down to the atomic scale (approximately 0.1 nm), with force sensitivity on the order of piconewtons, enabling detection of subtle surface features and molecular interactions. Applications include measuring on materials like thin films, where root-mean-square deviations as low as 0.1 nm can be quantified; studying biomolecular folding through single-molecule force , as demonstrated in the reversible unfolding of immunoglobulin domains under controlled tension; and , such as , where the AFM tip delivers ink molecules to create patterns with 30 nm resolution for fabricating nanostructures. Compared to other microscopy techniques, AFM's resolution surpasses optical limits while operating in ambient or environments without requiring . Despite its capabilities, AFM has limitations, including relatively slow scan speeds (typically micrometers per second) due to the mechanical feedback and dynamics, which can take minutes to hours for large areas. Tip artifacts, such as broadening from the tip's finite radius (5–10 ) or contamination, can distort images, and the technique is generally restricted to sample sizes up to about 100 μm laterally because of travel limits. Additionally, while versatile for insulators and biomolecules, quantitative measurements require careful to account for spring constants and .

Scanning Tunneling Microscopy

is a scanning probe technique that images conductive surfaces at the atomic scale by exploiting quantum tunneling of electrons. Invented in 1981 by and at Zurich Research Laboratory, the first successful observation of the exponential dependence of the tunneling current on tip-sample separation was achieved on March 16, 1981. The first atomic-resolution topographic images were obtained later in 1981 on a CaIrSn₄ surface. Their pioneering work, published in 1982, demonstrated topographic pictures of surfaces on an atomic scale and earned them the in 1986, shared with for contributions to electron microscopy. The principle of STM relies on the quantum mechanical tunneling effect, where electrons from a sharp metallic tip tunnel through the vacuum barrier to the sample surface when biased with a small voltage (typically 0.1–1 V). The resulting tunneling current I is highly sensitive to the tip-sample separation d, following the approximate relation I \propto e^{-2 \kappa d}, where \kappa = \sqrt{2 m \phi}/\hbar, m is the electron mass, \phi is the average work function of the tip and sample, and \hbar is the reduced Planck's constant. This exponential dependence—decreasing by about a factor of 10 per angstrom—increases the effective resolution, as small changes in distance produce large current variations. The current also depends on the local density of states near the Fermi level, allowing STM to probe both topography and electronic structure. The typical setup employs a sharp or platinum-iridium tip mounted on piezoelectric transducers for precise three-dimensional positioning with sub-angstrom accuracy. Operations occur in (typically <10^{-10} ) to minimize contamination and adsorption, often at low temperatures (4–77 K) using or for thermal stability and reduced thermal drift. is essential, achieved through springs, damping, or superconducting to suppress mechanical noise below 10^{-12} m/√Hz. A feedback electronics system monitors the tunneling current and controls the tip position accordingly. STM achieves lateral resolution of approximately 0.1 nm and vertical resolution of 0.01 nm, enabling visualization of individual atoms and lattice defects. This atomic-scale capability was first demonstrated on clean metal surfaces like Au(110), resolving the (1×2) missing-row . Two primary modes are used: constant-current mode, where a feedback loop adjusts the tip height to maintain a setpoint current, generating a from the height variations; and constant-height mode, where the tip scans at a fixed separation while recording current fluctuations for faster on flat surfaces. The former is standard for rough or varied topographies, while the latter suits atomically flat samples but risks crashes if features are abrupt. Applications of STM include detailed studies of surface reconstruction, such as the on Au(111) or the complex 7×7 structure on Si(111), which revealed atomic arrangements previously inferred only from data. It also maps adsorbate positions on metal surfaces, aiding research, and supports molecular electronics by imaging self-assembled monolayers and single-molecule conductance. In , STM has enabled atomic manipulation, such as positioning atoms to spell words on surfaces. Limitations of STM include its restriction to electrically conductive samples, such as metals or doped semiconductors, as insulators prevent sufficient tunneling current. Tip contamination or dulling from sample atoms can distort images, requiring frequent tip preparation, and the technique demands and cryogenic conditions for optimal performance, limiting in-situ studies.

Other Modalities

X-ray Microscopy

X-ray microscopy employs to achieve high-resolution of specimens, exploiting their short wavelengths and ability to penetrate dense or thick materials that are challenging for visible or microscopy. The technique relies on the principles of X-ray and , where X-rays interact with primarily through photoelectric , producing contrast based on differences in and . Wavelengths typically range from 0.6 nm to 10 nm, corresponding to energies of approximately 100 eV to 2 keV (soft X-rays), which enable in the "water window" (2.28–4.36 nm) for biological samples due to natural contrast between water and organic materials. Common types include full-field (TXM), analogous to but using for projection imaging, and scanning (STXM), which raster-scans a focused beam across the sample. In both, Fresnel zone plates serve as key focusing elements, diffracting to form images on a detector. The setup generally comprises an source—either laboratory-based (e.g., laser-plasma) or for higher brightness—and like zone plates with outermost zone widths of 25–50 nm, often paired with a condenser and scintillator detector. sources provide coherent, high-flux beams essential for advanced imaging, though laboratory systems enable more accessible setups. Resolution in microscopy typically reaches 10–50 nm laterally, with synchrotron-based systems achieving sub-30 nm through optimized and partial illumination; for instance, full-field TXM has demonstrated 50 nm without using total-reflection mirrors. The is around 1.5–23 μm depending on energy, supporting 3D via sample tilting up to ±60–79°. Brighter beams enhance signal-to-noise ratios (100–300) and enable finer details, surpassing limits. Applications span and , such as 3D of fossils at submicrometer to reveal embryonic structures without destruction, and in operando of cathodes to map chemical states and in lithium-ion systems. Phase-contrast variants, leveraging rather than , excel for soft tissues, providing high-contrast views of cellular ultrastructures like mitochondrial cristae or engineered tissues at low doses. These methods support chemical via near-edge structure (XANES) analysis. Limitations include , which can degrade biological samples at doses exceeding 10¹⁰ despite cryogenic mitigation, and the need for specialized facilities like synchrotrons, restricting accessibility and increasing operational complexity. Sample thickness is constrained to about two absorption lengths to maintain contrast, and tilt range limitations in can reduce z-resolution. environments and issues further challenge routine use. As of 2025, recent advances include methods for enhanced reconstruction and noise reduction in images, practical dark-field imaging for detecting tiny defects, and MEMS-based scanning for speeds up to several hundred kHz, improving throughput and resolution.

Infrared and Ultraviolet Microscopy

Ultraviolet microscopy utilizes wavelengths in the range of 200-400 to achieve higher compared to visible light microscopy, typically approaching 100 due to the shorter wavelength in the diffraction limit. Special , such as or fused silica lenses, are required because standard absorbs UV light, preventing transmission below approximately 350 . This enables detailed of structures invisible in visible light, such as protein distributions in cells. Key applications include biological analysis through UV-absorbing dyes, like those used for DNA staining, where fluorochromes such as DAPI absorb UV to visualize nucleic acids without additional labeling in some setups. In semiconductor manufacturing, UV microscopy supports photolithography by inspecting photoresist patterns and wafer defects at sub-micrometer scales during UV exposure processes. However, limitations arise from UV's high photon energy, which can cause photodamage to biological samples through mechanisms like DNA cross-linking and reactive oxygen species generation, often restricting exposure times. Additionally, UV light exhibits limited penetration depth in scattering media, such as tissues, confining imaging to surface or thin sections. Infrared microscopy operates across wavelengths from approximately 700 nm to 1 mm, with mid-infrared (2.5-25 μm) particularly suited for probing molecular vibrations through absorption spectroscopy. Integration with Fourier transform infrared (FTIR) spectrometers allows for chemical mapping by collecting spatially resolved spectra, revealing compositional variations at diffraction-limited resolutions of 5-10 μm. However, recent super-resolution techniques as of 2025, such as optical photothermal infrared (O-PTIR) spectroscopy and metasurface-enhanced methods, have achieved sub-micrometer to nanometer resolutions, enabling detailed chemical mapping in live cells and tissues. Applications encompass material science, such as identification, where spectra distinguish functional groups like C-H stretches in versus carbonyls in polyesters via or modes. In biomedical contexts, it aids by analyzing bands, for instance, the amide I band at 1650 cm⁻¹ corresponding to protein secondary structures, enabling differentiation of healthy versus diseased states like cancerous cells. Instrumentation typically employs IR-transparent windows, such as (KBr), for sample mounting, along with sensitive detectors like (MCT) for broadband detection from 4000 to 700 cm⁻¹. For opaque samples, modes, including (ATR), facilitate analysis without sectioning by measuring surface interactions.

Emerging Techniques

Photoacoustic Microscopy

Photoacoustic microscopy () is a biomedical imaging modality that leverages the to visualize optical absorption contrasts with high and depth penetration beyond traditional optical microscopy. This technique combines the high contrast of optical imaging with the superior propagation properties of , enabling non-invasive, label-free visualization of endogenous absorbers such as and in biological tissues. Pioneered in the early , PAM has evolved into a versatile tool for multiscale imaging, from cellular structures to whole organs, as detailed in foundational work on multiscale photoacoustic systems. The core principle of involves illuminating tissue with short-pulsed light, typically pulses, which is absorbed by chromophores, causing localized thermoelastic expansion and the emission of ultrasonic waves. These photoacoustic () waves are detected by ultrasonic transducers and reconstructed into images representing the of optical . The initial rise generated is proportional to the absorbed optical , governed by the equation p_0 = \Gamma \mu_a F, where \Gamma is the , \mu_a is the absorption coefficient, and F is the laser fluence. This mechanism allows to provide molecular specificity without exogenous labels, distinguishing it from purely optical methods. Typical setups employ a tunable near-infrared (wavelengths 500–1000 nm) for deep tissue penetration, an optical focusing system (e.g., microscope objective with up to 0.44), and single-element or array-based ultrasonic transducers operating at 20–50 MHz for signal detection. An optical-acoustic beam combiner, often using index-matching media like , co-aligns the excitation light and detection path to enable confocal-like scanning. Real-time imaging is facilitated by fast scanners or voice-coil mechanisms, achieving frame rates up to 20 Hz for dynamic processes like blood flow. PAM operates in two primary modes to balance and . In optical- PAM (OR-PAM), a tightly focused (spot size ~1–5 µm) determines the lateral (down to 0.22 µm), while acoustic detection uses diffuse , limiting depth to about 1 mm in scattering tissues. Conversely, acoustic- PAM (AR-PAM) employs diffuse optical illumination with ( width ~45 µm at 5 MHz), yielding lateral resolutions of tens of microns and depths up to 3–4 mm, suitable for larger vascular networks. Axial in both modes is typically 10–15 µm, set by the . Key advantages of PAM include its label-free nature, relying on strong endogenous contrasts from (absorption peaks at 532 nm and 760 nm for oxygenated and deoxygenated forms, respectively), which enables quantitative mapping of blood oxygenation (sO₂) with sensitivities down to 1% changes. Unlike pure optical techniques limited by scattering to ~1 mm depths, PAM achieves millimeter-scale penetration with minimal acoustic attenuation, offering a over 50 dB. In the , multimodal integrations with (OCT) have enhanced capabilities by combining PA's functional absorption data with OCT's structural scattering information, as demonstrated in systems for and . Applications of PAM span vascular imaging, where it resolves microvasculature down to capillaries in mouse models, providing non-invasive without iodinated contrasts. In , it facilitates detection by imaging melanin-rich tumors up to 4 mm deep, aiding in margin assessment during . For , PAM monitors brain , mapping cerebral blood flow and oxygenation responses to stimuli with sub-millisecond . These uses highlight PAM's role in preclinical research, with emerging clinical translations in and .

Expansion and Holographic Microscopy

Expansion microscopy (ExM) is a sample preparation technique that physically enlarges biological specimens to overcome the diffraction limit of conventional light microscopy, achieving effective resolutions down to approximately 50 nm by expanding samples isotropically by factors of 4 to 20 times. The process involves embedding fixed samples in a swellable hydrogel matrix, anchoring biomolecules such as proteins or nucleic acids to the polymer network, digesting the extracellular matrix and other structural components with proteinase K to reduce linkages, and then swelling the gel in water to physically separate labeled structures. This expansion allows imaging of the enlarged specimen using standard diffraction-limited microscopes, as the physical magnification decrowds molecules and enhances resolution without requiring specialized optics. A key variant, protein-retention expansion microscopy (proExM), preserves native protein structures by anchoring them directly to the using methacryloyl groups, enabling the use of conventional fluorescent antibodies or proteins post-expansion while maintaining compatibility with protocols. Recent 2025 advances from have extended proExM-like approaches to preserve membranes through ultrastructural membrane expansion microscopy (umExM), which incorporates innovative labels and optimized protocols to visualize boundaries and dynamics at nanoscale . Complementing this, multiplexed expansion revealing (multiExR) techniques now allow simultaneous imaging of over 20 protein sets in expanded samples, facilitating detailed mapping and analysis of molecular interactions, such as in neurodegenerative diseases. These developments enable isotropic expansion assumptions to hold for complex tissues, though challenges remain in ensuring uniform swelling across heterogeneous samples. Digital holographic microscopy (DHM) is a label-free interferometric technique that records the interference pattern between scattered object waves from a specimen and a reference wave, enabling numerical reconstruction of both amplitude and quantitative phase images for 3D visualization without physical sectioning. Pioneered in the late , DHM uses a coherent source to illuminate the sample, with the resulting hologram captured by a (CCD) camera; computational algorithms then back-propagate the to retrieve phase shifts, which quantify differences related to and thickness in transparent objects like live cells. Common implementations include off-axis holography, which records a single hologram tilted relative to the reference beam for instantaneous , and phase-shifting methods that acquire multiple images by modulating the reference phase for higher accuracy but requiring sequential exposures. In applications, DHM excels in quantitative phase imaging of living cells, providing non-invasive measurements of dry mass, , and without dyes or bleaching, as well as autofocus-free tracking of dynamic processes like or . For instance, it has been used to monitor morphological changes in cells and neuronal activity in , offering sub-micron axial through algorithms. However, DHM setups demand stable coherent illumination and high computational resources for hologram , which can introduce twin-image in off-axis configurations or sensitivity to vibrations in phase-shifting setups, limiting throughput for large fields of view.

Applications

Biological and Medical Uses

Microscopy plays a pivotal role in by enabling the visualization of subcellular structures such as organelles and , often through fluorescence-based techniques that highlight specific molecular components with high specificity.00239-8) For instance, microscopy allows researchers to track dynamic interactions between organelles like mitochondria and the , revealing how these elements coordinate cellular processes at the nanoscale.31308-4) Live-cell , a cornerstone application, facilitates real-time observation of events such as , where time-lapse microscopy captures segregation and without disrupting cellular viability.01227-8) In , microscopy remains essential for through the examination of slides, where light microscopy identifies malignant features in tissue sections stained with hematoxylin and . scanners employing whole-slide imaging have revolutionized this process by digitizing entire glass slides into high-resolution images, enabling remote consultation and quantitative analysis for improved diagnostic accuracy in . These systems support primary comparable to traditional light microscopy, as validated in multicenter studies. Neuroscience benefits from advanced microscopy modalities to map neural circuits and protein structures. provides optical sectioning to visualize three-dimensional neural circuits, allowing delineation of synaptic connections in tissue with reduced out-of-focus light. Cryo-electron microscopy (cryo-EM) has elucidated the atomic structure of synaptic proteins, such as receptors, offering insights into synaptic transmission and neurological disorders. Recent advancements in have enhanced the study of immune cell dynamics, enabling observation of molecular rearrangements during immune responses, such as T-cell activation at the . A notable 2024 development, FLASH-PAINT, introduces transient adapters for unlimited multiplexed , applied to detect multiple tumor markers in single cells for precise cancer profiling.00236-8) Clinical applications include endomicroscopy probes, which deliver imaging of the during , providing subcellular resolution for real-time detection of or without tissue . These tools, often based on confocal laser endomicroscopy, guide therapeutic interventions like polypectomy. Despite these advances, challenges persist in biological microscopy, particularly minimizing phototoxicity in live-cell imaging, where prolonged light exposure can induce cellular damage and alter physiological behaviors. Integrating for automated analysis addresses this by optimizing imaging parameters in real-time and accelerating data interpretation from large datasets, enhancing throughput in .

Materials and Forensic Science

In materials science, microscopy techniques enable the detailed characterization of material structures at the nanoscale, facilitating the understanding of properties such as strength, conductivity, and reactivity. Transmission electron microscopy (TEM), for instance, provides atomic-resolution imaging and spectroscopy to analyze defects, interfaces, and compositions in advanced materials like semiconductors and alloys. A seminal application involves the 3D atomic-scale reconstruction of nanoparticles, as demonstrated in early work using aberration-corrected scanning TEM (STEM) to map atomic positions with picometer precision, revealing strain distributions that influence material performance. Similarly, scanning electron microscopy (SEM) excels in visualizing surface morphology and topography, such as the agglomeration patterns in carbon quantum dots or the spherical shapes of metal nanoparticles ranging from 40 to 80 nm, aiding in quality control for nanomaterials production. Atomic force microscopy (AFM) complements these by offering non-destructive, three-dimensional surface profiling with sub-nanometer resolution, crucial for assessing roughness and thickness in thin films like graphene oxide, where values of 4.85–11.85 nm have been measured to correlate with mechanical properties. In battery research, electron diffraction tomography via TEM has elucidated crystal structures in cathode materials such as Li₂CoPO₄F, identifying phase transitions that impact efficiency. These techniques collectively drive innovations in , where TEM's high-resolution elemental mapping, for example, distinguishes core-shell distributions in bimetallic nanocrystals, informing catalytic applications. Transitioning to forensic science, microscopy serves as a cornerstone for trace evidence analysis, adhering to the Locard exchange principle that every contact leaves detectable residues. Light microscopy, including stereomicroscopes and comparison microscopes, is routinely employed to examine fibers, hairs, paint chips, and toolmarks, enabling side-by-side comparisons of bullets or fracture patterns in glass fragments to link suspects to scenes. For instance, identifies birefringent properties in synthetic fibers or minerals, distinguishing natural from manufactured materials with high specificity. Electron microscopy enhances forensic precision, particularly scanning electron microscopy coupled with energy-dispersive X-ray spectroscopy (SEM-EDX) for (GSR) detection, where characteristic particles containing lead, , and —often 1–10 μm in size—are identified to determine firing distances or weapon involvement. In one study, detected 447 GSR particles across eight samples, outperforming traditional chemical tests like sodium rhodizonate, which identified only 11. (TEM) further analyzes biological traces, such as myocardial ultrastructure in suspicious deaths, while (AFM) maps nanoscale degradation in latent fingerprints, revealing ridge detail evolution over 28 days on substrates and aiding individualization. These methods, integrated with techniques like , bolster evidence reliability in cases involving soil, tapes, or documents.

References

  1. [1]
    Microscopy - Medical Imaging Systems - NCBI Bookshelf - NIH
    Microscopes are the tools which enable us to extend our vision to the micro-world and, despite the prefix micro- in the name, to the nano-world, too.
  2. [2]
    The Microscope - University of Hawaii System
    Microscopes are designated as either light microscopes or electron microscopes. The former use visible light or ultraviolet rays to illuminate specimens.
  3. [3]
    Microscopy Basics | Historical Perspective - Zeiss Campus
    Microscope design and construction has significantly evolved over the past 350 years leading to advanced models that are capable of superresolution imaging.
  4. [4]
    Microscopy - an overview | ScienceDirect Topics
    Microscopy is defined as a technique that allows for the observation and analysis of small objects or materials, often utilized in various scientific fields ...
  5. [5]
    Microscopy: Intro to microscopes & how they work (article)
    A microscope is an instrument that magnifies objects otherwise too small to be seen, producing an image in which the object appears larger.
  6. [6]
    26.4 Microscopes - College Physics 2e | OpenStax
    Jul 13, 2022 · ... compound microscope composed of two lenses, an objective and an eyepiece. The objective forms a case 1 image that is larger than the object.
  7. [7]
    Basic Concepts in Microscopy - Zeiss Campus
    The objective is responsible for primary image formation at varying magnifications, while the eyepiece is used to observe the image created by the objective.
  8. [8]
  9. [9]
    Microscope Resolution: Concepts, Factors and Calculation
    Abbe recognized that specimen images are composed of a multitude of overlapping, multi-intensity, diffraction-limited points (or Airy discs).
  10. [10]
    Introduction to Electron Microscopy
    Thus, the wavelength of electrons is calculated to be 3.88 pm when the microscope is operated at 100 keV, 2.74 pm at 200 keV, and 2.24 pm at 300 keV.
  11. [11]
    Microscopy Basics | Enhancing Contrast in Transmitted Light
    Phase contrast (as illustrated in Figure 1(b)) employs an optical mechanism to translate minute variations in phase into corresponding changes in amplitude, ...
  12. [12]
    Introduction to Phase Contrast Microscopy - Nikon's MicroscopyU
    In effect, the phase contrast technique employs an optical mechanism to translate minute variations in phase into corresponding changes in amplitude, which can ...Wave Interactions in Phase... · Interpretation of Phase...
  13. [13]
    Microscopy - SERC (Carleton)
    May 30, 2007 · This process may involve immersing the sample (before or after fixation or mounting) in a dye solution and then rinsing and observing the sample ...<|control11|><|separator|>
  14. [14]
    An Introduction to Specimen Preparation - Leica Biosystems
    Understand the key steps in the preparation of specimens for brightfield microscopy in the histopathology laboratory with this introductory guide.
  15. [15]
    13.14: Lab Technique - Light Microscopy - Biology LibreTexts
    Jul 1, 2025 · Light Microscopy Preparations. Light microscopy uses two basic types of slide preparations: prepared slides and wet mounts. Prepared Slides.
  16. [16]
    What is Empty Magnification and How can Users Avoid it
    Nov 21, 2024 · Empty magnification occurs when magnification goes beyond the useful range, making structures appear larger without resolving new details.Missing: types | Show results with:types<|separator|>
  17. [17]
    Useful Magnification Range | Nikon's MicroscopyU
    The useful magnification range is 500 to 1000 times the objective's numerical aperture (NA), calculated as 500 to 1000 x NA.
  18. [18]
    What is Rayleigh Criterion for Microscope Resolution
    5 Abbe's 1873 paper is unusual in that it contained not a single equation. In it, Abbe profoundly stated that microscope resolution is determined by the ...Key Points · Rayleigh Criterion · Abbe Limit<|control11|><|separator|>
  19. [19]
    Finding, defining and breaking the diffraction barrier in microscopy
    Nov 1, 2012 · The limiting role of diffraction for microscopy is extensively discussed in the famous work by Ernst Abbe in 1873 (Abbe 1873). Here, the ...Missing: paper | Show results with:paper
  20. [20]
    Resolution measures in molecular electron microscopy - PMC
    For electron microscopes, the electron wavelength λ depends on the h2 accelerating voltage V. Within the classical approximation, λ = h 2 ...
  21. [21]
    Historical Development, Applications and Advances in Materials ...
    Sep 29, 2020 · Lenses used in vision correction have evolved from the initial glass reading stone of the 10th century to the currently anticipated planar ...Missing: rock | Show results with:rock
  22. [22]
    Science, Optics and You - Timeline - 1000 to 1599
    Nov 13, 2015 · 1590, Dutch lens maker Zacharias Janssen and his father, Hans, invent the first compound microscope. The device is made using a convex objective ...
  23. [23]
    Optical microscopy beyond the diffraction limit - PMC - NIH
    The reason for the limited resolution of an optical microscope is diffraction and, ultimately, the uncertainty principle: a wave cannot be localized much ...
  24. [24]
    Robert Hooke
    Hooke had discovered plant cells -- more precisely, what Hooke saw were the cell walls in cork tissue. In fact, it was Hooke who coined the term "cells": the ...
  25. [25]
    Anatomy of the Microscope - Introduction - Molecular Expressions
    Sep 10, 2018 · Compound microscopes developed during the seventeenth and eighteenth centuries were hampered by optical aberration (both chromatic and spherical) ...Missing: limitations 17th
  26. [26]
    The Microscope: A Crucial "Lens" of History
    Sep 25, 2020 · Anton van Leeuwenhoek (1632-1723), a Dutch cloth-merchant, made his own lenses, and his new lens tube had magnifying power of 270x. He later ...
  27. [27]
    Antonie van Leeuwenhoek: Father of Microbiology
    Aug 2, 2023 · In 1674, Antonie van Leeuwenhoek observed for the first time red blood cells and protozoa; in 1676, the 44-year-old amateur naturalist ...Missing: 270x | Show results with:270x
  28. [28]
    Antony van Leeuwenhoek (1632-1723)
    It was he who discovered bacteria, free-living and parasitic microscopic protists, sperm cells, blood cells, microscopic nematodes and rotifers, and much more.
  29. [29]
    From Animaculum to single molecules: 300 years of the light ...
    The Abbe diffraction limit, thought unbreakable for over one hundred years, has been circumvented by ever more inventive microscopy techniques which are now ...
  30. [30]
    Corrected Lenses and Objectives: An Illuminating History - Bitesize Bio
    Apr 29, 2025 · Lister began studying lenses in the mid-1820s, discovering that varying the distance between lenses could reduce aberrations. Lister published a ...
  31. [31]
    A note on early microscopes - Hektoen International
    Jul 11, 2023 · To correct these optical faults, in 1830 Joseph Jackson Lister with the instrument maker William Tulley made one of the first achromatic ...Missing: apochromatic | Show results with:apochromatic
  32. [32]
    Ernst Abbe and His Work
    **Summary of Ernst Abbe's Work:**
  33. [33]
    History of Oil Immersion Lenses
    In 1867, Tolles made immersion lenses for glycerin as well, the 1/16th inch immersion which resolved Nobert's 19th band, possibly for the first time.Missing: 1880s 0.2 μm
  34. [34]
    Mic-UK: The Oil Immersion Objective - Microscopy-UK
    They yield the highest powers of magnification of around x60 to x130, and highest resolution also, which averages at 0.2 um (130,000 dpi) from an aperture of ...Missing: history 1880s
  35. [35]
    Nobel Prize in Physics 1953
    - **Invention**: Frits Zernike invented phase-contrast microscopy.
  36. [36]
    The Nobel Prize in Physics 1953 - NobelPrize.org
    The Nobel Prize in Physics 1953 was awarded to Frits Zernike for his demonstration of the phase contrast method, especially for his invention of the phase ...Missing: 1930s | Show results with:1930s
  37. [37]
    Ernst Ruska – Facts - NobelPrize.org
    Ernst Ruska discovered that a magnetic coil could be used as a lens for electron beams and developed the first electron microscope in 1933.Missing: 1939 | Show results with:1939
  38. [38]
    Transmission electron microscopy in molecular structural biology
    During the 1950s the established metal shadowing technique continued strongly and was joined, if not overtaken, by negative staining through the 1960s and 70s, ...
  39. [39]
    Dennis McMullan Scanning Microscope - University of Cambridge
    Summary: This article gives an account of the origins of the scanning electron microscope (SEM) and traces its development up to 1965 when the first SEM was ...
  40. [40]
    Introduction to Confocal Microscopy - Evident Scientific
    The basic concept of confocal microscopy was originally developed by Marvin Minsky in the mid-1950s (patented in 1957) when he was a postdoctoral student at ...
  41. [41]
    Introduction to Confocal Microscopy - Molecular Expressions
    Nov 13, 2015 · The basic concept of confocal microscopy was originally developed by Marvin Minsky in the mid-1950s (patented in 1957) when he was a ...
  42. [42]
    What is Confocal laser scanning and how does it work?
    May 10, 2022 · Confocal laser scanning microscopy (or: CLSM) has been used since 1987. It is based on 1957s Marvin Minsky's confocal microscopy patent ...Missing: commercial 1980s history
  43. [43]
    Super-resolution microscopy inventors claim chemistry Nobel
    Oct 8, 2014 · Developed by Stefan Hell in 2000, the technique of stimulated emission depletion (STED) microscopy was the first to overcome Ernst Abbe's ...
  44. [44]
    The Nobel Prize in Chemistry 2014 - Popular information
    Hell and William E. Moerner are awarded the Nobel Prize in Chemistry 2014 for having taken optical microscopy into a new dimension using fluorescent molecules.Missing: invention | Show results with:invention
  45. [45]
    A guide to super-resolution fluorescence microscopy
    Jul 19, 2010 · These new super-resolution technologies are either based on tailored illumination, nonlinear fluorophore responses, or the precise localization of single ...
  46. [46]
    Cryo-EM Explained: Visualizing Life's Machinery
    These feats earned cryo-EM's early pioneers the 2017 Nobel Prize in Chemistry. ... At about 3 Å, or “near-atomic resolution,” particles in the image are ...
  47. [47]
    [PDF] THE DEVELOPMENT OF CRYO-ELECTRON MICROSCOPY
    Oct 4, 2017 · Frank and Richard Henderson the Nobel Prize for Chemistry 2017 for "developing cryo- electron microscopy for the high-resolution structure ...
  48. [48]
    Seeing more in expansion microscopy | MIT News
    Mar 3, 2025 · New methods light up lipid membranes and let researchers see sets of proteins inside cells with high resolution.
  49. [49]
    Unraveling cellular complexity with transient adapters in highly ...
    Mar 28, 2024 · We present a method for rapid multiplexed super-resolution microscopy that can, in principle, be applied to a nearly unlimited number of molecular targets.
  50. [50]
    Next-gen electron microscope at UCLA opens new chapter in ...
    Jan 23, 2025 · The next generation of the technology, the Thermo Fisher Scientific Krios G4, is now online and available to users from UCLA and researchers ...
  51. [51]
    Advancing biological super-resolution microscopy through deep ...
    In this brief review, we survey recent advances in using deep learning to enhance the performance of biological super-resolution microscopy.
  52. [52]
    Self-supervised machine learning framework for high-throughput ...
    Apr 2, 2025 · We introduce SHINE, the Self-supervised High-throughput Image denoising Neural network for Electron microscopy, accelerating minimally invasive ...
  53. [53]
    AI analysis of super-resolution microscopy: Biological discovery in ...
    Recent application of AI to reconstruction methods has the potential to reduce noise, improve spatial and temporal resolution, and automate high-throughput ...
  54. [54]
  55. [55]
    Specialized Microscopy Techniques - Darkfield Illumination
    Sep 11, 2018 · The cardioid condenser is a reflecting darkfield design that relies on internal mirrors to project an aberration-free cone of light onto the ...
  56. [56]
    Special Oil-Immersion Objectives for Dark-Field Microscopy - jstor
    FOR DARK-FIELD MICROSCOPY. DARK-FIELD microscopy was introduced by. Joseph Jackson Lister in 1830, and by the Rev. J. B. Reade in 1837. The optical principles.
  57. [57]
    Phase contrast microscopy of living cells within the whole lens - NIH
    Aug 1, 2012 · Current microscope technology for observing unstained living cells (such as phase contrast and differential interference contrast microscopy) ...
  58. [58]
    Fundamental Concepts in DIC Microscopy - Evident Scientific
    Although Nomarski prisms are widely employed as objective prisms in modern differential interference contrast microscopes, there are fewer spatial constraints ...
  59. [59]
    Comparison of Phase Contrast and DIC Microscopy - Evident Scientific
    The lateral (x-y) and axial (z) resolution in differential interference contrast microscopy dramatically exceeds that with phase contrast, which exhibits ...
  60. [60]
    Epi-Fluorescence Microscopy - PMC - PubMed Central - NIH
    Green emission light from the sample passes through the dichroic mirror and through a green band pass emission filter to the detector. The emission filter also ...
  61. [61]
    Fluorescence Microscopy: A Concise Guide to Current Imaging ...
    In addition to cameras, wide-field microscopy has also been improved by better light filters, mirrors, and objectives. Commercially available filters, for ...Missing: epifluorescence | Show results with:epifluorescence
  62. [62]
    An introduction to Performing Immunofluorescence Staining - PMC
    Here, we describe preparation of specimens preserved in different types of media and step-by-step methods for both direct and indirect immunofluorescence ...
  63. [63]
    Confocal Microscopy: Principles and Modern Practices - PMC
    The best resolution that can be obtained is ~ 0.2 μm laterally and ... Confocal z stacks were collected with an oil immersion objective and 0.8 μm slices.
  64. [64]
    Fluorescence microscope light source based on integrated LED
    Sep 12, 2023 · The LED industry has also grown significantly, resulting in increased production, reduced costs, and the development of new materials with the ...
  65. [65]
    Deep Tissue Imaging with Multiphoton Fluorescence Microscopy - NIH
    Sep 27, 2017 · We present a review of imaging deep-tissue structures with multiphoton microscopy. We examine the effects of light scattering and absorption.
  66. [66]
    Sub-diffraction-limit imaging by stochastic optical reconstruction ...
    Aug 9, 2006 · In this work we report a new high-resolution imaging technique, stochastic optical reconstruction microscopy (STORM), in which a fluorescence ...
  67. [67]
    Breaking the diffraction resolution limit by stimulated emission
    Breaking the diffraction resolution limit by stimulated emission: stimulated-emission-depletion fluorescence microscopy. Stefan W. Hell and Jan Wichmann.
  68. [68]
    Imaging cellular structures in super-resolution with SIM, STED and ...
    Jun 6, 2016 · Classically the resolution of the light microscope is limited by Abbe's Law to 200–250 nm in the lateral and 500–700 nm in the axial direction.Results · Discussion · Microscopy
  69. [69]
    Imaging Intracellular Fluorescent Proteins at Nanometer Resolution
    This technique, termed photoactivated localization microscopy (PALM), is capable of resolving the most precisely localized molecules at separations of a few ...
  70. [70]
    Nonlinear structured-illumination microscopy: Wide-field ... - PNAS
    This article experimentally demonstrates saturated structured-illumination microscopy, a recently proposed method in which the nonlinearity arises from ...
  71. [71]
    Expansion microscopy | Science
    We demonstrate ExM with apparent ~70-nanometer lateral resolution in both cultured cells and brain tissue, performing three-color superresolution imaging of ~10 ...
  72. [72]
    Dense, continuous membrane labeling and expansion microscopy ...
    Feb 12, 2025 · Protein and lipid expansion microscopy with trypsin and tyramide signal amplification for 3D imaging. Sci. Rep. 13, 21922 (2023). Article ...
  73. [73]
    UM School of Medicine Confocal Core Facility Offers Super ...
    Jul 1, 2025 · STED microscopy, also known as Stimulated Emission Depletion microscopy, is a powerful advancement in confocal imaging that provides clear views ...
  74. [74]
    Between life and death: strategies to reduce phototoxicity in super ...
    Here, we give an overview of new developments in hardware, software and probe chemistry aiming to reduce phototoxicity.
  75. [75]
    Transmission Electron Microscope (TEM) | Science Basics | Products
    The accelerating voltages commonly used are 100 to 200 kV (corresponding to wavelengths of 0.0037nm to 0.0025nm). This wavelength is far shorter than that ...
  76. [76]
    Transmission Electron Microscopy - Semitracks
    A TEM (transmission electron microscope) uses a highly energetic electron beam (100 keV - 1 MeV) to image and obtain structural information from thin film ...
  77. [77]
    Brief Introduction to Contrasting for EM Sample Preparation
    Oct 2, 2013 · The double contrast method of ultrathin sections with uranyl acetate (UA) and lead citrate is the standard contrasting technique for electron ...
  78. [78]
    Atomic-resolution transmission electron microscopy of ... - Science
    Jan 18, 2018 · High-resolution transmission electron microscopy is an invaluable tool for looking at the crystalline structures of many materials.
  79. [79]
    [PDF] Transmission Electron Microscopy - nanoHUB
    1) Imaging techniques in TEM mode a) Bright-Field TEM (Diff. contrast) b) Dark-Field TEM (Diff. contrast) c) Weak-beam imaging hollow-cone dark-field ...
  80. [80]
    Limitations of TEM - News-Medical
    Damages due to radiation. Ionizing radiation can always damage the specimens used in TEM. Polymers, organic materials, certain minerals, and ceramics are ...
  81. [81]
    [PDF] Radiation damage in the TEM and SEM - Frank Lab
    Abstract. We review the various ways in which an electron beam can adversely affect an organic or inorganic sample during examination in an.
  82. [82]
    Replacing critical point drying with a low-cost chemical drying ...
    The steps required to prepare any tissue for SEM include fixation, dehydration, critical-point drying (CPD), mounting, and coating with gold or palladium to ...
  83. [83]
    Generalized single-particle cryo-EM – a historical perspective - PMC
    This is a brief account of the earlier history of single-particle cryo-EM of biological molecules lacking internal symmetry, which goes back to the ...
  84. [84]
    Bringing Structure to Cell Biology with Cryo-Electron Tomography
    May 9, 2023 · To bring structure into cell biology, the modality of cryo-electron tomography has fast developed into a bona fide in situ structural biology ...
  85. [85]
    Automated vitrification of cryo-EM samples with controllable ... - Nature
    May 27, 2022 · Cryo fixation into vitreous water (amorphous ice) by fast freezing of biological samples can deliver nearly perfect structural preservation ...Missing: rapid | Show results with:rapid
  86. [86]
    Single-particle cryo-EM data acquisition by using direct electron ...
    Nov 6, 2015 · In this review, we discuss how the applications of direct electron detection cameras in cryo-EM have changed the way the data are acquired.
  87. [87]
    Cryo-EM University | Sample Preparation - Thermo Fisher Scientific
    Filling the ethane cup. Demonstration of how to carefully fill the ethane cup with liquid ethane for plunge freezing with the Vitrobot System.
  88. [88]
    High-resolution cryo-EM performance comparison of two latest ...
    Oct 12, 2022 · Here we benchmark the performance of two 300 kV latest-generation cryo electron microscopes, Titan Krios G4 from Thermofisher Scientific and CRYO ARM 300 from ...
  89. [89]
    Krios G4 cryo-electron microscope among first at a U.S. university
    The Krios G4 cyro-electron microscope is expected to help with new discoveries and innovations in fields such as cancer and Alzheimer's research.
  90. [90]
    Low-dose cryo-electron ptychography of proteins at sub-nanometer ...
    Sep 14, 2024 · Here we apply 4D-STEM and ptychographic data analysis to frozen hydrated proteins, reaching sub-nanometer resolution 3D reconstructions.
  91. [91]
    New tools for automated cryo-EM single-particle analysis in RELION ...
    We describe new tools for the processing of electron cryo-microscopy (cryo-EM) images in the fourth major release of the RELION software.
  92. [92]
    Recent advances in infectious disease research using cryo-electron ...
    Although diagnostic applications have diminished, purified viruses are still commonly used in cryo-EM studies to generate high resolution structures. Fewer ...
  93. [93]
    Challenges and opportunities in cryo-EM single-particle analysis
    Cryo-EM techniques open entirely new questions and raise novel challenges, especially pertaining to dynamic and heterogeneous assemblies.
  94. [94]
    Avoiding the pitfalls of single particle cryo-electron microscopy - PNAS
    Single particle cryo-electron microscopy is currently poised to produce high-resolution structures of many biological assemblies, but several pitfalls can ...
  95. [95]
    REliable PIcking by Consensus (REPIC) - Nature
    Oct 31, 2024 · Cryo-EM particle identification from micrographs (“picking”) is challenging due to the low signal-to-noise ratio and lack of ground truth ...
  96. [96]
    Atomic Force Microscope | Phys. Rev. Lett.
    The atomic force microscope, introduced in 1986, provided atomic-scale pictures of surfaces, with few limitations on the type of sample. See more in Physics ...
  97. [97]
    "Dip-Pen" Nanolithography - Science
    A direct-write “dip-pen” nanolithography (DPN) has been developed to deliver collections of molecules in a positive printing mode.
  98. [98]
    [PDF] Gerd Binnig and Heinrich Rohrer - Nobel Lecture
    We present here the historic development of Scanning Tunneling Microscopy; the physical and technical aspects have already been covered in a few recent.
  99. [99]
    Surface Studies by Scanning Tunneling Microscopy | Phys. Rev. Lett.
    Jul 5, 1982 · Surface microscopy using vacuum tunneling is demonstrated for the first time. Topographic pictures of surfaces on an atomic scale have been obtained.Missing: original | Show results with:original
  100. [100]
  101. [101]
    [PDF] Scanning Tunneling Microscope - UW–Madison Physics
    Apr 26, 2000 · This formula shows that for each angstrom change in separation, the probability that an electron tunnels decreases by an order of magnitude.
  102. [102]
    Scanning tunneling microscopy---from birth to adolescence | Rev
    Jul 1, 1987 · This collection marks the 35th anniversary of scanning tunneling microscopy (STM) and the 30th anniversary of atomic force microscopy (AFM).<|separator|>
  103. [103]
    Scanning Tunneling Microscopy (STM): An Overview
    Scanning Tunneling Microscopy allows researchers to map a conductive sample's surface atom by atom with ultra-high resolution, without the use of electron ...Missing: original | Show results with:original<|control11|><|separator|>
  104. [104]
    8.3: Scanning Tunneling Microscopy - Chemistry LibreTexts
    Aug 28, 2022 · Limitations. The majority of the limitations in this technique come from the materials and construction of the optical fibers and the prism used ...Principles of Scanning... · Applications of Scanning... · Adaptations to Scanning...
  105. [105]
    [PDF] Advanced Scanning Probe Microscopy I - nanoHUB
    Advanced Scanning Probe Microscopy 18. STM: Requirements and Limitations. • Sample and tip must be conductive. • Requires isolation from vibrations. • Precise ...
  106. [106]
    [PDF] Laboratory Soft X-Ray Microscopy and Tomography - DiVA portal
    Radiation damage clearly limits the attainable resolution of biological samples in x-ray microscopy. One approach to reduce the needed dose is to increase ...
  107. [107]
    Three-dimensional cellular ultrastructure resolved by X-ray microscopy
    Nov 14, 2010 · A key feature of the X-ray microscope is its ability to generate 3D views of intact cells at nanoscale resolution. The x-z slices through our ...
  108. [108]
    50-nm-resolution full-field X-ray microscope without chromatic ...
    Apr 13, 2017 · X-ray spectromicroscopy with a full-field imaging technique is a powerful method for chemical analysis of heterogeneous complex materials ...
  109. [109]
    Transmission X-ray Microscopy and Its Applications in Battery ...
    Jul 28, 2021 · This article presents a short review of the TXM, including its instrumentation, battery research applications, and the practical sample ...
  110. [110]
    Synchrotron X-ray tomographic microscopy of fossil embryos
    Here, we have applied synchrotron-radiation X-ray tomographic microscopy (SRXTM), obtaining complete three-dimensional recordings at submicrometre resolution.
  111. [111]
    X-ray phase-contrast microtomography of soft tissues using a ... - NIH
    We report on a compact and cost-effective system for x-ray phase-contrast microtomography. The system features high sensitivity to phase gradients and high ...
  112. [112]
    Advances in ultraviolet microscopy - IOPscience
    Aug 18, 2025 · Here we review the latest advances and trends in UV microscopy for applications in the life sciences, including histology, cell biology and ...
  113. [113]
    [PDF] TB62: The Suitability of Using Fluorescence Microscopy for Studying ...
    Nov 1, 1972 · If shorter wavelength ultra-violet radiation is required, quartz optics must be used (Needham, 1958). The absorption curve for lignin in the ...
  114. [114]
    Lithographic Processes for the Scalable Fabrication of Micro - NIH
    In the photolithographic process, UV light is used to transfer a pattern defined on a photomask onto a photosensitive resist coated on the substrate. Over the ...Missing: staining | Show results with:staining
  115. [115]
    Quantifying UV-induced photodamage for longitudinal live-cell ...
    Dec 19, 2024 · This work provides UV fluence thresholds that can serve as guidelines for nondestructive, longitudinal, and dynamic deep-UV microscopy experiments.
  116. [116]
    Light-induced cell damage in live-cell super-resolution microscopy
    Oct 20, 2015 · Total-internal reflection fluorescence (TIRF) microscopy can be used to lower the penetration depth to merely the basal cell membrane. In order ...Results · Online Materials And Methods · Irradiation And...Missing: limitations sample
  117. [117]
    FTIR-Microscopy | Infrared Radiation | Vibrational Modes - Jordi Labs
    FTIR-Microscopy is the tool of choice! During FTIR-Microscopy, molecules absorb the applied infrared radiation at their characteristic vibrational modes.
  118. [118]
    FT-IR Microscopic Analysis of Polymer Laminate Samples Including ...
    Aug 1, 2016 · This paper details the preparation of polymer laminate sample cross-sections and the collection of transmission and ATR spectra of various layers.
  119. [119]
    Exploring Fourier Transform Infrared (FTIR) spectroscopy for tissue ...
    Jun 17, 2024 · FTIR spectroscopy has emerged as a powerful analytical tool in medical research, offering non‐invasive and precise examination of the molecular composition of ...Missing: polymer ⁻
  120. [120]
    [PDF] Nicolet iN10 Infrared Microscope | Thermo Fisher Scientific
    Using a mercuric cadmium telluride (MCT) detector, motorized stage, or MicroTip attenuated total reflection (ATR) enables you to identify materials as small as.
  121. [121]
    Maximizing Your IR Microscope: Techniques for Improved ...
    Reflection Method. The IR reflection method is extremely simple, requiring little to no sample preparation. However, measurements are limited to certain sample ...Missing: setup MCT
  122. [122]
  123. [123]
  124. [124]
  125. [125]
    Quantitative phase imaging based on holography: trends and new ...
    Jun 27, 2024 · After 75 years of development, holographic imaging has become a powerful tool for optical wavefront measurement and quantitative phase imaging.
  126. [126]
    Review of quantitative phase-digital holographic microscopy
    Quantitative phase-digital holographic microscopy (QP-DHM) represents an important and efficient quantitative phase method to explore cell structure and ...Missing: paper | Show results with:paper
  127. [127]
    Digital Whole Slide Imaging Compared With Light Microscopy for ...
    Feb 14, 2020 · Digital whole slide imaging compared with light microscopy for primary diagnosis in surgical pathology: a multicenter, double-blinded, randomized study of 2045 ...
  128. [128]
    Whole Slide Imaging (WSI) in Pathology: Emerging Trends and ...
    One of the recent advancements in pathology is Whole Slide Imaging (WSI), a technique that involves scanning glass slides to produce high-resolution digital ...
  129. [129]
    Roche Digital Pathology Dx whole slide imaging system is ...
    Jun 10, 2025 · A randomized study comparing digital imaging to traditional glass slide microscopy for breast biopsy and cancer diagnosis . J Pathol Inform.
  130. [130]
    Visualizing Synaptic Connectivity Using Confocal and Electron ...
    Visualizing the synaptic connectivity of identified neuronal populations is an essential part of defining neuronal circuits in the central nervous system. In ...
  131. [131]
    Synaptic Imaging & Mass-Spectrometry for Synaptopathy
    Mar 15, 2023 · Electron microscopy is still considered the gold standard imaging technique for studying synaptic structure and has been used to study synapse ...
  132. [132]
    Super-resolving Microscopy in Neuroscience | Chemical Reviews
    Mar 22, 2021 · In this review, we describe how super-resolving microscopy has improved our understanding of neuronal and brain function and dysfunction in the last two ...
  133. [133]
    Clinical Application of Confocal Laser Endomicroscopy in the ...
    Sep 18, 2025 · Confocal laser endomicroscopy (CLE) is an endoscopic imaging method that uses a low-power laser beam (wavelength 488 nm) focused on a single ...
  134. [134]
    Confocal laser endomicroscopy in the “in vivo” histological ... - NIH
    This newly-developed technology has enabled endoscopists to collect real-time in vivo histological images or “virtual biopsies” of the gastrointestinal mucosa ...
  135. [135]
    Harnessing artificial intelligence to reduce phototoxicity in live imaging
    The integration of AI components directly into the fluorescence microscopy acquisition sequence shows great promise for minimizing photodamage in real-time and ...
  136. [136]
    Development of AI-assisted microscopy frameworks through realistic ...
    Sep 26, 2024 · The integration of artificial intelligence into microscopy systems significantly enhances performance, optimizing both image acquisition and ...
  137. [137]
    Recent Advances in Transmission Electron Microscopy for Materials ...
    The combination of high-resolution imaging and high energy resolution spectroscopies has proven to be a formidable tool to solve materials science problems.Missing: seminal | Show results with:seminal
  138. [138]
    Microscopic Techniques for Nanomaterials Characterization
    Jan 8, 2025 · There are a few microscopic techniques used to characterize nanomaterial morphology, AFM (atomic force microscopy), TEM (transmission electron microscopy), SEM ...
  139. [139]
    Forensic microscopy | McGraw Hill's AccessScience
    Edmond Locard (1877–1966) stated that every contact leaves a trace. This is known as the Locard exchange principle, the basis for much of forensic microscopy.
  140. [140]
    [PDF] Microscopy in Forensic Science - Abertay University
    This chapter examines the use of electron microscopy, atomic force microscopy and other analytical techniques in forensic investigation and research. These ...