Half-value layer

The half-value layer (HVL), also known as half-value thickness, is the thickness of a specified material that attenuates the intensity of a beam of ionizing radiation—such as X-rays or gamma rays—to exactly one-half of its initial value. This measure is fundamental in radiation physics and dosimetry, providing a practical indicator of a material's shielding effectiveness against radiation of a given energy, independent of the beam's initial intensity. The HVL varies with the type of radiation, its photon energy, and the absorbing material (e.g., lead, concrete, or aluminum), typically expressed in units of millimeters or centimeters. In and shielding design, the HVL is used to calculate the required thickness of barriers to achieve desired dose reduction levels, often in combination with the exponential law, where multiple HVLs correspond to successive halvings of (e.g., two HVLs reduce it to one-quarter). For diagnostic beams, regulatory standards mandate measurement of the HVL to assess beam quality and ensure compliance with limits, as higher-energy beams require thicker materials for the same . Similarly, in and , HVL values help quantify gamma-ray shielding for sources like , with empirical data showing, for instance, a lead HVL of approximately 1.2 cm for gamma rays (1.17–1.33 MeV). The concept extends to tenth-value layers (TVL), which reduce by a factor of 10 and equal about 3.32 HVLs, aiding in broader shielding calculations.

Fundamentals

Definition

The half-value layer (HVL) is the thickness of a specified absorbing material that reduces the intensity of a beam of , such as gamma rays or s, to exactly half its initial value. For beams, this is often quantified in terms of air kerma, the released per unit mass in air. The term originated in the early amid advancements in physics and technology, with its first documented use appearing in to describe beam absorption properties. HVL values are typically expressed in millimeters (mm) or centimeters (cm) of the absorber, such as aluminum for diagnostic X-rays or lead for higher-energy gamma rays. This measure applies to both monoenergetic beams, like those from radioactive isotopes, and polychromatic beams, such as those produced by tubes, though it is primarily relevant for penetrating .

Physical Significance

The half-value layer (HVL) serves as a key measure of a material's effectiveness in attenuating , quantifying the thickness required to reduce the intensity of an incident to half its original value, thereby providing insight into the 's penetrating power. A higher HVL indicates a "harder" with more penetrating , often associated with higher average energies, while a lower HVL signifies greater and thus softer ; this makes HVL an essential indicator of quality in applications like diagnostic imaging and shielding design. The value of HVL is intrinsically linked to the absorber's physical properties, including its , (Z), and , which influence the probability of interactions such as photoelectric absorption and . Materials with higher s, like lead (Z = 82), exhibit smaller HVLs for a given due to enhanced interaction cross-sections, making them more effective shields compared to lower-Z materials such as aluminum or . Similarly, higher and increase the number of interacting atoms per unit volume, further reducing the HVL by amplifying efficiency. HVL also demonstrates strong dependence on photon energy: as energy increases, the HVL generally lengthens because interaction probabilities decrease, particularly with the dominance of over photoelectric effects in the intermediate to high-energy range (e.g., above ~100 keV), allowing photons to penetrate more deeply. For polychromatic beams, such as those produced by tubes, the effective HVL reflects beam hardening, where initial preferentially removes lower-energy photons, raising the average beam energy and thereby increasing the measured HVL for subsequent layers. While versatile for photon-based radiation like X-rays and gamma rays, the HVL concept is specifically tailored to neutral particles that follow attenuation laws and does not directly apply to charged particles, such as electrons or protons, whose interactions involve significant and paths, nor to neutrons, which require distinct and capture mechanisms for shielding.

Mathematical Formulation

Exponential Attenuation Law

The exponential attenuation law, also known as Beer's law or the Beer-Lambert law in the context of radiation physics, describes how the intensity of a beam decreases as it passes through a material due to interactions that remove from the beam. This principle applies specifically to narrow, collimated beams where scattered radiation is minimized, ensuring that the observed reduction in intensity arises primarily from and coherent scattering without significant contributions from diffuse scatter. In such conditions, the transmitted intensity follows an , reflecting the probabilistic nature of interactions with matter. The mathematical expression for this law is given by: I = I_0 e^{-\mu x} where I is the transmitted after passing through a thickness x (in cm) of , I_0 is the incident , and \mu is the linear (in cm⁻¹), which quantifies the fraction of photons attenuated per unit length. This equation assumes a monoenergetic of photons, as polychromatic beams (common in sources) would require over the energy spectrum for accurate description. The linear \mu is related to properties by \mu = \rho (\mu / \rho), where \rho is the of the (in g/cm³) and \mu / \rho is the (in cm²/g), which depends on and atomic composition but is independent of . Values of \mu / \rho are extensively tabulated for various elements and compounds across photon energies from 1 keV to 20 MeV, enabling practical calculations for shielding and . This law holds under specific assumptions, including the use of monoenergetic photons, good to prevent buildup of scattered in the detector, and a homogeneous attenuating medium with no significant or secondary interactions altering the beam. It is valid for primary beam in narrow-beam configurations, where side-scattered photons are excluded, but deviates in broad-beam or scattering-inclusive setups. Historically, the exponential form was first formulated for visible absorption by in 1760 and refined by August Beer in 1852 to include concentration dependence; its application to X-rays emerged in the early 1900s through empirical observations following Wilhelm Röntgen's 1895 discovery, integrated with emerging quantum concepts to explain photoelectric and Compton effects.

Derivation and Calculation of HVL

The half-value layer (HVL) for monochromatic radiation is derived directly from the exponential attenuation law by identifying the thickness x that reduces the initial intensity I_0 to half its value. Substituting I = I_0 / 2 into the equation I = I_0 e^{-\mu x} (where \mu is the linear attenuation coefficient) yields $1/2 = e^{-\mu x_{\text{HVL}}}. Taking the natural logarithm of both sides gives \ln(1/2) = -\mu x_{\text{HVL}}, which simplifies to x_{\text{HVL}} = \ln(2) / \mu \approx 0.693 / \mu. This expression provides the theoretical basis for computing HVL in materials where \mu is known. For polychromatic beams, such as those produced by tubes, the HVL calculation requires an effective \mu_{\text{eff}} obtained by integrating the beam spectrum \phi(E) over E: the transmitted intensity is \int \phi(E) e^{-\mu(E) x} dE / \int \phi(E) dE = 1/2, solved iteratively for x. This effective \mu is often determined by matching the HVL to an equivalent monochromatic using tabulated coefficients, accounting for the spectrum's distribution. As an illustrative example, consider 100 keV gamma rays in aluminum (density \rho = 2.7 g/cm³). The \mu/\rho = 0.1704 cm²/g, so the \mu = (\mu/\rho) \cdot \rho \approx 0.460 cm⁻¹. The HVL is then x_{\text{HVL}} \approx 0.693 / 0.460 \approx 1.51 cm. Similar calculations apply to other energies and materials; for instance, representative HVL values for common gamma sources in shielding materials are shown below (values approximate narrow-beam geometry for monoenergetic equivalents).
Gamma SourceEnergy (MeV) (cm) (cm)Lead (cm)
Ir-192~0.384.51.30.5
Co-60~1.256.12.21.3
Cs-1370.6624.81.60.7
In polychromatic beams, multiple HVLs provide insight into beam quality. The first HVL (HVL₁) is calculated for the unfiltered beam using the initial effective \mu. The second HVL (HVL₂) is then determined after adding equivalent to HVL₁, reflecting the hardened . The homogeneity , defined as HVL₁ / HVL₂, quantifies polychromaticity; values near 1 indicate near-monoenergetic beams, while lower values (e.g., 0.7–0.9 for typical spectra) signal greater spectral breadth. This iterative approach refines beam characterization theoretically. Theoretical HVL calculations can introduce errors if certain effects are neglected. Scatter radiation increases the apparent transmitted intensity in non-ideal geometries, leading to an underestimation of HVL because the thickness required to achieve a true 50% primary reduction appears smaller when scatter contributions are unaccounted for. Similarly, beam hardening in polychromatic s shifts the spectrum toward higher energies with increasing thickness, raising the effective HVL; neglecting this (e.g., by assuming a fixed \mu) underestimates the HVL and overpredicts for thicker absorbers.

Determination Methods

Experimental Measurement

The experimental measurement of the half-value layer (HVL) typically employs an or suitable detector to quantify the exposure rate or of the beam, with absorbers incrementally added until the intensity is reduced to half its initial value. This approach ensures compliance with established protocols for assessing beam quality in diagnostic and therapeutic settings. For beams, aluminum sheets of high purity (≥99.9%) serve as the primary absorbers due to their effectiveness in attenuating lower-energy photons. The measurement setup requires a fixed source-to-detector , commonly 100 , to maintain consistent and . Narrow beam collimation, often limiting the field to 50 mm × 50 mm, minimizes scattered that could otherwise inflate the apparent HVL. All procedures adhere to international standards such as IEC 61267, which specifies the use of precise , including voltage dividers for accurate kVp determination and diaphragms for restriction, while ensuring minimal by restricting objects within the path. The standard procedure involves the following steps:
  1. Position the ionization chamber's reference point on the 's central axis at the specified distance, with no initial absorber in place, and measure the unfiltered rate X_0.
  2. Incrementally add aluminum absorbers of known thicknesses d_i (in steps of ≤0.5 , with ±0.01 accuracy) between the source and detector, recording the corresponding rates X_i for each configuration until the beam is attenuated by a factor of at least 6.
  3. Plot the as \log(X_i / X_0) versus d_i on semilogarithmic or via ; the HVL is the thickness d at which \log(0.5) = -0.3010, corresponding to half the initial intensity.
For gamma-ray beams, such as those from Cs-137 or Co-60 sources, higher (Z) materials like lead are used as absorbers to achieve efficient , with thicknesses typically measured in centimeters. Detectors such as Geiger-Müller counters or NaI(Tl) detectors replace ionization chambers, and the setup maintains a fixed similar to measurements. In broad beam conditions, corrections for the buildup factor—accounting for scattered photons increasing the effective dose—are essential to avoid underestimating the true HVL. Quality assurance protocols include measuring a second HVL (HVL₂), obtained by adding absorbers equivalent to the first HVL thickness before repeating the process; for polychromatic beams, HVL₂ is typically 20-30% greater than the first HVL (HVL₁), indicating beam hardening. Typical HVL values for diagnostic range from 2 to 5 mm of aluminum at tube potentials of 50-150 kVp, depending on and , with lower values signifying softer beams suitable for superficial . Challenges in HVL measurement arise from source variability, such as fluctuations in tube voltage or output due to aging components, which can introduce up to 5% error if not calibrated. Scatter correction remains critical, particularly in confined spaces like cabinet systems, where wall reflections can distort results without adequate collimation or . Modern digital detectors, including semiconductor-based systems, enhance precision by providing real-time data acquisition and reducing statistical uncertainties compared to traditional film or analog methods.

Theoretical Estimation

Theoretical estimation of the half-value layer (HVL) relies on computational methods and that predict attenuation without requiring physical measurements, particularly useful for monoenergetic photons or preliminary design in shielding applications. For monoenergetic radiation, the μ can be derived from mass attenuation coefficients (μ/ρ) obtained from such as the NIST , which provides tabulated values as a function of and Z for elements and compounds. The HVL is then calculated using the formula \text{HVL} = \frac{\ln 2}{\mu} = \frac{\ln 2 \cdot \rho}{\mu / \rho}, where ρ is the material density, enabling rapid estimation for pure materials under narrow-beam conditions. For polychromatic spectra, such as those from X-ray tubes, Monte Carlo simulations offer a robust approach by modeling photon interactions, transport, and energy deposition in materials. Tools like MCNP and GEANT4 simulate beam spectra, filtration effects, and geometry to compute effective HVL by iteratively determining the thickness that halves the transmitted intensity. Empirical formulas provide quicker approximations; for diagnostic X-ray beams, semi-empirical models based on tube voltage (kVp) and added filtration, such as those outlined in NCRP Report 147, estimate HVL by fitting transmission data to equations like the Archer form. These theoretical methods generally agree with experimental narrow-beam HVL values within 5-10%, with Monte Carlo simulations achieving discrepancies as low as 4% when spectrum assumptions are accurate, though larger errors arise from simplifications in beam hardening or scatter modeling. For advanced cases involving composite materials or partial volume effects, layered calculations apply the attenuation law successively across strata, using effective μ for each layer derived from mixture rules or simulations, or directly via to account for heterogeneous interactions.

Applications

Radiation Shielding

In radiation shielding, the half-value layer (HVL) serves as a fundamental parameter for determining the thickness of material required to attenuate gamma radiation to a specified reduction factor R, where R is the ratio of initial to final intensity. The number of HVLs needed, n, is calculated as n = \log_2 R, and the total shielding thickness is then x = n \times \text{HVL}. For instance, 10 HVLs reduce the intensity by a factor of $2^{10} = 1024, providing substantial attenuation for protective barriers around radiation sources. For broad beam conditions, which are common in practical shielding scenarios due to scattered , the buildup factor B (a unitless greater than 1) must be incorporated to account for secondary photons. The effective thickness becomes x_\text{eff} = \text{HVL} \times \log_2 (R B), adjusting for the increased radiation flux from and other interactions within the shield. This contrasts with narrow beam geometry, where B \approx 1 and simple applies without buildup correction. Material selection for shielding depends on the radiation energy and desired practicality, with high-density materials like lead preferred for compact designs and for cost-effective, structural barriers. For common isotopes such as cesium-137 (Cs-137, 0.66 MeV) and (Co-60, 1.17 and 1.33 MeV), HVL values guide thickness calculations; for example, offers HVLs of approximately 5 to 7 cm for Co-60 gamma rays, while lead provides around 1 cm. The following table summarizes representative HVLs for these materials and isotopes:
IsotopeEnergy (MeV)Concrete (cm)Steel (cm)Lead (cm)
Cs-1370.664.81.60.7
Co-601.17, 1.336.62.11.2
HVL calculations are integral to regulatory frameworks for ensuring compliance with dose limits, as outlined in National Council on Radiation Protection and Measurements (NCRP) Report No. 151 and (ICRP) Publication 103, which emphasize maintaining occupational and public exposures as low as reasonably achievable (ALARA). In hospital vault design for linear accelerators, NCRP 151 recommends using HVL to refine secondary barrier thicknesses, adding one HVL when patient-scattered and head-leakage contributions are comparable to conservatively meet weekly dose limits (e.g., 0.02 mSv for uncontrolled areas). Despite its utility, the HVL approach has limitations, as narrow-beam values ignore significant secondary radiation production in broad-beam setups, potentially underestimating required thicknesses; thus, designs must integrate buildup factors and adhere to the ALARA principle for comprehensive protection.

Diagnostic Radiology and Quality Control

In diagnostic radiology, the half-value layer (HVL) serves as a primary metric for evaluating beam quality, quantifying the penetrating power by indicating the thickness of aluminum required to attenuate the beam intensity to half its original value. This measure ensures that the beam is sufficiently hardened through inherent and added filtration to remove low-energy photons, which primarily contribute to patient skin dose without enhancing . Regulatory standards, such as those outlined in 21 CFR 1020.30, specify minimum HVL values based on tube potential (kVp); for instance, at 80 kVp, systems manufactured before June 10, 2006, require at least 2.3 mm Al, while later systems demand 2.9 mm Al to promote dose reduction and optimal imaging. Quality control (QC) protocols emphasize annual HVL testing to maintain beam integrity and , as recommended by the American Association of Physicists in Medicine (AAPM) and mandated by the (FDA). These assessments, typically conducted by medical physicists using ionization chambers or solid-state detectors with incremental aluminum attenuators, verify compliance with minimum standards and detect deviations that could arise from aging, filtration degradation, or generator inconsistencies. A deviation exceeding 10% from baseline or expected values often signals the need for , such as filter replacement or tube evaluation, to prevent suboptimal beam quality. Added filtration, commonly aluminum (Al) or copper (Cu), directly elevates HVL by preferentially absorbing low-energy X-rays, hardening the spectrum to improve deep tissue penetration while substantially lowering entrance skin dose. For example, increasing total filtration from 1 mm to 3 mm Al at 80 kVp can boost HVL by about 50%, reducing superficial exposure that does not aid diagnostics and thereby minimizing radiation risk without compromising overall image utility. This beam hardening is integral to protocols across general radiography and fluoroscopy, where Cu filtration (e.g., 0.1-0.2 mm) further refines quality for prolonged procedures. Clinically, higher HVL values facilitate effective imaging of thicker body parts, such as the abdomen or pelvis, by enhancing photon transmission through increased attenuation, which allows adequate detector exposure at lower tube currents and thus reduces overall patient dose; this penetration compensates for reduced subject contrast inherent to harder beams. In specialized applications like mammography, a molybdenum (Mo) target paired with a 0.03 mm Mo filter produces a characteristically low HVL of approximately 0.3-0.35 mm Al equivalent at 25-30 kVp, optimizing soft-tissue contrast for detecting subtle lesions while adhering to low-dose imperatives. Contemporary systems leverage HVL data for automated optimization, adjusting parameters to balance quality and dose across diverse anatomies. This approach evolved from the empirical tests of the 1920s, which relied on rudimentary measurements amid early concerns, to the standardized methodologies of the 1970s, when federal performance standards under the Radiation Control for Health and Safety Act of 1968 formalized HVL requirements and measurement techniques for diagnostic equipment.

Tenth-Value Layer

The tenth-value layer (TVL), also known as the decadelayer, is defined as the thickness of a specified shielding material that attenuates ionizing radiation, such as gamma rays or X-rays, to one-tenth (1/10) of its initial intensity under narrow-beam conditions. This measure is particularly relevant for monoenergetic or narrowly distributed photon beams following the exponential attenuation law, where the TVL is calculated as TVL = \ln(10) / \mu \approx 2.303 / \mu, with \mu denoting the linear attenuation coefficient. For beams obeying exponential attenuation, the TVL relates directly to the half-value layer (HVL) as TVL \approx 3.32 \times HVL, derived from the \ln(10) / \ln(2) \approx 3.3219; this holds exactly for monoenergetic narrow beams without significant scatter. The relation facilitates conversions between the two metrics when designing shields, though TVL values are tabulated separately to account for material-specific at given energies. TVL is preferred in radiation shielding calculations requiring substantial intensity reductions, such as factors exceeding 10 (e.g., multiple decades of ), where it simplifies determining total barrier thickness via logarithmic scaling: the number of TVLs needed is \log_{10}(I_0 / I), with I_0 and I as initial and final intensities. This approach has been a standard in since the mid-20th century, offering computational efficiency for broad-beam scenarios involving high-energy photons. Tabulated TVL data for common materials like , iron, and lead are widely used across photon energy ranges, enabling quick estimates without full simulations. Representative examples illustrate TVL's application: for cobalt-60 gamma rays (average energy ~1.25 MeV) in lead, the TVL is approximately 4.0 cm, compared to an HVL of 1.2 cm; in iron (or ), it is about 6.9 cm with an HVL of 2.1 cm; and in , around 37 cm with an HVL of approximately 11 cm, reflecting lower . These values underscore TVL's utility in scaling shield designs for decade-based reductions in high-energy environments.

Comparison with Other Attenuation Metrics

The half-value layer (HVL) is fundamentally linked to the (λ), a key metric in radiation physics that denotes the average distance a travels before interacting with in a medium. The is calculated as λ = 1/μ, where μ is the , providing insight into the probabilistic nature of interactions based on cross-sections. In comparison, the HVL represents the thickness required to attenuate the beam to 50% of its initial value, expressed as HVL = ln(2)/μ ≈ 0.693 λ. This makes the HVL shorter than the , as it accounts for the cumulative effect where half the photons are removed after traversing this distance, whereas λ describes the expected path to a single interaction. The relaxation length, another attenuation metric commonly employed in , is equivalent to the (λ = 1/μ) and corresponds to the distance over which beam intensity decreases to 1/e (approximately 37%) of its original value. Unlike the HVL, which offers a straightforward probabilistic halving for design purposes, the relaxation length emphasizes the characteristic and average interaction spacing, facilitating modeling of or gamma transport in complex systems. This distinction highlights the HVL's utility in empirical shielding assessments versus the relaxation length's role in theoretical particle diffusion simulations. Additional metrics include the quarter-value layer (QVL), defined as the material thickness that reduces to 25% of the initial value, calculated as QVL = ln(4)/μ ≈ 1.386/μ or approximately 2 HVL. In multilayer shielding, the concept of the homogeneous ray number—often interpreted as the equivalent number of HVLs traversed—enables additive calculations of total , where n HVLs reduce by a of 2^n. The HVL excels in intuitive applications for its halving simplicity and ease in , but it is less fundamental than the for deriving interaction probabilities from atomic cross-sections; conversely, the mean free path lacks the HVL's direct applicability to practical dose reduction estimates. For polychromatic or scattered beams, these metrics' interrelations break down due to energy-dependent , requiring spectrum-specific adjustments. HVL gained prominence in during the early as a standard for specifying beam penetrating power and shielding requirements, with early adoption in curve analyses by . In contrast, the originated in the quantum theoretical framework of the , building on photon-matter interaction models like to quantify microscopic attenuation processes.

References

  1. [1]
    [PDF] 32 - Shielding Radiation. - Nuclear Regulatory Commission
    A half value layer is the thickness of material that reduces the radiation intensity by one-half. Shielding Gamma Rays. 56. The following equation using HVL's ...
  2. [2]
    Half-Value Layer - Nondestructive Evaluation Physics : X-Ray
    The half-value layer (HVL) is the thickness of a material where 50% of the incident energy is attenuated, measured in distance (mm or cm).
  3. [3]
    Half-value layer | Radiology Reference Article - Radiopaedia.org
    Mar 15, 2025 · Half-value layer (HVL) is the thickness of a material required to reduce the air kerma of an X-ray or gamma-ray beam to half its original value.
  4. [4]
    Half-Value Layer (Shielding) - Radiation Safety - NDE-Ed.org
    The half-value layer (HVL) is the depth of a material where radiation intensity is half the surface value, and it's material and radiation energy dependent.
  5. [5]
    [PDF] Determination of the Tenth and Half Value Layer Thickness of ...
    The half value layers (HVL) and tenth value layers (TVL) are defined as the thickness of a shield or an absorber that reduces the radiation level by a factor ...<|control11|><|separator|>
  6. [6]
    Half-value layer | Radiology Reference Article - Radiopaedia.org
    Mar 15, 2025 · Half-value layer (HVL) is the thickness of a material required to reduce the air kerma of an X-ray or gamma-ray beam to half its original value.Question 623 · Question 744 · Question 2744Missing: origin | Show results with:origin
  7. [7]
    N.D. Admin Code 33.1-10-06-02 - Definitions | State Regulations ...
    "Half-value layer" means the thickness of specified material which attenuates x-radiation or gamma radiation to an extent such that the air kerma rate ...
  8. [8]
    Time-Line of the Evolution to the End of the 20th Century of Units ...
    Jul 12, 2016 · A time-line is presented of the evolution of X-ray units & quantities and X-ray measurements to the end of the 20th century.
  9. [9]
    Determination of half value layers of X-ray equipment using ...
    The HVL is the thickness of the aluminum (Al) absorber that is required to reduce the output to half of its initial value. The determination of these two ...
  10. [10]
  11. [11]
    Accuracy of measuring half‐ and quarter‐value layers and ... - NIH
    Jan 6, 2014 · Determination of the half‐value layer (HVL) and quarter‐value layer (QVL) values is not an easy task in X‐ray computed tomography (CT), ...
  12. [12]
    [PDF] BEAM CHARACTERIZATION FOR LOW-ENERGY X-RAYS ... - BIPM
    Half-Value Layer (HVL). The penetrating ability or 'quality' of a given x-ray beam is usually described by stating its Half-. Value Layer (HVL) for a chosen ...
  13. [13]
    Enhancing shielding efficiency of ordinary and barite concrete in ...
    The Half-Value Layer (HVL) and Tenth-Value Layer (TVL) depend on the energy of the radiation and material properties, such as density and atomic number. HVL ...
  14. [14]
    Beer-Lambert law | Radiology Reference Article - Radiopaedia.org
    Mar 25, 2025 · The Beer-Lambert law, also known as Beer's law, describes the relationship between the attenuation of light or other electromagnetic radiation ...
  15. [15]
    X-Ray Mass Attenuation Coefficients | NIST
    Sep 17, 2009 · The NIST database provides tables of X-ray mass attenuation coefficients (μ/ρ) and mass energy-absorption coefficients (μen/ρ) for elements Z=1 ...Water, Liquid · Air, Dry · Section 2 · Citation/Version History
  16. [16]
    Tutorial: Attenuation of X-Rays By Matter - Spectroscopy Online
    X-rays are attenuated as they pass through matter. That is, the intensity of an X-ray beam decreases the farther it penetrates into matter.
  17. [17]
    NIST: X-Ray Mass Attenuation Coefficients - Introduction
    Mass attenuation coefficients (μ/ρ) and mass energy-absorption coefficients (μen/ρ) are used to calculate photon penetration and energy deposition in materials.
  18. [18]
    [PDF] Review and history of photon cross section calculations* - SPRMN
    Jun 20, 2006 · Lambert (1760)–Beer (1852) exponential attenuation law. I(t) = I0 exp{−(µ/ρ)t}. (4) from which equation (1) follows. Calculations of photon ...
  19. [19]
    The exponential law | Radiology Key
    Mar 6, 2016 · The law is fundamental to an understanding of radioactive decay and the attenuation of certain electromagnetic radiations (eg X-rays and gamma-rays).
  20. [20]
    [PDF] Photon cross sections, attenuation coefficients, and energy ...
    ... Half-value layer= 0.69315/m, cm. Eq (1. —4b). = 0.69315/ (/i/p) g cm -2. Fig 1 ... exponential but is modified by two additional fac- tors. The first of ...
  21. [21]
    (PDF) Mathematical estimation of half‐value layer thicknesses
    ... half-value layer is given by halfvalue layer = 0,693/µ. Hubbell's mass absorption coefficient table was employed to compute the effective energy of the half ...
  22. [22]
    X-Ray Mass Attenuation Coefficients - Aluminum - NIST
    Aluminum Z = 13. HTML table format. Energy, μ/ρ, μen/ρ. (MeV), (cm2/g), (cm2/g) ... Aluminum Z = 13. ASCII format. Energy μ ...Missing: 100 | Show results with:100
  23. [23]
    Interaction with matter - Radiology Cafe
    Attenuated x-rays are those that are absorbed, transmitted with a lower energy or scattered. It is an exponential process and, therefore, the beam intensity ...Interactions With Matter · Compton Effect · Photoelectric Effect
  24. [24]
    [PDF] RESOURCE MANUAL for Compliance Test Parameters or ... - FDA
    Jul 15, 1999 · This manual covers compliance test parameters for diagnostic x-ray systems, including beam quality, entrance exposure rate, field limitation, ...
  25. [25]
    None
    ### Procedure for Measuring Half-Value Layer (HVL) per IEC 61267 Standard
  26. [26]
    [PDF] Gamma Ray Attenuation Properties of Common Shielding Materials
    Jun 18, 2018 · Half Value Layer (HVL).​​ Thickness of material required to reduce the intensity of radiation to one half of its original intensity (50% ...
  27. [27]
    [PDF] Beam Quality: Total Filtration and Half-Value Layer - CRCPD
    The concept of half-value layeror HVL is used to quantify the ability of an x-ray beam to penetrate the material being examined.Missing: HVL1 HVL2
  28. [28]
    Optimized Methodology for Accurate Half-Value Layer Measurement ...
    Results: Scattered radiation had significant impacts on the HVL measurements, particularly for the second HVL, resulting in measurement variations up to 1.8 ...
  29. [29]
    Calculation and comparison of mass attenuation coefficients of ...
    Mass attenuation coefficients (μ/ρ) of materials are used to calculate the half value layer (HVL) of X-ray radiation. They characterize how easily a mass of ...
  30. [30]
    Comparison between MCNP5, Geant4 and experimental data ... - NIH
    Aug 24, 2019 · Monte Carlo simulations, MCNP5 and Geant4 codes were developed to investigate radiation shielding properties of xPbO–(50-x) BaO–50B2O3 ...Missing: polychromatic | Show results with:polychromatic
  31. [31]
    Monte Carlo simulation of X-ray spectra in diagnostic radiology and ...
    The open-source object-oriented toolkit GEANT4 was used to simulate x-ray spectra in diagnostic radiology and mammography.
  32. [32]
    [PDF] Overview and Basis of Design for NCRP Report 147 - AAPM
    Jan 1, 2007 · NCRP Report 147 is about structural shielding design for X-ray imaging, a "best practice" for diagnostic x-ray devices, excluding dental and ...
  33. [33]
    Evaluation of X-Ray Beam Quality Based on Measurements and ...
    Therefore, determination of the half-value layer (HVL) is often used to describe the X-ray beam quality. The HVL of a beam is the thickness of material ...
  34. [34]
    Monte Carlo derivation of filtered tungsten anode X-ray spectra ... - NIH
    The greatest relative difference between the Monte Carlo calculated HVL values and experimental HVL values was 4%.Geometric Model · Figure 1 · Hvl Measurements
  35. [35]
    [PDF] Radiation Shielding Materials: Half-value layer determination for ...
    This study determined HVL for lead, concrete, water, polyethylene, and polyethylene mixtures with boric acid for neutron, photon, and simultaneous emissions. ...
  36. [36]
    An alternative method for calculation of half-value layers without the ...
    Half-value layer (HVL) is the thickness of a material required to reduce the air kerma of an x-ray or gamma ray to half its original value. It also determines ...
  37. [37]
    None
    ### Summary of Half-Value Layer (HVL) in Radiation Shielding from https://www.nrc.gov/docs/ML1122/ML11229A721.pdf
  38. [38]
  39. [39]
    NCRP-151: Linac Vault Shielding | Oncology Medical Physics
    NCRP-151 provides the standard formalism used in the United States to determine shielding requirements for megavoltage photon and electron beam sources.Missing: half- | Show results with:half-<|separator|>
  40. [40]
    21 CFR 1020.30 -- Diagnostic x-ray systems and their major components.
    ### Summary of Minimum Half-Value Layer (HVL) Requirements from Table 1 (21 CFR 1020.30)
  41. [41]
    [PDF] Performance Standard for Diagnostic X-ray Systems and - FDA
    Jun 10, 2005 · Half-value layer (HVL) means the thickness of specified material which attenuates the beam of radiation to an extent such that the exposure ...
  42. [42]
    [PDF] Quality control in diagnostic radiology - AAPM
    The purpose of QC testing is to detect changes that may result in a clinically significant degradation in image quality or a significant increase in radiation ...
  43. [43]
    [PDF] Handbook of basic quality control tests for diagnostic radiology
    This handbook is a compilation of tests that help to reveal key issues in the X ray units used in a diagnostic radiology department and provides a summary of ...
  44. [44]
    Filters | Radiology Reference Article - Radiopaedia.org
    Jan 18, 2025 · The change in the shape of the beam spectrum with filtration is referred to as beam hardening. This is due to the loss of lower energy photons ...
  45. [45]
    Breast Imaging Physics in Mammography (Part I) - PMC
    Oct 17, 2023 · Generally, additional filters used in mammography for Molybdenum targets are made of Molybdenum with a thickness of 0.03 mm, or for Rhodium ...
  46. [46]
    Radiographic Techniques, Contrast, and Noise in X-Ray Imaging
    Jan 23, 2015 · From a lesion visibility perspective, it is only the CNR that has to be increased, and increasing contrast is equivalent to reducing noise. The ...Missing: implications | Show results with:implications
  47. [47]
    [PDF] X-Ray measurements and protection, 1913-1964
    For35 years, he guided the Bureau's efforts in this area, a period during which a firm measurement base was constructed under his leadership.
  48. [48]
    Manual for Compliance Test Parameters of Diagnostic X-Ray Systems
    Dec 5, 2017 · The HVL of an x-ray beam is the thickness of absorbing material necessary to reduce the x-ray intensity to one-half its original value. By ...
  49. [49]
    [PDF] Safety Reports Series No.47
    TENTH VALUE LAYER (TVL) FOR 60Co AND X RAY. ENERGIES. (approximate values based on large attenuation). Co-60a. 4 MVb. 6 MVb. 10 MVb 15 MVb 18 MVb 20 MVb 24 MVb.
  50. [50]
    Introduction to Ionizing Radiation - OSHA
    Half-Value Layer (HVL): The shield thickness necessary to reduce intensity by half. HVL = 0.693 / µ. Similarly, a tenth-value layer reduces intensity by 10.
  51. [51]
    [PDF] Shielding - Gamma Constant - Point Source/Inverse Square
    HVL and TVL. ➢ The amount of shielding required to reduce the dose rate by 1/2 is called the half-value layer or HVL. ➢ HVLs for given materials are based ...
  52. [52]
    [PDF] Engineering Compendium on Radiation Shielding
    Analogous quantities are the half-value layer and one-tenth-value layer, I1/1 and tt/n0, in which the. 6. 'a. -a ½. ½ .6. ½. ½.. ½ a. Fig. 4.1.-1. Arrangement ...
  53. [53]
    Accuracy of measuring half‐ and quarter‐value layers and ...
    Jan 6, 2014 · For standard quality assurance testing and radiation dose reduction in patients, the half‐value layer (HVL) and quarter‐value layer (QVL) are ...Missing: shielding | Show results with:shielding