Relative fluorescence units

Relative fluorescence units (RFUs) are arbitrary, dimensionless units used to quantify the intensity of emitted by a sample in analytical techniques that rely on fluorescence detection, providing a relative measure proportional to the concentration of fluorophores or dye-labeled molecules. These units are generated by instruments such as spectrofluorometers, plate readers, or systems, where light at a specific induces emission that is detected and reported as RFUs, often normalized to factors like sample or optical for comparability. Unlike absolute fluorescence measurements, RFUs lack a universal and serve primarily as comparative values within an experiment or setup. RFUs find extensive application across scientific disciplines, particularly in and biochemistry, where they enable the monitoring of processes like (PCR) product accumulation in quantitative PCR (qPCR) or gene expression via reporter proteins such as (GFP). In enzyme kinetics assays, RFUs measure activities such as cleavage or reactions by tracking changes in fluorescence intensity over time. relies on RFU thresholds to distinguish true allelic peaks from background noise in short tandem repeat (STR) profiling electropherograms, ensuring reliable interpretation of genetic evidence. Additionally, in , RFUs assess algal pigments like in monitoring or detect bioavailable pollutants using biosensors, correlating signal intensity with contaminant levels.

Definition and Fundamentals

Definition of RFU

Relative fluorescence units (RFU) serve as an arbitrary measure to quantify the of emitted by fluorophores in biological assays, providing a relative scale rather than an absolute quantification tied to concentrations or standardized physical units. This unit is particularly useful in fluorescence-based detection systems, where it allows for straightforward comparisons between samples without requiring complex absolute calibrations. RFU values are derived from the output of detectors, which proportionally represent the collected emitted photons relative to a or within the . At its core, involves the of —molecules such as dyes or probes that absorb of at a specific , transitioning to a higher-energy . During the brief excited-state lifetime (typically 1–10 nanoseconds), the molecule relaxes, and a of lower energy (longer ) is emitted as the returns to the , producing the observable signal. This process, known as the , separates the and emission , enabling selective detection and minimizing from the source. In digital fluorescence detectors, such as photomultiplier tubes (PMTs) or charge-coupled devices (CCDs), the emitted intensity is converted into an electrical signal that is linearly scaled to generate RFU values, ensuring that the measurement reflects proportional changes in fluorophore concentration or activity. Unlike absolute fluorescence units, which might quantify emission in terms of photons per second or moles of fluorophore, RFU emphasizes practicality for intra- and inter-experiment comparisons in laboratory settings, where instrument-specific variations and sample matrix effects make absolute measurements challenging. This relative nature facilitates rapid analysis in assays like quantitative PCR, where RFU tracks signal changes over cycles without needing universal standardization. RFU emerged in the 1990s alongside automated fluorescence-based instruments, such as the ABI PRISM series sequencers, which revolutionized DNA analysis by replacing radioactive labeling with multicolor fluorescent detection for high-throughput applications.

Units and Calibration

In real-time (qPCR), relative fluorescence units (RFU) are calibrated using passive reference dyes, such as ROX (carboxy-X-rhodamine), to fluorescence signals and account for instrument-specific variations including differences in power, detector sensitivity, and . Calibration methods for RFU vary by application; for example, in , internal size standards are used for . ROX is an inert dye that emits a constant fluorescence signal throughout the , independent of PCR amplification, allowing the reporter dye signal to be divided by the ROX signal for . This process ensures reproducibility across experiments and platforms by compensating for well-to-well or run-to-run variations in fluorescence detection. The relative nature of RFU arises from the calculation, typically expressed as: \text{RFU} = \frac{\text{sample fluorescence} - \text{background}}{\text{scaling factor}} where the scaling factor adjusts the values to fit within the detector's , often 0–4000 RFU in many systems, though this can vary. In qPCR contexts, this often involves normalized reporter values (Rn) derived as the reporter fluorescence divided by the ROX reference, followed by subtraction of fluorescence to yield ΔRn, reported in arbitrary RFU. Instrument-specific scales are implemented in software for post-acquisition. For example, in Thermo Fisher Scientific's QuantStudio systems, RFU values are arbitrary units adjusted via ROX to span a typical range up to several thousand, ensuring consistent reporting across runs. Similarly, platforms, such as the 3500 Genetic Analyzer for fragment analysis, report RFU in arbitrary units after software-based scaling, with optimal signals often between 375 and 20,800 RFU to avoid saturation or noise. Baseline subtraction is crucial in RFU calibration to isolate the true amplification signal from inherent noise or non-specific background fluorescence, typically performed by averaging early-cycle fluorescence (before amplification) and subtracting it from all subsequent measurements. This step enhances accuracy, particularly in low-signal regions, and is automatically handled in instrument software to prevent overestimation of RFU values.

Measurement Techniques

Fluorescence Detection in PCR

Fluorescence detection in polymerase chain reaction (PCR) relies on the integration of fluorescent probes or dyes that bind to amplified DNA, enabling the measurement of relative fluorescence units (RFU) to monitor amplification in real time. Common probes include TaqMan hydrolysis probes, which consist of a fluorophore and quencher linked by an oligonucleotide sequence complementary to the target DNA; during PCR, the Taq polymerase's 5' nuclease activity cleaves the probe, separating the fluorophore from the quencher and generating a fluorescent signal proportional to the amount of amplicon produced. Another widely used method employs intercalating dyes like SYBR Green, which binds nonspecifically to double-stranded DNA and exhibits a substantial increase in fluorescence (up to 1,000-fold) upon binding, allowing detection of any amplified product without sequence-specific probes. These approaches ensure that fluorescence intensity, reported in RFU, correlates directly with the quantity of target DNA synthesized during each cycle. Real-time monitoring occurs at the end of each , typically during the annealing or extension , where optical systems in quantitative PCR (qPCR) instruments excite the fluorophores with specific wavelengths and detect emitted light to record RFU values. These readings are plotted as curves, with RFU on the y-axis and cycle number on the x-axis; the curves typically show an initial of minimal signal change, followed by an log where RFU rises sharply as amplicons accumulate, and a plateau where reagent limitations halt further increase. The reflects the core dynamics, enabling precise tracking of product formation. RFU measurements in this context facilitate the determination of the threshold (), the point where exceeds a predefined , which is crucial for quantifying initial template amounts. The underlying the amplification curve approximates RFU as a function of cycle number, fluorescence, , and initial concentration. Specifically, the signal can be expressed as: \text{RFU}_{\text{cycle}} = \text{RFU}_{\text{baseline}} + k \cdot N_0 \cdot (1 + E)^{\text{cycle}} where N_0 is the initial amount, E is the (ideally approaching 1 for 100% ), and k is a proportionality constant linking DNA quantity to detectable ; this highlights how RFU quantifies the phase and informs Ct calculation for relative quantification. qPCR instruments incorporate advanced optical systems, such as LED or excitation sources paired with dichroic mirrors, bandpass filters for excitation and wavelengths, and photodetectors (e.g., or tubes), to enable multi-wavelength detection in multiplex assays. These components allow simultaneous monitoring of multiple fluorophores with distinct spectra, such as FAM (excitation ~495 nm, ~520 nm) for one target and ROX (excitation ~585 nm, ~605 nm) as a passive reference for , ensuring accurate RFU readings across channels without . This hardware setup supports high-throughput, sensitive detection essential for real-time applications.

Data Acquisition in Electrophoresis

In capillary electrophoresis for DNA fragment analysis, fluorescently labeled DNA fragments, typically tagged with dyes such as 6-FAM or VIC at the 5' end of primers, are separated by size as they migrate through a polymer-filled capillary under an applied electric field. As the fragments pass the detection window, an excitation laser—often an argon-ion laser operating at 488 nm—induces fluorescence from the dyes, which is captured by a charge-coupled device (CCD) camera via a spectrograph and diffraction grating to separate wavelengths. This raw fluorescence signal is processed in real-time, yielding relative fluorescence units (RFU) that quantify the intensity of emission for each fragment. Software such as GeneMapper then converts these fluorescence data into an electropherogram, where peaks correspond to DNA fragments, with the x-axis representing fragment size in base pairs (bp) and the y-axis indicating RFU values that reflect allele abundance through peak height or area. The RFU scale is relative, normalized to internal standards to account for instrument variability, ensuring comparable measurements across runs. Accurate resolution of peaks requires of migration times, achieved by including size standards like GeneScan ROX, which contain known-length fragments labeled with a distinct (e.g., ROX) to generate a standard curve for aligning sample peaks based on relative migration speeds. This corrects for factors such as temperature, voltage fluctuations, and , enabling precise sizing within a typical range of 35–500 bp. Multiplexing allows simultaneous detection of multiple DNA targets by assigning different dyes to distinct color channels (e.g., blue for 6-FAM, green for VIC), with software applying a multicomponent matrix to deconvolute spectral overlaps and produce separate RFU traces for each channel. This approach supports efficient analysis of complex samples, such as those with multiple loci, while maintaining channel-specific RFU quantification.

Applications in DNA Analysis

Quantitative PCR (qPCR)

In quantitative PCR (qPCR), relative fluorescence units (RFU) quantify the real-time accumulation of amplified nucleic acids via intercalating dyes like SYBR Green or sequence-specific probes such as TaqMan, where fluorescence intensity correlates with product formation during each cycle. Amplification curves are generated by subtracting baseline fluorescence from raw RFU data to yield normalized values (ΔRn), and the cycle threshold (Ct) is calculated as the cycle number where the curve intersects a fixed threshold, typically set at 0.1–0.2 ΔRn above the baseline within the exponential phase. This threshold ensures reliable detection in the linear amplification region, avoiding noise from early cycles or plateau effects; a lower Ct value signifies a higher initial template concentration, as amplification reaches detectability sooner. Efficiency of qPCR reactions is assessed using standard curves constructed from serial dilutions of known template quantities, plotting Ct values against the logarithm of starting template concentration (log₁₀[N₀]); the slope of this linear regression determines amplification efficiency via the equation E = 10^{-1/\text{slope}} - 1, with ideal values of 90–110% (slope ≈ –3.1 to –3.6) indicating consistent doubling per cycle. While endpoint RFU can provide a proxy measure of final product yield proportional to initial template under optimal conditions, Ct-based standard curves are preferred for accuracy, as they account for kinetic differences before plateauing. Adherence to these metrics ensures reproducible quantification, as outlined in the MIQE guidelines for qPCR reporting. RFU-derived Ct values enable key applications in nucleic acid analysis, such as and quantification, where relative abundance is compared across samples. In studies, the ΔΔCt method normalizes target Ct to a reference within samples (ΔCt) and then to a calibrator sample (ΔΔCt), yielding fold changes via $2^{-\Delta\Delta C_t} assuming near-100% efficiency; this approach has become for assessing transcriptional differences in biological contexts like disease progression. For measurement, absolute quantification uses Ct against a curve calibrated to copy numbers, allowing precise monitoring of pathogens such as or in clinical samples. Post-amplification melt curve analysis further validates qPCR specificity by monitoring RFU as temperature increases from 60–95°C, revealing DNA dissociation profiles; a single sharp peak at the amplicon's melting temperature (Tm) confirms product homogeneity, while multiple or broad peaks indicate non-specific amplification like primer-dimers. This RFU-based derivative plot (–d(RFU)/dT vs. T) distinguishes true targets from artifacts, enhancing data reliability without additional gels, as recommended in qPCR best practices.

Short Tandem Repeat (STR) Profiling

In short tandem repeat () profiling, relative fluorescence units (RFU) quantify the intensity of fluorescent signals from PCR-amplified DNA fragments separated by , enabling the detection and of s at specific loci. Fluorescently labeled primers are used to amplify targeted STR loci, such as the 20 core CODIS loci established by the FBI, producing distinct peaks on electropherograms where peak height in RFU corresponds to allele dosage for distinguishing heterozygotes (two s with roughly equal RFU heights) from homozygotes (a single peak approximately twice the height of a heterozygote ). For accurate allele calling, peak height ratios (PHR) between the two alleles at a heterozygous locus are evaluated; balanced profiles typically exhibit PHR values of 60% or greater, with the smaller peak's RFU height being at least 60% of the larger peak's height, while ratios below this threshold may signal allelic imbalance due to DNA mixtures, degradation, or stochastic effects like allele dropout. Stutter artifacts, common in STR PCR amplification, manifest as minor peaks approximately 4 base pairs shorter than the parent allele due to slippage during DNA synthesis, with their RFU heights generally comprising less than 15% of the main allele peak and rarely exceeding 16%; these artifacts aid in mixture deconvolution by helping differentiate true minor contributor alleles from non-allelic PCR by-products.00003-8/fulltext) Forensic standards, as outlined in guidelines from the Scientific Working Group on DNA Analysis Methods (SWGDAM), recommend laboratory-specific validation of RFU thresholds, with many facilities setting minimum analytical thresholds for reportable alleles between 50 and 100 RFU to ensure reliable detection above baseline noise, though stochastic thresholds for interpreting low-level alleles are often higher (e.g., 150-200 RFU) to minimize false positives.

Interpretation and Standards

RFU Peak Thresholds

In forensic DNA analysis, the analytical threshold (AT) represents the minimum relative fluorescence unit (RFU) value above which a peak can be reliably distinguished from baseline noise, typically set in the range of 50-300 RFU varying by instrument and laboratory validation, such as 35-50 RFU for ABI 310 Genetic Analyzer and higher (e.g., ~350 RFU) for ABI 3500 series. This threshold is calculated as 3 to 5 times the standard deviation of the noise measured from negative control samples, ensuring a low probability (e.g., 10^{-2} to 10^{-6}) of false positives from instrumental artifacts or background signals. For instance, on instruments like the ABI 310 Genetic Analyzer, the AT is often estimated at approximately 35-50 RFU using 3 times the standard deviation or 2 times the peak-to-peak noise from replicate analyses. The stochastic threshold (ST), set higher at approximately 100-400 RFU depending on the instrument, such as 150-200 RFU for ABI 310 and ~750 RFU for ABI 3500 series, addresses challenges in low-template DNA samples where stochastic effects such as allele dropout or single-molecule drop-in contamination become significant. This threshold ensures that peaks exceeding it are less likely to result from random amplification inefficiencies or contaminant introduction, thereby supporting reliable genotyping in scenarios with limited input DNA, such as trace evidence. Laboratories determine the ST empirically through dilution series of known single-source samples, aiming for a confidence level where allelic dropout probability falls below 10^{-2} to 10^{-6}, often aligning with peak height ratios above 60% for heterozygotes. Reportable thresholds in short (STR) profiling are laboratory-specific and generally range from 50-200 RFU, below which peaks are not interpreted as true alleles to minimize erroneous inclusions in profiles. These thresholds guide the final reporting of genotypes, particularly in forensic contexts, where peaks under this level may indicate , stutter, or non-contributory artifacts rather than biological signal. Validation studies across labs confirm this range balances with specificity, often aligning the reportable threshold with or above the ST for conservative . Thresholds can be fixed or dynamic, with software like GeneMapper allowing adjustments based on run-specific parameters such as electrokinetic injection time or sample concentration to optimize peak detection across varying signal intensities. Fixed thresholds apply uniform RFU cutoffs per dye or locus, suitable for standardized high-quality samples, while dynamic approaches scale thresholds proportionally—for example, increasing from 75 RFU at 7-second injections to 100 RFU at 15-second injections—to account for enhanced signal strength without saturation. This flexibility, validated in instrument-specific protocols, improves allele recovery while reducing pull-up or off-scale artifacts in electrophoresis data.

Factors Influencing RFU Values

Several factors influence the accuracy and variability of relative fluorescence units (RFU) measurements in DNA analysis, stemming from sample characteristics, technical parameters, and environmental conditions. These variables can lead to inconsistent peak heights, reduced sensitivity, or imbalanced profiles across loci, necessitating careful control to ensure reliable interpretation. Sample-related factors, particularly DNA quantity and quality, profoundly affect RFU values. Degraded DNA, often resulting from post-mortem intervals or environmental exposure, leads to a statistically significant decrease in RFU, as fragmentation reduces efficient amplification and detection of shorter loci preferentially. For instance, in cadaveric samples, longer intervals between death and collection correlate with lower RFU, indicating impacts signal intensity. Inhibitors co-extracted from biological matrices, such as in , suppress by dyes like ROX to less than 50% intensity at concentrations equivalent to 5% , thereby reducing overall RFU and potentially causing allele dropout in forensic . Technical parameters in electrophoresis and amplification also contribute to RFU variability. In capillary electrophoresis, injection voltage and time directly influence the amount of DNA loaded, with higher settings (e.g., increased voltage and extended time) enhancing peak heights up to six-fold in low-level samples, though excessive injection can lead to saturation and off-scale RFU. Imbalances in dye or primer concentrations across loci exacerbate uneven RFU distribution; for example, deviations of ±20% from optimized primer levels in multiplex kits maintain full profiles but slightly lower peak heights and disrupt intracolor balance, affecting heterozygote peak ratios. Environmental factors introduce additional variability in RFU detection. Temperature fluctuations during alter DNA migration due to changes in buffer and , leading to shifts in , band broadening, and imprecise peak calling that can indirectly impact RFU interpretation. of fluorophores, induced by prolonged laser excitation generating , degrades signal intensity over time, reducing RFU in extended detection runs, particularly for sensitive dyes like Cy5. To mitigate RFU variability, normalization using internal standards, such as fluorescent size ladders (e.g., LIZ-500), standardizes peak heights and sizing across runs, ensuring consistent RFU scaling. Replicate amplifications and runs, typically in triplicates across serial dilutions, allow assessment of variability via (CV), with acceptable reproducibility often below 20% CV for peak heights to validate profile reliability.

References

  1. [1]
    Relative Fluorescence Units - an overview | ScienceDirect Topics
    Relative fluorescence units (RFUs) are defined as the measurements of emitted light energy that are proportional to the amount of light emitted by ...
  2. [2]
    Fluorescence Intensity Measurements | BMG LABTECH
    Consequently, fluorescence plate readers measure the light signal emitted by a sample in Relative Fluorescent Units (RFU).
  3. [3]
    Development and Characterization of a Green Fluorescent Protein ...
    Relative fluorescence unit (RFU) is defined as the culture fluorescence relative to culture biomass at OD660. Induction ratio is defined as the RFU of an ...<|control11|><|separator|>
  4. [4]
    Analytical Threshold | NIST
    Jun 12, 2023 · 2) A 'Relative Fluorescence Units' (RFU) level determined to be appropriate for use in the PCR/STR DNA typing process; a minimum threshold ...
  5. [5]
    What are these measurements? - Real-time water quality - USGS.gov
    Measurement units include relative fluorescence units (RFU) and micrograms per liter (ug/L). Chlorophyll sensor data provides information on relative ...
  6. [6]
    Fluorescence Fundamentals | Thermo Fisher Scientific - US
    The fluorescence excitation spectrum of a single fluorophore species in dilute solution is usually identical to its absorption spectrum. The absorption spectrum ...The Fluorescence Process · Fluorescence Detection · Fluorescence Output of...
  7. [7]
    The sequence of sequencers: The history of sequencing DNA - PMC
    This article traverses those years, iterating through the different generations of sequencing technology, highlighting some of the key discoveries.
  8. [8]
    The Purpose of ROX Reference Dye in Real-Time PCR (qPCR)
    Jun 3, 2021 · To calculate the normalized values (Rn), the reporter dye signal is divided by the ROX reference dye signal.
  9. [9]
    Normalization of Real-Time PCR Fluorescence Data with ROX ...
    ROX passive reference dye is an inert additive that provides a constant fluorescent signal for sample normalization throughout the real-time qPCR assay.
  10. [10]
    Real-Time PCR: Understanding Ct | Thermo Fisher Scientific - ES
    ROX passive reference dye. The Rn value is calculated as the ratio of the fluorescence of Applied Biosystems™ FAM™ Dye divided by the fluorescence of ROX dye.
  11. [11]
    [PDF] Using Passive Reference Dyes for Normalization and ... - Illumina
    Passive reference dyes are commonly used in qPCR reactions to normalize for non-PCR related fluorescence signal variation. Because.
  12. [12]
    Hello, could somebody explain the difference between relative ...
    Jun 3, 2022 · Fluorescence intensity measurements are reported as relative fluorescence units (RFU). The size of the numbers has no real meaning. They are arbitrary.What is the meaning of Relative Fluorescence Unit (RFU) at a ...How can I convert RFU units to fluorescence intensity? | ResearchGateMore results from www.researchgate.net
  13. [13]
    What is the optimal signal range for capillary electrophoresis of ...
    Probe Signal Range (RFU) Without Warnings*. Applied Biosystems, SeqStudio Genetic Analyzer, 375–20,800. SeqStudio Flex Series Genetic Analyzer, 375–20,800. 3500 ...
  14. [14]
    [PDF] Applied Biosystems 3500/3500xL Genetic Analyzer User Guide
    • Fragment: > 175 RFU. • HID: > 400 RFU. Sizing Precision. Standard ... • All Sizes for the software to analyze fragments of all sizes in the Analysis Range.
  15. [15]
    Pairwise efficiency: a new mathematical approach to qPCR data ...
    May 30, 2019 · Fluorescence data in Additional file 2: Dataset 1 are expressed as RFU (Relative Fluorescence Units) which is a term specific to Bio-Rad ...
  16. [16]
    Essentials of Real-Time PCR | Thermo Fisher Scientific - ES
    In the initial cycles of PCR, there is little change in fluorescence signal. This defines the baseline for the amplification plot. An increase in fluorescence ...
  17. [17]
    SYBR® Green for qPCR - Bio-Rad
    SYBR® Green fluorescence increases up to 1,000-fold upon intercalation with dsDNA. As amplification proceeds, the amount of DNA product increases, and hence the ...
  18. [18]
  19. [19]
  20. [20]
    CFX96 Touch Real-Time PCR Detection System - Bio-Rad
    This six-channel (five colors and one FRET channel) real-time PCR instrument combines advanced optical technology with precise temperature control.
  21. [21]
    [PDF] DNA Fragment Analysis by Capillary Electrophoresis User Guide ...
    For Research Use Only. Not intended for use in diagnostic procedures. DNA Fragment Analysis by Capillary Electrophoresis. Publication Number 4474504.
  22. [22]
  23. [23]
    Accurate DNA Fragment Sizing by Capillary Electrophoresis ... - NIH
    We report a new method for DNA fragment sizing using capillary electrophoresis with laser-induced fluorescence detection (CE-LIF).
  24. [24]
    [PDF] Evaluating analysis settings and assay verification for the TaqCheck ...
    Example Ct value distribution for expected RNase P–negative samples, using a ΔRn threshold of 0.2 and a Ct cutoff at cycle 32 for. RNase P. Cutoff and ...
  25. [25]
    [PDF] A Practical Approach to RT-qPCR — Publishing Data That Conform ...
    RFU, relative fluorescence units. Page 4. © 2015 Bio-Rad Laboratories, Inc ... earlier provide CT values for threshold cycle; however, MIQE guidelines ...
  26. [26]
    Minimum Information for Publication of Quantitative Real-Time PCR ...
    MIQE is a set of guidelines that describe the minimum information necessary for evaluating qPCR experiments. Included is a checklist to accompany the initial ...Abstract · Conceptual Considerations · Research vs Diagnostic... · qPCR
  27. [27]
    Efficiency of Real-Time PCR | Thermo Fisher Scientific - US
    The ideal structure for a standard curve intended to assess efficiency is 7-points with a 10-fold series (see Amplification Efficiency of TaqMan Gene Expression ...
  28. [28]
    Analysis of relative gene expression data using real-time ... - PubMed
    The 2(-Delta Delta C(T)) method is a convenient way to analyze the relative changes in gene expression from real-time quantitative PCR experiments.
  29. [29]
    Analysis of Relative Gene Expression Data Using Real-Time ...
    The 2 −ΔΔC T method is a convenient way to analyze the relative changes in gene expression from real-time quantitative PCR experiments.
  30. [30]
    Explaining multiple peaks in qPCR melt curve analysis | IDT
    Feb 2, 2023 · This program predicts melt curves and their derivatives for qPCR-length amplicons and is ideally suited to test for multiple peaks in a single ...
  31. [31]
    CODIS and NDIS Fact Sheet - FBI
    Jan 1, 2017 · For Forensic STR DNA analysis, the DNA profile consists of one or two alleles at the 20 CODIS Core Loci. 8. Is any personal information ...
  32. [32]
    STR Amplification | STR Analysis | STR kits - Promega Corporation
    STR loci are amplified using fluorescently labeled PCR primers that flank the hypervariable regions. One of the biggest strengths of PCR-based DNA typing is ...
  33. [33]
    [PDF] STR Results Interpretation - Identifiler and Yfiler - NYC.gov
    Jul 11, 2016 · data – 75 Relative Fluorescent Units (RFU's) – are labeled as alleles. A. Computer program processing steps for raw data: 1. Recalculating ...
  34. [34]
    [PDF] Procedure for Analysis and Interpretation of STR DNA Profiles 1.0 ...
    Jan 7, 2020 · heterozygote peak height ratios (PHR) greater than or equal to 60 % when both peaks are at/above the stochastic threshold. Low PHR may ...
  35. [35]
    STR Data Analysis and Interpretation for Forensic Analysts | Stutter
    Jul 11, 2023 · Stutter is a by-product of the amplification of STR loci whereby a minor product, typically one repeat smaller than the primary allele, is generated.
  36. [36]
    STR Data Analysis and Interpretation for Forensic Analysts
    Jul 11, 2023 · Only if the peak height, expressed in relative fluorescence unit (RFU), exceeds a standard value will it be accepted. National Institute of ...
  37. [37]
    [PDF] Fundamentals of CE-Based DNA Analysis and Related ...
    Thus for the 3500 one could set a stochastic threshold at 750 RFU and an analytical threshold based on the MDL of 350. The Limit of linearity is around 20,000 ...
  38. [38]
    How to Optimize Forensic Genotyping Settings for Accurate DNA ...
    Capillary Electrophoresis (CE) is a crucial tool in forensic DNA analysis, providing accurate methods for DNA profiling from crime scene samples. By separating ...Missing: acquisition | Show results with:acquisition
  39. [39]
    None
    ### Summary of Standards for Analytical and Stochastic Thresholds in Forensic DNA
  40. [40]
    The paradigm shift in DNA profile interpretation - ScienceDirect
    For a modern 3500 capillary electrophoresis (CE) machine the AT is often of the order of 50–200 rfu. Peaks can be up to 30,000 rfu in height. This suggests that ...
  41. [41]
    [PDF] Sample & Assay Technologies Investigator® DIPplex Handbook
    Thresholds are usually 50–200 RFU and should be determined individually by the laboratory. Recommendation: The minimal peak height should be 3 times as high ...
  42. [42]
    [PDF] 3500 Genetic Analyzer with a comparison of the Identifiler
    Conclusion: Analytical thresholds of 75 RFU for the 7s injection and 100 RFU for the 15s injection were chosen for the PP16HS amplification chemistry.
  43. [43]
    [PDF] Best Practice Recommendations for Internal Validation of Human ...
    This document outlines best practice recommendations for the internal validation of human short tandem repeat DNA profiling on capillary electrophoresis ...
  44. [44]
  45. [45]
  46. [46]
  47. [47]