Saturated calomel electrode

The saturated calomel electrode (SCE) is a reference electrode widely used in electrochemistry for providing a stable and reproducible potential in potentiometric measurements, consisting of a pool of elemental mercury in contact with a paste of mercury(I) chloride (Hg₂Cl₂, also known as calomel) immersed in a saturated aqueous solution of potassium chloride (KCl). Its half-cell reaction is Hg₂Cl₂(s) + 2e⁻ ⇌ 2Hg(l) + 2Cl⁻(aq), which establishes a fixed chloride ion activity due to the saturation of KCl, ensuring consistent electrode potential. The SCE exhibits a standard potential of +0.2444 V versus the standard hydrogen electrode (SHE) at 25°C, making it a reliable benchmark for calibrating indicator electrodes in analytical techniques such as pH measurements and voltammetry. Structurally, the SCE typically features an inner compartment containing the mercury-calomel paste connected via a porous or to an outer saturated KCl solution, which acts as a to minimize liquid junction potentials when coupled with the sample solution. This design allows for electrical contact through a wire embedded in the mercury pool, facilitating its integration into three-electrode systems. The saturation with KCl not only stabilizes the Cl⁻ concentration but also makes the electrode resistant to potential shifts from evaporation or minor volume changes in the . Despite its advantages, the SCE has limitations, including temperature sensitivity—its potential decreases to approximately +0.2376 V at 35°C due to increased KCl solubility—and potential contamination of the sample with leaking KCl, which can interfere with chloride-sensitive analyses (often mitigated by double-junction configurations). Additionally, concerns over mercury toxicity have led to alternatives like the silver/silver chloride electrode in modern applications, though the SCE remains a standard in many laboratory settings for its historical reliability and ease of preparation.

Composition and Construction

Electrode Components

The saturated calomel electrode (SCE) consists of a primary mercury pool overlaid with a paste of (Hg₂Cl₂) that acts as the contact layer for the couple, a saturated (KCl) solution, and a porous or that facilitates ionic contact with the external solution while minimizing contamination. Physically, the electrode is typically constructed within a glass tube body, often in an "H" shape or with a bent side arm to allow for the addition and saturation of KCl crystals, ensuring a constant chloride ion activity. The mercury pool is placed at the bottom of the wider arm, covered by a thin layer of calomel paste (approximately 5 mm thick), while the saturated KCl solution fills the remaining space up to near the top, with a platinum wire immersed in the mercury for external electrical connection. The porous junction, such as a Vycor® frit or ceramic fiber, is positioned at the tip to permit ion diffusion without allowing bulk mixing of solutions, thereby maintaining the electrode's internal stability. Standard dimensions include a tube length of about 10-15 cm and a diameter of 1-2 cm for the main body, though variations exist for specific applications. To ensure reliable performance, the mercury must be triple-distilled to remove impurities that could alter the , while the KCl is prepared from reagent-grade material to achieve with minimal contaminants, often involving baking at high temperatures to eliminate organics. A stopper or cap on the side arm allows periodic addition of KCl crystals to maintain as the is used.

Preparation Procedure

The preparation of a saturated calomel electrode (SCE) involves careful assembly to ensure electrical contact and stability, typically using a glass tube or body with a porous junction such as a ceramic frit or cracked bead for electrolyte diffusion. For laboratory-constructed electrodes, begin by purifying mercury through passage as a fine spray through dilute nitric acid, allowing it to settle, and then washing it repeatedly with distilled water followed by 1 N KCl solution to remove impurities. Next, prepare the calomel paste by adding concentrated hydrochloric acid to a solution of mercurous nitrate to precipitate Hg₂Cl₂ crystals, incorporating excess mercury (approximately 20 g), heating the mixture on a steam bath with stirring, and washing the resulting paste with distilled water before filtering by suction. Assemble the electrode by placing a layer of purified mercury (about 3/4 inch deep) at the bottom of the electrode tube, followed by an approximately 1-inch layer of the calomel-mercury paste to form the Hg/Hg₂Cl₂ interface; fill the remaining space with saturated KCl solution to ensure full saturation with calomel. For commercial electrodes, such as those with liquid-filled designs, remove any protective cap or insert, then add saturated KCl electrolyte (e.g., SP138 solution) via the fill hole until the level is about 1/4 inch below the cap, ensuring the internal element remains submerged. Install or verify the salt bridge or frit junction by checking for electrolyte flow: hold the electrode at a 45° angle, gently squeeze a fill bottle into the hole to form a liquid bead at the junction (up to 30 seconds for ceramic types), and repeat if necessary to clear any blockages. Once assembled, store the SCE in saturated KCl solution to maintain hydration of the and prevent drying, which can increase potential and degrade performance; for long-term storage, position it upright with a protective cap to minimize leakage. Periodic includes visually inspecting the level and replenishing with fresh saturated KCl if low, as well as checking for any precipitation buildup that might clog the —regenerate by soaking the tip in warm water for five minutes or longer until flow resumes, followed by a rinse. If the becomes irreversibly clogged, replace it or soak in slightly less than saturated KCl for at least one hour to re-wet thoroughly before reuse. Safety protocols are essential due to the toxicity of mercury and mercurous chloride; perform all preparation steps in a to avoid inhalation of mercury vapors, wearing gloves, safety goggles, and a lab coat to prevent contact, as mercury can be absorbed through the skin and cause neurological damage. In case of spills, use a mercury spill kit to contain and collect droplets without spreading, then ventilate the area thoroughly. Disposal must comply with EPA regulations under the (RCRA), which as of requires packaging mercury-containing waste in leak-proof containers labeled as hazardous, transporting it to a certified recycler or consolidation site, and avoiding releases into the environment to prevent in ecosystems.

Electrochemical Mechanism

Half-Cell Reaction

The saturated calomel electrode (SCE) functions as a in electrochemical cells by providing a stable half-cell potential through a well-defined , completing the circuit with the indicator to measure potential differences. This stability arises from the reversible reaction at the , where electrons are transferred without significant net change under standard conditions. The fundamental half-cell reaction for the SCE is the reduction of mercurous (calomel) to mercury metal, accompanied by the release of ions: \ce{Hg2Cl2(s) + 2e^- ⇌ 2Hg(l) + 2Cl^-(aq)} In this reaction, solid mercurous (\ce{Hg2Cl2}) serves as the oxidant, while liquid mercury (\ce{Hg}) acts as the reductant, with ions (\ce{Cl^-}) participating in the aqueous phase. The electrode is constructed such that a paste of finely divided mercury and is in intimate contact, forming the interface where occurs reversibly. This paste is immersed in a saturated (KCl) solution, which supplies the ions essential for the reaction. The saturation of KCl ensures equilibrium conditions by maintaining a constant activity of chloride ions through the presence of undissolved KCl , preventing variations in ion concentration that could shift the . This constant \ce{Cl^-} activity stabilizes the , as the reaction depends on the solubility and ionization of , which is governed by the concentration in . A porous or at the electrode's tip allows ionic conduction to the external while minimizing mixing, thus preserving the internal .

Potential Calculation

The potential of the saturated calomel electrode is derived from the applied to its half-cell reaction, which serves as the foundational electrochemical process. For the reduction half-reaction \ce{Hg2Cl2(s) + 2e^- ⇌ 2Hg(l) + 2Cl^-(aq)}, the at 25°C takes the form E = E^\circ + \frac{0.059}{2} \log \left( \frac{1}{[\ce{Cl^-}]^2} \right), where E^\circ is the standard reduction potential relative to the standard hydrogen electrode (SHE), and concentrations approximate activities under standard conditions. This simplifies to E = E^\circ - \frac{0.059}{1} \log [\ce{Cl^-}], highlighting the direct dependence on chloride concentration. The activities of solid \ce{Hg2Cl2} and liquid Hg are unity, eliminating their influence on the potential. In the saturated calomel electrode, the is a saturated KCl , which maintains a fixed concentration of approximately 4.6 M at 25°C, ensuring the potential remains constant regardless of minor dilution. This saturation stabilizes the activity, with the assumption that activity coefficients are near unity in the concentrated for theoretical derivation purposes. An equivalent derivation considers the in terms of the mercurous , \ce{Hg2^2+ + 2e^- ⇌ 2Hg(l)}, with the E = E^\circ_{\ce{Hg2^2+/Hg}} + \frac{0.059}{2} \log [\ce{Hg2^2+}] (again approximating activities with concentrations). The mercurous concentration is linked to the \ce{Hg2Cl2(s) ⇌ Hg2^2+(aq) + 2Cl^-(aq)}, governed by the solubility product K_{sp} = [\ce{Hg2^2+}] [\ce{Cl^-}]^2. Thus, [\ce{Hg2^2+}] = K_{sp} / [\ce{Cl^-}]^2, and substitution yields E = E^\circ_{\ce{Hg2^2+/Hg}} + \frac{0.059}{2} \log K_{sp} - 0.059 \log [\ce{Cl^-}], demonstrating that the potential depends only on the fixed chloride concentration under saturation, with the K_{sp} term absorbed into the effective standard potential relative to the SHE. This formulation underscores how the low solubility of calomel maintains negligible free \ce{Hg2^2+} while the high, constant [\ce{Cl^-}] ensures potential stability.

Standard Potential Value

Measured Value and Factors

The standard potential of the saturated calomel electrode (SCE) is +0.2444 V versus the (SHE) at 25°C when filled with saturated KCl solution. This value arises from the fixed chloride ion activity in the saturated electrolyte and has been empirically verified through precise measurements in controlled cells without liquid junctions. The chloride concentration significantly influences the , as it determines the in the half-cell reaction. For instance, a 1 M KCl calomel electrode exhibits +0.280 V vs. SHE, while a 0.1 M KCl version shows +0.336 V vs. SHE at 25°C, highlighting the stabilizing role of in maintaining consistent performance. Liquid junction potentials, which arise from ion diffusion across the interface, are minimized in SCE designs through the use of a frit or porous barrier, ensuring potential stability within ±1 mV under typical conditions. In neutral media, the SCE potential remains insensitive to pH variations, as the Hg/Hg₂Cl₂ redox couple does not involve H⁺ ions. Calibration of an SCE involves measuring its potential difference against a secondary standard, such as the saturated Ag/AgCl electrode, which typically yields +0.044 V at 25°C, confirming operational integrity.

Temperature Dependence

The potential of the saturated calomel electrode (SCE) varies with temperature due to changes in the activity of ions arising from the temperature-dependent of KCl in the saturated . This results in a of dE/dT ≈ -0.68 mV/°C. For instance, the is +0.2444 V versus the (SHE) at 25°C and +0.2376 V versus SHE at 35°C. To correct for deviations from 25°C, the potential at T (°C) is given by the formula: E(T) = E(25^\circ \text{C}) + \left( \frac{dE}{dT} \right) (T - 25) This stems from the van't Hoff relation applied to the dependence of the product for KCl, which influences the concentration and thus the Nernstian potential of the Hg₂Cl₂/, Cl⁻ half-cell. In precise electrochemical measurements, such as determinations or , temperature-compensated reference electrodes or isothermal jackets are essential to mitigate errors from this coefficient, which can accumulate to several millivolts over modest temperature ranges. Furthermore, the SCE is unsuitable for use above 80°C owing to the elevated of mercury, which risks electrode instability and mercury exposure hazards.

Practical Applications

Laboratory Uses

The saturated calomel electrode (SCE) serves as a primary in laboratory potentiometric measurements, providing a stable potential for accurate determination of concentrations in various electrochemical setups. In meters, the SCE is commonly paired with a indicator electrode to measure activity by establishing a fixed reference potential against which the pH-dependent potential is compared. Similarly, it is employed in systems for detecting specific ions such as or , where the SCE ensures reproducible baseline potentials during selective ion binding at the membrane surface. In experiments, the SCE functions as the reference in three-electrode configurations, allowing precise control and measurement of potentials at the , as seen in studies of kinetics for compounds like ascorbic acid. Specific laboratory applications highlight the SCE's versatility in analytical protocols. For redox titrations, such as those involving (IV) oxidants in for determining reducing agents like , the SCE acts as the against a indicator to monitor potential changes at equivalence points. In studies, the SCE is routinely used to measure corrosion potentials of metals like in inhibitor solutions, enabling evaluation of protective mechanisms through potentiodynamic polarization. For , the SCE provides a consistent in amperometric setups, such as those for glucose detection, where it facilitates accurate potential application during enzyme-mediated reactions. In laboratory handling, the SCE is often connected to the via a filled with saturated KCl to minimize ohmic () drop and ensure the reference potential is measured close to the without contamination. This setup is standard in academic protocols, where the SCE's stable potential—approximately +0.244 V versus the at 25°C—enables high-precision measurements.

Industrial Implementations

The saturated calomel electrode () is widely deployed in industrial process monitoring within chemical plants, where it serves as a stable reference for continuous and oxidation-reduction potential (ORP) measurements to optimize reaction conditions and ensure product quality. In facilities, SCE facilitates monitoring of pipelines and tanks by providing reliable potential references during electrochemical assessments, helping to mitigate degradation in aggressive environments containing chlorides and organics. For battery testing, particularly in lithium-ion , SCE is integrated into three-electrode setups to evaluate electrode performance and stability under operational stresses, enabling precise half-cell potential measurements that inform manufacturing scalability. In field applications, SCE-based probes are essential for of and , offering accurate and readings to assess levels in remote or harsh settings. Ruggedized versions, such as those with epoxy-sealed bodies and porous junctions, enhance portability and durability for on-site , resisting damage and maintaining in variable field conditions. These designs support prolonged deployment in electrochemical studies, where SCE contacts saturated KCl solutions to measure potentials without significant drift. Case studies highlight SCE's role in oil refineries, where it is used to determine corrosion potentials of API 5L pipeline steels in simulated crude and brine environments, aligning with API standards for internal corrosion control updated in 2024. For instance, potentiodynamic tests with SCE as the reference electrode reveal pitting tendencies in API 5L X65 under CO2-saturated conditions, guiding protective measures in refinery operations. In pharmaceutical quality control, SCE supports processes for medical devices and ensuring uniform metal deposition, with its stable potential enabling precise control of electrochemical parameters during . This application underscores SCE's contribution to reproducible in drug manufacturing and device fabrication.

Comparisons and Limitations

Versus Other Reference Electrodes

The saturated calomel electrode (SCE) serves as a secondary reference with a potential of +0.2444 V versus the standard hydrogen electrode (SHE) at 25°C, making it a convenient alternative to the SHE, which defines the zero potential scale but is impractical for routine use due to the need for continuous hydrogen gas flow, precise pressure control, and a platinum surface free of contaminants. In contrast, the silver/silver chloride (Ag/AgCl) electrode, with a saturated KCl filling, exhibits a potential of +0.197 V versus the SHE at 25°C, positioning the SCE at +0.0474 V relative to Ag/AgCl under the same conditions; the Ag/AgCl is often favored in chloride-free media because the SCE's saturated KCl bridge can introduce chloride ions that interfere with sensitive measurements, whereas Ag/AgCl maintains better stability without such leakage when properly designed. For non-aqueous solvents, where aqueous electrodes like the SCE are incompatible due to solvent immiscibility and junction potential issues, the ferrocene/ferrocenium (Fc/Fc⁺) couple is commonly employed as an internal standard with a formal potential of approximately +0.400 V versus the SHE in acetonitrile at 25°C, offering reversible behavior and solvent-independent referencing by convention setting Fc/Fc⁺ at 0 V. Selection of the SCE over alternatives depends on the system's tolerance to ions and aqueous conditions; it remains preferred in chloride-compatible aqueous environments for its reproducible potential, while Ag/AgCl suits broader applications avoiding mercury, and Fc/Fc⁺ is essential for organic electrochemistry. The following table summarizes standard potentials versus the SHE at 25°C for key reference electrodes:
Reference ElectrodePotential vs. SHE (V) at 25°C
Standard Hydrogen Electrode (SHE)0.000
Saturated Calomel Electrode (SCE)+0.2444
Silver/Silver Chloride (Ag/AgCl, saturated KCl)+0.197
/Ferrocenium (Fc/Fc⁺, in CH₃CN)+0.400
Historically, the SCE dominated laboratory practice from the mid-20th century until environmental regulations curbed mercury use; the has driven a shift to Ag/AgCl electrodes for their lower , with ongoing exemptions allowing continued use of mercury-containing reference electrodes as of 2025, though replacement is encouraged due to health and environmental concerns.

Advantages and Drawbacks

The saturated calomel electrode (SCE) provides highly reproducible potentials, typically within ±0.1 mV, thanks to the saturated KCl filling solution that stabilizes chloride ion activity against minor concentration variations. It is also economical, with commercial units costing less than $50 each, and straightforward to prepare using readily available reagents like elemental mercury, mercury(I) chloride (calomel), and potassium chloride. However, the SCE's reliance on mercury introduces significant drawbacks, including toxicity risks to human health and the environment, which has prompted a gradual phase-out under the . As of November 2025, while no outright ban exists for reference electrodes, regulatory frameworks encourage the adoption of mercury-free alternatives like Ag/AgCl electrodes, with exemptions under directives such as EU RoHS still permitting their use where necessary. In bioassays, it shows sensitivity to protein contamination, as adsorbed proteins can foul the electrode surface or liquid junction, leading to potential drift. Furthermore, the SCE is unsuitable for non-aqueous solvents, where the aqueous can intermix with media, destabilizing the reference potential. Mitigation strategies include using sealed electrode designs to contain mercury and prevent spills, alongside adopting mercury-free alternatives like Ag/AgCl electrodes in compliant laboratories. With proper , the SCE typically has a lifespan of 1-2 years.

References

  1. [1]
  2. [2]
  3. [3]
    [PDF] Section 4. Reference Electrodes - BASi
    Triple distilled mercury. Assembling the Calomel Reference Electrode. 1. Place the electrode body upright, using the small flask of KCl solution for support ...
  4. [4]
    [PDF] Chapter 11
    When using a saturated calomel electrode, the potential of the electro- chemical cell is. E. E. E cell. UO /U. SCE. 2. 4. = -. +. +. Substituting in known ...
  5. [5]
    Saturated Calomel Electrode - AlpHa Measure
    Jan 19, 2024 · Composition, Comprises mercury, calomel paste, and saturated potassium chloride solution. Consists of a silver wire coated with silver chloride.Missing: definition | Show results with:definition
  6. [6]
    [PDF] Lawrence Berkeley National Laboratory - eScholarship
    Two methods for purifying reagent grade KCl are offered. ... A Beckman #39270 saturated calomel electrode was used as reference. ... Sufficient water was added to 1 ...
  7. [7]
  8. [8]
    The Nernst Equation and Pourbaix Diagrams (all content) - DoITPoMS
    The saturated calomel electrode (SCE). The reaction is based on the reaction between this elemental mercury (Hg) and mercury(I) chloride (Hg2Cl2 ...
  9. [9]
    [PDF] Chapter 11
    When using a saturated calomel electrode, the potential of the electro- chemical cell is. Page 112. 778. Analytical Chemistry 2.0. E. E. E cell. UO /U. SCE. 2.
  10. [10]
    Accurate Potentials of Hg/HgO Electrodes: Practical Parameters for ...
    Jan 18, 2023 · This electrode potential tends to vary depending on the components of the internal solution in the Hg/HgO electrode and their concentrations.
  11. [11]
    Reference Electrode Potentials - eDAQ Wiki
    The calomel electrode (more accurately described as the 'mercury/calomel electrode') is usually constructed from a platinum wire inserted into a mixture of ...<|control11|><|separator|>
  12. [12]
    How can I calibrate a saturated calomel electrode? - ResearchGate
    Jun 10, 2015 · 1. Facing it with another equal and checking if the potential bias are close to zero. 2. Facing it in front of an Ag / AgCl / reference electrode and see if ...How to calibrate an Ag/AgCl reference electrode before performing ...How to calibrate Ag/AgCl reference electrode with respect to a ...More results from www.researchgate.net
  13. [13]
    Accurate Potentials of Hg/HgO Electrodes - ACS Publications
    HgO electrode of interest and a saturated calomel electrode. (SCE) was used (see the right side of Figure 1). In this system,. 0.1, 0.5, and 1 M NaOH/KOH ...
  14. [14]
    [PDF] Fundamentals of Electrochemistry
    ... temperature coefficient of the EMF of isothermal cells is related to the ... (0.65mV/K). Its main advantages are ease of preparation (an excess of. KCl ...
  15. [15]
    saturated calomel electrode (SCE) - Corrosionpedia
    A saturated calomel electrode (SCE) is a reference electrode based on the reaction between elemental mercury and mercury (I) chloride.
  16. [16]
    A Practical Beginner's Guide to Cyclic Voltammetry - ACS Publications
    Nov 3, 2017 · Some common reference electrodes used in aqueous media include the saturated calomel electrode (SCE), standard hydrogen electrode (SHE), and ...
  17. [17]
    [PDF] redox titration calculations - Chromedia
    A titration is performed with a saturated calomel reference electrode (S.C.E.) as the anode and a platinum wire as a cathode. Determine the cell potential ...
  18. [18]
    Studies on carbon steel corrosion in molybdate and silicate ...
    Studies on carbon steel corrosion in molybdate and silicate solutions as corrosion inhibitors ... saturated calomel electrode (SCE) as the reference electrode.
  19. [19]
    How to use Saturated Calomel Electrode (SCE) and Silver/Silver ...
    Sep 9, 2024 · One conventional way for using Ag/AgCl and SCE reference electrodes in an alkaline medium is using a Luggin capillary (It should be filled with either KCl or K ...
  20. [20]
    [PDF] Reference Electrode Overview and Care - Gamry Instruments
    To fill the electrode, slide the rubber grommet located at the top of the electrode down. This will reveal a small filling port where saturated KCl can be added ...
  21. [21]
    [PDF] Reference Electrodes and Their Usage - Metrohm
    This electrode consists of a Ag wire in contact with AgCl in a saturated KCl solution. This results in an electrode potential of 0.197 V vs. SHE at 25 °C.
  22. [22]
    Development of an Instantaneous Corrosion Rate Monitoring ...
    Stainless steel electrodes performed just as well as standard saturated calomel electrode while treated municipal wastewater was used as makeup water in the ...
  23. [23]
    Lithium Battery Materials Cell - Gamry Instruments
    The Lithium Battery Electrochemical Cell Kit has been designed for electrochemical testing of lithium battery ... Access your Saturated Calomel Electrode ...
  24. [24]
    A Guide to the Application and Customization of Three-Electrode ...
    Dec 13, 2024 · Commonly used reference electrodes for aqueous systems are saturated calomel electrode (SCE) ... Understanding dQ/dV Analysis for Lithium-Ion ...
  25. [25]
    What is Calomel Electrode? - Apure Instruments
    Mar 29, 2024 · Environmental monitoring: Calomel electrodes can be used to monitor the pH value of water in the environment, to monitor and control ...
  26. [26]
    REFERENCE ELECTRODE POROUS POLYMERjunction gel filled ...
    This item has been discontinued and is no longer available. View the product for possible alternatives or contact our Technical Support team on 056 260 260 ...
  27. [27]
    [PDF] AN ELECTROCHEMICAL APPROACH FOR INVESTIGATING ...
    A salt bridge that consisted of a. 8.25 cm diameter pine plug in a plastic tube filled with saturated KCl was employed to make electrical contact with the soil.
  28. [28]
    [PDF] Corrosion Mechanism of A333 and API 5L in 0.02 M Acidic Crude ...
    Jun 4, 2024 · Platinum and saturated calomel electrodes (SCEs) are the counter and reference electrodes, respectively. The polarization Tafel slope curves ...
  29. [29]
    [PDF] API PUBLICATIONS CATALOG
    API is pleased to present its 2024 Publications Catalog. The 2024 Edition lists API standards, recommended practices, and other technical documents to help the ...Missing: calomel | Show results with:calomel
  30. [30]
    [PDF] Corrosion Rates of API 5L X-52 and X-65 Steels in Synthetic Brines ...
    Mar 2, 2014 · This study shows the results of the measurements of the corrosion rates of type API 5L X-52 and API 5L X-65 steels in a multiphase system ...Missing: monitoring RP
  31. [31]
    Electrodeposition in deep eutectic solvents - ScienceDirect.com
    Electrodeposition in deep eutectic solvents (DES) offers a versatile and sustainable approach for both discovering new materials and creating sophisticated ...
  32. [32]
    23.1: Reference Electrodes - Chemistry LibreTexts
    Oct 26, 2022 · The potential of a calomel electrode that contains an unsaturated solution of KCl is less dependent on the temperature, but its potential ...Missing: composition | Show results with:composition
  33. [33]
    Other reference electrodes - DoITPoMS
    The saturated calomel electrode (SCE),; The silver/silver chloride ... The half-cell reaction is. AgCl + e- = Ag + Cl-. which gives a Nernst equation of.
  34. [34]
    Converting Reference Electrode Potentials - McAuley Group
    Learn how to convert between reference electrodes (SHE, SCE, Ag/AgCl, RHE, NHE) with worked examples and an online calculator.Missing: derivation | Show results with:derivation
  35. [35]
    Non-Aqueous Reference Electrode Overview
    Sep 25, 2024 · It is as stable and reliable as the saturated calomel reference electrode (SCE ... Ferrocene undergoes a reversible, one electron oxidation to ...Reference Electrode Stability · Modification for Non-Aqueous...
  36. [36]
    Conversion constants for redox potentials measured versus different ...
    Conversion constants for redox potentials measured versus different reference electrodes in acetonitrile solutions at 25°C are proposed.
  37. [37]
    A review of screen-printed silver/silver chloride (Ag/AgCl) reference ...
    Nov 1, 2017 · Ag/AgCl REs are widely used due to their environmental compatibility. The Mercury/Mercury Chloride (Hg/Hg2Cl2), calomel electrode, has ...
  38. [38]
    [PDF] Other electronic devices - Minamata Convention on Mercury
    The replacement of the calomel electrode by the silver chloride electrode is also recommended by the United States Pharmacopeial Convention (2017)16. Many ...<|control11|><|separator|>
  39. [39]
    Converting between common reference electrodes - solarchemist.se
    Nov 24, 2016 · The potentials of calomel electrodes can be reproduced to ±0.1mV ± 0.1 mV . ... Figure 5: Electrode potentials vs SHE for the two reference ...
  40. [40]
    Saturated Calomel Electrode - Beyond Battery
    In stockSaturated Calomel Electrode. Regular price $45.00. /. Model. Model. R0232. R0217 ... By continuing, I agree to the cancellation policy and authorize you to charge ...Missing: cost | Show results with:cost
  41. [41]
    Protein Adsorption and Its Effects on Electroanalytical Performance ...
    Jul 11, 2023 · Protein Adsorption and Its Effects on Electroanalytical Performance of Nanocellulose/Carbon Nanotube Composite Electrodes ... calomel electrode.
  42. [42]
    Reference Electrode, Saturated Calomel, Single Junction
    The standard-sized, single junction, saturated calomel (Hg 2 Cl 2 ) reference electrode is filled with 4M KCl. The refrerence electrode body is 9.5 mm OD.
  43. [43]
    Reference Electrodes - BASi
    Two widely used aqueous reference electrodes are the silver/silver chloride electrode and the saturated calomel electrode. These are now discussed in more ...