Automatic exposure control

Automatic exposure control (AEC) is a radiographic designed to automatically terminate exposure when a predetermined amount of reaches the receptor, ensuring consistent and across varying patient sizes and anatomies while minimizing unnecessary . This system is integral to modern diagnostic imaging equipment, including conventional , , and computed (CT) scanners, where it replaces manual timing to optimize both diagnostic efficacy and patient safety. AEC operates by employing detectors—typically ionization chambers positioned between the patient and the image receptor—that convert transmitted X-rays into an electrical signal proportional to the intensity. The selects one or more detectors based on the , such as the lungs or in chest imaging, and sets parameters like kilovoltage peak (kVp) and milliampere (mA); the system then modulates time in until the signal threshold is met, often displaying the effective milliampere-seconds (mAs) post-exposure for technique refinement. In CT applications, advanced AEC variants further adjust tube current angularly and longitudinally using scout scans to account for , achieving dose reductions of 10-60% compared to fixed protocols without compromising levels or diagnostic utility. Developed in the early 1940s with initial automatic timers using phototimers at institutions like the , AEC evolved from these rudimentary tube-based phototimers to more reliable chamber-based systems in the later . Its adoption has significantly curtailed "dose creep" and repeat exposures, with studies demonstrating up to 38% dose savings in routine head scans while preserving image quality. Today, AEC remains a cornerstone of guidelines from organizations like the American Association of Physicists in Medicine (AAPM), underscoring its role in balancing clinical needs with the ALARA (as low as reasonably achievable) principle.

Fundamentals

Definition and Purpose

Automatic exposure control (AEC) is an exposure termination that automatically ends the radiographic exposure when a preset level of reaches the image receptor, ensuring the appropriate amount of for . This system integrates sensors positioned in front of or behind the image receptor to monitor intensity and signal the to terminate the exposure once the threshold is met. The primary purpose of is to produce consistent image density and optical quality, regardless of variations in patient or thickness, thereby minimizing unnecessary to and eliminating the need for extensive manual adjustments of technique factors such as exposure time. By automating exposure termination, AEC helps standardize radiographic outcomes, reduces technologist error, and supports adherence to the ALARA (as low as reasonably achievable) principle for dose optimization. AEC was first developed in in the early 1940s to improve exposure consistency in early imaging systems. As imaging technology evolved, AEC systems were adapted for digital modalities, including computed radiography (CR) and direct digital radiography (DR), where they continue to regulate exposure to achieve optimal signal-to-noise ratios while accounting for the wider of digital detectors. This adaptation maintains AEC's core function but incorporates adjustments for digital processing capabilities, such as post-exposure contrast enhancement.

Underlying Principles

Automatic exposure control (AEC) operates on the principle that X-rays emitted from the tube are attenuated by the patient's body tissues through and , resulting in varying amounts of reaching the detector depending on anatomical thickness and . This attenuation reduces the intensity of the beam, with denser structures like absorbing more photons than softer tissues, thereby necessitating compensatory adjustments in exposure parameters to maintain consistent image quality across diverse morphologies. The core exposure termination mechanism in AEC involves real-time measurement of cumulative radiation at the detector plane, typically quantified in units of roentgens or air kerma, which triggers the cessation of the beam once a predefined is achieved to ensure adequate detector . This is calibrated to deliver a specific level that optimizes image receptor response, preventing under- or over while minimizing dose. The radiation reaching the detector is proportional to the X-ray tube output, which scales approximately with the tube current (mA), exposure time, and the square of the (kVp²), further modulated by patient . AEC primarily modulates the time component based on feedback to achieve the target exposure, adapting to variations in . In film-screen systems, AEC targets a consistent optical (typically in the range of 0.25–2.5) on the processed to ensure proper and visibility, as the film's response is directly tied to the silver halide exposure. Conversely, digital systems shift the focus to achieving an optimal (SNR) at the detector, where exposure indicators (e.g., equivalent air kerma values) guide termination to balance noise levels and diagnostic utility, accommodating the wider of digital detectors. Feedback loops form the operational backbone of , enabling continuous monitoring of intensity during and real-time adjustments to tube current or time via signals from the detector back to the , thus dynamically compensating for fluctuations. This closed-loop control ensures precise termination without manual intervention, enhancing reproducibility across exposures.

Components

Sensors and Detectors

In automatic exposure control (AEC) systems for , the primary sensors are ionization chambers, which consist of flat, parallel-plate devices filled with air or another gas and positioned between the patient and the image receptor to measure the ionization current produced by incident photons. These entrance-type detectors generate an electrical signal proportional to the intensity, enabling the system to terminate upon reaching a predetermined dose . Ionization chamber configurations typically feature 1 to 3 chambers per system, arranged in selectable fields such as central, lateral, or a combination thereof, allowing radiographers to position them appropriately for specific anatomical regions like the chest or . Some advanced systems employ up to 5 fields for greater flexibility in multi-region imaging. Alternative detectors in modern digital AEC systems include photodiodes or tubes (PMTs) paired with fluorescent screens, which convert X-ray-induced light to electrical signals and are often used for post-exposure feedback or in exit-type configurations behind the image receptor. Solid-state detectors are also emerging as options, offering compact designs with similar measurement capabilities. Calibration of these sensors is essential and involves adjusting to specific kilovoltage peak (kVp) settings and beam filtration to account for energy-dependent responses, ensuring accurate detection of exposure thresholds calibrated to air or standards. Service personnel typically perform this to align with departmental image quality requirements, with recalibration recommended for digital systems due to differences in detector materials like gadolinium oxysulfide versus cesium . Limitations of AEC sensors include their inherent sensitivity to beam quality variations, such as changes in kVp or , which can lead to inaccurate termination if not properly calibrated. Positioning errors, including misalignment of chambers relative to the or inter-chamber signal imbalances, further compromise performance, potentially resulting in underexposed or overexposed images. Additionally, minimum durations imposed by regulatory standards, such as less than 1/60 second or 5 , may be insufficient for certain high-speed digital applications.

Control Systems

Control systems in automatic exposure control (AEC) encompass the electronic and software elements responsible for processing sensor signals and precisely terminating X-ray exposures to achieve consistent image density. These systems receive input from radiation detectors, such as ionization chambers positioned between the patient and image receptor, and employ integrated circuits to analyze the incoming data in real time. The core of the control circuitry consists of amplifiers and integrators that transform the low-level current generated by the sensor—proportional to the incident X-ray intensity—into a measurable voltage signal representing the cumulative radiation exposure. Amplifiers boost the weak sensor output for reliable processing, while integrators accumulate this signal over time, providing a direct measure of the total exposure delivered to the image receptor. This analog-to-digital conversion ensures the signal accurately reflects variations in beam quality and patient attenuation without distortion. Termination logic within AEC systems utilizes microprocessor-based algorithms to continuously compare the integrated voltage signal against a user-preset threshold, often calibrated to a target density index or exposure value for optimal radiographic quality. Upon reaching this threshold, the logic generates a precise termination pulse, typically within milliseconds, to end the exposure and prevent over- or underexposure. These algorithms may incorporate compensation factors for kilovoltage peak (kVp) settings and field selection to maintain uniformity across diverse imaging scenarios. To enhance safety, incorporates backup mechanisms that automatically revert to a fixed time if the primary signal fails or falls below a minimum response level, typically set to 1.5 to 2 times the anticipated duration based on technique factors. This fallback prevents excessive radiation delivery to or damage to the , with limits often enforced at 600 mAs or less for systems operating above 50 . AEC control systems integrate seamlessly with the X-ray tube high-voltage generator through relay interfaces or direct electronic signaling, where the termination pulse interrupts the filament current or high-voltage supply, effectively cutting off the tube current (mA) and stopping X-ray production instantaneously. This synchronization ensures minimal latency, preserving the accuracy of the exposure control. In contemporary digital radiography (DR) and computed radiography (CR) systems, software enhancements refine AEC operation by applying algorithms that calibrate for the nonlinear response curves of flat-panel or photostimulable phosphor detectors. These adjustments, often derived from signal-to-noise ratio (SNR) targets, recalibrate kVp compensation curves to account for detector-specific energy absorption and processing characteristics, thereby optimizing dose efficiency and image quality in digital workflows.

Operation in Imaging Modalities

Projectional Radiography

In , automatic exposure control (AEC) operates through a standardized that begins with the selecting the kilovoltage peak (kVp) and milliamperage () settings based on the anatomical region, followed by positioning the patient such that the central area of interest aligns over one or more ionization chambers embedded in the image receptor bucky tray. The exposure is then initiated manually, during which the AEC system continuously monitors the transmitted through the patient and reaching the selected chambers; termination occurs automatically when the integrated signal from these chambers reaches a predetermined calibrated to the density () of the central , ensuring consistent image receptor exposure without manual timing. This process relies on the underlying principle of real-time detection to maintain optimal exposure levels across varying patient sizes and compositions. Chamber selection is critical for adapting AEC to the uniformity of the imaging field, with systems typically featuring three ionization chambers arranged linearly in the bucky tray. For non-uniform fields such as the chest, where lung density varies significantly from the mediastinum, a single central chamber or a combination of the central and one lateral chamber is often selected to prioritize the lungs and heart, avoiding overexposure from denser structures like the spine. In contrast, uniform fields like the abdomen utilize all three chambers to average attenuation across a more homogeneous region, promoting balanced exposure and reducing variability in milliamperage-seconds (mAs). Proper collimation to cover only the active chambers minimizes scatter and enhances accuracy, as including extraneous areas can skew the signal. In film-screen systems, AEC is calibrated to achieve a target optical density of 1.0 to 2.0 in the central image area, corresponding to the linear portion of the film's characteristic curve for optimal contrast and detail visibility. With the transition to digital radiography, AEC targets shift to exposure index (EI) values, typically aiming for 200 to 400 units (depending on the vendor's standard, such as Fuji's sensitivity number inverted or Canon's EI), where a deviation index near zero indicates appropriate receptor dose for post-processing without excessive noise or artifact amplification. This adjustment accounts for digital detectors' wider dynamic range, allowing AEC to maintain signal-to-noise ratios equivalent to film densities while enabling dose optimization. Common errors in AEC use often stem from patient or equipment misalignment, such as off-center positioning of the anatomy relative to the active chambers, which can result in overexposure if a low-density area (e.g., air-filled ) dominates the signal, leading to prolonged and unnecessary dose. Under-collimation, where the beam extends beyond the intended field to include grid lines or non-anatomic structures, may introduce scatter that prematurely terminates the exposure, causing underexposure in the . These issues are mitigated through timers set to 1.5 to 2 times the expected exposure and routine of chamber . AEC has been a standard feature in systems since the 1970s, evolving from early phototimer-based designs to modern configurations that provide greater reliability and integration with digital workflows.

Computed Tomography

Automatic exposure control (AEC) in computed tomography () was introduced in the mid-1990s to modulate the current (mA) dynamically during rotation, enabling dose reduction while preserving image quality across varying patient anatomies. This approach built on earlier concepts of anatomically adapted modulation introduced in the late 1990s, which demonstrated up to 50% dose savings in studies by adjusting mA based on regional . By the mid-2000s, AEC became a standard feature in multi-detector CT scanners from major manufacturers, responding to growing concerns over in volumetric imaging. In , AEC operation begins with a pre-scan topogram or scout view, a low-dose radiograph that maps the patient's along the z-axis and angular directions. This informs real-time of the tube current for each angle during helical scanning, increasing mA through denser regions (e.g., shoulders or ) and decreasing it in less attenuating areas (e.g., lungs) to optimize dose distribution. The system continuously adjusts mA within the scan plane and along the z-axis, ensuring uniform throughout the reconstructed volume without manual intervention. AEC algorithms in CT primarily employ noise-based control, targeting a user-defined constant image noise level by adjusting exposure parameters according to patient size and attenuation. For instance, the computed tomography dose index volume (CTDIvol) is scaled dynamically, with mA increased for larger patients to counteract higher attenuation and maintain noise targets. A key metric in this process is the effective milliampere-seconds (mAs), calculated as: \text{effective mAs} = \frac{\text{mA} \times \text{rotation time}}{\text{pitch}} where AEC varies mA to sustain the desired noise while accounting for helical pitch in volumetric acquisitions. Unlike projectional radiography, which relies on fixed-direction exposure termination, AEC in CT incorporates angular and z-axis modulation tailored for helical scans, enhancing dose efficiency in three-dimensional imaging by adapting to the full rotational and longitudinal path of the X-ray beam. This multidimensional adjustment minimizes overexposure in asymmetric anatomies, achieving 30-60% dose reductions compared to fixed mA protocols in routine exams.

Mammography and Fluoroscopy

In mammography, automatic exposure control (AEC) systems are dedicated to optimizing low radiation doses for imaging dense breast tissue, utilizing sensors integrated into the compression paddle to measure breast thickness and attenuation. These systems employ molybdenum (Mo) or rhodium (Rh) targets and filters to produce appropriate beam spectra that balance penetration and contrast while minimizing patient dose. The compression paddle sensors ensure precise positioning near the chest wall, guiding technique factor adjustments for varying breast densities. Operation in mammography involves an initial or trial to automatically select the material and combination based on characteristics, followed by termination of the main to achieve target image quality. For screen- systems, terminates at an optical of 1.4-1.6 in the parenchymal to ensure consistent across thicknesses from 2.5 to 8 cm. In full-field digital (FFDM), relies on value histograms from regions of interest in unprocessed images to provide post- , targeting signal-to-noise ratios and values for dose optimization without fixed optical goals. Challenges in AEC include risks of grid cutoff due to high variations from dense , which can result from misalignment or off-centering during , leading to uneven and artifacts. In , AEC operates in real-time as automatic (ABC) or automatic (ADRC), continuously adjusting kilovolt peak (kVp) and milliampere () per frame to maintain a constant video signal level at the receptor despite patient motion or varying thickness. This feedback loop uses photodetectors or chambers to monitor input, prioritizing either kVp increases for and lower dose or balanced /kVp adjustments for preservation. Such dynamic ensures stable during continuous low-dose procedures. A key challenge in fluoroscopic AEC is lag from image intensifiers, which introduces a 1-2 second delay in response to attenuation changes, potentially causing transient over- or underexposure during rapid movements.

Benefits and Limitations

Advantages

Automatic exposure control (AEC) ensures uniform at the image receptor regardless of variations in size, thickness, or habitus, thereby reducing inconsistencies that arise from technique selection. This consistency minimizes exposure variability across diverse populations, such as pediatric or obese individuals, leading to more reliable radiographic outcomes without the need for technologist adjustments during the procedure. A primary benefit of AEC is its role in dose optimization, which typically lowers patient radiation dose by 40-50% compared to fixed manual techniques, while preventing overexposure in thinner body parts. By modulating tube current and exposure time in real-time based on tissue attenuation, AEC adheres to the ALARA (as low as reasonably achievable) principle, tailoring the dose to the specific anatomical region and avoiding unnecessary radiation to radiosensitive areas. Studies confirm that this modulation maintains diagnostic image quality without compromising (SNR), particularly in (CT) scans where angular adjustments further enhance efficiency. AEC improves operational efficiency in radiographic workflows by eliminating technologist guesswork in selecting milliampere-seconds () values, which streamlines procedures and reduces the incidence of repeat examinations due to suboptimal exposures. Research indicates that proper AEC implementation can lower repeat rates by optimizing positioning and exposure accuracy, decreasing unnecessary patient radiation from retries in specialized applications like . This efficiency gain is especially valuable in high-volume settings, allowing faster throughput while upholding image standards. Recent advancements as of 2025 incorporate (AI) and into AEC systems, further optimizing dose and image quality for additional reductions in repeats and . In digital imaging systems, AEC enhances overall diagnostic quality by sustaining optimal SNR and optical density levels, which supports better lesion detectability and interpretive accuracy. By standardizing noise across slices or projections, AEC mitigates artifacts from inconsistent exposures, facilitating clearer visualization of subtle anatomical details. Evidence from clinical studies demonstrates that AEC counters "dose creep"—the gradual increase in radiation output during digital transitions—by enforcing protocol standardization in CT, resulting in sustained dose reductions without quality degradation.

Disadvantages

Automatic exposure control (AEC) systems in and computed tomography (CT) are susceptible to operator dependency, where errors in chamber selection or patient positioning can lead to significant under- or overexposure. For instance, incorrect selection of detectors has been shown to cause overexposures as high as 90 mGy in film-screen systems, while misalignment of the field can alter exposure by up to ±22% in chest phantoms. In complex anatomies, such as those involving varying densities, operator errors in positioning can result in suboptimal image quality, with AEC termination failures occurring due to improper setup. These issues underscore the need for precise operator training to mitigate risks associated with sensor configurations, though AI enhancements in recent systems (as of 2025) are reducing such dependencies. Artifacts represent another key drawback of AEC, particularly when dense structures like prosthetics or bones interfere with exposure termination. High-density materials, such as implants, can significantly alter gray values in digital intraoral radiography, leading to decreased contrast in and regions and potentially compromising diagnostic accuracy for conditions like caries. In , air gaps or metallic prosthetics may cause uneven beam attenuation, resulting in underexposed areas or visible grid lines on images. Such artifacts arise because AEC relies on integrated signal from detectors, which can be disproportionately affected by localized high-attenuation objects. Calibration requirements pose ongoing challenges for performance, necessitating regular testing to prevent sensor drift and inconsistent exposures. Digital systems must be calibrated using parameters like exposure index () rather than traditional optical density, with annual () recommended to ensure compliance with standards and maintain as low as reasonably achievable (ALARA) doses. Improper calibration can lead to incorrect exposures, as seen in systems where kVp variations cause deviations in , requiring service personnel intervention to realign detectors to departmental protocols. In modalities, over-reliance on can overlook detector saturation in high-contrast scenes, where the wide masks overexposures that appear clinically normal but deliver unnecessary . For example, exposures 5–10 times above normal may not visibly degrade images due to post-processing compensation, potentially increasing dose without diagnostic benefit. This limitation is exacerbated in small patients, where constant-noise AEC paradigms produce excessively noisy images (up to 435% relative noise compared to clinical charts), or in obese patients, leading to excessive doses (up to 549% relative to standards). Finally, introduces higher costs and complexity compared to manual exposure methods, including elevated expenses for advanced systems and the need for specialized . Modern scanners with features often command premium prices due to integrated tube current modulation hardware, while requires multidisciplinary optimization committees to dose reduction with image quality, adding operational overhead.

References

  1. [1]
    Automatic exposure control: a primer - PubMed
    AEC is a radiographic density control device that terminates the exposure when a predetermined amount of radiation has been reached.
  2. [2]
    Automatic Exposure Control | Radiology Key
    Feb 27, 2016 · AEC is a system used to consistently control the amount of radiation reaching the image receptor by terminating the length of exposure.
  3. [3]
    [PDF] Using Automatic Exposure Control in Digital Radiography - AAPM
    Initial acceptance testing. ▫. Sensor selector/location. ▫. Density correction. ▫. Screen sensitivity adjustment. ▫. Sensitivity vs. speed setting.<|control11|><|separator|>
  4. [4]
    [PDF] Automatic Exposure Control During Computed Tomography Scans ...
    However, automatic exposure control (AEC) is a feature that can be used to control radiation exposure in radiography and CT. Although AEC has been used in ...
  5. [5]
    An Evaluation of Automatic Exposure Control Equipment in ...
    History of Automatic Timer Development. Late in the summer of 1940, investigations were undertaken at the University of Chicago to develop equipment whereby ...
  6. [6]
    [PDF] Acceptance Testing and Quality Control of Digital Radiographic ...
    Acceptance Testing for X-Ray Generators and Automatic Exposure Control Devices. AAPM. Report 14. New York, NY: American Institute of Physics, 1985. [6] ...
  7. [7]
    Automatic exposure control | Radiology Reference Article
    Jun 7, 2024 · Automatic exposure control (AEC) is incorporated into radiographic and mammographic imaging systems. Its function automatically terminates exposure.
  8. [8]
    [PDF] Best Practices in Digital Radiography
    When systems have automatic exposure control (AEC), other variables such as AEC detector cell configuration and backup time also should be standardized.
  9. [9]
    An exposure indicator for digital radiography: AAPM Task Group 116 ...
    For film/screen images, the optical density of the image itself is used to indicate proper exposure according to the clinical preferences of the facility.
  10. [10]
    AEC Sensors Archives - Varex Imaging
    The Automatic Exposure Control sensors measure the dose rate of X-rays by conversion of X-ray photon energy to an electrical current or signal intended for AEC.
  11. [11]
    Ionization Chamber - OEM X-Ray Products and Components
    The 3-field and 5-field ionization chamber is used in X-ray diagnostic in automatic exposure control systems as a measuring detector.Missing: sensors | Show results with:sensors
  12. [12]
    [PDF] AAPM REPORT No. 14
    Automatic exposure control (A.E.C.) will be used to refer to any device used ... "Automatic Exposure Control in General Radiography",. A. D. Smythe ...Missing: history | Show results with:history<|control11|><|separator|>
  13. [13]
    Detector for automatic exposure control on an x-ray imaging system
    Automatic exposure control (AEC) is used in x-ray imaging equipment to control the exposure per image. The goal is to maintain image quality while minimizing ...Missing: radiology | Show results with:radiology
  14. [14]
    Automatic exposure control - Book chapter - IOPscience
    Technical Fundamentals of Radiology and CT. Chapter 8. Automatic exposure ... The first ionization chamber system was invented by Leo Szilard and is based on the ...
  15. [15]
    Automatic exposure control in general radiography - PubMed
    Automatic exposure control in general radiography. Australas Radiol. 1966 Aug;10(3):259-66. doi: 10.1111/j.1440-1673.1966.tb00808.x.
  16. [16]
    Effects of AEC Chamber Selection on Patient Dose and Image Quality
    Radiographers can reduce patients' ESE and maintain image quality by selecting the 2 outside AEC chambers.
  17. [17]
    The Importance of Patient Positioning in Radiography When ...
    Oct 17, 2025 · The AEC is commonly utilised for in-bucky imaging, such as for the chest, abdomen, pelvis and spine.
  18. [18]
    Effects of AEC chamber selection on patient dose and image quality
    Aug 8, 2025 · Radiographers can reduce patients' ESE and maintain image quality by selecting the 2 outside AEC chambers.
  19. [19]
    New Exposure Indicators for Digital Radiography Simplified for ...
    Digital radiography overcomes the dynamic range limitations of film-screen radiography because the receptor responds linearly to exposure, with a dynamic range ...
  20. [20]
    The standardized exposure index for digital radiography
    Apr 14, 2011 · The dose to the patient is determined by the X-ray technique factors (kV, mAs, grid, SID, filtration, beam collimation), X-ray beam ...Missing: equation | Show results with:equation
  21. [21]
    Exposure variations under error conditions in automatic ... - PubMed
    Improper automatic exposure control (AEC) termination may result in high overexposures on some radiographic systems.
  22. [22]
    Exposure variations under error conditions in automatic exposure ...
    Aug 6, 2025 · Automatic exposure control: A primer. Article. May 1988; Radiol Tech. S Sterling. AEC is a radiographic density control device that terminates ...
  23. [23]
    CT dose modulation using automatic exposure control in whole ...
    CT dose modulation using AEC may vary greatly depending on scout direction. Raising the arms tended to decrease radiation exposure.
  24. [24]
    Automatic Exposure Control Systems Designed to Maintain ... - NIH
    Objective. To compare computed tomography dose and noise arising from use of an automatic exposure control (AEC) system designed to maintain constant image ...
  25. [25]
    Relationship between size‐specific dose estimates and image ...
    May 4, 2018 · This suggests that the AEC technique that maintains a constant image quality (CARE Dose 4D) for each patient size corresponds well to the image ...
  26. [26]
    Methods for CT Automatic Exposure Control Protocol Translation ...
    This plot shows data measured from the 32 cm diameter PMMA phantom. Figure 3B: The measured values of CTDIvol and image noise are plotted for the AEC settings ...
  27. [27]
    Automatic exposure control in multichannel CT with tube current ...
    Jun 26, 2007 · Automatic exposure control (AEC) systems have been developed by computed tomography (CT) manufacturers to improve the consistency of image ...
  28. [28]
    [PDF] Pub1482_web.pdf
    Automatic exposure control on digital mammography systems. The design of the automatic exposure control (AEC) in digital mammography units differs from that in.
  29. [29]
    [PDF] EQUIPMENT REQUIREMENTS AND QUALITY CONTROL FOR ...
    Many of the current mammographic units operated in AEC mode have difficulty in yielding images of constant optical density as breast thickness and tube ...
  30. [30]
    [PDF] Handbook of basic quality control tests for diagnostic radiology
    Section 1 is an introduction to the subject. Section 2 outlines QC tests on radiography, Section 3 on fluoroscopy and angiography, Section 4 on mammography and ...
  31. [31]
    Recording of Fluoroscopic Images and Automatic Exposure Control
    Automatic Exposure Control. Automatic control of the radiation exposure to the image detector is needed for both fluoroscopy and image recording in most cases.
  32. [32]
    [PDF] The “Legacy” Fluoroscopy - AAPM
    The primary theme of this presentation is on the Automatic Brightness Control, or the updated version, called Automatic Dose Rate Control (ADRC) Logic which is ...
  33. [33]
    [PDF] Smart mA – Automatic Exposure Control (AEC) - Mayo Clinic
    One of the most impor- tant technologies used to reduce dose in CT is Automatic Exposure Control (AEC), which aims to automatically modulate the tube current to ...
  34. [34]
    Use of an Automatic Exposure Control Mechanism for Dose ...
    CONCLUSION: An automatic exposure control mechanism that is based on real-time anatomy-dependent tube current modulation delivers good image quality with a ...
  35. [35]
    Automatic Exposure Control (AEC) in Mammography: A new method ...
    [Show full abstract] routine clinical xeromammography, a repeat rate of 20% was eliminated by the use of AET. Read more. Conference Paper. Full-text available ...Missing: radiography | Show results with:radiography
  36. [36]
    [PDF] Avoidance of Unnecessary Dose to Patients While Transitioning ...
    The phenomenon of “exposure creep” in CR had been recognized for some time ... • that use of AEC significantly reduced patient dose for most examinations.<|control11|><|separator|>