Contact angle

The contact angle is a key parameter in interfacial science, defined as the angle formed between the tangent to the liquid-vapor interface and the solid surface at the three-phase contact line where a liquid droplet meets a solid substrate in the presence of vapor. It serves as a direct measure of wettability, with values below 90° indicating hydrophilic (wetting) behavior where the liquid spreads on the surface, and values above 90° signifying hydrophobic (non-wetting) behavior where the droplet beads up. This angle arises from the balance of interfacial tensions at the contact line, governed by adhesive forces between the liquid and solid versus cohesive forces within the liquid. The concept of the contact angle was first formalized by Thomas Young in 1805 through his equation, which relates the equilibrium contact angle \theta to the surface tensions of the solid-vapor (\gamma_{SV}), solid-liquid (\gamma_{SL}), and liquid-vapor (\gamma_{LV}) interfaces: \cos \theta = \frac{\gamma_{SV} - \gamma_{SL}}{\gamma_{LV}}. This relationship, known as Young's equation, provides a thermodynamic basis for understanding phenomena and remains central to the field despite later refinements accounting for factors like and chemical heterogeneity. In practice, contact angles are measured using techniques such as sessile drop analysis, where the shape of a deposited droplet is imaged to determine the angle, or dynamic methods like the , which assess advancing and receding angles during liquid movement to quantify —the difference between these angles due to surface imperfections or contaminants. is particularly important, as it reflects real-world deviations from ideal equilibrium and influences processes like droplet motion on inclined surfaces. Contact angles play a pivotal role across diverse applications in materials science, engineering, and biology, including the design of self-cleaning surfaces (e.g., lotus effect with angles near 180°), optimization of adhesives and coatings for better bonding, and control of ink spreading in printing technologies. In biomedical contexts, they guide the development of biocompatible implants by assessing protein adsorption and cell adhesion on surfaces, while in environmental science, they inform soil water retention and pollutant transport in porous media. Advances in nanotechnology have further expanded their utility, enabling precise tuning of surface energies for superhydrophobic materials that repel water effectively, with implications for anti-fogging, anti-icing, and microfluidic devices. Overall, the contact angle encapsulates the interplay of molecular forces at interfaces, driving innovations in wettability control for practical and scientific advancements.

Fundamentals

Definition and basic principles

The contact angle is defined as the angle, measured through the , between the - and the to the -vapor at the three-phase contact line, where the , , and vapor phases meet. This geometric configuration characterizes the shape of a droplet or on a under static conditions, with the contact angle denoted as \theta_{eq}. The contact angle arises from the balance of forces at this contact line, reflecting the interplay between the 's tendency to spread or bead up on the . Physically, the contact angle signifies the relative magnitudes of the interfacial tensions at the solid-liquid (\gamma_{sl}), solid-vapor (\gamma_{sv}), and liquid-vapor (\gamma_{lv}) interfaces, which determine the surface energy minimization at the contact line. A lower contact angle indicates stronger between the solid and liquid compared to the liquid-vapor , promoting ; conversely, a higher angle suggests weaker solid-liquid , favoring non-wetting behavior. This balance is quantitatively captured by Young's equation, which relates \theta_{eq} to these interfacial tensions. The underlying principles stem from the cohesive forces within the liquid and adhesive interactions with the solid, as first conceptualized by Thomas Young in his analysis of fluid . In terms of wettability, a contact angle \theta < 90^\circ classifies the surface as hydrophilic, where the spreads readily due to favorable solid-liquid interactions; \theta > 90^\circ indicates hydrophobic behavior, with the forming a more spherical shape to minimize contact with the solid. These distinctions are crucial for understanding phenomena like liquid spreading on materials in various applications. Basic prerequisites for contact angle formation include , defined as the cohesive force per unit length at a liquid-vapor that resists external deformation and minimizes surface area, and capillarity, the resulting curvature-driven rise or depression of liquids in confined spaces due to interfacial tension imbalances. These properties govern the overall interfacial dynamics without which the contact angle could not be observed.

Historical development

The concept of the contact angle emerged in the early through studies of capillarity and fluid . In 1805, Thomas Young described the contact angle as the angle formed at the three-phase junction where a meets a solid and its vapor, attributing it to the balance of cohesive forces within the and adhesive forces between the and solid. This foundational observation appeared in his essay "An Essay on the Cohesion of Fluids," published in the Philosophical Transactions of the Royal Society, marking the initial qualitative link between interfacial tensions and the angle's magnitude. Shortly thereafter, extended these ideas in his 1806 work on , part of the supplements to his "Mécanique Céleste." Laplace's contributions emphasized the at fluid interfaces, incorporating the contact angle into broader theories of and pressure differences across curved menisci, though without explicit focus on solid-liquid interactions. Later in the century, J. Willard Gibbs advanced the framework in his 1876-1878 treatise "On the Equilibrium of Heterogeneous Substances," providing a rigorous interpretation of interfacial energies and proposing a modification to Young's description to account for volumetric effects on the contact angle. Gibbs' analysis shifted the understanding from purely mechanical to energetic principles, laying groundwork for modern interfacial . The early 20th century saw refinements addressing practical complexities, particularly the phenomenon of contact angle hysteresis—the difference between advancing and receding angles during droplet motion. In the 1930s, N.K. Adam explored this hysteresis in his book "The Physics and Chemistry of Surfaces" (1930), attributing it to surface irregularities and pinning effects at the three-phase line, which expanded the theory beyond ideal smooth surfaces. Concurrently, R.N. Wenzel introduced in 1936 a model for how amplifies behavior, showing that roughness increases the effective contact angle on hydrophobic surfaces and decreases it on hydrophilic ones. Building on this, A.B.D. Cassie and S. Baxter developed in 1944 a complementary model for porous or heterogeneous surfaces, where air pockets trapped beneath the liquid lead to composite interfaces and higher apparent contact angles. Post-1950s developments marked a transition from empirical and macroscopic models to molecular-level interpretations, driven by advances in and instrumentation. The rise of techniques and in the late 1950s enabled detailed probing of solid-liquid interfaces, revealing atomic-scale influences on . By the , W.A. Zisman's critical surface tension concept further quantified solid surface energies via contact angle series, while subsequent decades incorporated and simulations to explain and dynamics at the nanoscale. This era solidified contact angle theory as a of , with ongoing refinements addressing line tension and non-equilibrium effects.

Theoretical foundations

Young's equation

Young's equation provides the fundamental relationship between the equilibrium contact angle and the interfacial tensions at the three-phase contact line where a liquid droplet meets a in a vapor . Originally described by Thomas Young in 1805, it equates the cosine of the contact angle \theta to the difference in solid-vapor and solid-liquid interfacial tensions divided by the liquid-vapor interfacial tension. The derivation arises from a force balance at the contact line, considering the interfacial tensions acting tangentially to their respective interfaces. In , the solid-vapor tension \gamma_{SV} pulling along the balances the solid-liquid tension \gamma_{SL} and the component of the liquid-vapor \gamma_{LV}. This yields: \gamma_{SV} = \gamma_{SL} + \gamma_{LV} \cos \theta Rearranging gives the standard form: \cos \theta = \frac{\gamma_{SV} - \gamma_{SL}}{\gamma_{LV}} where \gamma denotes the interfacial tension. This thermodynamic approach assumes reversible displacement at constant temperature and treats interfaces as zero-thickness surfaces with energies independent of curvature. Key assumptions underlying the equation include a smooth, homogeneous, and rigid solid surface that is chemically and physically inert to the liquid, ensuring a unique equilibrium contact angle exists. Additionally, gravitational effects and line tension are neglected, with the liquid drop small enough to maintain spherical cap geometry, and the system at thermodynamic equilibrium without hysteresis. Geometrically, interprets the contact angle as the orientation where the tangential component of the liquid-vapor interfacial tension \gamma_{LV} \cos \theta precisely counteracts the imbalance between the solid-vapor and solid-liquid tensions along the . While foundational, Young's equation has limitations in that it does not account for contact angle or , which can alter the apparent angle on real surfaces.

Molecular and thermodynamic interpretations

The equilibrium contact angle arises from the thermodynamic principle of minimizing the total of the system at the three-phase contact line, where the , , and vapor phases meet. The relevant free energy expression for the interfaces is G = \gamma_{SL} A_{SL} + \gamma_{SV} A_{SV} + \gamma_{LV} A_{LV}, with \gamma_{ij} denoting the interfacial between phases i and j, and A_{ij} the corresponding interfacial area. Minimizing G with respect to variations in the contact angle \theta yields the condition for , resulting in Young's equation as the macroscopic manifestation: \gamma_{SV} = \gamma_{SL} + \gamma_{LV} \cos \theta. This thermodynamic framework, derived rigorously from variational principles, underscores that the contact angle balances the competing interfacial energies without invoking specific molecular details. At the molecular level, the contact angle reflects the balance of intermolecular forces across the interfaces, primarily van der Waals dispersion forces, which dominate in non-polar systems, and specific interactions like hydrogen bonding in polar liquids such as . Van der Waals forces, arising from transient dipole-induced dipole attractions, contribute to the interfacial tensions by integrating pairwise potentials over the molecular densities at each interface, influencing the effective spreading coefficient. For , hydrogen bonding between liquid molecules and surface sites enhances on polar substrates, reducing \gamma_{SL} and thus lowering \theta. These molecular interactions collectively determine the interfacial free energies that minimize G, leading to the observed \theta. Specific examples illustrate these effects for . On clean metal surfaces, strong polar interactions, including electrostatic and coordination bonds with surface atoms, result in very low or zero , indicating complete to favorable \gamma_{SL}. In , on non-polar surfaces like , the absence of polar sites limits interactions to weak van der Waals forces, yielding high around 110°, promoting poor . Adsorption of molecules at the interfaces can further modulate these forces, altering local \gamma_{ij} values. Surface heterogeneity at the molecular scale, such as variations in or atomic-scale roughness, leads to local fluctuations in the contact angle by creating patches with differing intermolecular interactions. For instance, regions with polar functional groups exhibit lower local \theta due to enhanced hydrogen bonding, while non-polar domains increase local \theta via dominant dispersion forces. These variations do not alter the global thermodynamic minimization but influence the apparent macroscopic angle through averaged interfacial energies.

Advanced thermodynamic models

Modified Young's equation

The modified Young's equation extends the classical relation by incorporating the effects of line tension, an excess associated with the three-phase contact line in systems. This adaptation accounts for deviations observed when the contact line becomes significant, leading to the : \cos \theta = \frac{\gamma_{SV} - \gamma_{SL}}{\gamma_{LV}} - \frac{\tau}{\gamma_{LV} r} where \theta is the contact angle, \gamma_{SV}, \gamma_{SL}, and \gamma_{LV} are the solid-vapor, solid-liquid, and liquid-vapor interfacial tensions, respectively, \tau is the line tension (with typical magnitudes on the order of $10^{-11} to $10^{-9} J/m), and r is the of the contact line. Note that the sign of the line tension term can vary in literature depending on the definition of \tau; the minus sign is common when \tau > 0 leads to increased apparent contact angles for smaller drops. The term involving line tension arises from the seminal work of Vesselovsky and Pertzov, who first proposed this correction in to explain size-dependent contact angles in systems. The derivation of this equation follows from minimizing the total of a sessile droplet, which includes contributions from the three interfacial areas and an additional line energy term \tau times the perimeter of the contact line (approximately $2\pi r). For a geometry, the equilibrium condition is obtained by setting the of the free energy with respect to r to zero, yielding the balance of horizontal force components at the contact line augmented by the line tension's contribution, which scales inversely with r. This approach builds on Gibbs' foundational concept of line tension as an excess energy per unit length at the contact line. This modification is particularly relevant for nanoscale droplets or nanoparticles, where the contact line radius r is small (on the order of nanometers), making the \tau / (\gamma_{LV} r) term comparable to the classical interfacial tension ratio and causing measurable deviations in \theta. For instance, in systems involving on hydrophobic substrates, positive line tension can increase the apparent contact angle for smaller drops, influencing behavior in applications like or microelectronics fabrication. Experimental validation comes from microscale measurements, such as on nanodroplets or goniometry on micron-sized sessile drops, which show linear correlations between \cos \theta and $1/r when fitting data to the modified , confirming line tension values around -10^{-10} J/m for various liquid-solid pairs and highlighting deviations from classical predictions at scales below 10 \mum. These results underscore the 's utility in interpreting size-dependent anomalies without invoking surface heterogeneity.

Line tension and Laplace pressure effects

The basic Young's equation provides a macroscopic force balance at the three-phase contact line but neglects contributions from and differences in the liquid, which become relevant for small or curved interfaces. The line tension correction follows the modified Young's equation given above. Laplace pressure, arising from the curvature of the liquid-vapor interface, influences the overall drop shape through the Young-Laplace equation \Delta P = \gamma_{LV} (1/R_1 + 1/R_2), where R_1 and R_2 are the principal radii of curvature. For small sessile drops or curved menisci, this pressure affects the local force balance at the contact line indirectly by altering the drop profile, leading to size-dependent apparent contact angles. This effect is particularly notable in confined geometries or small droplets where the meniscus curvature deviates significantly from a flat interface. For non-spherical drops, where or other forces distort the shape from a perfect , models integrate line tension and Laplace pressure effects into the overall interfacial energy minimization. The full approach couples the modified Young's equation with the Young-Laplace equation applied along the drop surface, with the line tension term at the boundary. This resolves retention forces and shape deviations observed in experiments, quantifying how line tension modulates the contact line energy while Laplace pressure dictates the internal hydrostatic distribution. Such models are essential for accurate predictions in microscale phenomena, as demonstrated in studies of drop adhesion and spreading on solids. In micro- and nanoscale sessile drops, these effects lead to an increased apparent contact angle \theta compared to macroscopic values, as the high Laplace pressure enhances curvature and the line tension term dominates for small r. For instance, water drops on hydrophobic surfaces with radii below 100 \mum exhibit \theta deviations of several degrees, attributed to positive line tension values on the order of $10^{-11} to $10^{-9} J/m, coupled with \Delta P exceeding atmospheric pressure. This size dependence has been verified through systematic measurements showing \theta rising inversely with drop volume in the sub-millimeter regime. Experimentally, line tension and Laplace pressure corrections are negligible for drops larger than approximately 1 mm, where r > 0.5 mm ensures the terms fall below 1% of the primary interfacial ratios, allowing standard Young's equation to suffice. Below this threshold, precise control of volume and surface preparation is required to isolate these contributions, often using axisymmetric fitting of drop profiles to disentangle the effects.

Contact angle hysteresis

Contact angle hysteresis is defined as the difference between the advancing contact angle (\theta_\mathrm{adv}) and the receding contact angle (\theta_\mathrm{rec}), expressed as \Delta \theta = \theta_\mathrm{adv} - \theta_\mathrm{rec}. This discrepancy occurs when the three-phase line moves across a , resulting from the pinning of the line at microscopic defects or heterogeneities that prevent smooth motion. The primary mechanisms driving contact angle hysteresis involve during line motion and the formation of metastable states. As the line advances, it must depin from local maxima created by surface variations, requiring additional input, while receding involves overcoming barriers to expose new areas. These processes lead to a range of stable contact angles rather than a single value, as described in early models of defect-induced pinning. Thermodynamically, hysteresis arises from free energy barriers due to local heterogeneities in surface energy, which create multiple local minima in the system's potential energy landscape. The equilibrium contact angle \theta_\mathrm{eq} from Young's equation is often approximated as the average of the advancing and receding angles: \theta_\mathrm{eq} \approx (\theta_\mathrm{adv} + \theta_\mathrm{rec})/2, reflecting how the hysteresis loop brackets the true equilibrium state. This interpretation emphasizes the role of energy dissipation in sustaining non-equilibrium configurations. Factors influencing hysteresis include intrinsic effects at the molecular scale, such as variations in intermolecular forces between , , and vapor phases, and extrinsic effects from macroscopic surface features like roughness. For example, isolated dirt particles or chemical impurities can act as strong pinning sites, amplifying hysteresis by creating localized energy traps that the contact line must overcome. The presence of hysteresis complicates measurements, as the true equilibrium contact angle \theta_\mathrm{eq} is difficult to observe directly; instead, experimental values often reflect metastable states within the hysteresis range, requiring careful control of surface conditions to minimize pinning effects.

Surface effects and dynamics

Roughness and texture influences

Surface roughness and texture significantly influence the apparent contact angle observed on a solid substrate by altering the effective interfacial areas between the liquid, solid, and vapor phases. For hydrophilic surfaces (intrinsic contact angle θ < 90°), roughness enhances wetting, leading to smaller apparent angles, while for hydrophobic surfaces (θ > 90°), it promotes non-wetting behavior with larger apparent angles. This amplification arises because roughness increases the solid-liquid interfacial area relative to the projected area, modifying the balance of interfacial energies. The Wenzel model describes homogeneous regimes where the liquid fully penetrates the surface asperities, creating a uniform solid-liquid interface. In this state, the apparent contact angle θ* is related to the intrinsic angle θ by the equation: \cos \theta^* = r \cos \theta where r is the roughness factor, defined as the ratio of the actual surface area to the (r > 1). Since r > 1, the model predicts that roughness exaggerates the intrinsic wettability: hydrophobic surfaces become more hydrophobic (θ* > θ), potentially leading to superhydrophobicity, while hydrophilic surfaces become more wettable (θ* < θ). This effect is particularly pronounced on microscale roughened surfaces, where r can exceed 1.5, resulting in apparent angles up to 150° or more for moderately hydrophobic materials. In contrast, the Cassie-Baxter model applies to composite wetting states, where air pockets are trapped beneath the droplet in surface textures, forming a solid-liquid-air composite interface. The apparent contact angle θ* is given by: \cos \theta^* = f_{SL} \cos \theta + f_{SV} (1 + \cos \theta_{LV}) - 1 where f_{SL} and f_{SV} are the fractional areas of the solid-liquid and solid-vapor interfaces projected onto the horizontal plane (f_{SL} + f_{SV} = 1), and θ_{LV} is the liquid-vapor contact angle (typically 180° for air). This simplifies to cos θ* = f_{SL} (cos θ + 1) - 1 when θ_{LV} = 180°, emphasizing the role of air fraction f_{SV} in increasing θ* toward 180°. The model explains superhydrophobicity (θ* > 150°) on textured surfaces with low solid fraction f_{SL} < 0.1, as the droplet rests on composite interfaces with minimal liquid-solid contact. Transitions between Wenzel and Cassie-Baxter states can occur under external stimuli such as pressure, vibration, or impact, driven by energy barriers between the metastable Cassie state and the more stable Wenzel state on hydrophobic rough surfaces. For instance, on nanopillared surfaces, increasing pillar height or reducing spacing stabilizes the Cassie state, preventing spontaneous transition to Wenzel wetting. Rough surfaces in either state exhibit enhanced contact angle hysteresis compared to smooth ones, as topographic features increase contact line pinning, leading to larger differences between advancing and receding angles (up to 50° or more). A prominent natural example is the lotus effect, observed on Nelumbo nucifera leaves, where hierarchical micro-papillae (5–9 μm diameter) topped with nanometer-scale tubules trap air, promoting Cassie-Baxter wetting with θ* ≈ 160° and low hysteresis for self-cleaning. Artificial superhydrophobic surfaces mimic this using micropillar arrays (e.g., 10 μm height, 10 μm spacing) fabricated via lithography, achieving θ* > 150° and roll-off angles < 5° in the Cassie state for applications like anti-fogging coatings. Both models have limitations, as they assume static equilibrium without accounting for dynamic air entrapment or expulsion during droplet deposition or movement, which can lead to metastable states not predicted by the equations.

Dynamic contact angles

Dynamic contact angles refer to the apparent angles formed at the three-phase contact line when the liquid interface moves relative to the solid surface, differing from the equilibrium contact angle due to the motion-induced effects. These angles depend on the speed of the contact line, with advancing contact angles typically increasing and receding angles decreasing as the velocity rises, bounding the range of contact angle hysteresis in the static limit. The hydrodynamic approach, developed through the Cox-Voinov law, describes the dynamic contact angle for low capillary numbers (Ca = μ v / γ_LV ≪ 1), where μ is the liquid viscosity, v is the contact line velocity, and γ_LV is the liquid-vapor surface tension. The law states that the cubed dynamic contact angle θ_d relates to the equilibrium angle θ_eq by \theta_d^3 - \theta_{eq}^3 = \frac{9 \mathrm{Ca} \ln(1/\mathrm{Ca})}{4}, capturing viscous dissipation in the wedge near the contact line. This relation holds for smooth surfaces and low speeds, where inertial effects are negligible. In contrast, the molecular-kinetic theory posits that motion arises from thermally activated molecular jumps across the contact line, leading to a friction model where the unbalanced Young stress drives the contact line velocity: \gamma_{LV} (\cos \theta_{eq} - \cos \theta_d) = \zeta v, where \zeta is the contact line friction coefficient. For small deviations, the change in angle scales linearly with velocity. This theory applies particularly to nanoscale contact lines or systems where hydrodynamic models break down. These theories find applications in processes like droplet spreading, where faster motion yields higher advancing angles, and imbibition in porous media, where dynamic effects control liquid penetration rates. Influences such as liquid viscosity, which amplifies deviations at higher speeds, and surface mobility, which can reduce friction, modulate the response; at high Ca, inertial effects cause deviations from these low-speed predictions. The dynamic advancing and receding angles thus define the effective hysteresis limits under motion, extending the static hysteresis concept.

Contact angle curvature

In the context of contact angle measurements, the curvature of the liquid meniscus plays a critical role, particularly when gravitational effects distort the drop shape from an ideal spherical cap. For larger drops, gravity flattens the meniscus, leading to an apparent contact angle that deviates from the true equilibrium value predicted by Young's equation. This distortion arises because the hydrostatic pressure gradient within the drop increases with height, altering the local curvature at the three-phase contact line. The Bond number, defined as Bo = \frac{\rho g R^2}{\gamma_{LV}}, where \rho is the liquid density, g is gravitational acceleration, R is the drop radius, and \gamma_{LV} is the liquid-vapor surface tension, quantifies the relative importance of gravitational forces to surface tension. When Bo > 0.08, gravitational effects become significant, causing measurable changes in the apparent contact angle, as demonstrated in experimental studies on sessile drops. To accurately model these curvature effects, the Bashforth-Adams equation provides a numerical framework for solving the Young-Laplace equation, which describes the axisymmetric drop profile under gravity. Developed in 1883, this approach parameterizes the drop shape using tables or iterative solutions based on the dimensionless Bond number \beta = \frac{\Delta \rho g b^2}{\gamma}, where b is the radius of curvature at the drop apex and \Delta \rho is the density difference across the interface. By fitting the observed drop profile to these solutions, the contact angle \theta can be extracted at the contact line, accounting for gravitational deformation. This method is particularly useful for intermediate Bond numbers where analytical approximations fail, enabling precise determination of \theta from the fitted profile. For small drops where Bo \ll 1, the gravitational influence is negligible, and the drop shape approximates a , simplifying contact angle calculations via geometric relations such as \theta = \arccos\left(1 - \frac{h}{R}\right), with h as the drop height. This approximation holds for drop sizes much smaller than the length \ell_c = \sqrt{\gamma_{LV}/(\rho g)}, typically on the order of millimeters for , allowing direct measurement without corrections. However, as drop size increases, the apparent \theta varies systematically; for instance, larger sessile drops exhibit reduced apparent angles due to flattening, while correction methods like perturbation solutions to the Bashforth-Adams (accurate to O(\epsilon^3), where \epsilon = \ell / \ell_c) are applied to recover the true value. These curvature effects manifest differently in pendant versus sessile configurations. In sessile drops, flattens the upper , decreasing the apparent \theta for hydrophilic surfaces and increasing sensitivity to size at high numbers (e.g., Bo \approx 1). drops, suspended from a surface, experience elongation at the bottom due to gravitational pull, leading to higher apparent \theta and greater shape asymmetry, which complicates fitting but allows measurement of high angles on super-repellent surfaces. Such differences highlight the need for configuration-specific corrections to ensure accurate \theta determination across number regimes. The also ties to differences, \Delta P = \gamma_{LV} \left( \frac{1}{R_1} + \frac{1}{R_2} \right), influencing local at the line.

Measurement techniques

Static sessile drop method

The static sessile drop method measures the contact angle by placing a small droplet of , typically in the microliter range, onto a to form a sessile drop. The side profile of the drop is then captured using a high-resolution camera or a telescope-equipped positioned perpendicular to the surface. To minimize gravitational distortion, drop volumes are kept small (e.g., 1–5 μL), ensuring the shape approximates a . Analysis begins with fitting a tangent line to the liquid-vapor interface at the three-phase contact line (solid-liquid-vapor), where the contact angle θ is defined as the angle between the tangent and the solid surface through the liquid. For small drops with negligible Bond number (Bo << 1), θ can be measured directly from the image using manual or automated goniometric tools with accuracies of ±1–2°. Larger drops, where gravity influences the profile, require axisymmetric curve-fitting software (e.g., based on least-squares optimization) to reconstruct the drop shape and extract θ, often incorporating Laplace equation solutions for precision. This technique is advantageous for its simplicity and non-invasiveness, requiring minimal sample preparation and liquid volume (as low as 0.5 μL), which makes it applicable to diverse solid surfaces including powders compressed into pellets or thin films. It provides reproducible results with typical precisions of ±2° for θ > 20°, and its optical nature allows real-time imaging without physical contact. Limitations include the assumption of negligible , which can cause θ to decrease over time (e.g., by 1–5°/min for volatile liquids like ), necessitating rapid imaging within seconds of deposition. The method is also highly sensitive to surface cleanliness, as contaminants or residues can increase apparent θ by up to 10–20° on otherwise hydrophilic surfaces. Key error sources arise from volume effects on ; excessive volume (>10 μL) flattens the under (Bo > 0.1), leading to underestimation of θ by 5–15° compared to theoretical values. Additional inaccuracies stem from operator variability in tangent fitting (up to ±3°) and surface heterogeneity, which may cause local variations in θ along the contact line.

Pendant drop method

The pendant drop method measures the contact angle by depositing a droplet onto the underside of a horizontal or inclined solid substrate, where it hangs in a (teardrop) shape due to . This is captured using a high-resolution camera with illumination for of the . The contact angle is evaluated at the upper three-phase contact line, where the liquid-vapor interface meets the . This approach is particularly suited for studying behavior under on inverted surfaces, such as in anti-fogging applications or to avoid bubble interference in some systems. Analysis of the pendant drop shape relies on fitting the experimental profile to the Young-Laplace equation, which governs the equilibrium shape by balancing , gravitational forces, and interfacial tensions: \Delta P = \gamma_\mathrm{LV} \left( \frac{1}{R_1} + \frac{1}{R_2} \right) Here, \Delta P represents the hydrostatic pressure difference across the interface (arising from \Delta \rho g z, where \Delta \rho is the density difference, g is , and z is height), \gamma_\mathrm{LV} is the liquid-vapor , and R_1, R_2 are the principal radii of curvature. Numerical or iterative algorithms solve this to parameterize the drop shape, enabling simultaneous extraction of the contact angle \theta (via tangent fitting at the contact line) and \gamma_\mathrm{LV}. Densities of the and surrounding must be known a priori for accurate fitting, and the Bond number (Bo = \Delta \rho g R_0^2 / \gamma_\mathrm{LV}, with R_0 as the apical radius) should be sufficiently high (typically Bo > 0.1) to ensure measurable deformation. This method offers key advantages, including the ability to determine both \theta and \gamma_\mathrm{LV} from a single experiment without additional force-based measurements, which is invaluable for validating Young's equation applications. It excels with low-surface-tension liquids (e.g., surfactants or oils, where \gamma_\mathrm{LV} < 30 mN/m), as reduced tension amplifies gravitational distortion for clearer profile resolution, and requires minimal sample volumes (typically 5–50 \muL), making it ideal for precious or volatile liquids like solvents in pharmaceutical testing. The optical nature avoids direct mechanical contact with the solid, reducing contamination risks during measurement. However, limitations include the need for precise drop deposition to achieve stable hanging without detachment, which can introduce artifacts in the contact line profile. The imaging setup demands precise alignment and higher resolution than sessile drop methods to resolve the elongated shape accurately, particularly for small drops where optical distortions may arise. Accuracy diminishes at very low Bond numbers, where insufficient deformation hampers fitting reliability, and evaporation effects must be controlled for volatile samples through rapid imaging or environmental enclosures. Applications of the pendant drop method are prominent in fields requiring precise wetting characterization of limited or reactive samples, such as evaluating surfactant efficacy in emulsions or assessing adhesion in microelectronics coatings with volatile components. It is especially useful for low-volume liquids in biomedical research, where traditional methods might require excessive material, and supports dynamic studies by monitoring shape evolution during substrate tilting.

Dynamic sessile drop method

The dynamic sessile drop method measures advancing and receding contact angles on solid surfaces to characterize contact angle hysteresis, which reflects energy dissipation during contact line motion due to surface imperfections. This optical technique extends the static sessile drop approach by introducing controlled motion to the droplet, enabling direct observation of dynamic wetting behavior that simulates processes like spreading or retraction in practical applications. In the tilting procedure, a sessile droplet is deposited on a horizontal substrate, which is then gradually inclined until the droplet begins to slide or roll; the advancing contact angle is measured at the downhill contact line edge, and the receding angle at the uphill edge, just prior to motion. Alternatively, the volume variation procedure involves attaching a needle to the droplet and using a syringe to incrementally add or withdraw liquid, advancing the contact line outward for the advancing angle measurement or inward for the receding angle. Both approaches require a high-resolution camera synchronized with the tilting stage or syringe pump to capture images at rates sufficient to track the contact line position. Analysis typically employs image processing software to fit tangent lines at the three-phase contact points across multiple frames, averaging angles from several cycles to account for variability and improve reproducibility. The advancing angle often exceeds 90° on hydrophobic surfaces, while the receding angle is lower, with hysteresis values ranging from 5° to 50° depending on surface energy and topography. This method offers advantages in directly capturing hysteresis without external forces, providing visual insight into real-world dynamic wetting like droplet spreading on inclined planes. It closely mimics natural processes, such as raindrop retention on leaves, as originally modeled by in 1962. However, limitations include sensitivity to vibrations that can prematurely trigger droplet motion in tilting setups and evaporation effects during prolonged volume withdrawal, which may alter the measured angles by up to 10° on volatile liquids. A common variant is the needle-attached drop configuration, where the syringe needle remains embedded in the droplet throughout the experiment to enable precise, continuous volume control and minimize detachment risks during retraction. This setup is particularly useful for low-hysteresis surfaces, where small volume changes suffice to induce contact line motion.

Wilhelmy plate and fiber methods

The Wilhelmy plate method, originally developed by Ludwig Wilhelmy in 1863 for surface tension measurements, determines contact angles through force tensiometry by assessing the wetting force on a solid probe immersed in a liquid. This technique relies on the balance of forces at the three-phase contact line, where the solid probe interacts with the liquid-vapor and solid-liquid interfaces. It is particularly valued for providing equilibrium or dynamic contact angle data on solid surfaces without relying on optical imaging. In the standard procedure for plates, a thin, flat probe—typically made of platinum, glass, or another inert material with known dimensions—is attached to a high-precision electrobalance and vertically dipped into the test liquid, such as water or an aqueous solution. The weight change ΔW upon partial immersion is recorded as the probe advances into the liquid, capturing the capillary force due to wetting. This force arises from the surface tension pulling along the wetted perimeter, expressed as ΔW = γ_{LV} \cos \theta \cdot P + V \Delta \rho g, where γ_{LV} is the liquid-vapor surface tension, θ is the , P is the wetted perimeter of the plate, V is the immersed volume, Δρ is the density difference between liquid and air, and g is gravitational acceleration. The plate's sharp edges ensure a well-defined meniscus, and measurements are conducted under controlled conditions to minimize vibrations or evaporation effects. Analysis involves correcting for the buoyancy term, which is linearly dependent on immersion depth d. By plotting the measured force F (converted from ΔW) divided by the wetted length L against d and extrapolating to d = 0, the contact angle is derived from the y-intercept via \cos \theta = F / (L \gamma_{LV}) at zero immersion, isolating the pure wetting contribution. For dynamic assessments, the immersion or withdrawal speed is systematically varied—typically from 0.1 to 10 mm/min—to capture advancing contact angles during immersion and receding angles during emersion, revealing hysteresis effects related to surface energetics. This approach yields reliable θ values across the full range (0° to 180°) for smooth, homogeneous surfaces. The single-fiber variant adapts the method for cylindrical probes like textile or carbon fibers, where the fiber is suspended vertically and immersed similarly, with forces measured on the order of micro-Newtons using sensitive tensiometers. The capillary force is given by F_c = p \gamma_L \cos \theta, with p as the fiber perimeter (πD, D being diameter), and buoyancy is often negligible for thin fibers (e.g., 6–20 μm diameter) due to minimal displaced volume. Corrections for meniscus shape and any baseline force drift (e.g., from liquid adsorption) are applied by linear regression on force-depth data or using multiple fibers (2–4) to amplify the signal while maintaining single-fiber precision; fiber diameter is pre-measured via microscopy for accurate perimeter calculation. This enables static equilibrium angles alongside dynamic advancing and receding values, with immersion pauses (e.g., 60 s) ensuring meniscus stabilization. Key advantages include high throughput for repetitive measurements on multiple samples, spatial averaging of the contact angle over the probe's length to reduce local heterogeneity impacts, and suitability for irregular or porous materials like powders, textiles, and fibers where optical methods falter. It also facilitates direct assessment of hysteresis without specialized imaging setups. However, limitations arise from edge effects on plates, which can distort the meniscus and lead to inaccuracies if edges are not perfectly sharp or clean; the method assumes uniform wetting along the probe, failing for highly heterogeneous or rough surfaces. For fibers, errors in diameter measurement propagate to contact angle uncertainties, and extremely low forces demand vibration-free environments.

Capillary rise methods

Capillary rise methods provide an indirect approach to determining the contact angle by observing the spontaneous penetration of a liquid into narrow tubes or porous media, driven by capillary forces. This technique relies on the balance between surface tension and viscous forces, allowing the contact angle to be inferred from the rate of liquid rise. Originally derived for cylindrical capillaries, the method has been extended to porous materials, where the effective pore geometry plays a key role. The foundational equation for this process is the Washburn equation, which for porous media approximates the square of the liquid rise height h as h^2 = \frac{\gamma_{LV} \cos \theta \cdot r \cdot t}{2 \eta}, where \gamma_{LV} is the liquid-vapor surface tension, \theta is the , r is the effective pore radius, \eta is the liquid viscosity, and t is time. This relation stems from assuming Poiseuille flow in cylindrical pores and connects to Young's equation through the capillary pressure term \gamma_{LV} \cos \theta. In practice, the procedure involves packing a porous sample, such as a powder bed, into a capillary tube or using a straight tube, then measuring the liquid uptake—either by height or mass gain—over time. Data are plotted as h^2 (or mass squared, proportional to volume) versus time, yielding a linear slope from which \cos \theta is calculated after determining the geometric factor (effective r) using a fully wetting reference liquid like n-hexane. This enables contact angle estimation for materials where direct optical observation is challenging. Key advantages of capillary rise methods include their non-optical nature, making them ideal for opaque solids, powders, or translucent media without requiring transparent interfaces. They are particularly useful for bulk characterization of wettability in industrial powders, such as pigments or ceramics, providing reproducible results when sample preparation is controlled. However, limitations arise from the underlying assumptions of laminar in uniform cylindrical pores, which may not hold for irregular or rough pore structures, leading to inaccuracies in r and thus \theta. Additionally, in rough or heterogeneous pores, liquid intrusion effects—such as threshold pressures or partial wetting—can distort the rise dynamics, requiring corrections for real materials. A variant for assessing contact angle hysteresis employs forced intrusion, as in mercury intrusion porosimetry (MIP), where non-wetting mercury is pressurized into pores during intrusion and extrusion cycles. The hysteresis in intrusion-extrusion volume-pressure curves reflects differences between advancing and receding contact angles (typically 140° for intrusion and around 107° for extrusion on many solids), allowing indirect quantification of wetting hysteresis in porous media.

Practical aspects

Typical contact angles for materials

Contact angles provide a quantitative measure of wettability, with surfaces classified based on the equilibrium contact angle θ for water: ultrahydrophilic surfaces exhibit θ < 10°, hydrophilic surfaces range from 10° to 90°, hydrophobic surfaces from 90° to 150°, and superhydrophobic surfaces exceed 150°.[web:9][web:33] These categories reflect the degree to which water spreads or beads up on a material, influenced by surface chemistry and energy. For on common surfaces, clean glass is highly hydrophilic with θ ≈ 0°–30°, allowing near-complete spreading due to silanol groups on the silica surface.[web:51] Polymers like poly(methyl methacrylate) (PMMA) show intermediate wettability with θ ≈ 70°, indicating partial spreading suitable for applications like optics.[web:63] Paraffin wax, a hydrophobic organic material, has θ ≈ 104°–111°, where forms a distinct bead.[web:74] Lotus-inspired superhydrophobic surfaces, mimicking hierarchical roughness on lotus leaves, achieve θ > 150°, enabling droplets to roll off with minimal .[web:7] Contact angle values vary significantly with surface conditions such as cleanliness and temperature. For instance, clean surfaces exhibit θ ≈ 40°–70° for after typical cleaning procedures, showing moderate wettability. Further by hydrocarbons or prolonged air exposure can increase θ to 70°–90° or higher, reducing wettability. effects can alter θ by 0.1°–0.5° per °C, with higher temperatures generally decreasing θ for on metals due to reduced interfacial .[web:27] Other liquids, particularly organics, often exhibit lower contact angles on metals compared to , promoting better . For example, on has θ ≈ 0° (complete spreading), while on the same surface has θ > 60°.[web:104] This difference arises from lower of organics and favorable interactions with metallic substrates. The following table summarizes representative equilibrium contact angles (θ) for on selected materials, drawn from experimental across categories:
Material CategoryExample Materialθ (°)Notes/Source
CeramicsClean glass0–30Highly hydrophilic; silanol-rich surface. [web:51]
PolymersPMMA~70Intermediate; varies with molecular weight. [web:63]
PolymersParaffin wax104–111Hydrophobic; smooth organic surface. [web:74]
MetalsClean gold40–70Moderately hydrophilic to hydrophobic; practical values after cleaning.
MetalsContaminated gold70–90Increased hydrophobicity from adsorbates.
Bio-inspiredLotus-like>150Superhydrophobic; hierarchical structure. [web:7]
These values illustrate wettability ranges but should be measured under controlled conditions, as roughness and impurities can amplify effects via models like Wenzel or Cassie-Baxter, though here we focus on , cases.

Methods for controlling contact angles

Methods for controlling contact angles involve chemical modifications to alter surface chemistry and physical techniques to introduce , enabling tailored wettability for diverse applications. These approaches can achieve hydrophilic surfaces with contact angles below 90° or hydrophobic ones exceeding 90°, often extending to superhydrophilic (<10°) or superhydrophobic (>150°) regimes through combined strategies. Chemical methods primarily target surface energy by functionalizing substrates with polar or non-polar groups. Plasma treatment, such as oxygen or nitrogen plasma, introduces hydroxyl and other polar groups on polymer or silicon surfaces, reducing water contact angles from 70–100° to near 0–10° for hydrophilization; for instance, silicon substrates shift from 45–74° to <2° after brief exposure. Silane coatings, like octadecyltrichlorosilane (OTS), deposit alkyl chains to lower surface energy, increasing contact angles to 90–140° for hydrophobization; polyurethane composites with OTS-modified silica nanoparticles reach 91°. Fluorination, via perfluoroalkylsilanes or direct gas-phase treatment, yields low-energy surfaces with θ >120°, such as fully fluorinated polymers at ~120°, enhancing hydrophobicity by minimizing liquid-solid interactions. However, as of 2025, per- and polyfluoroalkyl substances (PFAS) used in these treatments are subject to stringent regulations, including mandatory reporting under TSCA and bans in several jurisdictions, prompting shifts to PFAS-free alternatives. Physical methods engineer to amplify intrinsic wettability, often referencing Wenzel and Cassie-Baxter regimes where micro/nanostructures increase apparent angles on hydrophobic bases or decrease them on hydrophilic ones. , including UV with SU-8 to form micropillars (e.g., 33 μm height, 42 μm spacing) or micromachining for pyramid-like features (~2 μm), creates hierarchical textures; combined with PTFE coatings, these yield superhydrophobic surfaces with apparent angles of 151–169°. fabricates rough metallic surfaces, such as dendritic or deposits, promoting Cassie-Baxter states for superhydrophobicity; electrodeposited structures on achieve 164° with sliding angles <5°, enabling robust self-cleaning. (Note: Specific paper from collection.) Recent advances post-2020 emphasize dynamic control via stimuli-responsive surfaces and nanomaterials. Stimuli-responsive coatings, such as pH-sensitive polyelectrolytes or temperature-responsive polymers, enable reversible switching; for example, pH changes from 4 to 10 can shift contact angles from 20° (hydrophilic) to 110° (hydrophobic), while thermal triggers around 32°C toggle between 60° and 140° for on-demand wettability in microfluidics. Nanomaterials like offer tunable wetting through doping or layering; n- or p-type doping shifts water contact angles by up to 13° (e.g., from 85° to 98°), with multilayer graphene exhibiting translucent wettability tunable via layer count. Self-assembled monolayers (SAMs), such as fluorinated phosphonic acids on nanowire forests, produce bio-inspired omniphobic surfaces repelling water (150–155°) and oils, surpassing traditional hydrophobicity for anti-fouling. These controlled contact angles underpin applications like anti-fogging coatings (low θ <30° via plasma-treated glass to spread water films), self-cleaning surfaces (high θ >150° on lithographed textiles to roll off contaminants), and (tunable θ via graphene-SAM hybrids for droplet manipulation without pinning).

References

  1. [1]
    Contact angle | KRÜSS Scientific
    The contact angle (wetting angle) is a measure of the wettability of a solid by a liquid. Background. In the case of complete wetting (spreading), the contact ...Dynamic contact angle · Static contact angle · 3D Contact Angle method · Hysteresis
  2. [2]
    Contact Angles - Chemistry LibreTexts
    Jul 30, 2024 · The contact angle is an angle that a liquid creates with a solid surface or capillary walls of a porous material when both materials come in contact together.
  3. [3]
    Contact Angle - an overview | ScienceDirect Topics
    Contact angle (CA) is defined as the angle formed between a liquid interface and a solid surface at the point of contact, which is crucial for accurately ...
  4. [4]
  5. [5]
    Young's equation | KRÜSS Scientific
    The equation is valid for three-phase systems in thermodynamic equilibrium for ideal (smooth and chemically homogenous) solids and pure liquids.
  6. [6]
    III. An essay on the cohesion of fluids - Journals
    An essay on the cohesion of fluids. Thomas Young. Google Scholar · Find this ... Contact Angle Amplification?, Langmuir, 10.1021/acs.langmuir.2c00478, 38 ...
  7. [7]
    Historical aspects of capillarity and capillary condensation
    In contrast, Laplace's 1806 work involved a more 'modern' energetic approach to capillary attraction, but avoided any direct reference to surface tension. The ...
  8. [8]
    Gibbsian Thermodynamics of Wenzel Wetting ... - ACS Publications
    Dec 23, 2019 · The contact angle (θ) of a drop in these states is commonly predicted by the Cassie–Baxter or Wenzel equations, respectively, but the accuracy ...
  9. [9]
    RESISTANCE OF SOLID SURFACES TO WETTING BY WATER
    Superhydrophobic Nanocoatings Fabricated by Materials with Low Young's Contact Angle. ... Energy Dissipation during Wenzel Wetting via Roughness Scale ...
  10. [10]
    Wettability of porous surfaces - Transactions of the Faraday Society ...
    Wettability of porous surfaces. ABD Cassie and S. Baxter. Abstract. The first page of this article is displayed as the abstract.
  11. [11]
    Surface Science – How it all began, from an experimental perspective
    Dec 18, 2017 · Pure surface physics got a boost in the late 1950s and through the 60s from developments in semiconductor technology and microelectronics. The ...
  12. [12]
    Contact angles: history of over 200 years of open questions
    Aug 7, 2025 · The origins and importance of the contact angle in analysis of three-phase systems date back to the famous works on cohesion and adhesion of ...
  13. [13]
    Local thermodynamic derivation of Young's equation - ScienceDirect
    Apr 15, 2004 · A derivation of Young's equation based on the energy balance near the contact line is presented. Our proposal is rigorous and avoids the errors identified in ...Missing: seminal | Show results with:seminal
  14. [14]
    [PDF] Local thermodynamic derivation of Young's equation
    Deviations of the contact angle from Young's equation are discussed in several cases: surfaces of high curvature and line tension. Finally, the relationship ...
  15. [15]
    [PDF] chapter 3 determination of talc's surface free energies from contact ...
    The derivation of the Young's equation assumes that the solid surface is smooth, homogeneous and rigid; it should also be chemically and physically inert with ...
  16. [16]
    The rigorous derivation of Young, Cassie–Baxter and Wenzel ...
    Jan 4, 2008 · The rigorous derivation of Young, Cassie–Baxter and Wenzel equations carried out in the framework of the unified thermodynamic approach is presented.Missing: seminal | Show results with:seminal
  17. [17]
    Surface energy and wettability of van der Waals structures
    The wetting behaviour of surfaces is believed to be affected by van der Waals (vdW) forces; however, there is no clear demonstration of this.
  18. [18]
    Molecular Dynamics Study on the Effect of Surface Hydroxyl Groups ...
    Hydroxyl groups on graphite cause a smaller water contact angle, due to hydrogen bonds, in oil-water-graphite systems. The contact angle changes with hydroxyl ...
  19. [19]
    Structure, Dynamics, and Wettability of Water at Metal Interfaces
    Oct 15, 2019 · Using the state-of-the-art simulations, we find that the clean metal surfaces are actually superhydrophilic and yield zero contact angles.
  20. [20]
    Interpretation of contact angle in solid-hydrocarbon-water system
    For these organic liquids and water, the values of the contact angle on paraffin wax, polytetrafluoroethylene, polyethylene, polyethylene terephthalate, and ...
  21. [21]
    Probing surface tension additivity on chemically heterogeneous ...
    Apr 1, 2011 · We use molecular simulations to measure the microscopic analogue of contact angle, θc, of aqueous nanodrops on heterogeneous synthetic and ...
  22. [22]
    Droplet slipperiness despite surface heterogeneity at molecular scale
    Oct 23, 2023 · Surface heterogeneity is generally acknowledged as the major cause of increased contact angle hysteresis and contact line friction of droplets.
  23. [23]
    Line-Tension-Based Modification of Young's Equation for Rock/Oil ...
    Oct 27, 2009 · Vesselovsky and Pertzov (1936) were the first to determine the line tension by studying the bubble-size dependence of contact angle and ...Missing: seminal | Show results with:seminal
  24. [24]
    Drop size dependence of soil-water contact angle in relation to the ...
    Dec 13, 2012 · The decrease in θ A with increasing drop size was explained as a result of the line tension effect using the modified Young's equation.Missing: seminal | Show results with:seminal
  25. [25]
    [PDF] arXiv:1805.02837v1 [cond-mat.soft] 8 May 2018
    May 8, 2018 · This modified. Young equation (1) is routinely used to deduce the magnitude of line tension from the computer simulation as well [7,8,17,18].
  26. [26]
    A critical assessment of the line tension determined by the modified ...
    Aug 8, 2018 · The modified Young's equation is a useful tool to predict the macroscopic contact angle based on a linear fit of the measured contact angles at the nanoscale.
  27. [27]
    Line Tension and Drop Size Dependence of Contact Angle at ... - NIH
    We show that the contact angle is indeed dependent on the drop size for small drop diameters and determine the magnitude of the apparent line tension.
  28. [28]
    Surface Science Letters Line energy, line tension and drop size
    Jul 15, 2008 · In this paper we quantify this relation of the retention force to the line tension, and the Laplace pressure. Apparently this way we can resolve ...
  29. [29]
    Dependency of Contact Angles on Three-Phase Contact Line - MDPI
    The contact angle was insensitive to droplet size, implying that the line tension effects were minimal. Two explanations were offered for this behavior. The ...<|control11|><|separator|>
  30. [30]
    Application of Young-Laplace with size-dependent contact angle ...
    The capillary pressure depends on the interfacial tension, contact angle, and throat size based on the Young-Laplace equation (YL) as follows: P c = 2 γ | cos ...
  31. [31]
    A model for contact angle hysteresis | The Journal of Chemical Physics
    Jul 1, 1984 · We discuss the behavior of a liquid partially wetting a solid surface, when the contact angle at equilibrium θ 0 is small, but finite.
  32. [32]
    Coexistence and transition between Cassie and Wenzel state on ...
    May 26, 2009 · We have studied 4 conditions that affect the transition between the Wenzel and Cassie state: (i) The height of nanopillars, (ii) the spacing ...
  33. [33]
  34. [34]
  35. [35]
    Effect of gravity on contact angle: An experimental investigation
    Sep 1, 1998 · This results in a. Bond number of 0.09. If the Bond number is less than 0.08, it has been previ- ously shown that the curvature of the liquid– ...
  36. [36]
    [PDF] Assessing the Accuracy of Contact Angle Measurements for Sessile ...
    Gravity-induced sagging can amplify variations in goniometric measurements of the contact angles of sessile drops on super liquid-repellent surfaces.
  37. [37]
    [PDF] Low-bond axisymmetric drop shape analysis for surface ... - CUHK EE
    The Young–Laplace equation for axisymmetric drops was first solved numerically in 1883 by Bashforth and Adams [9]. The authors tabulated solutions to (4) c ...
  38. [38]
    [PDF] Accuracy of Axisymmetric Drop Shape Analysis in Determining ...
    Aug 9, 2017 · Theoretical profiles of these drop or bubble configurations are described by the Bashforth−Adams equation. (eq 1), where the tangential angle (φ) ...
  39. [39]
    Spreading of drops on solid surfaces in a quasi-static regime
    which implies that at small Bond numbers the free-surface shape at every moment may be approximated by a spherical cap and so the dynamic contact angle can be ...
  40. [40]
    Modeling the effect of shape deformation induced by gravity on the ...
    Feb 21, 2024 · The Bond number ranges from 0.2 to 2 for drops (both sessile and pendant) on an hydrophilic substrate with a contact angles equal to 40°, while ...
  41. [41]
    Contact Angle Measurement of Small Capillary Length Liquid in ...
    Apr 7, 2017 · Comparing the water droplet and the small FC-72 droplet in the middle, they share the same Bond number (Bo = 0.73) and thus the same profile ...
  42. [42]
    Sessile Drop Method - an overview | ScienceDirect Topics
    The sessile drop method is defined as a straightforward and accurate technique for direct observation and measurement of apparent contact angles (CAs) of ...Missing: limitations | Show results with:limitations
  43. [43]
  44. [44]
    The debate on the dependence of apparent contact angles on drop ...
    In this review, we presented the views on the dependence of apparent contact angles on liquid/solid contact area beneath the drop or three-phase contact line.
  45. [45]
  46. [46]
  47. [47]
    Pendant drop method for surface tension measurements
    ### Summary of Pendant Drop Method for Surface Tension Measurements
  48. [48]
    Design and accuracy of pendant drop methods for surface tension ...
    In this paper, the design and accuracy of pendant drop methods is considered in detail. Dimensional analysis is used to describe similarity in pendant drop ...
  49. [49]
    Surface & Interfacial Tension - Nanoscience Instruments
    In the Young-Laplace equation, the parameter is adjusted to find the optimal drop shape is the drop shape factor β = ΔρgR02 / γ which contains two unknowns: the ...<|separator|>
  50. [50]
    Contact angle hysteresis: a review of fundamentals and applications
    Sep 14, 2012 · Contact angle hysteresis is an important physical phenomenon. It is omnipresent in nature and also plays a crucial role in various industrial processes.
  51. [51]
    Measurement of dynamic/advancing/receding contact angle by ...
    Aug 1, 1996 · By controlling the humidity of air phase surrounding the drop, this technique can measure the advancing and receding contact angles and monitor ...
  52. [52]
    Studies at phase interfaces. I. The sliding of liquid drops on solid ...
    The study examines how surface properties affect spray retention, using a theory based on drop size, surface tension, and contact angles to predict retention. ...
  53. [53]
    Wilhelmy Plate Method - an overview | ScienceDirect Topics
    Another method for measuring the surface tension is the Wilhelmy plate method, which was first used by the German chemist Ludwig Wilhelmy in 1863 [2]. In ...
  54. [54]
    [PDF] The Wilhelmy method: a critical and practical review - iris@unitn
    The Wilhelmy method, proposed in 1863, is an ancient technique for measuring surface tension and contact angle, using a plate method to evaluate the 'capillary ...Missing: seminal | Show results with:seminal
  55. [55]
    [PDF] Surface energy determination of fibres for Liquid Composite ...
    Nov 13, 2020 · Based on the meniscus mass measurement and following the Wilhelmy relationship (Eq. (1)) the fibre/liquid contact angle can be obtained by ...Missing: seminal paper
  56. [56]
    Contact angle measurement of single fiber using Wilhelmy plate method
    ### Summary of Single Fiber Contact Angle Measurement Using Wilhelmy Plate Method
  57. [57]
    Washburn method | KRÜSS Scientific
    The Washburn method is a method for measuring the contact angle and the surface free energy of porous substances such as bulk powder or pigments.
  58. [58]
    Advances in the Fabrication and Characterization ... - PubMed Central
    When θ is between 10° and 90°, a surface is considered to be hydrophilic; a value between 90° and 150° indicates that the surface is hydrophobic. A surface is ...Missing: paraffin | Show results with:paraffin<|control11|><|separator|>
  59. [59]
    [PDF] 1 Assessment of Physical Properties of Water Repellent Soils - arXiv
    Water contact angles are classified as ultra-hydrophilic (≈0º), hydrophilic (0º-. 90º), hydrophobic (90º-150º) and ultra-hydrophobic (>150º) (Chieng et al. 2019) ...
  60. [60]
    Systematic Study of Wettability Alteration of Glass Surfaces by ... - NIH
    Sep 28, 2023 · Iglauer et al. concluded that the contact angle on clean glass surfaces is relatively low (0–30°). The contact angles obtained in this study ( ...
  61. [61]
    [PDF] Droplet Behaviour on Superhydrophobic surfaces - UCL Discovery
    Specifically, those surfaces whose water contact angles are greater than 150° are named superhydrophobic surfaces, which means they are extremely water- ...Missing: paraffin | Show results with:paraffin
  62. [62]
    Hydrophilic and Hydrophobic Surfaces: Features of Interaction ... - NIH
    Aug 30, 2023 · This review paper describes methods for obtaining hydrophilic and hydrophobic surfaces. The features of the interaction of liquid droplets with such surfaces ...
  63. [63]
    Plasma Treatment to Create a Hydrophilic Surface on Silicon ...
    Jun 13, 2024 · See test below showing the initial contact angle of 45 to 74 degrees – after plasma treatment in the AutoGlow 1000 the contact angle was < 2 ...
  64. [64]
    Hydrophobic silanes-modified nano-SiO₂ reinforced polyurethane ...
    Aug 27, 2025 · WCA measurements show that the pure PU surface has a contact angle of 34°, which decreases to 5° with NS, increases to 91° with OTES-modified NS ...
  65. [65]
    Direct Fluorination as Method of Improvement of Operational ... - NIH
    Nov 28, 2020 · The values θ1 were chosen to be 120° for a fully fluorinated hydrophobic surface [93] and 0° for a surface that is completely wetted with water ...
  66. [66]
    Combination of Lithography and Coating Methods for Surface ...
    Jul 3, 2013 · We present two different methods for constructing surfaces with tunable wetting properties, both of them based on the concept of combining a lithographic ...
  67. [67]
    Bioinspired Surfaces With Switchable Wettability - Frontiers
    Aug 11, 2020 · In this minireview, we have summarized the typical stimulated-responsive surfaces with switchable wettability including thermal/photothermal, ...
  68. [68]
    Doping-Induced Tunable Wettability and Adhesion of Graphene
    Graphene's water contact angle changes by as much as 13° as a result of a 300 meV change in doping level. Upon either n- or p-type doping with subsurface ...Supporting Information · Author Information · Acknowledgment · References
  69. [69]
    Wettability of graphene, water contact angle, and interfacial water ...
    May 12, 2022 · Study on the surface energy of graphene by contact angle ... Water at hydrophobic surfaces: weak hydrogen bonding and strong orientation effects.
  70. [70]
    Development of omniphobic behavior in molecular self‐assembled ...
    Oct 13, 2015 · The wetting characteristics of self-assembled monolayers (SAMs) on three different surface ... SAMs, which displayed contact angles of 99.5 ...