Continuous mandatory ventilation

Continuous mandatory ventilation (CMV), also known as assist-control ventilation (ACV), is a fundamental mode of used in intensive care units to provide full respiratory support to patients with acute or other conditions impairing spontaneous breathing. In this mode, the delivers breaths at a predetermined rate (the "mandatory" component), ensuring a minimum , while allowing patient-initiated breaths (the "assist" component) that trigger full support at the same settings. CMV can operate in either volume-controlled (VC-CMV) or pressure-controlled (PC-CMV) variants: VC-CMV delivers a fixed with variable airway pressure, whereas PC-CMV applies a constant inspiratory pressure with resulting varying based on and resistance. Clinically, CMV is indicated for scenarios requiring precise control over , such as (ARDS), metabolic or , and postoperative respiratory support, where it has demonstrated mortality benefits in ARDS patients through protocols like low ventilation (6-8 mL/kg ideal body weight). Key settings include or inspiratory pressure, (typically 12-20 breaths per minute), (FiO2), and (PEEP) to maintain alveolar recruitment. Unlike intermittent mandatory ventilation (IMV), which permits unsupported spontaneous breaths between mandatory ones, CMV ensures every breath—whether patient- or time-triggered—is fully supported, minimizing the patient's but potentially leading to complications if not monitored. Advantages of CMV include enhanced patient-ventilator synchrony, reliable delivery for correcting acid-base imbalances, and reduced diaphragmatic workload compared to modes without full support. However, it carries risks such as breath stacking and auto-PEEP in tachypneic patients, which can cause and , particularly in volume-controlled settings with non-compliant s; pressure-controlled CMV mitigates peak pressure risks but may result in inconsistent tidal volumes. involves assessing peak/plateau pressures, tidal volumes, and end-tidal CO2 to adjust for optimal oxygenation and while avoiding ventilator-induced injury. Overall, CMV remains one of the most widely used modes due to its simplicity and efficacy in providing total ventilatory control during critical illness.

Fundamentals

Definition and Purpose

Continuous mandatory ventilation (CMV), also known as assist-control ventilation (ACV), is a mode of in which the delivers mandatory breaths at a predetermined rate using either a fixed or inspiratory pressure, while allowing patient-initiated breaths that trigger full support at the same settings. This ensures a minimum through time-triggered control breaths, supplemented by patient-triggered assist breaths if effort is detected. The primary purpose of CMV is to provide full or assisted respiratory support to patients with impaired breathing, ensuring consistent for adequate and oxygenation. By delivering supported breaths whether initiated by the patient or the machine, CMV minimizes the patient's and prevents , particularly in conditions where spontaneous efforts are unreliable or insufficient.

Basic Mechanisms

Continuous mandatory ventilation (CMV) operates through a series of fundamental mechanisms that govern breath delivery, with breaths being either time-triggered or patient-triggered. The breath cycle consists of three primary : , during which the delivers gas into the according to preset parameters; expiration, a passive where gas is expelled from the ; and a brief pause, often an inspiratory hold, allowing for stabilization and measurement of lung mechanics such as plateau pressure. The precisely controls the timing and flow patterns across these phases to maintain consistent ventilation. The triggering mechanism in CMV can be time-based for mandatory control breaths, where the ventilator initiates each breath at predetermined intervals set by the , or patient-based for assist breaths, where sensitivity to airflow or pressure changes detects patient effort to trigger delivery. This dual approach ensures a minimum while accommodating patient demand. Cycling, which terminates the inspiratory phase and initiates expiration, relies on predefined thresholds such as time, flow, or volume to end gas delivery and prevent prolonged inspiration. For instance, in time-cycled systems, inspiration concludes after a set duration, while volume- or flow-based cycling halts delivery once the target is reached, ensuring efficient breath completion. Limiting mechanisms establish upper bounds on key variables during inspiration to safeguard against lung injury, including overdistension or barotrauma. These include pressure limits to cap airway pressure, volume limits to restrict tidal volume delivery, and flow limits to moderate inspiratory gas speed, with alarms activating if thresholds are approached. The in CMV is determined by the clinician-set backup value, representing the minimum mandatory breaths per minute—for example, a of 12 breaths/min establishes a 5-second between control breaths, calculated as 60 divided by the set . The actual may exceed this if the patient triggers additional assist breaths. These mechanisms apply generally across volume-controlled and pressure-controlled implementations of CMV.

Primary Modes

Volume-Controlled CMV

In volume-controlled continuous mandatory ventilation (VC-CMV), the ventilator delivers a preset (VT) at a controlled during each mandatory breath, while allowing airway pressure to vary based on the patient's and . This mode ensures consistent alveolar ventilation by prioritizing volume delivery, making it suitable for full ventilatory support in controlled settings. Key parameters include (VT), typically set at 6-8 mL/kg of ideal body weight to promote lung protection; (RR), often 12-16 breaths per minute; inspiratory time (Ti), adjusted to achieve an inspiratory-to-expiratory (I:E) ratio of 1:2 or 1:3; peak inspiratory flow rate, commonly 40-60 L/min; and (PEEP), starting at 5 cmH₂O and titrated as needed. The serves as the primary variable, set at 30-40 cmH₂O to terminate inspiration early if exceeded, thereby preventing from excessive airway . Breaths in VC-CMV are volume-cycled, meaning inspiration ends once the target VT is delivered, with a time-cycled mechanism as a backup to ensure cycle completion. Flow waveforms can be configured as constant (square), decelerating to 50% of peak, or decelerating to zero, allowing optimization of gas distribution and inspiratory time based on patient needs—constant flow shortens Ti for faster delivery, while decelerating patterns may improve ventilation-perfusion matching. Minute ventilation (VE) is calculated as VE = VT × RR, for example, 500 mL × 12 breaths/min yields 6 L/min, providing a reliable measure of total gas exchange. Advantages of VC-CMV include guaranteed delivery of the set VT regardless of changing lung mechanics, which is particularly beneficial in patients with restrictive lung diseases or variable compliance, unlike pressure-controlled modes that may result in fluctuating volumes.

Pressure-Controlled CMV

In pressure-controlled continuous mandatory ventilation (PC-CMV), the ventilator delivers breaths by applying a predetermined inspiratory pressure (Pinsp), typically set at 15-25 cmH₂O above (PEEP), for a fixed inspiratory time, which results in a variable (VT) that depends on the patient's and resistance. This mode prioritizes pressure limitation to safeguard against excessive airway pressures, differing from volume-controlled approaches by allowing VT to fluctuate rather than maintaining a constant delivery. Key parameters in PC-CMV include the inspiratory pressure (Pinsp), inspiratory time (), respiratory rate (), PEEP, and inspiratory-to-expiratory (I:E) ratio, which collectively determine the breath profile and overall minute ventilation. Monitoring of delivered tidal volume is essential, as it can vary widely; alarms for high VT help prevent volutrauma in high-compliance scenarios, though inspiration remains time-cycled. Inspiration is time-cycled, concluding at the preset irrespective of the volume or flow achieved during the breath. The pressure waveform in PC-CMV features a square wave pattern, achieving a rapid rise to the target Pinsp followed by a sustained plateau, which produces a decelerating flow profile that mimics more natural respiratory dynamics. An approximate equation for the delivered tidal volume is: VT \approx (P_{insp} - PEEP) \times C where C represents lung compliance, typically ranging from 20-50 mL/cmH₂O in critically ill patients. A primary advantage of PC-CMV is its inherent limitation of peak inspiratory pressures, which reduces the risk of and -induced , making it particularly suitable for (ARDS) management where lung-protective strategies are essential.

Advanced Modes

Dual-Control Modes

Dual-control modes in continuous mandatory ventilation integrate targeting with regulation to deliver a predetermined (VT) while minimizing airway , thereby enhancing safety and adaptability in mechanical support. These modes function as time-cycled, mandatory breaths where the uses as the variable to adjust inspiratory breath-to-breath, ensuring consistent amid varying conditions. The core principle of dual-control modes, exemplified by pressure-regulated volume control (PRVC) and similar variants like volume control plus (VC+), prioritizes achieving a set VT through automated inspiratory pressure modulation, blending the reliability of volume assurance with pressure-limited delivery to protect against overdistension. In operation, the ventilator begins with a test breath—typically in volume control—to assess dynamic and establish an initial pressure estimate, followed by pressure-controlled breaths featuring decelerating patterns. Subsequent pressures are then fine-tuned based on the difference between delivered and target VT, with adjustments limited to incremental steps (e.g., up to 3 cmH₂O per breath in some systems) to avoid excessive changes. This adjustment follows a where the inspiratory for the next breath is updated as P_{\text{insp},n} = P_{\text{insp},n-1} + [\Delta P](/page/Delta), with [\Delta P](/page/Delta) proportional to the VT error (shortfall or excess) from the prior breath, constrained by user-defined upper and lower limits to maintain safety. Key parameters include the target VT (commonly 4–12 mL/kg ideal body weight), (RR, e.g., 15–30 breaths/min), inspiratory time (Ti, 0.6–1 second), (PEEP, often 5 cmH₂O), and limits (upper typically 30–40 cmH₂O to cap ). If the upper limit is hit and VT cannot be met, the reverts to fixed pressure-controlled CMV with an alarm indication, such as "Regulation Pressure Limited." These modes offer advantages by guaranteeing for stable while capping pressures to mitigate volutrauma and risks, often achieving equivalent VT to volume-controlled modes at 20–30% lower peak pressures and improving patient-ventilator synchrony with fewer alarms (e.g., reduced from 9.1 to 3.3 alarms/hour in comparative studies). Dual-control modes build on principles from volume- and pressure-controlled CMV by adding adaptive for optimization. Proprietary examples include PRVC and Adaptive Pressure Ventilation on Maquet/Getinge Servo-i ventilators, which adjust pressure in 3 cmH₂O steps up to 5 cmH₂O below the upper limit; AutoFlow on Dräger systems, which regulates flow and pressure in volume-targeted breaths to ensure minimum pressure delivery; and VC+ on platforms, employing similar incremental pressure titration for VT consistency.

Hybrid Variants

Hybrid variants of continuous mandatory ventilation (CMV) encompass specialized modes that augment the mandatory breath delivery with adaptive elements, such as targeting, while retaining a core mandatory framework to ensure reliable support. These variants incorporate hybrid triggers or automated adjustments to better synchronize with efforts, distinguishing them from purely mandatory or fully spontaneous approaches. They emerged as refinements to address limitations in traditional CMV, particularly in scenarios requiring precise control over delivery without excessive or exposure. A key example is volume guarantee pressure control (VGPC), a mode tailored for neonatal that hybridizes pressure-controlled CMV with volume assurance. In VGPC, the ventilator delivers pressure-limited breaths but dynamically adjusts the (PIP) on a breath-by-breath basis to achieve a preset (VT), based on measurements of expired VT from the previous breath. This mechanism mitigates variability in and resistance common in preterm infants, promoting more consistent alveolar and reducing the incidence of hypocarbia or hypercarbia. VGPC functions effectively even with moderate endotracheal tube leaks up to 50%, though accuracy diminishes beyond that threshold. Essential parameters in these modes include the target VT (4–5 mL/kg), backup (e.g., 30–70 breaths/min as a safety minimum), and protective limits such as maximum PIP (25–30 cm H₂O). These settings allow clinicians to tailor support while safeguarding against or volutrauma. Clinically, hybrid variants find niche applications in pediatric scenarios, where VGPC supports preterm neonates with respiratory distress by enabling gradual reductions and shorter durations. These modes relate briefly to dual-control foundations through their use of automated pressure-volume adjustments but emphasize neonatal-specific adaptations. Limitations of hybrid variants include risks of over-ventilation if expired VT measurements are inaccurate due to leaks, and challenges in environments with significant asynchronous triggering that could lead to inefficient or .

Clinical Applications

Indications and Settings

Continuous mandatory ventilation (CMV) is primarily indicated in clinical scenarios where patients require full ventilatory support due to inadequate spontaneous breathing, such as during general , neuromuscular blockade, severe (ARDS), status asthmaticus, or conditions carrying a high risk of apnea. Patient selection for CMV focuses on adults and children requiring full ventilatory support, often sedated or pharmacologically paralyzed; in awake or lightly sedated patients, it may lead to patient-ventilator asynchrony, necessitating careful monitoring and possible sedation adjustment. Initial ventilator settings for CMV typically include a of 10-20 breaths per minute, of 6-8 mL/kg predicted body weight in volume-controlled modes or inspiratory pressure of 15-25 cmH₂O in pressure-controlled modes, (PEEP) of 5-15 cmH₂O, and (FiO₂) titrated to maintain (SpO₂) between 88-95%. Ongoing monitoring during CMV involves arterial blood gas analysis to assess and of (PaCO₂), maintenance of plateau pressures below 30 cmH₂O to avoid , and evaluation for auto-PEEP, particularly in obstructive conditions like status asthmaticus. Ventilator alarms should be set for high and low pressure limits, volume discrepancies, and apnea detection to ensure and prompt intervention. In special populations such as those with ARDS, a lung-protective strategy employs lower volumes of 6 mL/kg predicted body weight, as established by the ARDSNet , to minimize ventilator-induced lung injury while permitting mild if necessary.

Outcomes and Considerations

Continuous mandatory (CMV) is designed to achieve stable by delivering preset breaths at controlled rates and volumes or pressures, particularly beneficial in patients where spontaneous breathing is inadequate. In such cases, CMV assumes the majority of the , supporting patient efforts or providing full breaths as needed to maintain oxygenation and , often improving overall respiratory stability in acute settings like postoperative recovery or . Despite these benefits, CMV carries risks of complications, including and volutrauma, which are more pronounced in volume-controlled CMV (VC-CMV) due to potential delivery of excessive volumes leading to alveolar overdistension. Pressure-controlled CMV (PC-CMV) may mitigate some pressure-related injuries but has been associated with a potentially higher risk in certain trials, though evidence quality is low. Ventilator-induced lung injury (VILI) remains a key concern across both modes, exacerbated by high plateau pressures or driving pressures, contributing to and multi-organ dysfunction. Additionally, high inspiratory pressures or (PEEP) in CMV can induce by impeding venous return and increasing right ventricular , particularly in patients with preexisting hemodynamic . Clinical considerations for CMV include the necessity for deep or neuromuscular to prevent patient-ventilator dyssynchrony, as spontaneous efforts against mandatory breaths can increase injury risk and discomfort; this often involves agents like or cisatracurium, targeting light to moderate levels once stabilized. Weaning from CMV poses challenges due to its full ventilatory support, requiring careful transition to assisted modes like pressure support to assess respiratory muscle and avoid abrupt failure. from randomized trials in (ARDS) suggests PC-CMV may offer a modest reduction in ICU mortality ( 0.84, 95% 0.71-0.99) compared to VC-CMV, potentially through better pressure limitation that lowers VILI incidence, though overall data on direct VILI reduction remain inconclusive and call for further high-quality studies. Effective management demands frequent monitoring, including arterial blood gases for over- or under-ventilation, plateau measurements to stay below 30 cm H2O, and hemodynamic assessments to detect right-heart strain from elevated intrathoracic pressures. Adjustments should prioritize lung-protective strategies, such as tidal volumes of 4-8 mL/kg predicted body weight, with prompt shifts to partial support modes as patient condition improves to facilitate recovery and minimize prolonged ventilation risks. Patient-centered factors, such as underlying cardiovascular status, further influence outcomes, as CMV's positive pressure can worsen and right ventricular function in vulnerable individuals.

Historical Context

Development History

The development of continuous mandatory ventilation (CMV) traces its roots to the mid-20th century, amid widespread epidemics that necessitated mechanical support for . In the 1950s, negative-pressure devices such as the iron lungs invented by Philip Drinker in 1928 and refined by John H. Emerson provided by simulating natural breathing through external chest enclosure; these were extensively used during outbreaks, but their limitations in mobility and efficacy prompted a shift to positive-pressure ventilation. Pioneering work by Albert Bower and V. Ray Bennett in 1949 introduced intermittent positive-pressure breathing via masks, achieving survival rates up to 84% in patients, while Danish anesthesiologist Bjørn Ibsen advanced the technique in 1952 by employing manual bag ventilation through tracheostomies, drastically reducing mortality in Copenhagen's epidemic from over 80% to 20% and establishing tracheostomy as a standard for prolonged support. By the 1960s, CMV evolved into a formalized mode with the advent of volume-controlled ventilators, exemplified by the Puritan-Bennett MA-1 introduced in 1967, which delivered preset tidal volumes at constant flow rates in a controlled manner, independent of patient effort, marking the transition to electrically powered, ICU-suitable devices that replaced earlier manual and bulky systems. The 1970s brought critical recognition of barotrauma risks associated with high-pressure volume delivery, as evidenced by studies documenting and other in ventilated patients, spurring design modifications to incorporate pressure limits and alarms for safer operation. In the , microprocessor integration revolutionized CMV, with systems like the Puritan-Bennett 7200 (1983) enabling precise gas delivery, real-time monitoring, and the feasibility of pressure-controlled CMV (PC-CMV) variants that prioritized airway pressure limits over fixed volumes to mitigate . The 1990s saw further advancements in CMV through dual-control modes, such as ' introduction of pressure-regulated volume control (PRVC) around 1991, which dynamically adjusted pressure breath-by-breath to achieve target volumes while minimizing peak pressures, enhancing adaptability for varying . Influential clinical trials, including the ARDSNet published in 2000, underscored the importance of lung-protective strategies in CMV, demonstrating that lower tidal volumes (6 mL/kg predicted body weight) reduced mortality by 22% compared to traditional higher volumes (12 mL/kg), influencing global protocols to emphasize volutrauma prevention. Regulatory milestones, following the FDA's 1976 Medical Device Amendments classifying as Class II or III devices requiring premarket approval, ensured CMV modes in ICU met safety standards by the late , with widespread clearances for microprocessor-based models facilitating their adoption in critical care.

Terminological Evolution

In the early days of , the term "continuous mandatory ventilation" (CMV) referred to a strictly controlled mode where all breaths were delivered by the at a preset and or , without the capability for triggering, often necessitating deep sedation or paralysis to prevent asynchrony. This original usage emphasized full control, distinguishing it from modes allowing any -initiated activity. However, as technology advanced in the mid-20th century, the began to blur, with CMV increasingly used interchangeably to describe assist-control (A/C) , a variant that incorporated -triggered breaths while maintaining a minimum mandatory . This shift reflected the evolution from rigid time-cycled delivery to more -synchronized support, but it introduced confusion as the strict non-triggered definition of CMV was often overlooked in clinical practice. By the 1990s, efforts to standardize ventilator mode nomenclature gained momentum, driven by the proliferation of manufacturer-specific terms and the need to differentiate full-control modes from assisted ones. Robert Chatburn's 1992 classification system proposed a structured based on control variables, breath sequences, and targeting schemes, explicitly defining CMV as a sequence of mandatory breaths without allowance for spontaneous between them, while highlighting A/C as its patient-triggerable counterpart. This work, along with broader discussions in respiratory care literature, aimed to resolve ambiguities, such as the mislabeling of synchronized intermittent mandatory ventilation (SIMV) as a CMV variant despite SIMV permitting spontaneous breaths between synchronized mandatory deliveries. Although no singular American Thoracic Society (ATS) guideline from the era fully codified these distinctions, the decade's publications underscored the importance of precise to avoid equating CMV with modes like SIMV that incorporate partial effort. Contemporary standards, as outlined by the American Association for Respiratory Care (AARC) and the European Respiratory Society (ERS), refine CMV to denote controlled mechanical ventilation where every breath is mandatory—either time- or patient-triggered—but without provisions for unsupported spontaneous breaths, reserving the term A/C or ACV specifically for triggered implementations. Chatburn's updated taxonomy, adopted in ERS educational materials and AARC resources, classifies CMV within breath sequences like VC-CMV (volume-controlled) or PC-CMV (pressure-controlled), emphasizing its role in full support scenarios while distinguishing it from hybrid modes. These clarifications build on 1990s foundations to promote interoperability across devices. Persistent terminological misuse, such as conflating CMV with SIMV or overlooking triggering distinctions, can result in suboptimal mode selection, potentially exacerbating patient-ventilator asynchrony or delaying in clinical settings. Standardized nomenclature thus remains critical for and interdisciplinary communication in critical care.

References

  1. [1]
    Assist-Control Ventilation - StatPearls - NCBI Bookshelf - NIH
    Assist-control ventilation is the most common setting used in the ICU and is the only setting with proven mortality benefits in patients with ARDS.
  2. [2]
  3. [3]
    Pressure Controlled Ventilation - StatPearls - NCBI Bookshelf
    Jul 31, 2023 · It is typically available in both pressure control-continuous mandatory ventilation (PC-CMV) as well as pressure control-intermittent ...
  4. [4]
    Modes of Mechanical Ventilation - OpenAnesthesia
    Mar 3, 2013 · Volume Modes. Assist-Control Ventilation (ACV). Also known as continuous mandatory ventilation (CMV). Each breath is either an assist or ...
  5. [5]
    Continuous Mandatory Ventilation - an overview - ScienceDirect.com
    Continuous mandatory ventilation (CMV) is defined as a mode of neonatal ventilation where each breath is delivered at predetermined intervals, ...
  6. [6]
    [PDF] Mechanical Ventilation - Cleveland Clinic
    Continuous Mandatory Ventilation (CMV) ............................94. Volume ... False; the definition of noninvasive ventilation is that it does not ...
  7. [7]
    History of Mechanical Ventilation. From Vesalius to ... - ATS Journals
    Mar 2, 2015 · The origins of modern mechanical ventilation can be traced back about five centuries to the seminal work of Andreas Vesalius.
  8. [8]
    Anesthesia Gas Machine- Ventilators - College of Health Professions
    Volume control- Continuous mandatory ventilation: A set number of breaths/minute (f or RR) which are time cycled and machine triggered. Tidal volume (VT) is ...
  9. [9]
    Mechanical Ventilation (Chapter 17) - Core Topics in Cardiothoracic ...
    CMV (Continuous Mandatory Ventilation). This ventilation mode is ventilator triggered and time cycled. Volume (VC-CMV) or pressure control (PC-CMV) mode is ...<|control11|><|separator|>
  10. [10]
    Mechanical Ventilation - StatPearls - NCBI Bookshelf
    Mechanical ventilation is a critical intervention to sustain life in acute or emergent settings, particularly in patients with compromised airways.
  11. [11]
    Overview of Mechanical Ventilation - Critical Care Medicine
    Minute ventilation is set by. Tidal volume. Respiratory rate. When tidal volume is set too high, overinflation is a risk; when it is set too low, it allows for ...
  12. [12]
    Barotrauma and Mechanical Ventilation - Medscape Reference
    Feb 1, 2024 · Limiting plateau pressures to less than 30 cm H2O is an effective approach for all patients. With volume ventilators, this is best accomplished ...
  13. [13]
    Evaluation of a Ventilation Strategy to Prevent Barotrauma in ...
    Feb 5, 1998 · The experimental ventilation strategy limited peak inspiratory pressure to no more than 30 cm of water and tidal volume to no more than 8 ml per ...
  14. [14]
    Basic Modes of Mechanical Ventilation - xlung.net
    Controlled respiratory cycles: tidal volume (VT), inspiratory flow (A, square type; B, decelerating to 50%, C, decelerating to 0%). Inspiratory time increases ...
  15. [15]
    Pulmonary mechanics • LITFL • CCC Ventilation
    Jul 6, 2024 · normal dynamic compliance during mechanical ventilation – 50-100mL/cmH2O; when paralysed and mechanically ventilated, peak airway pressure ...
  16. [16]
  17. [17]
    Pressure-Regulated Volume Control and ... - PubMed Central
    Pressure-regulated volume-control ventilation (PRVC) mode is defined as a dual-mode that combines the variable flow pattern of the pressure-control (PC) mode ...
  18. [18]
    [PDF] MODES OF VENTILATION SERVO-i INVASIVE AND NON INVASIVE
    The ventilator adjusts the inspiratory Pressure Control level to the lowest possible level to guarantee the preset Tidal. Volume, in accordance with the ...
  19. [19]
    Pressure-regulated volume controlled ventilation in acute respiratory ...
    Pressure-regulated volume control (PRVC) mode is a kind of dual-control ventilation that uses tidal volume as a feed back control for continuously adjusting the ...Missing: mechanism equation
  20. [20]
    A long-term clinical evaluation of autoflow during assist-controlled ...
    "Dual-controlled" modes, such as AutoFlow™, are supposed to improve patient- ventilator interfacing and could lead to fewer alarms. We performed a long-term ...Missing: mechanism | Show results with:mechanism
  21. [21]
    Introduction to Ventilator Management - Medscape Reference
    Apr 7, 2020 · Continuous mandatory ventilation. Breaths are delivered at preset intervals, regardless of patient effort. This mode is used most often in ...
  22. [22]
    Ventilation with Lower Tidal Volumes as Compared with Traditional ...
    May 4, 2000 · Traditional approaches to mechanical ventilation use tidal volumes of 10 to 15 ml per kilogram of body weight and may cause stretch-induced ...
  23. [23]
    Ventilator Technology and Management | Critical Care
    The goals of mechanical ventilation are to provide safe gas exchange, decrease the work of breathing, improve patient–ventilator interactions, minimize ...
  24. [24]
    Pressure‐controlled versus volume‐controlled ventilation for acute ...
    Patient comfort during pressure support and volume controlled‐continuous mandatory ventilation. ... Etiology and outcomes of ARDS in a rural‐urban fringe ...
  25. [25]
    Complications and Pharmacologic Interventions of Invasive Positive ...
    Continuous mandatory ventilation (CMV): a mode that provides only mandatory breaths and does not allow for any spontaneous breaths. Intermittent mandatory ...Data Synthesis · Fundamentals Of Ippv · Mode Of Ventilation
  26. [26]
    Mechanical Ventilation in Adult Patients with Acute Respiratory ...
    However, mechanical ventilation itself can cause and potentiate lung injury and may contribute to nonpulmonary organ failure and mortality in patients with ARDS ...
  27. [27]
    A history of home mechanical ventilation: The past, present and future
    Mar 21, 2024 · This state-of-the-art review provides an overview of the history of home mechanical ventilation (HMV), including early descriptions of mechanical ventilation.
  28. [28]
    Early ICU Ventilators - Virtual Museum
    An overview of the early generations of ICU ventilators from the 1950s – 1979. 1950s. The early positive pressure mechanical ventilators are featured.
  29. [29]
    A Brief History of Control and Spontaneous Modes
    The first version of the modern ventilation modes was a mode known as Continuous Mandatory Ventilation (CMV). It was created to artificially breathe for the ...
  30. [30]
    Pressure regulated volume control versus synchronized intermittent ...
    Nov 28, 2016 · In pressure regulated volume control (PRVC) mode which was introduced by Siemens Company, a certain volume is given to the patient. Thus ...
  31. [31]
  32. [32]
    Ventilatory Modes. What's in a Name? | Respiratory Care
    Dec 1, 2012 · Originally, assist/control meant volume-controlled continuous mandatory ventilation (CMV). Now it can also refer to pressure control (especially ...<|control11|><|separator|>
  33. [33]
    Ventilator Modes - CHEST Journal
    SIMV/PS was less likely to be selected if patients were ventilated because of asthma or coma, or if they developed complications such as sepsis or ...
  34. [34]
    [PDF] Ventilator Mode Nomenclature - AARC
    VC-CMV. Mechanical ventilation with preset tidal volume and inspiratory flow. Every breath is mandatory (ie, inspiration is patient or machine trigged.Missing: ERS | Show results with:ERS
  35. [35]
    are we ventilating our patients in the best possible way?
    Construction of the ventilation mode taxonomy suggested by Chatburn. The name of a ventilation mode results from three elements. CMV: continuous mandatory ...Missing: nomenclature | Show results with:nomenclature
  36. [36]
    A Taxonomy for Mechanical Ventilation: 10 Fundamental Maxims
    CMV, commonly known as assist control, is a breath sequence for which spontaneous breaths are not possible between mandatory breaths because every patient- ...
  37. [37]
    [PDF] STANDARDIZED VOCABULARY FOR MECHANICAL ...
    If the exhaled minute ventilation falls below the operator set value, the ventilator will trigger mandatory breaths or increase the inspiratory pressure until.