Duplex scanning

Duplex scanning, also known as duplex , is a noninvasive imaging modality that combines B-mode ultrasonography for visualization of anatomical structures with to assess blood flow velocity, direction, and turbulence within blood vessels. This integrated approach utilizes high-frequency sound waves emitted from a handheld , which reflect off tissues and moving red blood cells, enabling evaluation of vascular morphology and without the need for or contrast agents. The technique operates on the principle of the , where the frequency shift of reflected waves from moving blood cells is measured to quantify flow characteristics, often displayed in color (e.g., red for flow toward the , blue for away) overlaid on the B-mode image. Clinically, duplex scanning is widely applied to diagnose and monitor vascular conditions, including thrombosis, peripheral arterial disease, , venous insufficiency, and renal artery abnormalities, providing critical data for treatment planning such as surgical interventions or anticoagulation therapy. Its procedure typically involves applying conductive gel to and scanning targeted areas like the legs, , , or for 30 to 60 minutes, performed by a trained in an outpatient setting. Among its notable advantages, duplex scanning is portable, cost-effective, and safe for repeated use, including in pregnant patients, with minimal risks such as temporary skin discomfort from the gel or probe pressure. However, its efficacy can be limited by operator expertise, patient body habitus (e.g., obesity attenuating waves), and suboptimal insonation angles greater than 60 degrees, which may necessitate complementary imaging like CT angiography in complex cases.

Overview

Definition

Duplex scanning is a non-invasive technique that combines real-time B-mode grayscale ultrasonography, which provides detailed anatomical visualization of vascular structures, with to evaluate blood flow velocity and direction. This hybrid approach utilizes high-frequency sound waves reflected from tissues and blood cells to generate both structural images and flow data without requiring incisions or . The primary purpose of duplex scanning is to simultaneously display vascular anatomy alongside hemodynamic information, facilitating the of various vascular pathologies in a safe, patient-friendly manner that avoids contrast agents and their associated risks. By integrating these modalities, it offers a comprehensive of patency, wall integrity, and flow dynamics, making it a cornerstone for non-invasive vascular evaluation. In distinction from related techniques, standalone B-mode ultrasonography focuses solely on anatomical imaging without flow assessment, while continuous-wave or pulsed-wave Doppler alone measures blood flow but lacks structural context; duplex scanning uniquely merges both capabilities into a single scan for enhanced diagnostic accuracy. The term "duplex" specifically denotes this dual-mode operation, encompassing grayscale imaging and spectral Doppler waveform display to provide a unified view of vessel form and function.

History

The origins of duplex scanning trace back to the development of Doppler in the mid-20th century for detecting blood flow. In 1959, Japanese physicist Shigeo Satomura published the first measurements of the Doppler shift in ultrasonic signals reflected from a beating heart, laying the groundwork for non-invasive blood flow detection using . This innovation built on earlier industrial applications of and technology that Satomura explored starting in 1955. Concurrently, in the 1960s, surgeon D. Eugene Strandness Jr. at the pioneered the clinical application of Doppler to assess , establishing it as a tool for diagnosing arterial occlusive conditions through transcutaneous measurements. Strandness's work emphasized quantitative velocity assessments, transforming Doppler from a research curiosity into a practical diagnostic method in . A key advancement came in 1969 with the development of pulsed Doppler systems by bioengineer Donald W. Baker, also at the , which allowed range-gating to localize blood flow signals at specific depths without interference from surrounding tissues. Baker's landmark publications, including "Pulsed Ultrasonic Blood Flow Sensing" in 1969, introduced phase-coherent pulsed Doppler technology that improved accuracy for cardiovascular measurements. This pulsed approach addressed limitations of continuous-wave Doppler, enabling precise velocity profiling essential for later vascular imaging. The pivotal milestone occurred in 1974 when engineer Frank E. Barber, collaborating with Baker and Strandness, invented the duplex echo-Doppler scanner, which integrated real-time two-dimensional B-mode imaging with pulsed Doppler to provide anatomical visualization alongside flow data for vascular mapping. This combination allowed operators to position the Doppler sample volume on visualized vessels, revolutionizing non-invasive arteriography. During the late 1970s, , founded in 1969 by a team including , commercialized the first , with the Mark I model released in 1975 and the Mark V in 1978, making the technology accessible beyond research settings. Early adoption focused on duplex scanning in vascular laboratories for evaluating extracranial carotid arteries, with validation studies demonstrating its efficacy by 1980. The 1980s saw further evolution with the integration of color Doppler, which overlaid flow direction and velocity on B-mode images to create color duplex systems, though duplex remained the foundational technique for detailed . By the , duplex scanning had become a routine, non-invasive standard for vascular assessment in clinical practice, supplanting more invasive methods like in many applications.

Principles of Operation

B-mode Imaging

B-mode, or brightness mode, is a ultrasound imaging technique that utilizes reflected echoes from high-frequency sound waves, typically in the range of 2 to 18 MHz, to generate two-dimensional cross-sectional images of tissues and vessels. This mode forms the anatomical foundation of duplex scanning by displaying structures based on the intensity of returning echoes, where brighter shades of gray represent stronger reflections from tissue interfaces. The physics of B-mode imaging relies on piezoelectric transducers, which generate short pulses of waves through the rapid deformation of piezoelectric crystals that convert into mechanical vibrations. These pulses travel through tissue, and upon encountering boundaries between media of differing , a portion of the wave reflects back as an echo; the time-of-flight of the echo determines the depth of the reflecting structure, while the amplitude of the echo dictates the brightness level in the resulting image. in B-mode is influenced by several factors, including axial resolution, which measures the ability to distinguish objects along the beam path and is given by the formula: \text{Axial resolution} = \frac{\text{spatial pulse length}}{2} = \frac{\text{number of cycles} \times \text{wavelength}}{2}, where wavelength = speed of sound / frequency, and the speed of sound in soft tissue is approximately 1540 m/s. Lateral resolution, perpendicular to the beam, depends on the ultrasound beam width, which is narrower at the focal zone to enhance detail. In duplex scanning, B-mode provides essential static or real-time anatomical context by visualizing vessel walls, lumen dimensions, atherosclerotic plaques, and adjacent tissues, allowing clinicians to identify structural abnormalities such as or aneurysms prior to Doppler-based flow evaluation. Common artifacts in B-mode imaging include acoustic shadowing, where highly attenuating structures like calcified plaques produce a dark zone distal to the reflector due to reduced echo ; reverberation, caused by repeated reflections between parallel highly reflective surfaces, resulting in multiple equidistant bright lines; and posterior acoustic enhancement, appearing as brighter echoes behind low-attenuation areas like fluid-filled lumens, indicating uniform wave propagation through less scattering media. These artifacts can affect interpretation but also aid in identifying underlying when recognized.

Doppler Ultrasonography

In duplex scanning, the component employs either pulsed-wave or continuous-wave spectral Doppler to assess blood flow by detecting frequency shifts in the waves reflected from moving red blood cells. Pulsed-wave Doppler allows for range resolution by sampling echoes from a specific depth, making it suitable for localized flow measurements, whereas continuous-wave Doppler provides continuous sampling without depth selectivity, ideal for high-velocity flows. This modality integrates with B-mode imaging to correlate flow data with vascular anatomy. Color flow imaging, a form of pulsed Doppler, provides a real-time visual map of blood flow overlaid on the B-mode image, encoding mean and with colors—typically red for flow toward the and blue for flow away—based on the same Doppler shift principle. It uses multiple sample volumes across the beam to estimate average flow, aiding in quick identification of flow presence, , or absence, though it lacks the precise quantification of spectral Doppler. The fundamental principle relies on the , quantified by the equation: \Delta f = \frac{2 v f_0 \cos \theta}{c} where \Delta f is the frequency shift, v is the blood flow velocity, f_0 is the transmitted frequency, \theta is the angle between the beam and the direction of blood flow (angle of insonation), and c is the in (approximately 1540 m/s). This shift arises because red blood cells moving toward the reflect higher-frequency echoes, while those moving away reflect lower frequencies, enabling the calculation of velocity and direction. Spectral waveform analysis displays blood velocity against time, producing a graphical representation of flow dynamics. In laminar flow, the waveform exhibits a narrow spectral band with clear boundaries, reflecting uniform velocity profiles, whereas turbulent flow results in spectral broadening due to varying velocities across the vessel lumen. Key parameters include peak systolic velocity (PSV), the maximum velocity during systole, and end-diastolic velocity (EDV), the minimum velocity at end-diastole, which are used to grade the severity of vascular stenoses by comparing ratios such as PSV elevation proximal to a narrowing. Accurate measurement requires correction, as the \cos \theta term in the Doppler diminishes signal intensity at larger angles; an optimal beam-to-flow of 45° to 60° balances and , with angles approaching 90° yielding near-zero shifts and unreliable data. Operators adjust the cursor to align the parallel to within this range to minimize errors. Common artifacts include , where high velocities exceed the Nyquist limit (half the , PRF), causing wrap-around of the spectrum or color reversal in color ; this is mitigated by increasing PRF, shifting the , or switching to continuous-wave . artifacts occur when strong reflectors, such as vessel walls or gas interfaces, duplicate the spectral signal on the opposite side of the , potentially mimicking bidirectional flow.

Equipment and Procedure

Duplex Scanners

Duplex scanners are specialized systems that combine B-mode imaging for anatomical visualization with for blood assessment, enabling simultaneous display of grayscale images and data overlays. The core hardware includes an console housing the signal processor, which fuses B-mode and Doppler signals through digital beamforming and processing to generate integrated images. A probe, typically equipped with piezoelectric crystals, transmits and receives waves, while the display unit provides real-time visualization of grayscale alongside spectral Doppler waveforms or color maps. Probe specifications vary by application, with linear array transducers favored for superficial vessels like those in the carotid or peripheral arteries due to their rectangular scan format and high near-field resolution. These probes often operate at frequencies of 5-10 MHz to balance resolution and penetration for such structures, whereas lower frequencies (around 3-5 MHz) are used in curved linear or phased array probes for deeper abdominal vessels, employing multi-element arrays (typically 128-256 elements) for electronic beam steering in sector scan formats. Modern duplex scanners come in portable handheld or cart-based configurations, with portables offering enhanced mobility for point-of-care use while cart-based systems provide superior processing power for complex analyses. As of 2025, advancements include AI-powered diagnostics for automated detection and quantitative measurements, as well as miniaturized handheld devices improving in remote settings. Advanced integrations include power Doppler for sensitive low- detection without and contrast enhancement agents to improve visualization in challenging cases. Software features, such as automated angle correction algorithms, streamline velocity measurements by computationally estimating the Doppler beam-to- angle, reducing operator variability. The evolution to color duplex scanning introduced color flow mapping as an enhancement to traditional spectral duplex, overlaying qualitative and data in color on B-mode images to aid in rapid flow assessment without replacing core . in duplex scanning adheres to the ALARA (as low as reasonably achievable) principle, with output limits regulated to minimize and mechanical bioeffects, such as tissue heating below 1°C and thresholds, through adjustable acoustic power settings.

Scanning Technique

Preparation for a duplex scan begins with appropriate positioning, which varies by anatomical site to optimize and . For carotid and abdominal vascular assessments, the is typically positioned , while prone positioning is used for evaluating the , particularly the popliteal and tibial vessels. A water-soluble is applied to the skin overlying the area of interest to facilitate acoustic coupling between the and the body surface, minimizing air interference that could degrade image quality. Site-specific preparations may include for 4-6 hours prior to abdominal scans to reduce bowel gas and improve acoustic windows for visualizing deep vessels like the and iliac arteries. The step-by-step procedure starts with an initial B-mode ultrasound survey to map the vascular anatomy, identifying locations, diameters, and any obvious structural abnormalities such as plaques or aneurysms. Next, the sample volume of the pulsed-wave Doppler is placed within the lumen at the site of interest, ensuring it encompasses the full cross-section of flow. The angle is then adjusted to between 45° and 60° relative to the direction, as this range provides optimal alignment for accurate velocity measurements while avoiding the cosine error that occurs at angles greater than 60°. waveforms are subsequently acquired at multiple sites along the , including proximal, mid, and distal segments, as well as areas of suspected , to characterize flow patterns and velocities. Protocol variations incorporate both longitudinal and transverse views to fully assess patency and dynamics; for instance, longitudinal views are preferred for Doppler , while transverse views aid in measuring diameters and detecting eccentric plaques. In venous studies, maneuvers are applied proximally and distally to evaluate and augmentation, with additional techniques like Valsalva or distal used to provoke . Continuous monitoring allows for dynamic adjustments, such as repositioning the probe or altering gain settings, to capture physiological changes during the examination. Operator considerations emphasize to prevent fatigue during the procedure, including adjustable table heights and proper transducer grip to maintain consistent pressure without causing patient discomfort. Patient comfort is prioritized by explaining each step, using cushions for support, and minimizing time where possible; documentation involves capturing key images, waveforms, and measurements in a standardized report for clinical correlation. The typical duration of a duplex ranges from 30 to 60 minutes, depending on the vascular territory and complexity. Quality control is integral, focusing on achieving an adequate through optimized gain and frequency settings to ensure clear spectral displays without or clutter. Operators must recognize and address poor acoustic windows, such as those caused by , , or overlying gas, which can limit penetration; in such cases, lower-frequency transducers or alternative patient positions may be employed to improve results.

Clinical Applications

Arterial Imaging

Duplex scanning plays a crucial role in evaluating arterial structures and blood flow, particularly for detecting and assessing hemodynamic changes in various vascular beds. In arterial imaging, the technique combines B-mode for anatomical visualization with Doppler to measure blood velocity and waveform patterns, enabling non-invasive of conditions like . This approach is widely used due to its high for identifying hemodynamically significant lesions, such as those exceeding 50% reduction. Primary applications of duplex scanning in arterial imaging include the assessment of , peripheral arterial disease (PAD), and for . For evaluation, peak systolic (PSV) thresholds are employed to grade severity according to North American Symptomatic Carotid Endarterectomy Trial (NASCET) criteria, where a PSV greater than 125 cm/s indicates more than 50% narrowing at the , a common site for atherosclerotic plaque accumulation. In PAD, duplex scanning correlates with ankle-brachial index (ABI) measurements, revealing waveform progression from triphasic (normal, with forward systolic flow, brief reverse, and diastolic forward flow) in unobstructed arteries to monophasic (dampened, forward-only flow) in stenotic femoral or popliteal arteries, signaling disease severity. For renal arteries, duplex scanning targets ostial and proximal segments to detect contributing to , using elevated PSV or renal-aortic ratios to identify lesions greater than 60% diameter reduction. Diagnostic metrics in arterial duplex imaging emphasize velocity-based parameters and plaque features for precise interpretation. Velocity ratios, such as the internal carotid artery to common carotid artery (ICA/CCA) ratio exceeding 2 for 50-69% stenosis or greater than 4 for 70% or more, provide a reliable indicator of significant hemodynamic impact beyond absolute PSV values, which can vary by laboratory standards. Plaque characterization via B-mode assesses echolucency, where hypoechoic or echolucent plaques suggest instability and higher rupture risk due to lipid-rich necrotic cores and thin fibrous caps, contrasting with stable echogenic plaques. The resistive index (RI), calculated as RI = (PSV - EDV) / PSV where PSV is peak systolic velocity and EDV is end-diastolic velocity, evaluates distal vascular bed resistance; elevated RI values above 0.7 in renal or peripheral arteries may indicate parenchymal disease or microvascular impairment downstream of the imaged segment.30005-2/fulltext) Clinical scenarios for arterial duplex scanning encompass screening, surveillance, and real-time monitoring. It serves as a first-line tool for risk screening in asymptomatic patients with carotid bruits or risk factors, detecting clinically relevant with reasonable to guide interventions like . Post-intervention , such as after carotid or peripheral artery stenting, monitors for restenosis by tracking velocity changes and normalization over time. Intraoperative duplex scanning provides immediate feedback during vascular procedures, assessing restoration and graft patency in arteries like the femoral or renal to optimize outcomes. These applications highlight duplex scanning's utility in high-resistance arterial patterns, where pulsatile waveforms and velocity elevations are key to distinguishing pathology.

Venous Imaging

Duplex scanning plays a crucial role in evaluating the venous system, particularly for detecting deep vein thrombosis (DVT) and assessing (CVI). In DVT diagnosis, it identifies non-compressible veins and absence of flow augmentation, while for CVI, it measures valve reflux exceeding 0.5 seconds during maneuvers like Valsalva or distal augmentation. These applications leverage the non-invasive nature of to provide real-time imaging of venous patency and hemodynamics. The primary techniques in venous duplex scanning include grayscale imaging to evaluate vein compressibility, where normal veins exhibit 100% compression under probe pressure, indicating no intraluminal obstruction. Color Doppler assesses direction and detects abnormalities such as reversed or turbulent , while spectral Doppler evaluates phasic respiratory variation, with normal veins showing spontaneous augmentation and respiratory modulation. For reflux testing in CVI, provocative maneuvers provoke retrograde , quantified via . Common sites for venous duplex scanning encompass the lower extremity veins, including the femoral, popliteal, and calf veins (such as posterior tibial and peroneal), scanned from the to the ankle. In the upper extremity, it targets central veins like the subclavian, axillary, and brachial for catheter-related , using similar compression and augmentation techniques.30108-3/fulltext) veins, such as the and splenic, are also assessed for , though acoustic windows may limit visualization. Diagnostic criteria for DVT rely on direct visualization of —acute thrombi appear hypoechoic and expandable, while chronic ones are echogenic and rigid—combined with non-compressibility. Indirect signs include flow gaps on color Doppler, indicating obstruction, and lack of response to augmentation or respiratory variation on Doppler. In CVI, reflux is defined as retrograde flow greater than 0.5 seconds in or superficial veins. Special protocols enhance diagnostic accuracy; for reflux testing, patients are placed in the to distend veins, followed by for proximal veins or manual distal compression for lower segments. Evaluation of involves assessing residual , valvular incompetence, and vein wall thickening to guide management of chronic changes. Complete duplex protocols recommend serial imaging for untreated calf DVT to monitor propagation.

Advantages and Limitations

Benefits

Duplex scanning is a non-invasive diagnostic that avoids exposure to , nephrotoxic contrast agents, or surgical incisions, thereby reducing patient risks such as allergic reactions, injury, and procedural complications associated with alternatives like or computed tomography (). This approach enables safe evaluation of vascular structures in diverse patient populations, including those with renal impairment or implants. The modality provides real-time imaging and hemodynamic assessment, allowing clinicians to observe blood flow dynamics immediately during provocative maneuvers such as for venous evaluation or positional changes to assess arterial patency. Its portability facilitates bedside use in intensive care units or outpatient settings, supporting prompt diagnosis without the need for patient transport or hospitalization. Duplex scanning demonstrates high diagnostic accuracy, with reported sensitivity often exceeding 90% and specificity typically 80–95% for detecting compared to arteriography (based on older studies; recent data as of 2025 show variability, e.g., sensitivity 77–89%, specificity 63–92%), and similarly high values exceeding 90% for identifying (DVT) relative to . This reliability supports its role in initial screening and follow-up monitoring, such as evaluating patency or progression of . From a safety perspective, duplex scanning involves low bioeffects, with typical indices below 1.0 and mechanical indices under 1.9, minimizing risks of tissue heating or as per established guidelines. These parameters ensure the procedure is well-tolerated across serial applications without cumulative harm. Economically, duplex scanning is cost-effective, often costing less than angiography or while offering effective diagnostic yield, with studies showing substantial savings compared to these alternatives. Operator-dependent training enables efficient use by non-specialists, further avoiding complications like bleeding from invasive alternatives.

Challenges

Duplex scanning is highly operator-dependent, with results varying based on the technologist's skill level and experience, which can lead to misinterpretation or delayed . such as the Registered Vascular Technologist (RVT) credential is recommended to standardize performance and reduce errors in vascular procedures. Inter-observer variability is notable, particularly in waveform interpretation and peak systolic velocity measurements, where agreement can be as low as substantial but not perfect across independent observers. Technical limitations include poor acoustic windows in patients with or calcified vessels, which obscure vessel visualization and hinder accurate assessment. occurs in high-velocity flows when the Doppler shift exceeds the Nyquist limit, defined as half the (PRF/2), causing velocity wrap-around and inaccurate spectral displays. Common artifacts and errors arise from angle misalignment, where insonation angles greater than 70 degrees underestimate velocities due to the cosine factor approaching zero at 90 degrees, particularly in tortuous vessels. Bowel gas interference commonly affects abdominal scans, leading to incomplete evaluations, while limited depth penetration, typically less than 20 cm, restricts imaging of deeper structures. Diagnostic constraints involve poor visualization of intracranial vessels due to skull attenuation and lower resolution for small vessels under 2 mm, which reduces sensitivity for detecting subtle lesions like small ulcers. False positives and negatives are more frequent in acute settings, with error rates up to 13-21% attributed to patient factors, compensatory flow, or incomplete compression, potentially missing proximal or subclavian thrombi. Recent advances as of 2025, such as (CEUS) for better perfusion assessment and AI-assisted tools for reducing operator variability, are helping to mitigate some of these challenges. These challenges can be mitigated through rigorous training and certification programs to enhance operator proficiency, use of higher-frequency probes for superficial vessels to improve resolution, and adjunct imaging such as (CTA) for confirmation in equivocal cases.

References

  1. [1]
    Duplex Ultrasound - StatPearls - NCBI Bookshelf
    Mar 27, 2025 · Duplex ultrasound is a specialized imaging technique that integrates anatomic and flow ultrasonography to provide diagnostic information. This ...
  2. [2]
    Doppler Ultrasound: What Is It, Purpose and Procedure Details
    Duplex ultrasound: Combines traditional ultrasound pictures with Doppler ultrasound. It can check the width of blood vessels and can help show blockages ...Overview · What Is A Doppler Ultrasound... · Test Details
  3. [3]
    Duplex Ultrasound | Stanford Health Care
    Duplex ultrasound involves using high frequency sound waves to look at the speed of blood flow, and structure of the leg veins.Missing: principles applications
  4. [4]
    Duplex ultrasound : MedlinePlus Medical Encyclopedia
    May 10, 2024 · A duplex ultrasound is a test to see how blood moves through your arteries and veins. How the Test is Performed
  5. [5]
    Duplex Ultrasound | Society for Vascular Surgery
    Duplex ultrasound is a non-invasive evaluation of blood flow through your arteries and veins. This test provides information to help your vascular surgeon.
  6. [6]
    Shigeo Satomura: 60 years of Doppler ultrasound in medicine - PMC
    Dec 23, 2015 · We celebrate 60 years since Shigeo Satomura published the first measurements of the Doppler shift of ultrasonic signals from a beating heart.
  7. [7]
    Shigeo Satomura, history of doppler ultrasound
    In 1955, Satomura began using microwave and ultrasound in industrial research, shortly following his initial efforts in measuring vibrations on wooden boards.
  8. [8]
    Strandness found new way to diagnose vascular disease | UW News
    Apr 19, 2007 · In the 1960s, Dr. Eugene Strandness examined a patient with diabetes who was facing amputation of his leg because of his disease.
  9. [9]
    Donald Baker – UW Bioengineering pioneer of medical ultrasound
    Oct 26, 2020 · His idea was to use pulsed Doppler technology, rather than continuous sound waves. Not only did the technique provide clear images, the method ...
  10. [10]
    Alumnus Donald W. Baker - UWEE
    Baker published the landmark articles: "A phase coherent pulse Doppler ... Pulsed Ultrasonic Blood Flow Sensing" in 1969. Between 1973 and 1974 Baker ...
  11. [11]
    Ultrasonic duplex echo-Doppler scanner - PubMed
    Ultrasonic duplex echo-Doppler scanner. IEEE Trans Biomed Eng. 1974 Mar;21(2):109-13. doi: 10.1109/TBME.1974.324295. Authors. F E Barber, D W Baker, A W ...Missing: Frank | Show results with:Frank
  12. [12]
    From Doppler to duplex: A personal early history of the vascular ...
    Apr 8, 2021 · In the mid-1970s, a prototype B-mode scanning system was built at the University of Washington that contained three 5 MHz fixed-focus ultrasound ...
  13. [13]
    Duplex ultrasound imaging: vascular applications. | Radiology
    Feb 1, 1984 · This system was used in the evaluation of the extracranial carotid arteries, iliofemoral arteries, and hemodialysis shunts. The system obtains ...
  14. [14]
    From Doppler to duplex: A personal early history of the vascular ...
    Apr 8, 2021 · In the mid-1970s, a group of clinicians and bioengineers at the University of Washington, under the direction of Dr D Eugene Strandness, Jr, ...
  15. [15]
    Ultrasound principles and instrumentation - ScienceDirect.com
    In the B mode (or brightness modulation), the processor transforms the soundwaves into grayscale according to each echo's amplitude (A Mode). This allows for ...
  16. [16]
    Ultrasound Physics and Technical Facts for the Beginner - ACEP
    Jul 16, 2020 · B-mode or 'brightness mode' provides structural information utilizing different shades of gray (or different 'brightness') in a two-dimensional ...
  17. [17]
    Ultrasound Image Quality - Radiologica
    Mar 29, 2021 · Axial resolution = Spatial pulse length / 2. Where SPL = number of ... Hence any measure that shortens the length of the ultrasound pulse will ...
  18. [18]
    Propagation speed | Radiology Reference Article - Radiopaedia.org
    Aug 2, 2021 · Ultrasound machines assume sound waves travel at a speed of 1540 m/sec through tissue 1. In reality, the speed of sound is affected by the ...
  19. [19]
    Basic Principles of Ultrasound Physics and Artifacts Made Easy
    Acoustic shadowing occurs when ultrasound waves encounter a structure that has a high attenuation coefficient. You will most commonly encounter the acoustic ...
  20. [20]
    Acoustic shadowing | Radiology Reference Article - Radiopaedia.org
    Aug 11, 2025 · Acoustic shadowing (sometimes referred to as posterior acoustic shadowing) is a form of ultrasound artifact. It is characterized by the apparent lack of signal ...Question 2441 · Question 2835 · Question 2402
  21. [21]
    Sonographic artifacts and their origins. | AJR
    Artifacts are encountered daily in clinical sonography. They may be observed in B-mode gray-scale imaging, spectral pulsed Doppler imaging, and color Doppler ...
  22. [22]
    DOPPLER ULTRASOUND - PMC - NIH
    Spectral Doppler : This type of display is typically used in vascular flow imaging by Duplex scanning (B mode + Pulsed wave Doppler) where the Doppler signal is ...
  23. [23]
    Pulsed Wave Doppler – The Cardiovascular - ECGWaves
    PW Doppler sends short pulses of ultrasound and analyzes reflected sound waves between the pulses. CW Doppler sends and analyses ultrasound continuously.
  24. [24]
    Spectral Doppler (ultrasound) | Radiology Reference Article
    Mar 23, 2019 · Spectral Doppler refers to ultrasound modalities which yield graphical representations of flow velocity over time.
  25. [25]
    Doppler shift | Radiology Reference Article - Radiopaedia.org
    Oct 16, 2025 · Doppler equation. F = 2fo(v/c)cos(Q). where: F is Doppler frequency shift. fo is transmitted frequency from ultrasound probe. v is the velocity ...
  26. [26]
    Doppler Shift - The Angle Matters! - RK.MD
    Aug 14, 2023 · Doppler shift = [2*f*v*cos(θ)] / c where 'f' is the operating frequency, 'v' the velocity of blood, 'θ' the angle between the ultrasound beam ...<|separator|>
  27. [27]
    [PDF] Interpretation of peripheral arterial and venous Doppler waveforms
    Spectral Doppler waveforms demonstrate laminar. (A), disturbed (B), and turbulent (C) flow patterns. Figure 3. Waveform illustrations from authoritative ...
  28. [28]
    Normal Doppler Spectral Waveforms of Major Pediatric Vessels
    May 1, 2008 · Spectral waveforms reflect the physiologic status of the organ supplied by the vessel, as well as the anatomic location of the vessel in ...
  29. [29]
    The 60° Doppler Angle Correction Paradigm - Wiley Online Library
    Dec 23, 2020 · Based on the authors' data, measurements made at 60° are more variable than those made at lesser degrees of Doppler angle correction.
  30. [30]
    6. Spectral Doppler | 123sonography
    Dec 21, 2020 · Position your sample volume in the middle of the artery; The sample volume should be between 15 and 30mm; Optimize the angle correction (45-60°) ...
  31. [31]
    Aliasing phenomenon (ultrasound) | Radiology Reference Article
    Jan 31, 2025 · The artifact immediately disappears if the upper margin of the velocity scale is increased above the peak flow velocity. Color aliasing is ...
  32. [32]
    Sonography Doppler Flow Imaging Instrumentation - StatPearls - NCBI
    May 1, 2023 · Mirror image artifacts in ultrasonography can be caused by large reflectors and large discrepancies in acoustic impedance at soft tissue ...
  33. [33]
    Overview of Ultrasound Imaging Systems and the Electrical ...
    May 10, 2010 · A critical component of this system is the ultrasound transducer. A typical ultrasound imaging system uses a wide variety of transducers ...Missing: duplex core
  34. [34]
    3. Instrumentation and physical principles of carotid (Duplex ...
    Nov 19, 2020 · Aside from a good (high resolution B image) you will need color Doppler functionality and pulsed wave (spectral Doppler) to assess flow and to ...
  35. [35]
  36. [36]
    Peripheral Arterial Duplex Assessment, Protocols, and Interpretation
    Jul 20, 2023 · Higher frequency probes of 7.5 to 10 MHz are used for thin or small individuals when identifying superficial vessels. In contrast, lower ...
  37. [37]
  38. [38]
    Power Doppler US: a potentially useful alternative to mean ...
    Mar 1, 1994 · A new color Doppler (CD) ultrasonography (US) technique called power Doppler (PD), which displays the total integrated Doppler power in color.
  39. [39]
  40. [40]
    Computer Vision Approach for Ultrasound Doppler Angle Estimation
    In this paper, we will describe a computer vision approach to automate the Doppler angle estimation.Missing: software | Show results with:software
  41. [41]
    As Low As Reasonably Achievable (ALARA) Principle - AIUM
    May 19, 2020 · The as low as reasonably achievable (ALARA) principle should be observed when adjusting controls that affect the acoustic output and by considering both the ...
  42. [42]
    [PDF] EFSUMB Best Practice Guidance for the Safe use of Doppler ...
    Start with low indices and adjust the settings as needed using the ALARA-principle (As Low As Reasonable Achievable). - Keep Doppler exposure times and output ...
  43. [43]
    Sonography Vascular Peripheral Arterial Assessment, Protocols ...
    Mar 27, 2023 · Duplex ultrasound is used to image the direction and characteristics of blood vessels. Central, continuous blood flow is commonly observed in ...Continuing Education Activity · Introduction · Technique or Treatment
  44. [44]
    Duplex ultrasonography protocol for investigation of patients ...
    It consists of three methods of investigation: B-mode ultrasound, pulsed duplex ultrasound, and color duplex scanning.
  45. [45]
    Our Protocol for Transabdominal Pelvic Vein Duplex Ultrasound
    A summary of South Coast Vascular Laboratory's protocol for pelvic vein duplex ultrasonography, including equipment, patient positioning, ultrasound settings, ...
  46. [46]
    Duplex ultrasound for diagnosing symptomatic carotid stenosis in ...
    DUS is used in clinical practice as the first test to detect carotid artery stenosis, usually with the result confirmed by other more expensive and invasive ...
  47. [47]
  48. [48]
    Approach to the Patient With Peripheral Arterial Disease | Circulation
    Nov 12, 2013 · Duplex ultrasound incorporates 2-dimensional gray-scale ultrasound to provide an image of arterial structures and pulse-wave Doppler to sample ...
  49. [49]
    Critical analysis of renal duplex ultrasound parameters in detecting ...
    A PSV of 285 cm/s or an RAR of 3.7 alone can be used in detecting ≥60% RAS. Previously published data must be validated in individual vascular laboratories.
  50. [50]
    Ultrasound Doppler renal resistive index: a useful tool for the ... - PMC
    An RRI is calculated with the following formula: (peak systolic velocity – end diastolic velocity)/peak systolic velocity, and the mean value of three ...
  51. [51]
    Screening for Asymptomatic Carotid Artery Stenosis: USPSTF ...
    Feb 2, 2021 · Adequate evidence exists that duplex ultrasonography has a reasonable sensitivity and specificity profile for detecting clinically relevant ...
  52. [52]
    Arterial Doppler Imaging, ABI, and Plethysmography - NCBI - NIH
    Mar 6, 2023 · A typical peripheral artery at rest will have a triphasic or biphasic quality of sound, which corresponds to forward flow in systole, a brief ...
  53. [53]
    Ultrasound for Lower Extremity Deep Venous Thrombosis | Circulation
    Apr 3, 2018 · Complete duplex ultrasound (CDUS) is the preferred venous ultrasound test for the diagnosis of acute DVT. CDUS is compression of the deep veins ...
  54. [54]
    Duplex ultrasound in the assessment of lower extremity venous ...
    The purpose of this paper is to provide a focused overview of the key concepts in the assessment of lower limb chronic venous insufficiency (CVI)
  55. [55]
    The clinical application of venous ultrasound in diagnosis and follow ...
    Oct 26, 2023 · Complete duplex ultrasound (CDUS) is the preferred venous ultrasound protocol for diagnosis of acute DVT, with pooled sensitivity and ...
  56. [56]
    Splanchnic Venous Thrombosis - StatPearls - NCBI Bookshelf - NIH
    Jul 4, 2022 · Doppler ultrasound is the best initial diagnostic test in cases with a suspicion of splenic vein thrombosis when considering all these facts. A ...Continuing Education Activity · Epidemiology · Evaluation · Treatment / ManagementMissing: duplex | Show results with:duplex
  57. [57]
  58. [58]
    Cardiovascular Ultrasound: Expanding Frontiers in Vascular Disease
    The sensitivity of duplex ultrasound to detect carotid disease varies from 87% to 94% with a specificity of 88% to 93%. The accuracy of duplex examination ...
  59. [59]
    Statement on Biological Effects of Ultrasound in Vivo - AIUM
    Aug 19, 2021 · The mechanical index (MI) has been formulated to assist users in evaluating the likelihood of adverse nonthermal biological effects for ...Missing: duplex | Show results with:duplex
  60. [60]
    [PDF] Doppler ultrasound devices – Safety Aspects (update 2011) | EFSUMB
    Thermal TI and Mechanical MI Index. The on-screen display of two indices is mandatory for modern systems: these are the thermal TI and mechanical MI index.
  61. [61]
    A Review on Biological Effects of Ultrasounds: Key Messages for ...
    Feb 23, 2023 · This review work has made it possible to highlight the limits associated with the use of the estimated safety values of thermal and mechanical ...
  62. [62]
    Cost-effectiveness of diagnostic imaging modalities in symptomatic ...
    Sep 3, 2024 · In the basic case, the DUS diagnostic was the least expensive modality, at a cost of EUR 10,778, compared with EUR 10,804 for CTA, EUR 11,184 ...
  63. [63]
    Arterial Duplex Ultrasound is the Most Cost-Effective, Noninvasive ...
    Compared with other imaging modalities, duplex ultrasound is clearly the most cost efficient. The results of this study suggest that non-invasive, pre ...Missing: vascular | Show results with:vascular
  64. [64]
    Imaging of Carotid Arteries in Symptomatic Patients: Cost ...
    Duplex US performed without additional imaging is cost-effective in the selection of symptomatic patients suitable for endarterectomy.
  65. [65]
    [PDF] IAC Standards and Guidelines for Vein Center Accreditation
    Aug 15, 2025 · 1.1A. Must have an appropriate level of training and experience and must have a valid appropriate credential in vascular testing: i. Registered ...
  66. [66]
    Interobserver variation in duplex sonographic scanning in ... - PubMed
    In a prospective study we quantified the interobserver variation in duplex scanning for the diagnosis of peripheral vascular disease. Forty-one patients with ...Missing: variability waveform
  67. [67]
    Interobserver variability of carotid Doppler peak velocity ...
    This study was designed to investigate interobserver variability in the measurement of internal carotid artery (ICA) peak systolic velocity (PSV).
  68. [68]
  69. [69]
    Nyquist limit | Radiology Reference Article - Radiopaedia.org
    Aug 21, 2020 · The Nyquist limit represents the maximum Doppler shift frequency that can be correctly measured without resulting in aliasing in color or pulsed wave ...
  70. [70]
    The Limitations of Carotid Sonography Interpretive and Technology ...
    Tortuous vessels are known to cause difficulties with accurate duplex sonography [7]. It may be impossible to obtain an accurate angle of insonation and, ...Missing: challenges | Show results with:challenges
  71. [71]
    False-negative upper extremity ultrasound in the initial evaluation of ...
    Duplex ultrasound is limited in the initial evaluation of patients with suspected SCV thrombosis, with false-negative results in 21% of patients with proven ...Missing: negatives | Show results with:negatives
  72. [72]
    Operative Procedures for Ultrasound Assessment of Extracranial ...
    Results: This document outlines a comprehensive approach to the ultrasound evaluation of extracranial arteries, detailing techniques such as Brightness Mode (B- ...