Fact-checked by Grok 2 weeks ago

Defibrillation

Defibrillation is a that delivers a controlled electric to the heart through the chest wall to terminate life-threatening cardiac arrhythmias, such as (VF) or pulseless (VT), thereby restoring a normal . This intervention depolarizes the myocardium simultaneously, halting chaotic electrical activity and allowing the to resume coordinated contractions. It is a cornerstone of (ACLS) and is indicated exclusively for shockable rhythms during , with no absolute contraindications, even in patients with pacemakers or implanted defibrillators. The history of defibrillation began with animal experiments by Jean-Louis Prévost and Frédéric Batelli in 1899, followed by the first successful human open-chest defibrillation by Claude Beck in 1947; closed-chest techniques emerged in the 1950s with Paul Zoll and William Kouwenhoven, and direct-current devices were pioneered by Bernard Lown in 1962. Modern defibrillation employs biphasic truncated exponential waveforms, which are more effective and require lower energy than older monophasic devices; the 2025 American Heart Association guidelines recommend an initial shock of 120–200 joules for adults, escalating as needed, with higher settings (≥200 J) preferable for cardioversion of atrial fibrillation. Types include manual defibrillators for healthcare professionals, AEDs for public and out-of-hospital settings, wearable cardioverter-defibrillators for high-risk patients, and implantable cardioverter-defibrillators (ICDs) for long-term prevention of sudden cardiac death. In pediatric cases, energy dosing starts at 2 joules per kilogram of body weight, increasing to a maximum of 10 joules per kilogram. For refractory VF, double sequential defibrillation—using two devices in rapid succession—may be considered, though evidence remains limited. The procedure's success is time-sensitive, with survival rates declining by approximately 10% per minute of delay in out-of-hospital ; overall survival is about 10.8% out-of-hospital and 25.5% in-hospital. Risks include skin burns, myocardial , or proarrhythmia if mistimed, but benefits far outweigh hazards in arrest scenarios. Early defibrillation, combined with high-quality CPR, remains the most effective intervention for improving outcomes in sudden , which claims around 350,000 lives annually in the U.S.

Definition and Mechanism

Definition and Principles

Defibrillation is defined as the therapeutic delivery of an electrical shock through the heart to depolarize a critical mass of myocardial tissue, thereby terminating life-threatening arrhythmias such as ventricular fibrillation (VF) or pulseless ventricular tachycardia (VT). This intervention aims to halt the disorganized electrical activity that prevents effective cardiac contraction, restoring the potential for normal sinus rhythm. In , defibrillation works by overriding the erratic electrical impulses characteristic of these arrhythmias through a high-energy that simultaneously depolarizes a large portion of the myocardium. VF and pulseless VT represent chaotic "electrical storms" in the ventricles, where uncoordinated wavefronts propagate randomly, leading to ineffective quivering or rapid contractions that fail to produce a and can rapidly progress to if untreated. The disrupts these reentrant circuits and wavefronts, creating a brief period of that allows the to reestablish organized electrical control and resume normal heart rhythm. A key principle underlying successful defibrillation is the alteration of transmembrane potential across myocardial cells, where the induces (negative change in potential) and hyperpolarization, extinguishing excitable tissue and preventing reinitiation of . For efficacy, the must affect a of approximately 95% of the ventricular myocardium to ensure all fibrillation wavefronts are simultaneously rendered unexcitable, as smaller affected areas may allow residual activity to persist. This concept, first demonstrated in experimental models, highlights the need for sufficient strength to achieve uniform tissue quiescence.

Physiological Effects

Defibrillation shocks deliver a high-energy electrical that induces a rapid change in transmembrane potential across cardiac cell membranes, leading to simultaneous of a of myocardial cells. This widespread depolarization extinguishes re-entrant circuits underlying (VF) or pulseless (VT) by overriding disorganized electrical activity and allowing the to resume control. Following the shock, the myocardium undergoes a period of recovery characterized by reperfusion, which restores oxygen delivery to ischemic tissue but can contribute to transient dysfunction. Successful defibrillation often results in restoration of , though post-shock or may occur temporarily due to , where contractile function is impaired without permanent . This stunning is exacerbated by ischemia-reperfusion injury and the shock itself, leading to reduced myocyte shortening and altered intracellular calcium dynamics, particularly with higher energy deliveries. The minimum effective energy dose for defibrillation varies, with external biphasic shocks typically requiring 120-200 joules initially to achieve success rates comparable to higher monophasic energies. Transthoracic impedance, averaging 70-80 ohms in adults, significantly influences delivery and efficacy, as higher impedance reduces shock effectiveness and may necessitate energy escalation or techniques to lower it, such as firm pressure. Patient body weight shows no clear with required levels. Excessive shock energy can cause physiological complications, including myocardial injury manifested by troponin release and of cell membranes. occurs at field strengths above 25 V/cm, creating transient pores that lead to diastolic , reduced amplitude, and potential pro-arrhythmic effects, though it may also contribute to defibrillation success by extending refractoriness. High-energy shocks (e.g., 1-2 joules in experimental models) impair myocyte contractility and elevate biomarkers like , indicating subclinical damage that worsens with multiple or intense deliveries.

Clinical Use

Indications for Use

Defibrillation is primarily indicated for the treatment of shockable rhythms during cardiac arrest, specifically ventricular fibrillation (VF) and pulseless ventricular tachycardia (pVT), as these chaotic arrhythmias disrupt effective cardiac output and require electrical termination to restore organized rhythm. In hemodynamically unstable patients with non-perfusing tachycardias, such as pulseless VT, defibrillation serves as an urgent intervention to prevent deterioration into full arrest. According to the 2025 American Heart Association (AHA) Advanced Cardiovascular Life Support (ACLS) guidelines, defibrillation plays a central role in the cardiac arrest algorithm, with an immediate unsynchronized shock recommended upon identification of VF or pVT in witnessed arrests to maximize survival chances, followed by 2 minutes of high-quality CPR before rhythm reassessment. This integration emphasizes minimizing interruptions in chest compressions, with defibrillation delivered as soon as possible after CPR cycles in non-witnessed cases, and subsequent shocks using biphasic energies of 120-200 J or monophasic 360 J if needed. The European Resuscitation Council (ERC) 2025 guidelines align closely, advocating early defibrillation within the ALS algorithm for shockable rhythms alongside high-quality CPR with continuous compressions and asynchronous ventilations at 10 per minute when an advanced airway is in place. Defibrillation is contraindicated in non-shockable rhythms, including and (), where electrical therapy offers no benefit and may delay other resuscitative efforts, as well as in stable, organized rhythms without hemodynamic instability. Presence of a or does not preclude defibrillation, though pads should avoid direct placement over the device. In special populations, indications remain focused on VF and but with tailored approaches. For patients with left ventricular assist devices (LVADs), defibrillation is indicated for VF/, but prioritize assessing and attempting to restart LVAD function if a second rescuer is available, alongside standard shocks. For opioid-associated cardiac arrest, integrate administration (initial 2 mg IV/IO, repeat if needed) before or after shocks if overdose is suspected. For pediatrics, the Pediatric Advanced Life Support (PALS) guidelines recommend defibrillation at lower energy doses—2 J/kg for the first shock, escalating to 4 J/kg and ≥4 J/kg thereafter (maximum 10 J/kg or adult dose)—integrated with age-appropriate CPR. During , defibrillation is indicated without delay for shockable rhythms, with standard adult protocols applied alongside left uterine displacement to optimize CPR effectiveness and preparation for perimortem cesarean delivery if is not achieved within 5 minutes. In post-cardiac surgery scenarios, immediate defibrillation with up to three stacked shocks within 1 minute for VF or is prioritized before initiating CPR or emergency resternotomy, to address reversible causes like graft issues while adhering to ACLS principles.

Procedure and Application

The procedure for defibrillation in clinical settings follows standardized (ALS) protocols, such as those outlined by the (AHA), to ensure rapid and safe delivery during or elective . Preparation begins with confirming a shockable , such as (VF) or pulseless (pVT), through monitor analysis after a brief pause in high-quality (CPR), typically every 2 minutes. The patient is positioned on a firm, flat surface to facilitate chest compressions and application, while the team coordinates roles—one member continues compressions if possible, another prepares the defibrillator, and all communicate clearly to minimize interruptions exceeding 10 seconds. The chest is exposed by removing clothing, jewelry, and any transdermal patches to prevent arcing or burns; if the chest is wet, it is dried thoroughly to reduce transthoracic impedance. Self-adhesive defibrillation pads are applied to the bare chest in anterolateral positions—one over the right upper sternal and the other on the left mid-axillary line at the level of the cardiac apex—or an anteroposterior configuration if needed for cases. The defibrillator is charged to the initial while CPR continues, with biphasic devices set to 120–200 joules (J) based on manufacturer recommendations, or 360 J for monophasic devices. For patients in with VF or , unsynchronized shocks are delivered immediately after confirming the rhythm and ensuring no one is touching the patient or bed. In contrast, for elective of stable tachyarrhythmias like monomorphic VT with a , synchronized shocks are used, timed to the R-wave on the ECG to avoid inducing VF, starting at lower energies such as 100–200 J depending on the . The operator announces "all clear" loudly, visually confirms team safety, and presses the shock button to deliver the energy. Defibrillation integrates seamlessly with (BLS) and algorithms, where CPR resumes immediately after the shock for a full 2-minute cycle without pausing to check the rhythm or pulse, unless signs of (ROSC) are evident, such as rising end-tidal CO2. After the cycle, the rhythm is reassessed; if VF or persists, subsequent shocks are delivered at equal or higher energy levels, with epinephrine administered every 3–5 minutes and (300 mg IV/IO bolus) given after the third shock, followed by a 150 mg dose if needed. This sequence prioritizes minimizing peri-shock pauses to maintain coronary and cerebral . Safety protocols are paramount to protect rescuers and patients. Defibrillation is avoided in oxygen-enriched environments to prevent risks, with oxygen delivery tubes positioned at least 1 meter from the chest and exhaust directed away. Operators wear appropriate , including insulated gloves for manual paddle use if required, to guard against or electrical leakage. In wet conditions, such as around pools or during , the is deferred until the patient can be moved to a dry area, and all metallic objects on the patient or rescuers are removed to avoid conduction hazards.

Outcomes and Success Rates

The effectiveness of defibrillation is primarily measured by its ability to terminate shockable rhythms such as (VF) and restore a perfusing . In clinical settings, biphasic waveform shocks achieve initial VF termination rates of 80-90% in out-of-hospital cardiac arrest (OHCA) cases when delivered early. For patients with initial shockable rhythms in OHCA, (ROSC) rates typically range from 30-50%, while survival to hospital discharge falls between 25-40%, depending on system-wide factors like rapid response. These outcomes highlight defibrillation's role as a cornerstone intervention, though overall OHCA survival remains modest at around 10% across all rhythms due to the minority presenting with shockable arrhythmias. Several factors significantly influence defibrillation success. The time from collapse to first shock is critical, with each additional minute of delay associated with a 6-10% reduction in the likelihood of successful rhythm termination and ROSC, emphasizing the need for immediate intervention. Bystander-initiated CPR prior to defibrillation enhances outcomes by preserving myocardial viability, mediating up to 50% of the survival benefit through improved VF waveform characteristics and higher ROSC rates (odds ratio 1.72). Additionally, biphasic waveforms outperform monophasic ones, yielding higher termination rates (e.g., 82.5% vs. 71.2% for first shocks) and better long-term survival when used at energies of 150-200 J. Recent advancements, including drone-delivered automated external defibrillators (AEDs), have shown promise in improving outcomes, particularly in rural areas where traditional response times are longer. Data from 2024-2025 registries and trials indicate that systems can deliver AEDs in under 5 minutes (median 4:47 minutes), arriving before in 67% of cases and potentially boosting by enabling earlier shocks in remote settings. The Registry to Enhance (CARES) 2024 report notes overall OHCA survival to discharge at 10.4%, with shockable rhythms benefiting most from such innovations, though widespread adoption requires further validation. Despite these successes, defibrillation has notable limitations in certain scenarios. In prolonged cardiac arrest exceeding 10 minutes, efficacy declines sharply, with ROSC probability dropping by up to 19% per minute of intra-arrest delay due to myocardial deterioration, often necessitating like CPR or vasopressors. In hypothermic , standard defibrillation alone yields lower success rates (under 50% ROSC), as cold-induced arrhythmias resist termination; (ECMO) outperforms conventional rewarming, achieving survival rates of 71% versus 29% in the cardiac arrest subgroup.

Types of Devices

Manual External Defibrillators

Manual external defibrillators are devices operated by trained healthcare professionals to deliver controlled electrical shocks for terminating life-threatening arrhythmias such as or pulseless . These units allow for precise control over the defibrillation process, enabling operators to select levels and timing based on needs and clinical judgment. Key design features include adjustable energy delivery, typically ranging from 2 J for pediatric patients up to 360 J for adults, which permits escalation as required during efforts. They incorporate a synchronized mode that times the shock to the patient's R-wave on the ECG to reduce the risk of inducing in hemodynamically stable tachyarrhythmias. Integration with ECG monitoring is standard, providing real-time rhythm analysis directly on the device interface to guide treatment decisions. Operationally, these defibrillators require user-initiated controls for rhythm interpretation via the integrated ECG display and manual selection of shock delivery, ensuring shocks are administered only after confirming a shockable . They are commonly deployed in high-acuity environments such as emergency rooms, intensive care units, and ambulances, where skilled personnel can rapidly assess and intervene. These devices are compatible with self-adhesive electrodes for efficient application during emergencies. Manual external defibrillators offer high customization in energy selection and waveform delivery, allowing adaptation to diverse clinical scenarios, but their use demands significant operator expertise to avoid errors in rhythm recognition or timing. Modern models predominantly employ , which have been the clinical standard since the early 2000s due to their superior efficacy and ability to achieve defibrillation at lower energy levels compared to older , thereby reducing myocardial damage. Maintenance protocols emphasize regular calibration to verify energy output accuracy and battery checks to ensure operational readiness, with annual servicing recommended to include testing, reconditioning, and submission of tracking reports to manufacturers and the FDA as required under regulations.

Automated External Defibrillators

Automated external defibrillators (s) are portable devices designed for use by laypersons or minimally trained individuals in emergency situations, featuring integrated electrocardiogram (ECG) analysis algorithms that automatically detect shockable rhythms such as (VF) and pulseless (VT). These algorithms process ECG signals from self-adhesive pads applied to the patient's chest, analyzing characteristics like frequency and amplitude to determine if a defibrillatory shock is warranted, with rates exceeding 95% for VF detection in validated databases. Upon detecting a shockable rhythm, the AED charges its and delivers a biphasic truncated shock, typically ranging from 150 to 360 joules for adults, while prompting the user to ensure safety before delivery. Many AED models incorporate pediatric modes to adapt for younger patients, reducing delivered energy to approximately 50-75 joules or 2-4 joules per kg for children under 8 years old or weighing less than 55 pounds (25 kg), often activated via a dedicated button or specialized pediatric pads that attenuate the adult dose. This mode ensures safer application during in pediatric cases, where VF occurs in about 15-20% of out-of-hospital arrests, aligning with guidelines emphasizing early defibrillation to improve survival odds. Public access defibrillation programs have proliferated since the early 2000s, with over 4.5 million units deployed across the by 2024, strategically placed in high-traffic locations such as airports, schools, and sports venues to facilitate rapid access during out-of-hospital cardiac arrests. These initiatives, supported by legislation in all 50 states mandating placement in certain public buildings, have contributed to survival rates doubling when s are used within the first few minutes of collapse. Integration with mobile applications, such as the registry, enables real-time location tracking via GPS, allowing emergency dispatchers to guide bystanders to the nearest device and alert nearby CPR-trained volunteers. The of AEDs prioritizes simplicity for non-experts, providing step-by-step audio prompts that guide actions like turning on the device, applying in an anterior-lateral , and delivering the if advised, often supplemented by visual indicators such as illuminated diagrams or flashing lights for clarity in low-light or noisy settings. Built-in self-testing features perform daily, weekly, and monthly checks on battery status, integrity, and circuitry, alerting users via status indicators if is required, thereby ensuring device readiness without manual intervention. A key limitation of AEDs is their inability to deliver synchronized shocks for hemodynamically tachyarrhythmias, as they lack timing to align the discharge with the R-wave, potentially risking if used on organized rhythms like stable VT; in such cases, manual defibrillators are preferred. Recent advancements in the incorporate enhancements, including algorithms for improved ECG that filter artifacts and noise from motion or environmental interference, enhancing rhythm detection accuracy during uninterrupted CPR without pausing for analysis.

Implantable Cardioverter-Defibrillators

Implantable cardioverter-defibrillators (ICDs) are surgically implanted devices designed to monitor heart rhythm continuously and deliver electrical therapy to prevent sudden cardiac death in high-risk patients. Unlike external defibrillators used for acute , ICDs provide long-term prophylaxis by detecting life-threatening ventricular and responding automatically. These devices have evolved since the first implantation in 1980, becoming a cornerstone of management. ICDs consist of key components including lead systems that sense cardiac signals and deliver , a with and capacitors, and sophisticated circuitry for detection. The leads, typically transvenous wires positioned in the heart chambers, transmit electrical impulses and monitor rhythms, while the powers the device for an average lifespan of 4-6 years, depending on . Modern ICDs employ tiered algorithms, starting with antitachycardia pacing to terminate slower ventricular tachycardias non-invasively, escalating to low-energy or high-energy shocks for if needed; many also include pacing to support slow heart rates. Indications for ICD implantation fall into primary and secondary prevention categories. Primary prevention is recommended for patients with and reduced (HFrEF, LVEF ≤35%) in New York Heart Association class II-III symptoms, or post-myocardial infarction with LVEF ≤30% and at least 40 days elapsed, to avert first-time sudden . Secondary prevention applies to survivors of due to or hemodynamically unstable , not attributable to reversible causes, or those with structural heart disease and syncope of suspected arrhythmic origin. These guidelines, updated in 2025, emphasize patient selection to balance benefits against procedural risks. The implantation procedure involves placing the in a subcutaneous pocket, typically below the left , under and fluoroscopic guidance. Transvenous implantation, the traditional approach, inserts leads via the into the right ventricle and sometimes atrium for sensing and pacing; this method allows dual-chamber functionality but carries risks of vascular access complications. Subcutaneous implantation positions a single lead parallel to the without entering the vascular system or heart, reducing lead-related infections and extraction challenges, though it lacks pacing capabilities in standard designs. Post-implantation, devices are programmed to patient-specific zones and therapies, with follow-up involving in-clinic interrogations; remote monitoring uses home transmitters that wirelessly upload data nightly via radiofrequency to central servers, enabling early detection of arrhythmias, depletion, or lead issues without frequent visits. By 2025, advancements include leadless ICDs, which eliminate transvenous leads entirely by integrating sensing and shocking capabilities into a fully intracardiac capsule delivered via , significantly lowering infection risks associated with leads and pockets. Additionally, the extravascular ICD (EV-ICD), which places the lead outside the while maintaining defibrillation efficacy, and the OmniaSecure lead for pediatric applications were introduced, further reducing complications. MRI-conditional designs incorporate specialized circuitry and lead materials to minimize radiofrequency heating and induced currents, allowing safe 1.5T and 3T MRI scans under specific protocols, expanding imaging options for patients with comorbidities. These innovations enhance device longevity, reduce complications, and improve through better integration with diagnostic tools.

Electrode Systems

Paddle Electrodes

Paddle electrodes are handheld, reusable devices employed in external defibrillation, featuring circular metal plates with diameters typically ranging from 8 to 12 cm for use, mounted on insulated handles equipped with buttons that must be pressed simultaneously to deliver the . These electrodes connect to the defibrillator unit via durable, coiled cables, allowing the operator to maintain control during application. The design emphasizes direct contact with the patient's chest, requiring firm —approximately 8 to 12 kg per paddle—to optimize electrical and minimize transthoracic impedance. Prior to use, conductive must be applied liberally to the surfaces to facilitate current flow and prevent arcing or to the skin. These gels are predominantly saline-based electrolytes, such as those formulated with to reduce skin-electrode interface resistance from kilo-ohms to as low as tens of ohms, ensuring efficient energy delivery without excessive voltage requirements. Non-saline alternatives, like certain alpha-based pastes, have been used historically but are less common due to potential sparking risks if improperly applied; modern protocols prioritize gels that are water-soluble, bacteriostatic, and non-gritty for clean, reliable performance. Paddle electrodes find primary application in in-hospital settings, such as during advanced cardiovascular life support (ACLS) for rapid intervention in or pulseless , where trained personnel can quickly retrieve and deploy them from crash carts. Their design supports immediate use in controlled environments, often positioned in an apex-sternum configuration for optimal current path across the heart. Advantages include the ability to apply adjustable pressure for enhanced contact in challenging anatomies, potentially achieving lower and more consistent impedance compared to poorly adhered alternatives when is correctly used. However, disadvantages encompass higher variability in transthoracic impedance due to operator-dependent application, increased of skin burns from inadequate coverage or arcing, and potential operator exposure to the during shock delivery, which can interrupt chest compressions. Self-adhesive pads are preferred in modern protocols, including those from the European Resuscitation Council and , over paddle electrodes to improve safety and reduce CPR interruptions, though paddles remain available for specific in-hospital applications.

Self-Adhesive Electrodes

Self-adhesive electrodes, also known as disposable defibrillation pads, are pre-gelled, single-use devices designed for external defibrillation and cardioversion in both manual and automated systems. These pads facilitate rapid application and enable continuous electrocardiographic monitoring prior to shock delivery, making them a standard component in modern resuscitation protocols. These electrodes consist of multi-layer constructions optimized for conductivity and patient comfort. The backing is typically a flexible polyethylene foam layer that provides insulation and structural support, while the conductive element often incorporates silver/silver chloride (Ag/AgCl) or tin-coated materials to ensure low-offset voltage and reliable signal transmission. A key feature is the conductive hydrogel adhesive layer, which interfaces directly with the skin to enhance electrical contact and adhesion. Adult pads measure approximately 10-12 cm in diameter (or equivalent area of about 100-150 cm²), whereas pediatric versions are smaller, around 8-10 cm (or 70-90 cm²), to accommodate younger patients. Snap connectors at the edges allow secure attachment to defibrillator cables.80606-7/fulltext) Application involves a simple peel-and-stick process, where the pads are applied to the patient's chest for immediate and defibrillation readiness. This reduces transthoracic impedance to an average of 70-100 ohms, facilitating efficient energy delivery without the need for additional conductive gel. Regulated by the FDA as Class II devices, self-adhesive electrodes are approved for single-use to ensure sterility and prevent cross-contamination. They have a of 2-5 years, depending on the manufacturer, and are compatible with biphasic waveforms commonly used in contemporary defibrillators. The primary advantages include hands-free operation, which minimizes interruptions in chest compressions during and reduces rescuer exposure to electrical hazards. Additionally, their disposable nature lowers infection risk in scenarios requiring prolonged monitoring or repeated interventions, as opposed to reusable alternatives.

Placement Techniques

Proper electrode placement is crucial in defibrillation to ensure optimal delivery of electrical current through the heart, maximizing the likelihood of successful termination of arrhythmias such as . The goal is to position the s so that the current vector traverses a sufficient mass of the myocardium, particularly the ventricles, to achieve simultaneous . As per the 2025 and European Resuscitation Council guidelines, self-adhesive pads are recommended to minimize CPR interruptions, with placement techniques aimed at optimizing myocardial current flow. The standard anterior-lateral (antero-apical) position for external defibrillation involves placing one electrode pad immediately to the right of the upper , below the , and the second pad on the left side of the chest in the mid-axillary line, approximately at the level of the cardiac (fifth ). This configuration directs the current flow from the right shoulder region across the heart toward the left lateral chest, effectively encompassing the ventricular myocardium. An alternative anterior-posterior position places one pad on the anterior chest (typically over the left mid-axillary line or ) and the other on the posterior (below the left or mid-back), which is often used for implantable cardioverter-defibrillators (ICDs) during testing or in scenarios where anterior-lateral access is limited. The rationale for these positions stems from the need to minimize transthoracic impedance and align the electrical parallel to the heart's long axis, thereby reducing the defibrillation —the minimum required for successful defibrillation. In the anterior-lateral setup, the approximates the heart's anatomical orientation, promoting uniform distribution across the ventricles. The anterior-posterior approach can further optimize flow in cases where anterior attenuates the , as it bypasses superficial chest wall resistance. Adjustments to placement are necessary for specific patient characteristics to maintain efficacy. In obese individuals or those with large breasts, the anterior-lateral pads may be shifted laterally or under the breast tissue to improve contact and reduce impedance from adipose layers; alternatively, the anterior-posterior position is preferred to avoid excessive anterior tissue. For pediatric patients weighing more than 25 kg or older than 8 years, either anterior-lateral or anterior-posterior positions are suitable, compatible with automated external defibrillators (AEDs). In infants under 1 year or weighing less than 10 kg, where pads may overlap in the anterior-lateral position, an anterior-posterior configuration is recommended, with one pad on the center of the chest and the other on the center of the back, as posterior placement is practical due to the child's size. Verification of placement involves ensuring firm adhesion of self-adhesive pads to the bare chest, removing any excessive hair, moisture, or medicaments that could increase impedance, and confirming at least 8-10 cm distance from implanted pacemakers or ICD generators to prevent device damage or . Prior to shock delivery, the ECG rhythm is assessed via the defibrillator's leads to verify a shockable , indirectly confirming adequate contact through signal quality. Clinical evidence supports the impact of optimal positioning on outcomes, though guidelines consider both major configurations reasonable due to historical equivalence in success rates across multiple studies. However, a 2024 of out-of-hospital cardiac arrests found anterior-posterior placement associated with a 2.64-fold higher adjusted odds of (ROSC) compared to anterior-lateral (95% 1.50-4.65), suggesting up to a 17% relative in initial success. In a porcine pediatric model, anterior-posterior placement reduced the 50% defibrillation threshold by approximately 20% compared to anterior-lateral, highlighting potential energy savings and efficacy gains in vulnerable populations. These findings underscore the value of tailoring placement to anatomical factors for enhanced defibrillation performance.

Historical Development

Early Methods and Closed-Chest Techniques

The origins of defibrillation trace back to the late , when physiologists Jean-Louis Prévost and Frédéric Batelli conducted pioneering experiments on dogs. In 1899, they demonstrated that small electric shocks could induce , while larger (AC) shocks—delivered via electrodes placed in the mouth and rectum—could revert the and restore normal heart rhythm. These findings established the foundational principle that electrical countershock could depolarize fibrillating myocardium, allowing spontaneous organized activity to resume, though the work remained confined to animal models and was not immediately applied clinically. Human application of defibrillation began in the mid-20th century through invasive open-chest techniques. In 1947, American surgeon Claude S. Beck achieved the first successful defibrillation of a heart during on a 14-year-old boy whose ventricle had fibrillated intraoperatively. Beck applied 110-volt AC shocks directly to the exposed heart using internal electrodes, administering multiple pulses of 0.5 to 2 seconds each, combined with manual cardiac massage, to restore . This landmark procedure, performed at , marked the transition from animal experimentation to clinical use but was limited to surgical settings due to the need for . The development of closed-chest techniques emerged in the early , pioneered by cardiologist Paul M. Zoll, who adapted external defibrillation for non-invasive application. In 1956, Zoll reported the first successful use of transthoracic shocks on humans, delivering 480 to 1,000 volts of low-frequency (60 Hz) through large plate electrodes applied to the chest wall, often requiring multiple discharges to terminate . His bulky, wheeled device—typically around 150 kg—enabled without opening the chest, expanding potential use beyond the operating room. However, these early methods carried significant risks, including severe skin burns from high-energy delivery and induction of additional arrhythmias due to the prolonged, oscillating nature of the current, which could damage myocardial tissue.

Shift to Direct Current and Portable Devices

The transition from () to () defibrillation marked a pivotal advancement in the 1960s, addressing the limitations of earlier AC methods that often caused myocardial burns and arrhythmias. In 1959, cardiologist and his team developed a monophasic damped sinusoidal DC waveform defibrillator, which delivered controlled energy pulses of 100 to 400 joules, significantly reducing tissue damage while effectively terminating . This innovation, first clinically applied in 1962, shifted clinical practice toward safer, more reliable for both atrial and ventricular arrhythmias, establishing DC as the standard over AC by the mid-1960s. Parallel developments in portability transformed defibrillation from a hospital-bound procedure to one feasible in emergency settings. In 1966, Northern cardiologist introduced the first external portable defibrillator, a 70 kg unit powered by car batteries rather than , enabling its installation in ambulances for pre-hospital use in . By the , refinements led to more compact battery-operated models weighing around 25 kg, which facilitated widespread adoption in mobile coronary care units and further integrated defibrillation into out-of-hospital response protocols. Key contributions from researchers William B. Kouwenhoven, James R. Jude, and G. Guy Knickerbocker in the late and early integrated defibrillation with closed-chest (CPR), enhancing overall resuscitation efficacy. Their work demonstrated that combining external chest compressions with timely DC shocks improved circulation and rhythm restoration during , laying the foundation for modern protocols. These advancements enabled pre-hospital defibrillation, dramatically improving outcomes for out-of-hospital cardiac arrests; early programs with portable units in settings achieved survival rates up to 15-20%, compared to approximately 5% without prompt . Early paddle electrodes, typically 8-12 cm in and applied with conductive , supported this portability by allowing quick manual delivery of shocks in non-hospital environments.

Modern Advancements in Waveforms and Implantables

In the , defibrillation technology advanced significantly with the development of biphasic truncated exponential (BTE) waveforms, which reversed the direction of current flow during the shock to improve efficacy and reduce required energy levels compared to traditional monophasic damped sine waveforms. Clinical studies demonstrated that BTE shocks at 150-200 J achieved defibrillation success rates comparable to or exceeding those of 360 J monophasic shocks, with lower risks of myocardial injury and faster post-shock recovery. This shift, validated in multicenter trials, led to widespread adoption of biphasic waveforms in external and implantable defibrillators by the early 2000s, enhancing patient outcomes in both hospital and out-of-hospital settings. Implantable cardioverter-defibrillators (ICDs) represent a cornerstone of modern advancements, originating with Michel Mirowski's pioneering work; the first human implant occurred in 1980 at , successfully terminating ventricular arrhythmias in a patient with recurrent malignant episodes. Over the subsequent decades, ICDs evolved from bulky transvenous systems to subcutaneous models, which avoid intravascular leads to reduce and complication risks. By 2025, subcutaneous ICDs such as the EMBLEM MRI S-ICD system offer projected battery longevity of 8.7 years, enabling reliable long-term monitoring and therapy delivery for high-risk patients. These devices now incorporate advanced sensing algorithms, further minimizing inappropriate shocks while maintaining high efficacy rates. Recent innovations have integrated artificial intelligence (AI) into rhythm detection for both automated external defibrillators (AEDs) and ICDs, improving accuracy in identifying shockable arrhythmias like ventricular tachycardia and fibrillation amid noisy signals or CPR artifacts. Machine learning models, such as convolutional neural networks, achieve over 95% sensitivity and specificity in classifying rhythms from single-lead ECGs, reducing diagnostic delays and enhancing survival in cardiac arrest scenarios. Complementing this, drone delivery systems for AEDs have undergone trials across Europe from 2023 to 2025, demonstrating feasibility in rural and urban areas; for instance, a Swedish program achieved median delivery times of 5 minutes with a time advantage of approximately 3 minutes over ambulances in 67% of cases, while Danish studies reported median drone delivery of 4:47 minutes in feasible scenarios. To address global disparities, efforts in low-resource settings have focused on reusable cardiac implantable electronic devices, including defibrillators, to expand access without prohibitive costs; studies from high- and low-income contexts show no increased risks of or device failure when protocols for , testing, and sterilization are followed. These initiatives, supported by organizations, promote protocols for refurbishing explanted units, potentially extending their usability and aligning with broader goals for equitable sudden cardiac death prevention in underserved regions.

Societal and Cultural Impact

Public Access and Training Programs

Public Access Defibrillation (PAD) programs emerged in the United States during the 1990s, spearheaded by the American Heart Association (AHA), which established a task force in 1990 to promote widespread AED availability and training for lay responders. These initiatives expanded through federal funding, including the Cardiac Arrest Survival Act of 2000, which allocated grants for AED purchase and placement in public venues, and the ongoing Health Resources and Services Administration's Rural Access to Emergency Devices program, supporting AED deployment in underserved areas since the early 2000s. Training programs are central to PAD success, with the AHA offering standardized courses like Heartsaver First Aid CPR AED, which emphasize performing high-quality CPR in 2-minute cycles at a rate of 100-120 compressions per minute to minimize interruptions and integrate AED use seamlessly. These courses incorporate AED features, such as voice prompts guiding users through pad application and shock delivery. Mobile applications like PulsePoint further support bystander response by sending real-time alerts to CPR-trained individuals about nearby cardiac arrests, including AED locations from a national registry, enabling rapid intervention. Studies demonstrate the effectiveness of public access, with defibrillation within 3-5 minutes of collapse increasing out-of-hospital survival rates to 50-70%, compared to under 10% without prompt intervention, as evidenced by analyses from the and European Resuscitation Council. For instance, the Public Access Defibrillation Trial showed that trained lay responders using in community settings doubled survival to discharge rates versus CPR alone. Despite these advances, challenges persist in PAD implementation, including maintenance issues such as expired batteries, faulty , and inadequate monthly checks, which can render up to 20% of public inoperable and contribute to preventable failures. Urban-rural disparities exacerbate access gaps, with rural areas facing lower AED density—often fewer than half the units compared to cities—and longer emergency response times, resulting in survival rates 20-30% below urban benchmarks. Addressing these requires targeted funding for rural deployment and standardized protocols to ensure equitable PAD coverage. Good Samaritan laws provide legal protections for lay rescuers who administer defibrillation using automated external defibrillators (AEDs) in good faith during emergencies, shielding them from civil liability in the absence of . In the United States, all 50 states offer such immunity, with 48 states extending protections to untrained bystanders and only three requiring prior training. Internationally, similar protections exist in numerous countries, encouraging bystander intervention without fear of lawsuits, though requirements for training vary; for instance, unrestricted use by any citizen is permitted in 15 countries, while certification is mandated in three others. Ethical considerations in defibrillation encompass dilemmas around and , particularly in crises. For unconscious patients experiencing , implied is ethically justified for immediate life-saving interventions like defibrillation, as obtaining explicit permission is impossible and delay could be fatal; healthcare providers must act in the patient's best interest under principles of beneficence and non-maleficence. During the (2020-2022), resource shortages strained efforts, leading to protocols that prioritized patients based on survival likelihood and resource needs, sometimes limiting defibrillation or advanced care in overwhelmed settings to allocate ventilators and ICU beds equitably. These decisions raised concerns over fairness, with guidelines emphasizing transparent, evidence-based criteria to avoid . Economically, AED deployment involves upfront costs balanced against long-term health benefits. A single unit typically ranges from $1,200 to $3,000, depending on model and features, with home models averaging around $1,450 and professional units higher. Cost-effectiveness analyses demonstrate value in public settings, where bystander defibrillation yields an incremental cost of approximately $46,700 per (QALY) gained, reflecting improved survival and reduced morbidity from out-of-hospital cardiac arrests. As of 2025, regulatory updates emphasize workplace mandates and global disparities. In regions like , , legislation requires AEDs on construction sites with 20 or more workers lasting three months or longer, enhancing in high-risk environments. Similarly, mandates employer-provided AED by January 2026, with trained personnel available. On a global scale, equity remains challenged, with fewer public AEDs in low-income neighborhoods and regions compared to affluent areas, exacerbating survival gaps in low- and middle-income countries where is limited by and . Efforts to address these include calls for subsidized distribution to promote equitable deployment.

References

  1. [1]
    Defibrillation - StatPearls - NCBI Bookshelf - NIH
    Mar 23, 2025 · Defibrillation depolarizes the heart muscle by delivering a controlled electrical shock, allowing the sinoatrial node to reestablish a ...
  2. [2]
    Part 9: Adult Advanced Life Support
    Additionally, guidelines are provided for the use of double sequential defibrillation, head-up cardiopulmonary resuscitation, and point-of-care ultrasound ...
  3. [3]
    On Occasion of Seventy-five Years of Cardiac Defibrillation in Humans
    The first portable EMS defibrillators (used by paramedics) emerged in the early 1970s. In 1980 the automatic implantable cardioverter-defibrillator was invented ...
  4. [4]
    History of CPR | American Heart Association CPR & First Aid
    1947. In Cleveland, Ohio, cardiothoracic surgeon Dr. Claude Beck performs the first successful use of an electric defibrillator on an exposed human heart.
  5. [5]
    What is Defibrillation Used For? - Cleveland Clinic
    Defibrillation is the use of an electrical current to help your heart return to a normal rhythm when a potentially fatal arrhythmia (abnormal heart rhythm) is ...
  6. [6]
    Part 4: The Automated External Defibrillator | Circulation
    The term “defibrillation” means reversal of the action of fibrillation. Defibrillation is not a synonym for “shock.“ Thus, defibrillation should be understood ...
  7. [7]
    Mechanisms of Defibrillation - PMC - NIH
    In this review, we discuss recent work that demonstrates the various mechanisms responsible for defibrillation.
  8. [8]
  9. [9]
    [Electrical defibrillation and cardioversion] - PubMed
    This induces transmembrane potential in myocardium cells and results in synchronic depolarization of all myocardium.Missing: cardiac | Show results with:cardiac
  10. [10]
    Electroporation of the heart - PubMed - NIH
    According to the excitation theory of defibrillation, such shocks are aimed at depolarizing the membranes of most cardiac cells resulting in resynchronization ...
  11. [11]
    Part 9: Post–Cardiac Arrest Care | Circulation
    Treat Myocardial Stunning. ◦. Fluids to optimize volume status (requires clinical judgment). ◦. Dobutamine 5–10 mcg/kg per min. ◦. Mechanical augmentation ...
  12. [12]
    High-energy Defibrillation Impairs Myocyte Contractility ... - PubMed
    Higher defibrillation energy significantly impairs ventricular contractility at the myocyte level. Reductions in cardiomyocyte shortening and intracellular ...
  13. [13]
    Part 6: Advanced Cardiovascular Life Support : Section 2: Defibrillation
    The percentage of patients who fail to respond to a first or successive biphasic shock at a constant energy of ≤200 J remains unknown.
  14. [14]
    Effect of electroporation on cardiac electrophysiology - PubMed - NIH
    According to the excitation theory of defibrillation, such shocks are aimed at depolarizing the membranes of most cardiac cells, resulting in resynchronization ...
  15. [15]
    Myocardial injury secondary to ICD shocks: insights from patients ...
    Sep 2, 2013 · Troponin elevation occurred in the majority of our patients after inappropriate ICD discharges secondary to lead fracture.Missing: induced | Show results with:induced
  16. [16]
    [PDF] Adult Cardiac Arrest Algorithm (VF/pVT/Asystole/PEA)
    © 2025 American Heart Association. Start CPR. • Begin bag-mask ventilation and give oxygen. • Attach monitor/defibrillator. 1. CPR 2 min. • IV/IO access.Missing: indications | Show results with:indications<|control11|><|separator|>
  17. [17]
    Part 9: Adult Advanced Life Support: 2025 American Heart ...
    Oct 22, 2025 · The 2025 guidelines provide guidance for the treatment of cardiac arrest, including ventricular fibrillation, pulseless ventricular tachycardia, ...
  18. [18]
    Guidelines 2025 - English - European Resuscitation Council
    The European Resuscitation Council is excited to announce the launch of the Guidelines on Cardiopulmonary Resuscitation 2025 on October 22, 2025, in Rotterdam.Epidemiology In... · Adult Advanced Life Support · Newborn Life Support<|separator|>
  19. [19]
    [PDF] Pediatric Cardiac Arrest Algorithm
    Push hard (≥ 1/3 chest depth). • Push fast: 100-120/min. • Allow full chest recoil. • Minimize interruptions in compressions.
  20. [20]
    [PDF] Cardiac Arrest in Pregnancy Algorithm
    Provide continuous left lateral uterine displacement when the fundal height is at or above the umbilicus. • Use AED/defibrillator when indicated. Continue ALS.Missing: defibrillation | Show results with:defibrillation
  21. [21]
    Part 10: Adult and Pediatric Special Circumstances of Resuscitation
    Oct 22, 2025 · In these guidelines, the American Heart Association provides updated guidance for resuscitation of adults and children in cardiac arrest or ...Alternative Cpr Techniques · Cardiac Intervention... · Cardiac Surgery
  22. [22]
    [PDF] European Resuscitation Council Guidelines 2025 Adult Advanced ...
    Since 2015, ERC defibrillation guidelines have referred solely to biphasic energy waveforms and defibrillation pads. In this ERC Guideli- nes 2025 ALS, we ...
  23. [23]
    BIPHASIC Trial | Circulation - American Heart Association Journals
    These study findings suggest that patients in VF benefit from higher biphasic energy levels if multiple defibrillation shocks are required. Acknowledgments.
  24. [24]
    Temporal Patterns in Out-of-Hospital Cardiac Arrest Incidence and ...
    Jul 16, 2025 · Survival to hospital discharge was 42.6% among those presenting with a shockable rhythm and 8.8% among those presenting with a nonshockable ...
  25. [25]
    Part 9: Adult Advanced Life Support: 2025 American Heart ...
    Oct 21, 2025 · The 2025 guidelines provide guidance for the treatment of cardiac arrest, including ventricular fibrillation, pulseless ventricular tachycardia, ...
  26. [26]
    Association Between Delay to First Shock and Successful First ...
    Oct 27, 2024 · In patients with out-of-hospital cardiac arrest (OHCA), ≈10% to 40% have an initial shockable rhythm. It has been well established that survival ...
  27. [27]
    Insights From the Ventricular Fibrillation Waveform Into the ...
    Sep 25, 2021 · Bystander CPR correlated with more robust initial VF waveform measures, which in turn mediated up to one‐half of the survival benefit associated with bystander ...
  28. [28]
    Semi-autonomous drone delivering automated external defibrillators ...
    Feb 15, 2025 · This Danish drone AED-delivery study demonstrated the technical and operational feasibility of drone-delivered AEDs for real OHCA cases, ...
  29. [29]
    [PDF] CARES Summary Report - myCares™
    Apr 21, 2025 · Pre-hospital Outcome (%). N=137119. Pronounced in the Field. 59026 (43.0). Pronounced in ED. 13579 (9.9). Ongoing ... AED applied by bystander)Missing: drone | Show results with:drone
  30. [30]
    The impact of time to defibrillation on return of spontaneous ...
    Every one-minute delay in intra-arrest defibrillation is predicted to decrease the likelihood of achieving ROSC by 19%.Missing: decline per
  31. [31]
    Extracorporeal membrane oxygenation versus conventional ...
    Aug 24, 2022 · Among ED patients with severe hypothermia and cardiac arrest, survival was significantly higher with ECMO versus conventional rewarming.
  32. [32]
    Types of Defibrillator – Overview & Uses - Avive AED
    Nov 10, 2023 · Patients experiencing cardiac arrest at the hospital or in an ambulance might be treated with a manual external defibrillator. For example, if ...<|separator|>
  33. [33]
    Energy levels in manual defibrillation after prior AED shock - PubMed
    Sep 23, 2025 · Our regional manual defibrillators are by default set to deliver 200J for the initial shock and 360J for subsequent shocks. In case of a prior ...
  34. [34]
  35. [35]
    [PDF] LIFEPAK 15 monitor/defibrillator - Operating instructions - Stryker
    Manual mode—for performing manual defibrillation, synchronized cardioversion, noninvasive pacing, and ECG and vital sign monitoring. • Archive mode—for ...
  36. [36]
    Monophasic vs Biphasic – What's the Difference in AED Shocks | MDSI
    Jul 28, 2025 · Biphasic defibrillation is now the standard. It has completely replaced monophasic devices, and those are no longer made. Newer models like ...
  37. [37]
    What is manual defibrillator preventative maintenance?
    Apr 16, 2021 · Pacer is tested, analyzed, calibrated; Cosmetic condition check and unit cleaning; Tracking reports submitted to manufacturer and FDA for ...
  38. [38]
    Machine Learning Techniques for the Detection of Shockable ... - NIH
    Jul 21, 2016 · AED algorithms for VF-detection are customarily assessed using Holter recordings from public electrocardiogram (ECG) databases, which may be ...
  39. [39]
    Convolution Neural Network Algorithm for Shockable Arrhythmia ...
    Mar 21, 2023 · This machine learning algorithm tested within an automated external defibrillator demonstrated high performance to classify a shockable rhythm ...
  40. [40]
    Use of Automated External Defibrillators for Children: An Update
    Automated external defibrillators (AEDs) may be used for children 1 to 8 years of age who have no signs of circulation. Ideally the device should deliver a ...Rationale For Aed Use · Incidence Of Vf In Children · Clinical DataMissing: modes | Show results with:modes
  41. [41]
    What are pediatric modes on AEDs? - Safetymax
    Pediatric mode adjusts the shock energy and guidance for use on children under 8 years old or weighing less than 55 lbs. Some AEDs use special pediatric ...
  42. [42]
    FDIC 2024: Rethinking Public Access to Defibrillation
    Apr 5, 2024 · Despite the deployment of over 4.5 million AED units in the United States, an official government sponsored AED registry does not exist. Some ...<|separator|>
  43. [43]
    Next Generation AED Registry - PulsePoint
    PulsePoint AED is a simple yet sophisticated tool to record and share AED location information. You can help even before a life is in danger by easily ...
  44. [44]
    [PDF] AED Plus Administrator's Guide - accessdata.fda.gov
    The AED Plus performs a self test whenever batteries are installed and verifies the following functions: 1. Defibrillation Electrodes Connection: Verifies that ...Missing: calibration | Show results with:calibration
  45. [45]
    Part 6: Electrical Therapies | Circulation
    Although AEDs are not designed to deliver synchronized shocks (ie, cardioversion for VT with pulses), AEDs will recommend a (nonsynchronized) shock for ...
  46. [46]
    Role of artificial intelligence in defibrillators: a narrative review - PMC
    Jul 5, 2022 · In this review, we examine the use of machine learning in rhythm classification in automated external defibrillators (AEDs) particularly without interruption ...Missing: synchronize 2020s
  47. [47]
    Improvements in 25 Years of Implantable Cardioverter Defibrillator ...
    Components and Function​​ An ICD contains a battery, a capacitor to store and deliver charges, a microprocessor and integrated circuits for electrogram sensing, ...
  48. [48]
    Implantable Defibrillator - StatPearls - NCBI Bookshelf - NIH
    In some patients, the placement of an implantable cardiac defibrillator significantly reduces the risk of sudden cardiac death.
  49. [49]
    New strategies for energy supply of cardiac implantable devices - PMC
    ICD: 40 J per defibrillation battery life span: 4–6 years. Size and duration of the current pulses (continuous or intermittent operation), Corrosion resistant ...
  50. [50]
    Overview of implantable cardioverter defibrillator and cardiac ... - NIH
    When an ICD detects ventricular tachyarrhythmias based on physician-programmed algorithms, it initiates therapy by way of anti-tachycardia pacing and/or ...
  51. [51]
    ICD Therapy for Primary Prevention in HFrEF: Key Points
    Dec 11, 2024 · ICDs are also recommended (Class I) in patients ≥40 days after an AMI with LVEF <30%, even in the absence of symptoms. Patients with advanced HF ...
  52. [52]
    Transvenous versus subcutaneous implantable cardiac defibrillators ...
    An ICD system is implanted to monitor your heart rhythm and help treat dangerously fast or slow arrhythmias. Transvenous: implanted through a vein. Subcutaneous ...Missing: remote | Show results with:remote
  53. [53]
    Remote Monitoring of Implantable Cardioverter-Defibrillators ...
    Transtelephonic monitoring provides limited data on pacemaker function, including a real-time electrocardiogram, and battery longevity information through a ...
  54. [54]
    Current State of MRI With Cardiac Devices
    May 21, 2024 · MRI-conditional labeling is provided by the CIED manufacturers and relates to design features intended to reduce interaction with magnetic ...
  55. [55]
    [PDF] Defibrillation - LMS - Resuscitation Council
    When bystander CPR is given, the decrease in survival is more gradual and averages 3 - 4% per minute from collapse to defibrillation. Bystander CPR can ...<|separator|>
  56. [56]
    Electrical Therapy | Thoracic Key
    Jul 10, 2016 · When using hand-held paddles, the use of gels, pastes, or pregelled defibrillation pads aids the passage of current at the interface between the ...Missing: design advantages disadvantages
  57. [57]
    Defibrillators - Frank's Hospital Workshop
    To ensure a good connection and to minimise the electrical resistance the user has to apply conductive gel on the electrodes before placing the paddles on the ...Missing: disadvantages | Show results with:disadvantages
  58. [58]
    Signagel® Electrode Gel - Parker Laboratories
    Highly conductive, multi-purpose electrolyte meets all the standards of the ideal saline electrode gel. Recommended for ECG, defibrillation, biofeedback and EMG ...Missing: paddles type alpha
  59. [59]
    [PDF] Principles of External Defibrillators - Semantic Scholar
    Electrode pastes and gels can spread between the two paddles, creating the potential for a spark and should not be used. Modified prototype AEDs record ...
  60. [60]
    Self-adhesive pads in defibrillation or cardioversion: risks due to ...
    Self-adhesive pads can minimize the downtime of chest compressions before and after defibrillation and protect the patient from electric arc formation, skin ...Missing: design disadvantages
  61. [61]
    Defibrillation Pads and Paddles - LITFL
    Defibrillation pads are used to facilitate cardioversion and defibrillation; some allow ECG monitoring and external pacing; paddles are becoming obsolete.
  62. [62]
    Washable and Flexible Screen-Printed Ag/AgCl Electrode on ... - NIH
    Sep 6, 2023 · Ag/AgCl has proven to be one of the ideal materials for the electrode with low frequency recordings and exhibits very low offset voltage and ...
  63. [63]
  64. [64]
    Self-adhesive preapplied electrode pads for defibrillation ... - PubMed
    The transthoracic impedance using self-adhesive pads was 75 +/- 21 ohms (mean +/- standard deviation), which is similar to previously reported transthoracic ...
  65. [65]
    [PDF] summary of safety and effectiveness data (ssed) - accessdata.fda.gov
    Aug 7, 2020 · The results of the study demonstrated that the adhesive materials used in the defibrillation electrodes have sufficient adhesive performance to ...
  66. [66]
    [PDF] P200004 SSED - accessdata.fda.gov
    approved self-adhesive electrodes provided by the original equipment manufacturer ... PadPro. MFE devices will be distributed with a shelf life of three (3) years ...
  67. [67]
    [PDF] Hands-Free Defibrillation Pads: Clinical Effectiveness and Evidence ...
    Mar 29, 2010 · Four observational studies were identified regarding the clinical effectiveness of using hands- free defibrillation pads compared to paddles ...Missing: advantages infection risk
  68. [68]
    5 Reasons Why Single-Use Electrodes are Essential for Patient ...
    Jul 1, 2024 · Reusable electrodes, even with rigorous cleaning protocols, can harbor bacteria and pathogens that pose a risk of cross-contamination between ...
  69. [69]
    Part 6: Defibrillation | Circulation
    Oct 19, 2010 · In these reports, sinus rhythm with a pulse was restored using different pacing techniques. Treatment Recommendation. Electric pacing is not ...
  70. [70]
    Comparison of defibrillation efficacy between two pads placements ...
    The anterior–posterior pads placement lowered the defibrillation energy requirements and yielded a higher success rate when compared to the anterior–lateral ...
  71. [71]
    Part 4: Pediatric Basic and Advanced Life Support: 2020 American ...
    Oct 21, 2020 · Pediatric defibrillation: current flow is improved by using “adult” electrode paddles. Pediatrics. 1994;94:90–93. Crossref · PubMed · Google ...
  72. [72]
    The Ultimate AED Unit Pediatric Electrode Pad Guide
    May 9, 2023 · With our universal pads, users can press the “Child Button” on the AED, which switches the device between “adult” and “child” mode, to attenuate ...
  73. [73]
    Defibrillator Pad Position and Shockable Cardiac Arrest
    Sep 9, 2024 · In this cohort study of patients with OHCA and VF or pVT, AP defibrillator pad placement was associated with higher ROSC compared with AL placement.
  74. [74]
    Bernard Lown and defibrillation - ResearchGate
    Aug 7, 2025 · In 1959, under the supervision of Lown, a monophasic waveform defibrillator was developed. The first clinical use and documented electrical ...
  75. [75]
    Frank Pantridge • LITFL • Medical Eponym Library
    Jul 18, 2023 · Pantridge defibrillator (1965)​​ First 'portable' defibrillator operated from car battery and weighed 70kg – using a miniature capacitor ...Missing: batteryless | Show results with:batteryless
  76. [76]
    The introduction of defibrillation and external chest compressions ...
    Resuscitation great. Kouwenhoven, Jude and Knickerbocker: The introduction of defibrillation and external chest compressions into modern resuscitation.
  77. [77]
    Improved Out-of-Hospital Cardiac Arrest Survival Through the ...
    Apr 7, 1999 · Reported cardiac arrest survival rates range from 1% to 20%, and this variation can be attributed, in part, to community differences in the " ...
  78. [78]
    Biphasic truncated exponential waveform defibrillation - PubMed
    This paper presents data from studies that have compared the efficacies of biphasic truncated exponential (BTE) and monophasic damped sine (MDS) waveform ...
  79. [79]
    Relative Efficacy of Monophasic and Biphasic Waveforms for ...
    One major difference between the 2 biphasic waveforms is that the truncated exponential waveform has an abrupt onset, with the leading edge of the waveform ...
  80. [80]
    Multicenter Comparison of Truncated Biphasic Shocks and Standard ...
    We found that 130-J biphasic truncated transthoracic shocks defibrillate as well as the 200-J monophasic damped sine wave shocks that are traditionally used in ...
  81. [81]
    Termination of Malignant Ventricular Arrhythmias with an Implanted ...
    Aug 7, 1980 · Termination of Malignant Ventricular Arrhythmias with an Implanted Automatic Defibrillator in Human Beings. Authors: M. Mirowski, M.D., Philip R ...
  82. [82]
    EMBLEM™ MRI S-ICD System - Boston Scientific
    Projected Longevity, 8.7 years¹ ; Battery Chemistry, Boston Scientific Li/MnO2 ; Warranty, 6 years* ; Remote Patient Monitoring Capability, Enabled for use with ...
  83. [83]
    Convolution Neural Network Algorithm for Shockable Arrhythmia ...
    Mar 21, 2023 · We developed a 6‐layer CNN to detect shockable rhythms resulting from VT/VF, taking as input only the single‐lead ECG data without other patient ...
  84. [84]
    Drone delivery of automated external defibrillators compared with ...
    AED-equipped drones dispatched in cases of suspected out-of-hospital cardiac arrests delivered AEDs before ambulance arrival in two thirds of cases, with a ...
  85. [85]
    Reuse of cardiac implantable electronic devices to improve ... - NIH
    Jan 23, 2017 · We need to develop and validate protocols and standards for collecting, cleaning, electrically testing and sterilising CIEDs for reuse. The ...Missing: units | Show results with:units
  86. [86]
    Reuse of pacemakers, defibrillators and cardiac resynchronisation ...
    Jan 23, 2017 · Reuse of explanted pacemakers is a possible solution, but is still not widespread because of concerns regarding outcomes, especially infection.Missing: units | Show results with:units
  87. [87]
    American Heart Association Automated External Defibrillator ...
    Public access defibrillation (PAD) was first introduced by the American Heart Association (AHA) in 1990 with the creation of the Task Force on the Future of ...
  88. [88]
    CPR and AED Laws | Sudden Cardiac Arrest Foundation
    It authorized $30 million in federal grants to be made available to applying states and localities. Grant funds were used for the purchase and placement of AEDs ...
  89. [89]
    AL # 93.259 Rural Access to Emergency Devices Grant and Public ...
    The purpose of the Rural Access to Emergency Devices (RAED) Grant Program is to (1) purchase automated external defibrillators (AEDs) that have been approved, ...
  90. [90]
    Europe Automated External Defibrillator (AED) Market Outlook to 2030
    This directive is expected to result in the deployment of over 50,000 new AED units across Europe by 2025. The directive is part of a broader public health ...
  91. [91]
    Part 7: Adult Basic Life Support
    Early, high-quality CPR and prompt defibrillation are the most important interventions associated with improved outcomes in adult cardiac arrest. Despite this, ...
  92. [92]
    What is CPR | American Heart Association CPR & First Aid
    What is CPR and why is it so important? Learn about CPR steps, how to do CPR, and why AHA has a vision for a world where no one dies of cardiac arrest.CPR in Schools™ Training Kits · Hands-Only CPR · Cardiac Arrest vs Heart Attack
  93. [93]
    PulsePoint | Building informed communities
    The PulsePoint AED app, with native integration with the National Emergency AED Registry (NEAR), ensures that nearby AEDs can be placed in motion during a ...Join with us. Download today. · PulsePoint Respond · PulsePoint AED · Resources
  94. [94]
    The Role of Automated External Defibrillator Use in the Out-of ...
    Oct 26, 2023 · The chance of survival significantly increases to 50%-74% when immediate defibrillation with an on-site AED is available [2,3-4]. In recent ...
  95. [95]
    Make defibs accessible 24/7 to help increase survival rates across ...
    Oct 17, 2023 · Early defibrillation within 3–5 min of collapse can increase survival rates by as high as 50-70%, yet research shows access to this lifesaving ...
  96. [96]
    Public-Access Defibrillation and Survival after Out-of-Hospital ...
    Aug 12, 2004 · Our results show that use of AEDs by trained volunteers is safe and effective when initiated in public locations where there is at least a ...
  97. [97]
    Report on AED Failures: A Reminder that Maintenance is Vital
    Aug 30, 2011 · Over 1,000 deaths are linked to AED failures, with battery, pad, and connector issues. Most failures can be prevented with maintenance, and low ...
  98. [98]
    Barriers and facilitators to public access defibrillation in out-of ...
    Maintenance of AEDs was variable. One study reported that all but 1 of 206 AEDs were 'operable' and ready for use,50 but many AEDs were not maintained ...
  99. [99]
    Public access defibrillation is insufficiently available in rural regions
    Feb 28, 2018 · Our findings underline the need for a higher density of AED placement in rural areas. Since only 6 bystanders in 240 cases were not willing to ...Missing: maintenance disparities
  100. [100]
    Improving Access to Automated External Defibrillators in Rural and ...
    Jul 4, 2020 · These challenges are magnified in remote and rural areas, where population density is low and emergency medical service (EMS) response times are ...Missing: maintenance disparities
  101. [101]
    Public access defibrillation: challenges and new solutions - PMC
    Apr 24, 2023 · Defibrillation with on-site automated external defibrillators (AEDs) has been the conventional approach for public access defibrillation. This ...Missing: synchronize AI 2020s
  102. [102]
    Public Access Defibrillation (PAD) State Law Fact Sheet - CDC
    May 15, 2024 · In 48 states, untrained lay rescuers receive protection, and civil immunity is provided only to trained lay rescuers in three states.
  103. [103]
  104. [104]
    Good Samaritan Laws and CPR - Emergency First Response
    In most states, provinces and countries, Good Samaritan laws protect bystanders who attempt to save someone's life using CPR or an AED.
  105. [105]
    The Automated External Defibrillator: Heterogeneity of Legislation ...
    Oct 28, 2021 · The results revealed that every citizen can use an AED in 15 countries whereas a training certificate was required in three countries. In one ...
  106. [106]
    Part 3: Ethics | Circulation - American Heart Association Journals
    This section provides guidelines for healthcare providers who are faced with the difficult decision to provide or withhold emergency cardiovascular care.Ethical Principles · Dnar Orders In Ohca · Dnar Orders In Ihca
  107. [107]
    Scarce-Resource Allocation and Patient Triage During the COVID ...
    Health care systems have faced or will face difficult decisions about triage, allocation, and reallocation. These decisions should be guided by ethical ...
  108. [108]
    Part 3: Ethics: 2025 American Heart Association Guidelines for ...
    Oct 22, 2025 · In this chapter, the American Heart Association provides guidance on ethical considerations relevant to cardiopulmonary resuscitation and ...<|separator|>
  109. [109]
  110. [110]
    Evidence of Impact for Public Access Defibrillation
    One study found survival rates as high as 70% if AEDs are used within 2 minutes of collapse during the cardiac arrest.
  111. [111]
    Working For Workers Seven Act, 2025 | Ontario Newsroom
    May 28, 2025 · Requiring automated external defibrillators (AEDs) on construction projects with 20 or more workers that are expected to last three months or ...Missing: mandates | Show results with:mandates
  112. [112]
    Minnesota mandates employer access to automatic external ...
    Mar 20, 2025 · The bill mandates that employers provide employees with reasonable access to automatic external defibrillators (AEDs) and ensure that trained personnel are ...<|separator|>
  113. [113]
  114. [114]
    Gaps in Public Access Defibrillation: Analysis of International ...
    Feb 26, 2025 · Searches identified 419,248 legal records. Enabling Legislation regarding AED provision in public areas was present in 60% U.S. states, 8% ...