Elastography

Elastography is a non-invasive modality that quantifies the biomechanical properties of soft tissues, particularly and elasticity, by applying mechanical and measuring the resulting deformation or wave propagation. This technique leverages the principle that pathological tissues, such as tumors or fibrotic areas, often exhibit altered mechanical properties compared to healthy tissue, enabling differentiation based on elasticity rather than just anatomy. Developed in the late 1980s and early 1990s as an extension of and (MRI), elastography has evolved into a quantitative tool grounded in , which relates applied to through the . The core principles of elastography involve perturbing the tissue—either through quasi-static compression, harmonic vibrations, or transient acoustic pulses—and imaging the mechanical response, such as displacement, , or shear wave speed, to infer properties like (E) or (G). In shear wave elastography, a subtype of elastography, techniques typically use acoustic radiation force or external probes to generate shear waves, whose propagation velocity (c_s = √(G/ρ), where ρ is tissue density) indicates stiffness, with stiffer tissues showing higher speeds. Magnetic resonance elastography (MRE), on the other hand, employs low-frequency mechanical drivers (e.g., 50–100 Hz) synchronized with MRI phase-contrast sequences to map three-dimensional stiffness distributions. Both modalities assume tissue and for simplification, though real tissues may exhibit and , influencing accuracy. Elastography encompasses several subtypes, broadly categorized by imaging platform and perturbation method. Ultrasound elastography includes strain elastography (quasi-static, qualitative or semi-quantitative via strain ratio) and shear wave elastography (dynamic, quantitative subtypes like transient elastography (TE), point shear wave elastography (pSWE), and two-dimensional shear wave elastography (2D-SWE)), which measure wave speeds in meters per second or moduli in kilopascals. MRE provides volumetric assessment, particularly useful for deep organs, by inverting wave equations to produce elastograms. Optical elastography, employing light-based techniques such as optical coherence elastography, targets high-resolution imaging of superficial tissues. Emerging variants, such as surface acoustic wave elastography, target superficial tissues but remain less widespread. Clinically, elastography is widely applied for non-invasive disease assessment, with liver fibrosis staging being the most established use; for instance, TE achieves an area under the receiver operating characteristic curve (AUROC) of 0.89 for significant fibrosis (F≥2). In breast imaging, it differentiates benign from malignant lesions with sensitivities of 83–93% and specificities of 81–84%. Other applications include thyroid nodule malignancy detection (sensitivity 80–92%, specificity 85–90%), prostate cancer biopsy guidance (sensitivity up to 96%), and kidney allograft fibrosis evaluation (accuracy ~95%). Despite limitations like operator dependency, obesity-related failures, and artifacts from calcifications, elastography enhances diagnostic precision when integrated with conventional imaging.

Introduction

Definition and Basic Principles

Elastography is a non-invasive technique that quantifies the mechanical properties of soft tissues, particularly and , by applying controlled deformations and the resulting tissue responses to map spatial variations in ity. This approach visualizes contrasts between healthy and pathological tissues, where alterations in elastic properties can signal . Unlike conventional modalities that primarily depict anatomical structures or physiological functions, elastography focuses on tissue to provide complementary diagnostic . At its core, elastography relies on fundamental mechanical principles. Elasticity describes a material's ability to undergo reversible deformation and return to its original shape, governed by : \sigma = E \epsilon, where \sigma is (force per area, in pascals), \epsilon is (relative deformation, \Delta L / L), and E is , quantifying longitudinal stiffness. , in contrast, characterizes time-dependent resistance to deformation, often modeled alongside elasticity in viscoelastic tissues using frameworks like the Kelvin-Voigt model. is typically measured in kilopascals (kPa) or megapascals (), with elastography techniques producing either qualitative color-coded maps of relative stiffness or quantitative absolute values for precise assessment. Tissue stiffness serves as a key biomarker for pathology because diseased states, such as tumors, often exhibit increased rigidity compared to surrounding healthy , attributable to factors like elevated density, , and extracellular matrix remodeling. For instance, invasive breast carcinomas can display Young's moduli of approximately 490 kPa under compression, far exceeding the 20 kPa typical of . Elastography is adaptable to platforms including , , and , enabling broad clinical integration.

Historical Background

The concept of elastography emerged from early investigations into tissue mechanics using ultrasound, with foundational work in the 1950s at Wright-Patterson Air Force Base demonstrating that the speed of sound in biological tissues varies under compression, laying groundwork for later elasticity imaging techniques. Building on these principles, researchers in the 1980s began exploring strain imaging, culminating in the seminal 1991 paper by Ophir et al., which introduced "elastography" as a quantitative method to image tissue elasticity by measuring axial strain from ultrasound echoes before and after compression, marking the birth of quasistatic ultrasound strain elastography. In the , quasistatic elastography advanced through refinements in strain estimation and display, enabling visualization of relative tissue stiffness differences in real-time. A parallel milestone occurred in magnetic resonance elastography (MRE), with Muthupillai et al. demonstrating in 1995 a technique to directly visualize propagating acoustic in tissues using phase-contrast MRI synchronized to mechanical vibrations, allowing quantitative mapping of shear stiffness. The 2000s saw rapid expansion and commercialization of elastography modalities. Transient elastography was introduced in 2003 via the FibroScan device by Sandrin et al., which uses a single ultrasound transducer to generate and detect low-frequency shear waves for liver stiffness assessment, facilitating non-invasive fibrosis evaluation. That same year, Nightingale et al. developed acoustic radiation force impulse (ARFI) imaging, employing focused ultrasound pulses to induce localized tissue displacement and quantify stiffness through shear wave propagation analysis. In 2004, Bercoff et al. pioneered supersonic shear imaging (SSI), a method generating plane shear waves via sequential acoustic pushes to map 2D elasticity in real time, enhancing spatial resolution for clinical applications. During the 2010s, elastography gained regulatory and clinical traction, with the FDA approving FibroScan for liver stiffness measurement in 2013, followed by clearances for ARFI and 2D shear wave systems on various platforms. Integration into guidelines accelerated adoption; for instance, the 2015 EASL-ALEH Guidelines recommended non-invasive elastography as a first-line tool for staging liver fibrosis in chronic liver diseases, reducing reliance on . Post-2020 developments have incorporated to improve elastography processing, with AI algorithms enhancing strain quantification, reducing operator variability, and boosting diagnostic accuracy for conditions like tumors and liver , as evidenced in meta-analyses showing superior sensitivity over traditional methods. Concurrently, prototypes for wearable elastography have emerged, such as the 2025 mechano-acoustic sensing patch by Liu et al., which enables wireless, ambulatory monitoring of tissue stiffness through conformal skin contact and automated shear wave analysis.

Fundamental Principles

Inducing Tissue Deformation

In elastography, inducing tissue deformation is the initial step that applies to , enabling the subsequent assessment of their elastic properties, such as stiffness related to . This deformation mimics physiological loading conditions and must be controlled to ensure reproducibility and safety across various modalities. Deformation types primarily include compressive (axial strain), shear (lateral waves), and dynamic (vibrations). Compressive deformation involves applying normal stress along the axis of loading, resulting in axial strain that is particularly useful for superficial tissues. Shear deformation generates transverse waves through lateral forces, allowing evaluation of shear modulus in viscoelastic materials. Dynamic deformation employs harmonic vibrations to propagate waves throughout the tissue volume, facilitating three-dimensional stiffness mapping. External methods rely on forces applied from outside the body. Manual compression, often freehand, uses pressure to induce quasi-static , suitable for operator-controlled assessments. Physiological motion harnesses natural body movements, such as or , to generate passive deformation without additional . Mechanical actuators, including electromechanical drivers or piezoelectric devices, deliver precise (typically 50–500 Hz) for dynamic excitation, as seen in systems like FibroScan for liver evaluation. Internal methods generate deformation directly within the tissue using imaging hardware. Acoustic radiation force applies pulses to create localized displacement, producing shear waves with minimal skin contact. In magnetic resonance elastography, external actuators induce vibrations, while gradients primarily encode the resulting motion rather than directly causing deformation; optical forces from lasers remain experimental and limited to superficial or applications. Key factors influencing deformation induction include , , and tissue depth penetration. Low (static or near-static) suit compressive methods for uniform , while higher (e.g., 50–500 Hz in vibrations or pulses) propagate waves but increase . determines magnitude, typically limited to micrometers (10–20 μm) to maintain . varies inversely with and tissue , restricting deep imaging in obese patients or dense organs. Safety considerations prioritize non-invasive application to prevent discomfort or damage, with strain typically below 1% in ultrasound-based methods and acoustic pulse intensities adhering to FDA guidelines (e.g., derated spatial-peak pulse-average intensity < 190 W/cm² for non-fetal tissues). These constraints ensure patient tolerability while achieving measurable deformation.

Detecting Displacement and Strain

Detecting tissue is a fundamental step in elastography, where induced deformations are quantified to infer mechanical properties. In ultrasound elastography, displacement tracking commonly employs speckle tracking, which correlates pre- and post-deformation B-mode echo patterns to estimate local tissue motion with sub-wavelength accuracy. This method leverages the random patterns formed by backscattered waves from tissue microstructures. In magnetic resonance elastography (MRE), phase-contrast MRI sequences capture cyclic displacements by encoding motion into the of the MR signal, enabling visualization of propagating shear waves with micrometer sensitivity. For optical elastography, techniques such as Doppler (OCT) detect shifts in the fringes caused by tissue motion, akin to Doppler but at optical wavelengths, providing high-fidelity tracking of nanoscale displacements. Once displacements are tracked, is calculated to represent local deformation. Axial , defined as \varepsilon = \Delta L / L where \Delta L is the change in length and L is the original length, is derived by differentiating axial displacements along the direction, often using least-squares estimation on small windows of data. Shear , arising from wave propagation, is computed from the of transverse displacements, capturing off-axis deformations in viscoelastic . For more complex geometries, finite element methods solve the by minimizing the difference between measured and modeled displacements, yielding fields that account for boundary conditions and heterogeneity. Quantitative metrics from these measurements include strain maps, which visualize relative as color-coded overlays, and reconstructions of the elasticity . For incompressible soft tissues, the Young's E approximates $3\mu, where \mu is the shear estimated from strain distributions under known . These maps provide qualitative contrast for lesion detection, while modulus values offer absolute quantification, such as elevated E in fibrotic liver tissue exceeding 10 kPa. Challenges in detection include noise and artifacts that degrade accuracy. Signal-to-noise ratio (SNR) in displacement estimates is critical, with low SNR leading to decorrelation errors in speckle tracking or phase wrapping in MRI; typical elastographic SNR values range from 10-30 in clinical scans, improved by multi-frame averaging. Motion artifacts from physiological sources, like cardiac or respiratory movement, are mitigated using correction algorithms such as rigid-body registration or Kalman filtering to align pre- and post-deformation images. Resolution limits constrain the detail of displacement and strain maps. Spatial resolution in ultrasound elastography is typically 1-2 mm, limited by beam width and speckle size, while optical methods achieve 1-10 \mum via high numerical aperture focusing. Temporal resolution varies from 30-60 frames per second in real-time ultrasound systems to sub-millisecond rates in high-speed OCT, enabling dynamic wave imaging, whereas MRE is constrained to 10-20 Hz due to acquisition times.

Data Processing and Image Formation

Data processing in elastography begins with filtering the raw displacement data to mitigate , such as speckle artifacts in or in , often employing directional filters to isolate relevant wave propagation signals. Subsequent steps involve inverting these filtered displacements to estimate mechanical properties like the , typically using least-squares inversion of wave equations that solve the inverse elasticity problem under assumptions of local homogeneity. This inversion reconstructs 2D or 3D elasticity maps by iteratively fitting measured displacements to biomechanical models, enabling spatial mapping of tissue stiffness across the imaging volume. Image formation integrates these processed maps into interpretable visualizations, commonly overlaying color-coded elastograms on B-mode images, where hues like red indicate softer regions and blue denote stiffer areas for intuitive qualitative assessment. Quantitative outputs include region-of-interest (ROI) measurements of stiffness in kilopascals (kPa), derived from calibrated color scales that map or wave speed to values, alongside volume renders for volumetric analysis in complex anatomies. These approaches distinguish qualitative relative stiffness patterns from quantitative metrics, with reproducibility often exceeding an intraclass correlation coefficient () of 0.8 in controlled studies. Advanced algorithms enhance accuracy in heterogeneous tissues through finite element modeling, which simulates wave propagation in discretized domains to account for spatial variations in and boundary conditions during inversion. Post-2020 developments incorporate , particularly convolutional neural networks, to reduce artifacts by learning noise patterns from training data and refining displacement estimates without explicit physical modeling. Common output formats include 2D elastograms displaying strain distributions and wave speed maps, where shear wave velocity c relates to shear modulus \mu and tissue density \rho via c = \sqrt{\frac{\mu}{\rho}} facilitating direct stiffness quantification.

Ultrasound Elastography

Strain Elastography

Strain elastography is a quasistatic ultrasound-based imaging technique that assesses tissue stiffness by measuring relative axial strain induced by external compressive forces applied to the skin surface. The method relies on gentle, cyclic compression using either a freehand ultrasound transducer or a specialized mechanical device, which deforms the underlying tissue. Strain is derived from tracking the displacement of ultrasound echoes between uncompressed and compressed states, primarily through correlation of radiofrequency signals to detect local tissue motion and subsequent strain patterns. This approach visualizes softer regions as areas of higher strain (color-coded in red or green on elastograms) and stiffer areas as lower strain (blue), providing a qualitative map of tissue elasticity. Originally developed by and colleagues in 1991, strain elastography represented a pioneering effort to quantify elasticity non-invasively by distributions in soft s under external loading. The technique has since evolved, with early implementations focusing on manual freehand prone to variability, while contemporary systems integrate automated probes that deliver consistent, controlled cycles to enhance and reduce artifacts. These advancements include real-time feedback mechanisms and metrics, such as uniformity scores, to guide operators toward optimal acquisition conditions. Key advantages of strain elastography include its simplicity, as it requires no additional hardware beyond standard equipment, absence of , and capability for real-time visualization during examination, facilitating immediate clinical decision-making. However, it is limited by high operator dependence, as inconsistent compression force and angle can lead to unreliable estimates, and it offers only semi-quantitative results without direct measurement of absolute mechanical properties like . Artifacts from out-of-plane motion or heterogeneous tissue further challenge reproducibility. In clinical practice, strain elastography employs metrics such as the strain ratio, calculated as the ratio of strain in a healthy region to that in the (values greater than 1 indicate relative ), and the strain index, which averages strain within a selected area to quantify local elasticity variations. These semi-quantitative tools aid in differentiating benign from malignant lesions, for instance in or , where stiffer tissues often correlate with , though thresholds vary by organ and require site-specific validation.

Shear Wave Elastography

Shear wave elastography () is a dynamic technique that quantifies tissue stiffness by generating and tracking within the tissue. Acoustic radiation force, produced by push pulses, induces localized that propagate laterally from the excitation point. These are then detected and their propagation speed is measured using ultrafast imaging, which acquires frames at rates exceeding 5000 Hz to capture the rapid motion. This method allows for real-time visualization of wave and derivation of local stiffness properties. The (μ), a measure of tissue elasticity, is calculated from the shear wave speed (c) using the equation μ = ρ c², where ρ is the density, typically approximated as 1000 kg/m³ for soft s. This relationship assumes isotropic, incompressible materials and provides absolute quantification of in units of kilopascals (kPa), distinguishing SWE from relative strain-based methods. For example, in healthy liver , shear wave speeds typically range from 1.0 to 1.3 m/s, corresponding to shear moduli of approximately 1 to 1.7 kPa. A prominent variant is supersonic shear imaging (SSI), which employs multiple sequential push pulses from a linear to create waves that propagate faster than the beam, enabling the construction of two-dimensional stiffness maps over extended regions. In SSI, the push foci are positioned along a line, generating converging wavefronts whose patterns are analyzed to map elasticity with high (on the order of 1 mm). This approach facilitates comprehensive of tissue heterogeneity, such as in abdominal organs. SWE offers key advantages, including absolute quantification without external and depths up to 8 cm, making it suitable for evaluating both superficial and deeper structures like the liver or muscles. Limitations include sensitivity to confounders such as or , which can attenuate wave propagation. Recent post-2020 advances incorporate for enhanced analysis, enabling better characterization of tissue by modeling frequency-dependent wave speeds and . algorithms process dispersion curves to differentiate elastic and viscous components, improving diagnostic accuracy in conditions like .

Acoustic Radiation Force Impulse Imaging

Acoustic Radiation Force Impulse (ARFI) imaging is an -based that employs short-duration acoustic pulses, typically lasting less than 1 , to generate localized displacements and quantify mechanical properties at specific points. This method uses a single diagnostic to both apply the force via a focused "pushing" beam and track the resulting motion with high-frequency tracking beams, enabling integration with conventional B-mode . The acoustic induces shear waves within a region of excitation (ROE), and the response—such as peak displacement or recovery time—is monitored using correlation-based speckle-tracking algorithms to assess . In ARFI, stiffer tissues exhibit smaller displacements and faster recovery due to the inverse relationship between displacement magnitude and , allowing for point-wise elasticity estimation. Quantitative metrics include derived from displacement profiles, with examples showing values around 4 kPa in soft background and up to 58 kPa in stiffer lesions, providing contrast for focal abnormalities. Variants may measure shear wave speed outside the ROE using time-of-flight methods, further characterizing local viscoelastic properties. Commercially, this is implemented as Virtual Touch tissue quantification on systems like the ACUSON S2000 scanner, which automates ARFI-based stiffness measurements. ARFI offers advantages such as high comparable to B-mode , targeted probing of regions of interest without external compression, and reduced operator dependency compared to strain-based methods. It excels in scenarios requiring precise, localized assessment, such as evaluating focal lesions in organs like the liver or . However, limitations include its point-sampling nature, which does not produce full 2D elasticity maps, potential depth restrictions due to , and challenges with quantitative accuracy in heterogeneous tissues.

Transient Elastography

Transient elastography is a non-invasive -based primarily used for assessing liver stiffness through the propagation of low-frequency waves generated by a specialized . The FibroScan device, developed by Echosens, employs a combined and vibrator mounted on the probe's axis, which is placed on over the right liver lobe. The vibrator induces a at a of Hz, propagating a wave through the liver tissue, while the tracks the wave's propagation velocity along a line 25–65 mm below surface, covering a cylindrical volume approximately 1 cm wide and 4 cm long. The measured is converted to liver using the [formula E](/page/Formula_E) = 3 [\rho](/page/Density) v^2, where E is the in kilopascals (kPa), \rho is the (assumed 1.0 g/cm³), and v is the in meters per second; results are reported as a value from at least 10 valid acquisitions, typically ranging from 2.5 to 75 kPa. Normal liver is below 7 kPa, while values exceeding 12.5 kPa are indicative of , with diagnostic cutoffs varying slightly by (e.g., 12.5–14.6 kPa for advanced in chronic C). This method provides a one-dimensional profile of , focusing on a fixed depth range suitable for liver evaluation. Transient elastography offers several advantages, including rapid performance in under 5 minutes, painlessness, high with intra- and interobserver agreement exceeding 95%, and that enables widespread clinical use without . It is particularly valuable for serial monitoring in patients. However, limitations include reduced reliability in obese individuals ( >30 kg/m²), where failure rates can reach 20%, as well as in cases of or severe , which may confound results; an XL probe variant addresses some issues by extending measurement depth. Validation studies demonstrate strong correlation with for staging, with area under the curve (AUROC) values of 0.95–0.97 for detecting (F4) and over 0.90 for advanced (F≥3), enabling accurate exclusion of significant disease in low-risk cases and reducing the need for invasive procedures. Like other shear wave techniques, it derives from wave speed but is limited to one-dimensional assessment without .

Magnetic Resonance Elastography

Principles of Wave Propagation in MRE

In magnetic resonance elastography (MRE), mechanical waves are generated using external drivers, such as pneumatic actuators or piezoelectric devices, which apply harmonic vibrations to the at frequencies typically ranging from 50 to 200 Hz. These drivers are positioned outside the MRI bore to avoid interference, and the induced motion is precisely synchronized with the using a to ensure accurate capture of wave propagation. This setup allows for the creation of shear waves that propagate through the , providing a basis for mapping mechanical properties noninvasively. The propagation of these in biological is modeled using the viscoelastic , which accounts for both elastic and viscous behaviors inherent to soft . In the , this is often represented by the for harmonic , assuming homogeneity: \nabla^2 \mathbf{u} + k^2 \mathbf{u} = 0, where \mathbf{u} is the and k is the complex incorporating and . exhibit , where decreases due to from , and , where speed varies with , leading to broader wavelengths at lower frequencies for better . These effects are critical for interpreting patterns, as stiffer propagate faster with less . Detection of wave propagation relies on motion-sensitive gradients integrated into the MRI sequence, which encode tissue displacements into the of the signal. gradient pulses, oscillating at the driver , induce phase shifts proportional to the local magnitude, typically on the order of micrometers (\sim 1-10 \, \mum). Multiple images are acquired across the wave (e.g., 8 offsets over one ) and processed via transformation to extract fields in three orthogonal directions. This phase-contrast approach enables high-resolution of propagating waves without external reference frames. A key advantage of MRE's wave propagation principles is the ability to achieve whole-organ 3D imaging, facilitated by volumetric MRI acquisitions that capture wave dynamics throughout the volume. Unlike surface-limited techniques, MRE offers deep penetration (up to tens of centimeters) due to the non-ionizing nature of MRI and the diffusive propagation of low-frequency shear waves, making it suitable for internal organs like the liver or . Tissue models in MRE distinguish between isotropic assumptions, where mechanical properties are uniform in (common for homogeneous organs), and anisotropic models that account for directional variations, such as in fibrous or muscular tissues. Inversion processes derive the storage modulus, the real part of the complex G^* = G' + iG'' (reported in kPa), representing the storage component from observed wave speeds and attenuations. These models enable quantitative elasticity mapping while considering viscoelastic for more accurate property estimation.

Acquisition and Inversion Methods

In magnetic resonance elastography (MRE), acquisition methods rely on specialized MRI sequences designed to capture the propagation of shear waves within tissues. The most widely adopted approach uses phase-contrast gradient-echo (GRE) sequences that incorporate motion-encoding gradients (MEGs) synchronized with the mechanical driver to encode cyclic tissue displacements into the magnetic resonance phase. These MEGs are applied in multiple directions and offsets to sample the wave field comprehensively, typically acquiring data in 3D volumes for volumetric stiffness mapping. For assessing viscoelastic properties, multi-frequency acquisitions are employed, exciting tissues at frequencies ranging from 30 to 100 Hz; in liver MRE, a standard frequency of 60 Hz is used to balance wave penetration and resolution while minimizing attenuation. Inversion methods reconstruct mechanical properties from the acquired data by solving the wave equation, often assuming local homogeneity for initial estimates. Direct inversion techniques, such as local frequency estimation (LFE), estimate the local wave number k from the spatial content of the field using bandpass filtering, yielding the via the approximation \mu \approx \rho \left( \frac{2\pi f}{k} \right)^2, where \rho is and f is the excitation . This method provides rapid computation suitable for clinical workflows and has been validated in homogeneous phantoms and tissues, producing stiffness maps with values expressed in kilopascals (kPa). Advanced inversion strategies address tissue heterogeneity through 3D nonlinear inversion (3D MRE or NLI-MRE), which iteratively solves the full viscoelastic using finite element models to account for wave reflections, mode conversions, and spatial variations in . These methods improve accuracy in complex anatomies like the or fibrotic liver, where local assumptions fail. Recent post-2020 developments include open datasets of MRE wave images from phantoms, livers, and , acquired at multiple frequencies and used to train AI-based inversion models for enhanced robustness and speed. A 2025 dataset, for instance, provides 3D multifrequency data at 60 Hz for phantom and in vivo validation, facilitating approaches to inversion. Motion artifacts, arising from cardiac or respiratory influences, are mitigated during acquisition using navigator echoes, which monitor bulk tissue motion and enable retrospective or prospective corrections to preserve wave phase integrity. Final outputs include quantitative stiffness maps in kPa, often overlaid on anatomical images, along with confidence maps derived from inversion residuals to highlight reliable regions and exclude noisy areas.

Optical Elastography

Optical Coherence Elastography

Optical coherence elastography (OCE) is an advanced technique that leverages (OCT) to quantify the biomechanical properties of soft s at the microscale by detecting minute displacements induced by mechanical loading. It employs phase-sensitive detection within OCT systems to measure nanometer-scale deformations with high precision, enabling the mapping of elasticity in three dimensions. This method extends the capabilities of OCT beyond structural to include functional of stiffness, which is crucial for understanding pathological changes in diseases where mechanical properties alter, such as or tumors. The core technique of OCE involves exciting the tissue with external or internal stimuli to generate controlled deformations, followed by OCT tracking of the resulting displacements. Common excitation methods include non-contact air-puff loading, which induces surface for corneal assessment, and magnetic actuation using nanoparticles to create localized forces in deeper layers. -resolved OCT, particularly spectral-domain variants, detects these displacements with sub-nanometer by analyzing phase shifts in the signal, achieving axial resolutions of approximately 1-3 μm and lateral s of 10-15 μm. This high resolution makes OCE particularly suitable for superficial tissues like the ocular structures (e.g., and crystalline ) and , where it has been applied to evaluate stiffness variations in conditions such as or dermal disorders. For instance, corneal elastography using air-puff excitation has quantified values ranging from 577 to 1129 kPa, providing insights into biomechanical integrity post-treatment. Mechanical properties are quantified by deriving strain fields from displacement maps and applying models such as the Kelvin-Voigt framework for viscoelastic characterization. Static OCE computes from stress-strain relations, yielding values like 23-300 kPa in hydrated layers, while dynamic OCE assesses frequency-dependent responses to capture , such as shear wave speeds differing by 8-14% in tissue phantoms mimicking healthy versus diseased states. Post-2020 advancements have integrated OCE with light-sheet for volumetric microscale of extracellular matrix remodeling, enhancing throughput and for cellular-level studies in fibroblasts. These developments underscore OCE's advantages, including non-invasive, operation (with processing times reduced to ~1 second per frame via optimized algorithms) and compatibility with clinical settings for superficial tissues. Despite these strengths, OCE is limited by the shallow optical penetration depth of near-infrared light, typically 1-2 mm in scattering tissues like , which restricts its use to superficial applications and necessitates complementary techniques for deeper . Additionally, challenges in computational efficiency and excitation uniformity can introduce variability, though ongoing refinements in phase-stability and excitation methods continue to mitigate these issues.

Brillouin Elastography and Other Optical Techniques

Brillouin elastography, also known as Brillouin microscopy, is a label-free optical technique that probes the mechanical properties of materials through inelastic scattering from thermally excited acoustic phonons in the gigahertz range. This scattering, termed Brillouin light scattering (BLS), arises from the interaction between incident and high-frequency , resulting in a frequency shift that encodes information about the material's . The Brillouin frequency shift \Delta \nu is related to the v by the equation \Delta \nu = \frac{2 n v}{\lambda} \sin(\theta/2), where n is the , \lambda is the of the incident , and \theta is the scattering angle; from this, the longitudinal modulus M can be derived as M = \rho v^2, with \rho being the material . This approach enables non-contact, non-destructive mapping of viscoelastic properties at the micron scale, distinguishing it from excitation-based methods by relying on intrinsic . Key advantages of Brillouin elastography include its label-free nature, which avoids artifacts from exogenous agents, and its high , capable of acquiring mechanical images in seconds to minutes for cellular samples. Recent advances from 2022 to 2025 have focused on improving acquisition speed and signal-to-noise ratios through stimulated Brillouin scattering and optimized Fabry-Pérot interferometers, enabling sub-millisecond spectral acquisition for dynamic cellular studies. For instance, innovations in high-peak-power and have reduced light exposure while mapping elasticity in live cells, facilitating investigations into biomechanical responses during processes like activation. However, limitations persist, including the complexity of setups requiring high-resolution spectrometers and stable sources, as well as high to levels, where water content can dominate the measured over intrinsic . This hydration dependence arises because the longitudinal is acutely responsive to fractions, potentially interpretations in biological tissues. Beyond Brillouin methods, other optical techniques in elastography include photoacoustic elastography, which leverages hybrid opto-acoustic wave generation to assess stiffness. In this approach, illumination induces thermoelastic expansion, producing ultrasonic waves whose propagation and attenuation reveal viscoelastic parameters, often combined with for depth-resolved mapping up to several millimeters. Optical microrheology complements these by employing particle tracking to quantify local ; inert tracer particles embedded in the sample undergo , analyzed via video or to compute the complex G^*(\omega) from mean-squared displacements. Passive variants rely on , while active modes apply controlled forces, offering nanoscale sensitivity for intracellular environments without external mechanical excitation. These techniques find promising applications in biomedical research, particularly for elucidating mechanics in the , where Brillouin has revealed stiffness gradients in cancer spheroids that correlate with metastatic potential. For example, longitudinal tracking of tumor models has shown evolving elasticity during growth, highlighting heterogeneous regions influenced by remodeling.

Clinical Applications

Hepatic Disease Evaluation

Elastography has emerged as a cornerstone non-invasive tool for evaluating hepatic diseases, including , , and , by quantifying liver stiffness and fat content to stage disease severity and guide management. In chronic liver conditions such as non-alcoholic fatty liver disease (NAFLD) and , it provides quantitative metrics that correlate with histopathological findings, enabling serial monitoring without the risks of . Transient elastography () serves as the primary technique due to its accessibility and widespread validation, while shear wave elastography (SWE) and magnetic resonance elastography (MRE) offer advanced options for deeper tissue assessment in complex cases. For fibrosis staging using the METAVIR system, measures liver in kilopascals (kPa), with values below 7 kPa typically indicating no or mild (F0-F1), while greater than 7 kPa suggests significant (F2 or higher); (F4) is often indicated by readings exceeding 12 kPa. MRE, which excels in whole-liver coverage, reports normal values of 2.0-2.5 kPa, with thresholds of 3.0-4.0 kPa signaling advanced and over 5.0 kPa indicating . In steatosis assessment, incorporates the controlled attenuation parameter (CAP), measured in decibels per meter (dB/m), where values ≥248 dB/m indicate (S≥1), escalating to ≥290 dB/m for severe (S3). These metrics align with guidelines from the Society of Radiologists in Ultrasound (SRU), American Association for the Study of Liver Diseases (AASLD), and European Association for the Study of the Liver (EASL), which endorse for initial screening and longitudinal follow-up in at-risk populations. Meta-analyses confirm elastography's diagnostic performance against , with TE achieving an area under the curve (AUROC) of 0.84 for significant , 0.89 for severe , and 0.94 for , alongside sensitivities and specificities of 85-90% at optimized cutoffs. MRE demonstrates superior accuracy for advanced (summary AUROC 0.92) and (0.90), particularly in obese patients where ultrasound-based methods may falter. Post-2020 advancements include integration with (CEUS) to assess microvascular vascularity in fibrotic livers, enhancing prognostic evaluation in . As a non-invasive alternative to , elastography reduces procedural complications and costs, supporting risk stratification in guidelines for NAFLD and chronic hepatitis. Despite these strengths, limitations persist, as acute , (>5 times upper limit of normal), or hepatic congestion can falsely elevate stiffness readings, confounding assessment independent of or necro. Optimal results require standardized protocols, such as and quality metrics (e.g., / ≤30% for ), to mitigate variability across etiologies.

Oncologic Imaging

Elastography plays a significant role in by assessing tissue stiffness to aid in tumor detection and characterization across various organs. In evaluation, strain elastography and are integrated with the to differentiate benign from malignant lesions. Lesions exhibiting stiffness greater than 50 kPa on SWE are considered suspicious for malignancy, with studies reporting a of approximately 80% for this threshold when combined with conventional . This approach enhances specificity, particularly for BI-RADS 4 lesions, by identifying softer benign masses that may avoid . Post-2020 advancements, including refined SWE protocols, have positioned elastography as a valuable tool for early , as highlighted in a 2025 review emphasizing its role in improving detection rates in dense breasts. For thyroid nodules, acoustic radiation force impulse (ARFI) elastography quantifies to stratify risk, with values exceeding 30 kPa indicating higher suspicion for malignant lesions. This cutoff helps in triaging indeterminate nodules, where stiffer tissues correlate with papillary carcinoma, achieving diagnostic accuracies that support decisions. In , SWE and strain elastography detect stiff regions indicative of tumors, often combined with multiparametric MRI; for instance, SWE identifies lesions with elevated wave velocities greater than 3.5 m/s as potentially malignant, improving targeted yields. Other applications include and ovarian cancers, where elastography assesses stromal to characterize tumors, though evidence remains emerging compared to and uses. Clinical evidence demonstrates that incorporating elastography reduces unnecessary biopsies by 20-30% in without compromising cancer detection, particularly when categories are reclassified based on stiffness metrics. Combining elastography with further boosts diagnostic performance, with exceeding 85% for differentiating benign and malignant breast masses by evaluating both mechanical and vascular properties. Despite these benefits, challenges persist, such as false positives from benign stiff lesions like fibroadenomas or sclerosing adenosis in the breast, which can mimic and necessitate confirmation. Operator dependency and variability in tissue heterogeneity also limit in some cases.

Musculoskeletal and Cardiovascular Assessment

In musculoskeletal applications, shear wave (SWE) provides quantitative evaluation of tendon elasticity, enabling differentiation between healthy and pathological states. For the , normal stiffness values in asymptomatic non-athletic individuals average 103.6 ± 30.5 kPa, while athletes exhibit higher means of 183.8 ± 98 kPa due to adaptive changes from loading. In or post-injury scenarios, SWE detects reduced elasticity, with ruptured tendons showing values as low as 72 kPa compared to healthy ranges of 261–300 kPa. This technique supports post-injury monitoring by tracking stiffness recovery during ; for instance, surgically repaired Achilles tendons demonstrate progressive increases in , reaching 187.7 ± 23.8 kPa by 12 weeks postoperatively, correlating with improved mechanical integrity. Cardiovascular assessment leverages magnetic resonance elastography (MRE) to measure myocardial , a key indicator of in non-ischemic conditions. In healthy subjects, diastolic myocardial typically ranges from 4.5 to 5.1 kPa, whereas fibrotic myocardium in exceeds 12 kPa, with values up to 21.8 kPa in advanced cases. Shear wave elastography complements this by estimating arterial (PWV) to gauge vascular ; in the , PWV correlates directly with shear wave speed, where elevated values above 5 m/s signal increased cardiovascular risk in normotensive populations. Emerging evidence from 2025 highlights wearable mechano-acoustic sensors for real-time elastography, offering ambulatory monitoring of tissue through surface wave dispersion analysis without external actuators. These devices enable continuous, depth-resolved during daily activities, enhancing functional in degenerative conditions. A primary advantage of elastography in musculoskeletal and cardiovascular contexts is dynamic assessment during motion, allowing real-time quantification of and muscle elasticity under or movement, which improves diagnostic accuracy for functional impairments. Despite this, motion artifacts from or cardiac pulsation pose limitations, often requiring breath-holding or gating techniques to minimize signal degradation in both and MRE applications.

Limitations and Future Directions

Technical and Clinical Challenges

One major technical challenge in -based elastography modalities, such as (), is the limited depth penetration, typically restricted to less than 10 cm due to signal in , which hinders reliable of deeper structures like the liver in patients with or . in structure further complicates measurements by violating the isotropic assumption underlying many elastography algorithms, leading to variations in shear wave speed estimates at interfaces and potentially inaccurate quantification. variability introduces additional inconsistency, particularly in strain elastography where manual compression is subjective; while () targets below 10% are desirable for reproducibility, actual intra-operator can exceed this in heterogeneous s due to differences in and positioning. In magnetic resonance elastography (MRE), technical hurdles include trade-offs between and wave , as higher-frequency shear waves enhance resolution but decay rapidly, limiting applicability in stiff s like bone. Clinically, elastography measurements are confounded by physiological factors such as and , which elevate apparent independently of , thereby reducing specificity in acute conditions. Validation against gold standards like remains essential, yet discrepancies arise because elastography primarily captures composite rather than isolated pathological changes, complicating direct correlations. Reproducibility across systems poses a significant barrier, with inter-vendor differences in SWE platforms—such as those from and —resulting in up to 20-30% variability in liver stiffness measurements due to proprietary algorithms and hardware differences. Efforts toward , including the WFUMB guidelines, emphasize consistent protocols for acquisition and to mitigate these discrepancies, recommending multiple measurements and quality criteria like CV <10% for reliable results. Safety profiles are generally favorable, with low risk from non-ionizing ultrasound, but cumulative exposure from repeated scans raises concerns for thermal effects, particularly in vulnerable populations, prompting adherence to ALARA principles.

Emerging Technologies and Advances

Recent advancements in (AI) and machine learning (ML) have significantly enhanced elastography by enabling automated segmentation of tissue regions and improved dispersion imaging to assess alongside . For instance, AI-assisted elastography models have demonstrated high efficiency in classifying lesions, achieving a pooled of 90% and specificity of 88% in a 2025 meta-analysis of clinical studies using large datasets. In magnetic resonance elastography (MRE), frameworks like UNet-ResNet50 have automated liver segmentation from 2024-2025 imaging datasets, reducing processing time while maintaining diagnostic accuracy comparable to expert radiologists. These AI integrations address previous limitations in by providing real-time mapping through wave dispersion analysis, as evidenced in recent studies. Regulatory bodies like the FDA have issued guidance on AI/ML-enabled medical devices, emphasizing validation for elastography applications to ensure safety and efficacy. Hybrid imaging techniques combining elastography with other modalities have emerged to provide multimodal insights into tissue and mechanics. (CEUS) integrated with elastography allows simultaneous evaluation of vascular and , improving of focal liver lesions by correlating dynamic patterns with mechanical properties in post-2020 studies. Photoacoustic-optical hybrids further advance this by merging optical with elastographic , enabling high-resolution of microvascular and biomechanical changes in tumors, as demonstrated in interleaved photoacoustic-ultrasound systems achieving sub-second rates. These hybrids, such as spherical array optoacoustic-ultrasound setups, offer enhanced specificity for applications without invasive procedures. Wearable devices represent a major leap toward continuous, non-invasive monitoring, particularly with mechano-acoustic sensors introduced in 2025 for . These sensors generate and detect surface to perform elastography on skin-applied patches, enabling real-time cardiac monitoring of myocardial changes during daily activities, with demonstrated accuracy in detecting variations. Bioadhesive shear wave elastography wearables, developed in 2024, use acoustic radiation force to induce shear waves for portable liver and muscle evaluations, overcoming traditional scanner limitations for longitudinal studies. Such technologies facilitate home-based tracking of disease progression in cardiovascular conditions. Progress in three-dimensional () and four-dimensional () elastography has enabled volumetric and dynamic assessments critical for complex tissues. Volumetric MRE techniques now produce full stiffness maps of organs like the liver, incorporating for rapid inversion of wave propagation data to quantify heterogeneity in staging. optical coherence elastography (OCE) has advanced to phase velocity mapping, capturing spatiotemporal elasticity variations in heterogeneous biological tissues at high , as shown in 2020-2025 developments using asynchronous acquisition methods. These capabilities, including deep learning-enhanced OCE at 831 Hz volumes, support intraoperative guidance and personalized research. Looking ahead, portable elastography devices are poised for widespread adoption, driven by ongoing regulatory approvals and market expansion. Compact and MRE systems, cleared by bodies like the FDA for point-of-care use, promise broader accessibility in remote settings. The global elastography imaging market is projected to reach approximately $7.5 billion by 2032, fueled by these innovations and increasing demand for non-invasive diagnostics in chronic diseases.

References

  1. [1]
    Elastography: general principles and clincial applications - PMC
    The general principle of elastography can be summarized as follows: (1) perturb the tissue using a quasi-static, harmonic, or transient mechanical source.
  2. [2]
    Ultrasound Elastography: Review of Techniques and Clinical ...
    Here, we review the basic principles, foundation physics, and limitations of ultrasound elastography and summarize its current clinical use and ongoing ...
  3. [3]
    Ultrasound Elastography and MR Elastography for Assessing Liver ...
    Imaging-based elastography is an emerging technology that uses imaging to noninvasively assess mechanical tissue properties. Elastography techniques have ...
  4. [4]
    Ultrasonic surface acoustic wave elastography: A review of basic ...
    Apr 10, 2024 · Elastography, or elasticity imaging, is an imaging strategy that estimates tissue's mechanical properties, namely elastic modulus and/or ...
  5. [5]
    Biophysical Bases of Elasticity Imaging - SpringerLink
    Elasticity imaging is based on two processes. The first is the evaluation of the mechanical response of a stressed tissue using imaging modalities.
  6. [6]
    Introduction to ultrasound elastography - PMC - NIH
    Young's modulus, E = (stress)/(strain). Young's modulus E describes the longitudinal elasticity of a material (tissue) ΔL/L (the percentage change in its length) ...
  7. [7]
    Imaging the microscale mechanical properties of cancer using ...
    ... tumours are often markedly stiffer than surrounding healthy tissue. Over the past 20 years, studies have revealed the complex, scale-dependent biomechanics ...
  8. [8]
    [PDF] Elastography imaging: the 30 year perspective
    Dec 21, 2020 · Thus, a medical imaging approach—now generally termed elastography—was developed for objectively assessing the biomechanical properties of ...
  9. [9]
    a quantitative method for imaging the elasticity of biological tissues
    Elastography: a quantitative method for imaging the elasticity of biological tissues. Ultrason Imaging. 1991 Apr;13(2):111-34. ... J Ophir , I Céspedes, H ...
  10. [10]
    Elastography: history, principles, and technique comparison
    Aug 6, 2025 · 2 The first recognized strain elastography technique was innovated and implemented in 1991 by Ophir et al, from University of Texas Medical ...
  11. [11]
    Magnetic resonance elastography by direct visualization ... - PubMed
    A nuclear magnetic resonance imaging (MRI) method is presented for quantitatively mapping the physical response of a material to harmonic mechanical excitation.
  12. [12]
    a new noninvasive method for assessment of hepatic fibrosis
    This device is based on one-dimensional (1-D) transient elastography, a technique that uses both ultrasound (US) (5 MHz) and low-frequency (50 Hz) elastic waves ...Missing: introduction | Show results with:introduction
  13. [13]
    Shear-wave generation using acoustic radiation force: in vivo and ex ...
    Acoustic radiation force impulse (ARFI) imaging involves the mechanical excitation of tissue using localized, impulsive radiation force.Missing: elastography | Show results with:elastography
  14. [14]
    Supersonic shear imaging: a new technique for soft tissue elasticity ...
    Supersonic shear imaging (SSI) is a new ultrasound-based technique for real-time visualization of soft tissue viscoelastic properties.
  15. [15]
    [PDF] EASL-ALEH Clinical Practice Guidelines: Non-invasive tests for ...
    In order to optimize the value of liver biopsy for fibrosis evaluation, it is important to adhere to the following recommendations: (i) sample length >15 mm by ...<|separator|>
  16. [16]
  17. [17]
    Wireless, wearable elastography via mechano-acoustic wave ...
    This study introduces a mechano-acoustic wave sensing technology for automated, wireless elastography. The patch-form sensor maintains conformal ...Missing: prototypes | Show results with:prototypes
  18. [18]
  19. [19]
  20. [20]
  21. [21]
    Magnetic resonance elastography | Nature Medicine
    May 1, 1996 · Muthupillai, R. et al. Magnetic resonance elastography by direct visualization of acoustic strain waves. Science 269, 1854–1857 (1995). Article ...
  22. [22]
    Tissue Doppler optical coherence elastography for real time strain ...
    Oct 4, 2006 · The tDOCE system is capable of displaying the strain rates and strain maps of tissue subjected to a dynamic compression in real time. The system ...
  23. [23]
    A Quantitative Method for Imaging the Elasticity of Biological Tissues
    We describe a new method for quantitative imaging of strain and elastic modulus distributions in soft tissues. The method is based on external tissue ...
  24. [24]
    A finite element model for analyzing shear wave propagation ...
    We propose finite element (FE) modeling as a robust method to systematically study the effects of these parameters. An axisymmetric FE model was generated with ...
  25. [25]
    Ultrasound elastography: Principles and techniques - ScienceDirect
    In this paper, we introduce the principles of elastography and give a technical summary for the main elastography techniques: from quasi-static methods that ...Missing: review | Show results with:review
  26. [26]
    Estimating the elastographic signal-to-noise ratio using correlation ...
    Elastography, a technique of estimating strain using differential displacements of the tissue elements has been well established Ophir et al 1991, Ophir et al ...Missing: artifacts | Show results with:artifacts
  27. [27]
    A Novel Motion Compensation Algorithm for Acoustic Radiation ...
    A novel method of physiological motion compensation for use with radiation force elasticity imaging has been developed.
  28. [28]
    Application of Elastography for the Noninvasive Assessment of ...
    Elastography is a specialized imaging-based method available to spatially map strain fields and elastic properties, including stiffness or shear modulus, in ...<|control11|><|separator|>
  29. [29]
    Ultra-wideband optical coherence elastography from acoustic to ...
    Aug 16, 2023 · However, traditional elastography techniques have been limited by a maximum wave frequency below 1-10 kHz, which hampers temporal and spatial ...
  30. [30]
    Model-based elastography: a survey of approaches to the inverse ...
    This elastographic imaging technique uses radiation force to generate a shear wave of known spatial frequency, and then measures the temporal frequency response ...
  31. [31]
  32. [32]
    Quality Assurance of Point and 2D Shear Wave Elastography ... - MDPI
    Jul 31, 2024 · This study aims to establish baseline data using phantoms and monitor them for quality assurance in elastography.1. Introduction · 2. Materials And Methods · 3. Results
  33. [33]
  34. [34]
    Deep learning in ultrasound elastography imaging: A review - Li
    Jul 17, 2022 · In recent years, deep learning has begun to emerge in ultrasound elastography research. In this review, several common deep learning frameworks ...
  35. [35]
  36. [36]
    Ultrasound elastography: Advantages, limitations and artefacts of ...
    Ultrasound elastography is a technique currently under development. Its use in clinical practice is complicated because of the wide range of techniques used.Missing: metrics | Show results with:metrics
  37. [37]
    Ultrasound Elastography: Review of Techniques and Clinical ...
    Ultrasound elastography (USE) is an imaging technology sensitive to tissue stiffness that was first described in the 1990s.
  38. [38]
    Strain Elastography Ultrasound: An Overview with Emphasis ... - MDPI
    Feb 25, 2013 · This article presents an overview of strain elastography (SE) and its applications in breast cancer diagnostics.<|control11|><|separator|>
  39. [39]
    Supersonic shear imaging: a new technique for soft tissue elasticity ...
    Supersonic shear imaging (SSI) is a new ultrasound-based technique for real-time visualization of soft tissue viscoelastic properties.
  40. [40]
    Shear-Wave Elastography: Basic Physics and Musculoskeletal ...
    Shear wave elastography is a rapidly evolving US imaging technique that allows quantification of mechanical and elastic tissue properties.Missing: seminal | Show results with:seminal
  41. [41]
    Reverberant Shear Wave Elastography: A Multi-Modal ... - Frontiers
    Jan 13, 2021 · As observed, one advantage of using RSWE is that it produces stronger shear waves at deeper tissue regions of interest, and it can provide ...Abstract · Introduction · Applied Methodology in RSWE · Discussion
  42. [42]
    Artificial intelligence - based ultrasound elastography for disease ...
    Jun 2, 2023 · This review provides an overview of the basic concepts of USE and AI techniques for clinical radiologists and then introduces the applications of AI in USE ...Missing: post- | Show results with:post-
  43. [43]
    Shear-Wave Elastography and Viscosity PLUS for the Assessment ...
    Sep 2, 2022 · Various methods have been developed to assess the tissue viscoelastic properties by measuring the shear-wave dispersion and attenuation, ...
  44. [44]
    Acoustic Radiation Force Impulse (ARFI) Imaging: a Review - PMC
    Nightingale is a co-inventor of the ARFI imaging technology, and Duke University holds associated intellectual property rights. ARFI imaging technology has been ...
  45. [45]
  46. [46]
    Transient elastography (FibroScan®) with controlled attenuation ...
    Transient elastography (TE) with controlled attenuation parameter (CAP) is a fast, reliable, repeatable non-invasive method for the assessment of liver ...
  47. [47]
    MAGNETIC RESONANCE ELASTOGRAPHY: A REVIEW - PMC
    MRE obtains information about the stiffness of tissue by assessing the propagation of mechanical waves through the tissue with a special magnetic resonance ...
  48. [48]
    MR elastography: Principles, guidelines, and terminology - Manduca
    Dec 9, 2020 · Magnetic resonance elastography (MRE) is a phase contrast–based MRI technique that can measure displacement due to propagating mechanical ...Missing: seminal | Show results with:seminal
  49. [49]
    Magnetic Resonance Elastography by Direct Visualization ... - Science
    Muthupillai et al. ,. Magnetic Resonance Elastography by Direct Visualization of Propagating Acoustic Strain Waves.Science269,1854-1857(1995).DOI:10.1126 ...
  50. [50]
    Magnetic resonance elastography: basic principles, technique, and ...
    Elastography is a method of assessing the mechanical properties of tissues and can be performed with ultrasonography (US) and magnetic resonance imaging (MRI) ...<|separator|>
  51. [51]
    MR Elastography of the liver: qualitative and quantitative ... - NIH
    The most common pulse sequence used for liver MRE is based on Gradient Recalled Echo (GRE) (11). In this sequence, the MEGs are usually applied along the z-axis ...
  52. [52]
    Magnetic resonance elastography of the liver: everything you need ...
    For clinical liver MRE, stiffness thresholds have been established at 60 Hz, and the frequency should not be changed, as an increase (decrease) in frequency ...
  53. [53]
    Magnetic resonance elastography: non-invasive mapping of tissue ...
    Magnetic resonance elastography (MRE) is an MRI technique that measures tissue elasticity by visualizing acoustic strain waves, calculating shear modulus.Missing: frequency | Show results with:frequency
  54. [54]
    Mechanical Properties of Viscoelastic Media by Local Frequency ...
    Local frequency estimation (LFE) [21] has been used to locally extract the wavenumber κ = | k ¯ | from displacement wave fields in order to estimate the elastic ...Missing: formula | Show results with:formula
  55. [55]
    Phantom Evaluations of Nonlinear Inversion MR Elastography - PMC
    This study evaluated non-linear inversion MRE (NLI-MRE) based on viscoelastic governing equations to determine its sensitivity to small, low contrast inclusions ...
  56. [56]
    MR elastography datasets including phantom, liver, and brain - Nature
    Oct 23, 2025 · Magnetic resonance elastography (MRE) is a noninvasive way to measure biomechanical properties of soft biological tissues. It works by inducing ...Background & Summary · Methods · Data RecordsMissing: deformation | Show results with:deformation
  57. [57]
    Magnetic Resonance Elastography of the Brain using Multi-Shot ...
    Magnetic Resonance Elastography of the Brain using Multi-Shot Spiral ... It also provides a navigator for motion correction as discussed in the following section.
  58. [58]
    Comparison of inversion methods in MR elastography: An open ...
    Jun 12, 2022 · We introduce an expandable, open-access, webserver-based platform that offers multiple inversion techniques for multifrequency, 3D MRE data.
  59. [59]
  60. [60]
  61. [61]
    Brillouin Light Scattering: Applications in Biomedical Sciences
    May 1, 2019 · They are based on Brillouin light scattering from acoustic waves or phonons in the GHz range, providing a nondestructive contactless probe of ...
  62. [62]
    Consensus statement on Brillouin light scattering microscopy of ...
    Jul 3, 2025 · Brillouin light scattering (BLS) spectroscopy is a non-invasive, non-contact, label-free optical technique that can provide information on the mechanical ...
  63. [63]
    Stimulated Brillouin scattering microscopy with a high-peak-power ...
    Jul 10, 2025 · Brillouin microscopy has recently emerged as a promising technique for all-optical, non-invasive mechanical measurements of cells and tissues ...
  64. [64]
    Brillouin microscopy analysis of the fibroblast mechanical response ...
    May 13, 2025 · In this study, we non-invasively imaged the elastic properties of monolayer 4T1 murine fibroblast cells using Brillouin microscopy and analyzed their response ...
  65. [65]
    Brillouin scattering spectroscopy for studying human anatomy
    Brillouin light scattering (BLS) spectroscopy is a label-free method of measuring the GHz-frequency viscoelastic properties. The measured longitudinal modulus ...Missing: speed | Show results with:speed
  66. [66]
    Brillouin Microscopy: An Emerging Tool for Biomechanical Analysis ...
    Jul 1, 2025 · Brillouin microscopy has greatly advanced ocular biomechanical assessment, enabling precise in vivo diagnosis of keratoconus and non-destructive ...
  67. [67]
    Photoacoustic elastography imaging: a review - PMC
    This article gives a review of reported studies on PA imaging modalities for recovery, noninvasively and nondestructively, of elastic properties of soft ...
  68. [68]
    Photoacoustic Multispectral Elastography Based on the ...
    Aug 14, 2023 · Photoacoustic tomography is a hybrid imaging modality that combines the rich contrast of optical imaging and the high spatial resolution of ...
  69. [69]
    Passive and Active Microrheology for Biomedical Systems - Frontiers
    Microrheology encompasses a collection of techniques that fall into two broad categories, active and passive, defined by the force exerted on probe particles.Introduction · Microrheological Methodology · Microrheological Approaches...<|separator|>
  70. [70]
    Optical Tweezers Microrheology: From the Basics to Advanced ...
    Aug 5, 2018 · Optical tweezers offer a powerful platform for performing microrheology measurements and can measure rheological properties at the level of single molecules ...
  71. [71]
    Brillouin microscopy in cancer research: a review - PMC
    Oct 26, 2025 · The last two categories—the tumor microenvironment and its inherent heterogeneity—are closely intertwined and should be considered together when ...
  72. [72]
    Mechanical Evolution of Metastatic Cancer Cells in 3D ...
    Mar 21, 2025 · In this work, optical Brillouin microscopy is utilized to longitudinally acquire mechanical images of growing cancerous spheroids over the period of 8 days.
  73. [73]
  74. [74]
    Noninvasive Liver Disease Assessment - AASLD
    AASLD Practice Guideline on blood-based noninvasive liver disease assessments of hepatic fibrosis and steatosisMissing: EASL | Show results with:EASL
  75. [75]
    FibroScan and transient elastography - RACGP
    FibroScan is a non-invasive device that assesses the 'hardness' (or stiffness) of the liver via the technique of transient elastography.
  76. [76]
  77. [77]
    Normative values for magnetic resonance elastography-based liver ...
    May 31, 2019 · For women, the mean (SD) stiffness value was 2.14 (0.30) kPa (range, 1.37-2.66 kPa), and for men, 2.14 (0.25) kPa (range, 1.54-2.54 kPa).
  78. [78]
    Understanding Your Liver Elastography (FibroScan®) Results
    FibroScan is a special ultrasound technology that measures liver stiffness (hardness) and fatty changes in your liver.
  79. [79]
    Diagnostic accuracy of elastography and magnetic resonance ...
    May 13, 2021 · Conclusions: When elastography index tests are acquired successfully, they have acceptable diagnostic accuracy for advanced fibrosis and ...
  80. [80]
    Expanding Role of Contrast-Enhanced Ultrasound and ...
    Jul 1, 2025 · Beyond lesion characterization, CEUS has proved highly effective in evaluating the vascular architecture of liver lesions, an essential aspect ...
  81. [81]
    Quantitative shear wave ultrasound elastography: initial experience ...
    Dec 1, 2010 · A mean stiffness of over 50 kPa is highly suggestive of malignancy. Maximum and stiffness SD values show diagnostic accuracy and reproducibility ...
  82. [82]
    Diagnostic accuracy of shear wave elastography for prediction ... - NIH
    Jan 22, 2016 · The sensitivity of predicting malignancy in patients with breast lesions ranges from 80% to 98%, while the specificity ranges from 66% to 84%.
  83. [83]
    Ultrasound elastography: advances and challenges in early ...
    This review explores recent advances in ultrasound elastography for breast cancer detection, focusing on technological innovations, clinical validation, and ...
  84. [84]
    Evaluation of standard breast ultrasonography by adding two ...
    Aug 30, 2023 · After the addition of 2D SWE only for reclassification, reduction of benign biopsies in BI-RADS 4–5 lesions was 20~30% using an independent ...
  85. [85]
    Diagnostic performance of contrast-enhanced ultrasound combined ...
    Nov 24, 2023 · The combination of contrast-enhanced ultrasound (CEUS) and shear wave elastography (SWE) has high value in distinguishing between benign and malignant breast ...
  86. [86]
    Strain Versus 2D Shear-Wave Elastography Parameters—Which ...
    Nov 3, 2022 · ... 30 kPa that is easier to use ... Shear wave elastography versus real-time elastography on evaluation thyroid nodules: A preliminary study.
  87. [87]
    Shear Wave Elastography (SWE) of Asymptomatic Achilles Tendons
    Jan 14, 2019 · Results: Mean SWE-value for Achilles tendon was 183.8 kPa (± 98 kPa) in athletes and 103.6 kPa (± 30.5 kPa) in the nonathletic control group.
  88. [88]
    Current status of musculoskeletal application of shear wave ... - NIH
    When we use the formula of the Young modulus (E=3 μ=3 ρ cT2), ρ is the tissue density and it is assumed to be constant at 1,000 kg/m3, under the assumption of a ...
  89. [89]
    Evaluation of Elastic Stiffness in Healing Achilles Tendon After ... - NIH
    The mean values of elastic modulus of the repaired tendons were significantly different among the 3 time points postoperatively (12 weeks, 187.7±23.8 kPa, range ...<|separator|>
  90. [90]
    Myocardial Stiffness Evaluation Using Noninvasive Shear Wave ...
    Mar 14, 2018 · MS was significantly higher in HCM-HFpEF patients than in HV patients (mean MS = 12.68 ± 2.91 kPa vs. 4.47 ± 1.68 kPa, respectively; p < 0.01), ...
  91. [91]
    Cardiac magnetic resonance elastography imaging in hypertrophic ...
    Cardiac magnetic resonance elastography (MRE) showed increased LV stiffness (21.8 kPa; reference range, 7.2–9.8 kPa; Panel F). Notably, the locations of ...
  92. [92]
    Arterial Stiffness Estimation by Shear Wave Elastography
    The pulse wave velocity (PWV) technique is commonly used to estimate global arterial stiffness by measuring the velocity of the pulse wave generated by the ...
  93. [93]
    Arterial Stiffness Assessment by Shear Wave Elastography and ...
    Shear wave elastography and ultrafast imaging of the carotid artery pulse wave were performed in 27 normotensive participants and 29 age- and sex-matched ...
  94. [94]
    Wireless, wearable elastography via mechano-acoustic wave ...
    Sep 3, 2025 · It provides continuous, depth-sensitive estimation of subcutaneous tissue stiffness through real-time surface wave dispersion analysis.
  95. [95]
    Shear Wave Elastography for Carotid Artery Stiffness - MDPI
    Breath-holding minimizes motion artifacts caused by respiratory movement; therefore, participants should hold their breath for 3 s to ensure image stability [66] ...
  96. [96]
    What we need to know when performing and interpreting US ...
    As inflammatory changes, including edema, can increase the elasticity and viscosity, inflammatory activity has a major impact on liver stiffness.Missing: challenges | Show results with:challenges
  97. [97]
    Ultrasound Elastography and MR Elastography for Assessing Liver ...
    The purpose of the article is to review the diagnostic performance of ultrasound and MR elastography techniques for detection and staging of liver fibrosis.
  98. [98]
    Inter-platform Variability of Liver Elastography: Pairwise ...
    Aug 29, 2022 · This study was aimed at determining whether liver stiffness measurements by 2-D shear wave elastography using GE's (2D-SWE-GE) and Canon's ...
  99. [99]
    WFUMB Guideline/Guidance on Liver Multiparametric Ultrasound
    The guidelines are clinically oriented, and the role of shear wave elastography in both fibrosis staging and prognostication in different etiologies of liver ...Missing: integration | Show results with:integration
  100. [100]
    Overview of Diagnostic Ultrasound Safety - Wiley Online Library
    Oct 8, 2025 · This article provides a comprehensive, organ-, and modality-specific review of current guidelines for ultrasound exposure parameters, including ...
  101. [101]
    Bias in artificial intelligence for medical imaging - PubMed Central
    AI in medical imaging is at risk of being compromised by several types of biases, which could adversely affect patient outcomes. • Understanding that medical ...Missing: elastography | Show results with:elastography
  102. [102]
    Diagnostic accuracy of artificial intelligence models for imaging ...
    Oct 2, 2025 · AI-based diagnostic models showed excellent performance, with pooled sensitivity of 91% (95% CI: 84–95%), specificity of 92% (95% CI: 86–96%), ...
  103. [103]
    Rapid Liver Fibrosis Evaluation Using the UNet-ResNet50-32 × 4d ...
    Oct 20, 2025 · The aim of this study is to use image segmentation technology for the computational interpretation of MRE, enabling AI to automatically segment ...
  104. [104]
    Deep learning in ultrasound elastography imaging: A review
    Aug 9, 2025 · Ultrasound elastography is a technique to characterize tissue stiffness using ultrasound imaging either by measuring tissue strain using quasi‐ ...
  105. [105]
    [PDF] Artificial Intelligence in Ultrasound Imaging: A Novel Perspective
    ... 3D tissue strain and strain distribution maps. Hybrids of elastography and shear wave elastography have been developed by AI and offer exquisite lesion ...
  106. [106]
    Hybrid photoacoustic and fast super-resolution ultrasound imaging
    Here we developed an interleaved PA/fast ULM imaging technique that enables super-resolution vascular and physiological imaging in less than 2 seconds per frame ...
  107. [107]
    Hybrid spherical array for combined volumetric optoacoustic and B ...
    Mar 15, 2024 · We present a new, to our knowledge, hybrid optoacoustic ultrasound (OPUS) imaging approach employing a spherical array with dedicated segments for each ...
  108. [108]
    Wireless, wearable elastography via mechano-acoustic wave ... - NIH
    Sep 3, 2025 · Wireless, wearable elastography via mechano-acoustic wave sensing for ambulatory monitoring of tissue stiffness ... Received 2025 Apr 9; Accepted ...Missing: prototypes | Show results with:prototypes
  109. [109]
    Wearable bioadhesive ultrasound shear wave elastography - Science
    Feb 9, 2024 · We introduce wearable bioadhesive ultrasound elastography (BAUS-E), which can generate acoustic radiation force impulse (ARFI) to induce shear waves.
  110. [110]
    Four-dimensional (4D) phase velocity optical coherence ... - NIH
    Here we propose a method to evaluate phase velocity, wave velocity at various frequencies, in four-dimensional space (x, y, z, f), called 4D-OCE phase velocity.
  111. [111]
    4D deep learning for real-time volumetric optical coherence ...
    We propose a new approach for elastography estimations using a fast imaging device to acquire small image volumes at rates of 831 Hz.Missing: advances | Show results with:advances
  112. [112]
    Elastography Imaging Market Trends & Forecast to 2032
    Aug 21, 2025 · The global elastography imaging market is projected to grow from US$ 4.6 Bn in 2025 to US$ 7.5 Bn by 2032, recording a CAGR of 7.2% during ...
  113. [113]
    Ultrasound Tech Trends Every Sonographer Should Know
    Jun 16, 2025 · Emerging ultrasound trends include 3D/4D imaging for detailed visuals, AI for enhanced diagnostics, and portable devices for remote imaging.