Fact-checked by Grok 2 weeks ago

MRI pulse sequence

An MRI pulse sequence is a precisely timed series of radiofrequency (RF) pulses and gradients applied during (MRI) to excite and manipulate the and of protons in , thereby generating detectable signals that form images with specific contrast and . These sequences operate on the principles of , where protons in a strong static (B0) align and precess at the Larmor frequency; an RF pulse, typically at 90° or 180° flip angles, perturbs this alignment, tipping magnetization into the to induce a measurable in receiver coils via relaxation processes (T1 longitudinal and transverse). Key parameters include repetition time (), the interval between RF s, and echo time (), the duration from excitation to signal peak, which dictate image weighting: short and yield T1-weighted images emphasizing anatomical detail, while long and produce T2-weighted images highlighting pathology like . Common types include spin-echo and gradient-echo sequences, with advanced variants such as fast spin-echo and inversion recovery enhancing efficiency and specificity for clinical applications.

Historical Development

The foundations of MRI pulse sequences trace back to the discovery of (NMR) by in 1938. In 1950, Erwin Hahn introduced the technique, which became fundamental for refocusing signals in MRI. The transition to imaging occurred in 1973 when developed methods using magnetic field gradients to spatially encode NMR signals, enabling the first MRI images. Subsequent advancements in the and , including fast spin echo in 1986 by Hennig and colleagues, significantly improved imaging speed and clinical utility. Pulse sequences are fundamental to MRI's versatility, enabling optimization of waveforms for spatial encoding (slice selection, , and directions) while adhering to hardware limits like maximum strength and , thus minimizing artifacts and scan times in diagnostic protocols.

Introduction

Definition and Purpose

A (MRI) pulse sequence is defined as a predefined series of radiofrequency (RF) pulses, magnetic fields, and specific timing intervals designed to manipulate the spins of protons in tissues, thereby generating measurable transverse signals that form the basis for reconstruction. These sequences control the , evolution, and detection of the MR signal by precisely timing the application of RF pulses to tip spins away from the main and using gradients to encode spatial . The primary purpose of MRI pulse sequences is to enable selective control over image contrast, allowing differentiation of tissues based on intrinsic properties such as longitudinal (T1) and transverse (T2) relaxation times, proton density, diffusion coefficients, or blood flow characteristics, which is essential for identifying pathological changes in clinical diagnostics. By adjusting parameters like repetition time (TR), echo time (TE), and flip angles, sequences can emphasize specific contrast mechanisms—for instance, short TR and TE for T1-weighted images that highlight anatomical details, or longer values for T2-weighted images that reveal edema or inflammation. This versatility underpins MRI's non-invasive ability to provide high-resolution, multi-contrast imaging tailored to applications in neurology, oncology, and cardiology. In a typical pulse sequence workflow, the process begins with an excitation phase, often using a 90° RF pulse to rotate the net magnetization into the transverse plane, creating a detectable free induction decay (FID) signal. This is followed by a manipulation phase involving additional RF refocusing pulses or gradient applications to mitigate dephasing effects and enhance signal coherence, and concludes with a readout phase where frequency- and phase-encoding gradients are applied to acquire the spatial signal data during echo formation. The foundational concepts of MRI pulse sequences were first articulated in the 1970s by , who demonstrated spatial encoding via gradients in 1973, and , who advanced rapid imaging techniques, contributions that earned them the 2003 Nobel Prize in Physiology or Medicine for establishing the principles of MRI.

Historical Development

The foundations of MRI pulse sequences trace back to the discovery of (NMR) by and Edward Purcell in 1946, who independently demonstrated the absorption of radiofrequency energy by atomic nuclei in magnetic fields, laying the groundwork for signal generation in both NMR and later MRI applications. Their work earned them the 1952 Nobel Prize in Physics, establishing the basic principles of spin manipulation through radiofrequency pulses that would evolve into MRI sequencing. In 1950, Erwin Hahn introduced the technique in NMR, using a 90°-180° radiofrequency pulse pair to refocus dephased spins and form an echo, which became the cornerstone for early MRI sequences by mitigating field inhomogeneities. This method was first applied to medical diagnostics in 1971 by , who used NMR to measure relaxation time differences in normal and cancerous tissues, proposing its use for tumor detection and marking the first link between NMR and medical diagnostics. advanced this in 1973 by incorporating gradients with signals to encode spatial information, enabling the first 2D NMR images and transforming pulse sequences into true imaging tools. The 1970s saw further innovations for efficiency, with , Dieter Welti, and Richard Ernst introducing sequences in 1975, which replaced the 180° refocusing pulse with a reversal to generate echoes faster, facilitating transform-based and reducing scan times from hours to minutes. In 1977, developed echo-planar (EPI), an ultrafast variant that acquires an entire image plane in milliseconds via rapid switching after a single excitation, enabling real-time applications and earning him the 2003 in Physiology or Medicine shared with Lauterbur. Meanwhile, Edward Stejskal and James Tanner's 1965 pulsed method for measuring in NMR was adapted clinically in the , incorporating bipolar gradients into sequences to quantify water molecule motion for detecting strokes and tumors. Clinical adoption accelerated with the U.S. Food and Drug Administration's approval of the first commercial MRI scanners in 1984, primarily using sequences for whole-body at institutions like those employing Fonar systems. Contributions from David Hoult and Lauterbur in the late refined design and , optimizing radiofrequency coils and gradients for human-scale while addressing power deposition limits. By the 2000s, standardization efforts through the protocol ensured interoperability of sequence data across scanners, defining attributes like scanning sequence types (e.g., , ) to support consistent clinical workflows and multi-vendor integration. These developments were driven by demands for shorter scan times—from initial multi-hour acquisitions to seconds—and enhanced contrast for and , where sequences evolved to exploit T1/T2 relaxation and for tissue differentiation without .

Fundamental Principles

RF Pulses, Gradients, and Echo Formation

Radiofrequency (RF) pulses are essential components in MRI pulse sequences, serving to excite and manipulate the of atomic nuclei, primarily hydrogen protons. The primary types include excitation pulses, typically with a 90° flip angle to tip the longitudinal magnetization into the , and refocusing pulses, often 180° to reverse effects. The flip angle θ is determined by the γ of the nucleus, the amplitude of the RF magnetic field B₁, and the pulse duration τ, according to the relation θ = γ B₁ τ in the for short pulses. To achieve spatial selectivity, such as slice selection, RF pulses are combined with gradients; a frequency-modulated or shaped RF pulse applied during a slice-select (usually G_z) excites only within a specific corresponding to the desired slice thickness. Magnetic gradients play a crucial role in spatial encoding and signal manipulation within MRI pulse sequences. Three orthogonal gradients—G_x, G_y, and G_z—are used: G_z for slice selection during RF , G_y for encoding to impart position-dependent shifts across one , and G_x for encoding to create position-dependent variations during readout. These gradients induce by causing at different positions to precess at varying rates, which can be reversed through rephasing gradients of opposite to form echoes. and rephasing lobes in the gradient waveforms are precisely timed to control the and minimize artifacts. Echo formation is a fundamental process in MRI for generating detectable transverse signals after initial . Following a 90° RF , the transverse produces a (FID), a rapidly decaying signal due to T₂* relaxation and field inhomogeneities, which can be directly acquired but is limited by susceptibility effects. A is formed by applying a 180° refocusing RF at time /2 after ; this reverses the phase dispersion from static field inhomogeneities, refocusing the spins to produce a maximum signal at time , as originally demonstrated by Hahn in 1950. In contrast, a is generated without a 180° by applying a lobe followed by a rephasing readout of opposite , refocusing the gradient-induced more rapidly but remaining sensitive to T₂* effects. Key timing parameters govern the evolution of the and the resulting signal. The echo time (TE) is the interval from the RF pulse to the peak of the , influencing the degree of transverse relaxation, while the repetition time () is the period between successive , affecting longitudinal recovery. For sequences, the signal intensity at time t follows S(t) ∝ M₀ sin(θ) e^{-t/T₂*}, where M₀ is the equilibrium , θ is the flip angle, and T₂* is the effective transverse relaxation time, highlighting the dependence on efficiency and rapid decay. These elements—RF pulses, gradients, and echoes—combine to traverse k-space, the Fourier domain representation of the image, enabling spatial encoding for reconstruction. Gradients during phase and frequency encoding steps trace specific trajectories in (e.g., Cartesian lines or spirals), sampling the spatial frequency components of the object; the inverse then reconstructs the spatial image from this data, as pioneered by Lauterbur in 1973 using projection reconstruction. This framework forms the prerequisite for all subsequent pulse sequences, allowing efficient filling of to achieve high-resolution imaging.

Relaxation Processes and Contrast Mechanisms

In (MRI), contrast arises primarily from differences in tissue relaxation properties following radiofrequency excitation. Longitudinal relaxation, characterized by the time constant T1, describes the recovery of the net vector along the direction of the main B0 after . This process involves energy transfer from excited spins to the surrounding lattice (spin-lattice relaxation), with the rate depending on molecular tumbling rates matching the Larmor frequency. The of the longitudinal Mz is given by M_z(t) = M_0 \left(1 - e^{-t/T_1}\right), where M0 is the equilibrium magnetization and t is time since excitation, assuming an initial Mz=0. T1 values typically range from 200-3000 ms in biological tissues at 1.5-3 T fields, longer in fluids like cerebrospinal fluid and shorter in fat-rich tissues due to efficient energy dissipation. Transverse relaxation, governed by the time constant T2, refers to the decay of the magnetization component perpendicular to B0 due to spin-spin interactions, leading to dephasing from local field variations induced by neighboring spins. The transverse magnetization Mxy decays as M_{xy}(t) = M_0 e^{-t/T_2}, with T2 values generally shorter than T1 (e.g., 50-200 ms in gray matter at 3 T), reflecting irreversible loss of phase coherence. In practice, the observed decay is faster, described by T2*, which incorporates additional dephasing from magnetic field inhomogeneities and susceptibility effects: \frac{1}{T_2^*} = \frac{1}{T_2} + \frac{1}{T_2'} , where T2' accounts for reversible extrinsic factors. T2 is the intrinsic tissue property, while T2* dominates in gradient-echo acquisitions. Beyond relaxation, proton density (PD) provides a baseline signal proportional to the number of mobile nuclei per unit , contributing to in low-relaxation-weighting regimes (e.g., PD-weighted images highlight differences in between tissues like muscle and fat). Diffusion, the of water molecules, modulates T2 by enhancing dephasing in restricted environments, such as in or tumors, where reduced diffusivity alters local phase accrual. Perfusion introduces dynamic from blood flow, affecting signal via delivery of oxygenated/deoxygenated or contrast agents, which influence T2* through changes without direct relaxation modification. Contrast generation exploits these properties by adjusting repetition time (TR) and echo time (TE). Proton density weighting uses long TR (to allow full T1 recovery and minimize T1 effects) and short TE (to minimize T2 effects), emphasizing PD differences. T1 weighting employs short TR (to limit recovery and emphasize T1 differences) and short TE (to reduce T2 influence), highlighting tissues with short T1 (e.g., fat appears bright). T2 weighting requires long TR and long TE, accentuating tissues with long T2 (e.g., fluids appear bright). These parameters manipulate the steady-state signal as S ∝ PD · (1 - e^{-TR/T1}) · e^{-TE/T2}, enabling tissue differentiation based on biophysical variations. For quantitative imaging, accurate T1 and mapping requires correction for radiofrequency field () inhomogeneities, which cause flip-angle errors in variable flip-angle methods. mapping techniques, such as the Bloch-Siegert shift, improve T1 precision by up to 20-30% in regions like the breast or , ensuring reliable relaxometry. Post-2010 advancements in multi-parametric mapping integrate T1, , and PD acquisitions for comprehensive characterization, as in myocardial disease assessment, where mapping detects inflammation with reproducibility errors below 5%.

Conventional Sequences

Spin Echo

The spin echo sequence is a foundational in (MRI), first demonstrated by Erwin Hahn in through experiments showing that a refocusing radiofrequency (RF) pulse could regenerate transverse magnetization after initial dephasing. In this sequence, a 90° excitation pulse tips the longitudinal magnetization into the , initiating (FID) that dephases due to spin-spin interactions and field inhomogeneities. A subsequent 180° refocusing pulse, applied at time TE/2 (where TE is the echo time), reverses the phase dispersion of spins, causing them to rephase and form a detectable signal at time TE. This refocusing mechanism distinguishes from simple FID, as it corrects for static (B0) inhomogeneities and susceptibility-induced dephasing, yielding a signal that decays primarily according to the true transverse relaxation time rather than the faster T2*. For multi-echo acquisitions, additional 180° pulses can generate successive echoes, allowing simultaneous collection of proton density and T2-weighted data within a single repetition. The sequence timing can be visualized as follows: following slice-selective gradients, the 90° RF pulse is applied, followed by phase-encoding and frequency-encoding gradients during the echo formation; the 180° pulse occurs midway, with readout gradients centered on the echo peak at TE. Key parameters include the repetition time (TR), the interval between successive 90° excitations, and TE. For T2-weighted imaging, a long TR (>2000 ms) minimizes T1 effects, while a long TE (80-120 ms) emphasizes T2 contrast; for proton density weighting, a long TR pairs with a short TE (<30 ms); and for T1 weighting, a shorter TR (500-800 ms) with a short TE enhances differences in longitudinal recovery. These parameters render the spin echo insensitive to B0 inhomogeneities, unlike gradient echo sequences, making it robust against artifacts from metal implants or air-tissue interfaces. As the gold standard for T2 contrast, the spin echo sequence excels in providing high-quality images with reduced susceptibility artifacts, enabling clear visualization of tissue pathology such as edema or inflammation. It is routinely employed in clinical neuroimaging and spine imaging to detect lesions, demyelination, and cerebrospinal fluid characteristics. To accelerate acquisition without sacrificing contrast, the fast spin echo (FSE) or turbo spin echo (TSE) variant—introduced by Hennig et al. in 1986 as rapid acquisition with relaxation enhancement (RARE)—uses multiple 180° refocusing pulses per TR to fill multiple lines of k-space, with echo train lengths (ETL) typically of 8-16 echoes. This reduces scan times by factors of 8-16 compared to conventional spin echo, making it suitable for high-resolution T2-weighted imaging in routine protocols. The signal intensity in a spin echo sequence follows the equation: S = k \cdot \mathrm{PD} \left(1 - e^{-\mathrm{TR}/T_1}\right) e^{-\mathrm{TE}/T_2} where S is the signal, k is a proportionality constant, PD is proton density, T_1 is the longitudinal relaxation time, and T_2 is the transverse relaxation time. This derives from the Bloch equations describing magnetization evolution. After the 90° pulse, the longitudinal magnetization M_z recovers toward equilibrium M_0 as M_z(t) = M_0 (1 - e^{-t/T_1}) during TR, contributing the (1 - e^{-\mathrm{TR}/T_1}) term; at the next excitation, this recovered M_z is tipped into the transverse plane. The transverse magnetization M_{xy} then dephases during the initial TE/2 due to T2 effects, but the 180° pulse inverts the phases, allowing rephasing over the subsequent TE/2, resulting in an echo amplitude attenuated solely by e^{-\mathrm{TE}/T_2}—unlike FID, where irreversible T2* decay (including inhomogeneity contributions) governs the signal without refocusing. This derivation assumes ideal pulses and neglects diffusion or higher-order effects, but it establishes the core contrast mechanisms reliant on T1 recovery and T2 decay.

Gradient Echo

Gradient echo (GRE) sequences in magnetic resonance imaging (MRI) employ a partial flip angle excitation, typically less than 90°, followed by gradient reversal to form an echo, without the use of a 180° refocusing radiofrequency (RF) pulse. This design contrasts with techniques by relying solely on magnetic field gradients for echo formation, making the sequence inherently sensitive to relaxation effects, which include contributions from both true decay and magnetic field inhomogeneities such as susceptibility-induced dephasing. As a result, GRE sequences produce images weighted by contrast, which can highlight tissue differences due to local field variations but also introduce artifacts in regions with air-tissue interfaces or metal implants. Key parameters in GRE sequences include the echo time (TE), typically short at 2-5 ms to minimize T2* decay and maintain signal intensity, and the repetition time (TR), which can be varied from short (e.g., 5-20 ms) for rapid imaging to longer values for specific contrasts. The steady-state signal in GRE is governed by the , adjusted for T2* weighting: S = \mathrm{PD} \cdot \sin\alpha \cdot \frac{1 - e^{-\mathrm{TR}/T_1}}{1 - \cos\alpha \cdot e^{-\mathrm{TR}/T_1}} \cdot e^{-\mathrm{TE}/T_2^*} where PD is proton density, α is the flip angle, and T1 and T2* are the longitudinal and effective transverse relaxation times, respectively. This equation describes the transverse magnetization at steady state for low flip angles and short TR, optimizing signal through the Ernst angle (α_E = \arccos(e^{-\mathrm{TR}/T_1})), which balances excitation efficiency and T1 recovery. GRE variants include spoiled gradient echo (GRE) sequences, such as fast low-angle shot (FLASH), which apply RF spoiling or gradient crushing to eliminate residual transverse magnetization, yielding primarily T1-weighted images suitable for anatomical visualization. In contrast, balanced steady-state free precession (bSSFP) maintains a steady state without spoiling by balancing gradient moments, providing high signal-to-noise ratio (SNR) and T2/T1 contrast, often used in dynamic applications. The primary advantages of GRE sequences lie in their speed, enabling acquisition times of seconds per slice through short TR and TE, which facilitates 3D volumetric imaging and real-time applications like . This rapidity stems from the absence of a 180° pulse, allowing higher duty cycles for RF transmission. However, limitations include heightened susceptibility to magnetic field inhomogeneities, leading to signal loss and distortions, particularly in where off-resonance effects cause banding artifacts; recent advancements, such as interleaved radial linear combination with partial dephasing, have mitigated these in cardiac imaging near implantable devices, improving cine quality for functional assessment.

Inversion-Based Sequences

Inversion Recovery

The inversion recovery (IR) sequence serves as a preparatory module in magnetic resonance imaging (MRI) to enhance T1 contrast by manipulating longitudinal magnetization prior to image acquisition. It begins with a 180° radiofrequency (RF) inversion pulse that flips the net magnetization vector along the longitudinal axis (Mz) from +M0 to -M0, where M0 represents the equilibrium magnetization. Following this, an inversion time (TI) delay allows partial recovery of Mz toward equilibrium through T1 relaxation, after which a standard 90° excitation pulse tips the recovered magnetization into the transverse plane for signal readout. The readout module can employ either a spin echo (SE) or gradient echo (GRE) acquisition to form the image, enabling flexibility in TE and TR parameters while preserving the T1-weighted preparation. The signal evolution in IR is governed by the T1 relaxation process during the TI period, yielding a longitudinal magnetization of: M_z(\text{TI}) = M_0 \left(1 - 2 e^{-\text{TI}/T_1}\right) where the factor of 2 accounts for the initial inversion. This results in a null point—where signal from a tissue is completely suppressed—when TI = T1 \ln(2) \approx 0.69 T1, allowing selective nulling based on tissue-specific T1 values. For T1-weighted imaging, an intermediate TI (typically 300-500 ms at 1.5 T) is selected to capture differences in recovery rates between tissues, producing high contrast where short-T1 tissues (e.g., fat and white matter) appear bright and long-T1 tissues (e.g., gray matter, fluids) appear relatively dark. Conversely, longer TI values (e.g., >2000 ms) can suppress signals from fluids with prolonged T1, though this is tuned carefully to avoid over-recovery. These parameters leverage the inherent T1 relaxation differences outlined in fundamental contrast mechanisms. IR offers distinct advantages in clinical MRI, particularly for enhancing T1 contrast to improve detection in tissues with subtle relaxation differences, such as in or musculoskeletal applications. By providing superior gray-white matter differentiation compared to standard T1-weighted sequences, it aids in identifying pathologies like tumors or . Additionally, the inversion preparation module can be integrated as a transfer step in hybrid sequences, optimizing signal for subsequent acquisitions without altering core imaging parameters. Historically, IR was developed in the early as one of the foundational MRI techniques, initially for quantitative T1 mapping to characterize tissue properties accurately.

Variants for Specific Contrasts

Short tau inversion recovery (STIR) is an inversion recovery variant optimized for fat suppression by selecting an inversion time (TI) that nulls the signal, typically TI ≈ 140 at 1.5 T corresponding to the T1 relaxation time of (≈ 250 ). This sequence employs a long repetition time (TR > 2000 ) to allow recovery and is particularly effective at lower field strengths due to its field-independent fat suppression mechanism. Clinically, STIR is widely used to highlight edema and inflammation in musculoskeletal imaging, such as detecting bone marrow edema or abnormalities, by enhancing the of water-rich pathologies against suppressed background. Fluid-attenuated inversion recovery (FLAIR) adapts inversion recovery with a long TI (2000-2500 ) to suppress (CSF) signal, which has a prolonged T1, while using an extended (> 8000 ) to maintain T2-weighted contrast in . This results in dark CSF on images, preventing its hyperintensity from obscuring adjacent lesions, and is essential for protocols. FLAIR excels in detecting periventricular and subcortical lesions, such as those in or , by improving lesion conspicuity without CSF interference. Magnetization-prepared rapid acquisition of gradient echoes (MP-RAGE) combines inversion recovery preparation with a rapid readout to produce high-resolution three-dimensional T1-weighted images, typically using TI ≈ 900 ms and a low flip angle (α ≈ 10°) to optimize gray-white matter contrast. The inversion pulse enhances T1 weighting, followed by segmented acquisition for efficiency, enabling isotropic voxels for detailed structural brain imaging. It is a standard for volumetric analysis in , supporting applications like cortical thickness measurement and atrophy assessment in neurodegenerative diseases. Double inversion recovery extends the technique by applying two 180° inversion pulses to suppress signals from two distinct tissues simultaneously, such as CSF (TI1 ≈ 2000-3000 ms) and (TI2 ≈ 450 ms), on a T2-weighted background. This dual suppression improves the detection of gray matter lesions, particularly in , by reducing background noise from normal tissues. Clinical protocols often incorporate it for or infratentorial imaging, though artifacts from imperfect inversion—due to inhomogeneities or motion—can lead to incomplete suppression and require careful parameter adjustment.

Motion and Diffusion Sequences

Diffusion-Weighted Imaging

Diffusion-weighted imaging (DWI) is an MRI technique that sensitizes the signal to the random of water molecules in tissues by applying pairs of pulsed gradients. This method exploits the fact that is restricted in certain pathological states, such as cellular swelling, allowing for the detection of microstructural changes. The core principle relies on the Stejskal-Tanner equation, which describes the signal attenuation due to : S = S_0 e^{-b D} where S is the observed signal intensity, S_0 is the signal without diffusion weighting, D is the apparent diffusion coefficient, and b is the b-value that quantifies the degree of sensitization. The b-value is calculated as b = \gamma^2 \delta^2 g^2 (\Delta - \delta/3), with \gamma as the , \delta as the gradient duration, g as the gradient amplitude, and \Delta as the time between the leading edges of the gradients; typical b-values range from 0 to 1000 s/mm² to balance sensitivity and . The standard DWI sequence is based on a framework, where diffusion-sensitizing gradients are applied before and after the 180° refocusing pulse to encode motion-induced shifts. A echo-planar (EPI) readout is commonly used immediately following the to acquire the entire image plane rapidly, minimizing motion artifacts and enabling T2*-weighted contrast. Key parameters include an echo time (TE) of 60-100 ms to accommodate gradient timing while preserving signal, and acquisition of images at multiple b-values (e.g., b=0 and b=1000 s/mm²) to compute apparent coefficient (ADC) maps via pixel-wise fitting of the Stejskal-Tanner . For isotropic measurement, gradients are applied in three orthogonal directions and averaged, whereas directional sensitivity is achieved by varying gradient orientations to probe tissue anisotropy. Clinically, DWI excels in detecting acute ischemic , where cytotoxic restricts , yielding hyperintense signals on trace images and low values in the infarct core within minutes of onset. Diffusion tensor imaging (DTI), an extension of DWI, acquires data in at least six non-collinear directions to model as a tensor, enabling quantification of (FA)—a scalar measure (0-1) of directional in motion that highlights organized structures like tracts for fiber tracking in . Common artifacts in DWI include distortions from eddy currents induced by rapid gradient switching, which cause geometric warping, and bulk motion leading to signal loss or ghosting. These are mitigated through pre-emphasis calibration of gradients for eddy current compensation and navigator echoes or prospective motion correction to track and adjust for patient movement during acquisition.

Perfusion-Weighted Imaging

Perfusion-weighted imaging (PWI) in MRI quantifies tissue perfusion by measuring blood flow and volume using either exogenous contrast agents or endogenous blood water as tracers. Exogenous techniques rely on gadolinium-based contrast agents (GBCAs) injected as a bolus to track hemodynamic changes, while endogenous methods magnetically label arterial blood without contrast. These sequences are essential for assessing microvascular dynamics in clinical settings, providing parameters such as cerebral blood flow (CBF) and cerebral blood volume (CBV). Dynamic susceptibility contrast (DSC)-MRI is a primary exogenous method that uses T2*-weighted gradient-echo echo-planar (GRE-EPI) to detect transient signal drops caused by the effects of a GBCA bolus passing through capillaries. The signal intensity decreases due to local inhomogeneities induced by the , allowing measurement of the first-pass bolus dynamics. The transverse relaxation rate change is modeled as R_2^*(t) = R_2^*_0 + \Delta R_2^* [\text{Gd}(t)], where R_2^*_0 is the baseline rate, \Delta R_2^* is the relaxivity, and [\text{Gd}(t)] is the concentration over time; this equation enables derivation of concentration-time curves from signal changes. Dynamic contrast-enhanced (DCE)-MRI complements DSC by employing T1-weighted GRE sequences to evaluate vessel permeability and , tracking the T1-shortening effects of GBCA leakage into the extravascular space for parameters like the volume transfer constant K^\text{trans}. quantification in both requires an arterial input function (AIF) derived from major arteries, followed by deconvolution techniques such as (SVD) to compute CBF from the function; CBV is obtained by integrating the tissue concentration-time curve, often normalized to gray matter values. Arterial spin labeling (ASL) provides a non-invasive endogenous by inverting upstream of the imaging slice to create a diffusible tracer from protons. Flow-sensitive inversion recovery (), a pulsed ASL variant, applies slice-selective and non-selective inversion pulses to label blood selectively, while pseudo-continuous ASL (pCASL) uses a train of rapid RF pulses with gradients for efficient, multi-slice labeling and higher . The quantitative imaging of using a single subtraction () variant enhances accuracy by incorporating saturation pulses to control bolus length and multi-inversion time sampling for robust CBF estimation. Recent advancements in the include AI-accelerated ASL, such as networks that reduce required post-labeling delays by over 60% while maintaining quantitative CBF and arterial transit time measurements, aiding standardization efforts. Clinically, PWI aids in glioma grading by leveraging elevated relative CBV (rCBV > 1.75) in high-grade tumors to predict aggressiveness and guide targeting. In acute , it identifies the ischemic penumbra—hypoperfused but viable tissue—through perfusion-diffusion mismatch, where delays in time-to-peak (>6 seconds) indicate salvageable regions for intravenous up to 9 hours post-onset in select patients. These applications underscore PWI's role in distinguishing tumor recurrence from pseudoprogression and optimizing interventions.

Functional and Vascular Sequences

Functional MRI

Functional magnetic resonance imaging (fMRI) is a technique that maps activity by detecting changes in oxygenation levels, primarily through the blood oxygenation level-dependent (BOLD) contrast mechanism. The BOLD effect arises from the paramagnetic properties of deoxyhemoglobin, which increases and accelerates T2* decay in regions with lower oxygenation, leading to signal loss in gradient-echo sequences. During neural activation, cerebral flow rises disproportionately to oxygen consumption, reducing deoxyhemoglobin concentration and thereby increasing the T2*-weighted signal in active areas. This non-invasive method was first demonstrated in humans using sensory stimulation tasks, enabling real-time mapping of cortical responses without exogenous contrast agents. The standard pulse sequence for BOLD fMRI is echo-planar imaging (EPI) combined with (GRE), optimized for rapid whole-brain coverage and sensitivity to T2* changes. Key parameters include an echo time () of approximately 30-40 ms to maximize BOLD contrast near the T2* peak, and a repetition time () of 2-3 seconds for adequate temporal sampling of hemodynamic responses. Typical spatial parameters feature a matrix size of 64x64 to 128x128 voxels and slice thickness of 3 mm, balancing resolution with and scan time. These settings allow acquisition of multiple slices covering the in under 3 seconds, essential for capturing dynamic activity patterns. fMRI experiments employ two main paradigms: task-based designs, such as or event-related protocols where subjects perform alternating periods of and rest, and resting-state paradigms that analyze spontaneous low-frequency fluctuations in the absence of tasks. is critical and includes steps like motion correction to realign images across time points, spatial to a standard template, and smoothing to enhance signal detection. Statistical analysis typically uses the general (GLM) to fit voxel-wise against expected hemodynamic response functions, generating activation maps via t-tests or on parameter estimates. Recent advances have improved fMRI efficiency, particularly through multiband EPI, which simultaneously excites multiple slices to achieve higher acceleration factors (e.g., 4-8x) and sub-second whole-brain in the . Emerging sequences like zero echo time () fMRI provide artifact-free and quiet imaging alternatives, reducing distortions and acoustic noise for better and mapping. At ultra-high fields like 7T, enhanced BOLD sensitivity and enable finer mapping of subcortical structures and laminar cortical activity, though challenges such as increased artifacts persist. Further progress at 9.4T using 3D stack-of-spirals readouts has demonstrated even higher resolution for preclinical and advanced human studies as of 2025.

Magnetic Resonance Angiography

(MRA) encompasses a group of techniques designed for non-invasive visualization of blood vessels, primarily relying on flow-induced signal alterations rather than exogenous contrast agents. These methods exploit the differences in between stationary tissues and flowing blood to generate angiographic images, enabling assessment of vascular and without catheterization. Time-of-Flight (TOF) MRA is a cornerstone non-contrast technique that utilizes sequences with short repetition time () and echo time () to achieve flow-related enhancement. In this approach, repeated radiofrequency pulses saturate the longitudinal magnetization of stationary spins within the volume, while inflowing blood from outside the slice remains unsaturated and yields high signal intensity upon . This contrast mechanism is particularly effective for arteries with steady inflow perpendicular to the plane. TOF-MRA can be implemented in two-dimensional () mode, acquiring thin slices sequentially for better depiction of tortuous s, or three-dimensional () mode using slab for isotropic high-resolution of segments, though 3D acquisitions are more prone to intravoxel in longer flow paths. Key parameters for TOF-MRA include values under 50 ms to minimize of inflowing spins, as short as possible (typically 2-7 ms) to reduce T2* decay and artifacts, and flip angles of 20-30° to balance background suppression with blood signal optimization. Spatial presaturation bands are routinely applied upstream or downstream of the slab to suppress signals from veins or unwanted arterial branches, enhancing arterial specificity. These settings allow for rapid acquisitions, often under 5-10 minutes for intracranial studies, while maintaining vessel-to-background contrast. Phase Contrast (PC) MRA complements TOF by quantifying blood and direction through phase differences induced by motion-sensitive gradients, making it suitable for both qualitative and hemodynamic assessment. Bipolar velocity-encoding gradients are applied along one or more directions, creating a first-order moment that imparts a phase shift proportional to spin ; the phase φ is calculated as \phi = \gamma \int \mathbf{G} \cdot \mathbf{r} \, dt, where \gamma is the , \mathbf{G} is the , and \mathbf{r} is the of the moving . Magnitude and images are reconstructed from acquisitions with and without encoding, with the map directly reflecting . To avoid phase wrapping (aliasing), the velocity encoding parameter (VENC) is selected to match the anticipated peak flow velocity, typically 50-150 cm/s for or up to 200 cm/s for larger vessels; velocities exceeding VENC/2 cause , necessitating higher VENC values at the cost of for slower flows. PC-MRA supports multi-directional encoding for comprehensive flow mapping but requires longer scan times (10-20 minutes) due to multiple phase-encoding steps. Clinically, TOF-MRA excels in screening for , where it accurately depicts luminal narrowing with sensitivity over 90% for high-grade lesions, and intracranial s, providing non-invasive detection of saccular dilatations as small as 3 mm. PC-MRA adds value in evaluating flow dynamics across stenoses or in necks, aiding in rupture by quantifying or jet velocities. Both techniques are preferred for pediatric and renal-impaired patients due to the absence of risks. Despite their advantages, TOF-MRA suffers from signal voids in regions of slow, turbulent, or in-plane , potentially leading to overestimation of or underdetection of aneurysms with complex geometries. PC-MRA is limited by sensitivity to higher-order motion (e.g., ) and reduced accuracy in non-laminar flows, such as those in severe or aneurysms, where phase dispersion occurs. Scan times and susceptibility to patient motion remain challenges for both. Recent advances in MRA include applications at ultra-high fields like 7T, which enhance resolution for imaging small intracranial vessels and perforators, improving detection of microvascular pathologies such as , as of 2025. For cases where flow artifacts compromise non-contrast MRA, contrast-enhanced MRA using chelates with T1-weighted sequences provides robust vessel opacification by shortening blood T1 relaxation, though it introduces risks like in vulnerable populations.

Advanced Contrast Sequences

Susceptibility Weighted Imaging

(SWI) is a gradient-echo-based technique that exploits differences in among tissues to generate high-contrast images, particularly sensitive to paramagnetic substances such as deoxyhemoglobin, iron deposits, and calcium. The method combines magnitude and information from a fully flow-compensated, high-resolution three-dimensional gradient-echo sequence to amplify susceptibility-induced effects, enabling visualization of small venous structures, hemorrhages, and mineral accumulations that are subtle in conventional imaging. Unlike standard T2*-weighted gradient-echo sequences, which primarily rely on magnitude signal decay due to T2* relaxation, SWI incorporates post-processing steps to selectively enhance phase variations arising from local field inhomogeneities. The SWI sequence typically employs a three-dimensional spoiled gradient-recalled echo acquisition with echo times (TE) of 20-40 ms to accumulate sufficient phase accrual from susceptibility sources, repetition times (TR) of 25-50 ms, and flip angles of 12-20° (often 15°) to optimize signal-to-noise ratio while maintaining steady-state conditions. High spatial resolution, such as 0.5 mm isotropic voxels, is achieved through thin slices and matrix sizes up to 512×512, often at field strengths of 3T or higher for enhanced susceptibility contrast. Multiple echoes may be acquired in advanced implementations to improve phase reliability and enable multi-echo processing for better background suppression. The core post-processing involves creating a phase mask from the filtered phase image, obtained via homodyne high-pass filtering. This is done by dividing the original complex image by its low-pass filtered version (using a slice-by-slice 2D Hanning window of size e.g., 64×64), and taking the phase of the resulting high-pass filtered complex image to obtain the filtered phase \phi_f(\mathbf{r}). The phase mask m(\mathbf{r}) is then generated from the negative (paramagnetic) portion of the filtered phase, assuming a convention where negative phase indicates increased susceptibility: m(\mathbf{r}) = \begin{cases} \frac{\pi + \phi_f(\mathbf{r})}{\pi} & -\pi \leq \phi_f(\mathbf{r}) \leq 0 \\ 1 & 0 < \phi_f(\mathbf{r}) \leq \pi \end{cases} The final SWI image is the magnitude multiplied by the mask raised to a power (e.g., m^4) for contrast enhancement. This processing distinguishes SWI from basic gradient-echo by providing susceptibility-specific amplification without altering the acquisition. SWI finds primary applications in for detecting cerebral microbleeds, venous structures in acute , and lesions in , where it reveals perivenular iron deposition and demyelination-related susceptibility changes with superior sensitivity compared to . In and neurodegenerative diseases, it identifies from prior hemorrhages and abnormal iron accumulation in . An extension, quantitative susceptibility mapping (QSM), derives absolute susceptibility values in parts per billion by solving the ill-posed of dipole field , often using morphology-enabled regularization to handle streaking artifacts. Recent integrations of , such as deep learning-based acceleration and regularization in QSM pipelines, have improved reconstruction speed and accuracy for clinical deep brain nuclei evaluation as of 2025.

Magnetization Transfer and Fat Suppression

Magnetization transfer (MT) imaging enhances contrast by exploiting the of magnetization between free protons and those bound to macromolecules, such as proteins and in tissues. This technique employs an off- radiofrequency (RF) pulse that selectively saturates the protons in the immobile macromolecular pool, which have a broader linewidth due to their restricted motion. The saturation indirectly reduces the signal from the free pool through cross-relaxation and chemical , without directly exciting the free protons at their on- frequency. The magnetization transfer ratio (MTR), a common quantitative measure, is calculated as MTR = (M₀ - M_sat) / M₀, where M₀ is the signal intensity without the MT pulse and M_sat is the signal with the saturation pulse applied. This ratio reflects the efficiency of magnetization exchange and is typically expressed as a . In practice, the MT pulse—a series of high-power, off-resonance RF pulses—is added as a preparation module to standard (GRE) or (SE) sequences, allowing integration into routine imaging protocols without significantly extending scan times. MT imaging finds key applications in quantifying content and detecting pathological changes, such as in plaques where reduced correlates with demyelination. For instance, lower values in lesions indicate disrupted macromolecular integrity, aiding in early plaque identification and monitoring disease progression. Clinically, it improves visualization of brain tissue microstructure, particularly for myelin-sensitive contrasts in neurological assessments. Fat suppression techniques are essential in MRI to eliminate the bright signal from , which can otherwise obscure or mimic lesions due to its short T1 relaxation time. Several methods achieve this by targeting the difference between and protons, approximately 3.5 at 1.5 T, allowing selective manipulation of . selective (CHESS) saturation uses a frequency-selective 90° RF tuned to the resonance (offset by -3.5 from ), followed by a spoiler to dephase the excited , and sometimes preceded by a 180° for better homogeneity in variants like the 90°-180°-90° scheme. This method provides uniform suppression in homogeneous fields but is sensitive to B₀ and B₁ inhomogeneities, particularly at higher field strengths. Short tau inversion recovery (STIR), an inversion recovery-based approach, applies a non-selective 180° inversion pulse followed by a short inversion time (TI ≈ 140-170 ms at 1.5 T) to null the fat signal, as detailed in inversion recovery sequences. It offers robust suppression independent of field strength but suppresses all short T1 tissues, limiting its use with contrast agents. Spectral adiabatic inversion recovery (SPAIR) combines spectral selectivity with adiabatic inversion pulses, which are insensitive to B₁ variations, to invert only fat protons before allowing recovery to null the signal at a specific TI. This technique excels in regions with field inhomogeneities, such as the extremities, providing more homogeneous suppression than CHESS. The Dixon method, a phase-based water-fat separation , acquires images at echo times where fat and water signals are in-phase and opposed-phase, modeling the complex signal as S = |W + F e^{iφ}|, where W and F are water and fat magnitudes, and φ encodes the chemical shift phase evolution (φ = 2π Δf TE + ψ, with Δf ≈ 220 Hz at 1.5 T). Iterative decomposition solves for separate water and fat images, yielding robust suppression even in inhomogeneous fields and enabling fat fraction quantification. Seminal work by Dixon introduced this opposed-phase approach, with modern multi-point variants improving accuracy. In clinical practice, these fat suppression methods are vital in orthopedics for detecting and soft-tissue injuries, where CHESS or SPAIR enhances in T2-weighted images of the musculoskeletal system. In , Dixon and SPAIR techniques improve lesion conspicuity by reducing fat artifacts, facilitating accurate assessment of enhancing tumors in dynamic -enhanced protocols.

Specialized and Emerging Sequences

Neuromelanin Imaging

Neuromelanin imaging utilizes T1-weighted MRI sequences to detect neuromelanin, a dark pigment accumulated in catecholaminergic neurons of the substantia nigra pars compacta (SNc) and locus coeruleus (LC). The paramagnetism arising from neuromelanin-iron complexes shortens T1 relaxation times, producing hyperintense signals in these midbrain structures on T1-weighted images without requiring exogenous contrast agents. This endogenous contrast arises specifically from the melanin-bound iron, enabling visualization of regions vulnerable to degeneration in Parkinson's disease (PD). Common pulse sequences for imaging include T1-weighted turbo spin-echo (TSE) and gradient-recalled (GRE) with transfer (MT) preparation, typically acquired at . For TSE, standard parameters are repetition time () of 600 ms, time () of 10 ms, echo train length of 8, in-plane of 0.6 × 0.6 mm, and slice thickness of 2-3 mm over 15-20 axial slices centered on the . GRE sequences enhance volume coverage and incorporate MT pulses (e.g., Gaussian-shaped with flip angle 500°, duration 10 ms) to boost contrast, though they may use shorter (e.g., 55 ms) and (e.g., 2.2 ms) for efficiency. These parameters optimize the paramagnetic T1 shortening while minimizing artifacts from surrounding tissues. Clinically, imaging supports early diagnosis by quantifying signal loss and in the SNc and , markers of and noradrenergic degeneration. Signal intensity is often assessed via ratios (CR) or contrast-to-noise ratios (CNR) normalized to reference regions like or the pontine ; in , SNc CR decreases by approximately 20-30% compared to controls, correlating with disease severity. Volumetric analysis of the SNc further aids differentiation from , with thresholds below 440 mm³ indicating . measurement tracks progression, providing a noninvasive for monitoring therapeutic responses. Development of neuromelanin-sensitive sequences accelerated in the , building on postmortem validations of neuromelanin's MRI properties. Seminal in vivo demonstrations at were reported by Sasaki et al. in 2006 using TSE sequences, establishing reliable SNc visualization. Subsequent optimizations in the 2010s focused on reproducibility, MT weighting, and quantitative metrics like signal ratios. Recent 2020s advancements at 7T exploit higher field strength for improved and submillimeter resolution ( volumes ~0.24 mm³), enhancing LC depiction and diagnostic performance—achieving 100% sensitivity and 96.8% specificity for SNc volumetry in —while maintaining similar T1-based principles.

T1 Rho Imaging

T1ρ imaging, also known as spin-lattice relaxation in the , measures the relaxation of longitudinal in the presence of a spin-lock radiofrequency (RF) field, providing sensitivity to low-frequency molecular motions in tissues. The principle relies on applying a continuous or pulsed spin-lock field (B₁) along the after excitation, which locks the and suppresses dipolar interactions at low locking frequencies (typically 200–1000 Hz), making T1ρ relaxation time approximate at these low frequencies while being particularly sensitive to chemical exchange processes between protons and macromolecules like proteoglycans. This contrasts with standard T1 and relaxations by probing interactions on the order of the spin-lock , enabling detection of subtle biochemical changes not visible in conventional MRI. The pulse sequence begins with a 90° pulse to tip into the , followed by a spin-lock module consisting of (CW) RF pulses or composite pulses such as self-compensated spin-lock (SLR) trains to maintain the lock for varying durations (τ, or spin-lock time TSL). Multiple images are acquired at different τ values, often using multi-echo acquisition for signal averaging and fitting, with the signal intensity decaying mono-exponentially as S(\tau) = S_0 e^{-\tau / T_{1\rho}}, where S(\tau) is the observed signal, S_0 is the equilibrium signal, and T_{1\rho} is the relaxation time constant obtained via voxel-wise . For improved efficiency in 3D imaging, balanced steady-state free (bSSFP)-based sequences, such as T1ρ-prepared bSSFP, incorporate centric ordering and transient signal filtering to achieve rapid acquisitions (under 10 minutes for coverage) while preserving T1ρ contrast and reducing (SAR) concerns. T1ρ dispersion curves, plotting T1ρ values against varying spin-lock frequencies, reveal tissue-specific low-frequency relaxation behaviors influenced by macromolecular content, aiding in the characterization of pathological changes. Primary applications include quantifying early degeneration in (), where elevated T1ρ values (e.g., 55.4 ± 26.0 ms in advanced knee ) correlate with loss and outperform T2 in sensitivity; degeneration, detecting annulus fibrosus changes with steeper T1ρ gradients (e.g., -3.02 to -4.56); and amyloid detection in , with increased hippocampal T1ρ associated with and burden.00950-0/fulltext) Currently in transition from experimental to clinical use, T1ρ imaging faces challenges like limitations and scan time but benefits from acceleration techniques like parallel imaging, positioning it as a promising for early musculoskeletal and neurodegenerative pathologies without exogenous contrast.

References

  1. [1]
    Magnetic Resonance Imaging Physics - StatPearls - NCBI Bookshelf
    This sequence utilizes an initial 180 pulse instead of 90 pulses in the spin-echo sequence. This causes the initial inversion of the longitudinal ...
  2. [2]
    Optimization Methods for Magnetic Resonance Imaging Gradient ...
    Apr 27, 2020 · Magnetic resonance imaging (MRI) is driven by pulse sequences – a synchronized series of radiofrequency (RF) and gradient events that confer ...
  3. [3]
    The Basics of MRI
    A set of RF pulses applied to a sample to produce a specific form of NMR signal is called a pulse sequence. In the 90-FID pulse sequence, net magnetization is ...
  4. [4]
    [PDF] Basic Pulse Sequences I - Saturation & Inversion Recovery - UCLA
    Appreciate the definition of image contrast. • Explain what a T1 or T2-weighted image is. • Describe what a pulse sequence is. • Understand the saturation ...
  5. [5]
    Resonance Absorption by Nuclear Magnetic Moments in a Solid
    Phys. Rev. 69, 37 (1946) ... Resonance Absorption by Nuclear Magnetic Moments in a Solid. E. M. Purcell, H. C. Torrey, and R. V. Pound*.
  6. [6]
    Nuclear Induction | Phys. Rev. - Physical Review Link Manager
    It is shown that a radiofrequency field at right angles to the constant field causes a forced precession of the total polarization around the constant field ...
  7. [7]
    Tumor Detection by Nuclear Magnetic Resonance - Science
    Tumor Detection by Nuclear Magnetic Resonance. Raymond DamadianAuthors Info & Affiliations. Science. 19 Mar 1971 ... MRI Contrast Agents, Chemical & Biomedical ...
  8. [8]
    Spin Echoes in the Presence of a Time‐Dependent Field Gradient
    A derivation is given of the effect of a time‐dependent magnetic field gradient on the spin‐echo experiment, particularly in the presence of spin diffusion.
  9. [9]
    MR Imaging in the 21st Century: Technical Innovation over the First ...
    A major technical innovation for the clinical 3T MRI system was magnet technology with reasonable cost and the RF transmit, and receive system. In addition, ...
  10. [10]
    [PDF] RF Excitation RF Excitation - Center for Functional MRI
    Flip angle θ = ω1. 0 τ. / (s)ds where ω1(t) = γB1(t). Nishimura 1996 ... integrating the contributions of the field B1 τ( ) at the k - space point k τ,t.<|separator|>
  11. [11]
    Slice selection - Radiology Cafe
    This is known as “slice selection“. The way this is done is by using the RF pulse to select which slice to activate i.e. which slice will have the magnetic ...
  12. [12]
    Spatial encoding in MRI: selecting the slice plane - IMAIOS
    The RF wave associated with the slice selection gradient and the adapted resonance frequency, is called the selective pulse.
  13. [13]
    [PDF] Magnetic Field Gradients - LSA Course Sites
    x-gradient (Gx) y-gradient (Gy) z-gradient (Gz) ... Frequency Encoding. Page 36. Phase Encoding. Frequency. Encoding. Page 37. Signal Equation and Fourier MRI ...
  14. [14]
    Slice-selective excitation gradients - Questions and Answers ​in MRI
    All gradients should be drawn as trapezoids (or some similar shape) having a more gradual rate of rise and fall.
  15. [15]
    Free induction decay (FID) - Questions and Answers ​in MRI
    The FID is just one of four basic types of NMR signals produced in different ways. These are listed below and described in more detail in subsequent Q&A's.
  16. [16]
    Signal Generation
    A FID signal is formed by a single RF pulse and a spin echo by the interaction of two RF pulses. A stimulated echo, on the other hand, is formed by the ...Missing: Lauterbur | Show results with:Lauterbur
  17. [17]
    Spin echo (SE) - Questions and Answers ​in MRI
    In fact, spin echoes are formed when two successive RF-pulses of any flip angle are employed! Hahn, in his original paper, used two 90° pulses.Missing: Lauterbur | Show results with:Lauterbur
  18. [18]
    TR and TE - Questions and Answers ​in MRI
    TR and TE are basic pulse sequence parameters and stand for repetition time and echo time respectively. They are typically measured in milliseconds (ms).
  19. [19]
    The Basics of MRI
    Gradient Recalled Echo. S = k ρ (1-exp(-TR/T1)) Sinθ exp(-TE/T2*) / (1 -Cosθ exp(-TR/T1)) In each of these equations, S represents the amplitude of the signal ...Image Contrast · Image Processing
  20. [20]
    An Introduction to the Fourier Transform: Relationship to MRI | AJR
    The Fourier transform, a fundamental mathematic tool widely used in signal analysis, is ubiquitous in radiology and integral to modern MR image formation.
  21. [21]
    "Basic MR Relaxation Mechanisms & Contrast Agent Design" - PMC
    In this review we will detail the many important considerations when pursuing the design and use of MR contrast media.Missing: seminal | Show results with:seminal
  22. [22]
    A Comprehensive Introduction to Magnetic Resonance Imaging ...
    Oct 13, 2022 · This work reviews the origin of the image signal and contrast in MRI and the concepts of relaxometry and MRI contrast agents.Missing: seminal | Show results with:seminal
  23. [23]
    Bloch–Siegert B1-Mapping Improves Accuracy and Precision of ...
    This study evaluated the ability of Bloch–Siegert B 1 mapping to improve the accuracy and precision of VFA-derived T 1 measurements in the breast.Missing: T2 paper
  24. [24]
    T2 mapping in myocardial disease: a comprehensive review
    Jun 6, 2022 · Strong evidence supports the clinical use of T2 mapping in acute myocardial infarction, myocarditis, heart transplant rejection, and dilated cardiomyopathy.Missing: paper | Show results with:paper
  25. [25]
    Spin Echoes | Phys. Rev. - Physical Review Link Manager
    Reported at the Chicago Meeting of the American Physical Society, November, 1949 Phys. Rev. 77, 746 (1950) · F. Bloch, Phys. Rev. · E. L. Hahn, Phys. Rev. 77, 297 ...
  26. [26]
  27. [27]
  28. [28]
  29. [29]
  30. [30]
  31. [31]
    Inversion Recovery (IR) Sequence - MRI Questions
    Inversion recovery (IR) is a conventional spin echo (SE) sequence preceded by a 180° inverting pulse.
  32. [32]
  33. [33]
    Fat-Suppression Techniques for 3-T MR Imaging of the ...
    The major clinical limitations of STIR sequences are their relatively long imaging times, low SNR, and high SAR. Long imaging times may be reduced by using a ...
  34. [34]
    Fluid attenuated inversion recovery (FLAIR) MRI at 7.0 Tesla - NIH
    FLAIR imaging is nowadays a robust sequence at 1.5 T and 3 T, the clinically most commonly used field strengths [4]. Recently, ultra-high field strength magnets ...
  35. [35]
    Optimal combination of FLAIR and T2-weighted MRI for improved ...
    The main objective of this study is to investigate an optimal combination of T2w and FLAIR to improve the conspicuity of MS lesions.
  36. [36]
    Optimizing the Magnetization-Prepared Rapid Gradient-Echo (MP ...
    May 30, 2014 · The optimal parameters are flip angle of 12°, effective inversion time within 900 to 1100 ms, and delay time of 0 ms.The Mp-Rage Sequence · Optimizing Imaging... · Results
  37. [37]
    Radiopaedia.org
    **Summary of Magnetisation-Prepared Rapid Acquisition of Gradient Echoes (MPRAGE):**
  38. [38]
    Double inversion recovery sequence | Radiology Reference Article
    Nov 29, 2021 · The technique can be used to suppress signal from two different tissues or to suppress signal that moved between the two pulses. In the first ...
  39. [39]
    Utility of Double Inversion Recovery Sequences in MRI - PMC
    Aug 30, 2016 · DIR sequences suppress both cerebrospinal fluid (CSF) signal and normal white matter signal. This may make abnormal white matter associated with ...
  40. [40]
    Deep Learning-Based Prediction of PET Amyloid Status Using MRI
    Jun 27, 2025 · ... inversion recovery; FBP= 18 F-florbetapir; FBB= 18 F ... Materials and methods: Multi-contrast MRI and PET-based quantitative ...
  41. [41]
    Diffusion weighted imaging: Technique and applications - PMC
    Diffusion weighted imaging (DWI) is a method of signal contrast generation based on the differences in Brownian motion.
  42. [42]
    Evaluating corrections for Eddy‐currents and other EPI distortions in ...
    To propose a methodology for assessment of algorithms that correct distortions due to motion, eddy‐currents, and echo planar imaging in diffusion weighted ...
  43. [43]
    Arterial spin labeling for the measurement of cerebral perfusion and ...
    Arterial spin labeling (ASL) is an MRI technique that was first proposed a quarter of a century ago. It offers the prospect of non-invasive quantitative ...
  44. [44]
    Consensus recommendations for a dynamic susceptibility contrast ...
    After reviewing the basis for controversy over DSC-MRI protocols, this paper provides evidence-based best practices for clinical DSC-MRI as determined by the ...Missing: seminal | Show results with:seminal
  45. [45]
  46. [46]
    and DCE-MRI Parameters in Brain Tumor Patients: Theory and ... - NIH
    Dynamic contrast-enhanced (DCE) and dynamic susceptibility contrast (DSC) magnetic resonance imaging (MRI) are the perfusion imaging techniques most frequently ...Missing: seminal | Show results with:seminal
  47. [47]
    Dynamic contrast-enhanced magnetic resonance imaging
    DCE-MRI possesses an unparalleled capacity to quantitatively measure not only perfusion but also other diverse microvascular parameters such as vessel ...
  48. [48]
  49. [49]
    Deconvolution-Based CT and MR Brain Perfusion Measurement
    Deconvolution-based analysis of CT and MR brain perfusion data is widely used in clinical practice and it is still a topic of ongoing research activities.
  50. [50]
  51. [51]
  52. [52]
    Recommendations for quantitative cerebral perfusion MRI using ...
    Apr 9, 2024 · This guidelines article provides a comprehensive overview of quantitative perfusion imaging of the brain using multi-timepoint arterial spin ...
  53. [53]
    Deeply Accelerated Arterial Spin Labeling Perfusion MRI for ... - NIH
    Dec 5, 2023 · In this paper, we proposed a deep learning-based algorithm to reduce the number of PLDs and to accurately estimate ATT and CBF.Missing: AI 2020s
  54. [54]
  55. [55]
    Perfusion MRI: The Five Most Frequently Asked Clinical Questions
    Perfusion MRI is a promising tool in assessing stroke, brain tumors, and neurodegenerative diseases. Most of the impediments that have limited the use of ...
  56. [56]
  57. [57]
    Perfusion imaging in brain disease - ScienceDirect.com
    Perfusion MRI should be included in the assessment of any brain tumor, both at the time of the diagnosis as well as in the post-treatment monitoring. It is ...
  58. [58]
    Dynamic magnetic resonance imaging of human brain activity ...
    Dynamic magnetic resonance imaging of human brain activity during primary sensory stimulation. K K Kwong ... ArticleJune 15, 1992. Studies of the cloned 37-kDa ...Missing: first | Show results with:first
  59. [59]
    Functional connectivity in the motor cortex of resting human brain ...
    First published: October 1995 ... Regions of the sensorimotor cortex that were activated secondary to hand movement were identified using functional MRI ...
  60. [60]
    [PDF] Functional BOLD MRI (fMRI) sequence settings - Dartmouth
    Echo time (TE) and Repetition Time (TR). Echo Time ranges between 30ms to 35ms for BOLD scans of the whole brain. Value of TR depends on how many slices the ...
  61. [61]
  62. [62]
    Review The general linear model and fMRI: Does love last forever?
    In this review, we first set out the general linear model (GLM) for the non technical reader, as a tool able to do both linear regression and ANOVA within ...
  63. [63]
    Statistical parametric maps in functional imaging: A general linear ...
    Friston KJ, Worsley KJ, Frackowiak RSJ, Mazziotta JC, Evans AC (1994): Assessing the significance of focal activations using their spatial extent. Hum Brain ...Missing: fMRI | Show results with:fMRI
  64. [64]
    Vascular magnetic resonance angiography techniques - PMC
    In TOF MRA, repetitive RF pulses saturate the spin magnetization of the stationary structures in the imaging field, resulting in background signal suppression.
  65. [65]
    Accelerated Time-of-Flight Magnetic Resonance Angiography ... - NIH
    INTRODUCTION. Time-of-flight (TOF) magnetic resonance angiography (MRA) is a non-invasive, non-contrast-enhanced MRA technique that provides contrast between ...
  66. [66]
    Optimization of MR Parameters of 3D TOF-MRA for Various ... - NIH
    At first, we obtained three-dimensional (3D) TOF-MRA with parameters as follows: Spoiled gradient-echo sequence with TR/TE/FA, 24 msec/2.7 msec/20°; FOV/Matrix, ...
  67. [67]
    Clinical application of ultra-high resolution compressed sensing time ...
    Cerebral TOF MRA selectively visualizes arterial blood vessels by positioning a saturation slab superior to the imaging slab, thereby saturating the signal ...
  68. [68]
    Cardiovascular magnetic resonance phase contrast imaging - PMC
    When |v⊥i| exceeds VENC, velocity aliasing will occur. That means that ... Four-dimensional phase contrast MRI with accelerated dual velocity encoding.
  69. [69]
    Cardiovascular Applications of Phase-Contrast MRI
    Sep 27, 2008 · The peak velocity encoding (VENC) value and its sensitivity and direction are defined by the user. The amplitudes of the flow-sen- sitizing ...<|separator|>
  70. [70]
    Clinical Applications of Magnetic Resonance Angiography - PubMed
    We describe several case reports involving the use of MRA and discuss its advantages in evaluating patients for carotid artery stenosis, intracerebral aneurysms ...
  71. [71]
    Velocity-Selective Magnetization-Prepared Non-Contrast-Enhanced ...
    Moreover, TOF-MRA is prone to saturation of slowly flowing blood protons, resulting in signal loss within a severely compromised lumen or distal small arteries ...
  72. [72]
    [PDF] Susceptibility-Weighted Imaging: Technical Aspects and Clinical ...
    Haacke EM, Herigault G, Yu Y, et al. Observing tumor vascularity noninva ... , susceptibility-weighted magnetic resonance imaging of the brain: dose ...
  73. [73]
    Morphology Enabled Dipole Inversion for Quantitative Susceptibility ...
    Quantitative susceptibility mapping (QSM) of tissue requires the deconvolution of this measured magnetic field with the dipole kernel. This inverse problem is ...
  74. [74]
    Validation of deep-learning accelerated quantitative susceptibility ...
    Jan 22, 2025 · Abstract. Purpose. To test the feasibility and consistency of a deep-learning (DL) accelerated QSM method for deep brain nuclei evaluation.
  75. [75]
    [PDF] Magnetization transfer in MRI: a review
    As shown in. Fig. 1, it is possible to saturate the macromolecular spins preferentially using an off-resonance radio frequency pulse. The macromolecular spins ...
  76. [76]
    Magnetization Transfer - Questions and Answers ​in MRI
    This off-resonance pulse has sufficient power to saturate protons in the immobile pool without directly affecting those in free water. Following this MT ...
  77. [77]
    [PDF] Magnetization Transfer Basics - ISMRM
    The MT effect causes additional magnetization decrease (solid line MMT). The magnetization transfer ratio, MTR is a sum of Mdir and MMT contributions.
  78. [78]
    Magnetization Transfer - Quantitative MRI mOOC - qMRLab
    Oct 7, 2024 · A simple method for visualizing MT effects includes acquiring two images with and without an off-resonance MT pulse to calculate the MT ratio ( ...Missing: indirect | Show results with:indirect
  79. [79]
    Myelin Measurement Using Quantitative Magnetic Resonance Imaging
    A widely used myelin imaging method is based on magnetization transfer (MT), which is a physical process by which macromolecules and their closely associated ...
  80. [80]
    Quantitative magnetization transfer imaging in relapsing-remitting ...
    Apr 4, 2022 · Myelin-sensitive MRI such as magnetization transfer imaging has been widely used in multiple sclerosis. The influence of methodology and ...
  81. [81]
    Dixon techniques for water and fat imaging - Ma - Wiley Online Library
    Sep 5, 2008 · The primary objective of the Dixon techniques is to determine W and F from an acquired image or images that are represented by S in Equation [1] ...Missing: exp( | Show results with:exp(
  82. [82]
    Fat suppression techniques in breast magnetic resonance imaging
    Mar 16, 2015 · This review offers a critical comparison of various fat suppression techniques in breast MRI including spectral-selective excitation and saturation techniques.Missing: orthopedics | Show results with:orthopedics
  83. [83]
    Neuromelanin detection by magnetic resonance imaging (MRI) and ...
    Apr 10, 2018 · We discuss how MRI detects NM and how this approach might be improved. We suggest that MRI of NM can be used to confirm PD diagnosis and monitor disease ...
  84. [84]
  85. [85]
    [PDF] Normative Values of Neuromelanin‐Sensitive MRI Signal in Older ...
    NM-MRI images were collected via a turbo spin echo (TSE) sequence with the following parameters: repetition time (TR) = 600 msec; echo time (TE) = 10 msec; flip ...
  86. [86]
    Neuromelanin-sensitive MRI as a promising biomarker of ...
    MRI studies with NM-MRI sequences have consistently demonstrated the presence of a significant reduction of NM-related signal in the SN (SNc-VTA) and LC in ...
  87. [87]
    Reproducibility assessment of neuromelanin-sensitive magnetic ...
    Neuromelanin-sensitive MRI (NM-MRI) provides a noninvasive measure of the content of neuromelanin (NM), a product of dopamine metabolism that accumulates ...
  88. [88]
    Evaluation of Parkinson Disease and Alzheimer Disease with the ...
    Nov 1, 2013 · The signal intensity ratios in substantia nigra pars compacta and locus ceruleus on neuromelanin MRI positively correlated with the heart-to- ...
  89. [89]
  90. [90]
    Diagnostic utility of 7T neuromelanin imaging of the substantia nigra ...
    Jan 8, 2024 · Neuromelanin (NM) imaging is a promising biomarker for PD, but adoption has been limited, in part due to subpar performance at standard MRI ...
  91. [91]
    T1ρ magnetic resonance: basic physics principles and applications ...
    Dec 6, 2015 · This paper reviews the T 1ρ imaging's basic physic principles, as well as its application for cartilage imaging and intervertebral disc imaging.
  92. [92]
    T1ρ MR Imaging of Human Musculoskeletal System - PMC - NIH
    In the following sections of this article, we will review the basic principles of T1ρ MR imaging, implementation of different T1ρ pulse sequences, biochemical ...
  93. [93]
    T1ρ-Prepared Balanced Gradient Echo for Rapid 3D T1ρ MRI - NIH
    3D T1ρ maps of the human spine images are shown in Figure 9. The pulse sequence was modified for imaging of the spine to further reduce blurring. A segmented, ...Pulse Sequence · Figure 2 · Human Imaging Protocol
  94. [94]
    Comparison of T1rho MRI, Glucose Metabolism and Amyloid Burden ...
    Higher T1rho values were associated with a clinical diagnosis of cognitive impairment (i.e.,MCI or AD diagnosis), poorer neuropsychological test performance ( ...
  95. [95]
    MR-Imaging in Osteoarthritis: Current Standard of Practice and ... - NIH
    Aug 3, 2023 · T1-Rho Mapping. T1rho relaxation is also known as spin-lattice relaxation in the r(h)otational frame. In contrast to conventional T1 ...2. Osteoarthritis--A Whole... · 3.4. Fat Suppression · 6.3. T2 Mapping