Viscometer

A viscometer is an instrument designed to measure the of a , defined as the 's internal resistance to arising from intermolecular forces. quantifies this property as the ratio of (force per unit area) to (velocity gradient between adjacent layers). These devices apply controlled deformation to a sample and record the resulting force or motion, enabling precise characterization essential for understanding behavior under various conditions. Viscometers are categorized by their operational principles, broadly including , rotational, and falling-body types, each suited to different viscosities and conditions. viscometers determine kinematic by timing the of through a narrow under gravity, following Poiseuille's law for . Rotational viscometers, such as the Brookfield model, measure dynamic by detecting the needed to rotate a or geometry (e.g., cone-and-plate or ) within the sample, providing data on -dependent behavior. Falling-body viscometers assess from the terminal velocity of a or descending through the , ideal for opaque or high- samples. These instruments are indispensable across industries, including pharmaceuticals for drug formulation stability, for texture control, for and quality assessment, and for product consistency. In , they evaluate rheological properties of polymers and nanofluids to optimize processes. against standard reference materials, guided by international standards like ISO 3219 for rotational methods, ensures accuracy and . Advances in viscometry, such as MEMS-based and ultrasonic variants, continue to enhance portability and precision for real-time industrial monitoring.

Introduction

Definition and Purpose

A viscometer is an instrument designed to measure the of , quantifying their internal or to flow. represents a 's to deformation under an applied , arising from intermolecular forces that oppose relative motion between layers. This is expressed in units such as the poise (P) in the centimeter-gram-second (CGS) system or the pascal-second (Pa·s) in the (). Viscometers serve a vital purpose in assessing fluid behavior under shear, enabling precise characterization essential for various applications. In manufacturing, they ensure quality control by verifying consistency in products like lubricants and coatings, where deviations in viscosity can impact performance and processing. In the pharmaceutical sector, viscosity measurements guide formulation development for injectables, suspensions, and gels, influencing drug stability, delivery, and bioavailability. In fluid dynamics research, they provide data on flow properties critical for modeling complex systems, such as in petroleum reservoir analysis or material science simulations. Viscometers differ from rheometers, which evaluate a wider array of rheological behaviors including and yield stress across varying conditions, whereas viscometers target primarily steady-state under controlled . This focused capability makes viscometers suitable for routine, high-throughput testing in industrial settings.

Historical Development

The concept of as a measure of resistance to flow was first formalized by in his 1687 work , where he described the linear relationship between and for certain fluids, laying the groundwork for quantitative viscosity measurements. Earlier in the , contributed to the understanding of material deformation through his 1678 law relating stress and in solids, which indirectly influenced later rheological studies, though his work focused more on behavior than flow. In 1851, George Gabriel Stokes derived the drag force equation on a moving through a viscous fluid while studying motion in air, providing the mathematical foundation for the falling- viscometer, a practical for measuring by observing the terminal velocity of a in a column. This method, later refined and widely adopted, marked the advent of empirical viscometry tools. By the early , advancements in designs emerged, with chemist Leo Ubbelohde inventing a precise viscometer around 1910 to address limitations in earlier models like the Ostwald type, enabling accurate kinematic determinations across a wide range by minimizing corrections and effects. The 1930s saw the development of rotational viscometers, pioneered by Don Brookfield Sr. in 1934 with his dial-reading synchro-electric instrument, which measured on a rotating immersed in the and became a for applications like testing. Vibrational viscometers appeared in the mid-, with Bendix introducing an in-line model using a vibrating blade in the for process monitoring, evolving in the late to incorporate technologies for higher sensitivity in low- . Standardization efforts began with ASTM D445 in 1937 for kinematic testing using viscometers, building on earlier 1920s petroleum , while ISO established the ISO VG classification in 1975 to harmonize global grading. From the 1980s onward, and automated viscometers proliferated, integrating microprocessors for precise control of rates and output, as seen in controlled-stress rheometers from companies like Carrimed, enhancing accuracy for non-Newtonian fluids in industrial settings. Post-2010, microfluidic viscometers advanced rapidly, enabling low-volume (e.g., 10 μL) measurements with portable, single-use chips for biopharmaceuticals, such as solutions up to 320 mg/mL, validated against traditional rheometers with high correlation (R² = 0.96).

Fundamentals of Viscosity Measurement

Viscosity Concepts

is a measure of a fluid's resistance to deformation under , and dynamic , denoted as η, quantifies this property for Newtonian fluids as the constant ratio of τ to \dot{\gamma}, expressed by the equation \tau = \eta \dot{\gamma}. This relationship holds for fluids where the remains independent of the applied , such as or air at typical conditions. Kinematic viscosity, denoted as ν, is defined as the dynamic viscosity divided by the fluid density ρ, so ν = η / ρ, and it is particularly useful in scenarios involving gravitational flow where density effects are significant. Kinematic viscosity is often determined experimentally by measuring the time required for a to flow through a tube under , as this time is directly proportional to ν. The value of viscosity is influenced by several factors, including temperature, which typically causes a decrease in viscosity for liquids according to the : \eta = A \exp\left(\frac{E_a}{RT}\right) where A is a , E_a is the for viscous flow, R is the , and T is the absolute temperature. Pressure generally increases viscosity, particularly at high levels, while the chemical composition of the fluid—such as molecular weight or the presence of additives—can significantly alter its magnitude. In the (SI), dynamic is measured in pascal-seconds (·s), equivalent to N·s/m², while the centimeter-gram-second (cgs) system uses the poise (P), where 1 P = 0.1 ·s. Kinematic in SI units is square meters per second (m²/s), and in cgs, it is the stokes (St), with 1 St = 10^{-4} m²/s. For non-Newtonian fluids, where depends on , the term refers to the instantaneous ratio of to , η_app = τ / \dot{\gamma}, which varies with conditions rather than being .

Newtonian and Non-Newtonian Fluids

Newtonian fluids are characterized by a constant viscosity that remains independent of the applied shear rate at a given temperature and pressure. This linear relationship between shear stress and shear rate follows from the Newtonian postulate, making such fluids straightforward to measure using basic viscometer geometries. Common examples include water, air, and low-molecular-weight oils like mineral oil. In contrast, non-Newtonian fluids display viscosity that varies with shear rate, shear stress, or time, leading to complex flow behaviors that deviate from the linear stress-strain rate relationship. These fluids are prevalent in industrial and biological contexts, requiring specialized rheological analysis to understand their properties fully. Non-Newtonian fluids are categorized into several subtypes based on their response to . Shear-thinning, or pseudoplastic, fluids exhibit decreasing with increasing ; examples include paints, solutions, and . Shear-thickening, or , fluids show the opposite behavior, with increasing under higher ; representative cases are cornstarch slurries and certain clay suspensions. Bingham plastics behave as solids below a critical yield stress but flow as fluids above it, such as and drilling muds. Additionally, thixotropic fluids demonstrate time-dependent shear-thinning, where decreases progressively under sustained and recovers upon rest; examples encompass certain printing inks and synovial fluids. A widely used mathematical model for many time-independent non-Newtonian fluids is the power-law model, which describes the relationship between shear stress \tau and shear rate \dot{\gamma}: \tau = K \dot{\gamma}^n Here, K represents the consistency index, indicating the fluid's viscosity scale, and n is the flow behavior index, where n < 1 characterizes shear-thinning behavior and n > 1 indicates shear-thickening. This model provides a simple yet effective framework for approximating the rheological response of such fluids across a range of shear conditions. The measurement of non-Newtonian fluids in viscometers necessitates applying a spectrum of rates to map the profile accurately, as a single may yield misleading results due to the non-linear dependence. This approach is essential for characterizing fluids like polymers, , and products, where shear-dependent properties directly influence processing and performance.

Capillary Viscometers

U-Tube Viscometers

U-tube viscometers are a type of viscometer that measure the kinematic of Newtonian liquids by determining the time required for a to under through a calibrated tube. These instruments are widely used for their simplicity and precision in applications such as testing and solution analysis. The design features a U-shaped glass tube with two reservoirs connected by a narrow capillary bore, typically 0.3 to 3 mm in diameter, allowing laminar flow under hydrostatic pressure. The upper reservoir includes a filling bulb and timing marks (M1 and M2) etched on the capillary or adjacent sections to define the volume of liquid measured. Variants address specific sample requirements: the Ostwald viscometer uses a simple U-tube with bulbous extensions but is sensitive to filling volume due to variable hydrostatic pressure; the suspended Ubbelohde viscometer incorporates a leveling bulb and venting tube to make measurements independent of fill volume, ideal for volatile or foaming samples; and the Cannon-Fenske viscometer employs a reverse-flow configuration with a riser tube and multiple marks, suitable for transparent oils and opaque liquids. In operation, the viscometer is cleaned, dried, and filled with the test liquid to above the upper , ensuring no air bubbles are trapped. The liquid is then allowed to flow by through the at a precisely controlled , typically in a constant-temperature maintained at 20°C or 40°C with ±0.02°C accuracy, while the efflux time t between the timing marks is recorded using a or automated detector. Measurements are repeated at least twice, with results averaged if they agree within specified precision limits, such as 0.37% for oils at 40°C. is critical, as varies significantly with , and the instrument must be vertically suspended to ensure consistent flow. Kinematic viscosity \nu (in mm²/s) is calculated as \nu = C t, where C is the instrument's constant determined using standard reference fluids, and t is the flow time in seconds; for short flow times below 200 s, a correction may be applied as \nu = C (t - k), with k being the Hagenbach-Couette correction factor. Dynamic \eta (in mPa·s) is then obtained by multiplying \nu by the fluid density \rho (in g/cm³): \eta = \nu \rho. follows standards like ASTM D446, ensuring accuracy across viscosity ranges from 0.2 to 300,000 mm²/s.
VariantKey Design FeatureSuitable For
OstwaldSimple with bulbous reservoirsTransparent Newtonian fluids like solutions; sensitive to fill volume
UbbelohdeSuspended level with venting tubeVolatile or foaming samples; volume-independent measurements
Cannon-FenskeReverse-flow with riser tubeTransparent oils and opaque liquids per ASTM D445
These viscometers offer advantages in simplicity, low cost, and high accuracy for low- Newtonian fluids, achieving precisions better than 0.2% when properly calibrated. However, they are limited to transparent or semi-transparent samples (except Cannon-Fenske for opaques), require manual handling that can introduce errors from air bubbles or improper filling, and are unsuitable for non-Newtonian fluids where shear-dependent behavior invalidates the constant C. Additionally, flow times must exceed 200 s to minimize end-effect corrections, and the method demands rigorous temperature stability to avoid errors exceeding 1% per °C.

Slit and Bubble Variants

The rectangular-slit viscometer is a modified design featuring two parallel plates separated by a narrow rectangular slit, typically with a width w much larger than the height h, allowing to under controlled . Viscosity is determined by measuring the \Delta P across the slit length L and the Q, using the relation derived from the Hagen-Poiseuille for in a rectangular channel: \eta = \frac{\Delta P \, w \, h^3}{12 \, Q \, L} This setup enables precise calculation of dynamic \eta for Newtonian s, with corrections such as the Rabinowitsch-Mooney applied for non-Newtonian behaviors to account for shear-rate dependence. In operation, the slit viscometer is often integrated into high-pressure systems for online monitoring, where is pumped through the slit at varying rates while transducers record \Delta P and Q, facilitating real-time assessment under industrial process conditions like or . This design excels in applications, such as melt processing, where it provides data on shear-dependent to optimize and prevent equipment issues. Advantages of the rectangular-slit viscometer include its ability to handle non-Newtonian fluids more effectively than circular capillaries by directly measuring wall \sigma_w = \frac{h \Delta P}{2 L} and applying shear-rate corrections, which is crucial for complex fluids exhibiting shear-thinning or -thickening. It supports high shear rates up to $10^6 \, \mathrm{s}^{-1}, mimicking industrial conditions, and requires smaller sample volumes compared to traditional setups. The viscometer, a distinct type of kinematic viscometer using an empirical flow-based measurement, assesses kinematic by timing the rise of an air through a vertical column of the test in a graduated . The empirical relation links the t (in seconds, often called "bubble seconds") approximately to in stokes, calibrated against standard liquids at a fixed like 25°C. Operationally, the tube is filled to marked lines (e.g., 27 mm to 100 mm), inverted to release the bubble, and the travel time between lines is recorded, providing a quick, comparative assessment without complex equipment. This method, standardized under ASTM D1545 for transparent liquids, is particularly suited for field tests of lubricants in settings, where rapid checks ensure compliance with specifications for oils and fuels. Bubble viscometers offer simplicity and portability for on-site use, requiring no power source and minimal training, making them ideal for opaque or translucent fluids like certain lubricants where precise timing yields repeatable results after . Their low cost and ease of cleaning further enhance utility in industrial environments, though they are limited to approximate measurements and require for accuracy.

Gravity-Driven Viscometers

Falling-Sphere Viscometers

Falling-sphere viscometers measure the dynamic of transparent by observing the terminal velocity of a falling through the sample under gravity, relying on the principles of described by . The instrument typically consists of a vertical or slightly inclined filled with the test , into which a —often made of , , or another material of precisely known and —is released from rest at the top. The tube is transparent to allow visual or optical tracking of the 's descent over a marked distance, ensuring the remains at a constant temperature to maintain accurate conditions. In operation, the sphere initially accelerates due to the net gravitational force but quickly reaches a constant terminal velocity v_t once the viscous drag balances the buoyant and gravitational forces, typically after a short acceleration phase. Measurements are taken only during this steady-state fall to avoid inertial effects, with the flow regime confined to laminar conditions by maintaining a low Reynolds number (Re < 0.2) based on the sphere's velocity, radius, and fluid properties. The time for the sphere to traverse a fixed distance is recorded, often averaged over multiple trials, and the setup requires bubble-free fluid and precise sphere centering to minimize errors. The viscosity \eta is calculated directly from the terminal velocity using Stokes' law for the drag force on a sphere in a viscous medium: \eta = \frac{2}{9} \frac{(\rho_s - \rho_f) g r^2}{v_t} where \rho_s is the sphere's density, \rho_f is the fluid's density, g is gravitational acceleration, and r is the sphere's radius. This formula assumes an infinite fluid medium and negligible inertia, providing an absolute measurement without requiring calibration if all parameters are accurately known. Advantages of falling-sphere viscometers include their simplicity, high accuracy for Newtonian fluids of medium viscosity (typically 0.5 mPa·s to 10^5 mPa·s), and ability to yield absolute viscosity values independent of instrument constants. They are particularly suited for transparent liquids where optical observation is feasible and do not require complex mechanical agitation. The Höppler viscometer represents a widely adopted variant, featuring an inclined tube (typically at 80°) to facilitate the sphere's motion and using multiple standardized balls for different viscosity ranges, with manual timing but potential for semi-automation in modern setups. However, proximity to the tube walls introduces a retarding effect on the sphere's velocity, necessitating correction via , which adjusts the observed velocity by a factor C = 1 - 2.104(d/D) + 2.09(d/D)^3 - 0.95(d/D)^5, where d is the sphere diameter and D is the tube diameter; this correction is essential for ratios d/D > 0.05 but may overcompensate in certain high-temperature or non-ideal fluids.

Falling-Piston Viscometers

Falling-piston viscometers, also known as Norcross viscometers after their inventor Austin Norcross, consist of a vertical cylindrical tube filled with the sample fluid and a weighted piston that fits closely within the cylinder, creating a narrow annular gap. The piston, typically made of stainless steel or similar durable material, has a mass calibrated for the expected viscosity range and features guides to ensure concentric motion. As the piston descends under gravity, it displaces the fluid through the clearance between its outer surface and the inner wall of the cylinder, generating shear in the process. This design allows for precise control of the flow path, making it particularly effective for measuring dynamic viscosity in confined geometries. In operation, the is mechanically or magnetically lifted to an initial position at the top of the column, then released to fall freely over a predetermined marked on the . The time required for the to travel this is recorded, often using non-contact sensors such as magnetic switches or optical detectors that sense the piston's position without requiring visibility through the sample. Modern automated versions, like those from Norcross, integrate electronic timing and to maintain isothermal conditions, ensuring for samples tested at specified temperatures. The fall time directly correlates with the 's resistance to flow, providing a transient under gravity-driven . Viscosity is calculated from the fall time using a formula derived from Poiseuille's law adapted for annular flow in the gap, accounting for the of the : \eta = \frac{m g h}{8 \pi L t r^4} where \eta is the dynamic , m is the , g is , h is the fall , L is the effective length of the , t is the fall time, and r is the of the annular gap; corrections for end effects and exact are applied based on . This balances the gravitational force driving the against the viscous drag opposing it. These viscometers offer advantages for testing opaque or high-viscosity fluids, such as lubricating oils, where optical methods fail, as detection relies on the piston's physical passage rather than through the sample. Their rugged and make them suitable for and settings requiring reliable, low-maintenance measurements. As a piston-specific variant of gravity-driven methods analogous to the Saybolt viscometer's time-based efflux principle, they are employed for characterizing viscosities in products like engine oils, supporting in formulations with viscosities up to several thousand centipoise.

Oscillatory and Vibrational Viscometers

Oscillating-Piston Viscometers

Oscillating-piston viscometers feature a cylindrical chamber filled with the sample , containing a magnetically actuated that oscillates horizontally within the enclosed volume. The design typically includes electromagnetic coils surrounding the chamber to drive the piston's motion, with the annular space between the piston and chamber walls ensuring consistent application across the . This configuration allows for precise control of the and , often in the range of 1-10 Hz, while minimizing external influences such as or settling effects. In operation, the piston is driven at a fixed by an alternating , inducing oscillatory motion through the fluid. The viscous resistance of the fluid affects the speed of this motion, which is quantified by measuring the time required for the piston to travel a fixed distance using position sensors such as or optical detectors. This method adheres to standards like ASTM D7483, which specifies procedures for dynamic determination using such instruments. Viscosity is computed from the measured travel time using instrument-specific calibration constants that account for geometric factors and are established against known fluid standards. This approach follows from the balance of viscous drag forces in the annular gap, calibrated for absolute measurement. Commercial implementations, such as those from Cambridge Viscosity (e.g., ViscoLab series), integrate this computation electronically for direct output in centipoise. These viscometers offer advantages including minimal sample volume requirements (often 0.2-1 ), making them suitable for precious or limited samples, and robust performance with non-Newtonian fluids due to the controlled rates (typically 10-1000 s⁻¹). The enclosed design renders them insensitive to air bubbles or , enhancing reliability for opaque or translucent liquids. They find specific applications in the and pharmaceutical industries, such as viscosity monitoring of creams, lotions, and emulsions, where models like the ViscoPro series ensure process control and .

Quartz and Tuning-Fork Viscometers

Quartz crystal microbalances (QCMs) and tuning-fork viscometers utilize high-frequency or low-frequency vibrational principles to assess fluid through alterations in caused by viscous loading. These devices offer precise, contact-based measurements ideal for low-viscosity liquids and gases, enabling real-time analysis in and settings. By immersing a piezoelectric in the sample, they detect shifts in frequency or , which correlate with the fluid's hydrodynamic interactions at the sensor-liquid interface. In QCM design, a thin AT-cut disk, typically 0.17 mm thick with or silver electrodes, serves as the sensing element, excited piezoelectrically to oscillate in shear mode at fundamental frequencies of 5–10 MHz. Upon , the surrounding fluid exerts viscous drag, reducing the resonance frequency and broadening the peak width due to energy dissipation in the viscous . This setup allows for sensitive detection of mass and viscous effects, with the crystal's one-sided liquid exposure minimizing air-side interference. QCM operation involves applying an oscillating voltage to drive the crystal, while impedance analysis or frequency sweeps identify the resonance parameters. The frequency shift \Delta f arises from the fluid's shear wave penetration, and for low-viscosity Newtonian liquids, it follows the Kanazawa equation: \Delta f = -\frac{f_0^{3/2} \sqrt{\eta_f \rho_f}}{\sqrt{\pi \rho_q \mu_q}} where f_0 is the frequency in air, \eta_f and \rho_f are the fluid's and , \rho_q is (2.648 g/cm³), and \mu_q is (5.9 × 10^{10} g/(cm·s²)). is then derived if is independently known by rearranging for \eta_f \rho_f = \left( \frac{ -\Delta f \sqrt{\pi \rho_q \mu_q} }{f_0^{3/2}} \right)^2. Bandwidth broadening \Delta \Gamma further confirms viscous contributions via \Delta \Gamma \approx |\Delta f|. To arrive at the Kanazawa equation, start with the hydrodynamic wave equation in the liquid, \frac{d^2 u}{dz^2} = \frac{i \omega \rho_f}{\eta_f} u, solving for the velocity decay and matching boundary conditions at the crystal surface to the quartz's , yielding the frequency perturbation proportional to the fluid's shear impedance \sqrt{\eta_f \rho_f / (i \omega)}; for small loads, this simplifies to the given form after substituting properties and dependencies. QCM viscometry emerged in the mid-1980s, with and deriving the liquid-loading model that enabled practical use beyond applications. Commercial implementations, such as the QCM 200 from Stanford Research Systems, support inline monitoring with sub-ng/cm² sensitivity. Key advantages include high resolution for viscosities below 1 mPa·s, suitability for gases and thin films, and integration with for biochemical assays, though care is needed to avoid temperature-induced frequency drifts. Tuning-fork viscometers feature a U-shaped metallic probe, often , driven by piezoelectric actuators to vibrate laterally at low resonance frequencies of about 30 Hz, with prong amplitudes of 0.1–1 mm. The fork's immersion in fluid encounters resistive forces that dampen motion, requiring compensatory drive energy to maintain constant . This design supports robust, probe-style insertion into process lines without requiring transparent samples. During operation, the sensor oscillates sinusoidally, and electronic feedback adjusts voltage or current to stabilize displacement; the power input reflects the fluid's damping via mechanical impedance R_z = \pi \eta_f \rho_f f A, where f is vibration frequency and A is the effective wetted area. Viscosity is computed from this resistance, often alongside density from frequency shifts, enabling absolute measurements over shear rates of 10–2000 s⁻¹ by modulating amplitude. Advanced models distinguish Newtonian from non-Newtonian behavior through torque-viscosity profiles. To derive viscosity, calibrate the drive signal against known standards, relating observed impedance to \eta_f = R_z / (\pi \rho_f f A), assuming known density and geometry; iterative fitting refines for complex fluids. Tuning-fork technology for viscometry gained commercial traction in the , with A&D Company pioneering the SV series around 2000, achieving inclusion in JIS Z 8803:2011 standardization for methods. Instruments like the Micro Motion FVM provide multivariable outputs (, , ) in ranges up to 20,000 . Benefits encompass wide dynamic range (0.3–10,000 mPa·s), insensitivity to or bubbles, and seamless inline deployment for continuous monitoring in demanding environments like oil refining and .

Rotational Viscometers

Cone-and-Plate and Parallel-Plate Geometries

In rotational viscometers, the cone-and-plate geometry consists of a shallow positioned above a flat plate, with the sample filling the narrow gap between them. The cone angle \theta is typically small, less than 3°, to approximate simple shear conditions. The gap height h at a radial position r is given by h = r \theta, where \theta is in radians, ensuring a linearly increasing separation from the center. This configuration produces a \dot{\gamma} = \Omega / \theta that is constant and independent of r, with \Omega denoting the of the rotating cone. The viscosity \eta in the cone-and-plate setup is calculated using the measured torque M via the form factor relation \eta = \frac{3 M \theta}{2 \pi r^3 \Omega}, which accounts for the uniform distribution across the geometry. This formula derives from integrating the contributions over the plate radius r, assuming Newtonian behavior for direct computation. For non-Newtonian fluids, the constant simplifies data analysis by providing well-defined flow conditions. In contrast, the parallel-plate geometry employs two flat plates separated by a fixed h, with the sample sheared by of one plate. The varies radially as \dot{\gamma} = (r \Omega)/h, increasing linearly from the center to the edge, which introduces non-uniformity in the flow field. The M relates to the through over the plate area, as the stress distribution must be accounted for to determine overall viscous response. This yields the total resistive from the varying gradients across the . For parallel-plate measurements, the for Newtonian fluids is given by \eta = \frac{2 M h}{\pi r^4 \Omega}, which follows from the full integration and is equivalent to using the and rate for such fluids. This is particularly useful for assessments where are highest near the rim. Exact computations for Newtonian fluids incorporate the full radial dependence to refine accuracy. The cone-and-plate design excels in providing constant , ideal for precise of both Newtonian and simple non-Newtonian fluids under conditions, while the parallel-plate allows wider gaps suitable for heterogeneous or highly non-Newtonian materials that may not fit narrow truncations. Both geometries require only small sample volumes, typically on the order of 0.1–1 , minimizing material use in settings. Additionally, they facilitate easy sample loading and via the plate. Commercial instruments like the Brookfield LV series incorporate cone-and-plate configurations for low-viscosity s, enabling measurements with high in and . Edge effects, such as or inertial contributions at the periphery, are corrected through established protocols, including adjustments to data based on truncation or , to ensure reliable results.

Coaxial Cylinder and Specialized Designs

Coaxial viscometers, also known as Couette viscometers, consist of two concentric s with the fluid sample placed in the annular gap between them, where one rotates relative to the other to induce . The inner has radius R_i and the outer has radius R_o, with the gap width defined as d = R_o - R_i. For narrow gaps where d \ll R_i, the \dot{\gamma} is approximated as \dot{\gamma} \approx \Omega R_i / d, with \Omega as the of the rotating . There are two primary configurations: the Searle type, in which the inner cylinder rotates while the outer remains stationary, and the Couette type, where the outer cylinder rotates and the inner is fixed. These setups enable precise measurement of viscous torque, similar to torque-based shear in other rotational geometries. Viscosity \eta is calculated from the torque M, cylinder height h, and angular velocity \Omega using the formula \eta = \frac{M \left( \frac{1}{R_i^2} - \frac{1}{R_o^2} \right)}{4 \pi h \Omega}, which derives from the balance of viscous forces in the cylindrical gap. The electromagnetically spinning () viscometer represents a specialized variant, employing a small metal suspended in the sample within a . An external spins the sphere, and the required to maintain a constant rotational speed is determined from the driving current, allowing calculation based on the sphere's motion through the . This facilitates measurements in small sample volumes and opaque fluids. The Stabinger viscometer is another advanced design, featuring a rotating outer tube with a freely spinning inner bob immersed in the sample. As the tube rotates, viscous drag and centrifugal forces cause the bob to co-rotate, with the equilibrium speed yielding both dynamic and ; kinematic is then computed from these values per ASTM D7042. This method ensures a uniform measuring gap under centrifugal pressure, enhancing accuracy for products. Coaxial cylinder designs, including EMS and Stabinger variants, offer advantages such as coverage of a wide range from low to high values, precise via integrated heating or cooling, and suitability for testing lubricants and non-Newtonian fluids under controlled conditions.

Specialized and Industrial Viscometers

Krebs and Stormer Viscometers

The Krebs viscometer, developed in the 1930s by William K. Krebs, is a rotational instrument designed for empirical measurement of and consistency using a that rotates at a constant speed of 200 (rpm). The device measures the required to maintain this rotation in the fluid sample, converting the resistance into Krebs Units (KU), an empirical scale ranging from 0 to 141 KU, where higher values indicate greater viscosity. This method, standardized under ASTM D562, evaluates the consistency of paints and related coatings, particularly non-Newtonian fluids like inks and varnishes, by immersing the paddle in a 500 mL sample and recording the digitally or via a spring-loaded dial in manual models. The Stormer viscometer, an earlier design patented around 1908 and widely adopted in the coatings industry by the 1930s, operates on a similar rotational principle but uses falling weights to drive a paddle through the at a controlled speed. In operation, a weight is attached to a cord that rotates the paddle for exactly 100 revolutions, and the time taken is measured; this duration is then correlated to in centipoise () using charts specific to the type, making it suitable for Newtonian and non-Newtonian substances such as paints and clay slips. Modern variants may include digital timing for precision, but the core manual process remains valued for its simplicity in industrial settings. Both instruments are rugged and portable, enabling quick field assessments in paint manufacturing and quality control without requiring complex setups, though they provide relative rather than absolute values due to their dependence on specific rates (around 4.9 s⁻¹ for Krebs). Their empirical suits grading applications in the coatings sector, where under brushing or rolling conditions is prioritized over fundamental rheological data, but limitations include sensitivity to sample and non-reproducibility for highly thixotropic materials. These tools have been staples in the since the mid-20th century, with the Krebs model particularly dominant for KU-based specifications in architectural and industrial paints.

Viscosity Cups and Ultrasonic Viscometers

Viscosity cups, including , Zahn, and ISO types, provide a straightforward empirical method for assessing the kinematic of Newtonian or near-Newtonian , particularly in industrial settings like paints and varnishes. These devices consist of a with a calibrated at the base, where the efflux time—the duration required for a fixed volume of to drain under gravity—is measured and correlated to units such as centistokes () via standardized conversion charts. The operation is simple: the is dipped into the sample to fill it to a precise level, the is uncovered, and a captures the time until the continuous stream breaks, typically at 20–100 seconds depending on the and . This approach yields relative values suitable for , though it assumes constant (e.g., 20°C ± 0.5°C) and for accuracy. Ford cups, standardized under ASTM D1200, feature orifices ranging from 2 mm to 4 mm (e.g., No. 4 cup with 4.12 mm orifice for 70–370 mm²/s range), and are widely adopted for paints, lacquers, and related coatings due to their robust aluminum construction and stainless steel nozzles. Zahn cups, designed as portable dip-style instruments compliant with ASTM D4212, offer quick field measurements with nested cup designs (e.g., #2 cup for 40–230 cSt in under 90 seconds) and are favored for their ease in handling inks, syrups, and varnishes without a stand. ISO cups adhere to ISO 2431 and ASTM D5125, with interchangeable orifices from 3 mm to 8 mm (e.g., 4 mm for 25–130 mm²/s and 25–100 seconds efflux), enabling broader viscosity grading for global paint formulations. Conversion from efflux time to cSt relies on empirical tables, such as those provided by manufacturers, ensuring repeatability across three replicate measurements per sample. These cups are valued for their portability, low cost (often under $200), and minimal maintenance, making them ideal for on-site industrial checks despite limitations with non-Newtonian or highly viscous fluids. Ultrasonic viscometers employ high-frequency sound waves to evaluate fluid non-invasively, leveraging principles of wave propagation where influences or velocity through the medium. In these systems, an emits or longitudinal waves into the fluid, and the received signal's —quantified by the attenuation coefficient α—reveals viscous effects via energy dissipation. is derived from the coefficient, often approximating the fluid's resistance to deformation without direct contact between the and sample. This method suits inline industrial monitoring, as waves propagate through container walls or waveguides, enabling real-time assessment in pipelines or vessels. A common calculation relates dynamic viscosity η to as η ≈ k (α / f²), where f is the (typically 1–10 MHz), α is the in Nepers per unit length, and k is an empirical or theoretically derived constant incorporating fluid density ρ and sound speed c (e.g., from Stokes' relation adapted for : α ≈ (ω² η)/(2 ρ c³), with ω = 2πf). This quadratic frequency dependence isolates viscous contributions from other mechanisms, calibrated against known standards for precision within 1–5% error. Post-2000 advancements have integrated these viscometers into , such as monitoring (e.g., 0.97 mPa·s with 1.86 mPa·s accuracy via inline acoustic flowmeters) or high-viscosity gels (20–27,000 ·s) in and beverage lines, enhancing process efficiency without contamination risks. Their key advantages include non-contact operation for sterile environments, robustness in harsh conditions, and continuous capabilities, contrasting with batch-based methods.

Applications and Calibration

Industrial and Scientific Applications

Viscometers play a critical role in the for grading engine oils according to standards, where kinematic is measured at 100°C using ASTM D445 and cold-cranking simulator at -30°C via ASTM D5293 to ensure proper lubrication performance across temperature ranges. In pharmaceuticals, rotational viscometers like the Rheolab QC assess the rheological properties of oral syrups to maintain consistent texture and flow, influencing patient compliance and efficacy. The food sector employs viscometers to evaluate consistency during production, correlating inline measurements with Bostwick consistometer readings to optimize blending of pastes for desirable pourability and . For paints, the Stormer viscometer under ASTM D562 determines application in Krebs units, guiding adjustments to achieve uniform without sagging or poor leveling. In scientific research, viscometers facilitate studies by quantifying resistance to flow, enabling models of and in engineering simulations. relies on and rotational viscometers to derive from dilution series, correlating it to molecular weight via the Mark-Houwink equation for quality assessment in materials like . Biomedical applications use microfluidic viscometers to measure blood , providing insights into cardiovascular health; elevated values signal risks like , as seen in studies of non-Newtonian shear-thinning behavior under physiological conditions. Viscometers are integral to quality control in lubricant production, ensuring engine oils like 5W-30 meet SAE J300 specifications for low-temperature pumpability (maximum 6600 cP at -30°C) and high-temperature stability to prevent wear and maintain fuel efficiency. This monitoring verifies product stability, facilitating reliable flow through engines and pipelines while minimizing operational failures. Emerging microfluidic viscometers, such as the tesla valves and ultrasound waves-powered blood plasma viscometer, enable real-time analysis of biofluids with small sample volumes (∼0.2 mL), supporting point-of-care diagnostics for cardiovascular conditions through precise viscosity and coagulation profiling. Key challenges in viscometry include compensation to account for variations, as seen in lubricants where synthetic 5W-30 oils exhibit anomalous increases at elevated pressures and temperatures, requiring controlled environments for accurate readings. For non-Newtonian fluids prevalent in processes like food and biomedical flows, modeling shear-dependent demands advanced rheometers to capture yield stress and without inertial artifacts.

Calibration Methods and Standards

Calibration of viscometers ensures measurement accuracy by verifying instrument performance against known standards, typically involving the use of certified reference fluids at controlled temperatures such as 20°C or 40°C, where measured viscosities are compared to certified values. This process minimizes systematic errors and maintains traceability to international measurement systems. Certified reference materials, such as NIST-traceable mineral or silicone oils, serve as primary standards due to their Newtonian behavior and well-characterized viscosities, often provided by manufacturers like Cannon Instrument with uncertainties below 0.2%. These oils, including the S series like S3 to S30,000 from accredited suppliers, exhibit stable properties over time when stored properly, enabling reliable calibration across viscosity ranges from 0.5 mPa·s to over 100,000 mPa·s. Specific calibration methods vary by viscometer type. For viscometers, gravimetric techniques determine the calibration constant by measuring the volume of and its flow time under gravity, ensuring the product of time and constant yields the known kinematic . Rotational viscometers undergo using standard fluids, where the instrument's response is adjusted to match expected values at defined rates, often verified with Brookfield or similar certified oils. Vibrational or oscillating viscometers, such as or tuning-fork types, are calibrated by checking shifts in liquids, confirming sensitivity to changes within 1% accuracy. International standards govern these procedures to ensure consistency. ASTM D445 outlines calibration for viscometers measuring kinematic of petroleum products, requiring instruments calibrated to within 0.2-0.5% of certified values using reference oils. For rotational viscometers, ASTM D2196 specifies methods for non-Newtonian fluids, emphasizing torque verification with standards to assess and behavior. ISO 3104 provides guidelines for kinematic determination in transparent and opaque liquids via methods, with dynamic calculated by multiplying by , covering ranges from 0.2 mm²/s to 300,000 mm²/s. ISO 3219 addresses for polymers and viscous non-Newtonian fluids using rotational or setups, focusing on controlled conditions for . Laboratory error limits are typically maintained below 1% through these standards, ensuring high precision in industrial and research settings. Calibration frequency is recommended annually or after intensive use, with more frequent verification (e.g., quarterly) for critical applications to detect drift. All calibrations must demonstrate to units (Pa·s for dynamic ) via unbroken chains from national institutes like NIST, incorporating documented uncertainties for each step.

References

  1. [1]
    None
    ### Summary of Viscometer Information from https://cse.sc.edu/~gatzke/460/E-Rheology%203-07.pdf
  2. [2]
    [PDF] 1. theory and equations for brookfield viscometers
    The Brookfield viscometer measures the torque required to rotate the selected spindle in a fluid. This torque value is directly related to the viscosity of ...
  3. [3]
    Viscometers – Visual Encyclopedia of Chemical Engineering ...
    Apr 1, 2022 · Falling body viscometers are commonly used for low-viscosity substances in the pharmaceutical, food, chemical and mineral oil industries.
  4. [4]
    viscometers: Topics by Science.gov
    The viscometer involves the action of a rotating magnetic field on a metallic cylinder floating on the liquid to be studied. The principles of electromagnetism ...
  5. [5]
    The centrifugal viscometer - PMC - NIH
    It has been demonstrated that viscosity plays important roles in food, cosmetic, pharmaceutical, and chemical industries. For viscosity measurement, shear and ...
  6. [6]
    A comprehensive review on MEMS-based viscometers - ADS
    Viscosity is an important rheological parameter, which needs to be measured accurately in various industrial applications to improve the quality of the product.
  7. [7]
    ISO 3219:1993 - Plastics — Polymers/resins in the liquid state or as ...
    General specification for rotational viscometers for the determination of the viscosity of polymers in the liquid or liquidlike state.
  8. [8]
    Viscometer - an overview | ScienceDirect Topics
    The most common type of viscometers are capillary viscometers or also known as Ostwald's viscometers that require capillary tubes, which are held vertically in ...1 Introduction · Viscometers · 3.5 Others
  9. [9]
    What is Viscosity? Why is it important for fluids characterization?
    Viscosity is often referred to as the thickness of a fluid. You can think of water (low viscosity) and honey (high viscosity).
  10. [10]
    Common Units for Dynamic and Kinematic Viscosity - RheoSense
    The SI unit for dynamic viscosity η is the Pascal-second (Pa-s), which corresponds to the force (N) per unit area (m2) divided by the rate of shear (s-1). Just ...<|control11|><|separator|>
  11. [11]
    Viscometers: A Practical Guide - Machinery Lubrication
    Types of Viscometers · Portable Viscometers · Small Onsite Viscometers · Full-service Viscometers · Inline/Online Viscometers.
  12. [12]
    Pharmaceutical Applications of Viscosity Measurement
    Shear rate and temperature sweeps · Viscosity versus concentration · Intrinsic viscosity analysis · Hydrodynamic size estimation · Injectability analysis.
  13. [13]
    Viscometer Measurement: A Glimpse into Reservoir Fluid Dynamics
    Sep 26, 2023 · Viscometer measurement improves the process of oil reservoir analysis by offering a window into the viscosity and flow properties of these ...
  14. [14]
    Review on viscosity measurement: devices, methods and models
    May 5, 2023 · The article helps the researchers select the appropriate viscometer, method, and correlation to estimate their research's fluid viscosity.
  15. [15]
    Viscometer vs Rheometer: Discover Why a ... - TA Instruments
    Sep 30, 2024 · The ability to study viscosity over a range of temperatures is especially useful for materials such as oils, plastics, resins or greases, as it ...The Rheometer Can Show How... · Rheometer: Precision... · Other Rheometer Capabilities...<|control11|><|separator|>
  16. [16]
    Viscometer or Rheometer – which is best for me?
    Oct 3, 2020 · Rheometers, while generally more expensive than viscometers, are more versatile and have a much wider dynamic range of control and measurement parameters.
  17. [17]
    Chapter 2 History of rheology and Newton - ScienceDirect
    During the last 130 years, a number of instruments have been developed, which measure flow properties both on static as well as dynamic liquid systems.
  18. [18]
    What is Rheology? - RheoSense Blog
    Jun 29, 2021 · Hooke introduced his famous linear law relating stress and strain, while Newton described the flow behavior of fluids with a simple linear ...<|separator|>
  19. [19]
    Stokes' law, viscometry, and the Stokes falling sphere clock - PubMed
    Sep 4, 2020 · Stokes derived his law in the course of investigations to determine the effect of a fluid medium on the swing of a pendulum.Missing: 1845 | Show results with:1845
  20. [20]
    Ubbelohde's viscometer | Opinion - Chemistry World
    Aug 1, 2017 · Leo Ubbelohde (1877–1964). Invented a precise method of measuring absolute viscosity. Ubbelohde's parents were involved in the peat business ...
  21. [21]
    Brookfield Story
    In 1934, Don Brookfield Sr., a forward-thinking MIT graduate, set out to create an instrument that didn't yet exist—a better way to measure viscosity. That ...Missing: rotational 1930s
  22. [22]
    Industriology #2 - A brief history of viscometry and rheometry - rheonis
    The invention of the term viscometer is attributed to Charles Dollfuss, an industrialist in the textile sector at the time of the first industrial revolution.
  23. [23]
    [PDF] Kinematic Viscosity of Transparent and Opaque Liquids (and the ...
    Current edition approved March 10, 2003. Published April 2003. Originally approved in 1937. Last previous edition approved in 2001 as D 445–01.
  24. [24]
    ISO Viscosity Classification - Anton Paar Wiki
    In the 1970s, the well-known standardization institutes ISO (International Standards Organization), ASTM (American Society for Testing and Materials), ...
  25. [25]
    Portable Microfluidic Viscometer for Formulation Development and ...
    Aug 2, 2024 · Here we introduce a portable single-use analytical chip-based viscometer for determining the viscosity of protein solutions using low sample volumes of 10 μL.Experimental Section · Microfluidic Device Design... · Portable Viscometer For...Missing: post- | Show results with:post-
  26. [26]
    Probing Rate-Dependent Liquid Shear Viscosity Using Combined ...
    Jun 3, 2025 · Dynamic viscosity (η), defined as the ratio of shear stress (τ) to shear rate (γ̇): η = τ/γ̇, quantifies a fluid's resistance to deformation ...
  27. [27]
    [PDF] FLUID MECHANICS 203 - eng . lbl . gov
    The symbol η is also commonly used for dynamic viscosity. There are other ways of expressing viscosity and this is covered next. 1.2.2 KINEMATIC VISCOSITY ν.
  28. [28]
    [PDF] The accuracy of measurements of viscosity of liquids
    The viscosity (shear viscosity) of a Newtonian liquid is normally measured by a relative technique in an instrument calibrated using a liquid of known viscosity ...
  29. [29]
    Effects of temperature alteration on viscosity, polymerization, and in ...
    Jun 9, 2021 · This phenomenon could be interpreted based on the Andrade equation about temperature-dependent viscous models for liquids [15]: η = Aexp(Ea/RT) ...
  30. [30]
    [PDF] CHAPTER 3 THE CONCEPT OF VISCOSITY
    The viscosity of Newtonian fluids is affected by temperature, pressure, and, in the case of solutions and mixtures, by composition. The effect of pressure and ...
  31. [31]
    [PDF] Guide for the Use of the International System of Units (SI)
    Feb 3, 1975 · (a) The poise (P) is the CGS unit for viscosity (also called dynamic viscosity). The SI unit is the pascal second (Pa · s). (b) The stokes ...
  32. [32]
    [PDF] Non-Newtonian Fluid Math - BYU College of Engineering
    𝜏𝜏 = 𝜇𝜇 𝛾𝛾̇ (1) In contrast, for a non-Newtonian fluid, the viscosity is a function of 𝛾𝛾̇. A common way to. describe this is with an apparent viscosity 𝜇𝜇ap.
  33. [33]
    Rheology | NIST - National Institute of Standards and Technology
    A fluid is Newtonian if it has a constant viscosity (for a given temperature), regardless of any shearing forces applied to it. A common example of a Newtonian ...Missing: definition | Show results with:definition
  34. [34]
    [PDF] Pipe Flow of Newtonian and Non-Newtonian Fluids
    Fluids whose viscosity decreases when shear pressure is applied are known as shearthinning fluids. Examples include blood, toothpaste, or tomato sauce.
  35. [35]
    RHEOLOGICAL TESTING OF FOODS
    Thixotropic fluids. A thixotropic fluid is one in which the apparent viscosity decreases with time when the fluid is subjected to a constant shear rate.
  36. [36]
    A review on non-Newtonian fluid models for multi-layered blood ...
    Jan 31, 2023 · Power-law fluid model. Power-law fluids are a type of visco-inelastic fluids which exhibits a prominent characteristics of non-Newtonian fluids ...
  37. [37]
    [PDF] Non-Newtonian Fluids for Rheological Measurements
    Polymer melts and solutions often demonstrate non-Newtonian flow behavior [1-3]. Such fluids often have a viscosity that depends on the shear rate and display.Missing: implications viscometer
  38. [38]
    Comparative Response of Newtonian and Non-Newtonian Fluids ...
    Apr 24, 2021 · These fluids do not possess any elastic properties and they exhibit constant shear-viscosity in viscometric experiments. Newtonian fluids are ...Missing: implications viscometer measurements
  39. [39]
    [PDF] ASTM-D445-2017.pdf
    ... first published in 1991. See Guide D6074. D445 − 17a. 8. Copyright by ASTM Int'l (all rights reserved); Tue Jul 11 13:40:19 EDT 2017. Downloaded/printed by.
  40. [40]
    [PDF] Visco handbook - Xylem
    With UBBELOHDE viscometers, the transition from the capillary. (7) into the leveling bulb (6) is designed as a spherical cap. An additional venting tube (1) is.
  41. [41]
    [PDF] Rheometry FLOW IN CAPILLARIES, SLITS AND DIES DRIVEN BY ...
    (4) Determine ˙γw from equation (8-20a). (5) Determine viscosity η = σw/˙γw. 4. Page 5 ...
  42. [42]
    PART ONE How to Determine Viscosity Data Using a Slit-Die ...
    Mar 13, 2019 · This article describes how a slit-die viscometer connected to a laboratory extruder can be used to obtain accurate data on melt viscosity at various shear ...Missing: formula | Show results with:formula
  43. [43]
    [PDF] Practical Rheology Section 3 - Dynisco
    Here Newtonian and non-Newtonian shear rates are equal. The value of shear rate at this point is called the representative value. By using this value it is ...
  44. [44]
    Determination of shear viscosity and shear rate from pressure drop ...
    Jan 12, 2007 · In this study, we present a unique approach to calculate the shear viscosity and shear rate with the pressure drop and flow rate data from a channel having a ...Missing: formula | Show results with:formula
  45. [45]
    None
    ### Summary of Bubble Viscometers from BYK-Gardner PDF
  46. [46]
  47. [47]
  48. [48]
    [PDF] Design of a high precision falling ball viscometer - HAL
    Dec 15, 2007 · The viscometer measures viscosity by observing a falling ball's speed limit, corrected for effects like edge and inertial effects, to achieve a ...
  49. [49]
    Falling Ball Viscometer Principle - Inst Tools
    Jul 15, 2019 · The principle of the viscometer is to determine the falling time of a ball of known diameter and density through a close to vertical glass tube.
  50. [50]
    Falling-Ball Viscometer / Hoeppler manual - Anton Paar Wiki
    The falling-ball viscometer measures the time a ball takes to move through a liquid, using the Hoeppler principle, to calculate dynamic viscosity.
  51. [51]
    Is Faxén's correction function applicable to viscosity measurements ...
    The widely used Faxén correction function was found to significantly over-compensate the retarding effect of the crucible walls on the sphere velocity.
  52. [52]
    Design of a high precision falling-ball viscometer - AIP Publishing
    Jan 20, 2005 · This article describes the development stages of this viscometer. The bibliographical study highlights the various corrections, which must be ...
  53. [53]
    Viscometer Design and Application Review - Engineers Edge
    Stokes' law is the basis of the falling sphere viscometer, in which the fluid is stationary in a vertical glass tube. A sphere of known size and density is ...<|separator|>
  54. [54]
    Norcross Viscometers - Beacon Industrial Group
    Norcross viscometers use a sensor and controller, measuring viscosity by the time a piston falls through fluid, based on the 'Piston Time-of-Fall'.
  55. [55]
    Theoretical and experimental study of the falling cylinder viscometer
    A correlation equation from which the Geometry Number can be calculated is presented which can be used to design viscometers for particular applications.Missing: piston | Show results with:piston
  56. [56]
    Viscometers | Cambridge Viscosity
    Sensor Technology. CVI's sensors incorporate unique oscillating piston technology, for low-maintenance functionality, unmatched accuracy, and precise monitoring ...
  57. [57]
    How to measure viscosity | Anton Paar Wiki
    This article highlights traditional and technically advanced methods of how to measure viscosity and presents different types of viscometers. Learn more!
  58. [58]
    Creating Shear Sweeps with an Oscillating Piston Viscometer
    Mar 30, 2020 · The oscillating piston method, while not a rheometer, can produce rheometric data by providing viscosity values at known shear rates with non-Newtonian fluids.Missing: design advantages
  59. [59]
    Principles and Applications of the Quartz Crystal Microbalance (QCM)
    The quartz crystal microbalance is about 60 years old. At that time, people working with quartz resonators knew that one can lower the frequency by scribbling ...
  60. [60]
    Basics of a Quartz Crystal Microbalance - Gamry Instruments
    Knowing the liquids viscosity and density allows one to calculatethe expected frequency decrease upon immersing the crystal in that liquid using the equation.
  61. [61]
    Frequency of a quartz microbalance in contact with liquid
    Response of Quartz Crystal Microbalance to Liquid Electrical Properties. ... Kanazawa, Seimei Shiratori. Adsorption–Desorption Control of Fibronectin in ...
  62. [62]
    [PDF] QCM100- Quartz Crystal Microbalance Theory and Calibration
    The Sauerbrey equation relies on a linear sensitivity factor, Cf, which is a fundamental property of the QCM crystal. Thus, in theory, the QCM mass sensor does ...Missing: viscometer | Show results with:viscometer
  63. [63]
    Principles and Applications of the Quartz Crystal Microbalance (QCM)
    This influence is described by what today is called the Gordon-Kanazawa equation [23]. The Gordon-Kanazawa equation is a rediscovery. In slightly implicit ...
  64. [64]
    SV-A Series Tuning Fork Vibro Viscometers | Test & Measurement
    The SV-10/SV-10A series of tuning fork vibro viscometers are capable of continuous measurement over a range of viscosities from the extremely low (0.3 mPa·s) to ...
  65. [65]
    [PDF] Characteristics of Tuning-fork Vibration Rheometer RHEO-VISCO ...
    Tuning-fork vibration viscometers have a low resonance frequency of 30 Hz, and the amount of energy they transmit to materials through these slight vibrations ...
  66. [66]
    Micro Motion FVM Fork Viscosity Meter - Emerson Global
    $$20 deliveryA high performance, multivariable Fork Viscosity Meter that delivers reliable viscosity, density and temperature measurement in pipelines, bypass loops and ...
  67. [67]
    Demonstration of SV Tuning Fork Viscometer - Machinery Lubrication
    No more changing spindles and speeds to "find" the right range - this one sensor measures the whole range, from 0.3cP to 10,000cP.
  68. [68]
    [PDF] Rheology: An Introduction - TA Instruments
    Apr 2, 2019 · The shear rate and shear stress are constant throughout the gap with the cone-and-plate geometry. Parallel plate data can be corrected with ...
  69. [69]
    [PDF] Chapter 10: Rheometry
    Mar 18, 2014 · The introduction of the cone means that shear rate is independent of r. 0. Θ. Ω. = γq. Ω. Θ. = 3. 0. 2. 3. )( R. T π γ η q. No corrections ...Missing: viscometer | Show results with:viscometer
  70. [70]
    DVNext Cone/Plate Rheometer: Precision for Small Sample Volumes
    Get accurate viscosity measurements with the DVNext Cone/Plate Rheometer. Ideal for small samples, it offers precise shear rates, rapid temperature control, ...
  71. [71]
    Shear Rate in the Couette Viscometer
    Theory of the Concentric Cylinder Viscometer​​ = sRI. Here Tl = 111/27rRI 2L and S-2T 1 = 1I1/27rR22L are the shear stresses at the inner and outer cylindrical ...
  72. [72]
    An improved shear rate estimation using rotating coaxial cylinder ...
    Apelblat et al. (1975) proposed an approximate shear rate formula for the Couette viscometer with a narrow annular gap between two coaxial cylinders ...
  73. [73]
    [PDF] 2.2.10. VISCOSITY - ROTATING VISCOMETER METHOD
    Feb 2, 2010 · pascal-seconds is given by the following formula: M. = torque in newton-metres acting on the cylinder surface, ω. = angular velocity in ...
  74. [74]
    20.4.2: The Couette Viscometer - Physics LibreTexts
    Aug 7, 2022 · In equilibrium, this is equal to c ⁢ ϕ , where c is the torsion constant of the suspension and ϕ is the angle through which the inner cylinder ...
  75. [75]
    Electromagnetically Spinning Sphere Viscometer - IOPscience
    Jan 15, 2010 · An aluminum sphere revolves in a sample cell under a rotating external magnetic field, and the rotational speed of the sphere gives the sample ...
  76. [76]
    D7042 Standard Test Method for Dynamic Viscosity and Density of ...
    May 21, 2025 · Standard Test Method for Dynamic Viscosity and Density of Liquids by Stabinger Viscometer (and the Calculation of Kinematic Viscosity).
  77. [77]
    An Introduction to Cup and Bob Viscometer - Martests Instrument
    For many fluids, especially Newtonian fluids, the viscosity (η) is simply the ratio of shear stress to shear rate (η = τ/γ̇). In a cup and bob system, the ...
  78. [78]
    ASTM D7042 | Anton Paar Wiki
    ASTM D7042 is a standard test method for measuring dynamic viscosity and density of Newtonian fluids like liquid petroleum products and crude oils.
  79. [79]
    Rheometry and measuring instruments - BYK
    Stormer viscometers are based on a similar measuring principle. Here, the so-called KU values (Krebs Units), which are an indication of the processing ...
  80. [80]
    KU-3 Viscometer: Precise Viscosity Testing for Paints, Coatings, and ...
    The KU-3 Viscometer provides accurate, real-time viscosity measurements in Krebs units, grams, and cP. Ideal for paints, coatings, and inks testing with ...
  81. [81]
    Stormer Viscosimeter | Smithsonian Institution
    Stormer viscosimeters were soon being used for paint, clay slips, and pyroxylin solutions, as well as for corn, catsup, and cod liver oil. Stormer's original ...Missing: operation industry 1930s
  82. [82]
  83. [83]
    Standard Test Method for Viscosity by Ford Viscosity Cup - ASTM
    Jun 15, 2023 · This test method covers the determination of the viscosity of Newtonian or near-Newtonian paints, varnishes, lacquers, and related liquid materials.
  84. [84]
    Ford Viscosity Cup - BYK Instruments
    In stockThe BYK-Gardner Ford cup is a flow cup that complies with ASTM methods D333, D365, D1200. Body made of solid bar aluminum. Stainless steel orifice.
  85. [85]
  86. [86]
    ISO Flow Viscosity Cups - BYK Instruments
    In stockThe ISO Flow Cup has a 10-40 cSt range, 3mm orifice, 30-100 second efflux time, is compliant with ISO 2431 and ASTM D5125, and has a 3% calibration.
  87. [87]
    [PDF] Survey of Integrated Viscosity Sensing Methods for the ...
    Oct 13, 2022 · Viscosity measurements of food and beverage products help maximize both production and process-ability efficiency and ensures a consistently ...
  88. [88]
    [PDF] development of ultrasonic densimeter and viscometer for fluid ...
    In this work, an ultrasonic testing system was developed for fluid density and viscosity measurement. The proposed system was validated using water and honey. ...
  89. [89]
    Effect of Solvents on the Ultrasonic Velocity and Acoustic ...
    Apr 26, 2016 · (5). Ultrasonic attenuation (α/f2) and relaxation time (τ) were calculated using the following [16]:. (7). where η is the viscosity of the ...
  90. [90]
    [PDF] In-situ ultrasonic viscometry of lubricants under temperature and shear
    Dec 16, 2023 · The purpose of ultrasonic viscometry development is to overcome the limits of conventional viscometers.
  91. [91]
    SAE Motor Oil Viscosity Ratings - EngineeringTechnology.org
    ASTM standards D445 (kinematic viscosity) and D5293 (mini-rotary viscometer) are commonly used to measure the oil's viscosity at various temperatures. Test ...
  92. [92]
    Research of rheological properties of medicinal syrup for oral use
    The rheological (structural-mechanical) properties of the samples were determined using a Rheolab QC rotary viscometer (AntonPaar, Austria) with coaxial ...
  93. [93]
    Relationship between In‐line Viscosity and Bostwick Measurement ...
    Jul 31, 2009 · This work addresses the blending of tomato pastes, packed at different Bostwick readings, for use in tomato ketchup production. The objective of ...
  94. [94]
    Optimum Viscosity for Paint Application
    Different viscometers measure viscosity at different shear rate ranges. The Stormer instrument (see ASTM D562) often used to test architectural and maintenance ...
  95. [95]
    Role of Viscosity in Fluid Dynamics - Longdom Publishing
    Viscometers, devices designed directly for viscosity measurement, come in different types such as capillary viscometers, rotational viscometers, and falling ...
  96. [96]
    [PDF] Characterization of polyacrylamide (PAM) by viscometry and gel ...
    We shall study about three viscometers: falling ball viscometer, capillary (Ubbelohde) viscometer and Couette viscometer. The falling ball viscometer was ...
  97. [97]
    Development of a Microfluidic Viscometer for Non-Newtonian Blood ...
    Dec 20, 2024 · Our viscosity measurement model, specifically designed for non-Newtonian fluids, demonstrates significant advantages in accuracy, sample ...Missing: implications | Show results with:implications
  98. [98]
    Measuring Kinematic Viscosity of Engine Oils: A Comparison of Data ...
    Apr 4, 2021 · The aim of this paper is to compare the results of kinematic viscosity of lubricating oils measurements at 40 °C, obtained with three different rapid ...
  99. [99]
    Experimental Investigation of the Viscosity Parameters Ranges ...
    The SAE J300 classification (The Society of Automotive Engineers) describes the viscosity properties of multi-grade oils, which are widely used because they ...
  100. [100]
    A bio-fabricated tesla valves and ultrasound waves-powered blood ...
    Apr 23, 2024 · We fabricate a novel digital device, called Tesla valves and ultrasound waves-powered blood plasma viscometer (TUBPV) which shows capacities in both viscosity ...
  101. [101]
    Pressure- and Temperature-Dependent Viscosity Measurements of ...
    Anomalous increases in viscosity at specific temperatures were observed at similar pressures for the 5W-30 and 10W-30 lubricants for both synthetic (S) and ...Missing: control | Show results with:control
  102. [102]
    Viscosity Standards - High-Quality Reference Fluids
    Calibrate CAP Series Viscometers for accurate Cone/Plate Rheometer measurements. Available in various NIST-certified viscosities, with customizable options.General Purpose Silicone · High Temperature Silicone · Krebs Viscometer Oil Fluids
  103. [103]
    Viscosity & Flash Point Standards - CANNON Instrument
    All CANNON viscosity standards are traceable to a NIST calibration and prepared in our ISO 9001-registered, A2LA-accredited laboratory.
  104. [104]
    Viscosity Standard S60 - Stanhope-Seta
    Seta viscosity standards are manufactured in full compliance with ASTM D2162. Certification: Supplied with a Certificate which gives the certified value.
  105. [105]
    [PDF] SOP 14 Gravimetric Calibration of Volumetric Standards Using an ...
    The procedure uses an electronic balance and is suitable for all sizes of gravimetric calibrations only limited by the capacity and resolution of the balance.
  106. [106]
    FAQ - AMETEK Brookfield
    Accuracy is calculated as 1% of the full scale range (FSR) of the viscometer. FSR is defined as the highest achievable viscosity reading with a given spindle ...
  107. [107]
    How to Calibrate a Viscometer | HINOTEK
    This report will explore the fundamental principles of calibration, provide step-by-step instructions for the major types of viscometers, and offer expert ...Section 1: The Fundamentals... · Section 2: The Essential...
  108. [108]
    D445 Standard Test Method for Kinematic Viscosity of Transparent ...
    Apr 16, 2024 · This test method specifies a procedure for the determination of the kinematic viscosity, ν, of liquid petroleum products, both transparent and opaque.Missing: first | Show results with:first
  109. [109]
    D2196 Standard Test Methods for Rheological Properties of Non ...
    Dec 14, 2020 · These test methods cover the determination of the apparent viscosity and the shear thinning and thixotropic properties of non-Newtonian materials.
  110. [110]
    ISO 3104:2020 - Petroleum products
    The range of kinematic viscosities covered in this test method is from 0,2 mm2/s to 300 000 mm2/s over the temperature range ?20 °C to +150 °C. NOTE The result ...
  111. [111]
    Two Types of Viscometers and Tips to Calibrate Them
    Falling-Piston Viscometers. A falling-piston viscometer, or Norcross viscometer, calculates a piston's resistance to a fluid and the time it takes to move ...