Dynamic light scattering

Dynamic light scattering (DLS), also known as photon correlation spectroscopy (), is a non-invasive analytical that determines the size distribution, , and polydispersity of particles and macromolecules in suspension or solution by analyzing temporal fluctuations in the intensity of scattered due to . The method relies on illuminating a sample with a coherent laser beam, detecting the scattered light at a fixed angle (typically 90° or 173° for backscattering), and computing the function of the intensity fluctuations to extract the diffusion coefficient, which is then converted to via the Stokes-Einstein : D = \frac{kT}{6\pi\eta r_h}, where D is the diffusion coefficient, k is Boltzmann's constant, T is temperature, \eta is , and r_h is the . Applicable to particles ranging from less than 1 nm to several micrometers, DLS is particularly sensitive to trace aggregates and provides rapid, low-volume measurements without requiring sample separation or labeling. The theoretical foundations of DLS trace back to early observations of light scattering by colloidal suspensions, including John Tyndall's work in 1868 and Lord Rayleigh's scattering theory in 1871, with formalized by in 1905. Modern DLS emerged in the 1960s through experiments by H.Z. Cummins and colleagues in 1964, who demonstrated spectral broadening from diffusing particles, and Stephen Pecora's 1964 paper on Doppler shifts in scattered light from macromolecules. Key advancements included Eric Pike's development of the digital autocorrelator in 1969, enabling practical autocorrelation analysis, and subsequent refinements in laser technology and detectors during the 1980s. These innovations transformed DLS from a research curiosity into a standardized , now codified in international standards like ISO 22412:2025 for . DLS finds broad applications across fields such as , , and pharmaceuticals, where it characterizes protein homogeneity (e.g., with r_h \approx 3.5 nm), nanoparticle stability, and macromolecular interactions like protein-nucleic acid complexes. In biomedical research, it complements techniques like and by providing insights into aggregation states and conformational changes under varying solution conditions, such as pH or temperature. Despite its strengths, DLS requires dust-free, dilute samples (typically 0.01–1 mg/mL) and assumes spherical particles, with results interpreted via models like the cumulants or CONTIN algorithm to account for polydispersity. Commercial instruments from manufacturers like Malvern Panalytical and Wyatt Technology have made DLS accessible, supporting in industries ranging from formulation to production.

Fundamentals

Basic Principles

Dynamic light scattering (DLS), also known as photon correlation spectroscopy, is a non-invasive optical that measures the and of particles in by analyzing the time-dependent fluctuations in the intensity of light scattered from those particles. These fluctuations arise from the random of the particles in a liquid medium, where smaller particles diffuse more rapidly, leading to faster variations in the scattered light intensity compared to larger ones. The relationship between particle motion and scattered light can be understood through the diffusion of particles, which causes Doppler shifts in the scattered light's frequency spectrum or, equivalently, temporal fluctuations in . In brief, the scattered from an ensemble of particles is the sum of contributions from each moving particle, resulting in an interference pattern that fluctuates over time; the characteristic timescale of these fluctuations is inversely proportional to the particles' D, as the field function decays exponentially with a rate q^2 D, where q is the . Particle size is determined from the diffusion coefficient using the Stokes-Einstein equation, which relates D to the hydrodynamic radius r of a spherical particle: D = \frac{k T}{6 \pi \eta r} where k is Boltzmann's constant, T is the absolute temperature, and \eta is the solvent viscosity. This equation assumes spherical particles and no-slip boundary conditions at the particle-solvent interface. DLS relies on several key prerequisites for accurate measurements, including dilute suspensions where particle concentrations are low enough to neglect multiple events and interparticle interactions. Additionally, the technique operates in the regime, where particles are much smaller than the of the incident (typically a \ll \lambda/10), ensuring isotropic and negligible effects.

Historical Development

The theoretical foundations of dynamic light scattering originated in the early 20th century with Albert Einstein's seminal 1905 work on , which mathematically described the random displacement of particles suspended in a due to molecular collisions, establishing the relationship between and . This provided the essential link between observable particle dynamics and the fluctuations in scattered light intensity that DLS later exploits. In 1906, built upon Einstein's theory by deriving explicit expressions for the diffusion coefficient in colloidal suspensions, further quantifying how thermal motion governs particle transport in liquids. The experimental groundwork for dynamic measurements emerged in 1955 when A. T. Forrester and colleagues invented photon correlation , demonstrating photoelectric detection of intensity correlations in the scattered from incoherent sources, such as the Zeeman-split mercury line. This technique enabled the analysis of temporal fluctuations in , setting the stage for studying dynamic processes. The advent of continuous-wave lasers in the early 1960s facilitated its adaptation to colloidal systems; notably, H. Z. Cummins and co-workers conducted the first laser-based DLS experiment in 1964, measuring the spectrum of scattered by spheres undergoing . H. L. Swinney, collaborating with Cummins, advanced these methods through light-beating studies in the mid-1960s, applying them to in liquids and . Commercialization accelerated in the 1970s, driven by reliable sources and autocorrelators, which automated analysis and broadened accessibility beyond specialized labs. The 1976 book Dynamic Light Scattering by B. J. Berne and R. Pecora synthesized the field's theory and applications, serving as a foundational reference that spurred instrument development and interdisciplinary use in , , and physics. The first commercial DLS system, the Malvern Correlator, launched in 1971, marked the transition to routine particle sizing in and . Since 2000, innovations have enhanced DLS for challenging samples, including integration with for real-time, low-volume analysis of evolving colloidal systems during or . Advances in detector technology, such as adaptive, time-resolved photon correlation processing with high-sensitivity avalanche photodiodes, have improved signal-to-noise ratios, enabling reliable characterization of sub-10 nm particles that were previously undetectable due to weak scattering.

Experimental Setup

Instrumentation

Dynamic light scattering (DLS) instrumentation consists of several core components designed to illuminate a sample with coherent , collect the scattered , and analyze its fluctuations to infer particle dynamics. The primary source is typically a monochromatic , such as a helium-neon (He-Ne) operating at a wavelength of 632.8 nm, which provides the necessary coherence for detecting subtle phase shifts caused by Brownian motion. This beam is directed through an attenuator to control , ensuring optimal signal without overwhelming the detector. The sample is held in a quartz cuvette or capillary cell, typically 10–100 μL for standard cuvettes, with advanced systems supporting as low as 2 μL to minimize sample use while ensuring a stable measurement zone, positioned such that the scattering volume is focused within the sample. Scattered is captured by detection optics, commonly a photomultiplier tube (PMT) or avalanche photodiode (APD), which convert photon arrivals into electrical signals sensitive to variations on timescales from nanoseconds to seconds. These signals are then processed by a digital correlator, a or software module that computes the autocorrelation function of the fluctuations, enabling the extraction of diffusion coefficients. Optical geometries in DLS setups are configured to optimize the detection of scattered light, with two primary modes: homodyne and heterodyne. In homodyne detection, a single detector collects the self-beating of the scattered from multiple particles, producing intensity fluctuations proportional to the square of the field , which is standard for most routine particle sizing applications. detection, in contrast, mixes the scattered light with a (e.g., a portion of the incident ), resulting in linear beating and enhanced sensitivity to slow dynamics or directed motions like , though it requires more precise optical alignment. The angle is selected via or fixed optics, typically at 90° for polydisperse samples to balance signal strength and minimize contributions from larger particles, or at angles like 173° in modern instruments to reduce multiple effects in turbid samples. Key operational parameters influence the accuracy and range of DLS measurements. The laser's should exceed the difference (typically on the order of meters for standard lasers) to ensure that the scattered light maintains phase relationships over the , directly affecting the factor β in the analysis. Detector dead time, typically on the order of nanoseconds for APDs, limits the maximum count rate to avoid signal pile-up and distortion in the . Sample volume constraints, often 10–100 μL depending on the design, must accommodate the focused volume while preventing wall effects or . Contemporary multi-angle DLS systems, such as those employing rotating optics or multiple fixed detectors (e.g., spanning 30°–150°), enhance resolution for complex size distributions by acquiring data at varied angles, improving the of polydispersity without assuming spherical particles. Safety and procedures are essential for reliable -based DLS operation. protocols mandate the use of appropriate rated for the (e.g., OD 4+ for 632.8 ) and enclosure of the path to prevent exposure, in compliance with standards like ANSI Z136.1. involves precise steering with mirrors and lenses to center the incident on the sample and focus the collection on the volume, often automated in commercial systems like the Zetasizer series to maximize the factor and ; manual verification uses a visible or power meter to confirm <1% deviation in intensity across the setup.

Sample Preparation and Measurement

Sample preparation for dynamic light scattering (DLS) requires dilute suspensions to minimize multiple scattering effects, typically at concentrations of 0.01–10 mg/mL depending on the sample type (e.g., lower for nanoparticles, higher for proteins), ensuring the sample is clear or faintly opaque while providing sufficient scattering signal. Filtration through 0.2–0.45 μm membranes is essential to remove dust particles and large aggregates that could dominate the scattering signal, with the filter pore size chosen to retain the particles of interest. Refractive index matching between the sample solvent and the cuvette or surrounding medium enhances signal clarity, particularly in setups prone to unwanted reflections, by reducing interfacial scattering. Measurement protocols begin with sample equilibration, allowing 10–15 minutes at the desired temperature to achieve thermal stability and prevent convection currents that could distort Brownian motion readings. Temperature is precisely controlled using Peltier elements, maintaining stability within ±0.1°C, as viscosity and diffusion coefficients are highly temperature-dependent. Multiple runs, typically 3–10 per sample, are performed and averaged to improve reproducibility, with each run conducted at the optimal scattering angle selected based on particle size—low angles (e.g., 15°–30°) for larger particles exceeding 100 nm to capture forward-biased scattering patterns. Common pitfalls include sample aggregation, which increases apparent size due to enhanced scattering from clusters; sedimentation in denser suspensions, leading to inhomogeneous measurements; and contamination from dust or bubbles, which introduce spurious large-particle signals. Troubleshooting involves re-filtration or centrifugation for aggregates and sedimentation, sonication or degassing for bubbles, and using index-matching fluids (e.g., oils or solvents tuned to the sample's refractive index) to mitigate refractive mismatches causing artifacts. Data collection in DLS involves recording the time series of scattered light intensity fluctuations over 10–100 seconds per run to capture sufficient Brownian motion statistics for autocorrelation analysis. This duration balances signal-to-noise ratio while avoiding drift from sample instability, with raw intensity traces stored for subsequent processing.

Theoretical Foundations

Light Scattering Theory

Light scattering theory provides the foundational framework for understanding how coherent light interacts with particles in suspension, leading to the intensity patterns observed in dynamic light scattering (DLS) experiments. In static light scattering, the time-averaged intensity of scattered light, I(\mathbf{q}), depends on the number of particles N, the form factor P(\mathbf{q}) which describes intra-particle interference, and the structure factor S(\mathbf{q}) which accounts for inter-particle correlations. Under the Rayleigh-Gans-Debye (RGD) approximation, valid for optically soft particles where the refractive index difference between particle and medium is small (|\tilde{m} - 1| \ll 1, with \tilde{m} = m_p / m_m) and the phase shift across the particle is modest ($2k a | \tilde{m} - 1 | \ll 1, where k = 2\pi n / \lambda is the wavenumber, a is the particle radius, n the medium refractive index, and \lambda the wavelength), the scattered intensity simplifies to I(\mathbf{q}) \propto N P(\mathbf{q}) S(\mathbf{q}). This approximation extends Rayleigh scattering to larger, non-absorbing particles while neglecting multiple scattering and assuming dilute suspensions. The scattering vector \mathbf{q} is central to both static and dynamic analyses, defined as the difference between the scattered wave vector \mathbf{k}_s and incident wave vector \mathbf{k}_i, \mathbf{q} = \mathbf{k}_s - \mathbf{k}_i. For elastic scattering in a medium, |\mathbf{k}_s| = |\mathbf{k}_i| = (2\pi n)/\lambda, and the magnitude follows from the geometry: q = (4\pi n / \lambda) \sin(\theta/2), where \theta is the scattering angle. This expression arises because the path difference for rays scattered from points separated by a distance r along the \mathbf{q} direction contributes a phase factor \exp(i \mathbf{q} \cdot \mathbf{r}), leading to interference patterns that probe length scales on the order of $2\pi / q. The dependence on n, \lambda, and \theta implies that smaller q (longer \lambda or smaller \theta) resolves larger structures, while larger q probes finer details. Key assumptions underpin these formulations: scattering is isotropic within the particle ensemble, with no absorption (real refractive indices), and particles are in thermal equilibrium undergoing Brownian motion without external forces. These conditions ensure that the scattered field is a coherent sum of contributions from individual scatterers, and fluctuations arise solely from diffusive motion. The transition to dynamic light scattering involves analyzing time-dependent fluctuations in the scattered electric field. The first-order field autocorrelation function, g^{(1)}(\tau) = \langle E(0) E^*(\tau) \rangle / \langle |E|^2 \rangle, captures these fluctuations, where \tau is the delay time and E(t) is the scattered field. For non-interacting spherical particles in Brownian diffusion, g^{(1)}(\tau) = \exp(-D q^2 \tau), where D is the diffusion coefficient related to particle size via the Stokes-Einstein equation. This relation links the decay rate of correlations directly to diffusive dynamics at the scale set by q.

Dynamic Aspects and Autocorrelation

In dynamic light scattering (DLS), the technique captures the time-dependent aspects of particle motion primarily through fluctuations in the scattered light intensity, which result from the random of suspended particles. These intensity variations occur because particles undergoing cause phase shifts in the scattered wavefronts, leading to constructive and destructive interference that modulates the detected light intensity over time. The timescale of these fluctuations is inversely related to particle size, with smaller particles diffusing faster and producing more rapid intensity changes. The central observable in DLS is the intensity autocorrelation function, denoted as g^{(2)}(\tau), which quantifies the correlation of intensity at different time delays \tau. This function is experimentally measured using photon correlation spectroscopy and provides insight into the dynamics without directly resolving individual particle trajectories. Under the assumption of Gaussian field statistics, applicable to scattered light from many independent particles, the Siegert relation connects the intensity autocorrelation to the underlying electric field autocorrelation g^{(1)}(\tau): g^{(2)}(\tau) = 1 + |g^{(1)}(\tau)|^2 This relation, derived from the statistical properties of thermal light, allows extraction of dynamic information from intensity measurements alone. For a monodisperse system of spherical particles, the field autocorrelation function takes a simple exponential form: g^{(1)}(\tau) = \exp(-q^2 D \tau) where q is the magnitude of the scattering vector and D is the translational diffusion coefficient of the particles, related to their hydrodynamic radius via the Stokes-Einstein equation. This single-exponential decay reflects the uniform diffusion rate across all particles, enabling straightforward determination of D from the decay constant. In contrast, polydisperse samples exhibit a more complex, multi-exponential decay in g^{(1)}(\tau), as the overall correlation is a weighted superposition of exponentials from different size classes, each with its own D. The effects of polydispersity broaden the decay profile and reduce the apparent decay rate compared to a monodisperse equivalent, complicating size interpretation. To recover the particle size distribution G(R), the autocorrelation is inverted via the Laplace transform: g^{(1)}(\tau) = \int_0^\infty G(R) \exp(-q^2 D(R) \tau) \, dR where D(R) depends on the hydrodynamic radius R; this ill-posed inversion requires regularization techniques for practical implementation. Practical DLS measurements of autocorrelation functions are susceptible to several noise sources that can obscure the true dynamic signal. Shot noise, arising from the Poisson statistics of photon detection, introduces random baseline fluctuations that become prominent at low scattering intensities or short accumulation times, degrading the signal-to-noise ratio. Additionally, afterpulsing in single-photon avalanche diode (SPAD) detectors—caused by trapped charge carriers triggering false subsequent detections—generates correlated artifacts, particularly at short \tau, which mimic artificial correlations and bias polydispersity estimates. Mitigating these requires optimized detector selection, sufficient photon counts, and post-processing corrections to ensure reliable dynamic analysis.

Data Analysis

Cumulant Method

The cumulant method provides a straightforward, model-independent approach to analyze the normalized first-order electric field autocorrelation function g^{(1)}(\tau) obtained from dynamic light scattering measurements, yielding the average diffusion coefficient and a measure of polydispersity. Introduced by in 1972, this technique employs a cumulant expansion of the logarithm of the autocorrelation function: \ln g^{(1)}(\tau) = -\Gamma \tau + \frac{\mu_2}{2} \tau^2 + \ higher\ order\ terms, where \Gamma is the first cumulant representing the average decay rate, and \mu_2 is the second cumulant capturing the variance in decay rates. The expansion is typically truncated at the second order for practical analysis, as higher cumulants are prone to noise amplification. To compute these parameters, the initial portion of \ln g^{(1)}(\tau) versus \tau is fitted with a quadratic polynomial, often using least-squares regression on data up to where the function decays to about 10-20% of its initial value to minimize noise effects. The first cumulant \Gamma is determined from the negative slope of the linear term, while the second cumulant \mu_2 is obtained from the curvature of the quadratic term. The average diffusion coefficient is then calculated as \langle D \rangle = \Gamma / q^2, where q = (4\pi n / \lambda) \sin(\theta/2) is the scattering vector magnitude, with n the refractive index, \lambda the wavelength, and \theta the scattering angle. The polydispersity index (PDI) is given by PDI = \mu_2 / \Gamma^2, providing a dimensionless measure of the relative width of the size distribution; values below 0.05 indicate nearly monodisperse samples, while PDI > 0.3 suggests significant polydispersity. From the diffusion coefficient, the z-averaged hydrodynamic radius R_h is derived using the Stokes-Einstein equation: R_h = \frac{k_B T}{6 \pi \eta \langle D \rangle}, where k_B is the Boltzmann constant, T the absolute temperature, and \eta the solvent viscosity at the measurement temperature. This conversion assumes spherical particles undergoing Brownian motion in a dilute suspension. The method is particularly advantageous for rapid assessment of samples with low polydispersity (PDI < 0.2), as it requires minimal computational resources and is implemented in most commercial DLS instruments, often yielding results in seconds. However, it is limited for broad or multimodal distributions, where the expansion deviates from the true autocorrelation shape, leading to biased estimates of \Gamma and PDI; in such cases, PDI values may appear artificially low or negative, signaling the need for more advanced analysis. The approach has been standardized by the International Organization for Standardization (ISO 22412:2017) for consistent reporting of DLS results.

Size Distribution Recovery

In dynamic light scattering (DLS), recovering the particle size distribution from measured autocorrelation functions involves solving an inverse problem formulated as a Laplace transform inversion. The normalized electric field autocorrelation function g^{(1)}(\tau) for a polydisperse suspension of spherical particles undergoing Brownian diffusion is expressed as g^{(1)}(\tau) = \int_0^\infty A(R) \exp\left( -q^2 D(R) \tau \right) \, dR, where A(R) represents the amplitude-weighted size distribution function proportional to the number density times the scattering intensity per particle, q = (4\pi n / \lambda) \sin(\theta/2) is the magnitude of the scattering vector (with n the refractive index, \lambda the wavelength, and \theta the scattering angle), and D(R) = k_B T / (6 \pi \eta R) is the translational diffusion coefficient derived from the Stokes-Einstein relation (k_B is the Boltzmann constant, T the absolute temperature, \eta the solvent viscosity, and R the hydrodynamic radius). This inversion to obtain A(R) from experimental g^{(1)}(\tau) is inherently ill-posed, characterized by severe sensitivity to experimental noise and non-uniqueness of solutions, as small perturbations in the data can lead to exponentially amplified errors in the recovered distribution. To address these issues, regularization methods are essential, typically involving the minimization of a functional that balances data fidelity and solution smoothness, such as \min \left[ \| g^{(1)}(\tau) - \int A(R) \exp(-q^2 D(R) \tau) \, dR \|^2 + \lambda \mathcal{R}[A(R)] \right], where \lambda > 0 is a regularization parameter and \mathcal{R} is a penalty term (e.g., integral of the squared second derivative of A(R) to enforce smoothness). The choice of \lambda often relies on criteria like the L-curve method, which plots the residual norm against the regularization norm to identify an optimal balance avoiding under- or over-regularization. Non-parametric approaches to this inversion avoid assuming a specific functional form for A(R), instead discretizing the size domain into a histogram of bins or expanding A(R) in a basis set of functions (e.g., step functions for histograms or orthogonal polynomials). In histogram methods, the integral is approximated as a linear system \mathbf{g} = \mathbf{K} \mathbf{a}, where \mathbf{a} are the amplitudes in each bin, \mathbf{K} is the kernel matrix with elements K_{ij} = \exp(-q^2 D(R_i) \tau_j) \Delta R, and regularization is applied to solve for \mathbf{a} stably; basis set expansions similarly transform the problem into a regularized least-squares fit over coefficients. These techniques provide a flexible recovery of multimodal distributions but require careful selection of the number of bins or basis functions to mitigate artifacts from the ill-posed nature. While cumulant analysis offers quick estimates of average sizes and polydispersity via moments of A(R), full non-parametric inversion is necessary for detailed characterization of heterogeneous samples. Validation of recovered size distributions typically involves cross-comparison with orthogonal techniques, such as (TEM) for direct visualization of particle morphology and size histograms, or (SLS) for complementary determination of weight-average sizes and confirmation of overall polydispersity trends. For instance, in studies of polymer latexes or protein aggregates, DLS distributions have shown good agreement with TEM-derived number distributions when accounting for differences in weighting (intensity vs. number) and potential dehydration effects in electron microscopy. Such validations underscore the reliability of regularized inversions for quantitative analysis, though discrepancies may arise in concentrated or anisotropic systems.

Advanced Algorithms

Advanced algorithms in dynamic light scattering (DLS) address the challenges of inverting functions to recover size distributions in cases of high polydispersity, , or multimodality, where simpler methods like cumulants fail. These methods solve the ill-posed problem by incorporating regularization to stabilize solutions against data perturbations. The CONTIN algorithm, developed by Stephen W. Provencher, employs constrained regularization with quadratic smoothing to invert noisy linear integral equations, adapting the Jansson-van Cittert iterative deconvolution for DLS applications. It minimizes a functional combining the chi-squared misfit to the data and a smoothness penalty on the distribution, ensuring non-negativity and smoothness while avoiding oscillations common in unregularized inversions. The core optimization can be outlined in pseudocode as follows:
Initialize distribution A(Γ) as uniform or from cumulant estimate
Set regularization parameter α via L-curve or discrepancy principle
While convergence not reached:
    Compute predicted autocorrelation G_pred(τ) = ∫ A(Γ) exp(-Γ τ) dΓ
    Update χ² = Σ [ (G_meas(τ) - G_pred(τ)) / σ(τ) ]²
    Minimize J = χ² + α ∫ [d²A/d(ln Γ)²]² d(ln Γ)
    Constrain A(Γ) ≥ 0 and normalize
Output size distribution via D = Γ / q^2 and R_h = k_B T / (6 π η D), where Γ is decay rate, q the scattering vector, D the [diffusion](/page/Diffusion) coefficient, k_B the [Boltzmann constant](/page/Boltzmann_constant), T the temperature, and η the viscosity
This approach excels in producing smooth, physically plausible distributions for moderately polydisperse samples. The maximum entropy method (MEM) provides an alternative by maximizing the functional subject to data fidelity, favoring the most unbiased distribution consistent with the measurements. It maximizes S = -\int p(\ln R) \ln p(\ln R) \, d(\ln R), where p(\ln R) is the probability density in log-radius space, constrained by \chi^2 \approx N () to fit the within noise. MEM is particularly effective for or sparse datasets, as it avoids over-smoothing and naturally handles underdetermined problems by spreading probability over unresolved components. A outline is:
Initialize p(ln R) as flat or Jeffreys prior
Set Lagrange multipliers for χ² and entropy constraints
Iterate via gradient ascent or quadratic approximation:
    Update p(ln R) ∝ exp( ∑ λ_k K_k(ln R) ), where K_k are kernel basis functions
    Adjust λ to satisfy χ² = N and maximize S
    Transform to size distribution f(R) = p(ln R) / R
Converge when ΔS < ε
This method has been applied successfully to colloidal dispersions, resolving peaks separated by factors of 10 in size. Comparisons show CONTIN suits smooth, unimodal or mildly polydisperse distributions by enforcing curvature penalties, yielding stable results with lower computational cost for well-sampled data, whereas performs better on noisy or multimodal data by preserving sharpness without artificial broadening, though it requires more iterations. Software implementations include the suite from Wyatt Technology, which integrates CONTIN for multi-angle DLS analysis, and macros that interface with CONTIN executables or implement MEM via built-in optimization routines, enabling user-friendly distribution recovery with exportable fits. Post-2010 advances incorporate Bayesian frameworks for , treating the size distribution as a given the data and priors on polydispersity. These methods, such as sparse Bayesian learning, use hierarchical models to infer not only the distribution but also confidence intervals, improving robustness for low-signal or heterogeneous samples. For instance, Bayesian inversion via sampling has demonstrated reduced bias in nanoparticle sizing compared to deterministic regularization. From 2020 onward, techniques have emerged as powerful tools for DLS data analysis, particularly for tackling the ill-posed inversion in complex scenarios. Deep neural networks (DNNs), for example, can process raw time-resolved scattering signals to predict s with high accuracy, achieving errors below 1% for microparticles up to 150 μm by identifying key single-scattering regions, as demonstrated in studies on silica suspensions as of 2025. Additionally, the CORENN , developed around 2020, employs a constrained, regularized non-linear least-squares fit with optimization to extract multimodal distributions robustly, outperforming CONTIN for heterogeneous samples in colloidal and biological applications.

Advanced Considerations

Multiple Scattering Effects

In dynamic light scattering (DLS), multiple scattering effects arise when incident light undergoes more than one scattering event within the sample before reaching the detector, which becomes prominent in concentrated suspensions where the optical path length allows photons to interact with multiple particles. This phenomenon distorts the measured intensity autocorrelation function by introducing contributions from photons with altered trajectories and phase shifts, typically resulting in a faster apparent decay and thus an underestimation of particle sizes, as the effective diffusion appears accelerated compared to single scattering scenarios. In contrast to single scattering theory, where the autocorrelation strictly follows the Siegert relation linking field and intensity correlations, multiple scattering deviates from this, reducing the coherence factor β in the relation G(τ) = 1 + β |g(τ)|², where g(τ) is the normalized field autocorrelation. Theoretical models for multiple scattering in DLS often employ coupled dipole approximations to simulate interactions among scatterers, treating particles as induced s that collectively influence the scattered , particularly useful for predicting enhancements in perturbations in random . simulations provide another key approach, stochastically tracing photon paths to compute the autocorrelation function in regimes transitioning from to highly multiple , enabling comparisons between and backscattering geometries to quantify distortions. These models reveal that multiple increases with sample , often parameterized by the τ, where higher values lead to greater deviations in the apparent D_app from the true value. Quantification of multiple scattering can be achieved through turbidity measurements, which assess sample opacity via transmitted light intensity to estimate the likelihood of multiple events, with higher turbidity correlating to increased distortions in DLS signals. Deviations in the Siegert factor, observed as a drop in the autocorrelation intercept below ideal values (e.g., approaching 0.5 for dominant multiple scattering), serve as a diagnostic, allowing researchers to validate single scattering dominance. Modern correction techniques, such as photon cross-correlation spectroscopy, suppress multiple scattering by correlating intensity fluctuations from two spatially separated detection channels or beams, isolating single-scattered contributions and extending DLS applicability to turbid samples with transmissions as low as 1%. To mitigate multiple scattering, strategies include operating at low particle concentrations to keep the sample below the single scattering limit, typically ensuring transmissions above 90-95%. Index matching, by selecting solvents with refractive indices closely matching the particles (e.g., within 0.01 units), reduces the overall cross-section and thus the probability of multiple events. DLS further aids suppression, using parallel () polarization to favor single scattering while perpendicular (VH) components highlight multiples for diagnostic purposes, often combined with detection to shorten effective path lengths.

Non-Spherical Particle Scattering

Dynamic light scattering (DLS) measurements on non-spherical particles encounter significant challenges arising from their anisotropic properties. Unlike spherical particles, non-spherical ones exhibit anisotropic translational diffusion, with distinct diffusion coefficients along the principal axes, which complicates the decay of the intensity autocorrelation function beyond a simple single-exponential form. Orientation-dependent scattering further arises because the scattered field amplitude varies with the particle's alignment relative to the scattering vector \mathbf{q}, introducing angular averaging effects that broaden the correlation function. For shapes like rods or disks, these factors cause a departure from the Stokes-Einstein relation, as the effective hydrodynamic radius depends on the aspect ratio and cannot be directly equated to an equivalent sphere without shape-specific corrections. To model these behaviors, adaptations incorporate rotational diffusion via the Perrin equations, which extend the Stokes-Einstein-Debye framework to ellipsoids by providing expressions for the parallel and perpendicular rotational diffusion coefficients in terms of the semi-axes a, b, and solvent properties. For cylindrical particles, such as rods, the scattering requires form factor modifications to account for internal structure along the length. A key example is the orientationally averaged form factor for a thin rod of length L, given by P(q) = \left[ \frac{2 \mathrm{Si}(qL)}{qL} - \left( \frac{\sin(qL/2)}{qL/2} \right)^2 \right], where \mathrm{Si}(x) = \int_0^x \frac{\sin t}{t} \, dt is the sine integral function, and this modulates the dynamic structure factor in the intermediate scattering regime. These models enable prediction of the coupled translational-rotational dynamics observed in the autocorrelation. Extensions to data analysis combine translational and rotational contributions through cumulant expansions of the field autocorrelation function, where the first cumulant yields an average diffusion rate and the second captures polydispersity and anisotropy effects. Depolarized DLS (DDLS), which measures the perpendicular-polarized scattered light, isolates rotational relaxation by suppressing the dominant translational signal, thereby providing direct access to shape-sensitive rotational diffusion coefficients. For instance, in protein aggregates, DDLS reveals aspherical conformations through faster rotational decays compared to isotropic models, as demonstrated in studies of fibrinogen and monoclonal antibody oligomers where anisotropy correlates with aggregation pathways. Despite these advances, analyzing non-spherical systems introduces limitations, particularly the heightened complexity in inverting data to retrieve polydisperse and distributions, since multiple coupled parameters (e.g., aspect ratios and orientations) lead to underdetermined fits and require regularization techniques.

Applications

Biological and Pharmaceutical Uses

Dynamic light scattering (DLS) plays a crucial role in monitoring in biological and pharmaceutical contexts, particularly for therapeutic proteins like monoclonal antibodies (mAbs). In vaccine development, DLS detects early formation by measuring changes in the , enabling the assessment of stability under various formulation conditions. For instance, high-throughput DLS techniques have been employed to evaluate mAb aggregation kinetics, identifying critical quality attributes such as aggregate distributions that impact and . This approach is vital for ensuring the safety of biologics, as aggregates can trigger unwanted immune responses. In nanoparticle-based systems, DLS is widely used for sizing liposomes and viral vectors, which typically range from 20 to 200 nm and are essential for targeted therapies. For liposomes, DLS provides rapid, non-destructive of hydrodynamic and polydispersity , aiding in the optimization of encapsulation and release profiles for chemotherapeutic agents. Similarly, in , DLS quantifies adeno-associated virus (AAV) vector sizes and aggregation states, supporting the development of stable formulations for clinical use. These measurements help predict performance, such as biodistribution and cellular uptake. Electrophoretic DLS extends these applications by assessing , a key indicator of surface charge and colloidal stability, which correlates with in pharmaceutical formulations. This technique evaluates how charge influences interactions with biological membranes, reducing risks of or rapid clearance in systems like nanoparticles. For example, zeta potential measurements via electrophoretic light scattering have been integrated into testing protocols for liposomal carriers, ensuring electrostatic repulsion prevents aggregation in physiological environments. Case studies illustrate DLS's impact in specific pharmaceutical scenarios. In insulin formulations, DLS monitors aggregation under stress conditions, revealing how factors like and zinc influence oligomer sizes and long-term stability, which is critical for . For mRNA-lipid complexes in vaccines, DLS has characterized particle sizes around 80-100 nm, confirming uniformity and stability post-manufacturing, which directly affected in clinical trials. These examples underscore DLS's role in advancing .

Materials Science and Nanotechnology

In and , dynamic light scattering (DLS) is indispensable for characterizing the size distribution, aggregation behavior, and temporal evolution of , enabling precise control over parameters to achieve desired properties such as optical responsiveness or mechanical stability. By quantifying coefficients in colloidal suspensions, DLS reveals hydrodynamic diameters typically in the 1-100 range, which correlate with particle morphology and surface interactions critical for advanced . This non-invasive technique complements electron microscopy by providing ensemble-averaged data under native solution conditions, facilitating optimization of nanomaterial performance in composites, coatings, and energy devices. For polymer solutions, DLS elucidates molecular weight and chain conformation by measuring translational diffusion coefficients, which decrease with increasing polymer chain length due to larger hydrodynamic volumes. In dilute regimes, the diffusion coefficient D relates to the hydrodynamic radius R_h via the Stokes-Einstein equation, D = kT / (6\pi \eta R_h), where k is Boltzmann's constant, T is temperature, and \eta is solvent viscosity, allowing indirect determination of weight-average molecular weights from calibrated standards. This approach has been applied to polystyrene and polyethylene oxide solutions, revealing conformational transitions from coils to rods under varying solvent quality, with polydispersity indices indicating chain uniformity. DLS is widely used for sizing colloidal nanoparticles during synthesis optimization, particularly for gold and silica systems where uniform 1-100 nm particles enhance plasmonic or catalytic properties. In gold nanoparticle synthesis via citrate reduction, DLS monitors real-time growth from seeds to final assemblies, with hydrodynamic sizes correlating to reaction pH and temperature for yields of 10-50 nm spheres with low aggregation. Similarly, for silica nanoparticles produced by sol-gel methods, DLS optimizes parameters like precursor concentration and ammonia catalysis, achieving mesoporous structures with average diameters of 20-80 nm and narrow distributions essential for drug delivery scaffolds or photonic materials. In studies, DLS probes the dynamics of and formation, detecting critical aggregation concentrations through shifts in apparent coefficients as monomers form ordered nanostructures. For amphiphilic block copolymers, DLS reveals core-shell architectures with sizes evolving from 10 nm unimers to 50-200 nm aggregates, guiding the design of responsive for encapsulation. In-situ DLS monitoring tracks reaction kinetics in , capturing bursts and phases for like metal oxides. By integrating fiber-optic probes into autoclaves, DLS measures evolving particle sizes under elevated temperatures and pressures, optimizing dwell times for uniform 10-50 nm crystals in applications. This real-time feedback minimizes polydispersity, ensuring reproducible outcomes in scalable processes.

References

  1. [1]
    Dynamic light scattering: a practical guide and applications in ...
    Oct 6, 2016 · Dynamic light scattering (DLS), also known as photon correlation spectroscopy (PCS), is a very powerful tool for studying the diffusion behaviour of ...
  2. [2]
    Dynamic Light Scattering (DLS): Principles, Perspectives ... - NIH
    Focus a laser on dissolved particles and analyze the scattered light to reveal their size. This well established principle is used in dynamic light scattering ...Fig. 6.1 · Sample Preparation And... · Latest Methods And...
  3. [3]
  4. [4]
  5. [5]
  6. [6]
    The Malvern Correlator: The world's first commercial laser-based ...
    Dec 21, 2017 · The Malvern Correlator: The world's first commercial laser-based, particle-size analyzer ... On the 20th December 1977 (40 years ago) the Royal ...
  7. [7]
    Microfluidic in-line dynamic light scattering with a commercial fibre ...
    Apr 18, 2023 · We report the coupling of dynamic light scattering (DLS) in microfluidics, using a contact-free fibre-optic system, enabling the under-flow characterisation.
  8. [8]
    Improved Dynamic Light Scattering using an adaptive and ... - NIH
    Sep 18, 2019 · In this work we describe and assess a new DLS measurement and data treatment process which cuts the photon arrival time data from the detector ...
  9. [9]
    None
    ### Summary of Main Components of a Dynamic Light Scattering System
  10. [10]
    Dynamic Light Scattering (DLS) Particle Analyzer | MICROTRAC
    Of the two approaches, the heterodyne mode with "controlled reference" offers many advantages over the homodyne setup in a dynamic light scattering analyzer.
  11. [11]
    [PDF] Measurement of Dynamic Light Scattering Intensity in Gels - HAL
    Mar 17, 2015 · In practice, account should be taken of the dead time τd of the detector and its noise b. The extent to which the coherence factor depends on ...
  12. [12]
    Sample preparation: DLS - Polymer Chemistry Characterization Lab
    Filters 0.1 – 0.2 µm can be used on pure or prepared solvents, prior to combining with sample. Filters 0.2 – 0.45 µm are usually suitable for prepared samples.
  13. [13]
    Experimental Guidelines - DLS Sample Preparation - LS Instruments
    The measurement principle of Dynamic Light Scattering (DLS) is based on the Brownian motion of particles. This restricts the range of samples that can be ...
  14. [14]
    [PDF] DLS Data Interpretation | Entegris
    Sometimes the best approach is to make multiple runs with several analysis per run - then take an average of the multiple runs. It is not unusual to not ...
  15. [15]
    Pitfalls and novel applications of particle sizing by dynamic light ...
    The scattering angles typically utilized are 90 or 173°, i.e. large angles at which the influence of potential “dust contaminations” on the correlation ...
  16. [16]
    [PDF] DLS Sample Preparation | Entegris
    Aug 12, 2019 · This technical note guides a Nicomp® DLS user through the various steps that may be required to generate accurate and repeatable results.
  17. [17]
    Improved Dynamic Light Scattering using an adaptive and ... - Nature
    Sep 18, 2019 · Dynamic Light Scattering (DLS) is a ubiquitous and non-invasive measurement for the characterization of nano- and micro-scale particles in ...
  18. [18]
  19. [19]
    Dynamic Light Scattering: With Applications to Chemistry, Biology ...
    Dynamic Light Scattering: With Applications to Chemistry, Biology, and Physics Dover Books on Physics Series. Authors, Bruce J. Berne, Robert Pecora. Edition ...
  20. [20]
    Dynamic Light Scattering Microscopy. A Novel Optical Technique to ...
    Photon shot noise contributes to the amplitude of gT(τ), but only at τ = 0 since the photon arrival times are uncorrelated. Thus, to obtain an accurate ...
  21. [21]
  22. [22]
    Forward and inverse analysis for particle size distribution ...
    The main purpose of this review is to coherently present an updated theoretical background of forward and inverse analysis for particle size distribution ...Missing: seminal | Show results with:seminal
  23. [23]
    Particle Size Inversion Constrained by L∞ Norm for Dynamic Light ...
    Oct 13, 2022 · Particle size inversion of dynamic light scattering (DLS) is a typically ill-posed problem. Regularization is an effective method to solve the problem.
  24. [24]
    Dynamic light scattering and transmission electron microscopy in ...
    Sep 26, 2023 · Since DLS is more sensitive to larger particles, the scattering intensity will be proportional to the sixth power of their radius. In this case ...
  25. [25]
    The CONTIN algorithm and its application to determine the size ...
    Jun 17, 2015 · We review a powerful regularization method, known as CONTIN, for obtaining the size distribution of colloidal suspensions from dynamic light scattering data.
  26. [26]
    Maximum entropy analysis of quasielastic light scattering from ...
    May 1, 1986 · The maximum entropy method (MEM) is used to analyze quasielastic light scattering correlation functions. The central problem, in such data ...Missing: DLS | Show results with:DLS
  27. [27]
    [PDF] A Practical Minicourse in Dynamic Light Scattering *
    Scattering explained as diffraction from Fourier components selected by the scattering geometry. ... scattered light through the pinhole via the ocular of the DLS ...
  28. [28]
    DYNAMICS Software - Waters | Wyatt Technology
    DYNAMICS: the intuitive software for acquiring and analyzing size, molar mass and zeta potential of your DynaPro Plate Reader, NanoStar or ZetaStar.Missing: implementations CONTIN MEM IGOR Pro
  29. [29]
    CONTIN and Igor Pro
    Jul 19, 2010 · CONTIN, a Windows version, can be called from Igor. Igor can prepare data for CONTIN, but no direct port exists. Igor can use the Windows ...Missing: implementations MEM
  30. [30]
    Size distribution of nanoparticles by dynamic light scattering ...
    The diameter distribution of nanometric particles is estimated from multiangle dynamic light scattering (MDLS) measurements by solving an ill-conditioned ...Missing: post- | Show results with:post-
  31. [31]
    Sparse Bayesian learning for the Laplace transform inversion in ...
    For the analysis of dynamic light scattering (DLS) data, the use of a statistical Bayesian learning is suitable for making an inference about the sizes of ...Missing: seminal | Show results with:seminal
  32. [32]
    Multiple scattering effects on intercept, size, polydispersity index ...
    Dec 10, 2020 · Typically, the intensity-intensity autocorrelation function, G(τ), then shows a decay rate that is directly related to the Brownian movement of ...
  33. [33]
    Connecting field and intensity correlations: The Siegert relation and ...
    Oct 1, 2020 · The Siegert relation is derived in Sec. II for photons emitted by a large number of independent emitters. Experimentally, this can be achieved, ...
  34. [34]
    [PDF] arXiv:2105.02800v1 [physics.optics] 6 May 2021
    May 6, 2021 · Using a coupled dipole model we investigate how multiple scattering can enhance field perturbations arising in such random scat- tering based ...
  35. [35]
    Dynamic light scattering Monte Carlo: a method for simulating time ...
    Abstract. Few methods exist that can accurately handle dynamic light scattering in the regime between single and highly multiple scattering.Missing: dipole | Show results with:dipole
  36. [36]
    Monte Carlo multiple-scattering simulation and data correction in ...
    Jun 1, 1997 · Monte Carlo methods usually employed for γ- and x-ray problems have recently been utilized for simulating multiple Rayleigh light scattering ...Missing: dynamic theoretical models coupled dipole
  37. [37]
    Suppression of multiple scattering by photon cross-correlation ...
    This review will concentrate on multiple scattering suppression in turbid samples by cross-correlation. Diffusing wave spectroscopy was reviewed in this ...Missing: papers | Show results with:papers
  38. [38]
    Multiple scattering suppression in static light scattering by cross ...
    Nov 1, 1999 · Cross-correlation techniques have been used successfully to suppress multiple scattering in dynamic light-scattering experiments on turbid ...
  39. [39]
    Dynamic Light Scattering - Book - SpringerLink
    Book Title: Dynamic Light Scattering · Book Subtitle: Applications of Photon Correlation Spectroscopy · Editors: Robert Pecora · Publisher: Springer New York, NY.
  40. [40]
    [PDF] Dynamic Light Scattering: An Introduction in 30 Minutes
    The diameter that is obtained by this technique is the diameter of a sphere that has the same translational diffusion coefficient as the particle.
  41. [41]
    [PDF] Variations in Structure Explain the Viscometric Behavior of AOT ...
    to the form factor of an infinitely thin rod. To ... P(q) as the orientationally-averaged form factor ... rod, given in terms of the sine integral Si ...<|control11|><|separator|>
  42. [42]
    Size and Shape Determination of Proteins in Solution by a ...
    Jan 12, 2006 · We present here measurements with a variant of DLS that determines rotational diffusion as well as translation diffusion. This technique, called ...
  43. [43]
    Advanced scattering techniques for characterisation of complex ...
    Here we review the current state of the art in the characterisation of complex nanoparticles (non-spherical and multi-layered) using advanced scattering ...
  44. [44]
    High-Throughput Analytical Light Scattering for Protein Quality ...
    We present a review of high-throughput techniques for the characterization and quality control of proteins in the course of purification, evaluation, and ...
  45. [45]
    A New Approach to Study the Physical Stability of Monoclonal ...
    In this study, a dynamic light scattering plate reader is used to measure the aggregation (by means of the increase in the hydrodynamic radius [Rh]) of ...
  46. [46]
    Understanding and controlling the molecular mechanisms of protein ...
    In this review, we describe the theoretical molecular mechanisms for the aggregation of antibodies, and how it can be promoted by various stress factors.Research Review Paper · 3. Molecular Events In... · 6.2. Extrinsic Factors: Mab...
  47. [47]
    Liposomes characterization for market approval as pharmaceutical ...
    Nov 30, 2023 · The PDI of a dispersion and its mean hydrodynamic radius (Rh) instead can be accurately measured using dynamic light scattering (DLS) [67], ...
  48. [48]
    Comprehensive characterization and quantification of adeno ...
    Feb 4, 2021 · Various methods, including electron microscopy, dynamic light scattering, analytical ultracentrifugation, ELISA, and PCR are used to ...
  49. [49]
    Characterization of AAV vectors: A review of analytical techniques ...
    ... viral vectors that should be central in ensuring consistent characterization of viral vectors. ... virus with multi-angle dynamic light scattering (MADLS).
  50. [50]
    Development of chitosan, gelatin and liposome film and analysis of ...
    Liposomes size distribution and zeta potential were evaluated by using dynamic light scattering (DLS) and electrophoretic mobility (Zetasizer Nano ZS, Malvern).
  51. [51]
    Characterization of Sizes of Aggregates of Insulin Analogs ... - PubMed
    Characterization of Sizes of Aggregates of Insulin Analogs and the Conformations of the Constituent Protein Molecules: A Concomitant Dynamic Light Scattering ...
  52. [52]
    Impact of lipid nanoparticle size on mRNA vaccine immunogenicity
    Jul 10, 2021 · Available biophysical data for all LNPs were average particle size (determined by dynamic light scattering; DLS) and RNA entrapment (determined ...
  53. [53]
    mRNA-lipid nanoparticle COVID-19 vaccines: Structure and stability
    May 15, 2021 · ... dynamic light scattering; ESEM, environmental scanning electron microscopy; IEX-HPLC, ion-exchange high performance liquid chromatography ...
  54. [54]
    Model for Gold Nanoparticle Synthesis: Effect of pH and Reaction Time
    Jun 24, 2021 · The synthesis of gold nanoparticles is dependent on both the concentration of trisodium citrate dihydrate and the time that it interacts with tetrachloroauric ...Introduction · Results and Discussion · Synthesis Analysis · Experimental Section
  55. [55]
    Optimization of Mesoporous Silica Nanoparticles through Statistical ...
    Jan 31, 2021 · Abstract. The synthesis process or composition of mesoporous silica nanoparticles (MSNs) affects the physicochemical properties.
  56. [56]
    Insights into the Characterization of the Self-Assembly of Different ...
    Dec 4, 2023 · Insights into the Characterization of the Self-Assembly of Different Types of Amphiphilic Molecules Using Dynamic Light Scattering. Click to ...Introduction · Experimental Part · Results and Discussion · References
  57. [57]
    Tuning Perovskite Nanocrystal Synthesis via Amphiphilic Block ...
    The dynamic light scattering (DLS) measurements were performed using a ZEN2003 Zeta upgrade of MALVERN. UV–vis absorbance spectra were measured using a JASCO V- ...
  58. [58]
    Dynamic Light Scattering: A Powerful Tool for In Situ Nanoparticle ...
    As discussed, DLS is a very versatile approach that may be use in situ to characterize samples in a variety of phase states and physico-chemical environments.