Fluorometer

A fluorometer, also known as a fluorimeter, is an analytical instrument used to measure the emitted by a sample, quantifying parameters such as the intensity and distribution of in the after by a specific . This phenomenon, known as , occurs when molecules absorb photons at one and re-emit them at a longer , enabling highly sensitive detection of analytes at concentrations as low as . Fluorometers operate on the principle of Stokes fluorescence, where a light source—such as a lamp, LED, or —excites the sample with selected wavelengths, and a detector, often a , captures the resulting emission after passing through optical filters or monochromators to isolate signals. Early designs relied on basic components, but modern systems incorporate advanced features like automated wavelength selection and flow cells for continuous monitoring. The technology's sensitivity surpasses traditional by 1,000 to 500,000 times, making it ideal for trace analysis without requiring complex sample preparation. The development of fluorometers traces back to the mid-20th century, building on the discovery of in 1845 by , who observed light exciting solutions to emit blue light. Commercial instruments emerged in the 1950s, with companies like Farrand Optical producing early models for laboratory use, and the first spectrophotofluorometer invented in 1951 by Robert Bowman at the to enable precise quantification of biological substances. These advancements revolutionized trace detection, initially applied in fields like tracing with fluorescein dyes as early as 1877. Key applications of fluorometers span environmental monitoring, biotechnology, and industrial processes, including the detection of chlorophyll a in oceanic samples to assess algal blooms, quantification of DNA and proteins in molecular biology, and real-time analysis of pollutants like aromatic hydrocarbons in water. In ecological studies, portable fluorometers facilitate in situ measurements of phytoplankton fluorescence for water quality assessment, while benchtop models support high-throughput assays in research labs. Ongoing innovations, such as integration with microfluidic systems, continue to expand their utility in diagnostics and remote sensing.

Principles of Operation

Fluorescence Phenomenon

is the emission of light from a substance that has absorbed of light or other of higher , resulting in the release of lower-energy typically at longer wavelengths, a phenomenon known as the . This shift arises because the emitted carries less than the absorbed one, as some is lost through non-radiative processes like vibrational relaxation in the . The process begins with the of a —a or material capable of —from its ground (S₀) to a higher-energy (S₁ or S₂) upon absorption. The key steps in fluorescence are depicted in the , a representation of electronic and vibrational energy levels. Absorption rapidly promotes the to an excited vibrational level within S₁ or S₂ (on the order of 10⁻¹⁵ seconds), followed by ultrafast vibrational relaxation to the lowest vibrational level of S₁ (10⁻¹² to 10⁻¹¹ seconds). From this state, the returns to S₀ by emitting a (, lasting about 10⁻⁹ seconds) or through non-radiative decay pathways. This singlet-to-singlet transition distinguishes from , which involves a spin-forbidden to a (T₁) and subsequent emission with lifetimes ranging from milliseconds to seconds. Several factors govern the intensity and characteristics of . The (Φ) quantifies the efficiency of the process, defined as the ratio of the number of photons emitted to the number of photons absorbed by the . (ε), also known as the molar , indicates the strength of light absorption at a given and follows Beer's , where absorbance A = ε × c × l (c is concentration, l is path length). The lifetime (τ) represents the average time the remains in the before deactivation, typically on the scale for most . Common include dyes like fluorescein, which exhibits strong absorption around 494 nm and at 512 nm; quantum dots, valued for their tunable and high photostability; and fluorescent proteins such as (GFP), derived from and widely used in biological imaging due to its genetic encodability. Fluorescence efficiency can be diminished by quenching, where excited-state energy is dissipated without . Collisional (dynamic) quenching occurs when a quencher diffuses to and collides with the excited during its lifetime, promoting non-radiative decay. In contrast, static quenching involves the formation of a non-fluorescent ground-state between the and quencher prior to , preventing altogether. These mechanisms are crucial for understanding modulation in various environments.

Detection and Quantification

In fluorometers, the excitation-emission process begins with a monochromatic source directing onto the sample, prompting fluorophores to absorb photons and emit at longer wavelengths. The emitted is typically collected at a 90-degree relative to the to minimize scatter and stray interference, enhancing signal purity. The captured fluorescence undergoes signal processing where photodetectors, such as photomultiplier tubes, convert emitted photons into electrical signals proportional to photon flux. These signals are then amplified to boost weak emissions and passed through spectral filters or monochromators to isolate the fluorescence wavelength from residual excitation light or background noise. Quantification relies on measuring fluorescence intensity, which follows the relation I_f = k \Phi I_0 c l, where I_f is the fluorescence intensity, k is the instrument constant, \Phi is the , I_0 is the intensity, c is the concentration, and l is the path length; this holds for dilute solutions where is low./3:_Molecular_Luminescence/3.6:_Variables_that_Influence_Fluorescence_Measurements) Fluorometers achieve high , detecting certain fluorophores at nanomolar concentrations (e.g., limits of detection around 0.1–7 for optimized probes). Key error sources in detection include the inner filter effect, where high sample (exceeding 0.05 in a 1 cm path length) reabsorbs emitted light or attenuates , leading to nonlinear readings, and , the irreversible degradation of fluorophores under prolonged high- illumination, which reduces signal over time.

Instrument Design and Components

Optical System

The optical system of a fluorometer is designed to efficiently deliver to the sample and collect the resulting emission while minimizing interference from scattered or . Central to this system is the path, which begins with a light source capable of providing intense illumination across a broad spectral range, typically from 200 to 800 to match common wavelengths. Common light sources include arc lamps, which offer a continuous from to near-infrared and are suitable for both steady-state and pulsed operations in portable systems with power ratings of 5–60 W; light-emitting diodes (LEDs), which provide narrow-band emission (e.g., at 385 or 470 ) with high and low for targeted excitations; and lasers, such as 405 lasers with linewidths ≤1 , for monochromatic and coherent illumination in specialized applications. Wavelength selection in the path is achieved using monochromators, often grating-based for tunable selection with slit widths adjustable for versus trade-offs, or bandpass filters, such as those with 10–15 nm and ~40% , for fixed- setups that reduce complexity in portable devices (e.g., 440 nm filters). Following selection, the light is directed to the sample via lenses or fiber optics to ensure uniform illumination. The path collects at a right-angle (90°) to the beam, which orthogonally separates the weaker signal from the intense light and reduces scatter. Collection , including lenses and fiber , focus the emitted light onto wavelength-selective elements such as monochromators for scanning or filters (e.g., 525 nm bandpass or >500 nm longpass dichroic mirrors) to isolate specific bands while blocking residual . Beam splitters, often dichroic, further enable simultaneous routing of to detectors and channels for correction. Sample compartments are tailored to the analysis type, featuring holders for standard liquid samples. cuvettes are used for the UV-visible range (200–800 ) to minimize losses, while or cuvettes suffice for visible wavelengths; examples include square or round cuvettes with microcell options for small volumes. Flow cells, such as those made from (PDMS) in microfluidic setups, support continuous or high-throughput measurements, while integrations with microscopes or attachments enable imaging-based in biological samples. Optical alignment prioritizes efficient light throughput and rejection, achieved through collimating lenses and diffusers for even distribution, pinholes to limit , and dark enclosures or fiber optics (e.g., 200 m lengths in ) to guide light with minimal loss in portable models. Early fluorometer designs relied on slit-based monochromators for sequential scanning, but modern systems have evolved to incorporate array detectors like charge-coupled devices (CCDs) or complementary metal-oxide-semiconductor () sensors, enabling simultaneous multi- detection for faster spectral acquisition and improved in applications.

Electronic and Data Processing Elements

The electronic and data processing elements of a fluorometer handle the conversion, amplification, and analysis of weak signals generated from optical detection, ensuring high fidelity in measurement. Detectors are critical for capturing emitted photons with minimal loss. Photomultiplier tubes (PMTs) are widely used due to their high sensitivity, capable of detecting single photons through internal electron multiplication, making them ideal for low-light applications. Photodiodes offer cost-effectiveness and faster response times for moderate signal levels, while (CCD) arrays enable spectral imaging by simultaneously recording multiple wavelengths across a detector plane. Following detection, signals undergo and to enhance signal-to-noise ratios. Lock-in amplifiers synchronize with modulated excitation sources to extract the fluorescence signal at specific frequencies, effectively suppressing broadband noise and improving measurement precision in noisy environments. Analog-to-digital converters (ADCs) then digitize the amplified signals, typically with 16-24 bit resolution to preserve and resolve subtle intensity variations in fluorescence data. Data acquisition software manages the collection and initial processing of these signals, enabling real-time through algorithms that deconvolute emission profiles from contributions. Integration times range from milliseconds for rapid kinetics to seconds for accumulating weak signals, optimizing sensitivity based on sample properties. Output formats, such as files, facilitate export of spectra for further analysis in external tools like spreadsheets or specialized software. In portable fluorometers, microcontrollers integrate these elements for field deployment, controlling timing sequences, data logging, and distribution. These compact processors, such as Arduino-based systems, manage battery life by optimizing duty cycles and monitoring charge levels to support extended operation without mains . Safety features protect users and equipment from hazards associated with high-intensity light sources. Interlocks automatically disable the excitation lamp if access panels are opened, preventing exposure to or intense visible . Automated shutdown mechanisms monitor for overheating in detectors or power supplies, triggering -off to avoid damage or fire risks.

Types of Fluorometers

Steady-State Fluorometers

Steady-state fluorometers are instruments designed to measure the average of a sample under continuous, constant , providing -based quantification without resolving temporal dynamics. These devices rely on a steady source, typically a continuous xenon lamp, to illuminate the sample and detect the resulting steady-state emission signal, which reflects the time-averaged properties at . Steady-state fluorometers are broadly classified into two subtypes: filter fluorometers and spectrofluorometers. Filter fluorometers employ optical filters to select specific and wavelengths, offering broadband selectivity suitable for low-cost, routine measurements at fixed wavelengths; they are particularly advantageous for high-throughput applications involving consistent analytes due to their simplicity and affordability. In contrast, spectrofluorometers utilize scanning monochromators, often gratings, to achieve wavelength selection across full and spectra with resolutions typically between 1 nm and 10 nm, enabling detailed for complex samples. Since the , steady-state fluorometers have dominated fluorescence instrumentation, with early commercial models like the Farrand Optical Fluorometer Model A marking the onset of widespread adoption in settings. This prevalence stems from their reliability for routine quantitative assays, evolving from basic filter-based designs to advanced spectrofluorometers integrated into standard lab workflows. By the early , various research groups had developed these instruments for applications such as NADH monitoring, solidifying their role in . Performance characteristics of steady-state fluorometers include and wavelength ranges spanning the ultraviolet-visible (UV-Vis) spectrum from approximately 200 to 900 , extendable to the near-infrared (near-IR) up to 1650 with specialized detectors. They exhibit high , typically over 5 to 6 orders of magnitude in concentration, ensuring accurate quantification across a wide for analytes like nucleic acids or fluorescent dyes. Signal-to-noise ratios can exceed 800 (RMS) in high-sensitivity models, supporting detection limits down to picomolar concentrations. Representative examples include benchtop spectrofluorometers such as the F-7000, which features ultra-high-speed scanning at 60,000 nm/min and supports across UV-Vis to near-IR ranges for applications requiring full emission spectra. This model exemplifies the compact, high-sensitivity design prevalent in modern routine laboratories, with a surpassing 300 for reliable quantitative measurements.

Time-Resolved Fluorometers

Time-resolved fluorometers measure the lifetime, denoted as τ, which represents the average time a spends in the before returning to the and emitting a . This temporal analysis provides insights into and environmental interactions that steady-state intensity measurements cannot capture. Unlike steady-state approaches, time-resolved methods focus on the decay kinetics following pulsed , enabling the discrimination of fluorophores based on their microenvironmental influences. The primary techniques employed are time-correlated single photon counting (TCSPC) and . In TCSPC, a pulsed light source, such as (ns) to (fs) lasers or LEDs with repetition rates from 10 kHz to 100 MHz, excites the sample, and the arrival time of individual emitted is recorded relative to the excitation pulse using high-speed detectors like photomultiplier tubes (PMTs) or single-photon avalanche diodes (SPADs). The resulting of photon arrival times is fitted to a model, typically a for monoexponential decays: I(t) = I_0 \exp\left(-\frac{t}{\tau}\right) where I(t) is the fluorescence intensity at time t after excitation, I_0 is the initial intensity, and \tau is the lifetime. This method achieves high temporal resolution, often down to picoseconds, by iterative reconvolution to account for the instrument response function. Phase modulation, a frequency-domain , uses sinusoidally modulated light at frequencies typically from 1 to 200 MHz. The emitted experiences a shift (φ) and compared to the , from which the lifetime is calculated as \tau = \frac{1}{2\pi f \tan \phi}, where f is the modulation frequency. This approach is simpler in hardware but offers lower statistical precision than TCSPC for complex decays. Both techniques require careful control of photon flux to avoid artifacts like pile-up in TCSPC or harmonic distortions in methods. A key application of time-resolved fluorometry lies in distinguishing bound from free fluorophores through lifetime shifts induced by molecular interactions. For instance, free ethidium bromide exhibits a lifetime of approximately 1.7 ns, which lengthens to 23 ns when bound to DNA due to reduced quenching from intercalation into the double helix. Similarly, propidium iodide's lifetime increases from 1.3 ns (free) to 12.0 ns (bound), enabling selective detection in cellular assays without spectral overlap issues common in intensity-based methods. These shifts facilitate applications in biosensing, such as monitoring protein-DNA interactions or FRET-based proximity assays, where donor lifetime reduction signals energy transfer to an acceptor. Advancements since the 2000s have integrated time-resolved capabilities with and high-throughput formats. (FLIM), combining TCSPC or time-gating with confocal or wide-field setups, emerged prominently around 2005 with innovations like scanning for depth-resolved , allowing multi-point lifetime mapping in seconds. By the , FLIM extended to light-sheet for volumetric of large samples in minutes, enhancing studies of heterogeneity. Concurrently, automated FLIM plate readers, such as those developed in 2013, enabled unsupervised screening of 96-well plates with optical sectioning, supporting and high-content analysis. Recent progress includes video-rate FLIM at 20 frames per second (2019) and compressed ultrafast photography variants achieving 100 Hz acquisition (2021), driven by faster detectors and computational fitting. Despite these developments, time-resolved fluorometers face limitations in cost and complexity relative to steady-state instruments. High-precision systems require expensive components like fs lasers (often >$50,000) and specialized timing electronics, with full TCSPC setups exceeding $100,000, compared to sub-$10,000 for basic steady-state models. Operational complexity arises from the need for low count rates (e.g., <5% probability per pulse) to prevent distortions, prolonged acquisition times (minutes to hours for high signal-to-noise), and advanced via field-programmable arrays (FPGAs) or software for multi-exponential fitting. These factors limit accessibility for routine use, though low-cost alternatives using modulated LEDs have emerged to mitigate barriers.

Calibration and Measurement Techniques

Calibration Methods

Calibration of fluorometers is essential to ensure accurate wavelength selection, intensity measurement, and reproducibility across instruments and over time. This involves using certified standards and reference materials to align and wavelengths, quantify intensity, and account for instrumental artifacts such as inner filter effects. Procedures typically include both initial setup calibration and routine maintenance checks to maintain to primary standards. Standard fluorescent dyes serve as primary references for intensity and quantum yield calibration. Quinine sulfate dihydrate, often prepared in 0.1 M at concentrations of 0.1–10 μM, is widely used due to its high (approximately 0.54) and stability, enabling direct comparison for intensity and . 101, in solutions of similar concentration ranges, acts as a quantum counter standard for correcting response across 250–600 nm, ensuring uniform detection efficiency. These standards are measured under optically dilute conditions ( < 0.05) to minimize and self-quenching. Wavelength calibration aligns the and monochromators using reference filters or solutions. oxide (Ho₂O₃) glass filters, with certified absorption peaks at wavelengths such as 241.5 , 287.1 , and 536.4 (up to ±0.4 uncertainty), are placed in the emission path to verify spectral alignment, particularly for instruments operating in the UV-visible range. This method uses a diffuse reflector to simulate sample , allowing ratio measurements of with and without the filter for precise peak positioning. Intensity calibration corrects for instrument-specific response variations and sample geometry effects. Integrating spheres positioned at the sample compartment collect total emitted , independent of angular distribution, to determine absolute quantum yields without refractive index corrections; the formula for relative quantum yield is \Phi_\text{sample} = \Phi_\text{std} \times \frac{I_\text{sample}}{I_\text{std}} \times \frac{A_\text{std}}{A_\text{sample}} \times \frac{n_\text{sample}^2}{n_\text{std}^2}, where I is integrated , A is at , and n is . Inner filter effects, arising from sample reabsorption, are mitigated by simultaneous measurements at and s, applying correction factors derived from Beer-Lambert law deviations. Routine daily and periodic checks maintain calibration integrity. Dark current subtraction involves measuring baseline signal with excitation source blocked, subtracting it from sample readings to eliminate detector noise. Stray light tests assess off-wavelength light leakage by scanning with a narrow-band filter or blocked ports, ensuring rejection ratios below 0.1% for accurate spectral selectivity. Modern fluorometers incorporate software-automated routines for these verifications and quantum yield calculations, often integrating spectral correction files traceable to NIST standards like SRM 936a (quinine sulfate).

Common Measurement Modes

Fluorometers operate in several standard measurement modes to characterize fluorescent samples, each tailored to specific experimental needs such as identifying fluorophores or monitoring dynamic processes. These modes rely on precise control of and wavelengths, often requiring prior to ensure accurate intensity readings. In spectra mode, a fixed wavelength is selected, typically based on the sample's maximum, while the scans across a range of wavelengths, such as 300-700 , to record the intensity as a function of . This produces a that reveals the fluorophore's profile, aiding in its identification and distinguishing it from background signals or multiple fluorophores in a mixture. The resulting plot mirrors the vibrational structure of the , with peak often occurring at longer wavelengths than due to . Excitation spectra mode reverses this approach by fixing the at the fluorophore's maximum and scanning the across a relevant range to measure the intensity of emitted light. This mode identifies the optimal for maximum yield and confirms the sample's characteristics, as the spectrum closely resembles the spectrum when is independent of . It is particularly useful for optimizing experimental conditions in complex samples where multiple sources may be available. Anisotropy or polarization mode employs polarized to probe the of fluorophores, quantifying molecular size and mobility through the degree of in emitted . Vertically polarized excites the sample, and emission intensities are measured parallel (I∥) and perpendicular (I⊥) to the excitation axis; (r) is calculated as r = (I∥ - I⊥) / (I∥ + 2I⊥), where higher values indicate restricted due to larger complexes or viscous environments. This mode is valuable for studying protein-protein interactions or conformational changes, as rotational correlation times in the range reflect biomolecular dynamics. Förster resonance energy transfer (FRET) setups in fluorometers monitor proximity between donor and acceptor fluorophores through dual-wavelength detection, typically exciting the donor and measuring both donor and acceptor s simultaneously. Energy transfer occurs non-radiatively when molecules are within 1-10 , reducing donor and enhancing acceptor ; the is assessed by the of acceptor-to-donor signals after correcting for spectral overlap and bleed-through. This mode enables real-time assays for molecular interactions, such as protein binding or conformational shifts in biosensors. Kinetic modes track intensity over time to capture dynamic events like or binding , with down to approximately 1 ms in modern instruments. In assays, a substrate's releases a fluorescent product, and continuous monitoring of emission changes yields time-course data for parameters like reaction velocity (Vmax) and Michaelis constant (). For instance, assays for hydrolases use fluorogenic probes excited at wavelengths like 355 nm and monitored at 460 nm to quantify activity in , providing insights into inhibition or activation profiles.

Applications

Biomedical and Molecular Biology

In biomedical and research, fluorometers play a crucial role in protein quantification by enabling highly sensitive detection through fluorescent dyes that bind specifically to proteins. Assays such as the NanoOrange Protein Quantitation Kit utilize a merocyanine that exhibits enhanced upon binding to detergent-solubilized proteins, allowing detection in the range of 10 ng/mL to 10 μg/mL, which is particularly useful for analyzing low-abundance samples in cellular extracts. Similarly, the Protein Assay employs a fluorescent compatible with microvolume fluorometers, offering precise quantification down to 0.25 μg per with minimal sample volume, outperforming traditional colorimetric methods like or in sensitivity for dilute protein solutions. These fluorometric approaches provide advantages in and reduced interference from common contaminants, facilitating accurate protein determination in downstream applications like Western blotting or studies. For nucleic acid analysis, fluorometers are essential in DNA and RNA quantification, particularly in polymerase chain reaction (PCR) and quantitative PCR (qPCR) workflows. The PicoGreen dsDNA reagent, a high-affinity fluorescent dye, selectively binds double-stranded DNA (dsDNA) and enables detection as low as 25 pg/mL, allowing precise measurement of DNA concentrations in the presence of single-stranded DNA or RNA without prior purification. This sensitivity is critical for applications such as library preparation in next-generation sequencing or monitoring PCR amplicons. In qPCR setups, SYBR Green I dye intercalates into dsDNA amplicons during amplification, with fluorometers integrated into real-time instruments quantifying fluorescence increases cycle-by-cycle to determine initial nucleic acid amounts with high specificity and linearity over six orders of magnitude. These methods support gene expression analysis and pathogen detection by providing robust, reproducible quantification essential for molecular diagnostics. Fluorometers are integrated into cell imaging and for assessing cell viability and , leveraging fluorogenic probes that distinguish live from . The Viability/Cytotoxicity Kit uses calcein-AM for live cells (green upon esterase cleavage) and ethidium homodimer-1 for (red upon membrane compromise), enabling fluorometric or flow cytometric readout of viability ratios in populations, often achieving over 95% accuracy in distinguishing healthy cultures. For detection, caspase-specific fluorophores like DEVD-R110 substrates are cleaved by activated caspases-3/7, releasing rhodamine 110 for green measurable by plate-reading fluorometers, allowing high-throughput monitoring of in drug-treated cells with sensitivity down to single-cell levels. These assays are vital for evaluating effects in and . In , fluorometers facilitate (HTS) by measuring binding affinities through techniques like fluorescence resonance energy transfer (FRET) and . FRET-based assays monitor proximity between fluorophore-labeled proteins and ligands, with energy transfer donor emission upon binding, quantifiable via ratio-metric fluorometry to identify inhibitors with sub-nanomolar affinities in 384-well formats. assays, conversely, detect rotational changes in polarized emission from bound versus free tracers, enabling robust HTS of small-molecule libraries for targets like kinases, with Z' factors exceeding 0.7 for reliable hit identification. These homogeneous methods accelerate lead optimization by providing real-time kinetic data without washing steps.

Environmental and Oceanography

In environmental monitoring, fluorometers play a crucial role in assessing phytoplankton biomass through the measurement of chlorophyll-a fluorescence in natural water systems. In situ oceanographic fluorometers, such as those developed by Turner Designs, enable real-time quantification of chlorophyll-a concentrations, typically calibrated over ranges from 0.03 to 50 μg/L, to estimate primary productivity and algal blooms in coastal and open ocean environments. These devices are widely deployed in ocean observatories, where the 2-Wavelength Fluorometer (FLORD) specifically targets chlorophyll-a fluorescence alongside optical backscatter to profile phytoplankton distribution. Fluorometers also facilitate the tracking of dissolved (DOM) in aquatic ecosystems, providing insights into carbon cycling dynamics. Excitation-emission matrices (EEMs) generated by allow for the characterization of DOM components, including humic-like and protein-like substances, which are essential for understanding transport and microbial activity in rivers, estuaries, and oceans. In marine settings, such as the eastern , EEMs combined with parallel (PARAFAC) reveal spatial variations in DOM , linking it to sources like terrestrial runoff and production. For pollutant detection, fluorometers exploit the native of polycyclic aromatic hydrocarbons (PAHs) to monitor contamination in water bodies. Bench-scale fluorescence systems using EEMs can rapidly quantify multiple PAHs at parts-per-billion levels in surface and , aiding in the assessment of oil spills and industrial effluents without extensive sample preparation. This approach has been integrated into autonomous underwater vehicles (AUVs) equipped with submersible fluorometers for PAH detection during underwater oil-spill responses. Portable submersible fluorometers have expanded real-time since the 1990s, with models like the Sea-Tech fluorometer deployed on buoys and early AUVs for continuous . These rugged sensors, often rated for depths up to 2,000 meters, integrate with platforms such as floats and gliders to collect long-term data on parameters in remote marine areas. Steady-state fluorometers, valued for their portability, support such deployments by providing stable, low-power measurements. In climate research, fluorometers contribute to profiling (CDOM), a key indicator of biogeochemical responses to environmental stressors. CDOM measurements help trace carbon export and light attenuation in the , revealing impacts of on DOM cycling, where increased CO2 levels alter microbial processing and rates. For instance, integrated sensors on autonomous platforms like the Navis BGC simultaneously profile CDOM alongside to monitor acidification effects on coastal ecosystems.

Industrial and Food Science

In the industry, fluorometers play a crucial role in by enabling the detection of residues in , such as and oxytetracycline, using dual fluorescent aptasensors based on mesoporous silica nanoparticles that provide simultaneous and quantitative analysis with limits of detection of 2.2 ng/mL for and 1.23 ng/mL for oxytetracycline. These assays leverage enhancement upon binding to antibiotics, allowing rapid screening to ensure with regulatory limits and prevent risks from residues in animal-derived foods like . Additionally, fluorometers facilitate the of content through staining, a lipophilic fluorescent that specifically labels triglycerides in globules, enabling confocal microscopy-based quantification of distribution and globule size in products such as and cheese. This technique reveals structural changes in globules during processing, with intensity correlating directly to localization and aiding in optimizing product and . For concerning protein misfolding diseases, fluorometers employing Thioflavin T assays detect associated with transmissible spongiform encephalopathies (TSEs) in animal tissues, where the dye binds to fibrils and exhibits enhanced upon at 450 nm, enabling real-time monitoring of prion seeding activity in meat samples. This method, adapted in (RT-QuIC) protocols, achieves high sensitivity for prions in muscle tissue, supporting testing by identifying contamination at femtogram levels of prion protein. Such assays are vital for preventing TSE transmission through contaminated beef or , with readouts providing a non-invasive alternative to bioassays for routine industrial screening. In the analysis of oils and lubricants, fluorometers detect contamination by aromatic compounds, particularly polycyclic aromatic hydrocarbons (PAHs), through (LIF) techniques that excite samples at wavelengths to produce characteristic spectra for identifying petroleum-based pollutants in fluids. This approach quantifies with detection limits around 10 , distinguishing between lubricant degradation and external contaminants like crude oil spills in environments. By measuring intensity from aromatic fluorophores, fluorometers ensure machinery reliability and environmental compliance in sectors like automotive and maintenance. Agronomy benefits from fluorometers in monitoring plant via , where the Fv/Fm ratio—calculated as (Fm - F0)/Fm—serves as an indicator of efficiency, typically ranging from 0.78 to 0.84 in healthy but dropping below 0.7 under or . fluorometers measure this parameter non-destructively to assess health, correlating reductions in Fv/Fm with losses up to 20% in under severe deficit. This enables early intervention in field , such as adjustments, to mitigate impacts on photosynthetic performance. Inline process control in breweries utilizes fiber-optic probes integrated with fluorometers to evaluate viability during , employing fluorescent dyes like fluorescein diacetate that penetrate viable cells and produce green upon , allowing real-time monitoring of integrity without disrupting the process. These probes transmit and light via optical fibers inserted into fermenters, detecting viability rates above 90% in active populations and alerting to drops below 80% due to factors like accumulation. This enhances efficiency by optimizing pitching and timelines, reducing off-flavor risks in production.

Advantages and Limitations

Key Benefits

Fluorometers offer exceptional sensitivity, capable of detecting fluorophores at concentrations ranging from 10^{-9} to 10^{-12} M, which surpasses the typical limits of absorbance-based methods by 1,000 to 500,000 times due to the inherent signal amplification in fluorescence emission. This high sensitivity enables single-molecule detection in optimal conditions, making fluorometers invaluable for trace analyte analysis in fields like biomedical research. The specificity of fluorometers stems from their ability to use multi-wavelength excitation and emission selection, which minimizes interference from background substances in complex samples by targeting analyte-specific spectral signatures. This feature is particularly advantageous in biological matrices, where probes can be designed to bind selectively to targets, reducing false positives compared to broadband absorbance techniques. As a non-destructive , fluorometry allows real-time monitoring of samples without alteration, facilitating applications such as live-cell imaging to observe dynamic processes like protein interactions or cellular responses over extended periods. This preserves sample integrity, enabling continuous or repeated measurements in living systems. Fluorometers demonstrate versatility across formats, including adaptations for microplate readers for , capillary systems for micro-volume analysis as low as 2.9 μL, and fiber-optic probes for imaging in deep tissues or hollow organs. Such flexibility supports diverse experimental setups without compromising performance. In terms of cost-effectiveness, portable fluorometers have become accessible for routine and use since the , with models available under $5,000, including open-source multichannel designs costing around $1,400 for basic configurations. This affordability has broadened adoption in resource-limited settings while maintaining analytical reliability.

Potential Drawbacks

One significant limitation of fluorometers arises from sample , particularly autofluorescence in biological matrices such as tissues or cells, which can overlap with the emission spectra of target fluorophores and elevate , necessitating techniques to isolate specific signals. This is especially pronounced in complex samples like or cellular extracts, where endogenous fluorophores such as NADH or flavins contribute to non-specific emission, reducing the and complicating . Photobleaching represents another challenge, characterized by the irreversible degradation of fluorophores under prolonged , leading to a time-dependent decay in intensity that can skew measurements in kinetic or long-duration assays. This effect is exacerbated by high intensities and the presence of oxygen, resulting in the formation of non-fluorescent products and limiting the duration of reliable data collection. strategies include employing low-intensity sources to minimize rates while preserving sufficient signal strength. Instrument variability further impacts fluorometer performance, with photomultiplier tube (PMT) responses exhibiting to temperature fluctuations that alter gain and dark current, potentially introducing up to several percent variability in readings without proper controls. Maintaining stable environmental conditions, such as through temperature-controlled sample compartments or adjustments, is essential to ensure across experiments. Additionally, advanced models like spectrofluorometers often carry high initial costs, typically exceeding $50,000 for systems with broad spectral coverage and high , which can restrict accessibility for routine use. In the 2020s, critiques have highlighted fluorometers' limited penetration and accuracy in turbid samples, where and by attenuate and light, distorting concentration estimates in media like or soil extracts. This issue is particularly relevant in , as can reduce effective path lengths and inflate errors by 10-15% or more without correction. Recent advancements, such as ratiometric dyes that normalize signals against internal references, have addressed these constraints by compensating for optical heterogeneity in environments. effects, where collisional or static interactions diminish , can compound these challenges but are typically managed through protocols detailed elsewhere.

References

  1. [1]
    Benchtop and Modular Fluorometer Systems (Fluorimeters) - HORIBA
    A fluorometer (and fluorimeter) measures the fluorescence or light emitted by different fluorescing objects. Fluorescence occurs when light of specific ...
  2. [2]
    [PDF] An Introduction to Fluorescence Measurements - Turner Designs
    The instrument used to measure fluorescence is called a fluorometer or fluorimeter. A fluorometer generates the wavelength of light required to excite the ...
  3. [3]
    The History of Fluorescence Microscopy - News-Medical
    Nov 21, 2018 · Fluorescence was first discovered in 1845 by Fredrick W. Herschel. He discovered that UV light can excite a quinine solution (e.g. tonic water) ...
  4. [4]
    Fluorometer | National Museum of American History
    Farrand began manufacturing fluorometers in the 1950s. The inscription on this example reads “Fluorometer / MODEL A SER 202 / FARRAND OPTICAL / NEW YORK / PAT.
  5. [5]
    SPF 1951: Behold the First Spectrophotofluorometer!
    Jun 12, 2015 · Bowman's spectrophotofluorometer, or “SPF,” allowed scientists to use fluorescence as a way to identify and measure tiny amounts of substances in the body.
  6. [6]
    Field techniques for the determination of algal pigment fluorescence ...
    Dec 27, 2022 · The use of algal fluorometers by the U.S. Geological Survey (USGS) has become increasingly common. The basic principles of algal fluorescence, ...
  7. [7]
    [PDF] Fluorometric techniques for the measurement of oceanic chlorophyll ...
    This report gives a summary of the general concepts of fluorometry, examines important assumptions, presents an outline of an established fluorometer ...
  8. [8]
    The Working Principle of a Fluorometer - DeNovix
    Aug 2, 2023 · A fluorometer operates by quantifying biological analytes based on fluorescence. To begin the process, the sample of interest needs to be bound to a specific ...Missing: definition | Show results with:definition
  9. [9]
    [PDF] What is fluorescence?
    Jun 10, 2006 · First stab at a definition:​​ Fluorescence describes the emission of light from a. substance being irradiated by light of a different color. ...
  10. [10]
    Basic Concepts in Fluorescence - Molecular Expressions
    Nov 13, 2015 · Quantum yield (sometimes incorrectly termed quantum efficiency) is a gauge for measuring the efficiency of fluorescence emission relative to all ...
  11. [11]
    [PDF] Absorption, Emission and Fluorescence Spectroscopies
    ε = molar absorptivity (units of M-1cm-1). A = log(P0/P). Absorbance. Page 9. M + hν → M*. Absorption is the process that consumes a photon and puts the atom or ...
  12. [12]
    Introduction to Fluorescent Proteins - Zeiss Campus
    The principle fluorophore (most often termed a chromophore) in GFP is a tripeptide consisting of the residues serine, tyrosine, and glycine at positions 65-67 ...Missing: fluorescein dots
  13. [13]
    Semiconductor quantum dots as fluorescent probes for in vitro ... - NIH
    This paper is an overview of the basics of QDs and their role as fluorescent probes for various biological imaging applications.<|control11|><|separator|>
  14. [14]
    [PDF] Quenching
    Collisional quenching occurs when the excited fluorophore experiences contact with an atom or molecule that can facilitate non-radiative transitions to the ...
  15. [15]
  16. [16]
    Fluorescence Fundamentals | Thermo Fisher Scientific - US
    Fluorescent probes enable researchers to detect particular components of complex biomolecular assemblies, such as live cells, with exquisite sensitivity and ...
  17. [17]
    Principles and Theory of Fluorescence Spectroscopy - HORIBA
    Fluorescence spectroscopy analyzes fluorescence from a molecule based on its fluorescent properties. Fluorescence is a type of luminescence caused by photons ...
  18. [18]
    Design of a Portable and Reliable Fluorimeter with High Sensitivity ...
    As can be seen, wide linear ranges were obtained for all chemical compounds, ranging from nanomolar to micromolar concentration levels. Very low LOD (0.1–7.2 nM) ...
  19. [19]
    None
    ### Summary of Optical System in Fluorometers (from https://www.chem.uci.edu/~dmitryf/manuals/Fundamentals/Fluorescence%20Spectroscopy.pdf)
  20. [20]
    Review—Recent Progress in Portable Fluorescence Sensors
    Jan 7, 2021 · Portable fluorescence sensors have been developed for biochemical detection, water quality monitoring, biomedical sensing, and many other applications.
  21. [21]
    None
    ### Summary of Optical System in Fluorometers (http://web.mit.edu/solab/Documents/Assets/So-Fluorescence%20spectrophotometry.pdf)
  22. [22]
    Component Selection for Fiber-Optic Fluorometry - Sage Journals
    A fiber-optic fluorescence sensor testbed was constructed, and different optical designs and components were compared for sensitivity in detecting fluorescence.
  23. [23]
    [PDF] Recommendations and Guidelines for Standardization of ...
    4.5 Detector. For the detection of emission, a photomultiplier tube (PMT) or a CCD array is typically placed after the emission wavelength selector. The ...
  24. [24]
    [PDF] Lifetime Imaging Techniques for Optical Microscopy - UCI Chemistry
    These components often have ill-defined fluorescence spectra but are clearly distinguished by their fluorescence lifetime [70]. Fluorescence lifetime ...
  25. [25]
    Review of Fluorescence Spectroscopy in Environmental Quality ...
    Another type of detector used in fluorometers is the CCD with which the whole fluorescence spectrum can be analyzed at once without the need for wavelength ...
  26. [26]
    Ocular Spot Fluorometer Equipped With a Lock-In Amplifier for ...
    Ocular Spot Fluorometer Equipped With a Lock-In Amplifier for Measurement of Aqueous Flare ... This noise reduction ensures that the changes in ILS are ...
  27. [27]
    Canopy chlorophyll fluorescence applied to stress detection using ...
    May 25, 2019 · Nevertheless, the illuminated area was still of the same order than the original PAM fluorometer ... (ADC) resolution of 16 bits. One ...
  28. [28]
    Fluorescence Data Acquisition Software | DataStation from HORIBA ...
    DataStation is a measurement environment for time-domain fluorescence, integrating control of components and performing measurements like decay and anisotropy.
  29. [29]
    [PDF] SoftMax Pro Data Acquisition and Analysis Software User Guide
    You can set the integration delay before collecting data and the integration time of the fluorescence signal. The integration delay is the amount of time ...Missing: fluorometer | Show results with:fluorometer
  30. [30]
    Portuino—A Novel Portable Low-Cost Arduino-Based Photo - NIH
    A novel portable low-cost Arduino-controlled photo- and fluorimeter for on-site measurements has been developed. The device uses LEDs as a light source and ...Missing: management | Show results with:management
  31. [31]
    qByte: An open-source isothermal fluorimeter for democratizing ...
    External battery support, making the whole system portable. WiFi connectivity for remote control and data sharing. Sensitivity range able to measure DNA ...
  32. [32]
    [PDF] model 10-au-005-ce fluorometer user's manual - Turner Designs
    Jan 25, 1997 · When the sample compartment cover is removed for any reason, a safety interlock switch turns the lamp off. ... Your fluorometer will arrive in a ...
  33. [33]
    Fluorometer Control and Readout Using an Arduino Nano 33 BLE ...
    Oct 5, 2022 · Fluorometer Control and Readout Using an Arduino Nano 33 BLE ... fluorometer for vibration, acoustic noise, water leaks and overheating.
  34. [34]
    Fluorescence Measurements & Instrumentation | Spectrofluorometer
    Jul 31, 2018 · Steady state fluorescence measurements require a continuous excitation source. This is usually a continuous xenon lamp which generates intense ...
  35. [35]
    What are the different types of spectrometers used in fluorometry?
    Jan 13, 2023 · Filter fluorometers are typically used in experiments that consistently measure different substances since optical filters are cost-efficient ...
  36. [36]
    10.2: Fluorescence and Phosphorescence Instrumentation
    Aug 10, 2022 · A fluorometer is a filter based, fixed wavelength, instrument suitable for established quantitative fluorescence methods. A spectrofluorometer ...
  37. [37]
    What are the differences between fluorometers and ... - Kalstein
    Spectrofluorometers are equipment used to analyze the fluorescence of a sample. They employ diffraction grating monochromators to isolate incident light and ...
  38. [38]
    Mitochondrial function in vivo evaluated by NADH fluorescence
    Aug 30, 2006 · Since the early 1950s, differ- ent groups have developed a number of fluorometers adapted to their specific experimental protocols. As of today, ...
  39. [39]
    FS5 Spectrofluorometer | Absorption and Fluorescence
    The FS5 Spectrofluorometer can measure emission spectra up to 1650 nm when equipped with an NIR PMT or up to 2025 nm with an InGaAs detector. These upgrades ...
  40. [40]
  41. [41]
    Fluorescence Spectrophotometer F-7000 - Hitachi High-Tech
    Features · The highest sensitivity at S/N 800 (RMS) (S/N 250 Peak to Peak) in its class · The fastest ultra high-speed scan at 60,000 nm/min in its class ...
  42. [42]
    HITACHI F7000 Fluorescence Spectrophotometer - Best Price
    Powered by a 150W Xenon lamp and photomultiplier tube, the F7000 guarantees high sensitivity with a signal-to-noise ratio exceeding 300. This enables precise ...
  43. [43]
    Fluorescence Lifetime Techniques: TCSPC, FRET, TRES, and SSTD
    TCSPC is typically used to measure fluorescence lifetimes from the picosecond to microsecond time scale.Missing: fluorometer phase modulation
  44. [44]
    [PDF] Time-Correlated Single Photon Counting - PicoQuant
    Classical TCSPC for fluorescence lifetime mea- surements only uses the short term difference be- tween excitation and emission. It was soon realized that ...
  45. [45]
    [PDF] Overview of Fluorescence Lifetime Measurements in Flow Cytometry
    Average fluorescence lifetimes range from 1 to 30 ns, and individual fluorophores might experience a shift in their average lifetime by an addition or reduction ...
  46. [46]
    Fluorescence lifetime imaging microscopy: fundamentals and ... - NIH
    May 13, 2020 · Conventionally, we define fluorescence lifetime ( τ ) as the average time that a fluorophore remains in its excited state. In this interval, ...
  47. [47]
    Advancements in fluorescence lifetime imaging microscopy ... - NIH
    In the measurement of fluorescence lifetime, a time-domain FLIM imager can operate in either time-gated or time-correlated single photon counting (TCSPC) mode, ...Missing: fluorometer | Show results with:fluorometer
  48. [48]
    Automated fluorescence lifetime imaging plate reader and its ... - NIH
    We present here an automated multiwell plate reader able to perform rapid unsupervised optically sectioned FLIM of fixed and live biological samples.Missing: advancements fluorometers
  49. [49]
    Innovating lifetime microscopy: a compact and simple tool for life ...
    May 1, 2006 · This novel FLIM setup is used for cellular and FRET imaging, and for high-throughput and fast imaging applications. The all-solid-state design ...
  50. [50]
    Advantages and Limitations of Fluorescence Lifetime Measurements ...
    May 18, 2022 · The total measurement time cannot be shorter than the FL of the analyzed fluorophore. Another limiting factor for the total measurement time is ...
  51. [51]
    Development of Low-Cost Instrumentation for Single Point ...
    May 25, 2017 · Here we present the development of a low-cost time-resolved fluorometer implemented in a single point measurement instrument that utilises a ...
  52. [52]
    Hardware Selection and Performance of Low-Cost Fluorometers
    Mar 17, 2022 · Modern laboratory fluorometers range in cost from USD 6000 to USD 39,000 with detection limits ... To assess the ability of the low-cost sensors ...
  53. [53]
    [PDF] Standard Guide to Fluorescence Instrument Calibration and Correction
    If 90-degree detection is only possible, then a right-triangular cuvette can be used with the excitation beam at 45-degree incidence to the hypotenuse side ...
  54. [54]
    Fluorescein, Quinine Sulfate and Green Fluorescent Protein - NIH
    Jul 26, 2018 · Quinine sulfate has been used as a fluorescence standard for more than 50 years [11, 13] and the photophysical properties of various quinine ...
  55. [55]
    Inner Filter Effect Correction for Fluorescence Measurements in ...
    May 2, 2022 · The inner filter effect (IFE) hinders fluorescence measurements, limiting linear dependence of fluorescence signals to low sample ...Introduction · Results and Discussion · Conclusions · Supporting Information
  56. [56]
  57. [57]
    Measurements of Absolute Values in Biochemical Fluorescence ...
    The true, or absolute, fluorescence emission spectrum of a substance is a plot showing the relative number of photons emitted at different wavelengths; i.e., dq ...
  58. [58]
    What is Fluorescence Spectroscopy? - HORIBA
    A fluorescence excitation spectrum is when the emission wavelength is fixed and the excitation monochromator wavelength is scanned. In this way, the spectrum ...
  59. [59]
    Fluorescence Anisotropy and Resonance Energy Transfer
    The method of fluorescence anisotropy (FA) or polarization allows for determination of the rotational correlation time of the fluorescent probe and hence ...Missing: fluorometers | Show results with:fluorometers
  60. [60]
    [PDF] Fluorescence Anisotropy Theory Method and Data Analysis
    For this reason, polarisation anisotropy measurements are frequently performed using excitation and emission filters rather than monochromators. Optical filters ...
  61. [61]
    Methodology and Applications of Förster Resonance Energy ... - NIH
    Förster resonance energy transfer (FRET) is a process by which energy is transferred from one fluorophore (the donor) to a second fluorophore (the acceptor) in ...
  62. [62]
    Fluorescence-Based Enzyme Activity Assay - PubMed Central - NIH
    This review provides a comprehensive overview of the principles behind these assays, the various substrates and fluorophores used, and advances in assay ...Missing: modes | Show results with:modes
  63. [63]
    Real-time enzyme kinetics monitored by dual-color fluorescence ...
    A method for sensitively monitoring enzyme kinetics and activities by using dual-color fluorescence cross-correlation spectroscopy is described.Materials And Methods · Results And Discussion · Enzyme KineticsMissing: modes resolved fluorometer
  64. [64]
    a fluorescence-based assay of proteins in solution - PubMed
    The NanoOrange assay is a fluorescence assay using a dye that increases fluorescence with detergent-coated proteins, detecting 10 ng/mL to 10 micrograms/mL  ...
  65. [65]
    [PDF] Accurate and sensitive protein quantitation - Fisher Scientific
    The assay has an optimal range of 1.25–25 µg/mL (0.25–5 µg) in the assay tube (initial stock concentrations, 12.5–5 mg/mL) and is provided in a simple kit ...
  66. [66]
    Characterization of PicoGreen reagent and development ... - PubMed
    Jul 1, 1997 · The PicoGreen assay allowed the detection of 25 pg/ml dsDNA, surpassing the sensitivity achieved with Hoechst 33258 by 400-fold. The linear ...
  67. [67]
    SYBR® Green for qPCR - Bio-Rad
    SYBR® Green is a dsDNA-binding dye that intercalates nonspecifically into dsDNA, allowing measurement of the amount of PCR product. SYBR® Green fluorescence ...
  68. [68]
    LIVE/DEAD Cell Viability Assays | Thermo Fisher Scientific - US
    Thermo Fisher Scientific LIVE/DEAD cell viability assays provide for easy and sensitive differentiation of live, dead, and cytotoxic populations.
  69. [69]
    Caspase-3 DEVD-R110 Fluorometric HTS Assay Kit - Biotium
    A fluorometric endpoint assay kit with DEVD substrate for detection of caspase-3 activity in cells by high throughput screening (HTS).
  70. [70]
    Fluorescence polarization assays in high-throughput screening and ...
    Assays based on FP/FA technology have been widely utilized in high-throughput screening (HTS) and drug discovery due to the homogenous format, robust ...
  71. [71]
    Fluorescence anisotropy imaging in drug discovery - ScienceDirect
    FP/FA have also been used in assays to enable high-throughput screening of small molecule libraries for drug discoveries [17]. Optical imaging technologies ...
  72. [72]
    Fluorogenic CRISPR for genomic DNA imaging - Nature
    Jan 31, 2024 · With fCRISPR, we image various genomic DNA in different human cells with high signal-to-noise ratio and sensitivity.
  73. [73]
    Chlorophyll - Fluorometer Parameters - Turner Designs
    A procedure for measuring extracted chlorophyll a free from the errors associated with chlorophyll b and pheopigments.
  74. [74]
    Fluorometer - Ocean Observatories Initiative
    A Fluorometer is a device used to measure patterns of fluorescence. The 2-Wavelength Fluorometer (FLORD & FLNTU) specifically examines chlorophyll-a ...
  75. [75]
    Fluorescence Characteristics of Chromophoric Dissolved Organic ...
    Sep 30, 2021 · The objective of this study was to better characterize DOM in the eastern Indian Ocean. To better understand the underlying mechanisms, seawater samples were ...
  76. [76]
    Fluorescence excitation emission matrices for rapid detection of ...
    A bench-scale fluorescence system with on-line monitoring capabilities was used to quantify several polycyclic aromatic hydrocarbons (PAHs) and pesticides in ...Missing: native fluorometer<|separator|>
  77. [77]
    AUV based mobile fluorometers: System for underwater oil-spill ...
    Along these lines, this paper analyzes application of Autonomous Underwater Vehicles (AUV) with integrated submersible fluorometer for underwater detection of ...
  78. [78]
    [PDF] Oceanography
    Chlorophyll fluorometers. a Sea tech fluorometer, circa 1990, (background) and a pair of Wet labs eCO-puck fluorometer/backscatter sensors (foreground). The ...
  79. [79]
    Cyclops-7F Submersible Fluorescence/Turbidity Sensors
    Turner Designs recommends a minimum flow rate of 50 mL per minute. When you are measuring in vivo chlorophyll it is recommended that the flow rate not exceed ...
  80. [80]
    Impacts of Global Change on Ocean Dissolved Organic Carbon ...
    We review the research showing impacts of global change stressors on the DOC cycling, specifically: ocean warming and stratification, acidification, ...Missing: fluorometers | Show results with:fluorometers
  81. [81]
    Biogeochemical Monitoring of the Oceans Using Autonomous ...
    Mar 19, 2013 · The ability of the Navis BGC float to provide integrated CTD, oxygen, fluorescence, backscatter, and CDOM measurements as well as external ...
  82. [82]
    Dual fluorescent aptasensor for simultanous and quantitative ... - NIH
    Dec 21, 2022 · We developed a dual fluorescent aptasensor based on mesoporous silica to simultaneously detect sulfadimethoxine (SDM) and oxytetracycline (OTC) in animal- ...Missing: industry | Show results with:industry
  83. [83]
    Europium Fluorescent Nanoparticles-Based Multiplex Lateral Flow ...
    Dec 20, 2021 · A fluorescent immunoassay based on europium nanoparticles (EuNPs-FIA) was developed for the simultaneous detection of antibiotic residues.
  84. [84]
    Using Confocal Laser Scanning Microscopy to Probe the Milk Fat ...
    Nile Red is an oxazone dye used to stain lipid droplets (39). It is specific to triglycerides. The double-staining of the reconstituted buttermilk powder ...
  85. [85]
    Mechanisms for the interaction of the milk fat globule membrane with ...
    The Nile Red fluorescent probe (5H-Benzo α-phenoxazine-5-one, 9-diethylamino) was used to label the triacylglycerol (TAG) core of MFG as described by Lopez and ...
  86. [86]
    A field-deployable diagnostic assay for the visual detection of ...
    Jul 18, 2022 · Our newly developed assay, MN-QuIC, enables both naked-eye and light-absorbance measurements for detection of misfolded prions.
  87. [87]
    Comparison of conventional, amplification and bio-assay detection ...
    ... thioflavin T(ThT) into growing amyloid fibrils. RT-QuIC has been shown to detect amyloid seeding activity in a variety of biological tissue and fluid ...Missing: safety | Show results with:safety
  88. [88]
    [PDF] Chronic Wasting Disease- where it is
    Moreover, MN-QuIC may have potential food-safety test applica- tions given the recent documentation of RT-QuIC-based detection of CWD prions in WTD muscles used ...
  89. [89]
    Site Characterization and Analysis Penetrometer System (SCAPS ...
    The LIF method provides data on the in situ distribution of petroleum hydrocarbons based on the fluorescence response induced in the polycyclic aromatic ...
  90. [90]
    Fluorometric Detection of Oil Traces in a Sea Water Column - MDPI
    Mar 5, 2022 · This study focuses on broadening the knowledge of a fluorometric index to improve the detection of oil substances present in the marine environment.
  91. [91]
    On-Site Analytical Tool Based on Crude Oil Fluorescence and ...
    Jan 31, 2024 · Many reported studies showed the possibility to detect and identify crude oils by recording the direct fluorescence emission of oils upon ...
  92. [92]
    The Chlorophyll Fluorescence Parameter Fv/Fm Correlates with ...
    Jan 18, 2023 · F v /F m can be used as a proxy for estimating the yield performance of wheat after severe drought at anthesis.
  93. [93]
    Chlorophyll Fluorescence Imaging for Early Detection of Drought ...
    Mar 21, 2023 · This study developed a CFI system to measure the maximum quantum efficiency of photochemistry (Fv/Fm).
  94. [94]
    [PDF] EVALUATION OF CHLOROPHYLL FLUORESCENCE ... - OAKTrust
    stress but the Fv/Fm ratio did not change. The authors found that when plants were exposed to high irradiance, in addition to salt stress, both control and ...
  95. [95]
    Sensors and Techniques for On-Line Determination of Cell Viability ...
    Dec 3, 2022 · Fiber-optic waveguides are commonly used to transmit light [28]. Cell culture analysis can also be performed using an optical window or a flow ...
  96. [96]
    Fiber-Optic Chemical Sensors and Fiber-Optic Bio-Sensors - PMC
    Abstract. This review summarizes principles and current stage of development of fiber-optic chemical sensors (FOCS) and biosensors (FOBS).
  97. [97]
    [PDF] Novel Rapid Molecular Detection and Processing Approaches for ...
    Beer: pitching yeast wash. LAB (Lactobacillus and. Pediocococcus). 25.0-37.5. Beer: post fermentation ... A novel FRET-based optical fiber biosensor for rapid ...
  98. [98]
    [PDF] RF-20A Fluorescence Detector Basics and Applications - Shimadzu
    Fluorescence detection has the following features: 1) Selectivity is high because measurement is conducted using specific excitation and emission wavelengths.
  99. [99]
    SELECTIVITY AND SPECIFICITY OF SMALL MOLECULE ... - NIH
    Fluorescent dyes are widely used in the detection of labile (free or exchangeable) Zn2+ and Ca2+ in living cells. However, their specificity over other ...
  100. [100]
    Live-Cell Imaging | Thermo Fisher Scientific - US
    Fluorescence imaging of live cells is used to observe dynamic cellular processes and track cellular biomolecules and structures over time.Example: mitochondrial activity... · Tips to detect changes in live...Missing: destructive | Show results with:destructive
  101. [101]
    Glass capillary systems for micro-volume fluorometry - ScienceDirect
    Jan 30, 2025 · Capillary-based fluorometry can be considered a promising alternative to cuvette-based methods, enabling low-volume measurements while reducing analyte ...
  102. [102]
    Fiber-optic fluorescence imaging - PMC - PubMed Central - NIH
    Fiber-optic fluorescence imaging systems include portable handheld microscopes, flexible endoscopes well suited for imaging within hollow tissue cavities.Missing: capillaries | Show results with:capillaries
  103. [103]
    A Low-Cost, Open-Source, Multichannel In Situ Fluorometer Applied ...
    Apr 14, 2025 · We introduce the PIXIE, an open-source, multichannel, in situ fluorometer that performs high-quality fluorometry at a low cost.<|control11|><|separator|>
  104. [104]
    Hardware Selection and Performance of Low-Cost Fluorometers - NIH
    Mar 17, 2022 · While the optical filter did improve performance, components A–C, E, F, and H achieved a higher r2 for less than 10% of the cost of the optical ...
  105. [105]
    [PDF] General and technical considerations for background subtraction in ...
    Fluorescence background includes autofluorescence from tissue molecules, light excitation, and probe clearance. Background subtraction is needed to avoid ...
  106. [106]
    Excitation Light Dose Engineering to Reduce Photo-bleaching and ...
    Aug 3, 2016 · Assays are presented to measure photo-bleaching of fluorophores and photo-toxicity to living cells under different illumination conditions.
  107. [107]
    How to Minimize Photobleaching - Enzo Biochem
    Feb 6, 2023 · To minimize photobleaching, reduce light intensity and time exposure, reduce oxygen, use antioxidants, and use more frequent pulses of lower- ...
  108. [108]
  109. [109]
    Buy Spectrophotometer, UV-Vis, Fluor For Sale, New & Used Prices
    Prices for new systems generally range from $3,000 to $60,000, depending on the wavelength range, sensitivity, and additional features. High-end ...
  110. [110]
    Quantitative Fluorescence Spectroscopy in Turbid Media: A Practical ...
    A quantitative fluorescence model (QFM) was proposed to explicitly model the effects of the scattering and absorption on fluorescence measurements.
  111. [111]
    Recent advances in ratiometric fluorescence imaging of enzyme ...
    This review highlights recent advances using fluorescent probes that are processed by enzymes to induce photophysical changes that can be monitored by ...