Sample preparation

Sample preparation is the essential preliminary process in and related fields, involving the modification, extraction, concentration, and purification of raw samples—whether solid, liquid, or gaseous—to render them suitable for by removing interferences, ensuring representativeness, and optimizing detectability. This step transforms a sample into an analytical one, addressing challenges such as matrix complexity and low concentrations to enable precise quantification and identification. The importance of sample preparation cannot be overstated, as it often accounts for approximately 80% of the total time in analytical workflows and serves as a of errors if not executed properly. By isolating target analytes from interferents that could mask signals, cause instrument damage, or alter measurement accuracy, it enhances the reliability of techniques like , , and across applications in , , pharmaceuticals, and biomedical research. Key steps typically include sampling for representativeness, homogenization or size reduction, cleanup to eliminate effects, extraction for isolation, concentration or dilution to adjust levels, derivatization to improve compatibility with detection methods, and final transfer to analytical vessels. Common techniques encompass both conventional and advanced approaches, such as liquid-liquid extraction (LLE) for partitioning analytes between immiscible solvents, solid-phase extraction (SPE) utilizing sorbents like silica or polymers for selective adsorption, and miniaturized variants including and stir-bar sorptive extraction (SBSE) to reduce solvent use and enhance efficiency. Emerging methods, such as microwave-assisted extraction (MAE), pressurized liquid extraction (PLE), and the integration of novel sorbents like metal-organic frameworks (MOFs), address challenges like time consumption and environmental impact by improving recovery rates (often 70-120%), lowering detection limits (e.g., down to 0.03 μg/L), and promoting principles through reduced reagent volumes. Despite these advances, persistent hurdles include sorbent instability in aqueous matrices, in extraction systems, and the need for method validation to ensure across diverse sample types.

Introduction

Definition and Purpose

Sample preparation in refers to the process of transforming raw samples—whether solid, liquid, gas, or biological—into a homogeneous, contaminant-free form that is compatible with analytical techniques such as or . This involves physical and chemical manipulations to create representative subsamples suitable for accurate measurement, while minimizing losses or alterations during handling. The goal is to produce a sample that preserves the original composition and enables reliable downstream analysis. The primary purpose of sample preparation is to ensure the representativeness of the , enhance detection , minimize matrix effects that could interfere with , and improve the overall of analytical results. Core objectives include the isolation of target analytes from complex , concentration of trace-level components when necessary, and removal of potential interferences, all without significantly altering the of the analytes. These steps address challenges posed by heterogeneous or contaminated samples, ultimately supporting precise quantification in the broader analytical workflow. For instance, in environmental analysis, samples are prepared to detect pollutants by homogenizing and sieving to achieve uniformity and eliminate . Similarly, biological samples like are processed for drug screening to separate analytes from proteins and salts, thereby reducing effects and enabling sensitive detection.

Historical Development

The history of sample preparation in began in the with rudimentary manual techniques tailored for basic qualitative and quantitative analyses. Chemists relied on mechanical grinding of solid samples to achieve homogeneity and facilitate , often followed by simple solvent extractions to isolate analytes from complex matrices. A pivotal advancement came in 1879 with the invention of the by Franz von Soxhlet, which enabled efficient, continuous solvent extraction of and other compounds from solids, marking a shift toward more standardized procedures for organic analysis. In the mid-20th century, sample preparation evolved toward more systematic separation methods. Liquid-liquid extraction (LLE) gained prominence in the 1940s through the development of countercurrent distribution apparatus by Lyman C. Craig, which improved the partitioning of compounds between immiscible solvents for enhanced purity in biochemical separations. Precursors to (SPE) emerged in the 1960s with the use of filters and silica-based columns for adsorbing organics from aqueous samples, laying the groundwork for sorbent-based cleanup. A key milestone occurred in 1978 with the commercialization of SPE cartridges by , utilizing in disposable formats to streamline extraction and reduce solvent volumes compared to traditional LLE. The late saw accelerated innovation driven by instrumentation. Microwave-assisted digestion was introduced in 1985 by CEM Corporation in collaboration with the National Bureau of Standards, accelerating acid decomposition of intractable samples like soils and tissues for analysis while minimizing contamination. The 1990s marked the rise of , with robotic systems and automated pipetting stations enabling high-throughput preparation, such as in 96-well formats for pharmaceutical screening. These developments coincided with the broader adoption of hyphenated techniques, including the integration of SPE and LLE with liquid chromatography-mass spectrometry (LC-MS) in the early 2000s, which allowed online sample cleanup and direct injection for sensitive . Entering the , sample preparation shifted toward microscale and eco-friendly approaches to address environmental concerns and reduce reagent use. Techniques like single-drop microextraction (SDME), first described in 1996 but refined post-2000, utilized microliter volumes of solvent suspended from a for rapid extractions, promoting green principles. This era emphasized and , with methods such as dispersive liquid-liquid microextraction gaining traction for their low waste generation in environmental and food analyses.

Importance and Principles

Role in Analytical Workflow

Sample preparation occupies a central position in the analytical workflow as the initial major step following sample collection, where raw materials are processed into a form compatible with subsequent instrumental analysis. This phase bridges the gap between sample receipt and advanced techniques, such as chromatographic separation and mass spectrometric detection, ensuring that analytes are isolated, concentrated, or purified to minimize matrix interferences before injection or introduction into the instrument. In a typical workflow, the process begins with sample receipt and initial handling, progresses through extraction or digestion, incorporates cleanup steps like filtration, and culminates in the preparation of a final extract ready for analysis, effectively serving as the foundational bridge to reliable detection. The integration of sample preparation into the broader analytical process requires alignment with method validation protocols, where preparation efficiency directly influences key performance metrics such as the limit of detection (LOD) and limit of quantification (LOQ). Inefficient preparation can introduce or loss, thereby elevating the LOD—the lowest concentration at which an can be reliably distinguished from the blank—and compromising the LOQ, the threshold for accurate quantification with acceptable precision. Validation ensures that preparation steps maintain method robustness across diverse sample matrices, preventing variability that could skew overall analytical outcomes. Despite its critical role, sample preparation accounts for 60-80% of the total analysis time and a substantial portion of errors in analytical procedures, often leading to false positives or negatives if not executed properly. These errors typically arise from , incomplete , or matrix effects during preparation, which propagate through the and undermine result reliability. By optimizing this stage, laboratories can enhance overall efficiency and accuracy, reducing the risk of misleading interpretations in fields like and pharmaceutical testing.

Key Principles

Sample preparation in relies on the principle of representativeness, which ensures that the prepared sample accurately reflects the composition of the original material. This is achieved through homogenization techniques, such as grinding or milling, to create a of analytes, thereby minimizing sampling errors associated with heterogeneous materials. from homogenized batches further maintains this representativeness, with smaller particle sizes requiring less material to achieve low uncertainty levels, such as 0.1 g for 15% uncertainty at 0.5 mm particle size compared to 56 g at 5 mm. Minimizing contamination is essential to prevent the introduction of exogenous analytes that could compromise analytical accuracy. This involves using clean laboratory tools, such as acid-soaked labware, and running blanks to detect and quantify potential contaminants. Inert materials like Teflon or are selected to avoid chemical reactions or , ensuring the sample remains free from artifacts during handling and processing. Preservation of integrity focuses on maintaining the chemical and physical stability of the sample throughout preparation. Techniques include controlling , often below 1 with strong acids to reduce adsorption losses, and managing temperature to avoid degradation of volatiles, such as refrigerating or freezing biological samples. Antioxidants or buffers may be added to protect labile compounds, ensuring recovery without alteration. Efficiency and scalability in sample preparation balance high analyte recovery rates, typically targeting over 90%, with increased throughput for practical applications. Methods like automated or microwave-assisted processes achieve recoveries of 76-117% with relative standard deviations under 15%, enabling handling of multiple samples simultaneously for routine analyses while allowing customization for research needs. This approach optimizes resource use without sacrificing reliability. Compliance with standards such as ISO 17025 ensures , validation, and in sample preparation protocols. Adherence involves detailed standard operating procedures, use of internal standards for tracking losses, and method validation through and accuracy testing, aligning with guidelines from organizations like the EPA and AOAC. This framework supports across analytical workflows.

Sample Types and Initial Handling

Classification of Samples

Samples in are primarily classified by their physical state, which determines the initial approach to preparation. Solids, such as soils or biological tissues, often require mechanical disruption like grinding to increase surface area for . Liquids, including or , typically involve simpler manipulations such as dilution or to achieve homogeneity. Gases, like ambient air, necessitate trapping or adsorption techniques to concentrate analytes for . Hybrid forms, such as aerosols, combine particulate and gaseous phases, demanding specialized collection methods like impaction or to separate components. Matrix complexity further refines , distinguishing matrices from ones. matrices, such as pure solvents, contain minimal interferents and require little preprocessing beyond dilution. In contrast, matrices, like biological fluids rich in proteins or environmental samples laden with , introduce interferences that necessitate selective strategies to isolate target analytes. Samples are also categorized as biological or non-biological based on origin and composition. Biological samples, including tissues or fluids, often feature macromolecules that may require enzymatic to disrupt cellular structures and release analytes. Non-biological samples, such as minerals or soils, typically involve or to break down inorganic lattices. Classification is influenced by analyte properties, including volatility, solubility, and concentration levels. Volatile analytes in gaseous or headspace samples risk loss during handling, favoring cryogenic or sorbent-based stabilization. Solubility governs phase partitioning in extractions, with polar analytes suiting aqueous media and non-polar ones organic solvents. Concentration—ranging from trace levels (e.g., ng/L) to bulk—dictates whether preconcentration or dilution is needed to match instrumental detection limits. Initial preservation, such as cooling or acidification, may be referenced briefly to maintain integrity post-classification.

Initial Steps

Following sample collection, the initial steps in sample preparation are critical to preserve the sample's and prevent or , ensuring reliable downstream . Collection considerations begin with the use of sterile tools, such as autoclaved or pre-sterilized , syringes, and containers, to minimize the introduction of exogenous microbes or that could alter the sample's . Immediate labeling of samples with details like collection date, time, location, and collector's initials is essential to maintain chain-of-custody and avoid mix-ups during transport or storage. Additionally, incorporating field blanks—clean containers exposed to the sampling but not filled with the target material—helps detect potential from equipment or ambient conditions, allowing for baseline subtraction in analytical results. Storage protocols must be implemented promptly to halt biological, chemical, or physical changes. For liquid samples, at is standard to slow microbial growth and enzymatic activity while preserving volatile compounds, with many protocols recommending within 7-14 days to avoid significant loss. Biological samples, such as tissues or cells, are typically frozen at -20°C to inhibit , though stability can vary; for instance, unfixed tissues should be fixed or frozen as promptly as possible, ideally within 1 hour of collection, to prevent autolysis. Volatile samples may require headspace sampling or storage in sealed vials under inert atmospheres to retain gaseous components, with preferred over freezing to avoid issues. Preservation methods often involve targeted additives to stabilize specific analytes. In water samples for metal analysis, acidification with (typically to <2) solubilizes metals and prevents adsorption to container walls or precipitation, ensuring accurate trace element quantification. For biological fluids containing trace elements, chelating agents like EDTA are added to bind metals and inhibit clotting or oxidation, as seen in blood collection tubes designed for such analyses. These preservatives must be applied immediately post-collection in compatible containers to maximize efficacy without introducing interferences. Basic homogenization follows to ensure representativeness before subsampling. Initial mixing, such as gentle vortexing for liquids or manual stirring for solids, distributes heterogeneous components evenly, while portioning the homogenized sample into smaller aliquots (e.g., 1-10 mL or small fractions of the total volume) allows for multiple analyses without repeatedly disturbing the original sample. This step is particularly vital for non-uniform samples like soils or foods, where inadequate mixing can lead to biased results; for example, creating representative aliquots requires thorough blending to average out particle size variations.

Preparation Techniques

Extraction Techniques

Extraction techniques are essential in sample preparation for isolating target analytes from complex matrices, enabling subsequent analysis by minimizing interferences and concentrating compounds of interest. These methods rely on principles such as partitioning or adsorption to selectively separate analytes based on their physicochemical properties. Common approaches include solvent-based partitioning and sorbent-based retention, which are chosen depending on the sample type and analytical goals. Liquid-liquid extraction (LLE) operates on the principle of partitioning analytes between two immiscible liquid phases, typically an aqueous sample phase and an organic solvent phase, driven by differences in solubility. The process begins with the addition of an immiscible organic solvent, such as or , to the sample, followed by vigorous shaking to facilitate mass transfer and equilibration of the analyte between phases. Phase separation is then achieved through settling or centrifugation, yielding an organic extract containing the concentrated analyte. Recovery in LLE is quantified using the formula: \% \text{Recovery} = \frac{(C_{\text{aq}} \cdot V_{\text{aq}} + C_{\text{org}} \cdot V_{\text{org}})}{C_{\text{initial}} \cdot V_{\text{initial}}} \times 100 where C_{\text{aq}} and C_{\text{org}} are the analyte concentrations in the aqueous and organic phases, respectively, and V denotes volumes. This technique is particularly effective for moderately polar to non-polar analytes in liquid samples like biological fluids or environmental waters. Solid-phase extraction (SPE) involves the retention of analytes on a solid sorbent material, offering a versatile alternative to LLE with reduced solvent consumption and improved selectivity. The process comprises four main steps: activation of the sorbent with a conditioning solvent to expose functional groups; loading the sample onto the sorbent, where analytes adsorb via interactions like hydrophobic bonding or ion exchange; washing with a solvent to remove matrix interferences while retaining the analytes; and elution with a strong solvent, such as methanol, to release the analytes for analysis. Reversed-phase sorbents, such as C18-modified silica, are widely used for non-polar analytes due to their hydrophobic surfaces that retain compounds through van der Waals forces. SPE excels at trace-level detection by providing preconcentration factors that lower limits of detection to ng/L ranges, making it ideal for complex matrices like food or wastewater where matrix effects must be minimized. Supercritical fluid extraction (SFE) utilizes supercritical carbon dioxide (SC-CO₂) as a solvent, which exhibits gas-like diffusivity and liquid-like solvating power, particularly for non-polar analytes. The technique involves passing SC-CO₂, often with a polar cosolvent like methanol to enhance solubility, through the sample matrix under elevated pressure (e.g., 15 MPa) and temperature (e.g., 318 K), followed by depressurization to collect the extract. In pesticide residue analysis, SFE with SC-CO₂ has achieved recoveries above 70% for a range of organochlorine and organophosphate pesticides from soil samples, demonstrating its efficacy for environmental monitoring without excessive organic solvents. Selection of an extraction technique depends primarily on analyte polarity and matrix complexity to optimize recovery and minimize interferences. For non-polar analytes in aqueous matrices, non-polar solvents in LLE or reversed-phase SPE sorbents like C18 are preferred to match partitioning coefficients; polar analytes benefit from polar solvents or normal-phase sorbents. Complex matrices, such as those with high organic content or emulsions, favor SPE or SFE for their ability to handle interferences through selective retention or high diffusivity, respectively.

Dissolution and Digestion Techniques

Dissolution and digestion techniques are essential for converting solid or complex sample matrices into homogeneous liquid forms suitable for elemental analysis, particularly when total analyte recovery is required. These methods involve chemical breakdown using acids or fluxes to solubilize inorganic components, destroying organic matter and refractory structures to ensure complete matrix decomposition. Unlike selective extraction, dissolution aims for total digestion, enabling accurate quantification of elements via techniques such as (ICP-MS) or (ICP-OES). Acid digestion employs strong acids to oxidize and dissolve samples, commonly using nitric acid (HNO₃) alone or in combination with hydrochloric acid (HCl) for metals analysis. A mixture of HNO₃ and HCl, known as (typically 1:3 ratio), effectively dissolves noble metals and other recalcitrant forms by providing both oxidizing power from HNO₃ and chloride complexation from HCl. The general reaction proceeds as metal + acid → soluble salt + gas, for example, Zn + 2HCl → ZnCl₂ + H₂↑, releasing gases like NO₂ or Cl₂ during oxidation of organic components. Open-vessel digestion, performed on hot plates at around 95°C for 10-30 minutes per step, allows venting of fumes but risks analyte loss through volatilization, while closed-vessel variants minimize contamination and enhance efficiency. These procedures are standardized for sediments, sludges, and soils, using 1-2 g samples with stepwise addition of acids and hydrogen peroxide to achieve near-complete breakdown. Microwave-assisted digestion accelerates the process through pressurized, controlled heating in closed vessels, promoting faster and more uniform decomposition compared to conventional methods. Typically, 5 mL concentrated HNO₃ is added to a 45 mL sample aliquot, heated to 170-200°C for 10-30 minutes (including a 10-minute ramp-up), reaching pressures up to 200 bar to facilitate complete breakdown of inorganic matrices. This technique is particularly effective for aqueous liquids, wastes, and solids with suspended particles, reducing digestion time from hours to minutes while improving recovery rates above 90% for elements like Al, Cd, and Pb. Parameters such as temperature and duration are optimized based on sample type, with higher settings (up to 250-280°C) for refractory inorganics to ensure total solubilization without residue. Fusion techniques involve melting the sample with an alkali flux at high temperatures to dissolve highly refractory solids, such as silicates, that resist acid attack. Lithium metaborate (LiBO₂) or tetraborate fluxes are commonly used, mixed with the sample at a 1:4 ratio and heated to 900-1200°C (often around 1050°C) in platinum-gold crucibles for 5-10 minutes until a molten bead forms. The cooled glass-like bead is then dissolved in dilute acid (e.g., 5% HNO₃) to yield a solution for analysis, achieving total recovery of major and trace elements like Si, Al, and Zr in geological materials. This method minimizes selective dissolution issues but requires careful flux selection to avoid introducing contaminants, with non-wetting agents like LiBr aiding bead release. These techniques find primary application in environmental monitoring, where acid digestion and microwave variants prepare soil, sediment, and water samples for / to detect total concentrations of metals such as As, Cr, and Pb, ensuring regulatory compliance for contamination assessment. Fusion is favored for silicate-rich matrices in geochemical studies, providing reliable total analyte data for pollution tracking and ecosystem health evaluation.

Cleanup and Modification

Filtration and Separation

Filtration and separation techniques are essential physical methods in sample preparation to remove particulates, separate phases, and purify samples for subsequent analytical procedures, ensuring clarity and preventing instrument contamination. These processes typically follow initial extraction or dissolution steps, targeting the isolation of analytes from interfering solids or macromolecules in environmental, biological, or chemical matrices. Membrane filtration employs porous membranes, commonly with a 0.45 μm pore size, to retain particulates larger than this threshold while allowing soluble analytes to pass through, a standard practice in water and liquid sample preparation to eliminate suspended solids. For small sample volumes, typically up to 10 mL, syringe filters with nylon or similar membranes provide a convenient, low-hold-up volume option (<20 μL) for rapid filtration, minimizing analyte loss and suitable for laboratory-scale processing. Centrifugal filtration, often using spin columns or 96-well plates, enables high-throughput processing of multiple samples simultaneously, ideal for proteomic or environmental analyses requiring efficient volume reduction and purification. Centrifugation separates sample components based on density differences, commonly applied post-extraction to pellet solids or separate immiscible phases, with typical speeds ranging from 3000 to 10,000 rpm for 5-15 minutes depending on sample viscosity and particle size. This method achieves effective phase separation in biological extracts or solvent-based preparations, with relative centrifugal force (RCF) values around 2000-10,000 × g ensuring minimal emulsion formation while preserving analyte integrity. For biological samples, dialysis utilizes semi-permeable membranes to remove small molecules from macromolecules like proteins through diffusion, often over 12-24 hours in buffer exchanges, facilitating purification without denaturation. Ultrafiltration, a pressure- or centrifugation-driven variant, concentrates proteins or generates protein-free filtrates in biological matrices, retaining molecules above 3-100 kDa cutoffs while excluding smaller contaminants, commonly used in plasma or serum preparation. Efficiency in these techniques is gauged by retention rates, typically exceeding 95% for target particulates or analytes under optimized conditions, with pre-washing of filters using solvents or buffers preventing clogging by removing manufacturing residues and improving flow rates in turbid samples. Such practices minimize sample loss and maintain high recovery, critical for trace-level analyses.

Derivatization

Derivatization involves the chemical modification of analytes in sample preparation to enhance their suitability for analytical techniques, particularly by improving volatility, thermal stability, or ionization efficiency. This process is essential for polar or non-volatile compounds that are incompatible with methods like , where silylation converts hydroxyl groups in polar analytes, such as steroids or carbohydrates, into volatile trimethylsilyl ethers, thereby enabling effective separation and detection. In , derivatization can introduce chromophores or fluorophores to boost sensitivity, or enhance ionization for better mass spectrometric response. Derivatization is classified into pre-column and post-column types based on timing relative to chromatographic separation. Pre-column derivatization occurs prior to injection, allowing reactions to proceed under controlled conditions, such as esterification of carboxylic acids to form more lipophilic esters for reversed-phase LC; this approach often integrates with extraction steps for efficiency but requires complete reaction to avoid artifacts. Post-column derivatization, conversely, takes place after separation but before detection, demanding rapid kinetics to prevent band broadening, and is commonly used in HPLC for amino acid analysis with reagents like o-phthaldialdehyde (OPA). Common reagents include N,O-bis(trimethylsilyl)trifluoroacetamide (BSTFA) for silylation in GC, which reacts with active hydrogens in alcohols, amines, and acids to form stable derivatives. A representative silylation reaction is the conversion of an alcohol to its trimethylsilyl ether: \ce{R-OH + (CH3)3SiCl -> R-OSi(CH3)3 + HCl} This reaction exemplifies the replacement of labile protons with silyl groups to reduce polarity. Optimizing reaction conditions is critical for quantitative yields, with temperature typically ranging from to 70°C to accelerate without degrading analytes, and pH adjusted to 9–11 for basic-catalyzed reactions like to favor . Side reactions, such as over-silylation or in the presence of , can lead to incomplete derivatization or multiple peaks, necessitating conditions and excess . Recovery validation involves assessing derivatization efficiency through spiked standards, ensuring >90% to maintain analytical accuracy, often confirmed by comparing peak areas before and after modification.

Automation and Advances

Automated Systems

Automated systems in sample preparation utilize robotic and instrumental technologies to enhance efficiency in analytical workflows, particularly for high-throughput processing in and laboratories. These systems replace manual handling with programmable devices that perform repetitive tasks such as , , and dilution, ensuring consistency across large sample volumes. By integrating hardware like robotic manipulators and software controls, minimizes variability introduced by human operators and supports seamless transfer to downstream instruments such as liquid chromatography (LC) or () systems. Key types of automated systems include robotic arms designed for (SPE) and liquid-liquid extraction (LLE). Robotic arms, such as dual-arm configurations, handle complex sequences like pipetting, vortexing, and in SPE and LLE protocols, often feeding prepared samples directly into mass spectrometers. For instance, the SPE system employs a robotic platform for flexible SPE , accommodating various cartridge formats and solvents. Another type involves automated digesters integrated with technology, which accelerate acid for . Systems like the Milestone ultraWAVE use microwave-assisted heating in a single-reaction-chamber design to process multiple vials simultaneously, achieving complete in minutes while maintaining and . Benefits of these systems encompass reduced through precise, reproducible operations; continuous 24/7 functionality without fatigue; and direct integration with autosamplers for /, enabling end-to-end workflows. Automation enhances precision and accuracy by standardizing conditions like volume dispensing and timing, while lowering exposure risks to hazardous reagents. Examples include 96-well plate systems for , such as the APP96, which prepares up to 96 samples daily using filter-plate technology for , , and cleanup. Software platforms, like the Agilent 7696A Sample Prep WorkBench, allow method programming for tasks including derivatization and addition, with user-defined protocols stored for . Implementation involves balancing initial setup costs against long-term labor savings, with validation required under (GLP) standards. Automated systems typically require upfront investments of tens to hundreds of thousands of dollars for hardware and software, but they can yield significant labor reductions for routine tasks, with often achieved within a few years through decreased personnel needs and fewer errors. GLP compliance demands installation qualification (IQ), operational qualification (OQ), and performance qualification (PQ) to verify system reliability, , and , ensuring automated processes meet regulatory requirements for non-clinical safety studies.

Green and Modern Methods

Green and modern methods in sample preparation prioritize by adhering to principles that minimize hazardous use, energy consumption, and overall waste generation, aligning with the broader framework of . These approaches aim to replace traditional techniques that rely on large volumes of toxic organic solvents with eco-friendly alternatives that enhance efficiency while reducing environmental impact. Key tenets include selecting safer solvents, reducing sample size requirements, and integrating to lower energy needs, as outlined in the ten principles of green sample preparation. Such methods not only comply with regulatory pressures for greener practices but also improve analytical performance through higher selectivity and lower detection limits. Ultrasound-assisted (UAE) exemplifies these green principles by employing high-frequency sound waves to generate bubbles that disrupt sample matrices, facilitating analyte release with minimal and input. Typically, UAE operates in short cycles of 10-20 minutes, significantly shortening extraction times compared to conventional methods like Soxhlet extraction, which can require hours, while using up to 50-80% less solvent. This technique is particularly effective for solid samples such as plant materials or foods, yielding higher extraction efficiencies—often exceeding 90% recovery—for bioactive compounds without excessive heating, thereby preserving analyte integrity. Modern techniques like dispersive liquid-liquid microextraction (DLLME) advance solvent minimization by using only microliters (typically 10-100 μL) of extraction solvent dispersed into fine droplets within the aqueous sample, achieving rapid partitioning and enrichment factors up to 100-500. Compared to traditional liquid-liquid (LLE), which often consumes 10-100 mL of per sample, DLLME reduces organic usage by over 90%, making it a cornerstone of protocols for trace-level analysis in complex matrices like or biological fluids. Similarly, the QuEChERS method—acronymous for Quick, Easy, Cheap, Effective, Rugged, and Safe—has become a standard for pesticide residue in food samples, involving a simple salting-out step with followed by dispersive solid-phase cleanup, which eliminates the need for large volumes and laborious evaporation, achieving recoveries of 70-120% with minimal waste. Recent advances incorporate nanomaterial-based sorbents, such as carbon nanotubes or , which provide exceptionally high surface areas (up to 1000 m²/g) and tunable selectivity, enhancing adsorption efficiency in microextraction formats by factors of 2-10 over conventional sorbents while requiring sub-milligram quantities. These materials enable faster kinetics and reusability, further aligning with green goals by reducing material consumption. As of 2025, further innovations include deep eutectic solvents and molecularly imprinted polymers enhancing selectivity and sustainability. Complementing this, on-line sample preparation systems integrate pretreatment modules directly with analytical instruments like liquid chromatography or , automating extraction and minimizing transfer steps to cut solvent use by 50-70% and eliminate manual errors. Overall, these innovations demonstrate a shift toward scalable, low-impact methods that maintain analytical rigor.

Applications

Environmental and Forensic Analysis

In environmental , sample preparation is essential for detecting trace pollutants in complex matrices such as , , and s, where techniques must achieve high sensitivity to meet regulatory limits. (SPE) is widely employed for isolating polycyclic aromatic hydrocarbons (PAHs) from samples, enabling the concentration of analytes from large volumes to detect levels as low as nanograms per liter. For instance, EPA Method 8272 utilizes (SPME) to separate PAHs from pore , minimizing interference and facilitating subsequent gas chromatography-mass () analysis. Similarly, EPA Method 525.2 employs liquid-solid , a form of SPE, to preconcentrate PAHs and other organics from using octadecyl (C18) sorbents, achieving recoveries of 70-130% for most target compounds. Microwave-assisted is a standard technique for preparing samples to quantify , rapidly breaking down the matrix with acids like nitric and hydrochloric under controlled pressure and temperature. EPA Method 3051A outlines this process for sediments, sludges, and soils, using microwave energy to dissolve metals such as lead, , and mercury, with times reduced to 10-20 minutes compared to conventional heating. For semivolatile organic compounds, including PAHs and polychlorinated biphenyls, EPA Method 8270E specifies followed by cleanup, often via Soxhlet or pressurized liquid , to prepare solid wastes and soils for GC-MS determination at parts-per-billion (ppb) levels. The QuEChERS (Quick, Easy, Cheap, Effective, Rugged, and Safe) method has become a cornerstone for monitoring residues in environmental samples, particularly in soils and food-related matrices for assessment. This approach involves extraction, salting-out with and , and dispersive cleanup using sorbents like primary secondary amine (PSA) to remove matrix components. A validated QuEChERS variant for agricultural soils achieves recoveries of 70-120% for over 200 , with limits of detection in the low ppb range, making it suitable for under frameworks like the European Union's maximum residue limits. In , QuEChERS minimizes use and handles high matrix loads effectively, outperforming traditional methods in speed and reproducibility for multiclass analysis. In forensic , sample preparation ensures the integrity and admissibility of evidence from biological specimens, integrating rigorous protocols to trace contaminants or of abuse. Liquid-liquid (LLE) is commonly applied for blood determination, where is partitioned from using organic solvents like n-propanol or , followed by GC-flame ionization detection (FID) to quantify concentrations accurately. A validated LLE-GC-FID method for postmortem blood achieves from 0.01-4.0 g/L with recoveries exceeding 95%, addressing volatility and matrix effects in forensic casework. For drugs in , derivatization enhances volatility and detectability for GC-MS , converting polar compounds like amphetamines or opioids into more stable derivatives using such as N-methyl-N-(trimethylsilyl)trifluoroacetamide (MSTFA). Comparative studies of derivatization methods, including and , report improved peak shapes and sensitivities down to 10-50 ng/mL for benzodiazepines and metabolites in . Chain-of-custody procedures are integral to forensic sample preparation, documenting every handling step from collection to to prevent tampering or contamination, as mandated by standards from the . This includes sealed containers, tamper-evident seals, and electronic tracking during and derivatization to maintain evidentiary value. Specific challenges in these fields include detecting analytes at low concentrations, often in the ppb range, which necessitates preconcentration steps like SPE or LLE to amplify signals without introducing artifacts. Matrix interferences from in environmental samples, such as soils and waters, can suppress signals or cause ion suppression in ; cleanup with graphitized or silica-based sorbents in QuEChERS or SPE protocols mitigates this by selectively removing humics while retaining targets. These tailored preparations address regulatory demands, ensuring reliable quantification for pollution control and .

Pharmaceutical and Biomedical Applications

In pharmaceutical analysis, solid-phase extraction (SPE) is widely employed to isolate active pharmaceutical ingredients (APIs) and detect impurities in tablet formulations, enabling the removal of excipients that could interfere with chromatographic detection. For instance, SPE cartridges with reversed-phase sorbents (e.g., Oasis HLB) have been used to analyze dexamethasone and its impurities in low-dose formulations, achieving recoveries of approximately 100% while minimizing matrix effects in UHPLC-UV assays. Dissolution testing, governed by United States Pharmacopeia (USP) <711> standards, involves preparing samples by filtering the dissolution medium to separate undissolved particles from solubilized drug, ensuring accurate measurement of release profiles in quality control. This preparation step uses apparatus such as paddles or baskets in specified media (e.g., pH 1.2 to 6.8 buffers) to simulate gastrointestinal conditions, with filtration preventing clogging and maintaining assay linearity. In biomedical applications, serves as a rapid cleanup method for samples in , depleting high-abundance proteins like to enrich low-level analytes for downstream (). Organic solvents such as or are commonly added to , yielding precipitation efficiencies that enhance peptide identification depth by up to 50% in bottom-up workflows. Enzymatic digestion, typically with , follows to cleave precipitated proteins into suitable for analysis, optimizing conditions like pH 8 and 37°C incubation for complete in 4-16 hours. complements this by concentrating proteins through membranes (e.g., 10-30 kDa), removing small molecules and salts to improve proteomic resolution without denaturing sensitive biomolecules. Derivatization enhances the volatility and thermal stability of steroids in () for biomedical quantification, often using silylating agents like N,O-bis(trimethylsilyl)trifluoroacetamide to form trimethylsilyl ethers that boost sensitivity by 10-100 fold in detection. This technique is essential for analyzing endogenous steroids in , where it facilitates structural elucidation and trace-level monitoring in . Regulatory frameworks, such as FDA bioanalytical validation guidelines, require accuracy within ±15% and within ±15% CV for samples, with efficiency evaluated for across the analytical range, ensuring reliable pharmacokinetic data. Automated systems in pharmaceutical labs can streamline these preparations, reducing variability across batches.

Challenges and Best Practices

Common Challenges

One of the primary challenges in sample preparation is , which can introduce extraneous substances that interfere with analytical accuracy. particulates, such as or aerosols in the , represent a significant source, particularly for trace-level analyses where even low concentrations can skew results. Reagent impurities, including metal contaminants in acids or solvents, further exacerbate this issue by directly adding unwanted analytes during or steps. To mitigate these risks, conducting preparation in cleanrooms or controlled environments helps minimize exposure, ensuring particulate levels remain below critical thresholds for sensitive determinations. Analyte loss or instability during sample preparation often compromises quantification, especially for volatile or reactive compounds. Volatilization can occur during drying or evaporation steps, where heat or reduced pressure causes analytes to escape, leading to underestimation in subsequent analyses. Adsorption to container surfaces is another common problem, as analytes bind to glass or plastic walls, reducing recovery rates; for instance, mercury species in acidic solutions exhibit notable instability and loss when stored in non-inert materials like borosilicate glass without proper stabilization. These losses highlight the need for material selection, such as polytetrafluoroethylene (PTFE) vessels, to preserve analyte integrity. Matrix effects pose substantial hurdles in sample preparation for techniques like (), where co-eluting compounds from the sample matrix alter ionization efficiency. Ion suppression, caused by these interferents competing for charge in the , can reduce signal intensity by up to 50% or more, directly impacting quantification accuracy and leading to biased results. Such effects are prevalent in complex matrices like biological fluids or environmental extracts, underscoring the importance of thorough cleanup to isolate the target . Variability in sample preparation materials and processes introduces inconsistency across replicates or batches, undermining method reliability. Batch-to-batch differences in sorbents, such as variations in or surface chemistry in media, can result in recovery fluctuations, leading to inconsistent recovery rates across batches, which signals potential quality issues in analytical workflows.

Safety and Quality Control

Safety in sample preparation begins with the implementation of appropriate (PPE), including chemical-resistant gloves, safety goggles, face shields, and laboratory coats, to protect against splashes and spills from corrosive acids commonly used in and processes. Operations involving volatile or toxic reagents, such as solvents or gases, must be performed in certified chemical fume hoods to ensure adequate ventilation and prevent inhalation exposure. Particular caution is required when employing (HF) for the digestion of or silicate-containing samples, as it can cause severe, delayed burns to and tissues, corrode glassware, and lead to systemic toxicity; laboratories must maintain antidote gel nearby and conduct such procedures only in well-ventilated enclosures with specialized PPE like or gloves. Quality control protocols are integral to verifying the integrity of prepared samples and minimizing errors from contamination or loss. (CRMs), standardized samples with known analyte concentrations provided by organizations like NIST, are routinely incorporated to assess the accuracy and precision of preparation methods, ensuring results align with established values. Spike recovery tests, involving the addition of known analyte quantities to blank or real matrices, provide a measure of efficiency, with acceptance criteria typically set at 80-120% recovery to confirm reliable analyte retention without significant matrix interferences. Blank samples, processed identically to actual specimens but without analytes, are analyzed to identify background contamination from reagents, containers, or environmental sources, thereby safeguarding . Comprehensive documentation underpins and in sample preparation workflows. Standard operating procedures (SOPs) detail step-by-step instructions for handling, , and , promoting uniformity and reducing variability across personnel. trails, maintained through logs or chain-of-custody forms, all manipulations, including dates, operators, and deviations, enabling retrospective and regulatory audits. For laboratories pursuing formal recognition, adherence to ISO/IEC 17025 accreditation processes establishes robust systems, encompassing proficiency testing, internal audits, and continual improvement to meet international standards for competence in testing and . Troubleshooting post-preparation focuses on ensuring analytical instrument compatibility through targeted to mitigate issues like signal suppression or drift caused by residual . curves, constructed using standards matching the prepared sample's , verify and before , with adjustments made if deviations exceed predefined limits. In cases of incompatibility, such as high salt content affecting nebulization in ICP-MS, dilution or matching is applied, followed by re- to restore performance and accuracy.

References

  1. [1]
    Introduction to Sample Preparation - Chemistry LibreTexts
    Aug 29, 2023 · Sample preparation involves handling and modifying samples for analysis, including extraction, derivatization, and removing interfering ...
  2. [2]
    Sample Preparation - an overview | ScienceDirect Topics
    Sample preparation converts a lab sample into an analytical sample by removing non-relevant parts, ensuring the analytical portion is representative.
  3. [3]
    Overview of Liquid Sample Preparation Techniques for Analysis ...
    The primary role of sample preparation for analysis is usually concentration, the isolation of analytes from the sample matrix, and the removal of some or all ...
  4. [4]
    Sample Preparation for Bioanalytical and Pharmaceutical Analysis
    Oct 25, 2016 · A form of sample preparation prefaces virtually every analysis of a complex sample. Converting the sample into a format that is compatible with ...Solid Phase Extraction of... · Magnet-Based Sample... · Biographies
  5. [5]
    [PDF] Chapter 12, Laboratory Sample Preparation
    On first impression, sample preparation may seem the most routine aspect of an analytical protocol. However, it is critical that analysts realize and ...
  6. [6]
  7. [7]
    Introductory Chapter: Evolution of Sample Preparation - IntechOpen
    Dec 22, 2021 · Specialized sample preparations, such as self-assembly of analytes on nanoparticles for surface enhancement, have also evolved in parallel.
  8. [8]
    Origins of Sample Preparation Technologies | LCGC International
    Sep 1, 2021 · We look at five modern treatments of chromatographic sample extraction, explaining their initial conception and development to shed light on ...
  9. [9]
    Fifty years of solid-phase extraction in water analysis - ScienceDirect
    Since the first experimental trials using activated carbon filters that were performed 50 years ago, solid-phase extraction (SPE) has become an established ...
  10. [10]
    Microwave Digestion - Sample Preparation for Metals Analysis
    CEM invented microwave digestion in conjunction with the National Bureau of Standards(NBS) now NIST back in 1985 and received a special IR 100 award as a result ...
  11. [11]
    LC-MS/MS in the Clinical Laboratory – Where to From Here? - NIH
    Jan 18, 2011 · LC systems have also improved in user-friendliness and integration with the MS, and sample clean-up or extraction has begun to be integrated ...
  12. [12]
    Single-drop microextraction - ScienceDirect.com
    After its introduction just over twenty years ago, single-drop microextraction (SDME) became a widely applied solvent-based microextraction procedure as a ...
  13. [13]
    None
    Below is a merged summary of the key principles of sample preparation in analytical chemistry, consolidating all information from the provided segments into a comprehensive response. To retain maximum detail and ensure clarity, I will use a structured table format in CSV style for each principle, followed by a consolidated text summary of useful URLs. This approach allows for a dense representation of the data while maintaining readability and completeness.
  14. [14]
    A tutorial for computing limits of detection and quantification in ...
    May 29, 2022 · The present tutorial aims to review the most frequently reported criteria for the calculation of the limits of detection (LOD) and quantification (LOQ) in ...
  15. [15]
    What is Sample Preparation? | Teledyne LABS
    Sample preparation transforms a sample (solid, liquid, or gas) into a form that can be directly placed into an analytical instrument.<|control11|><|separator|>
  16. [16]
    Smart materials for sample preparation in bioanalysis: A green ...
    It is still the most labour-intensive and time-consuming analytical step, accounting for 60–80% of the total analysis time (Novakova, 2016). These steps can ...
  17. [17]
    Preparing samples for analysis - The key to analytical success
    ... Sample pretreatment is usually the weakest link of the analytical chain, providing till 60% of error sources of the whole procedure and consuming around 60- ...
  18. [18]
    Sample Preparation: A Comprehensive Guide - Organomation
    Sample preparation is the preliminary step in analysis, processing raw samples to make them suitable for analysis, ensuring accuracy and reliability.
  19. [19]
    Principles of Sample Preparation by Grinding or Comminution
    Oct 13, 2023 · Sample processing involves reducing the material size to ensure samples for homogeneity and extraction into a suitable matrix for analysis. The ...
  20. [20]
    Modern Sample Preparation Techniques: A Brief Introduction
    This introductory chapter intends to draw the readers' attention to the importance of sample preparation, the procedures of sampling and the source of errors.
  21. [21]
    Perspective on sample preparation fundamentals - ScienceDirect.com
    Sample preparation stands a pivotal stage in the analytical process. Unfortunately, the optimization of associated parameters often relies on trial and error ...
  22. [22]
    7: Obtaining and Preparing Samples for Analysis
    Sep 11, 2021 · It describes various sample types, including solutions, gases, and solids, and the methods used for their collection, preservation, and ...
  23. [23]
    Overview of Sample Preparation and Chromatographic Methods to ...
    Usually, several steps are required, involving up to about 80% of the analytical time. Figure 1 shows the main phases of the sample preparation methods. ...
  24. [24]
    [PDF] CHAPTER 4 - SAMPLING - FDA
    designate one sample as the "lead" sample. Enter that sample number in this field of the collection record for each related sample. Other related sample ...
  25. [25]
    [PDF] Method 1669: Sampling Ambient Water for Trace Metals at EPA ...
    Jul 10, 1996 · 2.10 Acid preservation of samples is performed in the field or in the laboratory. Field preservation is necessary for determinations of ...Missing: additives EDTA
  26. [26]
    [PDF] COLLECTION OF WATER SAMPLES - USGS Publications Warehouse
    Collect a field blank, if required, after sampling equipment has been ... Example procedure for collecting a field-blank quality-control sample.
  27. [27]
    Pre-analytical Practices in the Molecular Diagnostic Tests, A ... - NIH
    Viral infections can be detected in urine using PCR, using sterile containers for urine sample collection and transport at 4°C (159). ... 20 C by rapid phenol- ...
  28. [28]
    Evaluation of preservation techniques for trace metals and major ...
    There is a need to study the effectiveness of various amounts of nitric acid preservation for trace metals and major cations in surface water and groundwater.
  29. [29]
    [PDF] TRACE ELEMENTS SPECIMEN COLLECTION GUIDE
    Do not use preservatives, as they may cause contamination. Refrigeration is sufficient to preserve specimens. • Transport urine in trace element-free transport ...Missing: additives acids
  30. [30]
    [PDF] 2.2 Food Homogenization - FDA
    Food homogenization is needed for most foods to obtain representative analytical portions, as most foods are not inherently homogeneous.
  31. [31]
    Liquid-Liquid Extraction - an overview | ScienceDirect Topics
    Liquid-liquid extraction is a mass transfer operation where a liquid solution is mixed with an immiscible solvent, separating solutes based on solubility ...
  32. [32]
    A review of the modern principles and applications of solid-phase ...
    Oct 5, 2022 · Solid-phase extraction (SPE) is an effective way to prepare samples in chemistry. A sample collection and analysis gap can be closed using SPE ...
  33. [33]
    Extraction of pesticides from soil using supercritical carbon dioxide ...
    In this work we used the SFE to remove pesticides in soil using SC-CO2 as solvent added with methanol as cosolvent (SC-CO2/MeOH) at different pressures, ...
  34. [34]
    [PDF] Method 3050B: Acid Digestion of Sediments, Sludges, and Soils ...
    Method 3050B provides two digestion procedures for sediments, sludges, and soils, one for GFAA/ICP-MS and one for ICP-AES/FLAA analysis.
  35. [35]
    Understanding the Chemistry of Acid Digestion - Inorganic Ventures
    May 9, 2023 · Acid digestion serves as a critical step in sample preparation, enabling complex matrices to be converted into simplified, homogeneous solutions ...Missing: review | Show results with:review
  36. [36]
    [PDF] Method 3015A: Microwave Assisted Acid Digestion of Aqueous ...
    Digests produced by the method are suitable for analysis by inductively coupled plasma mass spectrometry (ICP-MS), inductively coupled plasma atomic emission ...
  37. [37]
    Microwave Digestion for ICP Analysis - Anton Paar Wiki
    Appropriate sample weight and heating time in microwave digestion for ICP analysis · A very low sample weight (100 mg to 300 mg) · A flat heating ramp (e.g., ...
  38. [38]
    Using Borate and Peroxide Fusions for Dissolution in ICP - AZoM
    Aug 27, 2020 · The borate fusion consists of mixing a powder sample with a fusion flux inside a platinum-5% gold crucible, heating to about 1000 – 1150 °C with agitation.
  39. [39]
    Borate Fusion Sample Prep – How It Works & Benefits | Katanax
    In sample preparation by fusion, the sample never actually melts. It is merely dissolved into a solvent. This solvent, generally a lithium borate flux, is solid ...
  40. [40]
    ICP-OES Sample Preparation | Thermo Fisher Scientific - US
    Learn how ICP-OES sample preparation is used to analyze solid, aqueous, organic, viscous, volatile, and other compounds.
  41. [41]
    Sample Preparation Guidelines - Center for Applied Isotope Studies
    For the analysis of dissolved analytes, the sample should be filtered through a 0.45 µm membrane filter. Acidify the filtrate with trace-metal grade nitric acid ...Missing: techniques | Show results with:techniques
  42. [42]
    [PDF] 5.2.2.B. Syringe-Filter Procedure for Processing Samples for ...
    A, use of the syringe-filter technique requires a relatively small volume of sample (10 mL) for analysis. Tandem mass spectrometry, as is used in schedules ...
  43. [43]
    Sample Preparation | Eldridge Reeves Johnson Foundation
    Resuspend small molecules in same dialysis buffer the day of the experiment. All buffers should be made fresh and filtered 0.2 µm. Purity: Protein samples ...
  44. [44]
    [PDF] Centrifugal Devices | SERC (Carleton)
    AcroPrep™ Advance 96- and 384-well Filter Plates are an excellent platform for a wide variety of molecular biology, analytical, and high throughput sample ...
  45. [45]
    [PDF] Unit Technical Manual - City of San Diego
    Jun 1, 2017 · Microcentrifuge the sample extract at maximum speed for 3-5 minutes. 5. The ... A minimum of 15 minutes of extraction time is recommended.
  46. [46]
    [PDF] Solution Processing and Optical Properties of 2D Transition Metal ...
    Dec 11, 2019 · The swollen sediment was re-dispersed with 50 mL of DI water, manually agitated (hand shaken) for 5-15 minutes, and centrifuged at 3500 rpm for ...
  47. [47]
    Chapter 3: Investigating Proteins - Chemistry
    The process of dialysis separates dissolved molecules by their size. The biological sample is placed inside a closed membrane, where the protein of interest is ...
  48. [48]
    Proteomic Challenges: Sample Preparation Techniques for ...
    Feb 5, 2015 · In order to obtain protein from a biological specimen, the sample must first be harvested from the organism, culture, or patient. Samples can be ...
  49. [49]
    Water pre-filtration methods to improve environmental DNA ... - NIH
    May 7, 2021 · We suggest that pre-filtration is an effective method of preventing filter clogging and contributes to increased eDNA yield, high sensitivity, ...Missing: rates | Show results with:rates
  50. [50]
    Analyte Derivatization as an Important Tool for Sample Preparation
    Jan 1, 2018 · Some precolumn derivatization schemes allow for the reactions to take place during sample extraction or inside the heated GC inlet. Post-column ...
  51. [51]
    The Pre‐ and Post‐Column Derivatization on Monosaccharide ...
    Jun 10, 2024 · Derivatization reactions can be conducted either pre-column to influence compound separation or post-column to impact only the detection step, ...
  52. [52]
    About Post-Column HPLC - Pickering Laboratories
    Post-column derivatization, also known as post-column reaction, renders visible certain compounds that are normally invisible.
  53. [53]
  54. [54]
    A Pre-Column Derivatization Method for the HPLC-FLD ... - MDPI
    Nov 22, 2024 · Temperature. Increasing the temperature often enhances the yield of the derivatization and shortens the reaction time, while the combination ...
  55. [55]
    The Importance of Derivatizing Reagent in Chromatography ... - NIH
    Jan 2, 2020 · A stable pH is necessary for biogenic amine recoveries, where the ideal pH is 11.5. Anything above pH 11.5 reduces the recovery of tyramine.
  56. [56]
    Modern automated sample preparation for the determination of ...
    Automation of sample preparation improves throughput, precision, and accuracy while reducing health risks associated with hazardous chemicals or biological ...
  57. [57]
    Application of a Dual-Arm Robot in Complex Sample Preparation ...
    The robot is flexibly used for automated processes in sample preparation and also in feeding the analytical instruments, such as mass spectrometers coupled with ...
  58. [58]
    The FREESTYLE SPE/EVA the ultimate in solid phase extraction
    The most flexible solution for automated Solid Phase Extraction (SPE), is LCTech's FREESTYLE SPE robotic system. Unique possibilities for sample prep.
  59. [59]
    ultraWAVE 3 Microwave Digestion System | Milestone SCI
    Samples are weighed into vials and suitable reagents are added. The vials are placed in a rack, which is automatically lowered into the reaction chamber. The ...
  60. [60]
    An Update on Current Trend in Sample Preparation Automation in ...
    Jul 1, 2024 · Automation in sample preparation improves accuracy, productivity, and precision in bioanalysis. Moreover, it reduces resource consumption for ...
  61. [61]
    APP96 System | Bruker
    The APP96 system automates LC-MS sample preparation, boosting lab efficiency with PreOmics iST technology. Process up to 96 samples daily with superior ...
  62. [62]
  63. [63]
    Automated Sample Preparation | Cut Impact, Save Time & Costs
    Sep 6, 2025 · Automation of the lab significantly reduces human errors that plague sample preparation outcomes. Manual processes handled by laboratory ...<|control11|><|separator|>
  64. [64]
    Labor Cost Savings from Automation: Stat Breakdown | PatentPC
    Oct 9, 2025 · Automated systems can operate at 60–80% less than the cost of human labor for the same task. Let's break that down. If it costs $20 an hour for ...
  65. [65]
    [PDF] Selection and Validation of Automated Sample Preparation Systems ...
    We herein endeavor to provide thorough guidance on the validation process, including Installation Qualification (IQ), Operational Qualification (OQ), and ...Missing: GLP | Show results with:GLP
  66. [66]
    The ten principles of green sample preparation - ScienceDirect.com
    The ten principles of GSP are presented with the aim of establishing a road map toward the development of overall greener analytical methodologies.
  67. [67]
    Green Approaches to Sample Preparation Based on Extraction ...
    The goal of sample preparation is to provide a representative, homogenous sample that is free of interferences and compatible with the intended analytical ...
  68. [68]
    A comprehensive review of ultrasonic assisted extraction (UAE) for ...
    This innovative technique not only reduces the extraction time but also enhances the quality and yield of the extract. The UAE exhibits exceptional advantages ...
  69. [69]
    Ultrasound assisted extraction of food and natural products ...
    Ultrasound extraction enhances extraction yield, economy and green impacts. Ultrasound acts directly in the plant cell to release metabolites.
  70. [70]
    Dispersive Liquid–Liquid Micro Extraction: An Analytical Technique ...
    The main advantages of this technique include its simplicity, environmentally friendly process, disposal costs, reduced solvent consumption, and short shelf ...
  71. [71]
    Advancements in overcoming challenges in dispersive liquid-liquid ...
    Microwave electromagnetic radiation and ultrasonic waves have been combined with DLLME to reduce or eliminate the necessity for dispersive solvent [45].
  72. [72]
    Review of the QuEChERS method for the analysis of organic ...
    QuEChERS is a quick, easy, cheap, effective, rugged, and safe extraction method, originally for pesticides, now used for multi-residue analysis of pollutants.Abstract · Introduction · Quechers Method<|separator|>
  73. [73]
    Analysis of pesticide residues using the Quick Easy Cheap Effective ...
    The Quick Easy Cheap Effective Rugged and Safe multiresidue method (QuEChERS) has been validated for the extraction of 80 pesticides belonging to various ...
  74. [74]
    Nanosorbents as Materials for Extraction Processes of ... - MDPI
    The application of NMs as sorbents in the extraction of organic pollutants in environmental samples allows for providing better sensitivity, repeatability, ...
  75. [75]
    Current advances and applications of online sample preparation ...
    Apr 12, 2022 · This review describes the most common online miniaturized LC configurations, and the main applications of current online sample preparation techniques.
  76. [76]
    Current Trends in Fully Automated On-Line Analytical Techniques ...
    Feb 1, 2019 · This critical review presents the current trends in sample preparation techniques based on solid phase extraction miniaturization, automation and on-line ...
  77. [77]
    Incorporating solid-phase extraction into compendial procedures for ...
    Oct 25, 2019 · In this work solid-phase extraction (SPE) was used to enrich analytes and minimize the interference of excipients, despite the fact that the ...<|control11|><|separator|>
  78. [78]
    The Dissolution Procedure
    Filtration is a key sample-preparation step in achieving accurate test results. The purpose of filtration is to remove undissolved drug and excipients from the ...
  79. [79]
    [PDF] 〈711〉 DISSOLUTION - US Pharmacopeia (USP)
    Dissolution test smooth, vertically reciprocating cylinder. A device is used procedures using a flow-through cell must be characterized that allows the ...
  80. [80]
    Comparison of protein precipitation methods for sample ... - PubMed
    To find the most efficient method for sample preparation, we used human plasma and compared four widely applied precipitation methods, using trichloroacetic ...
  81. [81]
    A simple, time- and cost-effective, high-throughput depletion strategy ...
    Mar 29, 2023 · The method is based on selective protein precipitation by perchloric acid (perCA) (4). Our implementation of the method, which is built upon ...Results And Discussion · Materials And Methods · Sample Preparation
  82. [82]
    Multiple-Enzyme-Digestion Strategy Improves Accuracy and ...
    Oct 18, 2018 · The MED FASP method using three different enzymes: LysC, trypsin and chymotrypsin increased the no. of identified peptides and proteins and ...Experimental Section · Figure 1 · Discussion
  83. [83]
    Ultrafiltration for proteomic sample preparation - PubMed
    Ultrafiltration offers an alternative sample preparation technique that enables more thorough analysis of a proteome.
  84. [84]
    [PDF] Bioanalytical Method Validation - Guidance for Industry | FDA
    May 24, 2018 · Recommendations and Acceptance Criteria for Bioanalytical Method Validation and In-Study Conduct. Parameters. Validation Recommendations. In ...
  85. [85]
    [PDF] M10 BIOANALYTICAL METHOD VALIDATION AND STUDY ... - FDA
    This guidance describes the validation of bioanalytical methods and study sample analysis that are expected to support regulatory decisions. The guidance is ...
  86. [86]
    Clean Chemistry for Elemental Impurities Analysis of ...
    Nov 19, 2015 · This article focused on sample contamination during sample preparation. Contaminations from environment, reagents, and lab apparatus were investigated for ...Results And Discussion · Icp-Ms Calibration And... · Contamination From Sampling
  87. [87]
    [PDF] Clean laboratories and clean rooms for analysis of radionuclides ...
    One of the main sources of contamination of samples is through particulate matter from the laboratory environment. The laboratory determination of elements at ...
  88. [88]
    [PDF] Guide for environmental sample analysis by ICP-MS
    Environmental analysis uses techniques to assess pollution and toxicity in water, soil, air, and industrial wastes, driven by environmental laws.Missing: humics | Show results with:humics<|separator|>
  89. [89]
    Methods and recent advances in speciation analysis of mercury ...
    Mar 7, 2016 · Furthermore, analyte loss due to adsorption on surface of the containers can be minimized using PTFE vessels.[34]. In addition, sample ...
  90. [90]
    Matrix effects demystified: Strategies for resolving challenges in ...
    Oct 28, 2023 · This review aims to highlight common challenges associated with matrix effects throughout the entire analytical process with emphasis on gas chromatography- ...
  91. [91]
    Needle Trap Devices in Analytical Chemistry: A Critical Review of ...
    Mar 27, 2025 · Variability in sorbent materials, extraction conditions, and analytical methods can lead to inconsistent results across studies. Varying sorbent ...
  92. [92]
    Working with Chemicals - Prudent Practices in the Laboratory - NCBI
    Use laboratory chemical hoods and other ventilation devices to prevent exposure to airborne substances whenever possible. Do not underestimate hazards or risks.
  93. [93]
    Fume Hoods - Environment, Health & Safety
    Fume hoods provide a safe, enclosed work area for chemical manipulations and provide critical ventilation for the user and laboratory.Using A Fume Hood · Specialty Fume Hoods · Fume Hoods With Motion...<|control11|><|separator|>
  94. [94]
    [PDF] Rapid Method for Acid Digestion of Glass-Fiber
    Oct 22, 2012 · This procedure uses hydrofluoric acid (HF) to digest glass filters and any siliceous material in the sample. Do not use glass beakers in ...
  95. [95]
    Hydrofluoric Acid (HF) - Division of Research Safety
    Aug 20, 2025 · Hydrofluoric acid (HF) is extremely corrosive, causing deep tissue damage, bone destruction, and systemic effects. Even small exposure can be ...Missing: analytical | Show results with:analytical
  96. [96]
    [PDF] Development and use of reference materials and quality control ...
    The application of certified reference materials (CRMs) in analytical chemistry for quality control purposes is well recognized and recommended by a wide ...
  97. [97]
    What is the acceptance spike recovery range in order to validate ...
    Jul 4, 2016 · The spike recovery range for pharma industry should be 80-120 for finished product and 70-130 % for content uniformity.Accepted range of relative recovery? - ResearchGateIs Recovery an essential parameter for method validation?More results from www.researchgate.net
  98. [98]
    A Practical Guide to Immunoassay Method Validation - PMC
    Calculate the recovery using formula (7). Note: acceptance range for the recovery is usually 80–120%. % Recovery = Measured concentratio n spiked sample ...
  99. [99]
    The Vital Role of Blanks in Sample Preparation | LCGC International
    Aug 1, 2018 · Blank samples, along with replicates, standards, and spiked samples, are crucial to quality control and the development of robust quantitative ...
  100. [100]
    Quality Assurance (QA) - Quality Control (QC) | US EPA
    Sep 8, 2025 · Use of dilution or internal standards, or both, increases the likelihood of a passing matrix spike recovery above 50%, removing the need for the ...
  101. [101]
    [PDF] Guidance for Preparing Standard Operating Procedures (SOPs)
    Apr 6, 2007 · SOPs describe both technical and fundamental programmatic operational elements of an organization that would be managed under a work plan or a ...<|control11|><|separator|>
  102. [102]
    Achieving ISO 15189 Accreditation with Scispot | Tips and Tricks
    Sep 16, 2025 · It maintains audit trails for all actions and changes, providing transparency and traceability required for ISO 15189 certification.
  103. [103]
    Preparing for an ISO/IEC 17025 Accreditation Assessment
    Ensure your lab meets ISO/IEC 17025 standards. Explore steps for internal audits, addressing non-conformities, and improving lab practices.
  104. [104]
    Calibration: Effects on Accuracy and Detection Limits in Atomic ...
    Aug 1, 2021 · Attaining accurate quantitative results in atomic spectroscopy requires proper calibration, especially when measuring low-level ...Missing: troubleshooting | Show results with:troubleshooting
  105. [105]
    Calibration – an under-appreciated component in the analytical ...
    Calibration should be performed whenever modifications are made to the reagent (e.g., using a fresh batch or changing the lot) and/or the instrument (due to ...