Ion chromatography

Ion chromatography (IC) is a form of (HPLC) specifically designed to separate and quantify ions and polar molecules in solution based on their electrostatic interactions with a charged stationary phase, typically an , while a mobile phase elutes the analytes for detection. The technique primarily targets inorganic anions (such as , , and ) and cations (such as sodium, , and calcium), as well as certain organic acids, enabling analysis at trace levels down to parts-per-billion (ppb) concentrations with high sensitivity and precision. The foundational principles of IC rely on ion-exchange mechanisms, where analytes are retained on the through reversible electrostatic and selectively displaced by competing ions in the eluent, with separation efficiency governed by factors like charge, size, and . Additional modes include ion-pair chromatography, which uses hydrophobic interactions with paired counterions, and ion-exclusion chromatography, which separates weak acids or bases via Donnan exclusion effects on a cation-exchange . Detection is most commonly achieved through suppressed or nonsuppressed , where the eluent's background conductance is minimized to enhance signal-to-noise ratios, though alternatives like UV-Vis or are employed for specific analytes. IC was pioneered in 1975 by Hamish Small, Timothy S. Stevens, and William C. Bauman at Dow Chemical, who introduced a novel system combining ion-exchange separation with low-background conductimetric detection using a suppressor column to convert the eluent into a weakly conducting form. This innovation addressed limitations of earlier ion-exchange methods, which suffered from high background noise due to conductive eluents, marking the birth of modern IC as a practical analytical tool. Subsequent developments, such as nonsuppressed conductivity detection in 1979, expanded its versatility and led to widespread commercialization. Key components of an IC system include a high-pressure for eluent delivery, an injection for sample introduction, a separation column packed with functionalized (e.g., polystyrene-divinylbenzene copolymers with quaternary ammonium or groups), a suppressor or guard column to reduce background, and a detector coupled to software for generating chromatograms. Stationary phases vary by application, with low-capacity resins for trace and higher-capacity ones for complex matrices, while eluents like for anions or for cations are optimized for selectivity and flow rates typically between 0.5–2.0 mL/min. Advances in column technology, such as pellicular beads with superficial ion-exchange layers, have improved and reduced times to under 30 minutes for routine separations. IC finds extensive applications in for assessing , including the determination of anions in rainwater, surface waters, and to detect pollutants like nitrates and sulfates. In the , it quantifies additives, preservatives, and content in products like beverages and ; in pharmaceuticals, it ensures purity by analyzing ionic impurities; and in clinical settings, it measures electrolytes in biological fluids. Its advantages include simultaneous multi-analyte detection, robustness for diverse matrices (aqueous, solid extracts), and compliance with standards like EPA Method 300.0 for regulatory testing, making it indispensable for and research.

Introduction and Background

Overview of Ion Chromatography

Ion chromatography is a form of (HPLC) that separates ions and polar molecules based on their affinity to the ion exchanger using an aqueous eluent. This technique leverages the physiochemical interactions between analytes and a stationary phase to achieve high-resolution separations of charged species in solution. Common analytes include inorganic anions such as (Cl⁻) and (NO₃⁻), cations like sodium (Na⁺) and calcium (Ca²⁺), as well as certain organic acids. The core process involves , where analytes compete with eluent ions for binding sites on the stationary phase. The general entails injecting the sample into the chromatographic , followed by separation on an ion-exchange column and detection, most commonly via to measure ionic as they elute. This approach enables efficient analysis of complex samples with minimal . Ion chromatography is vital for trace-level analysis, achieving detection limits from parts per million (ppm) to (ppb), such as 5–10 μg/L for anions. It supports in , pharmaceuticals, and , as well as fundamental research in .

Historical Development

The foundations of ion chromatography trace back to early developments in ion exchange technology during the 1930s and 1940s. In 1935, British chemists Basil A. Adams and Eric L. Holmes synthesized the first organic ion exchange resins through the condensation of phenolsulfonic acids with formaldehyde, creating materials capable of selectively exchanging cations or anions in aqueous solutions. These phenol-formaldehyde-based resins marked a significant advancement over natural exchangers like zeolites, enabling more controlled and efficient ion separation processes in water softening and analytical applications. By the 1940s, these synthetic resins were commercialized, laying the groundwork for future chromatographic techniques by demonstrating the practical utility of ion exchange in separating ionic species. Modern ion chromatography emerged in the 1970s as a specialized form of chromatography optimized for conductometric detection. In 1975, Hamish Small, Timothy S. Stevens, and William C. Baumann at developed the technique, introducing a suppressor column to reduce eluent background and enable sensitive detection of ions. Their seminal paper described a system using a low-capacity anion exchanger with a sodium bicarbonate/sodium carbonate eluent, followed by a suppressor that converted the eluent to while transforming sample anions into detectable acids, achieving separations of common inorganic anions at parts-per-million levels. This innovation, patented by Dow, revolutionized ion analysis by combining high-efficiency separation with low-noise detection, distinguishing it from prior methods. Commercialization accelerated in the late 1970s and 1980s, driven by the formation of Dionex Corporation in 1975 to license and market the technology. Dionex released the first commercial suppressed ion chromatograph, the Model 10, in 1975, which was publicly demonstrated at the meeting that year and quickly adopted for environmental and industrial monitoring. By the 1980s, Dionex (later acquired by ) expanded the market through improved instrumentation, leading to widespread use in laboratories worldwide for routine anion and cation analysis. A key milestone in the technique's evolution came in 1979 with the introduction of nonsuppressed ion chromatography, offering an alternative to suppressor-based systems. Donald T. Gjerde, James S. Fritz, and Gabor Schmuckler published a method using low-capacity resins and dilute aromatic acid eluents, such as p-hydroxybenzoic acid, to achieve low background without a suppressor, enabling direct detection of anions at sub-ppm concentrations. This approach simplified and gained traction in the , particularly for applications requiring minimal setup. During the and , suppressed ion chromatography remained dominant due to its superior sensitivity, but nonsuppressed variants co-evolved alongside suppressor improvements, such as micromembrane designs, broadening the method's versatility and adoption in diverse analytical fields.

Principles of Operation

Basic Mechanism

Ion chromatography relies on the process, a reversible interaction between ions in the sample and the stationary phase, which consists of a with fixed charged functional groups. The stationary phase, typically cross-linked beads, binds counterions that can be displaced by sample ions of the same charge type through electrostatic attraction and other intermolecular forces. This exchange allows for the selective separation of ionic species as the sample is carried through the column by the mobile phase. The selectivity of separation in ion chromatography is primarily governed by the charge, hydrated size, and of the ions, with higher-charged ions generally exhibiting stronger retention due to greater electrostatic interactions with the . For instance, divalent ions like are retained longer than monovalent ions like under similar conditions. The underlying for this process can be represented as \ce{R-X + M+ ⇌ R-M + X}, where \ce{R} denotes the , \ce{X} is the initially bound to the , and \ce{M+} is the sample cation; a similar applies for anions. This determines the extent of ion binding and thus the separation efficiency. The eluent, or mobile phase, plays a crucial role by introducing competing ions that displace the sample ions from the , facilitating their in order of decreasing affinity. For , a common eluent is (NaOH), which provides ions to compete with anionic analytes, with concentrations typically ranging from 1 to 100 mM to control retention. This competition ensures that ions with weaker interactions elute first, producing distinct peaks based on their relative affinities. Retention times in ion chromatography are influenced by the distribution coefficient K_d, defined as the ratio of the amount of analyte in the stationary phase to that in the mobile phase (K_d = \frac{[\text{analyte}]_{\text{stationary}}}{[\text{analyte}]_{\text{mobile}}}). Higher K_d values correspond to longer retention times, as ions spend more time bound to the resin; factors like eluent strength and resin capacity modulate this coefficient to optimize separations. The capacity factor k' = K_d \cdot \frac{V_s}{V_m} (where V_s and V_m are stationary and mobile phase volumes, respectively) further quantifies retention, ideally maintained between 1 and 10 for baseline resolution. Ion chromatography operates in two primary modes: anion-exchange, which uses a positively charged stationary phase (e.g., quaternary ammonium groups) to retain and separate anions such as chloride and nitrate, and cation-exchange, which employs a negatively charged stationary phase (e.g., sulfonic acid groups) to separate cations like sodium and calcium. In anion-exchange, the resin's positive charges attract negative sample ions, while in cation-exchange, negative charges on the resin bind positive ions; eluent composition is adjusted accordingly, such as using methanesulfonic acid for cations. These modes enable versatile analysis of both anionic and cationic species in complex matrices.

Detection and Suppression

In ion chromatography, the primary detection method is conductivity detection, which measures changes in the electrical conductance of the eluent caused by the separated ions as they elute from the column. This technique relies on the fact that ions in solution conduct electricity, producing a signal proportional to their concentration, with the detector cell typically consisting of two electrodes separated by a known volume of eluent. Conductivity detection is universal for ionic species but can be limited by high background conductance from the eluent itself. To improve sensitivity, suppressed ion chromatography employs an ion suppressor device placed between the column and the conductivity detector, which chemically reduces the background conductance of the eluent while preserving the conductance of the analyte ions. In anion analysis with a hydroxide eluent, for example, a cation-exchange membrane suppressor exchanges sodium ions (Na⁺) from the eluent for hydronium ions (H⁺), converting NaOH to water (with Na⁺ removed to waste): NaOH + H⁺ → H₂O + Na⁺ (waste), resulting in a low-conductivity effluent (typically <1 µS/cm) and enabling detection of weakly conducting analyte acids. Membrane-based suppressors, such as self-regenerating types, offer continuous operation with high capacity and minimal noise (0.2–0.5 nS), outperforming earlier packed-bed suppressors that required periodic regeneration. In contrast, nonsuppressed ion chromatography avoids suppressors and instead uses low-conductivity eluents, such as dilute solutions of benzoate or phthalate salts (e.g., 3–4 mM), to minimize background interference and allow direct conductivity measurement. This mode produces higher baseline noise (5–10 nS) and limits sensitivity compared to suppressed systems, as the eluent background remains elevated (around 1100 µS/cm), but it simplifies instrumentation for routine analyses of common ions. Suppressed methods generally achieve 10-fold better minimum detection limits (MDLs), such as 0.19 µg/L for Li⁺ versus 2.0 µg/L in nonsuppressed mode. Alternative detection methods complement conductivity for ions lacking strong conductance or requiring higher selectivity. UV-Vis detection is applied to UV-absorbing species like nitrate or iodide, or via post-column derivatization (e.g., with 4-(2-pyridylazo)resorcinol for transition metals, absorbing at 500–520 nm), achieving MDLs around 0.21–0.5 µg/L for specific analytes. Amperometric detection, often in pulsed mode with a gold electrode, targets redox-active ions such as nitrite or carbohydrates, measuring current from oxidation or reduction (e.g., at +50 mV for glucose), with sensitivities enhanced by ~10% analyte turnover. Overall sensitivity for common ions in suppressed conductivity detection typically reaches 0.05–1 ppm (µg/mL), depending on the analyte and system, enabling trace-level analysis in environmental and pharmaceutical samples.

Components and Techniques

Ion Exchangers

Ion exchangers serve as the stationary phase in ion chromatography, facilitating the separation of ions based on their affinity for charged functional groups attached to a solid support. These materials are typically cross-linked polymer resins designed to withstand the chemical conditions of chromatographic separations, with selectivity determined by the type and density of ionic sites. Strong ion exchangers maintain a constant charge density across a wide pH range, typically from 1 to 14, due to their fully dissociated functional groups. For anion exchange, quaternary ammonium groups, such as trimethylammonium or dimethylethanolammonium, provide a permanent positive charge. For cation exchange, sulfonic acid groups (–SO₃⁻) remain ionized regardless of pH, offering consistent performance without buffering effects. These exchangers are preferred in ion chromatography for their stability and ability to handle diverse sample matrices. In contrast, weak ion exchangers exhibit pH-dependent ionization, with charge varying based on the of their functional groups, limiting their operational range but allowing tunable selectivity. Weak cation exchangers often use carboxylic acid groups (–COOH, ≈ 4–5), which deprotonate above this pH to become negatively charged. Weak anion exchangers typically feature tertiary or secondary amine groups ( ≈ 9–10), which protonate below this pH to gain positive charge. This pH sensitivity enables applications where mild conditions are needed, though it requires careful buffer selection to maintain capacity. The backbone of most ion exchangers in ion chromatography is a polystyrene-divinylbenzene (PS-DVB) copolymer, chosen for its mechanical stability, chemical resistance across pH 0–14, and ability to support high cross-linking (4–16%) for durability under pressure. Functional groups are covalently attached to this matrix via processes like chloromethylation followed by amination for anions or sulfonation for cations. Particle sizes are optimized at 5–10 μm to balance resolution and flow rates in packed columns. Ion-exchange capacity, measured in milliequivalents per gram (meq/g) or microequivalents per gram (μeq/g), quantifies the number of available charged sites, typically low (0.01–0.1 meq/g) in ion chromatography to minimize background conductivity and enhance detection sensitivity. Selectivity arises from differences in ion affinity, influenced by charge density and size, while Donnan exclusion repels co-ions (those with the same charge as the resin) via electrostatic repulsion at the resin-eluent interface, preventing non-target penetration and improving separation efficiency. Preparation of ion exchangers involves suspension polymerization of styrene and divinylbenzene to form microbeads, followed by surface functionalization; for instance, PS-DVB beads are sulfonated with sulfuric acid for strong cation exchangers or grafted with quaternary amines for strong anion types. In IC, columns are maintained by periodic equilibration with eluent; for heavily fouled columns, mild cleaning protocols (e.g., dilute acid/base rinses) may be used, but replacement is common due to low-capacity design.
PropertyStrong ExchangersWeak Exchangers
Charge BehaviorConstant over pH 1–14pH-dependent (cation: active >pKa 4–5; anion: active <pKa 9–10)
Functional Groups (Anion)Quaternary ammonium (e.g., –N(CH₃)₃⁺)Tertiary/secondary amines (e.g., –NR₂)
Functional Groups (Cation)Sulfonic acid (e.g., –SO₃⁻)Carboxylic acid (e.g., –COO⁻)
Typical Capacity (IC)0.01–0.1 meq/g (low for sensitive detection)0.1–1 meq/g (less common in IC, higher in specialized applications)
MaintenanceEluent equilibration; mild cleaning if neededpH adjustment or mild acids/bases for equilibration

Instrumentation

Ion chromatography systems consist of integrated hardware designed to deliver eluent, inject samples, separate ions, suppress background conductivity, and detect analytes with high precision. The core setup typically includes a high-pressure pump, injection valve, separation column with guard column, suppressor, detector, and supporting automation and safety features. These components are often constructed from inert materials like or titanium to withstand corrosive eluents such as sodium hydroxide or carbonate solutions. The high-pressure pump is essential for delivering the eluent at controlled flow rates, typically ranging from 0.5 to 3 mL/min, to ensure consistent separation without excessive back pressure (typically up to 1000–3000 psi, though traditional systems may operate at 300–500 psi). These pumps, often dual-piston or isocratic models, maintain pulse-free flow to minimize baseline noise and are compatible with low-conductivity eluents. The injection valve, available in manual or automated configurations, introduces precise sample volumes via sample loops of 10–100 μL, allowing reproducible injections with minimal band broadening. Separation occurs in the analytical column, which features dimensions such as 4.6 mm inner diameter (ID) and lengths of 25–250 mm, packed with ion-exchange resin particles (3–9 μm). Guard columns, shorter versions (e.g., 10–50 mm length, same ID), are placed upstream to protect the analytical column from particulates, multivalent ions, or contaminants, extending column lifespan. Suppressors and detectors are integrated post-column for enhanced sensitivity. Micromembrane suppressors, such as those using cation-exchange membranes, chemically convert the eluent (e.g., NaOH to H2O) to reduce background conductivity to below 1 μS/cm, enabling detection limits in the ppb range. Conductivity cells serve as the primary detectors, featuring low-volume flow paths (e.g., 0.5–3 μL) and temperature control (≤0.001 °C stability) to measure analyte conductivity changes accurately. System automation is facilitated by dedicated software that controls gradient elution—progressively varying eluent composition for complex separations—and handles real-time data acquisition, peak integration, and method optimization. Inline filters (e.g., 2 μm) and calibration routines ensure reliability. Safety and maintenance features include eluent degassing systems (using helium, nitrogen, or vacuum) to remove dissolved gases and prevent bubble formation, which could disrupt flow. Leak detection sensors monitor fittings and tubing, alerting users to potential issues via salt deposits or pressure anomalies, while modular designs allow easy replacement of consumables like suppressors and columns.

Analytical Procedure

The analytical procedure for ion chromatography begins with sample preparation to ensure compatibility with the system and to minimize contamination or column damage. Samples are typically filtered through a 0.45 μm membrane to remove particulates larger than this size, which could clog the column or interfere with separation. Dilution is performed as needed using high-purity water or eluent to bring the sample within the linear range of the detector and to reduce matrix effects, often at ratios such as 1:10 or higher depending on initial conductivity. If the sample matrix contains high levels of interfering substances, matrix elimination techniques, such as offline solid-phase extraction or ion-exchange cartridges, may be applied to isolate target analytes. System setup involves selecting an appropriate eluent and equilibrating the column for stable performance. For anion analysis, common eluents include hydroxide-based solutions, such as 10-35 mM NaOH, or carbonate/bicarbonate mixtures like 1.7 mM NaHCO₃/1.8 mM Na₂CO₃, chosen based on the analytes and column type to achieve optimal selectivity and low background conductivity. The column, typically packed with ion exchangers, is equilibrated by pumping the eluent at the operating flow rate (e.g., 0.7-2.0 mL/min) until a stable baseline is achieved, which usually takes 30-45 minutes and confirms consistent retention times. The run sequence commences with sample injection, followed by separation and detection. A precise volume, such as 10-100 μL, is injected via an autosampler loop into the eluent stream, ensuring reproducibility across runs. Separation occurs as ions interact differentially with the stationary phase, typically requiring 5-30 minutes per run depending on the column length and flow rate, after which peaks are identified by comparing retention times to those of known standards run under identical conditions. Data processing entails peak integration and quantification to determine analyte concentrations. Chromatographic software integrates peak areas or heights, which are then plotted against a calibration curve constructed from multiple standards (e.g., 0.1-10 mg/L range) exhibiting linear response with correlation coefficients ≥0.995. Concentrations are calculated using the curve equation, adjusted for dilution factors, providing results in units such as mg/L. Quality control measures are integrated throughout to validate accuracy and precision. Blanks (e.g., high-purity water) are analyzed at the start and every 10 samples to detect contamination, with analyte levels required to be below 10% of the lowest calibration standard. Spikes, such as matrix or laboratory control samples at known concentrations, are run periodically to assess recovery (typically 80-120%), while system suitability tests, including initial and continuing calibration verifications, ensure deviations remain within ±10-20% of expected values. Duplicates are also performed to check relative percent differences, limited to ≤15% for mid-range concentrations.

Specialized Variants

Membrane Ion Chromatography

Membrane ion chromatography utilizes ion-exchange membranes for separation and purification, particularly effective for large biomolecules like proteins through convective mass transport, which reduces diffusion limitations compared to traditional packed columns. In ion chromatography (IC) workflows, these membranes are commonly employed for inline sample preparation, such as matrix elimination and preconcentration, rather than as primary stationary phases for separating small inorganic ions, where packed resin columns remain standard due to higher binding capacity and resolution. Designs typically feature flat-sheet or hollow-fiber configurations with embedded ion-exchange groups, such as quaternary ammonium for anion exchange or sulfonate for cation exchange. Flat-sheet membranes are often stacked in modular cassettes, while hollow-fiber formats support radial or annular flow, enabling efficient contact with minimal pressure drops—typically below 100 psi (about 7 bar)—even at higher flow rates, thanks to through-pore convection rather than diffusion-dominated transport. Advantages include support for flow rates up to 10 mL/min, which shortens processing times, and low dead volumes that permit microliter-scale samples, ideal for limited biological specimens. Inline preconcentration via Donnan dialysis selectively accumulates target ions across the membrane before elution, boosting sensitivity in trace analyses. These properties enhance throughput in bioprocessing and IC sample handling. In IC applications, membrane systems are valuable for matrix elimination, such as removing interfering alkaline earth metals from seawater, and sample cleanup to isolate anions or cations prior to separation on packed columns. Anion-exchange membranes can retain matrix ions while permitting analytes to permeate, simplifying preparation for environmental or industrial matrices and preconcentrating dilute species to lower detection limits without manual steps. Limitations include lower resolution for complex mixtures, where packed columns provide superior theoretical plate counts via extended diffusion paths, and reduced binding capacity from constrained surface area, rendering them unsuitable for high-concentration or multi-component small-ion separations. They serve best as complementary devices for pretreatment in intricate IC analyses. Commercial examples include Sartobind ion-exchange membrane adsorbers from Sartorius and Mustang membranes from Cytiva (formerly Pall), which integrate into IC and bioprocessing systems for sample pretreatment and purification tasks. As of 2025, advances in membrane materials, such as UV-grafted hydrophilic polymers, have improved capacity and stability for broader applications in downstream bioprocessing.

Protein Separation Techniques

Ion exchange chromatography (IEC), a specialized variant of ion chromatography adapted for macromolecules, enables the separation of proteins based on their net surface charge, which varies with pH relative to their isoelectric point (pI). Weak ion exchangers, such as diethylaminoethyl (DEAE) cellulose for anion exchange or carboxymethyl (CM) cellulose for cation exchange, are particularly suited for pH-based fractionation, as their charge capacity changes with buffer pH, allowing proteins to bind at one pH and elute as the pH shifts toward their pI. This approach, often implemented via chromatofocusing, creates a stable pH gradient along the column, resolving proteins into narrow bands according to their pI values, with resolutions improved by up to 1.5-fold compared to salt gradients in some cases. The Gibbs–Donnan effect further influences protein retention by establishing an electrochemical equilibrium that unevenly distributes ions across the charged resin-protein interface. Gradient elution strategies enhance selectivity in protein separations, typically employing increasing salt concentrations (e.g., 0–1 M NaCl) to disrupt electrostatic interactions or pH ramps (e.g., from pH 9 to 4) to neutralize protein charges sequentially. In salt gradients, competing ions shield the protein-resin binding, eluting less charged proteins first, while pH gradients exploit titration of ionizable groups for orthogonal separation, often yielding higher peak capacities for complex mixtures like cell lysates. Column selection emphasizes high-capacity resins, such as perfusive particles like or tentacle-based exchangers like , which accommodate large biomolecules (up to 150 kDa) without pore exclusion and minimize denaturation through hydrophilic matrices and mild operating conditions (e.g., pH 4–9). Despite these advantages, challenges in IEC for proteins include the need for low flow rates (0.1–1 mL/min) to prevent shear-induced denaturation and ensure mass transfer equilibrium in viscous samples, as well as careful buffer selection to maintain compatibility and avoid precipitation (e.g., using low-conductivity zwitterionic buffers like MES at 20–50 mM). These adaptations have proven effective in practical applications, such as the purification of monoclonal antibodies (mAbs) via cation exchange to remove aggregates and host cell proteins, achieving >95% purity in capture steps, or enzyme isolation like using anion exchange gradients for high-yield recovery.

Applications

Environmental Analysis

Ion chromatography plays a pivotal role in by enabling the precise quantification of inorganic ions in various matrices such as , , and soil extracts, ensuring compliance with regulatory standards for . In analysis, it is routinely applied to determine key anions including (F⁻), (Cl⁻), (NO₂⁻), (Br⁻), (NO₃⁻), orthophosphate (PO₄³⁻), and (SO₄²⁻), following U.S. Environmental Protection Agency (EPA) Method 300.1, which specifies suppressed detection for achieving detection limits as low as 0.01–0.5 mg/L depending on the anion. This method supports the assessment of suitability for consumption by identifying potential contaminants that affect taste, corrosion, or health. For cations such as (NH₄⁺) and (K⁺), ion chromatography provides reliable separation and detection in environmental waters at trace levels (typically 0.01–1 mg/L), complementing anion analysis in comprehensive evaluations. In and extracts, ion chromatography is essential for detecting like (NO₃⁻) and orthophosphate (PO₄³⁻), which are critical indicators of risk in aquatic systems. EPA Method 9056A outlines the procedure for these anions in samples, allowing simultaneous determination with detection limits around 0.1–0.5 mg/L, and is adaptable for extracts prepared by aqueous to evaluate runoff impacts. For instance, elevated levels in effluents can be quantified to track loading from agricultural or sources, while analysis in extracts helps assess for uptake or potential. Ion chromatography also facilitates the speciation of disinfection byproducts and in environmental samples. Haloacetic acids, formed during , are analyzed using EPA Method 557, which employs ion chromatography coupled with for direct injection and detection of nine species at limits of 0.02–0.11 μg/L, aiding in the monitoring of carcinogenic byproducts in treated water. Similarly, for speciation, such as (Cr(VI)) versus total chromium, EPA Method 218.7 uses ion chromatography to differentiate toxic forms in and at concentrations below 1 μg/L, supporting remediation efforts in contaminated sites. These applications underscore ion chromatography's versatility in identifying species-specific risks rather than bulk elemental concentrations. Sample handling is crucial for maintaining ion integrity in environmental analyses; on-site filtration through 0.45 μm membranes removes particulates, while preservation differs by analyte—cooling to 4°C for anions and acidification to <2 with nitric acid (HNO₃) for cations to prevent adsorption or precipitation. Regulatory compliance is enforced through limits such as the EPA's maximum contaminant level of 10 mg/L for nitrate (as nitrogen) in drinking water to prevent methemoglobinemia, with ion chromatography providing the validated quantification needed for routine testing. This aligns with standard procedures for water sample preparation, emphasizing rapid analysis within 48 hours for unstable nutrients like nitrate and phosphate.

Pharmaceutical and Clinical Uses

Ion chromatography (IC) plays a crucial role in pharmaceutical quality control and clinical diagnostics by enabling the precise separation and quantification of ionic species in drug formulations and biological samples. In the pharmaceutical sector, IC is employed to assess counterions, excipients, and degradation products, ensuring compliance with regulatory standards such as those outlined in the . Clinically, it facilitates electrolyte analysis in serum to diagnose imbalances associated with various disorders. Counterion analysis using IC is essential for verifying the identity, purity, and assay of active pharmaceutical ingredients (), particularly in salt forms where counterions enhance solubility and stability. For instance, IC determines chloride (Cl⁻) content in hydrochloride salts like metformin hydrochloride, a common antidiabetic drug, by employing suppressed conductivity detection on anion-exchange columns. This method aligns with USP guidelines, including <621> Chromatography, which supports IC for such determinations in monographs like . The technique offers high for both inorganic counterions (e.g., Cl⁻, SO₄²⁻) and ones (e.g., benzenesulfonate in amlodipine besylate), with limits of detection () typically in the low range, minimizing interference from the . In testing, IC quantifies anionic species such as and in tablet formulations to ensure they meet pharmacopeial limits and do not contribute to instability or toxicity. , often present in excipients like , is separated using eluents on high-capacity anion-exchange columns, with detection via suppressed for trace-level accuracy. Similarly, levels in buffering agents are monitored to prevent excessive accumulation that could affect drug release profiles. These analyses support monograph requirements for inorganic impurities, providing specificity and robustness in complex matrices. Clinically, enables electrolyte profiling in for diagnosing ionic imbalances, such as or , which are critical indicators of conditions like , renal , or cardiac issues. The method accurately measures sodium (Na⁺), (K⁺), and (Cl⁻) concentrations, with Cl⁻ reference intervals of 98–106 mmol/L; deviations can signal disorders like . Using cation- or anion-exchange columns with suppressed , IC achieves linearity over physiological ranges (e.g., 0–150 mmol/L for Na⁺) and precision with coefficients of variation (CV) below 1%, serving as a reference method traceable to NIST standards. For stability studies, identifies and quantifies ionic degradation products, such as originating from systems in formulations, to evaluate drug shelf-life and forced under stress conditions like oxidation or . In buffered solutions, levels are monitored post-stability testing using ion-exclusion or anion-exchange modes, detecting increases that may arise from buffer breakdown or API . This stability-indicating approach ensures that does not exceed ICH limits (e.g., <0.5% for specified impurities), with examples including the analysis of in hemodialysis solutions transferred to patients. Validation of IC methods for pharmaceutical quality control emphasizes parameters like , , and limit of quantification (LOQ) to meet USP <621> and ICH Q2(R1) guidelines. Linearity is typically demonstrated over 0.1–150% of the target concentration with correlation coefficients >0.999, while and LOQ for ions like Cl⁻ reach 0.1 µg/mL and 0.37 µg/mL, respectively, ensuring reliable detection of impurities at trace levels. These validations confirm method robustness, accuracy (recovery 97–100%), and precision ( <2%) across diverse formulations. In drug development, IC also supports brief assessments of protein purification, complementing specialized techniques for ionic contaminant removal in biologics.

Industrial Applications

Ion chromatography plays a vital role in industrial settings for monitoring ionic species that impact process efficiency, product quality, and equipment integrity across manufacturing sectors. In power generation, food processing, semiconductor production, hydrometallurgy, and general process control, it enables precise detection of trace ions, ensuring compliance with stringent standards and optimizing operational parameters. In power plants, ion chromatography is essential for boiler water analysis to detect corrosive ions such as sodium (Na⁺) and silicate (SiO₃²⁻), which can lead to turbine deposits and stress corrosion cracking if concentrations exceed parts-per-billion levels. Online ion chromatography systems monitor these ions in feedwater, boiler water, and steam cycles, allowing real-time adjustments to water treatment to maintain ionic purity and prevent costly downtime. For instance, trace sodium levels below 1 ng/L are routinely quantified to safeguard high-pressure components. In the food processing industry, ion chromatography facilitates the monitoring of preservatives like benzoate and sorbate in beverages, ensuring they remain within regulatory limits (e.g., up to 150 ppm for benzoate and 300 ppm for sorbate in soft drinks) to prevent microbial growth while avoiding over-preservation. It also quantifies mineral content, including cations such as calcium, magnesium, sodium, and potassium, and anions like chloride and sulfate, which influence taste, stability, and nutritional labeling in products like beer and fruit juices. These analyses support quality control during production, with methods achieving detection limits in the low ppm range for direct injection of liquid samples. The semiconductor industry relies on ion chromatography for ultrapure water (UPW) testing, where trace ionic contaminants must be below 1 ppb to avoid defects in wafer fabrication and etching processes. Capillary ion chromatography with suppressed conductivity detection separates and quantifies anions like fluoride, chloride, nitrate, and sulfate, as well as cations including sodium, ammonium, and transition metals, in UPW used for rinsing and chemical dilution. This ensures resistivity exceeds 18 MΩ·cm and prevents yield losses from ionic residues. In hydrometallurgy, ion chromatography supports metal ion recovery by profiling impurities in leachates and process streams, enabling optimization of extraction and purification steps for metals like nickel, cobalt, and rare earth elements. It detects trace anions and cations that interfere with solvent extraction or precipitation, such as chloride or sulfate at µg/L levels, allowing for targeted removal to achieve high-purity products (e.g., >99% recovery efficiency). This analytical role complements preparative techniques in and mining operations. For process optimization across industries, inline ion chromatography systems provide monitoring of ionic compositions in streams, facilitating immediate adjustments to maintain quality and reduce waste. Automated analyzers, such as those with electrolytic eluent generation, operate continuously from ng/L to percent levels, integrating with control systems for parameters like and in chemical manufacturing. Standard industrial , like , is often applied upstream to ensure reliable inline performance.

Advances and Limitations

Recent Innovations

In the 2020s, advancements in stationary phases for ion chromatography (IC) have focused on enhancing separation efficiency and speed, particularly through the adoption of core-shell particles. These particles feature a solid core surrounded by a porous shell, reducing diffusion paths and enabling higher flow rates while maintaining low backpressure. Thermo Fisher Scientific's IonPac series, including high-efficiency columns like the AS-HC variants, further exemplify this trend by incorporating smaller particle sizes (e.g., 4 μm) and optimized chemistries for faster gradient separations of anions such as and . Hyphenation techniques, especially the coupling of IC with mass spectrometry (IC-MS), have seen significant progress for ion speciation in environmental samples. Electrospray ionization (ESI) interfaces have been refined to handle high-ionic-strength eluents, enabling sensitive detection of trace species without extensive sample pretreatment. For instance, short-column IC-ESI-MS/MS methods have achieved detection limits of 0.14 μg/L for bromate in drinking water, offering interference-free quantification even in matrices with high chloride levels. This integration enhances specificity for oxyhalides and disinfection byproducts, as demonstrated in applications monitoring water quality. Automated and portable IC systems have emerged to support field-deployable analysis, with capillary IC platforms reducing eluent consumption to microliter-per-minute flows (5–30 μL/min) for on-site monitoring. Devices like the University of Tasmania's portable ion chromatograph, developed in 2025, facilitate real-time anion analysis in remote environments, such as and samples, with battery-powered operation and integrated conductivity detection. Complementing these, ()-driven algorithms have improved data processing in complex chromatograms from , minimizing manual intervention. Efforts toward in IC emphasize through low-eluent systems and recyclable components. Capillary and microbore formats, as in the Thermo Scientific Dionex Inuvion system launched in 2024, minimize reagent use by up to 90% compared to conventional analytical-scale IC, aligning with eco-friendly analytical practices. Electrolytic suppressors with eluent regeneration capabilities, such as those in reagent-free IC-eluent regeneration (RFIC-ER) setups, recycle suppressed eluents by electrolytically converting them back to the original form, reducing waste and operational costs while maintaining suppression efficiency. Multidimensional approaches have gained traction in the for tackling samples in , combining techniques like with orthogonal methods such as or to boost peak capacity. These configurations separate ionic and neutral species sequentially, achieving improved for ions in matrices, as highlighted in recent reviews. These trends, driven by phase design innovations, underscore IC's evolution toward higher throughput and versatility in diverse analytical challenges.

Advantages and Challenges

Ion chromatography offers several key advantages that make it particularly suitable for the of ionic species. It provides high for detecting ions at trace levels, often in the parts-per-billion (ppb) to parts-per-trillion () range, enabling the quantification of low-concentration analytes in environmental and samples. This excels in multi-analyte capability, allowing the simultaneous separation and determination of multiple anions or cations, such as , , and , within a single run, which enhances efficiency for routine monitoring. Additionally, its robustness for aqueous samples stems from the use of dilute, non-organic eluents like acids, bases, or salts, which require minimal and support automated operation with low error rates. Despite these strengths, ion chromatography faces notable challenges that can impact its performance in certain scenarios. Matrix interferences, such as contamination in samples, may reduce sensitivity for specific analytes like by altering peak shapes or baseline noise. Eluent compatibility issues arise when samples contain high levels of incompatible ions or organics, potentially requiring additional pretreatment to prevent degradation of separation efficiency. Column fouling is a common problem in dirty or complex samples, where particulates or hydrophobic compounds accumulate on the stationary phase, leading to increased backpressure and reduced column lifespan. In comparisons to other techniques, ion chromatography demonstrates distinct advantages for certain applications. Relative to (HPLC), ion chromatography is superior for analyzing non-UV-absorbing ions, as it employs detection that does not rely on chromophores, while using simpler, aqueous-based mobile phases without organic solvents. Compared to inductively coupled plasma mass spectrometry (ICP-MS), ion chromatography is more cost-effective and operationally simpler for routine anion analysis, avoiding the need for high-vacuum systems and extensive sample digestion, though it lacks ICP-MS's elemental specificity. The cost structure of ion chromatography systems reflects a balance between accessibility and specialization. Initial setup costs typically range from $20,000 to $50,000, encompassing the chromatograph, suppressor, and columns, which is moderate for analytical but higher than basic spectrophotometers due to components. Running costs remain low, primarily from inexpensive eluents and minimal consumption, making it economical for high-throughput laboratories. Future mitigation of these challenges is being addressed through innovations such as advanced resins with improved selectivity and resistance to , which enhance overall performance and extend column life.

References

  1. [1]
    Ion Chromatography - SERC (Carleton)
    Ion chromatography, a form of liquid chromatography, measures concentrations of ionic species by separating them based on their interaction with a resin. Ionic ...
  2. [2]
    [PDF] Ion Chromatography - IIT Kanpur
    Ion Chromatography (IC) is a liquid chromatography used to separate and analyze ions, especially anions and cations, in aqueous solutions.Missing: definition | Show results with:definition
  3. [3]
    Determination of selected anions in water by ion chromatography
    Ion chromatography is a rapid, sensitive, precise, and accurate method for the determination of major anions in rain water and surface waters.<|control11|><|separator|>
  4. [4]
    [PDF] Monograph – Practical Ion Chromatography - Metrohm
    Ion chromatography (IC) involves basic principles like ion exchange, ion pair formation, and ion exclusion. It is used to extract information from mixtures.
  5. [5]
    Novel ion exchange chromatographic method using conductimetric ...
    Ultra-High-Pressure Ion Chromatography with Suppressed Conductivity Detection at 70 MPa Using Columns Packed with 2.5 μm Anion-Exchange Particles.
  6. [6]
    Ion Chromatography Applications in Wastewater Analysis - MDPI
    Feb 26, 2018 · The majority of applications of ion chromatography in wastewater analysis concern determinations of inorganic anions and cations [42,43].
  7. [7]
    Application of ion chromatography in clinical studies and ... - PubMed
    This paper provides a short review of the ion chromatography applications for determining different inorganic and organic substances in clinical and ...
  8. [8]
    Ion Chromatography - an overview | ScienceDirect Topics
    This chapter reviews the underlying principles of ion chromatography and its application in pharmaceutical analysis. It provides an overview of eluent systems, ...
  9. [9]
  10. [10]
    [PDF] Application of Ion Chromatography to the Investigation of Real ...
    Ion chromatography is a specific implementation of a more general technique, high performance liquid chromatography. Ion chromatography is used extensively in ...
  11. [11]
    [PDF] The Past, Present and Future of Ion Chromatography | UNT Chemistry
    Modern Definition of the Term Ion Chromatography. Ion Chromatography is an umbrella term for all liquid chromatographic techniques that are suitable to ...
  12. [12]
    Ion chromatography for the pharmaceutical industry | Metrohm
    Nov 18, 2024 · Ion chromatography is a type of high-performance liquid chromatography (HPLC) used to separate and quantify ionic and polar analytes ...<|control11|><|separator|>
  13. [13]
    History of ion exchange - François de Dardel
    Oct 1, 2011 · Synthetic ion exchange resins (phenol-formaldehyde polycondensates) invented by Adams & Holmes. 1940, First commercial phenol-formaldehyde ...Missing: early 1940s
  14. [14]
    [PDF] Ion Exchange Resins
    Synthetic ion exchange materials based on coal and phenolic resins were first introduced for industrial use during the 1930's. A few years later resins ...Missing: Adams Holmes
  15. [15]
    David Colin Sherrington. 5 March 1945—4 October 2014
    Sep 13, 2017 · A major step forward came in 1935, when Adams and Holmes described the preparation of some phenol-formaldehyde particles and their use for ion ...
  16. [16]
    [PDF] Ion Chromatography - Semantic Scholar
    Ion Chromatography (IC) was introduced in 1975 by Small, Stevens and Baumann as a new analytical method. Within a short period of time, ion chromatography ...
  17. [17]
    HISTORY OF CHROMATOGRAPHY Hamish Small - PubMed Central
    In 1971, Small and his colleagues William Bauman and Tim Stevens made the breakthrough in ion chromatography, eluent suppression, that revolutionized much of ...Missing: Horvath Yale
  18. [18]
    Innovations of 1975 and the Birth of the First Commercialized ...
    Sep 25, 2025 · Explore the groundbreaking innovations of 1975, including the first commercialized suppressed ion chromatograph, and their lasting impact to ...
  19. [19]
    Landmarks in the Evolution of Ion Chromatography
    (5) H. Small, T.S. Stevens, and W.C. Bauman Anal. Chem. 75, 1801–1809 (1975). (6) H. Small, ...Missing: Baumann | Show results with:Baumann
  20. [20]
    Anion chromatography with low-conductivity eluents - ScienceDirect
    This system uses a low-capacity resin and a dilute aromatic acid salt eluent, allowing detection of anions below 1 ppm without a suppressor column.
  21. [21]
    Discovery and Early Development of Non-Suppressed Ion ... - PubMed
    This is a personal account of the initial discovery and early development of non-suppressed ion chromatography. ... non-suppressed ion chromatography is discussed ...
  22. [22]
    [PDF] Handbook of Ion Chromatography - Thermo Fisher Scientific
    The most important characteristic of an LC–MS interface is the transfer of analyte molecules from the separator column into the high vacuum of a mass.
  23. [23]
    [PDF] Ion Chromatography - Monograph
    Theory of Ion Chromatography. 8. 1.1. Introduction. 8. 1.2. Ion exchange as a separation mechanism. 9. 1.3. Chromatographic characteristics.
  24. [24]
    [PDF] Advanced Detection Techniques in Ion Chromatography
    Ion chromatography is a determination method used in ion analysis that has undergone great com- mercial expansion in the last thirty years and has also ...
  25. [25]
    [PDF] Comparison of Suppressed to Nonsuppressed Conductivity ...
    The Dionex CSRS suppressor improved the ease of use of the IC system, provided low baseline noise, and therefore enhanced detection sensitivity for cations.
  26. [26]
    Determination of common inorganic anions and cations by non ...
    Feb 29, 2008 · Detection limits calculated at S/N = 3 were 0.05–0.58 ppm (mg/l) for the anions and 0.05–0.38 ppm for the cations, whereas repeatability values ...
  27. [27]
    [PDF] Ion Exchange Resins Fundamentals of Ion Exchange - DuPont
    The majority are prepared in spherical (bead) form, either as conventional resin with a polydispersed particle size distribution from about 0.3 mm to 1.2 mm (50 ...
  28. [28]
    None
    ### Summary of Weak and Strong Ion Exchangers, Resin Materials, Capacity, Preparation
  29. [29]
    Comparison of chromatographic ion-exchange resins V. Strong and ...
    Strong and weak cation-exchangers were compared for a number of chromatographic parameters, ie pH dependence, efficiency, binding strength, particle size ...
  30. [30]
    Preparation and characterisation of anion exchangers with ...
    Poly(styrene-divinylbenzene) (PS-DVB) is a common matrix for the preparation of the anion exchangers due to its chemical stability in the broad pH range (0–14), ...
  31. [31]
    The Donnan potential revealed - PMC - NIH
    Oct 6, 2022 · The Donnan potential acts to exclude co-ions from entering the membrane, where larger absolute Donnan potential values lead to stronger co-ion ...
  32. [32]
    [PDF] A Practical Guide to Ion Chromatography - Nest Group
    Foreword. This guide will show you some of the most com- mon practical issues that occur during analysis of anions by ion chromatography. Many users find.
  33. [33]
    [PDF] Method 9056A: Determination of Inorganic Anions by Ion ...
    1.2. The lower limit of quantitation (LLOQ), the lowest concentration level that can be measured within stated accuracy limits, varies for each individual ...
  34. [34]
    Standard Methods: 4110B: Anions in Water by Ion Chromatography
    A water sample is injected into a stream of eluent and passed through a series of ion exchangers. The anions of interest are separated on the basis of their ...
  35. [35]
    Standard operating procedure for the analysis of major ions in ... - NIH
    Jun 14, 2023 · This standard operating procedure (SOP) describes an ion chromatography (IC) procedure for the major cations and anions in hydrothermal fluids.
  36. [36]
    Practical and Popular Sample Preparation Techniques for Ion ...
    The most commonly used sample preparation technique for IC other than filtration is matrix elimination, with and without off-line or in-line concentration. In ...
  37. [37]
    Ion exchange membranes in ion chromatography and related ...
    Nov 1, 2019 · Ion chromatography made its commercial debut with a carbonate/bicarbonate eluent. It is still one of the most popular eluents used. Except ...Eluent Cleanup · Eluent Generation Strategies... · Donnan Dialysis Separation...
  38. [38]
    [PDF] Protein Separation with Ion-exchange Membrane Chromatography
    membrane chromatography has a lower pressure drop, higher flow rate and higher ... shape membranes including flat sheet, hollow fiber and radial flow (Figure 1.4) ...
  39. [39]
    [PDF] Macroporous ion-exchange membrane adsorbers - DuEPublico
    Membrane adsorber devices are available as stacked flat sheet in flow through or radial flow type, pleated as well as hollow fibers devices (4). Page 19. 2.
  40. [40]
    [PDF] Considerations on the flow configuration of membrane ...
    All runs were operated at a flow rate between 1-10 ml min-1. In some experiments, external loading of the tangential flow membrane adsorber was performed. This.
  41. [41]
    Bioseparation using membrane chromatography: Innovations, and ...
    Mar 15, 2025 · The main advantage of membrane chromatography continues to be its convection dominant transport mechanism, the resultant benefit being fast and scalable ...
  42. [42]
    Membrane-based sample preparation for ion chromatography ...
    Jan 6, 2017 · The respective features of the various techniques are outlined with respect to the possibilities of matrix removal and selectivity enhancement.Missing: elimination | Show results with:elimination
  43. [43]
    Elimination of matrix interferences in ion chromatography by the use ...
    Topics discussed include solid-phase extraction for sample clean-up and preconcentration, dialytic methods, combustion methods, matrix-elimination IC, ...
  44. [44]
    [PDF] Applied
    Mar 21, 2005 · A major limitation of membranes is its relatively low resolution ... Cation exchange membrane chromatography has been used to separate complex ...
  45. [45]
    [PDF] Guide to Ion-Exchange Chromatography - Harvard Apparatus
    Ion exchange chromatography is the reversible adsorption of charged molecules to immobilized ion groups on a matrix of an opposite charge. Separation can be.
  46. [46]
    [PDF] Ion Exchange Chromatography & Chromatofocusing
    gives a brief overview of this approach which is recommended for any multi-step protein purification. The Protein Purification Handbook from GE Healthcare ...
  47. [47]
    Developments in Ion Exchange Chromatography–Mass ...
    Sep 21, 2025 · Based on the chemical properties of the ion exchangers, AEX and CEX can be further classified into weak and strong exchangers. Weak cation ...
  48. [48]
    Gibbs-Donnan Equilibria in Ion Exchange Resin Systems
    Ion Exclusion Chromatography: Model Development and Experimental Evaluation. ... A Gibbs-Donnan-Based Analysis of Ion-Exchange and Related Phenomena. 2017 ...
  49. [49]
    Modeling of salt and pH gradient elution in ion-exchange ... - PubMed
    The separation of proteins by internally and externally generated pH gradients in chromatofocusing on ion-exchange columns is a well-established analytical ...
  50. [50]
    [PDF] Exploration of pH-Gradient Ion-Exchange Chromatography for High ...
    The work shown here describes the application of pH-gradient IEC as compared to salt-gradient IEC for the separation of proteins from various sources. High ...
  51. [51]
  52. [52]
    Ion Exchange Chromatography Resins | Thermo Fisher Scientific - US
    Ion exchange chromatography (IEX) is a purification technique that allows for the separation of molecules based on their surface charge.
  53. [53]
    Ion-Exchange Chromatography of Proteins
    Ion-exchange chromatography is the most widely used technique in protein chromatography (1). This is because it is nearly always possible to develop successful ...
  54. [54]
    [PDF] Tips for successful ion exchange chromatography - Cytiva
    May 17, 2017 · It is important to use a low flow rate during cleaning (about half of the flow rate used for separations). Also, rinse the column with water ...
  55. [55]
    Purification of Monoclonal Antibodies Using Chromatographic ... - NIH
    The separation process includes three steps: enzymatic enrichment to remove impurities, intermediate purification to achieve 40%-90% purity, and polishing to ...
  56. [56]
    Protein Purification by Ion Exchange Chromatography | Sino Biological
    Ion exchange chromatography (IEX) uses protein charge for separation, is widely used in biopurification, and involves binding, washing, and eluting.
  57. [57]
    [PDF] EPA Method 300.1, Revision 1.0: Determination of Inorganic Anions ...
    These different injection volumes are required in order to compensate for the relative concentrations of these anions in drinking water and maintain good ...
  58. [58]
    Cation Analysis | Thermo Fisher Scientific - US
    Cation analysis include analysis of alkaline metals cations, ammonium cations, and amines in municipal drinking water, wastewater, or industrial wastewater.
  59. [59]
    Simultaneous Determination of Nitrate, Chloride, Sulfate, and ...
    Apr 1, 1983 · Ion chromatography (IC) is used to determine NO3-, Cl-, SO42-, and PO43- in water in about 10 minutes, detecting as little as 0.1 mg/L.
  60. [60]
    Method 557: Determination of Haloacetic Acids Bromate ... - epa nepis
    Method 557 is a direct-inject, ion chromatography, negative-ion electrospray ionization, tandem mass spectrometry (IC-ESI-MS/MS) method for the determination ...
  61. [61]
    National Primary Drinking Water Regulations | US EPA
    National Primary Drinking Water Regulations ; Nitrate (measured as Nitrogen), 10, 10 ; Nitrite (measured as Nitrogen), 1, 1 ; Selenium, 0.05, 0.05 ; Thallium ...Ground Water Rule · Surface Water Treatment Rules · Chemical Contaminant Rules
  62. [62]
    Ion chromatography as candidate reference method for the ... - NIH
    Jul 24, 2020 · Ion chromatography is a simple sample treatment procedure for the determination of chloride in human serum with high sensitivity and specificity.
  63. [63]
  64. [64]
  65. [65]
    Ion chromatography testing for pharmaceuticals and materials
    Oct 9, 2025 · It is used to quantify inorganic ions, such as sodium, potassium, halides, phosphate, and sulfate, in pharmaceutical products to ensure ...Missing: tablets | Show results with:tablets<|control11|><|separator|>
  66. [66]
    Analyzing Pharmaceuticals by Ion Chromatography - US
    Analysis of Pharmaceuticals by Ion Chromatography. Drug characterization and quantitation, and purity, counterion, and excipient analyses.
  67. [67]
    Ion chromatography applications | Metrohm
    Ion chromatography (IC), or ion exchange chromatography is a powerful analytical technique used to separate and quantify ions in a sample.
  68. [68]
    An improved reference method for serum cations measurement by ...
    Ion chromatography with a simple sample treatment procedure for the determination of cations in human serum with high sensitivity and specificity was developed.
  69. [69]
  70. [70]
    Ion Chromatography: The All-Rounder in Pharmaceutical Analysis
    Apr 29, 2019 · According to USP 621, the US Pharmacopeia recommends ion chromatographic azide determination after direct injection. In this method, a ...Missing: counterion | Show results with:counterion
  71. [71]
    Ion Chromatography: Beyond Water Analysis | Lab Manager
    Both are liquid chromatography techniques, but they differ in their stationary and mobile phases, and the types of molecules they analyze. Ion chromatography ...
  72. [72]
    Simple and Sensitive Stability-Indicating Ion Chromatography ... - NIH
    This research paper describes the stability-indicating nature of the method by performing the forced degradation studies to drug substances as an active ...
  73. [73]
    <621> Chromatography - US Pharmacopeia (USP)
    The new harmonized standard for <621> Chromatography has been approved by the Pharmacopeial Discussion Group (PDG) as described in its PDG Stage 4 sign-off ...Missing: ion counterion APIs
  74. [74]
  75. [75]
    [PDF] Ion Chromatography – Power Plant Water Applications Summary ...
    Ion chromatography is used in power plants to maintain ionic purity, identify contaminants, and determine trace anions, with low detection limits.
  76. [76]
    2060 IC Process Analyzer: online monitoring of ionic compounds
    The 2060 IC Process Analyzer is a complete and flexible system for online monitoring of ionic compounds in aqueous media from ng/L to % concentrations.
  77. [77]
    Use of on-line ion chromatography in controlling water quality in ...
    The determination of low ng/l sodium in the power industry is critical in identifying and preventing corrosive conditions in many power plant components. To ...
  78. [78]
    [PDF] Determination of Silicate in High-Purity Water Using Ion ...
    To develop an online sample preparation and IC method that accurately determines silicate in. UHPW. Determination of silicate in high-purity water using ion ...
  79. [79]
    Online monitoring of sodium in industrial power plants | Metrohm
    Sodium monitoring is commonly used to detect salt contaminants in the water demineralization (DM) plant (Figure 1). In power plants, ion exchange demineralizers ...
  80. [80]
    [PDF] Determination of Benzoate in Liquid Food Products by Reagent ...
    In this application note, we describe a simple ion chromatography method for the direct determination of benzoate in liquid food products. This method.Missing: sorbate | Show results with:sorbate<|separator|>
  81. [81]
  82. [82]
    Anions and cations in beer | Metrohm
    This Application Note presents ion chromatography as a precise method to analyze anions in beer as well as cations with non-suppressed conductivity.Missing: mineral | Show results with:mineral
  83. [83]
    [PDF] Determination of Trace Anions in Ultrapure Water Using Capillary ...
    Trace anion determinations in ultrapure water (UPW) are important for the power and electronics industries. Ionic contamination at µg/L and ng/L concentrations ...
  84. [84]
    [PDF] Semiconductor workflows—Trace contaminant analysis application ...
    Ion chromatography with suppressed conductivity detection is the preferred method for determining ionic species, especially at µg/L and ng/L concentrations.
  85. [85]
    The recent progress of ion exchange for the separation of rare ...
    Hydrometallurgy (HM) has been used as the principal route for recovering REEs (Jha et al., 2016), and the conventional procedure consists of three steps: ...
  86. [86]
    Application of ion-exchange fibers towards the chromatographic ...
    Jul 15, 2024 · This study aimed to separate and recover nickel (Ni), cobalt (Co), cerium (Ce), lanthanum (La), and neodymium (Nd) from waste NiMH batteries using ion-exchange ...
  87. [87]
    Unmatched flexibility in online ion analysis: The 2060 IC Process ...
    Mar 8, 2021 · One of the most effective ways to bridge laboratory analyses to the process environment is to employ real-time analysis monitoring. Product and ...
  88. [88]
    Synthesis and chromatographic evaluating of the core-shell SiO2 ...
    May 10, 2025 · The development of core-shell materials for achieving more efficient and rapid chromatographic separations has remained a key objective in ...
  89. [89]
    Ion Chromatography Columns | Thermo Fisher Scientific - US
    Our shorter 150 mm columns allow you to perform analyses at higher flow rates to increase productivity without sacrificing performance. Learn more. Fast ...Missing: 2020s | Show results with:2020s
  90. [90]
    Rapid bromate determination using short-column ion ...
    A short-column IC-ESI-MS/MS method was developed for bromate analyse. ... Adding ACN to samples improves the sensitivity and anti-interference ability. ... The ...
  91. [91]
    [PDF] An introduction to ion chromatography mass spectrometry (IC-MS)
    IC-MS is a robust, sensitive, and selective technique used for the determination of polar contaminants like inorganic anions, organic acids, haloacetic acids, ...
  92. [92]
    Portable ion chromatograph brings real-world practice to classrooms
    Sep 17, 2025 · University of Tasmania scientists develop portable ion chromatograph for field use, enhancing student learning and real-world applications.Missing: 2020s | Show results with:2020s
  93. [93]
    Artificial Intelligence On Artificial Intelligence In Separation Science
    Jun 3, 2024 · Overall, the use of AI can significantly improve peak deconvolution and peak assignment in HRMS/MS and HRIM–MS, allowing for more accurate ...Missing: portable | Show results with:portable
  94. [94]
    PeakBot: machine-learning-based chromatographic peak picking
    A machine-learning-based approach entitled PeakBot was developed for detecting chromatographic peaks in LC-HRMS profile-mode data.
  95. [95]
    Ion Chromatography Through the Ages with Alexandra Zatirakha
    Aug 29, 2025 · The launch of Thermo Scientific's Dionex Inuvion IC system in February 2024 marks another milestone in IC evolution. What are some key features ...
  96. [96]
    IC and RFIC Suppressors FAQs | Thermo Fisher Scientific - US
    They generate their own regenerant inside the suppressor with water, either in recycled eluent mode or external water mode. In recycled eluent mode, the water ...
  97. [97]
    Reagent-Free Ion Chromatography Systems With Eluent Regeneration
    These systems use the electrolytic suppressor to regenerate the starting eluent after suppression. They provide benefits similar to RFIC-EG systems, offering ...Missing: green | Show results with:green
  98. [98]
    Multi‐dimensional technology – Recent advances and applications ...
    Mar 12, 2024 · This review provides an overview of the latest advancements and applications in multi-dimensional liquid chromatography coupled with mass ...
  99. [99]
    Comprehensive analytical approaches: The use of LC-MS and IC ...
    This review article provides a comprehensive and evaluative overview of liquid chromatography-mass spectrometry (LC-MS) and ion chromatography-mass ...
  100. [100]
    Ion Chromatography: An Overview and Recent Developments
    With zwitterionic columns and chromatographic conditions, the retention is due to a salt-exchange process simultaneously involving both the anion and cation.
  101. [101]
    Ion Chromatography Advantages v Disadvantages
    Aug 16, 2023 · IC can detect ions in trace amounts, often at parts-per-billion (ppb) or even parts-per-trillion (ppt) levels, rendering it highly sensitive.
  102. [102]
    When HPLC fails: unleashing the potential of ion chromatography
    Key benefits of Metrohm ion chromatography · Hardly any influences of the sample matrix on the separation · Metal-free system · Organic solvents are not required ...Missing: advantages | Show results with:advantages
  103. [103]
    Assessing ICP-MS vs IC for Ionic Element Quantification in Water
    Sep 19, 2025 · IC systems generally require less capital investment and maintenance than ICP-MS, making them more accessible to smaller laboratories. IC's ...
  104. [104]
    Shop Ion Chromatography Instruments For Sale, New and Used Prices
    Free delivery over $1M 30-day returnsPrices for new ion chromatography systems generally range from $20,000 to $100,000, depending on the features, configuration, and capabilities of ...Missing: initial | Show results with:initial