Fact-checked by Grok 2 weeks ago

Peak expiratory flow

Peak expiratory flow (PEF), also known as peak expiratory flow rate (PEFR), is the maximum speed at which air can be forcefully expelled from the lungs during a vigorous exhalation, typically measured in liters per minute (L/min) using a portable handheld device called a peak flow meter. This measurement provides a simple, non-invasive assessment of airway patency and lung function, primarily used to monitor and manage obstructive respiratory conditions such as asthma. It reflects the degree of airflow limitation in the larger airways and serves as an early warning indicator for exacerbations before symptoms become severe. PEF measurement involves taking a deep breath and then exhaling as forcefully and rapidly as possible into the peak flow meter, with the highest of three consecutive attempts recorded as the reading. Normal values vary by age, sex, height, and ethnicity, but are individualized based on a patient's "personal best"—the highest reading obtained over 2–4 weeks during stable health, often ranging from 400–700 L/min in adults and 150–450 L/min in children. Readings are interpreted using color-coded zones relative to the personal best: (80–100%, good ), (50–80%, caution and potential need for intervention), and red (below 50%, indicating a ). Proper technique is essential, as inconsistent effort or device calibration can lead to inaccurate results. In clinical practice, regular PEF monitoring is recommended for individuals with moderate to severe , enabling self-management through personalized action plans that guide adjustments, trigger avoidance, and timely medical consultation. It is particularly valuable in settings where full is unavailable, offering quick insights into asthma control and response to therapy, though it is less reliable in young children due to challenges with . Peak flow meters adhere to standards set by organizations like the to ensure consistency across devices.

Definition and Function

Definition

Peak expiratory flow (PEF) is defined as the maximum achieved at the during a forced maneuver, starting from full inflation. This metric represents the highest speed at which air can be expelled from the s under maximal effort. It is conventionally expressed in units of liters per minute (L/min). The primary purpose of PEF is to provide a rapid evaluation of airflow limitation within the airways, serving as a key indicator of lung function. In contrast to forced expiratory volume (FEV), which measures the volume of air exhaled over a defined time period such as one second, PEF focuses exclusively on the peak velocity of expiration. This distinction allows PEF to offer a simple, immediate insight into potential airway narrowing.

Physiological Role

Peak expiratory flow (PEF) represents the maximum rate achieved during the initial of a forced expiratory , starting from total capacity, and serves as a key indicator of large airway patency. This effort-dependent measure relies on the coordinated of expiratory muscles to generate high pleural pressures, propelling air through the central airways at the highest possible velocity before flow limitation occurs. In healthy individuals, PEF typically reflects the wave speed mechanism, where flow is constrained by the physical properties of the airways rather than solely by muscular effort once maximal is reached. The physiological underpinnings of PEF are closely tied to the interplay between elastic recoil and . pressure, which is maximal at full inflation, provides the primary driving force for expiration, while low upstream minimizes frictional losses and allows rapid acceleration of airflow. Airway wall compliance and cross-sectional area at the flow-limiting "" further modulate PEF by influencing the onset of dynamic in the larger bronchi. Disruptions in these , such as increased from narrowed airways, directly attenuate peak flow rates. Reductions in PEF are a sensitive signal of or obstruction, as they indicate compromised airway caliber that elevates resistance and diminishes the effectiveness of elastic recoil in generating high initial flows. In conditions involving airway narrowing, such as exacerbations, PEF drops progressively with the degree of obstruction, often falling below 80% of predicted values before more severe spirometric changes become evident. This makes PEF a valuable proxy for detecting early alterations in large airway dynamics, though it is less sensitive to small airway involvement.

Measurement Techniques

Procedure

The measurement of peak expiratory flow (PEF) follows a standardized protocol to ensure reliable and reproducible results, typically performed using a handheld peak flow meter. The patient should stand upright or sit straight with the back supported to optimize expansion and expiratory force. Proper patient training is essential, as suboptimal technique can lead to underestimation of PEF; clinicians often demonstrate the and supervise initial attempts to correct errors such as incomplete lip sealing or hesitant . To perform the measurement, first resets the meter's indicator to zero or the bottom of the scale. They then take a maximal inspiration, filling the lungs completely while keeping the nose unobstructed (no nose clips are typically required). The mouthpiece is placed firmly between the teeth, with sealed tightly around it to prevent air leaks, ensuring the does not obstruct the path. The patient exhales as forcefully and rapidly as possible in a single blast, using abdominal and to achieve maximal effort within the first 1-2 seconds of expiration. This maneuver is repeated in quick succession, with short rests between attempts to avoid , and the highest of the three values is recorded as the PEF reading. Coughing, incomplete efforts, or air leaks invalidate a blow, necessitating repetition. For immediate clinical interpretation during , PEF values are often classified into color-coded zones based on the of the patient's personal best or predicted value: the represents 80-100% (indicating good control), the yellow zone 50-80% (signaling caution and potential need for ), and the below 50% (indicating severe obstruction requiring urgent action). These zones guide self-management in conditions like , but accuracy depends on consistent technique and regular calibration of the device.

Devices and Scales

Peak expiratory flow (PEF) is measured using portable peak flow meters, which are handheld devices designed to quantify the maximum speed of air expulsion during a forced . The original device, known as 's peak flow meter, was a mechanical instrument invented by Dr. B.M. in the as a large, circular, clock-faced apparatus that used a spring-loaded vane to indicate flow rates up to 1000 L/min. This mechanical design relied on physical components for measurement without electronic components, making it simple and cost-effective but prone to variability in calibration. Subsequent developments led to miniaturized mechanical versions, such as the Mini-Wright peak flow meter introduced in the , which retained the vane-based mechanism but offered improved portability in a compact, format measuring up to 880 L/min for adults or 400 L/min for low-range models suited to children and smaller adults. Digital or electronic peak flow meters emerged later, incorporating sensors and microprocessors to provide precise readings, often with additional features like , Bluetooth connectivity for transmission to healthcare providers, and the ability to measure other parameters such as forced vital capacity (FVC) in some handheld spirometers. These electronic devices typically cost more than mechanical ones but enhance accuracy and user convenience through automated calculations and reduced operator error. PEF measurement scales have evolved to address inaccuracies in early calibrations, with the original scale (also called Wright-McKerrow) serving as the foundational standard but over-reading flows in the mid-range (300-500 L/min) by up to 80 L/min due to limitations in testing methods at the time. The scale, standardized under EN 13826:2003, was introduced in 2004 to ensure greater accuracy and consistency across devices, specifying requirements for peak expiratory flow meters (PEFMs) used in spontaneous breathing assessments, including flow ranges from 50-800 L/min and compliance testing with computerized pumps for precise waveform simulation. This standard mandates for legal sale in the EU and has influenced global manufacturing, replacing non-compliant Wright-scale devices. The NHANES III reference equations, derived from a large U.S. of over 7,400 asymptomatic nonsmokers aged 8-80 years, provide a standardized framework for interpreting PEF values across ethnic groups (, African-American, Mexican-American), emphasizing predicted norms based on age, height, and sex rather than device-specific scaling. Conversion between the and scales is necessary for historical data comparability, with correction formulas developed using on simulated flows; for Mini- meters, the equation is PEF_{corrected} = 0.00090 \times (PEF_{recorded})^2 + 0.373 \times PEF_{recorded} + 47.4, yielding a deviation of 7 L/min when aligned to EN 13826 . For larger meters, a similar adjustment applies: PEF_{corrected} = 0.00075 \times (PEF_{recorded})^2 + 0.585 \times PEF_{recorded} + 53.2 ( deviation: 6 L/min). These conversions account for the original scale's systematic overestimation, enabling alignment with modern -calibrated devices. NHANES III values can be integrated by applying these corrections to -based predictions before comparison to reference equations. Calibration of peak flow meters is critical for reliability, with mechanical devices like the Mini-Wright individually assembled and tested against master standards to meet ISO 23747 or EN 13826 criteria, ensuring accuracy within ±10% across the measurable range through or pump simulations of expiratory flows. Digital meters undergo electronic sensor validation, often self-calibrating via built-in diagnostics, though no universal method exists to directly calibrate between different brands due to inherent design variations. Modern devices prioritize portability, weighing under 200 grams and fitting in a pocket, with disposable mouthpieces for and battery-powered operation in lasting months, facilitating frequent home monitoring without compromising .

Reference Values and Interpretation

Normal Ranges

Peak expiratory flow (PEF) reference values are established to represent normal lung function in healthy individuals, derived from large-scale population studies such as the , which analyzed data from over 7,000 asymptomatic, lifelong nonsmokers aged 8 to 80 years across Caucasian, African-American, and Mexican-American ethnic groups. These predicted values account for key demographic factors including age, sex, height, and ethnicity, as PEF naturally varies with growth in youth, peaks in early adulthood, and declines gradually with age, while taller stature and male sex generally yield higher values. The use of ethnicity-specific adjustments in these equations has become controversial, with recent guidelines from the American Thoracic Society recommending race-neutral approaches to avoid potential underdiagnosis in minority populations. For adults, representative normal PEF ranges typically fall between 550 and 700 L/min for and 350 to 500 L/min for , though these can adjust downward by 10-15% for African-American or Mexican-American individuals compared to of similar age and . For instance, a 30-year-old of 175 cm might have a predicted PEF of approximately 630 L/min, while a 40-year-old of 165 cm could expect around 430 L/min, illustrating the combined influence of these factors. Personalized predictions are often obtained using nomograms, standardized charts, or online calculators that implement NHANES III-derived equations, allowing clinicians to input patient-specific details for accurate assessment. These tools, endorsed by organizations like the American Thoracic Society, facilitate quick determination of an individual's normal range, with typical diurnal variability up to 20% considered within healthy limits.

Factors Influencing Values

Peak expiratory flow (PEF) values are influenced by several demographic factors that contribute to variations from predicted norms. plays a significant role, with PEF typically peaking in the 20-30 years age group at approximately 430 L/min in healthy women, followed by a progressive decline thereafter. This decline accelerates after age 40, averaging about 3.8 L/min per year in men and 2.0 L/min per year in women, reflecting age-related reductions in elasticity and muscle strength. Sex differences also affect PEF, with males generally exhibiting higher values than females—such as mean PEF of 367 L/min in men versus 253 L/min in women—primarily due to larger airway diameters in males. correlates positively with PEF, increasing by roughly 1.5 L/min per centimeter in men and 1.1 L/min per centimeter in women, as taller stature is associated with greater volume and airway size. Environmental and behavioral factors further modify PEF readings. , for instance, reduces PEF in a dose-dependent manner, with smokers showing values around 410 L/min compared to 513 L/min in non-smokers, and the effect worsening with age and pack-years due to airway inflammation and obstruction. Altitude influences PEF through changes in air density; at high altitudes (e.g., 2150 m), values are significantly higher—such as 265 L/min in boys versus 245 L/min at —because lower air density decreases , facilitating greater expiratory flow. Temperature and impact meter accuracy rather than physiological PEF directly; lower temperatures (e.g., 7°C) can reduce uncorrected meter readings by up to 5-10%, while high has minimal effect, necessitating corrections for environmental conditions during measurement. Diurnal variations and short-term influences like recent exercise also cause fluctuations in PEF. PEF exhibits a , with values lowest in the morning (e.g., around 8.7 L/s at 8:00 AM) and highest in the afternoon or evening (e.g., 8.9 L/s at 5:00 PM), resulting in about 6% variability in healthy individuals due to natural cycles in airway tone and levels.

Clinical Applications

Asthma Management

Peak expiratory flow (PEF) monitoring plays a central role in management by enabling patients to perform daily home assessments, which facilitate early detection of exacerbations before severe symptoms emerge. Patients are advised to establish a personal best PEF value through consistent measurements over 2-4 weeks during stable periods, using the same device and recording the highest of three consecutive readings each time. This personal best serves as the reference point for ongoing monitoring, typically conducted twice daily (), allowing individuals to track subtle declines in lung function that may signal impending attacks. Integration of PEF into written action plans, as recommended by the Global Initiative for Asthma (GINA), structures management around color-coded zones based on percentages of the personal best: the (80-100%) indicates good control with maintenance therapy; the yellow zone (50-79%) prompts increased reliever medication or step-up in controller therapy; and the red zone (<50%) requires immediate intervention, such as additional short-acting beta-agonists or oral corticosteroids. These plans empower patients to adjust treatments proactively, reducing emergency visits and improving control, particularly for those with moderate to severe persistent . GINA guidelines emphasize providing all adults and adolescents with such personalized plans to guide self-management. Evidence from clinical studies demonstrates that PEF variability—calculated as the difference between morning and evening readings or over several days—effectively predicts exacerbations, with variability exceeding 20% often preceding attacks by days. For instance, longitudinal of home PEF has shown that increasing variability correlates with heightened , enabling timely interventions. In the , advancements in digital tracking have enhanced this utility, with smartphone-connected PEF meters and apps allowing logging, trend , and alerts for variability thresholds, as validated in recent systematic reviews of digital biomarkers for control.00060-1/fulltext)

Other Respiratory Conditions

Peak expiratory flow (PEF) monitoring plays a role in managing (COPD) by providing a simple, home-based measure to track disease progression and assess responses to . In COPD patients, serial PEF measurements can detect daily variations in function, helping identify and predict before they become severe. For instance, declines in PEF have been associated with increased exacerbation risk, allowing for timely interventions like adjusting therapy. Studies show that PEF can reliably identify moderate to severe obstruction in COPD, with a peak flow below 80% predicted serving as a sensitive indicator, though it is less specific than . Additionally, post- improvements in PEF, often exceeding 10-15% of baseline, help evaluate treatment efficacy in reducing expiratory flow limitation. Beyond COPD, PEF is applied in and allergic reactions through serial monitoring to detect work-related or allergen-induced airway variability. In , at least two weeks of twice-daily PEF recordings, compared between work and non-work periods, aids diagnosis by revealing patterns of diurnal and occupational decline, with sensitivity around 72% when interpreted against objective standards. This approach is particularly useful in field settings for workers exposed to irritants, as recommended by occupational health guidelines emphasizing portable PEF meters for serial assessment. For allergic reactions, such as those in or acute exposures, PEF variability correlates with airway hyperresponsiveness, enabling early detection of in atopic individuals. In , PEF serves as a screening and monitoring tool to assess pulmonary involvement, particularly in resource-limited settings where full is unavailable. Portable PEF meters allow for regular evaluation of expiratory flow, helping track disease severity and response to therapies like mucolytics or antibiotics. Guidelines from respiratory societies, including the 2022 ATS/ERS standards on pulmonary function test interpretation, endorse PEF for field assessments in various respiratory disorders due to its simplicity and accessibility, though it should complement rather than replace comprehensive testing. This utility extends to diverse conditions by providing objective data on airflow limitation in outpatient or community-based evaluations.

Limitations and Comparisons

Potential Errors and Variability

Peak expiratory flow (PEF) measurements are susceptible to user errors, primarily stemming from poor , insufficient effort, or inconsistent timing of assessments. Improper or , such as failing to take a full breath or not providing maximal effort, can result in significantly lower readings than actual values. Additionally, accelerating using the or cheeks instead of relying solely on lung force leads to inaccurate results. Inconsistent timing exacerbates variability, as PEF exhibits natural diurnal swings, with variations up to 20% observed in individuals with due to circadian rhythms in airway caliber. Device-related issues further contribute to measurement inaccuracy and inconsistency. Peak flow meters may experience calibration drift over time, leading to systematic errors in readings, although some models demonstrate stability after repeated use. Turbine-type meters are particularly sensitive to low temperatures, where readings can decrease due to changes in gas and response, potentially underestimating true PEF in cold environments. Regular validation against standards is essential to mitigate these effects. Population-specific factors limit the reliability of PEF interpretations when reference norms are not adjusted appropriately. In obese individuals, excess body mass restricts chest wall , resulting in systematically lower PEF values compared to non-obese peers, which can confound assessments if standard predictions are applied. Among the elderly, age-related declines in muscle strength and lung elasticity often yield diminished PEF, increasing the risk of misclassification for respiratory impairment. Similarly, ethnic differences necessitate tailored reference values; for instance, equations derived from populations overestimate PEF by 12-15% in individuals of descent, leading to potential underdiagnosis of limitation.

Relation to Other Lung Function Tests

Peak expiratory flow (PEF) measurement offers a quick and portable alternative to , particularly for monitoring large airway function in conditions like , as it requires minimal equipment and can be performed by patients without specialized training. However, PEF is less sensitive to small airway dysfunction compared to spirometry parameters, which better detect early or peripheral airflow limitations. The correlation between PEF and forced expiratory volume in one second (FEV1), a key spirometry metric, typically ranges from 0.7 to 0.8 in clinical studies, indicating moderate agreement but highlighting PEF's limitations as a standalone proxy. While PEF integrates well with comprehensive pulmonary function tests (PFTs) for ongoing assessment, it cannot substitute for the FEV1/forced (FVC) ratio, which is essential for diagnosing obstructive diseases and grading severity. Full PFTs, including , provide a broader evaluation of and flow rates, making them indispensable for initial diagnosis, whereas PEF excels in serial home monitoring to track variability and response to . Recent reviews from 2024 have explored combining PEF with mobile applications for remote monitoring, showing improved control and adherence compared to traditional in-clinic , though these digital approaches complement rather than replace formal testing. For instance, app-integrated PEF tracking enables real-time data sharing with clinicians, potentially reducing exacerbations, but validation against remains crucial for accuracy.

History and Development

Early Development

The measurement of peak expiratory flow (PEF) originated in the mid-1950s through the work of British bioengineer Basil Martin Wright at the in , . Wright designed the first dedicated peak flow meter in 1956 to provide a practical alternative to cumbersome existing devices for assessing lung function. This instrument, a lightweight and portable tool, allowed for quick, repeatable measurements of maximum forced expiratory flow rate during a single breath. The invention was motivated by the need for an accessible method to monitor respiratory conditions, particularly as the use of inhalers rose following the introduction of the first pressurized in 1956. At the time, traditional equipment was bulky, required trained operators, and fatigued patients, limiting its utility for routine clinical or home-based assessment of airflow obstruction in . Wright's device addressed this by enabling patients to self-measure PEF easily, facilitating better evaluation of treatment responses amid growing reliance on aerosolized therapies. The first peak flow meter was publicly introduced in 1959, detailed in a seminal publication co-authored with C.B. McKerrow. Early clinical validation involved testing the meter on healthy subjects and patients with respiratory impairment, comparing PEF readings to forced and demonstrating its sensitivity to effects; for instance, in asthmatic individuals, PEF increased significantly after inhalation of adrenaline , confirming its value in detecting reversible airway obstruction. These initial trials established PEF as a reliable, objective index of ventilatory capacity, paving the way for its adoption in respiratory .

Modern Advancements

In the 1970s, the introduction of the Mini-Wright peak flow meter marked a significant advancement in portability and accessibility for peak expiratory flow (PEF) measurement. Developed as a compact, lightweight version of the original meter, it enabled by patients at home and in clinical settings, particularly benefiting pediatric and geriatric populations with its ease of use and lower cost. A key standardization effort occurred in 2004 with the adoption of the (EU) scale under the EN 13826 standard, which replaced the older scale to enhance measurement accuracy and consistency across devices. This non-linear scale, calibrated for better alignment with actual airflow dynamics, became mandatory for CE-marked meters sold in , facilitating global comparability and reducing variability in PEF readings between instruments. The saw the rise of digital integration in PEF monitoring through smart meters equipped with connectivity, allowing real-time data transmission to applications for enhanced tracking and clinician oversight. Devices such as the MIR Smart One spirometer, introduced around 2015, combined PEF measurement with app-based logging to support remote management by alerting users to declines in function. Similarly, platforms like the Smart Peak Flow system, emerging in the late , integrated -enabled meters with user-friendly apps to visualize trends and predict exacerbations. Advancements in during the have further refined PEF applications, with studies developing predictive models to forecast airflow variations in patients. For instance, hybrid approaches combining and algorithms have demonstrated improved accuracy in predicting daily PEF rates by analyzing historical data and environmental factors, aiding proactive intervention. These AI-driven tools, often integrated into digital platforms, enhance personalization in respiratory care. Recent guidelines have emphasized PEF's role in , promoting its use for remote amid the expansion of services. The 2023 and 2024 Global Initiative for Asthma (GINA) reports highlight PEF variability as a diagnostic and management tool in virtual consultations, recommending integrated apps for tracking to reduce exacerbations through timely adjustments in therapy. Complementing this, the American Academy of Allergy, & Immunology (AAAAI) guidelines on remote endorse Bluetooth-enabled PEF devices to enable continuous , improving outcomes in chronic respiratory conditions via platforms.

References

  1. [1]
    Peak Flow Rate Measurement - StatPearls - NCBI Bookshelf - NIH
    Oct 6, 2024 · Peak expiratory flow (PEF), a key indicator of lung function, offers clinicians valuable insights into asthma management.
  2. [2]
    Peak Flow Meter: How To Use, Measurement & Results
    A peak flow meter measures how fast you can exhale. It helps manage asthma. Use your meter at least once a day or as directed by a healthcare provider.Missing: reliable sources
  3. [3]
    Measuring Your Peak Flow Rate | American Lung Association
    Oct 23, 2024 · A peak flow meter is a portable, inexpensive, hand-held device used to measure how air flows from your lungs in one fast blast.Missing: reliable sources
  4. [4]
    physiological determinants of peak expiratory flow - PubMed
    Peak expiratory flow (PEF) can be defined as the maximum flow at the mouth achieved during an expiration, delivered with maximum force starting from the ...
  5. [5]
    Peak expiratory flow measurement - NIH Common Data Elements ...
    NINDS; Definition: Peak expiratory flow (PEF) value measured in liters/minute. PEF is the maximal expiratory flow rate achieved during the forced vital capacity ...
  6. [6]
    Blood and the respiratory system: 3.1 Spirometry | OpenLearn
    Peak expiratory flow (PEF) measures the maximum speed at which air is forcefully expired (litres per second). The forced expiratory volume ... and PEF ...
  7. [7]
    Peak Expiratory Flow Rate Measurement Periprocedural Care
    Jan 29, 2024 · The PEFR is conventionally measured with the patient standing, though an upright seated position is also suitable. The mouthpiece of the PFM is ...<|control11|><|separator|>
  8. [8]
    How to use your peak flow meter: MedlinePlus Medical Encyclopedia
    Mar 2, 2024 · A peak flow meter is a small device that helps you check how well your asthma is controlled. Peak flow meters are most helpful if you have moderate to severe ...Missing: definition sources
  9. [9]
    [PDF] Peak Flow Meters | Mercury Medical
    Education Program Technical Standards for Peak. Flow Meters, based on American Thoracic Society. Spirometry Waveform 24. Performance Data. Calibration Data.
  10. [10]
    2 Independent Studies Prove Mini-Wright Accuracy - Flexicare
    Apr 8, 2024 · Mini-Wright is the only mechanical Peak Flow Meter that meets all required accuracy standards.1,2. Mini-Wright peak flow meters (PFMs) are the ' ...
  11. [11]
    Peak Flow Meters: Our Top 5 Picks - Healthline
    Apr 27, 2023 · Both digital and mechanical peak flow meters can be purchased for under $100. Digital meters tend to be more expensive. For this roundup, we ...What it's used for · Types · Our picks · How to use
  12. [12]
    [PDF] Peak expiratory flow meter scale changes
    Figure 2 shows that if the mean within-day PEF is around. 200 l/min the variability is recorded reasonably faithfully. However, when below this level the ...
  13. [13]
  14. [14]
    Spirometric Reference Values from a Sample of the General U.S. ...
    Dec 22, 1997 · This analysis of the NHANES III spirometry data was conducted to develop reference equations to describe normal pulmonary function for three ...
  15. [15]
    Estimated/Expected Peak Expiratory Flow (Peak Flow) - MDCalc
    The Estimated/Expected Peak Expiratory Flow (Peak Flow) quantifies asthma exacerbation severity.
  16. [16]
    Variations in the Peak Expiratory Flow Rate with Various Factors in a ...
    It is well known that age, height and weight are the main factors which affect the Peak Expiratory Flow Rate (PEFR), the Forced Expiratory Volume in the first ...Missing: sex | Show results with:sex
  17. [17]
    Reference Values and Related Factors for Peak Expiratory Flow in ...
    Aug 20, 2021 · This study aimed to determine age- and sex-specific reference values of PEF in the middle-aged and elderly Chinese population.
  18. [18]
    Effect of Cigarette and Cigar Smoking on Peak Expiratory Flow Rate
    Sep 10, 2013 · Our present findings suggest that both cigarette and cigar smoking have their deleterious effects on lung function causing reduction in PEFR with advancing age ...
  19. [19]
    Peak Expiratory Flow Rate of Healthy School Children Living at High ...
    The mean peak expiratory flow rate value of boys at high altitude (265 ± 92.6 L/min) was significantly higher than those in plain areas (245 ± 82.1 L/min).
  20. [20]
    Portable peak flow meters: physical characteristics, influence of ...
    Little is known about the linearity of portable peak flow meters, or about physical gas factors affecting peak expiratory flow (PEF) readings.
  21. [21]
    Diurnal Variation in Peak Expiratory Flow and Forced ... - NIH
    Oct 1, 2015 · Diurnal variability in airways has most commonly been reported using peak expiratory flow (PEF) and forced expiratory volume in one second (FEV1).
  22. [22]
    Pulmonary functions and anthropometric parameters of young male ...
    It is concluded that moderate aerobic exercise improved respiratory functions (FEV1, FVC and PEFR) in both male and female subjects with greater improvement in ...
  23. [23]
    [PDF] Summary Guide for Asthma Management and Prevention
    1. Measure PEF morning and night for 1–2 weeks using same PEF meter (each time, measure PEF 3 times and record only the highest). 2. Calculate daily ...
  24. [24]
    Create an Asthma Action Plan | American Lung Association
    Sep 5, 2025 · Your peak flow rate can show if your asthma is getting worse, even before you feel symptoms. Your peak flow rate is measured with a peak flow ...Missing: expiratory | Show results with:expiratory
  25. [25]
    [PDF] CDC Asthma Action Plan
    GO to the hospital or call 911 NOW! If you use a peak flow meter you can use these scores to determine your current zone: Your best score.
  26. [26]
    [PDF] Peak expiratory flow rate monitoring can predict asthma ...
    Aug 21, 2024 · Global initiative for asthma (GINA) guidelines suggest to use PEFR monitoring to assess response to treatment, to.
  27. [27]
    Digital markers of asthma exacerbations: a systematic review - PMC
    Dec 16, 2024 · Digital predictors are markers of asthma attacks which can be identified using digital technologies and may help reduce morbidity from asthma ...<|control11|><|separator|>
  28. [28]
    Peak expiratory flow rate and COPD exacerbation
    Peak expiratory flow rate (PEFR) monitoring could provide a daily objective measurement of lung function in COPD patients at home.
  29. [29]
    Monitoring peak expiratory flow could predict COPD exacerbations
    Peak expiratory flow (PEF) can be safely used instead of spirometry which can assess the severity of COPD as a standard tool. We hypothesized that monitoring ...<|control11|><|separator|>
  30. [30]
    Effectiveness of adding a peak flow meter for the identification ... - NIH
    May 17, 2025 · Peak expiratory flow (PEF) measurement is useful for detecting moderate and severe chronic obstructive pulmonary disease (COPD).
  31. [31]
    Effect of bronchodilation on expiratory flow limitation and resting ...
    Bronchodilator drugs improve lung emptying, and this leads to variable increases in forced expiratory volume in 1 s (FEV1), mainly by reducing lung volume ...
  32. [32]
    Full article: The Volumetric Response to Bronchodilators in Stable ...
    Jul 2, 2009 · Peak expiratory flow (PEF) ... Volume effect and exertional dyspnoea after bronchodilator in patients with COPD with and without expiratory flow ...<|control11|><|separator|>
  33. [33]
    Peak expiratory flow rates in possible occupational asthma - PubMed
    With the objective diagnoses as the gold standard, the sensitivity of the PEFR interpretations was 72 percent for OA; specificity for no asthma was 53 percent.
  34. [34]
    Diagnosing Work-related Asthma - CDC
    Feb 25, 2025 · Serial measurements of peak expiratory flow might be useful for evaluating work-related asthma. Provide instructions to the patient on how to ...
  35. [35]
    [PDF] A Systematic Review of Serial Peak Expiratory Flow Measurements ...
    This paper systematically reviews literature on the application of serial peak expiratory flow (PEF) measurements in the diagnosis of occupational asthma and ...
  36. [36]
    Variability of peak expiratory flow rate in allergic rhinitis and mild ...
    Background: Previous studies have demonstrated a significant correlation between airway sensitivity to pharmacologic agents and daily peak expiratory flow ...
  37. [37]
    Screening Asthma among Allergic Rhinitis Patients with Peak ...
    RATIONALE: We aimed to study atopic status, prevalence of asthma and the benefit of using peak expiratory flow (PEF) follow-up in the diagnosis of asthma at ...<|control11|><|separator|>
  38. [38]
    [Peak expiratory flow rate in the management of cystic fibrosis]
    To verify the role of peak expiratory flow, measured through portable meters, as an alternative test to assess pulmonary disease in cystic fibrosis.Missing: screening | Show results with:screening
  39. [39]
    ERS/ATS technical standard on interpretive strategies for routine ...
    In 2005, the American Thoracic Society (ATS) and European Respiratory Society (ERS) jointly adopted technical standards for the interpretation of PFTs. We aimed ...Bronchodilator... · Classification of physiological... · The future of pulmonary...
  40. [40]
    Effects of equipment and technique on peak flow measurements
    Jun 20, 2006 · This difference (up to 19%) was in line with a previous laboratory study which showed "error" rates up to 26% in some devices [1].
  41. [41]
    Incorrect use of peak flow meters: are you observing your patients?
    Examples of errors included failure to inhale fully or give maximum effort on exhalation, accelerating air with the tongue and buccal musculature, and ...Missing: poor | Show results with:poor
  42. [42]
    Relation of peak expiratory flow rate to body mass index in young ...
    This study concluded that PEFR declines with increase in BMI, and there is negative correlation between BMI and PEFR.Missing: reliability ethnic
  43. [43]
    PEAK EXPIRATORY FLOW AS A PREDICTOR OF SUBSEQUENT ...
    In our elderly cohort, we found that a diminished PEF, when expressed as an SR-percentile, is independently associated with subsequent disability and death.
  44. [44]
    Reference Values for Peak Expiratory Flow Rate in Adults of African ...
    It has also been noted that applying prediction formulae derived from a Caucasian population overestimated the PEF values in Black Africans by 12%–15%.
  45. [45]
    Peak flow meter with a questionnaire and mini-spirometer to ... - NIH
    May 9, 2017 · Compared to spirometry, peak flow measurements are less-time consuming, are not dependent on trained manpower, easy for patients to perform and ...Missing: advantages | Show results with:advantages
  46. [46]
    Can peak expiratory flow measurements estimate small airway ...
    Background: Asthma is characterized in part by small airways dysfunction. Peak expiratory flow (PEF) measurement has been suggested by all international ...
  47. [47]
    Peak expiratory flow is a reliably household pulmonary function ...
    Mar 19, 2022 · Peak expiratory flow (PEF) is another useful pulmonary function index, which was defined as the maximum expiratory flow per minute.Missing: units | Show results with:units<|control11|><|separator|>
  48. [48]
    Peak expiratory flow rate as a surrogate for forced expiratory volume ...
    This study reveals a marked variability in agreement between the measurements of % predicted FEV1 and % predicted PEFR values in COPD patients. The results ...
  49. [49]
    Asthma management in the digital age - PMC - PubMed Central
    Sep 3, 2024 · Digital tools for monitoring treatment focus on tracking medication use, inhalation technique, and physiological markers such as peak-flow rate ...
  50. [50]
    Maximum Forced Expiratory Flow Rate as a Measure of Ventilatory ...
    Wright B M, McKerrow C B. Maximum Forced Expiratory Flow Rate as a Measure of Ventilatory Capacity Br Med J 1959; 2 :1041 doi:10.1136/bmj.2.5159.1041. BibTeX ...
  51. [51]
    Martin Wright | | The Guardian
    Mar 22, 2001 · Seven million peak flow meters have been distributed since production began. In 1957, recognising his unique skills, the MRC employed Wright to ...Missing: Fleming | Show results with:Fleming
  52. [52]
    The History of Therapeutic Aerosols: A Chronological Review - PMC
    In 1956, Riker Laboratories, Inc., (now 3 M Drug Delivery Systems) introduced the first pressurized metered dose inhaler (MDI).
  53. [53]
    Wright Peak Flowmeter - Wood Library-Museum of Anesthesiology
    Dr. Wright introduced a miniature version in the 1970s and in time it came to replace the original device.Missing: mechanical | Show results with:mechanical
  54. [54]
    Martin Wright invented pocket peak-flow meter - The Irish Times
    Mar 31, 2001 · The device's invention in 1956 was followed by publication of results in 1959, and peak flow rapidly became a standard measure of respiratory ...
  55. [55]
    Maximum forced expiratory flow rate as a measure of ... - PubMed
    Maximum forced expiratory flow rate as a measure of ventilatory capacity: with a description of a new portable instrument for measuring it. Br Med J. 1959 ...
  56. [56]
    Mini Wright Gold Standard Peak Flow • Clement Clarke - Flexicare
    Mini-Wright was introduced in the 1970's to enable the benefits of peak flow monitoring to be accessed in doctor's surgeries and at patient's homes all around ...Missing: date | Show results with:date
  57. [57]
    What Meter Do I Have? - Flexicare
    The original Wright peak flow meter was a large, circular clock-shaped device and was very heavy. Mini-Wright peak flow meters can be recognised from their grey ...Missing: mechanical | Show results with:mechanical
  58. [58]
    Smart One: App-Based Personal Spirometer (PEF, FEV1)
    Brand-new personal device for accurate heart check and arrhythmias detection. Start listening to your heart!
  59. [59]
    Asthma Control In Your Pocket - Smart Peak Flow
    Rating 4.8 (126) · 60-day returnsTrack your results easily in the app, share your results with your GP or asthma nurse, see how the AI app predicts your condition for tomorrow or suggests you ...
  60. [60]
    A hybrid approach for forecasting peak expiratory flow rate in ... - NIH
    Aug 21, 2025 · In this work, we propose a novel hybrid approach using machine learning models and similarity measurement technique with the aim of precise peak ...
  61. [61]
    [PDF] GINA Pocket Guide 2023 - Global Initiative for Asthma
    For example, a ≥20% improvement in PEF 15 minutes after giving 2 puffs of albuterol (salbutamol), or improved symptoms and PEF after a 4-week therapeutic trial ...
  62. [62]
    Remote Monitoring
    All the current inhalers utilize Bluetooth technology to transmit information to a patient APP or to secure provider portals. The attached digital inhalers have ...Missing: meters | Show results with:meters