Selected area diffraction

Selected area (SAED) is a crystallographic performed in (TEM) that enables the collection of patterns from a localized of a thin specimen, typically 100 nm to several micrometers in diameter, to analyze and orientation. This method relies on the of a parallel beam by the , producing patterns of spots or rings that reveal parameters, composition, and defects in materials. Pioneered in the by Soviet scientist B. K. Vainshtein using early elektronograf instruments with 50–100 keV electrons, SAED evolved alongside TEM development to facilitate qualitative of both organic and inorganic substances. By the 1980s, it played a pivotal role in landmark discoveries, such as Dan Shechtman's identification of quasicrystals in rapidly solidified Al-Mn alloys through icosahedral patterns, and L. Bendersky's of decagonal phases with 10-fold rotation axes. These advancements highlighted SAED's capacity for high-resolution studies down to 100 areas, supported by modern achieving atomic-scale resolutions like 1.65 Å at 300 kV. In practice, SAED involves inserting a selected-area aperture into the objective lens's to restrict electrons to the target region, then switching to mode by focusing the imaging lens on the back focal plane, where the pattern is observed and interpreted via and Ewald sphere constructions in . The technique requires electron-transparent samples to minimize multiple scattering effects, which can introduce errors like double , and is best suited for parallel illumination to ensure accurate reciprocal lattice projections. SAED finds broad applications in for phase identification, lattice matching at interfaces, defect analysis, twinning, and structural characterization of such as nanorods or metallic alloys. It excels in studying complex systems like quasicrystals and intergrowths, providing essential data for , , and , though limitations include and the need for precise sample positioning, historically restricting selectable areas to 1–5 μm.

Introduction

Definition and purpose

Selected area diffraction (SAD), also referred to as selected area electron diffraction (SAED), is a crystallographic technique utilized in (TEM) to examine a localized region of a thin sample, typically spanning 100–1000 in diameter, by recording the pattern generated from that area. This method employs a selected area aperture to isolate electrons scattered from the chosen region under parallel illumination, producing a pattern that maps the and provides detailed insights into the atomic arrangement. The primary purpose of SAD is to deliver high-resolution crystallographic information at the nanoscale, enabling the determination of structures, orientations, and defects in materials that are too small or heterogeneous for techniques such as . Unlike methods, which require larger sample volumes and may overlook local variations, SAD probes very small sample volumes—typically corresponding to areas of 100–1000 in and specimen thicknesses of 20–100 —with exceptional sensitivity to structural nuances like ordering and modulations due to the strong interaction of electrons with matter. In and , SAD is essential for phase identification and precise measurement of parameters, facilitating the characterization of complex microstructures. For instance, it has been pivotal in identifying unknown phases in alloys, such as the icosahedral in aluminum-manganese systems, where traditional diffraction methods failed to detect the aperiodic order.

Historical background

The discovery of in 1927 by and Germer at Bell Laboratories demonstrated the wave nature of through the observation of patterns from a nickel crystal, confirming Louis de Broglie's hypothesis and laying the groundwork for electron-based crystallographic techniques. This breakthrough, published in , marked the first experimental verification of electron waves interacting with crystalline matter in a manner analogous to . The invention of the transmission electron microscope (TEM) in 1931 by Ernst Ruska and Max Knoll at the Technical University of Berlin enabled both imaging and diffraction modes by focusing electron beams with magnetic lenses, achieving resolutions beyond light microscopy. Ruska's subsequent refinements in the early 1930s, including the first practical TEM prototype in 1933, facilitated the integration of diffraction capabilities, though initial instruments lacked precise area selection. Selected area diffraction (SAD) emerged in the and as TEM apertures were refined to isolate from specific sample regions, pioneered in the by Soviet scientist B.K. Vainshtein using early elektronograf instruments. Early implementations appeared in instruments like the Elmiskop I introduced in 1953, which featured a double condenser for routine studies. Pioneering work by Robert Heidenreich in applied TEM to thin metal foils, enabling the first widespread use of in post-World War II for analyzing microstructures and compositions. By the 1960s, SAD had become a standard tool for materials analysis, allowing identification and orientation mapping in polycrystalline samples through aperture-limited beam selection typically down to 1-5 μm. In the , advancements in via detectors and software for indexing revolutionized SAD, with tools like the package enabling automated simulation and analysis of patterns for complex structures. These developments, including early applications at the National Bureau of Standards for discovery, enhanced accuracy in construction and reduced reliance on manual interpretation, while building on foundational principles like for spot positioning.

Theoretical Foundations

Electron diffraction principles

Electron diffraction arises from the wave-particle duality of electrons, a fundamental concept introduced by in 1924, which posits that particles possess wave-like properties characterized by a de Broglie given by \lambda = [h](/page/H+) / [p](/page/P′′), where [h](/page/H+) is Planck's and [p](/page/P′′) is the electron's momentum. For accelerated electrons in typical (TEM) setups, this is on the order of picometers; at an accelerating voltage of 200 kV, \lambda \approx 2.5 pm, making it comparable to interatomic distances in crystalline materials and enabling phenomena. This wave nature was experimentally confirmed through the observation of patterns by Davisson and Germer in 1927, demonstrating that electrons scattered from a produce interference maxima analogous to . The primary mechanism underlying is , where incident electrons interact with the periodic potential of the crystal lattice without energy loss, resulting in coherent scattering that produces patterns. In this process, the electrons' wavefunctions superimpose constructively or destructively depending on the phase differences introduced by the lattice spacing, leading to discrete diffraction spots or rings that encode information about the . dominates in thin specimens under high vacuum conditions, as inelastic events (e.g., or excitations) are minimized, preserving the necessary for sharp patterns. To interpret diffraction conditions, the Ewald sphere construction provides a geometric framework in reciprocal space, representing the incident electron wavevector as the radius of a sphere centered at the origin of the . occurs when reciprocal lattice points lie on the surface of this sphere, corresponding to momentum transfer vectors that satisfy the for constructive interference; for electrons, the sphere's large radius (due to short \lambda) approximates a plane, simplifying the analysis of accessible reflections in TEM. In TEM, parallel illumination of the sample ensures that the diffracted beams converge to form the interference pattern in the back focal plane of the objective , where the lens acts as a Fourier to spatially separate the zero-order transmitted beam from the diffracted orders. Selected area diffraction (SAD) applies these principles to isolate and analyze from specific regions of the specimen.

Bragg's law and reciprocal lattice

Bragg's law describes the condition for constructive in from a crystal lattice, given by the equation n\lambda = 2d \sin\theta, where n is an representing the order, \lambda is the of the incident , d is the interplanar spacing of the diffracting crystal planes, and \theta is the angle between the incident beam and the planes. This law arises from the path length difference between waves scattered from adjacent atomic planes in the crystal. Consider two parallel crystal planes separated by distance d; an incident wave at angle \theta to the planes strikes the first plane at point A and the second at point B. The scattered waves from A and B travel to a distant detector, and the extra length for the wave from B is $2d \sin\theta. For constructive , this path difference must equal an multiple of the , yielding $2d \sin\theta = n\lambda. In selected area diffraction (SAD) using electrons, the de Broglie \lambda of the electrons determines the applicable scale for these interferences. The reciprocal lattice provides a geometric framework for understanding diffraction patterns in crystals, where each point in the reciprocal lattice corresponds to a set of diffracting planes in the direct (real-space) lattice. The reciprocal lattice vectors \mathbf{g}_{hkl} are defined perpendicular to the (hkl) planes, with magnitude |\mathbf{g}_{hkl}| = 1 / d_{hkl}, linking the spacing d_{hkl} directly to the diffraction condition via Bragg's law, as |\mathbf{g}| = \frac{2 \sin\theta}{\lambda}, where \hat{\mathbf{s}} is the scattering direction. In SAD patterns, the observed diffraction spots represent intersections of the Ewald sphere with these reciprocal lattice points; a spot appears when a reciprocal lattice point lies on the sphere of radius $1/\lambda centered at the origin in reciprocal space. This construction ensures that spots arise only for planes satisfying Bragg's condition, transforming the three-dimensional crystal structure into a two-dimensional pattern of points whose positions encode the lattice geometry. Indexing SAD patterns involves assigning (hkl) to diffraction spots by analyzing their geometric arrangement relative to the pattern center and systematic row symmetries. The denote the reciprocals of the intercepts of the plane with the crystal axes, scaled to integers; for a spot, (hkl) is determined by measuring angles and distances to identify the and plane normals consistent with the crystal symmetry. Higher-order spots (n > 1) align along the same radial direction as the fundamental (hkl) spot but at distances scaled by n. The radial distance from the pattern center to a diffraction spot is proportional to $1/d_{hkl}, reflecting the inverse relationship between spot position and interplanar spacing in the . In the typical for , this distance r_{hkl} satisfies r_{hkl} = L \lambda / d_{hkl}, where L is the effective camera length, allowing direct inference of parameters from measured spot separations.

Experimental Methods

Instrumentation in TEM

Selected area diffraction (SAD) in (TEM) relies on specialized instrumentation to generate and capture patterns from specific regions of a thin sample. The core setup involves an source, illumination , imaging lenses, and detection systems configured to project the diffraction pattern from the back focal plane of the objective lens. The serves as the primary source, producing a beam of accelerated electrons typically at energies of 100-300 keV, with wavelengths on the order of picometers suitable for atomic-scale . Common types include thermionic emitters such as (LaB6) cathodes, which provide stable, high-current beams for routine SAD experiments, and field emission guns (FEGs), including Schottky or cold field emitters, which offer higher brightness and coherence for enhanced pattern resolution in advanced setups. The beam is then conditioned by condenser lenses, which focus it into a near-parallel illumination to minimize and ensure uniform illumination across the selected area, typically achieving angles of less than 1 mrad for sharp diffraction spots. The objective lens plays a central role by forming the diffraction pattern in its back focal plane, where diffracted electrons from the sample converge based on their scattering angles. To enable SAD, a selected area aperture is inserted into the conjugate image plane of the objective lens, typically with diameters ranging from 10 to 50 μm, allowing diffraction data to be collected from a localized specimen region of approximately 0.5-2 μm in diameter depending on the magnification. Switching to diffraction mode involves adjusting the intermediate lens to project this back focal plane onto the detection system, effectively magnifying and focusing the pattern while defocusing the real-space image. Detection of the diffraction pattern is achieved through various systems positioned at the end of the column. Traditional phosphor screens provide real-time visual observation by converting impacts into visible light, while modern charge-coupled device () cameras capture digital images with sufficient sensitivity for low-dose conditions. Increasingly, direct detectors, such as hybrid pixel arrays, are employed for their superior , enabling noise-free recording of weak diffraction signals at high frame rates. Calibration of the camera length L, the effective distance from the sample to the detection plane, is essential for and follows the relation r = L \lambda / d_{hkl}, where r is the radial distance of a diffraction spot from the central , \lambda is the electron wavelength, and d_{hkl} is the interplanar spacing. This setup requires thin samples on the order of 100 nm to ensure sufficient transmission and minimize multiple scattering effects.

Sample preparation techniques

Sample preparation for selected area diffraction (SAD) in (TEM) requires creating electron-transparent specimens, typically thinner than 100 nm, to minimize multiple scattering and ensure clear diffraction patterns. This thickness constraint arises because electrons interact strongly with matter, necessitating thin regions for sufficient while preserving structural integrity. For metallic and inorganic materials, is a widely adopted technique, involving anodic dissolution in an to achieve uniform thinning to transparency. milling, often using ions at low keV energies, serves as a complementary or final step for precise removal of material, particularly after initial mechanical dimpling to create a concave surface. An example for metals includes dimpling to approximately 20-30 μm followed by Ar to reach <100 nm thickness. Organic and biological samples, such as polymers or tissues, are commonly prepared via ultramicrotomy, where a diamond knife sections embedded materials into thin slices of 50-100 nm. For site-specific analysis, focused ion beam (FIB) milling employs a gallium ion beam to extract and thin targeted regions with nanometer precision, ideal for heterogeneous materials or cross-sections. Prepared specimens are mounted on 3 mm diameter copper grids overlaid with holey carbon support films, which provide mechanical stability and reduce background scattering in diffraction experiments. For crystalline materials, such as 2D protein crystals in structural biology, negative staining with heavy metals like uranyl acetate enhances contrast and flatness on carbon films. Nanoparticles are typically dispersed from suspension onto these grids, allowing random orientation for ring pattern analysis in polycrystalline ensembles. The selected area aperture in the TEM then isolates the region of interest for diffraction.

Diffraction Pattern Generation

Procedure for acquiring patterns

The procedure for acquiring selected area diffraction (SAD) patterns in transmission electron microscopy (TEM) begins with sample insertion and initial alignment in imaging mode. The sample, typically a thin specimen mounted on a grid, is loaded into the TEM column under vacuum conditions, and the microscope is operated at an accelerating voltage between 100 and 300 kV to balance resolution and sample stability. The electron beam is spread to a low-intensity mode using the brightness control to locate and center an area of interest, often at magnifications of 5,000–20,000×, ensuring the region is eucentric (aligned along the column axis) via fine adjustments with the stage tilt and translation controls. Next, the selected area is isolated using the objective or selected area (SA) aperture, which is inserted into the intermediate image plane to define the diffracting region, typically 0.5–5 μm in diameter depending on the desired spatial resolution. The aperture is centered over the target area by adjusting the beam shift or image wobble, with the objective aperture removed if necessary to avoid extraneous scattering. Magnification is then optimized (e.g., 10,000–50,000×) to precisely position the aperture while visualizing the sample features. To generate the diffraction pattern, the TEM is switched from imaging to diffraction mode, which projects the back focal plane of the objective lens onto the viewing screen or detector. The beam is recentered using deflection and stigmator controls to ensure the direct beam aligns with the optical axis, and parallel illumination is established by defocusing the condenser lenses (e.g., reducing beam convergence angle to near zero) to prevent spot blurring from angular spread. The camera length is adjusted via the magnification/camera length knob to scale the pattern appropriately for analysis, and the pattern is focused using the diffraction focus control until spots are sharp. If needed for specific orientations, the sample is tilted using the goniometer stage to align a zone axis parallel to the beam, often guided by preliminary imaging or known crystal symmetry to maximize diffraction spot intensity. For pattern capture, the viewing screen is lowered, and an exposure time of 0.1–1 s is set on the digital camera to record the pattern while minimizing beam damage, particularly for beam-sensitive materials like organics or beam-labile nanomaterials, where intermittent blanking or low-dose modes are employed. The intensity is fine-tuned by adjusting beam brightness and focus to avoid saturation of the central beam while capturing weaker reflections.

Factors affecting pattern quality

The quality of selected area diffraction (SAD) patterns in transmission electron microscopy (TEM) is highly sensitive to sample thickness, as thicker specimens introduce dynamical scattering effects that deviate from the kinematic approximation. In samples thinner than approximately 100 nm, electron scattering is predominantly single and elastic, producing sharp, well-defined diffraction spots with intensities proportional to the structure factor squared. However, when thickness exceeds 100 nm, multiple scattering events occur, leading to dynamical interactions that redistribute electron intensities, weaken higher-order reflections, and distort the pattern symmetry. Beam convergence also degrades pattern clarity in SAD, where a parallel incident beam is ideally required to generate discrete spots. Any convergence angle greater than about 10^{-3} radians transforms the spots into overlapping disks, as seen in convergent beam electron diffraction (CBED) modes, reducing resolution and complicating spot identification due to the spread of diffracted intensities. Contamination and beam-induced damage further compromise pattern reliability, particularly in organic or beam-sensitive materials. Hydrocarbon buildup from ambient residues forms a carbon layer under the electron beam, causing amorphous halos that obscure spots, while radiolysis in organics breaks bonds and alters lattice structure, fading reflections over time. Cryo-TEM at liquid nitrogen temperatures mitigates these by immobilizing contaminants and slowing damage kinetics. Objective lens astigmatism and defocus introduce aberrations that blur diffraction patterns, with astigmatism causing direction-dependent focus shifts that elongate or distort spots anisotropically. These effects are exacerbated at higher magnifications or defocus values, leading to reduced contrast and inaccurate intensity measurements. In thicker samples where dynamical scattering dominates, Kikuchi lines emerge as a diagnostic feature, arising from inelastic scattering and Bragg diffraction of thermally diffuse electrons, providing a qualitative indicator of specimen thickness through their sharpness and density.

Pattern Interpretation

Spot patterns in single crystals

In selected area electron diffraction (SAED) of single crystals, the resulting patterns consist of discrete, sharp spots arranged in a symmetrical array that directly reflects the underlying crystal lattice symmetry. For instance, close-packed structures such as hexagonal close-packed (HCP) metals exhibit hexagonal spot arrangements corresponding to their lattice geometry. These spots represent projections of reciprocal lattice points onto the Ewald sphere, with the central spot denoting the undiffracted beam (000 reflection). Analysis of these spot patterns involves measuring the distances and inter-spot angles from the center to identify the zone axis [uvw] and index the reflections (hkl). Distances r from the center to a spot are used to calculate the interplanar spacing d_{hkl} via the relation d_{hkl} = \frac{\lambda L}{r}, where \lambda is the electron wavelength and L is the effective camera length (sample-to-screen distance). Indexing proceeds by comparing measured angles and distance ratios to known values for the crystal structure, often applying the Weiss zone law hu + kv + lw = 0 for reflections in the zero-order Laue zone. Software tools such as ProcessDiffraction facilitate this by automating the determination of consistent zone axes across multiple patterns from the same crystal. Certain spots may be absent, known as forbidden reflections or systematic absences, arising from space group symmetries such as screw axes or glide planes that cause destructive interference. These absences provide key information for confirming the space group, as they follow specific reflection conditions (e.g., h00 absent if h is odd for certain centering). A representative example is the SAED pattern from a face-centered cubic (FCC) metal like copper along the zone axis, displaying prominent {111}-type spots arranged in a rectangular array with 90° angles between them, while mixed-parity (hkl) reflections are systematically absent due to the FCC structure factor.

Ring patterns in polycrystals

In (SAED) of polycrystalline samples, the diffraction pattern consists of concentric rings formed by the superposition of spot patterns from numerous small crystallites with random orientations within the selected area. These rings arise because the statistical distribution of grain orientations averages out the discrete spots into continuous circles centered on the transmitted beam spot. The radius of each ring is inversely proportional to the interplanar spacing d_{hkl}, with the relationship given by r_{hkl} = \lambda L / d_{hkl}, where r_{hkl} is the ring radius, \lambda is the electron wavelength, and L is the camera length. Analysis of ring patterns involves measuring the radii of the rings and converting them to d-spacings using the calibrated camera constant \lambda L, which allows determination of average lattice parameters and phase identification analogous to the Debye-Scherrer method in X-ray powder diffraction. The rings are indexed by comparing measured d-spacings to known values from crystal structure databases, providing information on the constituent phases without resolving individual grain orientations. The intensity distribution along each ring is governed by the structure factor F_{hkl}, where the diffracted intensity I_{hkl} \propto |F_{hkl}|^2, modulated by factors such as atomic scattering factors and multiplicity of reflections. In samples with preferred crystallographic orientations (texture), the uniform rings may exhibit intensity variations or deform into arcs, reflecting the non-random distribution of grain orientations and enabling assessment of texture strength. For example, in nanocrystalline thin films such as , sharp rings indicate well-defined lattice spacings from nanoscale grains, while amorphous materials produce diffuse halos due to short-range order lacking discrete reflections.

Applications

Materials characterization

Selected area diffraction (SAD) is widely employed in transmission electron microscopy (TEM) for phase identification in inorganic materials, enabling the distinction of polymorphs in ceramics and alloys by analyzing diffraction spot patterns corresponding to specific crystal structures. For instance, in steel alloys, SAD patterns differentiate austenite (face-centered cubic) from martensite (body-centered tetragonal) based on interplanar spacing and symmetry, which is crucial for understanding heat-treated microstructures. In ceramics, such as oxide phases in advanced composites, SAD confirms the presence of distinct polymorphs like cubic versus tetragonal zirconia by indexing the diffraction rings or spots to known lattice parameters. SAD also facilitates defect studies in engineering materials, where deviations from ideal spot patterns reveal crystallographic imperfections. Stacking faults appear as streaked or shifted diffraction spots due to local disruptions in atomic planes, while dislocations produce extra reflections or broadening in patterns, allowing quantification of defect density in alloys and ceramics. These features are particularly useful for analyzing deformation mechanisms in metals, such as dislocation interactions in plastically deformed titanium alloys. In nanomaterials, SAD enables orientation mapping of structures like nanowires and thin films by correlating local diffraction patterns with crystal axes, revealing preferred growth directions and texture. For example, in silicon nanowires, SAD patterns confirm single-crystalline nature and <111> orientation along the growth axis, aiding in device optimization. Similarly, in thin films such as epitaxial layers, SAD identifies epitaxial relationships and misorientations across nanoscale regions. A landmark application of SAD occurred in the 1980s discovery of quasicrystals, where patterns from rapidly solidified Al-Mn alloys exhibited icosahedral symmetry with five-fold rotational symmetry, challenging traditional crystallographic rules. This observation, made by in 1982 and published in 1984, used SAD to reveal sharp, forbidden diffraction spots indicative of quasi-periodic order without . SAD is often integrated with () in TEM for comprehensive materials characterization, combining structural data from diffraction with chemical composition from spectra. This correlation identifies phases in multiphase alloys or ceramics, such as confirming inclusions in corroded metals by matching SAD patterns to EDS elemental maps.

Structural biology

Selected area diffraction (SAD) plays a pivotal role in by enabling the analysis of beam-sensitive biological macromolecules, particularly through electron crystallography of two-dimensional () crystals. This technique has been instrumental in elucidating the structures of proteins that readily form arrays but resist three-dimensional () crystallization suitable for methods. A landmark application was the determination of the bacteriorhodopsin structure in the purple of in 1975, where SAD patterns from unstained, tilted specimens yielded a 7 resolution map, revealing the protein's seven transmembrane helices and marking the first visualization of a protein's secondary structure via electron microscopy. In modern applications, SAD has evolved into variants like microcrystal electron diffraction (MicroED), which extends the method to 3D nanocrystals of proteins, achieving atomic resolutions below 2 . For instance, MicroED has resolved structures of amyloid-forming peptides and protein fragments, such as a 1.4 model of the α-synuclein NACore from nanocrystals as small as 200 , allowing observation of atoms and advancing understanding of neurodegenerative diseases. These high resolutions are possible due to the technique's ability to collect data from vanishingly small, cryo-preserved samples, minimizing the need for large crystals. Key advantages of SAD in include cryo-preservation, which immobilizes biomolecules in vitreous ice to prevent radiation-induced damage, and low-dose imaging strategies that limit electron exposure while capturing sufficient diffraction intensities. These features address the challenges of beam sensitivity in biological samples, enabling native-state analysis without or . The foundational contributions to cryo-electron (cryo-EM), including SAD-based methods, were recognized by the 2017 , awarded to , , and Richard Henderson for developing techniques that revolutionized high-resolution imaging of biomolecules. A practical example of SAD's utility is in determining the of , where spot patterns reveal icosahedral arrangements and aid in reconstructing quasi-atomic models. In studies of human adenovirus type 5, SAD combined with cryo-EM confirmed the T=25 icosahedral of the , facilitating insights into assembly and host interaction without requiring full .

Limitations and Comparisons

Technical challenges

One of the primary technical challenges in selected area diffraction (SAD) arises from dynamical diffraction effects, where multiple events within the sample distort the intensities of diffraction spots, deviating from the kinematic that assumes single . This violation becomes significant in samples thicker than approximately 50 nm, as electrons undergo repeated interactions that alter relative intensities and can even excite kinematically forbidden reflections, complicating accurate structural . Area selection in SAD is constrained by the physical limitations of the selected area aperture and objective lens aberrations, typically restricting the minimum analyzable region to around 200 in . Smaller apertures lead to increased overlap from adjacent grains or features, introducing extraneous contributions that obscure the pattern from the intended area, particularly in polycrystalline or nanostructured materials. Radiation damage poses a severe limitation, especially for organic and biological specimens, where the high-energy electron beam induces , bond breakage, and mass loss, rapidly degrading crystallinity and fading diffraction spots within seconds of exposure. This effect restricts to low-dose regimes, often limiting total exposure to less than 20 electrons per square to preserve sample integrity. Quantitative analysis of SAD patterns is hindered by the unreliability of relative intensities due to dynamical effects and factors, necessitating advanced simulations such as the multislice to model multiple and predict accurate factors. Without such corrections, refinements yield poor agreement with known structures, as observed in cases where simulated patterns using multislice algorithms achieve better fits to experimental data than kinematic models alone. Beam-induced drift further complicates SAD experiments, as prolonged illumination causes sample movement from charging, heating, or structural changes, blurring patterns and requiring real-time tracking software to adjust the beam position dynamically during acquisition. This drift is particularly problematic in sensitive , where automated tracking can maintain pattern quality by compensating for shifts on the order of nanometers per second.

Relation to other diffraction techniques

Selected area electron diffraction (SAD) employs a parallel to probe larger areas of the sample, typically on the order of 1–10 µm in diameter, making it suitable for analyzing grains or regions greater than 100 nm. In contrast, convergent beam electron diffraction (CBED) utilizes a convergent with a range of incident wave vectors, focusing on much smaller areas (~10–100 nm), which enables detailed local crystallographic , including determination through higher-order Laue zone (HOLZ) lines. While SAD offers simplicity in setup and is ideal for routine orientation mapping in polycrystalline materials, CBED provides enhanced for nanostructures but requires more complex instrumentation to account for the beam convergence. Compared to X-ray diffraction (XRD), SAD targets nanoscale volumes within transmission electron microscopy specimens, allowing examination of minute crystallites that are inaccessible to bulk XRD methods requiring micrometer-sized or larger samples. Electrons in SAD interact more strongly with matter—by a factor of 10⁶ to 10⁷—resulting in shallower penetration depths and higher sensitivity to light elements via Coulombic scattering, though this also amplifies dynamical scattering effects not as prominent in XRD's weaker interactions. SAD's shorter electron wavelengths (e.g., ~1.97 pm at 300 keV) yield larger Ewald spheres and more accessible reflections at low angles (<2°), facilitating rapid pattern acquisition for phase identification in thin samples, whereas XRD excels in comprehensive structure refinement of bulk materials with greater penetration. Precession electron diffraction (PED) extends SAD by introducing a precessing in a hollow cone around the , which mitigates dynamical effects inherent in standard SAD through integration over multiple orientations, approximating kinematic conditions for more accurate intensity measurements. This advancement in PED supports structure solving using zero-order Laue zone reflections and conventional phasing algorithms, particularly for complex inorganic materials like quasicrystals, where SAD's strong multiple often necessitates prior models for interpretation. PED thus enhances SAD's utility for at the nanoscale, though it demands additional for beam precession. MicroED builds directly on SAD principles as a precursor for three-dimensional determination, employing continuous sample rotation during data collection to merge multiple two-dimensional SAD-like patterns into a full dataset, thereby overcoming SAD's limitation to static projections. This rotation method in MicroED enables high-resolution (e.g., 1.4 ) atomic models from beam-sensitive protein nanocrystals as small as hundreds of nanometers, using ultralow doses to minimize , in contrast to SAD's primary role in qualitative without rotational averaging. While SAD provides foundational two-dimensional for routine crystallographic screening, MicroED's sophistication allows de novo solving of biomolecular structures unsuitable for methods due to crystal size constraints. Overall, SAD distinguishes itself through its straightforward implementation for everyday identification and in , prioritizing accessibility over the enhanced , reduced artifacts, or full structural capabilities of advanced techniques like CBED, PED, and MicroED, which involve greater instrumental complexity for specialized applications.