Nebulizer

A nebulizer is a medical device that converts liquid medication into a fine aerosol mist for direct inhalation into the lungs, facilitating the delivery of therapeutic agents to treat respiratory conditions. Nebulizers are particularly useful for patients who have difficulty coordinating inhalation with metered-dose inhalers, such as young children, the elderly, or those with severe breathing difficulties, and they are commonly prescribed for conditions like asthma, chronic obstructive pulmonary disease (COPD), cystic fibrosis, and acute respiratory infections. Medications delivered via nebulizer include bronchodilators like albuterol to open airways, corticosteroids to reduce inflammation, and mucolytics to thin mucus in the lungs. The primary types of nebulizers include jet nebulizers, which use or oxygen to create the ; ultrasonic nebulizers, which generate through high-frequency vibrations; and vibrating mesh nebulizers, which employ a perforated vibrated at ultrasonic frequencies for more efficient and portable . Each type varies in efficiency, portability, and suitability for specific medications, with jet nebulizers being the most traditional and widely available, while mesh models offer quieter operation and faster treatment times. The origins of nebulization trace back over 4,000 years to ancient practices of inhaling herbal vapors, but the modern emerged in the 19th century with the invention of pressurized atomizers, and the term "nebulizer" was coined in 1872 from the Latin word for . Significant advancements occurred in the , including the introduction of ultrasonic models in 1949 and vibrating mesh technology in recent decades, improving output and reducing treatment duration. Today, nebulizers are regulated as Class II medical devices by the U.S. , ensuring safety and efficacy for home and clinical use.

Fundamentals

Definition and Purpose

A nebulizer is a medical device that converts liquid medication into a fine mist or aerosol, enabling direct inhalation into the lungs. This process facilitates the delivery of therapeutic agents either locally to the respiratory tract or systemically through absorption in the lungs, providing a direct route that avoids gastrointestinal absorption and first-pass metabolism for a more rapid onset of action. The primary purpose of nebulizers is to treat various respiratory conditions, including , chronic obstructive pulmonary disease (COPD), , and pulmonary infections, by administering medications such as bronchodilators (e.g., albuterol), corticosteroids, or antibiotics. The generated typically features particles in the 1-5 micron range, which optimizes deposition in the lower . As a non-invasive , nebulizers offer key benefits, particularly for patients unable to effectively use metered-dose inhalers or dry powder inhalers, such as young children, the elderly, or individuals experiencing severe dyspnea. This accessibility enhances treatment adherence and efficacy in vulnerable populations by allowing passive inhalation without requiring coordinated breathing techniques. Nebulization originated in the , with early powered devices emerging in the mid-1800s, though contemporary designs prioritize portability, efficiency, and integration with modern respiratory care.

Basic Mechanism of Operation

A nebulizer operates by atomizing medication into a fine of droplets suitable for , enabling direct delivery to the . The core process involves converting the —typically a or —into droplets through the application of , such as pressurized gas, ultrasonic vibration, or mesh perforation, followed by transport of the via a patient interface like a mouthpiece or for breathing. Key components of a nebulizer include a or to hold the , a nebulization chamber where the occurs, an energy source to drive the process (e.g., , piezoelectric elements, or electrical vibration), and an output interface to direct the to the patient. The fundamental principles rely on disrupting the liquid's to form droplets: this is achieved via shear forces from high-velocity fluid interactions or bubbles that collapse and fragment the liquid. During generation and transit through the device, partial of the (usually ) from the droplets reduces their size, enhancing respirability while potentially concentrating the active drug. Output is influenced by several factors, including (typically 5–10 L/min for gas-driven systems), fill volume (2–5 mL to ensure consistent performance), and nebulization duration (5–15 minutes until the liquid is depleted or begins). For , nebulizers commonly incorporate one-way valves to prevent exhaled humid air from re-entering and diluting the , as well as inline filters on the gas supply to capture bacterial contaminants and maintain sterility. Nebulization efficiency, defined as the proportion of the initial nominal dose emitted as respirable (droplets <5 μm in aerodynamic diameter), is generally low at 10–20%, with the remainder lost to deposition within the device or as non-respirable particles. This respirable fraction is critical for therapeutic deposition in the lungs.

Clinical Applications

Medical Indications and Guidelines

Nebulizers are primarily indicated for the delivery of aerosolized medications in the management of acute and chronic respiratory conditions, including exacerbations, (COPD), , , and acute respiratory infections such as . In , they are used to administer bronchodilators like during severe attacks to rapidly relieve and improve . For COPD, nebulizers facilitate the inhalation of bronchodilators and mucolytics to alleviate symptoms in patients with frequent exacerbations or those unable to coordinate use. In and , they enable the administration of antibiotics (e.g., nebulized gentamicin or tobramycin) and hypertonic saline to reduce bacterial load, thin , and prevent infections. For acute respiratory infections like , nebulizers deliver corticosteroids or bronchodilators to reduce and support in hospitalized patients. Clinical guidelines from authoritative bodies emphasize nebulizers in specific scenarios rather than as routine therapy. The Global Initiative for Asthma (GINA) recommends nebulizers for acute management in settings where patients, particularly young children or those with poor technique, cannot effectively use metered-dose inhalers or dry powder inhalers, but advises against their routine use in stable due to equivalent efficacy of handheld devices. Similarly, the American Thoracic Society (ATS) supports nebulizer use in COPD for patients experiencing dyspnea or breathlessness who struggle with coordination, positioning them as an alternative in acute exacerbations or for home maintenance therapy in severe cases. Both organizations prioritize on device selection, with nebulizers favored in emergency departments or for intubated individuals. Dosing protocols for nebulized medications vary by condition but follow standardized regimens to ensure safety and efficacy. For acute or COPD exacerbations, a typical dose of albuterol is 2.5 mg administered via nebulizer every 20 minutes for the first hour, followed by 2.5–5 mg every 4–6 hours as needed; this can be integrated with by using an oxygen-driven nebulizer at 6–8 L/min to deliver both medication and supplemental oxygen. In or , protocols often include 300 mg of nebulized tobramycin twice daily for 28 days in cycles to target infections. Frequency and duration are tailored to clinical response, with monitoring for side effects like . Patient selection for nebulizer therapy focuses on those who benefit most from passive aerosol delivery. They are ideal for infants, young children, elderly patients with low inspiratory flow rates, or critically ill individuals such as those who are intubated or unconscious, where coordinated is impossible. Nebulizers are not recommended as first-line for stable outpatients with adequate technique due to higher costs, longer treatment times (10–15 minutes per session), and potential for contamination. The includes nebulized (5 mg/mL solution) on its Model List of for antiasthmatic and COPD treatment, underscoring its global accessibility. Contraindications are limited but include to the medication, its preservatives, or excipients, which may cause allergic reactions. Post-2020 updates in respiratory guidelines have highlighted nebulizers' role in home-based management during pandemics like , particularly for patients with respiratory distress requiring bronchodilators or supportive therapy, with precautions such as using filters to mitigate transmission risks. The ATS and other bodies advise preferring metered-dose inhalers in shared clinical spaces but endorse home nebulization for isolated use to reduce visits.

Efficacy and Clinical Evidence

Nebulized bronchodilators have been shown to be as effective as metered-dose (MDIs) with spacers for relieving acute symptoms in both adults and children, according to a Cochrane and of 39 randomized controlled trials involving over 5,000 participants. In settings, nebulizers facilitate faster for patients unable to coordinate inhaler use, though overall clinical outcomes such as hospital admission rates remain comparable between the two methods. Clinical studies demonstrate that nebulizer therapy provides symptom relief in approximately 70-80% of acute cases, with significant improvements in forced expiratory volume in one second (FEV1) of 15-20% within 30-60 minutes post-treatment. For instance, albuterol nebulization in moderate-to-severe exacerbations leads to measurable bronchodilation and reduced respiratory distress in the majority of patients, supporting its role in initial . Despite these benefits, nebulizers exhibit lower efficiency, with only 10-30% of the administered dose reaching the lungs due to losses in the device and upper airways, compared to up to 50% efficiency with dry powder inhalers. Reusable models also carry a of bacterial , with studies reporting contamination rates of up to 70% in home-use nebulizers among pediatric patients with chronic respiratory conditions if not properly disinfected after each session. Evidence gaps persist regarding long-term chronic use, where data on sustained efficacy and safety are limited, often due to challenges in adherence and device maintenance. Ongoing clinical trials in the 2020s are exploring nebulized antivirals for (), such as investigational candidates in human challenge studies, to address acute viral infections in vulnerable populations. Recent 2023 research highlights advantages of vibrating mesh nebulizers, which improve treatment efficacy by approximately 20% over jet models in terms of output and deposition, leading to faster symptom resolution in pediatric . In children, nebulized therapies like ipratropium have reduced hospitalization rates by 9-25% in acute exacerbations compared to standard care alone. Post-2020 evidence from the underscores nebulizers' role in telemedicine-supported home management of respiratory conditions, with remote monitoring reducing emergency visits while minimizing transmission risks through single-use devices.

Aerosol Physics

Particle Generation and Characteristics

Nebulizers produce aerosol droplets through distinct physical mechanisms tailored to their design. In nebulizers, compressed gas induces pneumatic forces on the liquid formulation, fragmenting it into primary droplets. Ultrasonic nebulizers rely on piezoelectric vibrations to create bubbles in the liquid, which collapse and eject fine droplets from the surface. Vibrating nebulizers force the liquid through a perforated via ultrasonic vibrations, generating droplets directly from the apertures. These processes typically yield initial droplet diameters of 5-10 microns, which subsequently decrease to 1-5 microns as water evaporates during transit, concentrating the solute and reducing the overall . Key characteristics of nebulizer-generated aerosols include the mass median aerodynamic diameter (MMAD), which represents the aerodynamic diameter at which 50% of the aerosol mass resides, ideally ranging from 2-4 microns for effective pulmonary delivery. The geometric standard deviation (GSD) quantifies the polydispersity of the size distribution, typically falling between 1.5 and 2.5, indicating a moderately narrow spread suitable for consistent dosing. Aerosol size distributions often follow a log-normal pattern, where the MMAD is calculated as \text{MMAD} = e^{\mu}, with \mu denoting the mean of the natural logarithm of the particle diameters. The output rate, or dose delivered per unit time, can be expressed as \text{Dose output} = \frac{\text{fill volume} \times \text{efficiency}}{\text{time}}, where efficiency accounts for the fraction of liquid aerosolized. Several factors influence these aerosol properties. Higher liquid impedes droplet formation and ejection, reducing the overall output rate and potentially increasing residual volume in the nebulizer. affects droplet stability post-generation; elevated promotes coalescence and larger droplets, while can lower it to enhance fine particle production. The respirable fraction—defined as the proportion of particles with aerodynamic diameters below 5 microns—should exceed 50% to ensure adequate penetration. Upon , hygroscopic particles in nebulizer aerosols undergo growth due to water uptake in the humid respiratory environment, enlarging by 20-50% depending on the solute composition and relative . Recent investigations using have revealed that these particles often exhibit spherical with smooth surfaces, contributing to predictable aerodynamic behavior and minimal aggregation. Vibrating mesh nebulizers, for instance, tend to produce aerosols with narrower size distributions compared to other types.

Deposition in the Respiratory System

Aerosol particles generated by nebulizers deposit in the respiratory system through three primary mechanisms: inertial impaction, gravitational sedimentation, and Brownian diffusion. Inertial impaction predominates for larger particles exceeding 5 microns, causing them to collide with and adhere to the walls of the upper airways and oropharynx due to their momentum during inhalation. Gravitational sedimentation affects particles in the 1-5 micron range, allowing them to settle onto airway surfaces in the tracheobronchial tree and alveoli under the influence of gravity, particularly during breath-holding. Brownian diffusion governs the deposition of very fine particles below 1 micron, where random molecular motion enables them to reach and attach to alveolar walls. For nebulizers, total deposition in the respiratory tract is typically around 20% of the nominal dose, with approximately 10-20% in the oropharynx due to impaction and 10% in the lungs. Of the lung deposition, 50-80% occurs in the tracheobronchial region, and a smaller fraction targets the alveoli for deeper delivery. Alveolar deposition is particularly enhanced by slow inhalation flows around 30 L/min, which minimizes impaction losses and promotes sedimentation in peripheral lung regions. These patterns ensure that therapeutic agents can be directed toward specific sites, such as the conducting airways for bronchodilators or the alveolar space for anti-inflammatory drugs. Several physiological factors influence deposition efficiency, including breathing patterns and . Optimal deposition occurs with tidal volumes of approximately 500 mL and respiratory rates of 12-15 breaths per minute, as these allow sufficient time for sedimentation without excessive that promotes impaction. Upright facilitates greater deposition in the lower lobes compared to supine positioning by enhancing gravitational settling in dependent regions and improving overall penetration. The (ICRP) model provides a standardized framework for predicting regional deposition of inhaled aerosols, dividing the into compartments such as the extrathoracic, bronchial, bronchiolar, and alveolar-interstitial regions to estimate particle fate based on size, flow rates, and . Total dose is calculated as the product of the inhaled dose and the deposition fraction, which for nebulizers typically ranges from 10-25%, accounting for losses in the upper airways and device inefficiencies. This model has been widely adopted for simulating nebulizer performance in both healthy and diseased . In pediatric patients, upper airway losses are notably higher, reaching up to 80% of the nebulized dose due to narrower nasal passages and shorter inhalational paths that enhance impaction. Following deposition, plays a key role in particle removal, transporting approximately 50% of deposited aerosols from the tracheobronchial region toward the oropharynx within a few hours via coordinated ciliary action and flow. A 2017 SPECT-CT study demonstrated that vibrating mesh nebulizers can achieve up to six times higher pulmonary deposition compared to traditional jet nebulizers, attributed to more uniform and reduced oropharyngeal losses, thereby improving to distal airways in clinical settings.

Types of Nebulizers

Jet Nebulizers

Jet nebulizers, also known as pneumatic nebulizers, operate using compressed gas, typically air or oxygen, supplied from an external at a of 6-8 L/min. The design incorporates a Venturi system where the high-velocity gas passes through a narrow , generating that draws liquid medication from a into the airstream via a tube. This liquid is then sheared into fine droplets by the turbulent airflow, with a baffle mechanism positioned downstream to separate particles: larger droplets impinge on the baffle and recycle back to the for re-atomization, while smaller, respirable droplets proceed toward the mouthpiece. In operation, the continuous airflow through the atomizes the medication, producing an cloud that the patient inhales over a typical duration of 10-15 minutes, until the is nearly empty. The baffle system enhances efficiency by minimizing waste of non-respirable droplets, though a portion of the may escape during if no one-way is present. These devices require an external power source for the , making them dependent on or portable gas supplies, and they generate noticeable from the air and motor. Performance characteristics include a mass median aerodynamic diameter (MMAD) of 3-5 microns for the generated aerosols, which is optimal for deposition in the lower . The output rate generally ranges from 0.2-0.5 mL/min, depending on the model and formulation , though overall lung delivery efficiency remains low at 10-15% of the loaded dose due to losses in the device, exhaled air, and upper airways. Introduced in the as an advancement over earlier nebulizers, nebulizers are compatible with a wide range of medications, including viscous solutions and suspensions that may not suit other types. Advantages of jet nebulizers include their low cost, typically $20-50 for basic reusable units, and versatility with various drug formulations. However, they are bulky, require an external source, produce significant (around 60-70 ), and have prolonged times compared to newer alternatives. Inadequate poses a of bacterial and growth, potentially leading to respiratory infections; daily with mild soap and warm water, followed by air drying, is recommended after each use to mitigate this.

Ultrasonic Nebulizers

Ultrasonic nebulizers represent a type of electrical nebulizer that employs piezoelectric technology to generate without the need for . The core component is a piezoelectric crystal, which vibrates at high frequencies typically ranging from 1 to 3 MHz when electrically activated, producing ultrasonic waves that induce in the medication liquid. This process creates bubbles that collapse and fragment the liquid into fine droplets, forming an aerosol suitable for . These devices, first developed in , are designed for portability and often operate on batteries, making them ideal for home use. Nebulization typically occurs over 5 to 10 minutes per treatment session, delivering a continuous through a mouthpiece or mask. The resulting particles have a mass median aerodynamic diameter (MMAD) of 2 to 4 microns, which facilitates deposition in the lower . Ultrasonic nebulizers perform particularly well with low-viscosity solutions, achieving higher output rates compared to more viscous formulations. Key advantages of ultrasonic nebulizers include their quiet operation, compact size, and lack of dependence on external gas sources, enhancing user convenience. However, the vibrational energy can heat the liquid reservoir, with temperatures rising up to 40°C during use, potentially degrading heat-sensitive drugs such as proteins or antibiotics through denaturation. Overall efficiency is approximately 15-25% of the loaded dose reaching the lungs, influenced by factors like and patient breathing patterns. For , these devices require disassembly of components followed by thorough rinsing to prevent residue buildup.

Vibrating Mesh Nebulizers

Vibrating nebulizers utilize advanced micropump technology featuring a microporous plate perforated with to 4000 precisely formed apertures, typically 3 to 5 microns in diameter, which is vibrated by a piezoelectric at frequencies of 100 to 200 kHz. This extrudes directly through the apertures, producing a fine without the need for compressed gas or excessive heat. The devices are compact, handheld, and battery-powered, operating silently during treatments that typically last 2 to 5 minutes, while generating minimal drug waste through low residual volumes of less than 0.1 mL. These nebulizers offer high efficiency, delivering 50 to 70% of the nominal dose as respirable aerosol, and effectively preserve the integrity of sensitive drugs like proteins and biologics by avoiding thermal degradation. However, they are relatively expensive, often costing more than $100 per unit, and the mesh can clog when nebulizing highly viscous fluids, potentially reducing output. The technology emerged from patents filed in the 1990s, with notable commercial models including Aerogen's AeroNeb and Omron's NE-U22, which produce aerosols with a mass median aerodynamic diameter (MMAD) of 2 to 3 microns suitable for deep lung deposition. Vibrating mesh nebulizers have also received FDA clearance for use in clinical trials evaluating insulin nebulization, demonstrating their compatibility with macromolecules. In terms of performance, vibrating mesh nebulizers deliver consistent doses across varying patterns, with reduced to inspiratory rates compared to traditional designs. The use of disposable or replaceable es further minimizes risks by preventing cross-use of components exposed to . This portability and efficiency contribute to superior aerosol deposition in the , enhancing therapeutic outcomes for conditions requiring precise delivery.

Usage and Accessories

Instructions for Safe Use

To ensure effective and hygienic delivery of medication via nebulizer, users must follow precise preparation steps. Begin by washing hands thoroughly with and for at least 20 seconds to prevent . Assemble the clean components, including the nebulizer cup, tubing, mouthpiece or , and connect to the as per the manufacturer's instructions; always use only the prescribed and fill the reservoir with the exact dose to avoid overfilling, which can lead to inefficient or spillage. During usage, sit in an upright position to optimize lung deposition and prevent liquid pooling. Turn on the compressor and hold the nebulizer upright while breathing normally through the mouthpiece or mask for the full treatment cycle, typically 5 to or until the mist stops and sputtering occurs, indicating low liquid levels. At the end of nebulization, inhale deeply several times and gently tap the cup to dislodge any residual medication for complete delivery. If sputtering begins prematurely, it signals insufficient liquid and requires refilling or stopping to avoid dry running. Safety considerations are essential to minimize risks during operation. Perform nebulization in a well-ventilated area to reduce secondhand exposure to others, particularly during treatments with beta-agonists that may cause or other systemic effects; monitor for side effects such as rapid heartbeat, tremors, or , and discontinue use if severe symptoms arise, seeking medical advice immediately. For pediatric patients, adaptations like blow-by masks positioned 1-2 inches from an infant's face allow delivery without direct contact, ensuring comfort while maintaining efficacy. Hygiene protocols prevent microbial growth and . After each use, disassemble parts and rinse the medicine cup, mouthpiece, and with warm and mild dish ; disinfect by soaking in a 1:3 vinegar- solution for 30 minutes or boiling for 5 minutes if heat-tolerant, then rinse thoroughly and air- on clean towels to inhibit . Replace tubing and filters monthly or as specified, and store all components in a , dust-free .

Attachments and Maintenance

Nebulizers are equipped with various attachments to facilitate effective aerosol delivery tailored to patient needs and settings. Standard components include mouthpieces for adult users, which allow direct inhalation into the oral cavity, and pediatric masks designed for children, often featuring fun shapes like elephants to improve treatment compliance and reduce anxiety during use. T-piece adapters enable integration with mechanical ventilators, permitting inline aerosol administration without interrupting ventilation circuits. Air intake filters on compressor-based systems capture particulates from ambient air, ensuring cleaner aerosol generation. These attachments are typically affordable, ranging from $5 to $20 per unit depending on material and reusability. Proper extends longevity and ensures consistent performance. Routine involves weekly deep : disassemble the nebulizer cup, mouthpiece or , and top piece, then soak in a 1:1 solution of white and water for 30 minutes to remove residue and disinfect, followed by thorough rinsing with and air drying. Annual servicing of the by a is recommended to inspect internal components and replace worn parts, such as air filters every six months or as specified by the manufacturer. Users should regularly check for physical wear, including cracked baffles in jet nebulizers, which can impair output and require immediate replacement. Infection control is paramount, as nebulizers can harbor if not maintained rigorously. Always use distilled or sterile water for rinsing and filling to prevent mineral buildup and bacterial growth, such as , which poses a significant risk in immunocompromised patients like those with . In hospital environments, single-use disposable kits are preferred to minimize cross-contamination, and any device suspected of contamination should be discarded promptly. Troubleshooting common issues involves monitoring output and component integrity. For vibrating mesh nebulizers, if the aerosol output falls below 0.3 mL/min, the mesh should be replaced to restore efficiency. Moving parts, such as in ultrasonic models, require sparing only if recommended by the manufacturer to avoid residue accumulation.

Historical Development

Early Inventions

The origins of nebulizer technology emerged in the mid-19th century amid efforts to deliver medications directly to the for conditions like . In 1858, Dr. Jean Sales-Girons patented the first powered nebulizer, known as the pulverisateur, a portable pump-operated device that used a manual handle to draw liquid medication from a and atomize it into a fine mist through a and impaction plate. This invention marked a significant advancement over earlier manual methods, enabling more effective generation for therapeutic use. Subsequent developments in the late built on this foundation. In 1864, German physician Hermann Siegle introduced the steam spray , which utilized steam from a small to produce fine droplets for , further refining the process. By the early , nebulizers were employed in medical practice, including for administering vapors during surgical , where consistent mist generation was crucial for safe delivery. However, these early devices suffered from notable limitations, such as low portability due to their bulk and dependence on heat sources or manual operation, as well as inconsistent output influenced by variable and liquid properties. Key milestones in the and expanded nebulizer applications. Oxygen-driven nebulizers appeared in the for , facilitating the controlled delivery of anesthetic gases in clinical settings. In the , glass jet nebulizers powered by hand bulbs gained popularity for administration, such as epinephrine solutions, exemplified by devices like the Glaseptic nebulizer. The Pneumostat, an early electric compressor nebulizer developed in and marketed by Riddell, Ltd., in 1935, improved reliability by using for , though its 9-pound weight limited mobility. The mid-20th century saw a pivotal shift from manual to electric compressors in the , enhancing output consistency and enabling broader home and clinical adoption of nebulizers for respiratory .

Modern Innovations and Advancements

The Babington nebulizer , described in 1972, marked a significant advancement in generation for medical applications in the early , enabling more reliable delivery of therapeutic mists by utilizing a thin film of liquid spread over a surface and atomized by high-velocity gas. This improved upon earlier pneumatic designs by reducing clogging and enhancing consistency for viscous medications. Ultrasonic nebulizers, first developed in for humidification and adapted for nebulization in the using high-frequency piezoelectric vibrations to generate without compressed gas, represented an early electronic advancement, though initial models were bulky. In the , portable ultrasonic nebulizers emerged as a key innovation, leveraging piezoelectric vibrations to produce fine without compressors, making them suitable for home and mobile use while minimizing noise and power requirements. The saw the rise of vibrating mesh technology, exemplified by the PARI eFlow nebulizer, which was approved in 2004 and featured an electronically controlled vibrating membrane with laser-drilled apertures for precise droplet sizes (typically 3-5 μm), achieving higher drug output and shorter treatment times compared to traditional jet models. Breath-enhanced models, introduced around the same period, further optimized efficiency by synchronizing release with , potentially doubling to the lungs and reducing waste by up to 50% in clinical evaluations. Recent innovations in the include smart nebulizers equipped with connectivity for real-time adherence tracking via mobile apps, allowing patients and providers to monitor usage patterns and dosage compliance remotely. During the , clinical trials demonstrated the feasibility of nebulized vaccines, such as live-attenuated formulations delivered via nasal nebulizers, which provided robust mucosal immunity in animal models and preliminary human studies for vaccination. Emerging patents and prototypes as of 2025 incorporate algorithms to optimize dosing by analyzing breathing patterns and adjusting parameters dynamically, enhancing personalization for conditions like . Nebulizers are classified as Class II medical devices by the FDA, requiring 510(k) premarket notification to ensure and through standardized testing for performance. In the , they must obtain under the to verify compliance with essential requirements for design and manufacturing. Updated 2023 FDA guidance on reprocessing reusable devices emphasizes rigorous cleaning protocols to mitigate infection risks, recommending disposable components for single-patient use in high-risk settings amid concerns over bacterial . The global nebulizer market reached approximately USD 1.22 billion in 2024, fueled by rising demand for home-based respiratory care and post-COVID integration with platforms for virtual monitoring of treatment adherence.