Photometer

A photometer is an instrument that measures the or of , often focusing on such as , , or , and can encompass from the to spectra, including the visible range. These devices typically employ photodetectors like photodiodes, photoresistors, or photomultipliers to convert into electrical signals for quantification, enabling precise assessment of light's physical or perceptual characteristics. In photometry specifically, measurements account for human visual sensitivity through luminosity functions, distinguishing it from broader radiometric approaches that ignore perceptual factors. The origins of photometry trace back to ancient astronomy, where in the BCE developed a system to classify stellar qualitatively. Quantitative measurement advanced in the , with credited as the inventor of the first photometer around 1729, a device that compared light intensities by equalizing illumination on a surface. Subsequent developments included visual comparison instruments in the , such as those by Zöllner in , and the transition to photoelectric methods in the early , which introduced objective precision using cells and later tubes. By the mid-20th century, photometers had evolved into standardized tools for industrial and scientific use, with institutions like NIST establishing standards for units such as the and since the early 1900s. Photometers find essential applications across diverse fields, including and evaluation to ensure compliance with standards for human vision, such as workplace illuminance levels around 500 . In astronomy and , they measure celestial light, , or photosynthetic in aquatic systems to assess ecological impacts. Chemical analysis employs spectrophotometric variants for concentration determination via absorption, as in the Beer-Lambert law, while industrial uses span testing, automotive uniformity, and research. Modern digital photometers, often portable or integrated with like integrating spheres, support high-precision tasks in vision science, signaling, and .

Definition and Principles

Definition and Classification

A photometer is an instrument designed to measure the intensity of , particularly , , or , across wavelengths such as , (UV), and (IR). These measurements quantify how interacts with surfaces or passes through media, providing essential data for fields like , astronomy, and . Photometers convert into readable signals, enabling precise assessment of light properties without altering the source itself. The term "photometer" originates from the Greek words phōs () and metron (measure), reflecting its purpose as a device for quantifying . This was formalized in through Johann Heinrich Lambert's seminal 1760 publication Photometria sive de mensura et gradibus luminis, colorum et umbrae, which introduced systematic methods for measurement and described an early photometer . Photometers differ from related instruments: spectrometers resolve into its wavelength components for , while luxmeters represent a specialized subset of photometers focused exclusively on in visible . Photometers are classified by several criteria to suit diverse applications. By detection method, they include visual types relying on human eye comparison for brightness equality, photographic variants using light-sensitive films to record intensity variations, photoelectric models employing photocells or photomultiplier tubes for electrical signal conversion, and digital systems utilizing charge-coupled devices (CCDs) or complementary metal-oxide-semiconductor (CMOS) sensors for high-resolution data capture. Classification by measured light property encompasses intensity (luminous or radiant power per unit solid angle), flux (total light output), and, in extended cases, spectral content (though this overlaps with spectrophotometry). By wavelength range, photometers are categorized as visible (optimized for human-perceived light), UV-Vis (covering ultraviolet to visible spectra), and IR (infrared-focused for thermal radiation). Key units distinguish photometric (vision-weighted) from radiometric (energy-based) measurements. Photometric units include the (lm/m²) for , representing light received on a surface, and the (cd) for , denoting light emission directionally. Radiometric counterparts use watts per square meter (W/m²) for , focusing on total without visual sensitivity weighting. This duality ensures photometers align with both human perception and physical energy quantification.

Fundamental Operating Principles

Photometers operate by detecting and quantifying light through interactions between photons and detector materials, fundamentally rooted in the principles of and photometry. measures the total across all wavelengths, focusing on physical energy quantities like , independent of human perception. In contrast, photometry weights these measurements by the human eye's , emphasizing visible light (approximately 380–780 nm) to assess perceived brightness. This distinction is crucial, as photometers for visual applications incorporate the photopic luminosity function V(λ), which peaks at 555 nm and models the eye's response to different wavelengths. When light encounters a detector in a photometer, photons interact via , , , or . occurs when photons transfer energy to electrons in the detector material, generating a measurable electrical signal proportional to the incident ; this is the primary mechanism in semiconductor-based detectors like photodiodes. allows photons to pass through the detector without interaction, while redirects them away from the surface, and disperses them in multiple directions due to surface irregularities or internal particles, potentially reducing measurement accuracy. These interactions determine the detector's efficiency in converting into a quantifiable output. A core principle in photometer operation is the conversion of radiance to radiant flux, which quantifies the total power emitted or received by a source or detector. Radiance L(θ, φ) represents the power per unit area per unit in direction (θ, φ), and the Φ is obtained by integrating over the Ω subtended by the detector: \Phi = \int_{\Omega} L(\theta, \phi) \cos \theta \, d\Omega Here, cos θ accounts for the of the detector to the incoming rays, and dΩ = sin θ dθ dφ integrates across the or relevant . This equation ensures accurate flux measurement regardless of angular distribution, forming the basis for calibrating photometers against known sources. For photometric applications, quantities are adjusted using the luminous efficacy function V(λ) to mimic human vision. E_v, the luminous flux per unit area on a surface, is calculated as: E_v = 683 \int_{380}^{780} E_e(\lambda) V(\lambda) \, d\lambda where E_e(λ) is the spectral irradiance ( per area per ), V(λ) weights the contribution of each wavelength to perceived illumination (normalized to 1 at 555 nm), and 683 lm/W is the exact maximum K_m fixed by the 2019 SI redefinition of the . Detector response in photometers must exhibit —output signal proportional to input over a wide —to ensure accurate quantification, with sensitivity defined as the signal per unit flux (e.g., A/W for photodiodes). traces to the 2019 SI-defined via the fixed luminous efficacy; practical realizations often use standard sources such as blackbody radiators at 2042 K (corresponding to the freezing point of ) for broad-spectrum applications. Common error sources in photometer measurements include , which introduces extraneous from unintended paths and can inflate readings by 1–5% in poorly baffled systems; detector noise, arising from thermal or effects that limit in low-light conditions; and spectral mismatch, where the detector's response deviates from V(λ), causing up to 10% errors with non-standard sources like LEDs. These are mitigated through against reference sources, such as lamps at 2856 K, and computational corrections based on spectral characterization.

Historical Development

Early Visual Photometers

The earliest systematic efforts to measure light intensity relied on human visual comparison, marking the inception of visual photometry in the 18th century. In 1760, Johann Heinrich Lambert published Photometria, sive de mensura et gradibus luminis, colorum et umbrae, which laid the foundational principles for quantifying light through experimental visual assessments conducted between 1755 and 1760. Lambert's tentative photometer involved comparing the brightness of illuminated white surfaces, using standardized wax candles as reference sources to match perceived illumination levels from unknown lights. This approach built on earlier work by Pierre Bouguer, who in 1729 described brightness matching between surfaces, but Lambert formalized it into a comprehensive system of photometric laws, emphasizing the inverse square law for light propagation and the cosine law for diffuse reflection. Toward the end of the century, , Count Rumford, advanced these methods with his photometer introduced in 1798, a modification of focused on shadow extinction. Rumford's instrument used a translucent screen with a grease spot, where two light sources illuminated opposite sides; the intensities were compared by adjusting distances until the or spot contrast vanished, indicating equal illumination based on the principle that perceived brightness equates when balances. This grease-spot method simplified visual matching for practical applications, such as evaluating lamp efficiencies in public lighting, and was applied as early as 1792 in assessments of artificial lights. Early visual photometers also hinted at reciprocity in human light perception, where the eye's response to total light exposure appeared proportional to intensity multiplied by duration, foreshadowing later formalizations. However, these devices faced inherent limitations due to the subjective nature of human judgment; observer variability in sensitivity, fatigue, and adaptation led to inconsistencies, with estimates of brightness differing by up to 20-30% between individuals under similar conditions. Such inaccuracies, compounded by the lack of standardized viewing conditions, motivated the shift toward more objective mechanical and chemical methods in subsequent decades.

19th-Century Advancements

In the , Scottish physicist William Ritchie developed a photometer that advanced visual comparison techniques by incorporating shadow projection principles. This device utilized a grease spot on a translucent screen illuminated from opposite sides by two sources, with movable screens allowing adjustment of distances to achieve balance when the spot's visibility vanished, indicating equal intensities. The ratio of light intensities was determined by the inverse square of the distances from the sources to the screen, providing a more precise mechanical method than earlier subjective estimates. The method of extinction of shadows, refined throughout the from earlier designs, employed parallel sources positioned on adjustable stands to cast overlapping shadows of opaque rods or objects onto a white screen, typically 6-8 feet away. Observers varied the distances of the sources until the shadows' intensities equalized and the boundary between them became imperceptible, applying the where intensity I \propto \frac{1}{d^2} to compute relative luminous fluxes as \frac{I_1}{I_2} = \left( \frac{d_2}{d_1} \right)^2. This procedure, conducted on long photometric benches for accuracy, bridged early visual photometers toward standardized industrial measurements despite challenges with parallel rays in applications like lighthouses. By the , the Lummer-Brodhun photometer marked a significant leap in precision for comparing diffuse light sources, featuring a cube-shaped assembly that merged images from two sides of an opaque white screen into a central spot and surrounding ring viewed through an . Developed by Otto Lummer and Eugen Brodhun at the Physikalisch-Technische Reichsanstalt, it enabled adjustments until the images matched in brightness, achieving up to eight times the accuracy of grease-spot designs and becoming the standard for in laboratories and gas-electric industries. Chemical photometers emerged in the mid-19th century, leveraging the light sensitivity of silver halides like for quantitative measurement. French physicist Alexandre-Edmond Becquerel's galvanic photometer, for instance, used two silver plates coated with immersed in an ; exposure to light induced a proportional to intensity, offering an early objective alternative to visual methods and influencing photographic development. These devices, though limited by chemical instability, paved the way for actinometry in astronomy and exposure control. Standardization efforts culminated in 1881 when the International Electrical Congress in adopted the spermaceti candle—burning at 120 grains per hour—as the global unit of , harmonizing disparate national standards like Britain's Parliamentary Candle and facilitating consistent photometric comparisons across gas and emerging electric lighting sectors. This unit, refined to 7.8 grams per hour with tolerances up to 5%, supported regulatory testing despite variability critiques, setting the stage for later refinements.

20th-Century Transitions to Electronic Methods

The transition to electronic methods in photometry during the was profoundly influenced by Albert Einstein's 1905 explanation of the , which described light as discrete packets of energy (photons) capable of ejecting electrons from a metal surface, thereby enabling the design of devices that convert light directly into electrical signals. This theoretical foundation paved the way for photometers, which replaced subjective visual comparisons with objective electrical measurements. One of the earliest practical implementations occurred in 1907 at , where Joel Stebbins and F.C. used a cell connected to a to detect moonlight, marking the birth of photoelectric photometry. By the 1910s, commercial cells, produced by firms like Elster and Geitel, became available and were integrated into photometers for astronomical observations, offering improved sensitivity over mechanical predecessors despite limitations in low-light detection. A key milestone in the 1920s was the introduction of barrier-layer cells, which featured a thin selenium layer sandwiched between metal electrodes to generate a photovoltaic voltage without external power, revolutionizing portable meters in . These cells, refined for greater light sensitivity, enabled the first battery-free electric meters, such as early models from Electrical Corporation, allowing photographers to measure incident light accurately in the field and reducing reliance on visual estimation. In the 1930s and 1940s, photomultiplier tubes (PMTs) emerged as a breakthrough for high-sensitivity photometry, invented by Soviet Leonid A. Kubetsky in 1930 through a using a photocathode and multiple to amplify weak photocurrents by factors of thousands. amplification worked by accelerating photoelectrons from the photocathode onto secondary emission surfaces, where each impact released additional electrons, repeated across several stages for exponential . The overall gain G is given by G = \delta^n where \delta is the secondary emission coefficient per (typically 3–5) and n is the number of dynode stages (often 10–14), achieving gains up to $10^8 for detecting faint light sources in and astronomy. The mid-20th century saw the rise of solid-state detectors, with semiconductor photodiodes developed in the early , which generated current proportional to incident light via the in a p-n junction. By the 1960s, the shift to silicon-based sensors accelerated, as these offered superior stability, lower noise, and compact form factors compared to vacuum tubes, facilitating portable photometers for field and laboratory use in and . Digital integration transformed photometers in the with the advent of analog-to-digital converters (ADCs), which digitized continuous light-induced signals for precise, noise-resistant , evolving from 8-bit to 12-bit resolutions to interface seamlessly with emerging microprocessors like the Intel 8080. This enabled real-time measurement and data logging, as microprocessors allowed automated calibration and computation of photometric quantities, markedly improving accuracy and speed in applications such as and industrial .

Measurement Techniques

Transmission Photometry

Transmission photometry measures the intensity of that passes through a transparent or translucent sample, providing quantitative data on the sample's such as or . The fundamental principle relies on the Beer-Lambert law, which states that the absorbance A of by a sample is directly proportional to the concentration c of the absorbing , the path length l through the sample, and the molar absorptivity \epsilon at a specific : A = \epsilon l c. This relationship assumes monochromatic , a dilute sample, and negligible interactions between absorbing molecules. The basic setup involves a source emitting a beam through the sample holder, followed by a detector that records the transmitted intensity relative to the incident intensity. In ultraviolet-visible (UV-Vis) transmission photometry, a broadband light source such as a deuterium or tungsten-halogen lamp illuminates the sample, with a selecting specific wavelengths for across a typical range of 200-800 . The , often employing a , disperses the light to isolate narrow bandwidths, enabling precise measurement of electronic transitions in molecules. This technique is widely applied in , where the at selected wavelengths correlates with color intensity or concentration of colored species, such as in of dyes or biochemical assays. Infrared (IR) transmission photometry extends the measurement to longer wavelengths, typically from 700 nm to 1 mm, probing molecular vibrations and rotations that produce characteristic absorption bands. Fourier transform infrared (FTIR) spectrometers are commonly integrated, using an interferometer to generate an interferogram that is Fourier-transformed into a spectrum, allowing simultaneous detection across the IR range for identifying functional groups in organic compounds. A key challenge in IR transmission is atmospheric absorption by water vapor and carbon dioxide, which can obscure sample signals and requires purging with dry nitrogen or using sealed cells to minimize interference. Instrumentation in transmission photometry varies by configuration to enhance accuracy and stability. Single-beam setups direct light sequentially through reference and sample positions, offering simplicity and higher but susceptible to source fluctuations. Double-beam configurations split the using a or , simultaneously measuring reference and sample paths to compensate for drifts in intensity or detector response, thus improving long-term stability for quantitative work. Detectors are selected based on : photomultiplier tubes (PMTs) for UV-Vis due to their high gain and to low-light levels via multiplication, and lead sulfide (PbS) detectors for , which provide room-temperature operation and broad response in the near- to mid- region. Calibration ensures reliable measurements by establishing a baseline transmittance. Neutral density filters, which uniformly attenuate without wavelength dependence, are used to verify photometric linearity and correct for instrumental response across intensity ranges. Error correction for , which can mimic in turbid samples, involves subtracting baseline spectra or applying mathematical models to isolate true losses.

Reflectance Photometry

Reflectance photometry involves measuring the reflected from a surface to determine its , particularly in the . The fundamental principle is defined by the reflectance ratio R = \frac{I_r}{I_i}, where I_r is the of the reflected and I_i is the of the incident . To achieve accurate measurements of diffuse , integrating spheres are commonly employed, as they collect and spatially integrate the through multiple internal reflections, enabling hemispherical averaging of the reflected . This setup is particularly suited for non-specular surfaces, where the sphere's high- coating (typically with reflectance r > 0.94) minimizes losses and accounts for the sphere multiplier effect to enhance signal uniformity. In visible light applications, reflectance photometry is essential for assessing and color, as it quantifies how surfaces interact with wavelengths between 400 and 700 nm. For pigmented layers, such as those in coatings or fabrics, the Kubelka-Munk theory provides a foundational model by relating and to observed , expressed as \frac{K}{S} = \frac{(1 - R)^2}{2R}, where K is the , S is the , and R is the at infinite thickness. This equation allows for the prediction of color development in opaque materials under diffuse illumination, facilitating formulation adjustments for consistent visual appearance. Instrumentation in reflectance photometry includes goniophotometers, which measure the angular dependence of to capture bidirectional scattering distribution functions (BSDF) or bidirectional reflectance distribution functions (BRDF) across a wide angular range. These devices use motorized stages and collimated sources to simulate varied illumination geometries, providing data on how varies with incidence and observation angles. Integration with CIE standards, such as the 1931 , converts spectral measurements into tristimulus values () for standardized color assessment, using the CIE color matching functions and illuminants like D65 to mimic daylight conditions. Applications extend to material science, where reflectance photometry evaluates opacity by comparing reflected intensities against incident to determine hiding power in coatings. In quality control for paints, it ensures batch-to-batch color consistency by measuring reflectance in geometries like 0°/45°, aligning with human . Similarly, in textiles, it verifies uniformity and vibrancy, supporting non-destructive analysis of fabric surfaces for aesthetic and functional standards. Limitations arise from surface , as roughness from or environmental factors can alter diffuse and introduce variability. Correction methods involve calibrating against white standards, such as or references with near-100% , to normalize data and account for instrumental drift, often performed periodically to maintain accuracy.

Absorption Photometry

Absorption photometry, specifically atomic absorption photometry, relies on the absorption of light by in a gaseous sample to determine elemental concentrations with high specificity. The principle involves free atoms absorbing at characteristic wavelengths corresponding to electronic transitions from the to higher energy levels, enabling of over 70 elements. This technique was pioneered by Alan Walsh in the , who recognized its potential for sensitive chemical analysis using atomic absorption spectra. In photometry, the primary light source is a filled with the element of interest, which emits sharp, element-specific lines when electrically excited, ensuring selective by matching vapor wavelengths. The sample is introduced via atomizers such as flame systems, where aspiration into a burner produces a gaseous cloud, or furnace atomizers, which electrothermally vaporize small sample volumes (typically 5–20 μL) in a heated tube for enhanced . The transmitted light passes through a to isolate the desired and reaches a detector, often a , which measures the intensity reduction due to . Modern instruments achieve detection limits down to (ppb) for many elements, particularly with furnace atomization./6%3A_Atomic_Spectroscopy/6.2%3A_Atomization_Sources/6.2B%3A_Electrothermal_Atomization__Graphite_Furnace) The is quantified using the Beer-Lambert law, adapted for absorption:
A = \log\left(\frac{I_0}{I}\right) = \epsilon b c
where A is , I_0 and I are incident and transmitted intensities, \epsilon is the absorptivity for the , b is the path length through the atomic vapor, and c is the concentration of ground-state atoms. This relationship, similar to the general form referenced in transmission photometry, allows calibration curves for accurate quantification.
To address non-specific absorption from matrix interferences or molecular species, background correction techniques are essential. The deuterium lamp method employs a continuum source to measure broadband absorption separately, subtracting it from the total signal via electronic modulation. Alternatively, the Zeeman effect correction applies a magnetic field to split the atomic absorption line, measuring analyte-specific polarized absorption while the shifted background components are isolated and deducted. These methods improve accuracy in complex samples. Applications of absorption photometry focus on trace metal detection, such as lead, , and mercury in environmental waters, soils, and biological tissues like or , supporting regulatory monitoring and health assessments. Graphite furnace variants excel in analyzing limited sample volumes from clinical or ecological sources, achieving ppb sensitivity without preconcentration. Walsh's invention spurred widespread adoption, revolutionizing in fields like and since the technique's commercialization in the 1960s.

Specialized Methods and Applications

Photon Counting in Low-Light Detection

in low-light detection relies on the statistical detection of individual arrival times, enabling precise measurement of faint light sources where the average flux is low. This method treats arrivals as a process, in which the variance of the detected count equals the mean number of photons, σ² = N, providing a fundamental limit on the due to . Such statistical principles ensure that even at very low intensities, the detection remains reliable by accumulating counts over time to overcome inherent quantum fluctuations. Key detectors for photon counting include avalanche photodiodes (APDs) operated in linear mode and single-photon avalanche diodes (SPADs) in Geiger mode, which amplify the initial through an internal avalanche process to produce a detectable for each absorbed . The η, defined as the ratio of detected counts to incident s, typically ranges from 10% to 80% depending on wavelength and material, with silicon-based SPADs achieving high values in the . These devices offer single-photon sensitivity, distinguishing them from bulk detectors by resolving discrete events rather than integrated intensity. A prominent technique is time-correlated single photon counting (TCSPC), which measures the time delay between a periodic and the arrival of individual s to reconstruct temporal profiles, such as lifetimes, with . For measurement in astronomy, integrates counts over exposure times to determine stellar or cosmic light es, allowing detection of sources with rates below 1 per second while maintaining and low . This integration approach is particularly valuable for faint object photometry, where it outperforms analog methods by avoiding read-out . The primary advantages of lie in its exceptional sensitivity, capable of operating at fluxes less than 1 per second, which enables applications in systems for imaging in extreme low-light conditions and in for experiments involving entangled s or single-photon sources. These capabilities stem from the digital nature of the detection, which provides and immunity to certain analog distortions. However, limitations include dead time—the brief period after a detection event during which the detector is insensitive, typically 10–100 ns in SPADs—and afterpulsing, where trapped charges trigger false subsequent counts, reducing accuracy at higher rates. Calibration often involves attenuated lasers to simulate low-flux conditions and verify quantum efficiency and linearity, ensuring traceability to primary standards.

Photographic and Imaging Applications

In , photometers serve as exposure meters to measure scene and guide optimal camera settings, ensuring accurate light capture on or sensors. Early photoelectric exposure meters emerged in the 1930s using cells to generate current proportional to , marking a shift from visual estimation methods. By the 1950s, companies like Sekonic introduced dedicated models such as the P-1, which relied on similar photoelectric principles for handheld portability. The adoption of () cells in the 1960s revolutionized built-in camera photometers, as these photoresistors decreased resistance with increasing light, enabling compact integration into single-lens reflex (SLR) viewfinders for through-the-lens () metering. Photographers like integrated photometers with the , a tonal mapping technique developed in 1940 to visualize and control across nine zones from deep shadows (Zone 0) to bright highlights (Zone IX). By metering key scene elements and adjusting to place them in desired zones—such as shadows at Zone III for detail—Adams used incident or reflected light readings to previsualize negative density, compensating for film's limited . Reciprocity failure occurs in photographic emulsions at exposure extremes, where the linear relationship between E and exposure time t breaks down, requiring adjustments to total H = E \times t. The basic in logarithmic form is \log(H) = \log(E) + \log(t), but at low light levels (e.g., shutter speeds beyond 1 second), efficiency drops, necessitating longer times or higher to achieve equivalent . For instance, films like Tri-X may require 2-3 times the calculated at 10 seconds, with developers adjusted to restore contrast. In , (CCD) and (CMOS) sensors function as array photometers, with millions of pixels acting as individual photodiodes to quantify light intensity across the scene. CMOS arrays, dominant since the 2000s due to on-chip amplification and lower power use, enable real-time readout for applications like burst . analysis of these arrays visualizes tonal distribution, revealing —the span from noise floor to saturation—typically 12-14 stops in modern sensors, allowing post-exposure corrections for over- or underexposure without reciprocity concerns. Beyond exposure, photometers support film densitometry, where transmission devices measure negative opacity (optical density) to assess development uniformity and characteristic curves, ensuring consistent gamma () around 0.6 for films. In machine vision for , photometry-based systems inspect products like automotive symbols by quantifying and color uniformity, detecting defects with sub-pixel precision far exceeding manual checks. The evolution continues with ambient light sensors, often (ALS) calibrated for measurements, serving as accessible photometers via apps that approximate incident readings with ±10-15% accuracy in controlled conditions. These integrate principles for scene evaluation, bridging traditional tools like 1930s meters to mobile imaging workflows.

Spectroscopic and Analytical Uses

employs photometers capable of measuring across a full , enabling detailed analysis of substance composition by wavelength-specific or . These instruments, known as spectrophotometers, integrate detectors (DADs) that simultaneously capture data from multiple wavelengths in the UV-Vis range, typically from 190 to 1100 nm, in under one second. This rapid acquisition facilitates applications in chemical identification and quantification, where the Beer-Lambert law relates to concentration without the need for sequential scanning. In , flame photometry serves as a specialized emission-based technique for detecting metals such as sodium and . When samples are aspirated into a , atoms are excited and emit characteristic ; for instance, sodium emits at 589 , while emits at 766 , allowing quantification through . This method operates in mode, contrasting with modes in , and is particularly effective for trace analysis in clinical and samples due to its and sensitivity for these elements. Environmental monitoring utilizes photometers based on differential optical spectroscopy (DOAS) to detect atmospheric pollutants like (NO2). These systems measure differential in the UV-Vis along paths through the atmosphere, isolating gases from ; for NO2, key bands around 400-450 nm enable column density retrievals with high . Multi-axis DOAS (MAX-DOAS) configurations enhance vertical profiling, supporting air quality assessments in urban areas. In biomedical applications, transmission photometers underpin pulse oximeters for non-invasive blood (SpO2) measurement. These devices transmit red light at 660 nm and near-infrared light at 940 nm through , exploiting differential absorption: oxygenated absorbs more at 940 nm, while deoxygenated form absorbs more at 660 nm, yielding SpO2 via the ratio of ratios calculation. This dual-wavelength approach provides real-time monitoring with accuracy typically within 2-4% of gas values in individuals with lighter skin tones, though biases up to several percent overestimation have been reported in those with darker skin tones, as noted in FDA guidance and recent studies (as of 2025). As of January 2025, the FDA has proposed updated recommendations to improve pulse oximeter performance across diverse skin tones, mandating testing on broader pigmentation ranges to mitigate biases. Photometers integrate seamlessly with (HPLC) as UV-Vis detectors, particularly diode array types, to analyze separated analytes in real time. As eluents pass through the flow cell, absorption spectra are recorded, and software performs peak integration—calculating areas under chromatogram peaks proportional to concentration via baseline subtraction and trapezoidal approximation—for quantitative determination in pharmaceutical and environmental assays.

References

  1. [1]
    Photometers - RP Photonics
    Photometers are instruments for measuring light properties such as irradiance or illuminance, or optical properties of objects.
  2. [2]
    Photometry | NIST - National Institute of Standards and Technology
    Thus, photometry is essential for evaluation of light sources and objects used for lighting, signaling, displays, and other applications where light is intended ...
  3. [3]
    A light history of photometry: from Hipparchus to the Hubble Space ...
    In 1844, Ludwig Seidel began to put the Steinheil prism photometer through its paces. Eventually in 1866, Seidel published a catalogue of208 stars, being a ...
  4. [4]
    Note on the Invention of Photometry - AIP Publishing
    It has been suggested that François d'Aguillon (1566–1617) made a photometer, thus anticipating Pierre Bouguer (1698–1758). It is shown that D'Aguillon was ...
  5. [5]
    Discoveries in Modern Astronomy
    In 1861, Karl Friedrich Zöllner (1834-82) in Berlin introduced the visual photometer. Further improvement was achieved by introducing photoelectric cells for ...
  6. [6]
    [PDF] PRACTICAL TECHNIQUES AND THEORY OF ASTRONOMICAL ...
    In was not until the invention of the photoelectric photometer and the photomultiplier tube in the early 1900's, allowing for precise, objective measurements of.
  7. [7]
    Candela | NIST - National Institute of Standards and Technology
    May 16, 2019 · When work in photometry was initiated at the National Bureau of Standards (NBS) around 1903, a group of carbon-filament lamps calibrated against ...
  8. [8]
    Photometer - an overview | ScienceDirect Topics
    A photometer is defined as an instrument used to measure the intensity of light, which is essential for assessing irradiance levels that affect ...
  9. [9]
    Photometer - Definition and References - AZoOptics
    A photometer is an instrument for measuring the light intensity, or optical properties, of optical sources or detectors.
  10. [10]
    photometer - Wiktionary, the free dictionary
    English. English Wikipedia has an article on: photometer · Wikipedia. Etymology. From photo- +‎ -meter. Pronunciation. (General American) IPA: /fəˈtɑmɪtɚ/, ...
  11. [11]
    Foundational study on the measurement of light, Johann Lambert ...
    In the Photometria he described his photometer and propounded the law of the absorption of light named after him.Missing: etymology | Show results with:etymology
  12. [12]
    [PDF] Photometers - Instrument Systems
    High precision photometer used to measure illuminance, luminous intensity, and luminous flux in laboratory or field applications as well for control purposes ...
  13. [13]
    Photometer - an overview | ScienceDirect Topics
    Photometry is used to describe the assessment of the properties of light sources, such as radiation characteristics, color temperature, and color rendering ...
  14. [14]
    Photometry: The Answer to How Light Is Perceived - Photonics Spectra
    Imaging photometers are used for fast and accurate uniformity testing of cockpit displays and automotive dashboards. They are suitable for ISO ergonomics ...
  15. [15]
    Radiometry and Photometry | Ossila
    Radiometry and photometry are techniques used to measure light. Light measured radiometrically involves the whole electromagnetic spectrum, whereas light ...
  16. [16]
    [PDF] Radiometry and Photometry
    All light measurement is considered radiometry with photometry being a special subset of radiometry weighted for a typical human eye response. Photometry.
  17. [17]
    [PDF] Fundamentals of Radiometry and Photometry
    Radiometry is the process of measuring electromagnetic radiation. Radiometry deals with the measurement of the energy transferred by a source through a ...
  18. [18]
  19. [19]
    II.8 Reflection, Transmission, and Absorption - Light Measurement
    Both processes can be accompanied by diffusion (also called scattering), which is the process of deflecting a unidirectional beam into many directions. In this ...
  20. [20]
    1.5 Calculation of Radiometric Quantities (Examples) - Gigahertz-Optik
    Calculating the surface's exitance Me from Equation 4 using the relation dΩ = sin(ϑ) dϑ dφ gives: Me = ∫ Le cos(ϑ) dΩ = Le, 2π, ∫, π/2, ∫ cos(ϑ) sin(ϑ) dϑ dφ = ...
  21. [21]
    luminous flux - Illuminating Engineering Society
    ... ∫Φe,λ​V(λ)dλ ,. where: Φv = luminous flux, (lumens, lm) Φ e,λ = watts per ... For scotopic vision, the corresponding spectral luminous efficiency function, V'(λ ...
  22. [22]
    National Institute of Standards and Technology detector-based ...
    [5] The blackbody radiation at the gold point (1337.58 K in IPTS-68) was used to calibrate a variable temperature blackbody, which provided the NBS scale of ...
  23. [23]
    Spectral Mismatch Correction Factor a* / F - Gigahertz-Optik
    The a* factor corrects for spectral mismatch in photodiodes, compensating for errors in the nominal weighting function, and is used to correct measurement ...
  24. [24]
    [PDF] Spectral responsivity-based calibration of photometer and ...
    Jun 20, 2013 · The spectral mismatch errors can be especially high when narrowband SSL sources are measured. The performed matrix-based color corrections ...
  25. [25]
    Photometrie. Photometria, sive De mensura et gradibus luminis ...
    Feb 14, 2016 · Photometria, sive De mensura et gradibus luminis, colorum et umbrae (1760). by: Lambert, Johann Heinrich, 1728-1777; Anding, E. (Ernst), 1860 ...
  26. [26]
    Photometria sive de mensura et gradibus luminis, colorum et umbrae
    The first appraisal of Photometria appeared in 1776 in Georg Klügel's German translation of Priestley's 1772 survey of optics. An elaborate reworking and ...Missing: etymology | Show results with:etymology
  27. [27]
    Bouguer and Lambert's pioneering contributions to ...
    Written independently and published in the same year, 1760, Bouguer's Traité d'optique and Lambert's Photométria are the foundations of photometry.<|separator|>
  28. [28]
    Light's Measure: A History of Industrial Photometry to 1909
    Apr 13, 2025 · A summary history is provided of the development of industrial photometry from its earliest beginnings to 1909.<|control11|><|separator|>
  29. [29]
    (PDF) A notion or a measure the quantification of light to 1939
    PDF | This study, presenting a history of the measurement of light intensity from its first hesitant emergence to its gradual definition as a scientific.<|control11|><|separator|>
  30. [30]
    Bouguer's photometer. | Download Scientific Diagram - ResearchGate
    The practitioners of visual photometry in the 18th and 19th centuries were convinced that the eye was reliable and capable of making accurate judgments in ...
  31. [31]
    (PDF) Light's Measure: A History of Industrial Photometry to 1909
    Industrial photometry evolved significantly from 1792 to 1909, paralleling technological advances in lighting. Photometry quantified luminous values, ...
  32. [32]
    [PDF] A Notion or a Measure: The Quantification of Light to 1939 - CORE
    From the 1920s, the long-established visual methods of observation were increasingly replaced by physical means of light measurement, a process initially.
  33. [33]
    [PDF] A History of Light and Colour Measurement - Science in the Shadows
    The 18th century polymath Johann Lambert (1728–77) made his own study of ... An early success was an improvement in visual photometers. Otto Lummer.
  34. [34]
    [PDF] 5-History of Photo-Parrington - Chemistry
    The 'galvanic photometer' of A. E. Becquerel consisted of two pure silver plates coated with silver iodide (sometirfies chloride or bromide) immersed in.
  35. [35]
    The standardisation of light and photometry – A historical review
    This paper investigates the technological challenges and solutions to defining a reproducible standard for light together with its measurement, from the 18th ...
  36. [36]
    Einstein and The Photoelectric Effect - American Physical Society
    Jan 1, 2005 · This became known as the photoelectric effect, and it would be understood in 1905 by a young scientist named Albert Einstein.
  37. [37]
    A Brief History of Photometers - Labmate-Online
    Oct 5, 2014 · One of the earliest recorded uses of a detector is in 1907 at Harvard University. A selenium cell was connected to a galvanometer and used by ...Missing: invention | Show results with:invention
  38. [38]
    Selenium cell · Birth of Photoelectric Photometry
    By June 1907, the photometer, with a selenium cell manufactured by Elster and Geitel at its heart, was ready for testing. On the first night Stebbins operated ...Missing: history | Show results with:history
  39. [39]
    the age of selenium begins - phsc.ca
    Jan 31, 2019 · Selenium cells stayed in use into the 1960s when cadmium sulfide (CdS) cells became common – requiring a tiny battery once again. Far more ...
  40. [40]
    On the history of photomultiplier tube invention - ScienceDirect.com
    On 4 August 1930 Soviet-Russian physicist and engineer LA Kubetsky proposed new method and device to amplify weak photocurrents by dozens and hundreds ...
  41. [41]
    [PDF] PHOTOMULTIPLIER TUBES - Hamamatsu Photonics
    ... photomultiplier tubes whose gain is less dependent on voltage (the slope of gain-voltage curve is not sharp) are less affected by the dynode voltage change.
  42. [42]
    Bell Labs announces the phototransistor, March 30, 1950 - EDN
    Mar 30, 2015 · Shive's invention used a single collector wire and its end rested in a small dimple ground into one side of a germanium disk. Light focused ...Missing: N. | Show results with:N.
  43. [43]
    Photodetectors | Particle Detectors: Fundamentals and Applications
    As in the field of electronics, semiconductors have revolutionised the field of light generation and detection since the 1950–1960s. The electron tube technique ...
  44. [44]
    History of Analog-to-Digital Converters (ADCs) - Dewesoft
    May 3, 2023 · An analog-to-digital converter, or ADC, is a system that converts analog phenomena into digital signals that can be processed and stored on computers.Missing: photometers | Show results with:photometers
  45. [45]
    [PDF] Beer's Law: Determining the Concentration of a Solution
    linearity of this plot arises from the Beer-Lambert law (or Beer's law, for short) which states that the absorption of light by a substance is proportional ...
  46. [46]
    [PDF] Standardization in transmission spectrophotometry in the visible and ...
    Apart from this, the Lambert-Beer law is often not obeyed, especially if the measuring beam is not collimated so that the pathlength is not exactly equal to ...
  47. [47]
    UV-Visible Spectroscopy - MSU chemistry
    Each monochromatic (single wavelength) beam in turn is split into two equal intensity beams by a half-mirrored device. One beam, the sample beam (colored ...
  48. [48]
    [PDF] Spectrophotometric and colorimetric determination of the colors of ...
    Doubtful samples were also examined for excitation by each portion of the visible spectrum by energy from the monochromator of a spectrophotometer. More ...
  49. [49]
    [PDF] Infrared Spectroscopy in Conservation Science - Authentication in Art
    Apr 5, 2025 · Molecular vibrations account for the strongest absorption bands in an ... Developing Fourier transform infrared spectroscopy (FTIR) as an ...
  50. [50]
    A Rapid and Simplified Approach to Correct Atmospheric ... - NIH
    Oct 31, 2024 · These temporal fluctuations cause absorption variations that interfere with the sample's spectrum, making accurate analysis challenging. While ...
  51. [51]
    UV-Visible absorption spectrophotometer - Terpconnect
    UV-Visible absorption spectrophotometer with blackbody light source, grating monochromator, and photomultiplier detector. Assumptions: light source is imaged ...
  52. [52]
    Optical Spectroscopy | Cornell Center for Materials Research (CCMR)
    Cary 5000 UV-Vis-NIR Spectrophotometer · Detectors: Scintillating from 175nm-800nm; PbS from 800nm-3300nm · Optics: Schwarzschild coupling · Sources: Standard ...<|control11|><|separator|>
  53. [53]
    [PDF] An absolute light scattering photometer: II. Direct determination of ...
    Diffusers are illuminated by a relatively large beam, usually of the order of 10 mm in width, when they are used in light scattering calibration. Table. 2 shows ...
  54. [54]
    Errors in Spectrophotometry and Calibration Procedures to Avoid ...
    The study is limited to photometers for UV, VIS, and near IR. A good survey of the literature published until 1972, especially regarding tests for photometric ...
  55. [55]
    [PDF] Integrating Sphere Theory and Applications - labsphere
    The sphere as part of a radiometer or photometer can directly measure the flux originating from lamps and lasers or the flux density produced from hemispherical ...
  56. [56]
    Kubelka-Munk Theory - an overview | ScienceDirect Topics
    The Kubelka–Munk theory is defined as a model used to predict the optical properties of paper by analyzing the interactions of light with various materials ...
  57. [57]
    pab BSDF/BRDF goniophotometer
    Our gonio-photometers measure the angular transmittance/reflectance of arbitrary materials (BSDF) and the angular light distribution of small lamps.
  58. [58]
    CIE Fundamentals for Color Measurements | NIST
    Sep 30, 2000 · The paper first overviews the CIE system of colorimetry, covering CIE 1931 color matching functions, XYZ tristimulus values, the x, ...Missing: photometry | Show results with:photometry
  59. [59]
    Best Spectrophotometer for Measuring Color Reflectance - HunterLab
    Sep 19, 2024 · It has several applications, including: Textiles; Paint and coatings; Food; Chemicals; Paper; Building products. The Agera Spectrophotometer ...
  60. [60]
    Reflectance Measurements in the Paint Industry*
    The spectrophotometer is used in the reproduction of colors for successive batches in preference to either wet or dry standards. The record used for such color ...
  61. [61]
    [PDF] 7. reflectance spectrophotometry and colorimetry
    Measurements can be calculated based on either the 2° or 10° standard observer and any of 11 illuminants (CIE standard illuminants A, C, D50, and D65 and.Missing: photometry | Show results with:photometry
  62. [62]
    Atomic absorption spectroscopy - CSIROpedia
    Feb 21, 2011 · The concept of atomic absorption spectroscopy (AAS) came to CSIRO scientist Alan Walsh in a flash of inspiration as he was gardening at his ...
  63. [63]
    Atomic Absorption Spectrometry - an overview | ScienceDirect Topics
    The technique offers excellent detection limits permitting trace element analysis (∼0.1–1 ng ml−1 for a furnace and 10 ng ml−1 –1 μg ml−1 for a flame atomizer).
  64. [64]
    The Development of Atomic Absorption Spectroscopy (AAS)
    Jun 16, 2021 · It all began in the 1950s with a pioneering article by Alan Walsh. Today, it is an established standard technology in the field of ...Missing: invention | Show results with:invention
  65. [65]
  66. [66]
    Background correction in atomic absorption utilizing the Zeeman effect
    Zeeman atomic absorption spectrometry using high frequency modulated light polarization. ... Atomic Absorption Spectrometry and Deuterium Background Correction.Missing: lamp | Show results with:lamp
  67. [67]
    [PDF] Method 7010: Graphite Furnace Atomic Absorption ...
    1.1. Metals in solution may be readily determined by graphite furnace atomic absorption spectrophotometry (GFAA). The method is simple, quick, and applicable to ...
  68. [68]
    Statistical treatment of photon/electron counting; extending the linear ...
    Nov 29, 2010 · Photon counting is a well-established method providing low detection limits in sensitive optical measurements. However, conventional approaches ...
  69. [69]
    [PDF] Chapter 5 Photon Counting
    The purpose of this section is to develop a simple model photon detection. The detector will be a device that responds to the number of photons that.
  70. [70]
    Avalanche Photodiodes – APD, single-photon detection, Geiger ...
    ... quantum efficiency, i.e., not every incident photon can trigger an avalanche. Single-photon APDs are called SPADs, which means single-photon avalanche diodes.What is an Avalanche... · Detection Noise · Geiger Mode for Single-photon...
  71. [71]
    By the Numbers: Photon Counting and Number Resolution
    Photon counting offers several advantages over analog detection. Because the technique discriminates between the electrical pulses based on their amplitudes — ...
  72. [72]
    [PDF] Time-Correlated Single Photon Counting - PicoQuant
    Time-correlated single photon counting records time-dependent light intensity after a laser pulse, using fast electronics to measure time and increment ...
  73. [73]
    [PDF] Photon Counting in Astrophotometry. Fundamentals and Some ...
    Photon counting in astrophotometry counts pulses per time interval, independently of pulse height, and is not subject to electron multiplication noise.
  74. [74]
    High Sensitivity Photodetector for Photon-Counting Applications
    Photon counting is a statistically accurate technique to measure faint light signals, based on the measurement of a random sequence of pulses generated by ...
  75. [75]
    Low noise imaging photon counter for astronomy
    Photon counting provides significant benefits in terms of stability, linearity, dynamic range, lack of reciprocity failure, and interpretive clarity for many ...
  76. [76]
    Photon Counting
    Photon counting is an alternative to intensity measurements, enabled by single-photon detectors. It can involve time tagging, histogram generation and ...
  77. [77]
    Time-Resolved Calibration of Photon Detection Efficiency ... - MDPI
    The primary limitation of InGaAs/InP SPADs' performance is the afterpulsing effect. Afterpulses are transient error counts created by carriers trapped in ...
  78. [78]
    Exposure metering - the last word
    Aug 6, 2025 · The 1930s brought the first real revolution in metering with the introduction of photoelectric exposure meters. Selenium cells, which generated ...
  79. [79]
    About Us - Sekonic
    Established in June 1951 as Seiko Electric Industries Co., Ltd., the first Sekonic brand exposure meter introduced was the revolutionary Sekonic P-1. ‌.
  80. [80]
    A Beginner's Guide to Exposure Meters
    Optical or Extinction meters use some sort of visual aid to help determine exposure. One type used printing-out-paper, which visibly turns dark when exposed to ...
  81. [81]
    The Zone System Of Planned Photography | Aperture | Spring 1955
    Now let us integrate the twofold nature of the zones. This can be done visually on an exposure meter. Take the print grey scale and its Roman Numerals in the ...
  82. [82]
    The Photographic Reciprocity-Law Failure for Radiations of Different ...
    The reciprocity-law failure of a number of dye-sensitized emulsions has been studied for radiations of different wave-length. The curves showing the ...
  83. [83]
    [PDF] Selection guide / CCD/CMOS image sensors - Hamamatsu Photonics
    The CCD image sensors realize high quantum efficiency in the ultraviolet, visible, and near infrared regions. The CMOS image sensors realize low price, low ...
  84. [84]
    What is a histogram, and how do you use it? - DPReview
    Oct 3, 2025 · Histograms are not specific to photography and are simply graphs that show the distribution of continuous numerical data. They help visualize ...
  85. [85]
    [PDF] Understanding Graphic Arts Densitometry - X-Rite Photo
    Black and white transmission densitometers are used for film inspection, as well as scanner and imagesetter calibration and linearization. The basic ...
  86. [86]
    Choosing a Photometry-Based Imaging System for Inspection
    Imaging photometers feature high-resolution sensors and advanced optical components to capture images in greater detail than a typical machine vision camera.
  87. [87]
    Smartphones as a Light Measurement Tool: Case of Study - MDPI
    This paper discusses the possibility of using smartphones as an accessible and accurate tool, focusing on the measurement of light.Missing: photometer | Show results with:photometer
  88. [88]
    Spectroscopy Market Profile: Photodiode Array UV-Vis ...
    The advantage of the PDA over scanning UV–vis spectrophotometers is its speed, allowing the capture of the full spectrum (190–1100 nm) in under 1 s. The PDA ...Missing: photometers | Show results with:photometers
  89. [89]
    What is a diode array spectrophotometer?
    The diode array detector is the secret to fast spectra collection. A diode array instrument can collect a full range UV/Vis spectrum in milliseconds to seconds ...
  90. [90]
    Flame Photometry - an overview | ScienceDirect Topics
    Flame photometry is defined as a method used to determine the concentration of elements, such as potassium, in samples by measuring the emitted light when ...
  91. [91]
    Camping Burner-Based Flame Emission Spectrometer for ...
    Aug 7, 2014 · The obtained emission spectrum is characteristic of each individual atom, such as sodium (589 nm) and potassium (766 nm), or molecule, such as ...
  92. [92]
  93. [93]
    Remote sensing of NO2 emission from the central urban area of ...
    Jul 7, 2012 · The central urban area encircled by the mobile DOAS field measurements undergoes heavy local NO2 pollution from vehicle emissions, whereas the ...Missing: photometer | Show results with:photometer
  94. [94]
    Pulse Oximetry - StatPearls - NCBI Bookshelf - NIH
    Jul 30, 2023 · The normal oxygen levels displayed on a pulse oximeter commonly range from 95% to 100%. Hypoxemia is an oxygen saturation of less than 90%.Missing: photometer | Show results with:photometer
  95. [95]
    Pulse oximetry: Understanding its basic principles facilitates ...
    Exploiting this difference in light absorption properties between O2Hb and HHb, pulse oximeters emit two wavelengths of light, red at 660 nm and near-IR at 940 ...Missing: photometer | Show results with:photometer
  96. [96]
    How It Works: UV Detection for HPLC | LCGC International
    UV detection in HPLC uses light absorption by molecules. Variable wavelength detectors use a grating, while diode array detectors use a fixed grating and ...Missing: photometer | Show results with:photometer
  97. [97]
    5.1: High performance liquid chromatography - Chemistry LibreTexts
    Aug 19, 2025 · For example, it is common for HPLC with a UV-vis detector to be coupled with a mass spectrometer so that analytes can be identified based on ...