X-ray machine

An is a device that generates , a form of high-energy , to create images of the internal structure of opaque objects. It produces these images by directing an X-ray beam through the object, where varying degrees of create contrast that is captured by a detector, such as or a . are widely used in diagnostics to visualize bones, tissues, and organs; in security screening to detect concealed items in and ; and in industrial to inspect materials for defects without damage.

Fundamentals

Principles of X-ray production

X-rays occupy a position in the between and gamma rays, with energies typically ranging from 100 to 100 keV. In X-ray machines, these s are generated through the acceleration of electrons in a , where propels electrons from a toward a positively charged target. Upon impact with the , the electrons interact with the target's atomic nuclei and electrons, producing X-rays via two primary mechanisms: and characteristic . Bremsstrahlung, or "braking radiation," occurs when the high-speed electrons are decelerated by the of the anode's atomic nuclei, converting into a . This process yields a continuous of wavelengths from a minimum value determined by the accelerating voltage up to approximately 10 nm, corresponding to energies of about 0.12 to 120 keV. Characteristic radiation, in contrast, arises when an incoming electron ejects an inner-shell electron from a target atom, creating a vacancy that is filled by an outer-shell electron; the energy difference is emitted as discrete , forming sharp line spectra superimposed on the . These line energies are specific to the anode material's atomic structure. The output of X-rays is influenced by several key factors. Tube voltage (kVp) sets the maximum and increases output approximately proportional to its square, enhancing beam penetrability. Tube current () and exposure time determine the number of electrons available, with output directly proportional to their product (), thereby scaling the quantity of photons linearly. The anode material, often due to its high (Z=74) and of 3422°C, affects both the efficiency of production and heat dissipation; higher Z boosts yield. Overall X-ray I follows the approximate relation I \propto Z \cdot V^2 \cdot i, where V is the tube voltage and i is the current.

Types of X-ray machines

X-ray machines are classified primarily by their imaging modality, design configuration, and intended clinical purpose, encompassing systems that produce static or dynamic images for diagnostic and interventional applications. Conventional film-based radiography uses photographic film to capture a single static image, relying on the differential absorption of X-rays by tissues to produce high-contrast projections suitable for initial evaluations of fractures or lung conditions. This type represents the traditional approach, where the film is developed chemically after exposure. Computed radiography (CR) systems bridge analog and digital methods by employing photostimulable phosphor plates that store X-ray energy as a latent image, which is then scanned by a laser to generate digital data for processing and display. These machines allow reuse of imaging plates and integration with existing X-ray generators, facilitating a transition to digital workflows without full hardware replacement. Direct digital radiography (DR) advances this further with flat-panel detectors that convert X-rays directly into electrical signals, enabling immediate image acquisition and higher spatial resolution for detailed anatomical visualization. DR systems reduce processing time and radiation dose compared to CR, making them ideal for high-volume settings. Fluoroscopy units provide real-time dynamic imaging by continuously emitting s and capturing images on a monitor, often using image intensifiers or flat-panel detectors to guide procedures such as insertions. Computed tomography (CT) scanners, as specialized machines, acquire multiple projections from various angles to reconstruct cross-sectional slices, offering three-dimensional views of internal structures with enhanced contrast for soft tissues. These systems typically involve rotating tubes around the patient, differing from in their volumetric data output. Design configurations distinguish fixed installations, such as room-mounted or units for stable, high-throughput environments, from portable variants like mobile DR systems for bedside imaging in intensive care. Dental intraoral machines, often compact and fixed or handheld, focus on localized high-resolution imaging of teeth and using small sensors placed inside the . C-arm fluoroscopes, with their adjustable arc-shaped connecting the source and detector, enable intraoperative imaging during surgeries like orthopedics or placements. Hybrid systems incorporate specialized features for targeted applications; units, a variant of , include compression paddles to flatten breast tissue for uniform X-ray penetration and improved detection. systems, typically fluoroscopy-based, integrate contrast injection mechanisms to visualize blood vessels in real-time, supporting vascular interventions. Selection of X-ray machines depends on requirements, with high-resolution or dental systems preferred for fine details and lower-resolution screening options like portable for general surveys. Factors such as , procedural needs, and radiation dose minimization guide choices, ensuring alignment with diagnostic goals.

History

Discovery and early experiments

The discovery of X-rays is credited to German physicist Wilhelm Conrad Röntgen, who made an accidental observation on November 8, 1895, while conducting experiments with at the . Working in a darkened room, Röntgen noticed that a nearby screen coated with barium platinocyanide began to fluoresce, even though the cathode ray tube was covered with black paper to block light, indicating the emission of some unknown penetrating radiation from the tube. This phenomenon, initially termed "X-rays" due to their mysterious nature, marked the birth of . Röntgen's early experiments utilized Crookes tubes—early vacuum tubes designed for studies—to generate the rays, confirming their key properties through systematic tests over the following weeks. He verified that X-rays were invisible to the , capable of penetrating materials like paper, wood, and while being absorbed by denser substances such as bone and metal, and that they could induce in certain screens. A pivotal demonstration came on December 22, 1895, when Röntgen produced the first X-ray image, or röntgenogram, of his wife Anna Bertha's hand, revealing its skeletal structure and a on her finger after a 15-minute on a . These experiments established X-rays as a previously unknown form of penetrating distinct from , later identified as short-wavelength electromagnetic waves, laying the groundwork for their scientific validation. The immediate impact of Röntgen's findings was profound, beginning with his publication of "Über eine neue Art von Strahlen" (On a New Kind of Rays) on December 28, 1895, in the proceedings of the Physico-Medical Society, which detailed the rays' properties and included the famous hand image. This work earned him the first in 1901, recognizing the discovery's transformative potential for science and medicine. News spread rapidly across and the via telegraphs and press reports; for instance, the Wiener Presse announced it on January 5, 1896, and followed on January 16, 1896, sparking widespread interest. Early demonstrations occurred in , with Röntgen presenting to scientific societies, while in the U.S., researchers like quickly replicated the effects using modified bulbs. Initial medical applications emerged swiftly, as physicians within weeks used X-rays to visualize bone fractures and locate foreign bodies like bullets, revolutionizing diagnostics for skeletal injuries.

Key developments and commercialization

Following Wilhelm Röntgen's discovery of X-rays in 1895, early experimental devices evolved rapidly into practical machines through key engineering advancements in the early . A pivotal improvement came in 1913 when , working at , invented the hot-cathode , which used a heated filament to emit electrons in a high , providing stable and controllable X-ray output far superior to the unreliable gas-filled tubes like Crookes devices that had dominated previously. This innovation enabled consistent beam intensity and reduced variability, marking a shift from erratic experimental setups to reliable diagnostic tools. In the and , further refinements addressed power and heat limitations, with the development of high-voltage generators allowing voltages up to 200,000 volts by 1922 for penetrating thicker materials, and later reaching 1,000,000 volts by 1931 through General Electric's efforts. To manage the intense heat generated during prolonged use, rotating designs emerged in the early , featuring a spinning disk that distributed thermal load and permitted higher intensities without anode meltdown; Siemens introduced a commercial version in 1933 with an all-tungsten disk rotating at high speeds. These enhancements coincided with the creation of portable X-ray units for military applications, starting with Marie Curie's "Little Curie" mobile setups during , which brought to battlefield hospitals and treated over a million soldiers. By , compact field units like the British MX2 and U.S. Army Picker models were deployed widely, enabling on-site diagnostics in combat zones. Commercialization accelerated in the as major firms entered the market, with producing Coolidge tubes for widespread sale starting in 1913 and developing early systems from 1896 onward, including the "Reform Apparatus" in the for therapeutic applications that standardized equipment design. By the , film-screen systems became standardized, featuring double-emulsion films on flexible supports paired with intensifying screens to boost sensitivity and reduce exposure times, as pioneered by DuPont's blue-tinted base films that improved image viewing and became industry norms. Amid this growth, radiation safety concerns emerged in the early 1920s following reports of skin damage and cancers among early users, prompting initial regulations and the first International Congress on Radiology in 1925, which advocated for basic shielding in machine housings. This led to self-rectifying, grounded tubes with inherent lead shielding by the mid-1920s, reducing operator exposure and marking the onset of formalized protection standards.

Design and Components

X-ray tube

The is the primary component responsible for generating X-rays through the interaction of high-speed electrons with a material. It consists of a and enclosed in a vacuum-sealed . The includes a heated filament that emits electrons via when current is applied. These electrons are accelerated toward the by a high-voltage potential difference, typically 30–150 . The , often made of or a , serves as the where electrons decelerate, producing X-rays via and characteristic . Rotating , driven by an , dissipate heat and allow higher power outputs. The tube is housed in a lead-lined metal to contain and cool the components with oil or air circulation.

Power supply and control systems

The power supply and control systems provide the electrical energy needed to operate the X-ray tube and regulate exposure parameters. The high-voltage generator converts low-voltage input to high-voltage , supplying 20–150 across the tube for electron acceleration. Generator types include single-phase, three-phase, and high-frequency inverters, with high-frequency offering better voltage ripple (under 1%) for consistent X-ray output. The filament supply heats the cathode (3–5 A, 5–12 V) to electron emission and thus tube current (mA). Control systems allow selection of kilovoltage peak (kVp) for beam energy, milliampere-seconds () for exposure duration and intensity, and timers for precise . Modern systems include microprocessor-based interfaces for automated exposure (AEC) to optimize dose and image quality.

Accessories and shielding

Accessories in machines are essential for precise control, optimization, and safety, allowing operators to limit to the area of while minimizing scatter and unnecessary dose. These devices include collimators and filters that the , as well as grids that enhance by reducing scattered . Shielding components, such as aprons and barriers, protect patients and from stray . receptors, often housed in cassettes, capture the attenuated for diagnostic purposes. Collimators are adjustable devices typically made of lead or other high-attenuation metals that restrict the beam to the specific region of clinical interest, thereby reducing scatter and dose. By limiting the field size to match the image receptor, collimators prevent spillover of beyond the target area, which improves and complies with regulatory requirements for all radiographic examinations. In systems, automatic collimators adjust the field dynamically with changes in source-to-image distance, ensuring efficient beam utilization. Rectangular collimation, for example, can reduce the exposed area by approximately 50% in , further minimizing scatter. Filters, often constructed from aluminum or , are placed in the beam path to remove low-energy photons, a process known as beam hardening that enhances image contrast by reducing the contribution of soft X-rays that primarily increase patient dose without diagnostic value. These spectral shaping filters are user-selectable or automatic in modern systems, with materials like aluminum providing inherent filtration while adds further hardening for thicker body parts. In , wedge-shaped filters can equalize beam intensity across varying densities, lowering skin dose in high-attenuation regions. must meet FDA standards to optimize dose and image quality across applications like and general . Shielding protects against stray and scattered using lead or lead-equivalent materials to attenuate the beam effectively. Lead aprons for staff and patients typically range from 0.25 to 0.5 mm in thickness, with 0.5 mm serving as a standard for adequate during procedures; integrity checks are recommended to ensure no defects compromise safety. shields, also lead-based, are used for reproductively capable patients to safeguard sensitive areas, while shields are required for children and recommended for adults when they do not interfere with the . Facility walls in controlled areas are designed with shielding equivalent to 1–2 mm of lead to keep occupational doses well below the limit of 50 mSv, with design goals such as 5 mGy/year for controlled areas per NCRP standards, often incorporating lead sheets or barriers in veterinary or interventional setups. These measures align with federal guidelines to minimize biological risks from . Anti-scatter grids and cassettes facilitate high-quality image acquisition by managing scatter and capturing the beam. Anti-scatter grids, composed of thin lead strips separated by radiolucent materials like aluminum or , absorb scattered X-rays while permitting primary rays to reach the receptor, thereby improving contrast in thick-body such as the . Common grid ratios range from 6:1 to 10:1, with higher ratios offering greater scatter reduction at the cost of requiring increased ; they are essential in systems sensitive to scatter. Cassettes house image receptors, including traditional screen- combinations where phosphors convert X-rays to for film , or flat-panel detectors using materials like cesium iodide and for direct signal conversion. These receptors determine overall image fidelity, with formats allowing reuse after signal erasure and providing wider for low-dose . Grids and cassettes are positioned between the patient and receptor, and grids should be inspected annually for damage to prevent artifacts.

Operation

X-ray generation process

The process of generating in a medical X-ray machine begins with the operator selecting key parameters via the control console, including tube voltage (kVp, typically 40-150 kV for diagnostic imaging), tube current (mA, often 100-1,000 mA), and exposure time (usually less than 100 milliseconds for standard ). These settings determine the beam's energy and output, tailored to the anatomical region being imaged. Once parameters are set, the in the is heated by applying a (around 10 V and 3-7 A), causing where electrons are released from the into the of the . A potential (matching the selected kVp) is then applied across the and , accelerating these electrons at high speeds toward the positively charged target, usually made of . Upon striking the , the electrons interact with its atomic nuclei, primarily producing X-rays through (deceleration radiation) and characteristic radiation, with only about 0.9% of the electron's converted to X-rays at 100 , the rest dissipating as heat. The resulting X-rays exit the tube through a specialized or , forming a divergent beam directed outward. The characteristics of the X-ray beam are precisely controlled to optimize image quality and patient dose. Tube voltage (kVp) governs beam quality, influencing the energy spectrum and maximum (up to the kVp value, with average energy about one-third to one-half of that maximum), which affects tissue penetration. Beam quantity and intensity are regulated by milliampere-seconds (, the product of and exposure time), which scales the number of electrons emitted and thus the fluence. For effective imaging, the patient is positioned relative to the and detector, with the beam aligned to the target area using adjustable collimators that restrict the field size, minimizing unnecessary to surrounding tissues. Exposure can be a single pulse for static or continuous/pulsed for dynamic , with typical durations ranging from 0.001 to 10 seconds depending on the procedure and equipment settings.

Image acquisition and processing

In traditional X-ray imaging, film-screen systems capture the X-ray beam using radiographic coated with crystals, typically , embedded in a . When X-rays interact with these crystals, they eject electrons that migrate to sensitivity sites, forming a through the Gurney-Mott mechanism, where silver ions are reduced to metallic silver specks too small to be visible without . An intensifying screen, often containing materials like calcium , amplifies the effect by converting X-rays to visible light, which further exposes the film, reducing the required radiation dose. The chemical development process then converts the into a visible one: a developer solution reduces exposed grains to black metallic silver, while unexposed grains remain intact. This is followed by fixing, where a fixer removes the unexposed using , halting development and making the permanent, and washing, which rinses away residual chemicals to prevent degradation. These steps produce an analog negative where varying correspond to densities, with denser areas appearing lighter due to less penetration. Digital image acquisition has largely replaced film-screen systems with methods like and direct radiography (DR). In CR, photostimulable phosphor (PSP) plates, coated with materials such as fluorohalide, store X-ray energy as trapped electrons in color centers upon exposure. The plate is then scanned with a in a reader unit, stimulating the release of blue proportional to the stored energy, which is detected by a and converted into a . Direct radiography (DR) employs flat-panel detectors with thin-film transistor (TFT) arrays, typically using or layers. In indirect DR, a like cesium iodide converts X-rays to light, which is then detected by photodiodes in the TFT array, generating electrical charges read out row by row. Direct DR uses photoconductive materials like amorphous to produce charge directly from X-rays, offering higher . These methods allow immediate preview and eliminate chemical processing. Analog processing concludes with drying the film after washing to yield the final radiograph, while digital processing involves several computational steps to enhance usability. Raw digital signals undergo amplification to boost weak detections, followed by analog-to-digital conversion. Histogram equalization redistributes pixel intensities to improve contrast by expanding the dynamic range, particularly in low-contrast regions like soft tissues. Noise reduction techniques, such as Gaussian filtering or transforms, suppress quantum and electronic without significantly blurring edges. Image quality in X-ray acquisition is evaluated through metrics like , which measures the difference in between adjacent structures; , referring to overall blackness in analog films or value averages in digital images; and , quantified in line pairs per millimeter (lp/mm), where higher values indicate sharper detail separation, typically 2-5 lp/mm for diagnostic systems. These metrics ensure diagnostic utility, with digital systems often achieving superior and through post-processing adjustments.

Applications

Medical diagnostics

X-ray machines are extensively used in medical diagnostics for imaging bones, chest, and soft tissues via plain ; real-time imaging in for procedures like ; and cross-sectional images in computed (CT) for detailed anatomy. Specialized systems include for breast cancer detection and dental X-rays for oral health assessment. These applications rely on varying X-ray energies (typically 40-150 ) to balance image quality and radiation dose.

Security and nondestructive testing

X-ray machines play a crucial role in security screening by enabling the non-invasive detection of concealed threats in , , and personnel at airports and facilities. Dual-energy X-ray systems are widely employed in scanners to differentiate materials based on their , distinguishing organic substances like explosives or drugs from inorganic such as metals. These systems operate by emitting X-rays at two distinct levels—typically low- and medium-energy levels in the keV , such as 140 kV and 160 kV—and analyzing the patterns to compute effective atomic numbers (Z_eff) and densities, allowing operators to identify potential hazards without opening containers. In (NDT), X-ray machines facilitate the inspection of such as welds and to detect internal flaws without compromising structural integrity. For weld inspection, radiographic testing uses X-rays to reveal defects like cracks, voids, or inclusions in pipeline welds, ensuring compliance with safety standards during construction and maintenance. Real-time enhances this process by providing dynamic imaging through fluorescent screens and digital detectors, enabling immediate visualization of flaws in pipelines via continuous X-ray exposure at low doses, which supports high-throughput applications in industrial settings. Key techniques in security applications include and imaging, each suited to different material properties. Transmission imaging passes X-rays through the object to a detector on the opposite side, effectively highlighting high-density materials like metals by their strong , making it standard for screening. In contrast, reflects X-rays off the object back to a nearby detector, excelling at detecting low-density organic materials such as plastics or explosives due to enhanced from light elements, and requiring access from only one side for personnel or vehicle screening. To aid threat identification, many X-ray security scanners employ color-coding based on material composition derived from attenuation data. Organic materials with low atomic numbers, such as explosives or fabrics, appear in orange, while high atomic number metals like weapons show in blue; intermediate densities are rendered in green, and very dense areas in black, allowing screeners to quickly assess risk levels without detailed analysis. The primary benefits of X-ray machines in these contexts are their non-contact nature and rapid processing capabilities, which minimize physical handling and expedite security checks. For instance, in screening, -based systems process up to 1000 bags per hour with automated threat detection, reducing false alarms and secondary inspections. In personnel screening, units enable quick full-body scans without pat-downs, enhancing throughput for millions of daily travelers while maintaining through image anonymization.

Industrial and scientific uses

X-ray machines play a crucial role in industrial , particularly for detecting defects in manufactured components such as castings and . In the automotive and industries, X-ray radiography is employed to identify internal flaws like voids, cracks, or inclusions in metal castings without disassembling the parts, ensuring structural integrity before assembly. For , X-ray laminography—a technique that generates layered images of circuit boards—allows precise inspection of solder joints and hidden components for defects such as bridging or misalignment, which is essential in high-reliability sectors like production. In scientific research, X-ray machines are fundamental to and materials analysis. X-ray diffraction () uses X-rays to probe the atomic and molecular structure of crystals, enabling scientists to determine parameters and phase compositions in materials like pharmaceuticals and semiconductors, as pioneered in the work of and the Braggs in the early 20th century. Additionally, X-ray fluorescence () spectroscopy, which excites atoms with X-rays to emit characteristic fluorescent radiation, facilitates non-destructive elemental composition analysis of samples ranging from geological specimens to alloys, providing quantitative data on trace elements with detection limits down to parts per million. Beyond manufacturing and core sciences, X-ray machines support diverse applications in and preservation. In the , inline X-ray systems scan packaged products to detect contaminants like metal fragments, glass shards, or bone in items such as or nuts, enhancing safety compliance with standards from organizations like the FDA. For art restoration, portable X-ray equipment reveals underdrawings, alterations, or layers in paintings and sculptures, aiding conservators in authenticating works and planning non-invasive treatments, as demonstrated in analyses of masterpieces by institutions like the Getty Conservation Institute. Specialized X-ray machines, such as those equipped with microfocus tubes, enable high-resolution () for detailed 3D imaging of small components. These tubes produce a finely focused beam with spot sizes as small as 1-5 micrometers, allowing sub-millimeter resolution in scans of intricate parts like turbine blades or microchips, which supports advanced and in .

Safety and Regulations

Radiation hazards and protection

X-ray machines produce , which can pose health risks through effects like increased cancer risk from DNA damage at low doses, and deterministic effects such as skin burns or cataracts at high acute doses. Protection follows the ALARA (As Low As Reasonably Achievable) principle, minimizing exposure via reducing exposure time, increasing distance from the source (), and using shielding materials like lead aprons, walls, or collimators. Occupational workers wear personal dosimeters (e.g., thermoluminescent dosimeters) to cumulative dose, with annual limits of 20 mSv averaged over 5 years (not exceeding 50 mSv in any year) for workers, and 1 mSv for the public, per international standards. Facilities require area and interlocks to prevent unintended exposures.

Operational standards and guidelines

Operational standards and guidelines for X-ray machines ensure safe and effective use by establishing regulatory frameworks, operator training requirements, protocols, and emergency response measures. These standards are developed by and national bodies to minimize while maintaining diagnostic accuracy across medical, industrial, and security applications. The (IAEA) provides comprehensive safety standards for the operation of X-ray equipment, emphasizing in both medical and industrial settings. IAEA Safety Standards Series No. SSG-46 outlines requirements for diagnostic radiology, including the safe operation of X-ray generators and image-guided procedures, with a focus on occupational exposure control. In the United States, the (FDA) enforces performance standards under 21 CFR Part 1020, which specifies technical requirements for diagnostic X-ray systems and their components, such as tube housing assemblies, controls, and high-voltage generators, to limit leakage and ensure beam quality. These regulations mandate labeling warnings on equipment, such as "This X-ray unit may be dangerous to and unless safe exposure factors, operating instructions, and maintenance schedules are observed," to promote adherence during use. Operator training and certification are critical components of these standards, requiring radiographers to demonstrate competency in equipment handling and . In many countries, including the U.S., is provided by organizations like the American Registry of Radiologic Technologists (ARRT), which mandates completion of an accredited educational program—typically an with clinical training—followed by passing a national examination with a minimum score of 75%. Certified operators must renew their credentials biennially through 24 hours of , covering topics like and protocols. IAEA guidelines reinforce this by recommending specific training courses for all staff using X-ray , integrated with practices to ensure ongoing proficiency. Maintenance schedules, often outlined in manufacturer manuals and regulatory codes, require periodic inspections—such as quarterly checks for electrical integrity and annual performance verifications—to prevent operational failures. Quality assurance (QA) programs form the backbone of operational reliability, involving routine testing to verify and quality. Acceptance testing for new X-ray machines, conducted prior to clinical use, assesses key parameters like output, alignment, and filtration, as detailed in IAEA's Handbook of Basic Quality Control Tests for Diagnostic . Daily calibrations, performed by technologists, include visual inspections of safety interlocks, exposure reproducibility tests using phantoms, and checks for timer accuracy, following protocols from the American Association of Physicists in Medicine (AAPM) Report No. 4. Ongoing QA, such as monthly for film processors or constellation tests for digital systems, ensures consistent results and compliance with standards like those in 21 CFR 1020.31 for radiographic . These measures prioritize conceptual consistency over exhaustive metrics, focusing on thresholds like a maximum 10% variation in exposure to maintain diagnostic utility. Emergency procedures address malfunctions in systems, such as unintended or control failures. Operators must immediately cease use, secure the area, and notify the radiation safety officer or regulatory authority, as per IAEA Safety Standards Series No. GSR Part 3. For systems involving radioactive sources (e.g., in some applications), additional protocols for spills include isolating the area, donning protective equipment, and containing the spill with absorbent materials before professional , while for spread. These protocols, aligned with FDA requirements under 21 CFR 1020.30, emphasize rapid response to mitigate risks without delving into biological effects.

Modern Developments

Digital and computed radiography

Computed radiography (CR) represents an early digital alternative to traditional film-screen systems, utilizing photostimulable phosphor plates to capture images. Introduced by in 1983, CR systems store energy in the form of a within the material, typically fluorohalide doped with (BaFBr:Eu²⁺). These plates are exposed to during imaging, trapping electrons in higher energy states, and then scanned by a beam in a raster pattern to release the stored energy as visible light, known as . The emitted light is detected by a and converted into an electrical signal, which is digitized to form the image. After readout, the plates are erased using intense white light or additional exposure to remove residual energy, allowing reuse for hundreds of cycles. Digital radiography (DR) advances beyond CR by directly acquiring images without intermediate scanning steps, offering faster processing and higher efficiency. DR systems are categorized into direct and indirect conversion types. Direct conversion detectors employ amorphous selenium (a-Se) as a photoconductor layer, where X-rays generate electron-hole pairs that are collected as electrical charges by an underlying array of thin-film transistors (TFTs). This method provides high due to minimal light spreading. In contrast, indirect conversion systems use a layer, such as cesium iodide (CsI) structured in needle-like columns, to convert X-rays into visible light, which is then captured by an amorphous silicon array coupled to TFTs for charge readout. Wireless DR detectors, introduced in the late 2000s, integrate power and to send images directly to workstations, enhancing portability while maintaining wired . Both and offer significant advantages over analog radiography, including instantaneous image availability for immediate review and reduced patient doses by 20–50% through optimized techniques and higher (DQE). Digital formats enable extensive post-, such as edge enhancement via algorithms, which sharpens boundaries between tissues to improve diagnostic visibility without additional . These capabilities also support digital archiving, via picture archiving and communication systems (PACS), and workflow efficiencies that eliminate chemical . The transition to digital technologies accelerated post-2000, with achieving widespread adoption in hospitals by the early 2000s as a bridge from systems. followed, gaining prominence in the 2010s due to its superior speed and image quality, leading to the phasing out of radiography in many facilities by the early 2020s. This shift has enhanced overall efficiency in radiographic imaging while prioritizing through lower doses and better image manipulation.

Advanced and portable systems

Portable X-ray units have advanced significantly, enabling greater mobility for specialized applications in veterinary and dental fields. Battery-powered handheld devices, such as the EzRay AirVet, weigh under 4 pounds and incorporate rechargeable batteries for extended operation, allowing veterinarians and dentists to perform intraoral without fixed installations. Other portable battery-powered units, like the MinXray TR90+, provide mobility for veterinary use despite weighing around 14 pounds. These systems feature double to minimize and produce high-resolution digital images wirelessly transferable to computers or tablets. For remote inspections, drone-mounted systems like the Pacific NDT DroneX integrate high-energy sources with custom unmanned aerial vehicles, enabling safe evaluation of conductor sleeves without disrupting service or requiring ground access. This innovation supports in inaccessible areas, with wireless image acquisition via tablet for real-time analysis. Integration of (AI) into X-ray systems enhances automation and diagnostic accuracy, with FDA approvals for such technologies accelerating since 2018. AI algorithms, such as those in Aidoc's platform, enable automated anomaly detection in chest X-rays, identifying conditions like or pulmonary emboli with performance comparable to radiologists, thereby aiding in settings. For positioning, GE Healthcare's Critical Care Suite uses AI to detect endotracheal tube placement in chest X-rays, providing alerts for suboptimal positioning to reduce procedural errors. Similarly, United Imaging's uDR Aurora CX incorporates for real-time image quality control and automated adjustments during acquisition, streamlining workflows in clinical environments. By mid-2025, the FDA had authorized over 1,000 AI-enabled devices, with applications comprising the majority, including more than 100 for X-ray interpretation and automation. Hybrid modalities combine traditional with advanced techniques for enhanced visualization, particularly in dental and analysis. Cone-beam computed tomography (CBCT) systems, such as those from NewTom, employ a rotating cone-shaped beam to generate images of dentomaxillofacial structures, offering superior detail for implant planning and pathology detection compared to radiographs. These hybrid units integrate and capabilities in compact designs suitable for dental offices. Low-dose spectral imaging, utilizing dual-energy spectra, improves differentiation by exploiting energy-dependent , as seen in ' systems that distinguish materials like bone and with reduced radiation exposure. This approach enables precise characterization of abnormalities while maintaining doses comparable to conventional . Recent advances in technology leverage (CNT) cathodes to create more compact and efficient sources, significantly impacting mobile systems. CNT-based tubes, developed by companies like Micro-X, operate via field emission at , eliminating the need for bulky heating elements and enabling instantaneous on-off switching. In mobile units, this results in devices such as Micro-X's , weighing around 75 kg—significantly lighter than traditional systems (300-600 kg)—while producing equivalent output, facilitating easier transport for bedside or field use. Such innovations support distributed multi-beam arrays for stationary CT without mechanical rotation, further reducing overall system weight and size.

References

  1. [1]
    Electromagnetic Spectrum - Introduction - Imagine the Universe!
    X-ray photons have energies in the range 100 eV to 100,000 eV (or 100 keV). Gamma-rays then are all the photons with energies greater than 100 keV.
  2. [2]
    [PDF] Production of X-rays
    When electrons hit the anode, they decelerate or brake and emit Bremsstrahlung (meaning braking radiation in ... another, characteristic X-rays are produced.
  3. [3]
    X-ray Imaging - Medical Imaging Systems - NCBI Bookshelf - NIH
    Note that the wavelength of most X-rays lies in the range of 0.01 nm up to 10 nm. This corresponds to an energy range of 100 keV down to 100 eV. Figure 7.1.
  4. [4]
    X-ray Image Production Procedures - StatPearls - NCBI Bookshelf
    Oct 17, 2022 · Current and exposure time are often reported together in the following way: current (mA) x time (s) = mA-seconds (mAs). Changes in mAs affect ...
  5. [5]
    CT Radiographic Techniques | Radiology | SUNY Upstate
    X-Ray Tube Voltage (kV)​​ The increase is greater than the linear relationship that is observed when the mA is changed; doubling of the x-ray tube voltage would ...Missing: anode | Show results with:anode
  6. [6]
    Electronic Sources of Ionizing Radiation
    Because x‐ray production is directly proportional to the atomic number of the target and the accelerating voltage of the device, reducing both variables can ...
  7. [7]
    Medical X-ray Imaging - FDA
    Feb 21, 2023 · Computed tomography (CT), fluoroscopy, and radiography ("conventional X-ray" including mammography) all use ionizing radiation to generate images of the body.Radiography · Pediatric X-ray Imaging · Medical Imaging
  8. [8]
    Radiography | FDA
    ### Summary of Radiography Types and Selection Based on Resolution
  9. [9]
    The Selection of Patients for Dental Radiographic Examinations - FDA
    Jun 20, 2019 · HAND-HELD X-RAY UNITS. Hand-held, battery-powered x-ray systems are available for intra-oral radiographic imaging. The hand-held exposure ...
  10. [10]
    21 CFR 892.1600 -- Angiographic x-ray system. - eCFR
    An angiographic x-ray system is a device intended for radiologic visualization of the heart, blood vessels, or lymphatic system during or after injection of a ...
  11. [11]
    Wilhelm Conrad Röntgen – Biographical - NobelPrize.org
    Learn more about X-rays, which were discovered by Wilhelm Röntgen in 1895. He was awarded the very first Nobel Prize in Physics in 1901 for his work.
  12. [12]
    [PDF] Early History of X Rays - SLAC National Accelerator Laboratory
    Roentgen turned to a new interest in October of 1895: the study of cath- ode rays. In the course of repeating the experiments of Hertz and Lenard, he happened ...
  13. [13]
    Coolidge X-Ray Tube | National Museum of American History
    William David Coolidge (1873-1975), an American physicist working for General Electric, introduced a new hot cathode x-ray tube in 1913.
  14. [14]
    William D Coolidge | Radiology Reference Article | Radiopaedia.org
    Nov 26, 2020 · The Coolidge tube rapidly replaced all other x-ray tubes on the market, although gas tubes remained in use for some niche applications for many ...Early life · Coolidge tube: a revolution in... · Later life
  15. [15]
    Coolidge X-Ray Tubes | Museum of Radiation and Radioactivity
    This reflects the fact that William Coolidge's inspiration for the hot cathode X-ray tube was his earlier work developing tungsten filaments for incandescent ...
  16. [16]
    History of Radiography - NDE-Ed.org
    In 1922, industrial radiography took another step forward with the advent of the 200,000-volt X-ray tube that allowed radiographs of thick steel parts to be ...
  17. [17]
    History of X-rays - 125 years in the making (pt 2) - Excillum
    Nov 8, 2020 · As in all future X-ray tube developments, the Coolidge tube allowed for improved X-ray quality and quantity for medical and other applications.
  18. [18]
    [PDF] x-ray tubes development - iomp history of medical physics
    Apr 29, 2018 · Figure 17 (a) Rotating anode system of a Siemens PantixTMtube, equipped with an all-tungsten high temperature anode disk and produced from 1933 ...
  19. [19]
    How Marie Curie Brought X-Ray Machines To the Battlefield
    Oct 11, 2017 · During World War I, the scientist invented a mobile x-ray unit, called a “Little Curie,” and trained 150 women to operate it.
  20. [20]
    Radiology and radiography in the 1940s
    By 1939, good portable X-ray diagnostic apparatus was available and the British Army was supplied with the MX2. This apparatus could be used for both ...Missing: WWI | Show results with:WWI<|separator|>
  21. [21]
    US Army X-Ray Field Unit | WW2 US Medical Research Centre
    Picker X-Ray Corporation was the sole manufacturer of x-ray units for the United States Army during WW2. Smaller, less complex items were also produced by ...Missing: WWI history
  22. [22]
    Pioneers of radiotherapy - Siemens Healthineers MedMuseum
    Apr 1, 2023 · In the early days of radiotherapy in the 1910s, Siemens developed the “Reform Apparatus.” This device pioneered the use of numerous techniques ...
  23. [23]
    [PDF] Screen film processing systems for medical radiography
    This article traces the evolution of screen-film processing systems, which advanced radiographic imaging. Early developments included the gelatin silver ...
  24. [24]
    [PDF] film-screen radiography receptor development a historical perspective
    1930s. DuPontintroduced an x-ray film with blue tinted base to improve viewing comfort that became the standard type of film throughout the industry.
  25. [25]
    Evolution of radiation protection for medical workers - PMC - NIH
    The use was further precipitated by the 1913 development at the Research Laboratory of the General Electric Company (Schenectady, NY) of the hot-cathode X-ray ...Missing: 1900s | Show results with:1900s
  26. [26]
    Radiology in the 1920s
    The radiation protection of the tubes improved dramatically with the use of self-protected tubes however there was still an element of electrical danger before ...
  27. [27]
    A history of radiation shielding of x-ray therapy rooms - PubMed
    In this report the history of shielding for radiation treatment rooms is traced from the time of the discovery of x-rays to the present.
  28. [28]
    Equipment | IAEA
    The principal components of a system for X ray projection radiography are: X ray generator, X tube and housing with collimation system, patient table, anti- ...Missing: accessories shielding
  29. [29]
    Modern Fluoroscopy Imaging Systems
    The key components include an X-ray tube, spectral shaping filters, a field restriction device (aka collimator), an anti-scatter grid, an image receptor, an ...Missing: accessories | Show results with:accessories
  30. [30]
    [PDF] Radiation Protection Guidance for Diagnostic and Interventional X ...
    ... CR and DR) – an x-ray imaging method in which a digital sensor, rather than photographic film, is used to capture an x-ray image. DDR is a cassette-less ...<|separator|>
  31. [31]
  32. [32]
    X-ray Production, Tubes, and Generators - Radiology Key
    May 16, 2021 · This chapter describes the x-ray production process, characteristics of the x-ray beam, x-ray tube design, x-ray generator components, and ...<|control11|><|separator|>
  33. [33]
    X-ray Production - StatPearls - NCBI Bookshelf
    Jul 31, 2023 · Simply understood, the generation of X-rays occurs when electrons are accelerated under a potential difference and turned into electromagnetic ...
  34. [34]
    Basics of X-ray Physics - X-ray production - Radiology Masterclass -
    X-rays are generated via interactions of the accelerated electrons with electrons of tungsten nuclei within the tube anode. There are two types of X-ray ...
  35. [35]
    [PDF] Radiographic For Biolo - Southern Research Station
    A theory proposed in 1938 by Gurney and. Mott to explain the formation of a latent image by silver halide after photon capture. The silver halide crystal is ...
  36. [36]
    Recent Development in X-Ray Imaging Technology - PubMed Central
    The chemical process to capture an X-ray image using a radiographic film involves the formation of a latent image and then developing an X-ray image [29, 30].
  37. [37]
    Modeling the Performance Characteristics of Computed ...
    Computed radiography (CR) that uses photostimulable phosphor (PSP) plates is an important radiographic medical imaging technology [1]. PSP is also referred to ...
  38. [38]
    Computed radiography: photostimulable phosphor image plate ...
    Computed radiography is emerging as a digital imaging modality for use in conventional radiography. It is based on photostimulable phosphor image plate ...Missing: CR | Show results with:CR
  39. [39]
    Using deep learning for pixel-defect corrections in flat-panel ... - NIH
    Purpose: Flat-panel radiography detectors employ thin-film transistor (TFT) panels to acquire high-quality x-ray images. Pixel defects occur due to circuit ...
  40. [40]
    Advances in Digital Radiography: Physical Principles and System ...
    Digital chest radiography with a solid-state flat-panel x-ray ... Dual-exposure technique for extending the dynamic range of x-ray flat panel detectors.
  41. [41]
    The Digital Imaging Workstation - PMC - NIH
    The digital image was processed with a contrast-limited adaptive histogram equalization algorithm (CLAHE), which mapped each 12-bit pixel to an 8-bit value.Missing: fixing amplification
  42. [42]
    Digital radiography image quality evaluation using various ...
    Nov 7, 2022 · To investigate the effect of the exposure parameters on image quality (IQ) metrics of phantom images, obtained automatically using software or from visual ...
  43. [43]
    Effect of Pixel Size on Resolution | Radiology | SUNY Upstate
    The limiting spatial resolution is approximately 3 lp/mm. The radiograph shown above was obtained using a large cassette (35 cm x 43 cm), with a pixel size of ...Missing: metrics contrast density
  44. [44]
    [PDF] Dual-Energy X-ray Radiography and Computed Tomography - OSTI
    Jun 25, 2018 · X-ray CT security scanners are deployed world-wide to detect explosives in checked luggage and are very similar to CT scanners used today in ...
  45. [45]
    A review of airport dual energy X-ray baggage inspection techniques
    It discusses multiple X-ray imaging inspection systems used in airports for detecting dangerous objects inside the baggage. Moreover, it also explains the dual ...
  46. [46]
    [PDF] AI-Enabled Paradigms for Non-Intrusive Screening
    Jun 27, 2024 · The X-ray image acquisition process involves passing baggage through a gantry equipped with an X-ray source and dual-energy detectors. X-ray ...
  47. [47]
    Radiography – RT Inspection Services - MISTRAS Group
    Radiography is a common technique to inspect newly-constructed pipelines to ensure weld quality. MISTRAS provides expert inspection technicians to complete ...
  48. [48]
    What is fluoroscopy and how are image intensifiers used?
    Fluoroscopy, also known as radioscopy, is a technique whereby “real-time” detection of defects is achieved by the use of specialized fluorescent screen ...Missing: pipelines | Show results with:pipelines
  49. [49]
    X-Ray Scanning explained in brief - analyticon
    The X-ray backscatter method differs significantly from the transmission method commonly used in medicine or baggage screening. In transmission, the X-ray ...
  50. [50]
    Using Eye Movements to Understand how Security Screeners ... - NIH
    Jun 4, 2019 · When an orange, green, or blue hue is assigned to each part of the image, it is based on an average of the atomic numbers of the different ...
  51. [51]
    [PDF] Baggage/Cargo/People Screening Program - Homeland Security
    That means fewer secondary inspections of checked and carry-on bags, fewer passenger pat downs, less divestment of personal items at the checkpoint, and reduced ...Missing: non- personnel
  52. [52]
    [PDF] IAEA Safety Standards Radiation Safety in Industrial Radiography
    The overall responsibility for radiation safety lies with the operating organization that is authorized to carry out industrial radiography work. Specific.
  53. [53]
    21 CFR Part 1020 -- Performance Standards for Ionizing Radiation ...
    “Warning: This x-ray unit may be dangerous to patient and operator unless safe exposure factors, operating instructions and maintenance schedules are observed.”.21 CFR 1020.40 · 1020.30 · Title 21 · 1020.33
  54. [54]
    [PDF] IAEA Safety Standards
    This Safety Guide considers the occupational exposure of workers who operate X ray generators and other types of radiation source that are used for inspection ...
  55. [55]
    21 CFR 1020.30 -- Diagnostic x-ray systems and their major ... - eCFR
    C-arm fluoroscope means a fluoroscopic x-ray system in which the image ... (1) Mobile x-ray equipment means x-ray equipment mounted on a permanent base ...
  56. [56]
    ARRT Certification and Registration
    To become certified and registered, candidates must meet certain professional standards in medical imaging or radiation therapy. In the foreground, a nuclear ...
  57. [57]
    Education Requirements-Primary Eligibility Pathway - ARRT
    To meet the primary pathway education requirement, you must have earned an associates degree and have completed an ARRT-approved educational program.
  58. [58]
    [PDF] Performance Standard for Diagnostic X-Ray Systems and ... - FDA
    To which products does this rule (21 CFR 1020.30; 1020.31; 1020.32;. 1020.33) apply? The rule applies to all medical diagnostic x-ray systems and their major.
  59. [59]
    [PDF] Handbook of basic quality control tests for diagnostic radiology
    Routine performance tests for radiography, fluoroscopy, angiography, mammography and computed tomography (CT) are described. The tests require only the use of ...Missing: dental | Show results with:dental<|separator|>
  60. [60]
    [PDF] BASIC QUALITY CONTROL IN DIAGNOSTIC RADIOLOGY - AAPM
    This document describes a quality assurance protocol for diagnostic x-ray equipment at the radiologic technologist level. A series of tests are described using ...
  61. [61]
    Guide for Radiation Safety/Quality Assurance Programs
    This guide describes the type and extent of information and standards by which the New York State Department of Health will evaluate a facility's Radiation ...
  62. [62]
    The Evolution of Digital Radiography | Imaging Technology News
    Feb 3, 2014 · Wireless detectors entered the U.S. market a few years ago, giving radiologists the versatility and mobility of CR detectors and analog film ...
  63. [63]
    Using Edge Enhancement to Identify Subtle Findings on Soft-Copy ...
    Application of edge enhancement to soft copies of neonatal chest radiographs helps radiologists to identify small pneumothoraces, vascular catheters, and ...
  64. [64]
    The History of the Digital X-Ray - Windsor Imaging
    Apr 1, 2020 · By the early 2000s, most hospitals and imaging centers had transitioned from analog film to computer radiography. This form of imaging ...
  65. [65]
    From Films to Digital, The Evolution of Imaging Records
    Jan 21, 2022 · The work to obtain higher-quality images quickly, reliably and safely is a never-ending journey, with steady progress being made by ...
  66. [66]
    EzRay AirVet Handheld Portable Dental X-ray System
    The EzRay AirVet Handheld Portable Dental X-ray System is more than 30% lighter than other handheld devices weighing less than 4 lb.
  67. [67]
  68. [68]
    ImageWorks Veterinary's Handheld X-rays
    The Image-VET FLEXX is a portable, compact, one-handed handheld x-ray system with a high capacity battery, double lead shielding, and low dose.
  69. [69]
    DRONEX - Pacific NDT
    Using a custom drone, we have successfully mounted x-ray imaging machine to safely inspect conductor sleeves on power lines. DroneX™ has been acquiring data ...
  70. [70]
    Pacific NDT DroneX Powerline Inspection System
    The DroneX system successfully uses a mounted x-ray imaging machine to safely inspect conductor sleeves on power lines.
  71. [71]
    A First in the World of Radiology: Aidoc Receives FDA Clearance to ...
    Aug 8, 2018 · Aidoc, a leading provider of AI solutions for radiologists announced today that it was granted Food and Drug Administration (FDA) clearance.
  72. [72]
    Deep learning improves physician accuracy in the comprehensive ...
    Oct 24, 2024 · The FDA-cleared AI system demonstrated strong standalone performance, detecting chest abnormalities on X-ray on par with radiologists, and the ...
  73. [73]
    [PDF] Critical Care Suite with Endotracheal Tube Positioning AI Algorithm
    Oct 29, 2021 · Critical Care Suite is a suite of AI algorithms for the automated image analysis of frontal chest X-rays acquired on a digital x-ray system.
  74. [74]
    United Imaging Announces the uDR ® Aurora CX
    The embedded AI quality control system, which is integrated throughout the entire X-ray procedure, provides real-time image quality control feedback and ...
  75. [75]
    FDA has approved over 1,000 clinical AI applications, with most ...
    Jan 10, 2025 · FDA has approved over 1,000 clinical AI applications, with most aimed at radiology · Dozens of new artificial intelligence-enabled products have ...
  76. [76]
    Dental X-ray devices - Newtom
    Thanks to Cone Beam CT (CBCT) technology, our dental X-ray ... The most comprehensive hybrid CBCT for 2D and 3D imaging and dental: get high definition images.
  77. [77]
    Dental Cone-beam Computed Tomography - FDA
    Sep 28, 2020 · The CBCT systems used by dental professionals rotate around the patient, capturing data using a cone-shaped X-ray beam.
  78. [78]
    Dual Energy CT Spectral Imaging - Siemens Healthineers USA
    Dual Energy CT scanners produce image sets with two different X-ray spectra to characterize materials with precision while highlighting abnormalities in the ...
  79. [79]
    Spectral Computed Tomography: Fundamental Principles and ... - NIH
    Spectral CT is an emerging technology offering superior material differentiation, which is achieved using the energy dependence of X-ray attenuation in any ...
  80. [80]
    X-ray generation using carbon nanotubes - Nano Convergence
    Jan 6, 2015 · This review provides a summary of the current state-of-the-art in carbon nanotube-based field emission X-ray sources.
  81. [81]
    Are carbon nanotubes the future for digital x-ray systems? | AuntMinnie
    Nov 18, 2024 · The digital CNT x-ray source was about 75% smaller, weighed about 50% less, and consumed 17% less energy producing x-rays of the same energy ...
  82. [82]
    Carbon nanotube-based multiple source C-arm CT system
    Dec 15, 2023 · To extend the field of view while reducing dimensions of the C-arm, we propose a carbon nanotube (CNT)-based C-arm computed tomography (CT) ...