Capsule endoscopy

Capsule endoscopy is a noninvasive technique in which a swallows a small, pill-sized capsule containing a miniature camera, light source, and wireless transmitter to capture thousands of images of the , particularly the , as the capsule travels through the digestive system via natural . Developed in the late and first approved for clinical use by the FDA in , this procedure revolutionized the diagnosis of small bowel disorders by providing high-resolution visualization without the need for invasive scopes or . The technology originated from a 1981 collaboration between Israeli engineer Gavriel Iddan and gastroenterologist Eitan Scapa, who envisioned a imaging device to address the challenges of examining the , which is difficult to access with traditional . Key advancements included the adoption of complementary metal-oxide-semiconductor (CMOS) image sensors for miniaturization and energy efficiency, along with light-emitting diodes (LEDs) for illumination, enabling the capsule to transmit over 50,000 images during an approximately eight-hour journey through the body. Patients wear an external data recorder to capture the signals, which are later compiled into a video for review, with the disposable capsule typically expelled naturally within 24 hours. Primarily used to investigate obscure , inflammatory conditions such as and celiac disease, polyps, ulcers, tumors, and celiac disease, capsule endoscopy offers superior detection rates for small bowel pathologies compared to other noninvasive tests like or MRI enterography. Preparation involves fasting for 12 hours, consuming clear liquids, and sometimes laxatives to ensure optimal visualization, while post-procedure care restricts strenuous activity to avoid dislodging the recorder. Although generally safe, risks include rare capsule retention in narrowed areas (e.g., due to strictures from or tumors), which may necessitate endoscopic retrieval or , occurring in about 1-2% of cases. Results are typically available within a week, guiding further treatment decisions and improving patient outcomes in .

History

Invention and Early Development

The concept of capsule endoscopy originated in 1981 when Israeli electro-optical engineer Gavriel Iddan, during a in from his position at , met gastroenterologist Eitan Scapa. Scapa highlighted the limitations of existing fiber-optic endoscopes in visualizing the small bowel, inspiring Iddan to explore adapting the miniaturized camera technology he had developed for guided missiles at into a swallowable, for non-invasive gastrointestinal imaging. During a second sabbatical in 1991, Iddan revisited the idea, initially proposing a charged-coupled device (CCD) camera tethered by an electrical cord, but soon shifted to a fully wireless capsule concept comprising a miniature camera, external recorder, and image-processing software. By 1993, he had refined the design into a three-part system, and in 1994, with funding from entrepreneur Gavriel Meron, assembled a team of engineers to build prototypes. In 1998, Meron founded Ltd. to further develop and commercialize the technology. These early 1990s efforts focused on shrinking wireless camera components to pill size while ensuring the device could navigate the via natural , capturing and transmitting images externally. Development faced significant hurdles, including limited battery life—initial CCD prototypes allowed only about 10 minutes of operation due to high power consumption—and challenges in achieving sufficient image quality amid the dark, fluid-filled environment of the gut. Wireless transmission feasibility was another key obstacle, as signals needed to penetrate the body reliably without an , compounded by issues like signal strength and potential . These were addressed by 1997 through Iddan's for the system and the adoption of complementary metal-oxide-semiconductor () imagers, which offered lower energy use and better resolution, enabling working prototypes by 1999.

Key Milestones and Approvals

The first commercial capsule endoscope, known as the M2A (mouth-to-anus) system, developed by , received for use in in 2001, marking its entry into clinical practice in the region. Shortly thereafter, in August 2001, the U.S. (FDA) granted clearance for the M2A capsule as an adjunctive diagnostic tool for evaluating suspected small bowel disease, particularly obscure , enabling its adoption in American medical centers. These approvals followed the initial prototype development by engineer Gavriel Iddan and gastroenterologist Eitan Scapa in the late 1990s. The first in-human trial was successfully conducted in October 1999 by gastroenterologist Paul Swain. Early clinical validation came through pivotal studies in the early 2000s that demonstrated the device's efficacy for obscure . A 2002 pilot study by and Swain involving 20 patients with suspected small intestinal found that capsule endoscopy provided excellent visualization of the small bowel, identifying lesions in 11 cases where traditional methods failed, with high patient tolerability and safety. This was expanded in a 2004 multicenter study by Pennazio et al., which examined 100 consecutive patients with obscure and reported a diagnostic of 66% for identifying potential bleeding sources, significantly higher than prior endoscopic techniques, thus solidifying capsule endoscopy's role in clinical guidelines. Post-approval expansion occurred rapidly, with the technology gaining regulatory clearance in additional regions including , , and parts of by the mid-2000s, facilitating global adoption for small bowel evaluation. In 2014, Covidien acquired Given Imaging for approximately $860 million, integrating the PillCam brand—renamed from M2A—into its portfolio and enhancing distribution. Covidien itself was subsequently acquired by in 2015, further supporting the continued evolution and market penetration of capsule endoscopy products.

Technology

Device Design and Components

Capsule endoscopy devices are typically designed as small, ingestible capsules resembling pills, with dimensions of approximately 11 mm in diameter and 26 mm in length to facilitate easy swallowing and passage through the . The exterior consists of a biocompatible shell that encases all internal components, ensuring safety and durability during transit through the digestive system. This material choice minimizes risks such as allergic reactions or irritation while protecting the from bodily fluids. The core hardware includes an optical dome at one end, which serves as a transparent window for , paired with a complementary metal-oxide-semiconductor () image that captures high-resolution video. Illumination is provided by multiple white light-emitting diodes (LEDs) arranged around the system to evenly light the surrounding mucosa, enabling clear visualization in the low-light environment of the intestines. The assembly focuses light onto the , typically achieving a between 140° and 170° for single-camera models. Power is supplied by a compact , which sustains operation for 8 to 12 hours, sufficient to traverse the small bowel in most patients. transmission occurs via an onboard wireless transmitter (such as radio frequency (RF) or human body communication (HBC)) that sends images to an external worn by the patient, or in some designs, images are stored in internal memory for later retrieval. Design variations exist to enhance coverage, particularly with some advanced models, such as the , incorporating four laterally positioned cameras for a 360° panoramic view of the intestinal walls, reducing blind spots compared to traditional single-camera configurations. These multi-camera systems maintain similar overall dimensions but adjust internal layouts to accommodate additional sensors and optics.

Imaging and Data Processing

Capsule endoscopy systems capture images of the using miniaturized image sensors illuminated by LEDs, producing a continuous stream of visual data as the capsule travels through the digestive system. Modern devices, such as the PillCam SB3, operate at an adaptive of 2 to 6 frames per second, adjusting based on the capsule's to optimize coverage without excessive . has advanced to 320 × 320 pixels, providing clearer visualization of mucosal surfaces compared to earlier models. The raw images are processed onboard the capsule before wireless transmission to minimize requirements and power consumption. Key steps include color enhancement techniques, such as adaptive functions, to accentuate subtle variations and improve for better detection. White balance adjustments are applied to correct illumination inconsistencies from the LED sources, ensuring natural color reproduction of the mucosa. Compression algorithms, often based on variants adapted for color filter arrays, reduce file sizes while preserving diagnostic quality, enabling efficient handling of the high-volume data stream. Transmitted via signals, the processed images are received by an external data recorder strapped to the patient's or , which stores the full for later review. A standard 8- to 12-hour procedure generates approximately 50,000 to 100,000 images, depending on the and transit time, with the recorder utilizing solid-state memory to manage this volume without loss. This setup allows for real-time monitoring of signal strength but defers detailed analysis until post-procedure downloading to a .

Procedure

Patient Preparation and Ingestion

Patients undergoing capsule endoscopy must follow specific pre-procedure instructions to ensure optimal visualization of the and safe ingestion of the device. Typically, individuals are required to fast for at least 12 hours prior to the , abstaining from all food and drink to empty the . On the day before the examination, a clear is often prescribed after noon, consisting of transparent fluids such as , clear , or , while solid foods are avoided. Additionally, patients may need to discontinue certain medications, such as iron supplements, at least seven days in advance to prevent interference with capsule transit or imaging quality. Bowel preparation is a key component of pre-procedure readiness to cleanse the and enhance image clarity. Guidelines recommend the use of 2 liters of (PEG)-based laxatives, often administered as a split dose the evening before the procedure, which has been shown to improve small bowel visualization quality and diagnostic yield without affecting completion rates. Simethicone may be added to the regimen to reduce gas bubbles and further optimize views, though prokinetics are not routinely advised due to lack of benefit. Patients with swallowing disorders or risk factors like advanced age should undergo a swallow evaluation, such as bedside assessment with monitoring, to identify and mitigate risks, which occur in less than 2% of cases but can be higher in vulnerable populations. The of the capsule occurs in a clinical setting under medical supervision to monitor for immediate complications. The patient swallows the vitamin-sized capsule, which measures approximately 26 mm by 11 mm and features a slippery , while upright and with a of to facilitate passage into the . A standardized may involve initial positioning checks using real-time imaging to confirm proper entry into the digestive tract, particularly for those with known difficulties, where alternative placement via can be considered if needed. Following successful ingestion, patients are advised to remain upright briefly to aid transit and avoid strenuous activity that could dislodge recording equipment.

Post-Procedure and Data Analysis

After ingesting the capsule, patients typically resume normal daily activities for approximately 8 to 12 hours while wearing an external data recorder attached to the via a or sensors, which captures the transmitted images. During this period, patients are advised to avoid strenuous exercise, bending over repeatedly, or exposure to strong electromagnetic fields, including (MRI) scans or powerful magnets, to prevent interference with data transmission. The disposable capsule is naturally excreted from the body via , usually within 24 to after , though patients are often instructed to check their to confirm and avoid scheduling MRI examinations until is verified. If the capsule is not observed in the stool within this timeframe, an may be performed to assess its location and rule out retention. Once the recording period ends, the external recorder is returned to the medical facility, where the captured data—comprising thousands of images—is downloaded to a dedicated using provided by the device manufacturer. This software enables sequential viewing of the images in a video-like format, allowing gastroenterologists to perform a preliminary by scrolling through frames, identifying key anatomical landmarks, and annotating notable findings for further clinical . The typically involves an initial rapid overview to assess transit completeness before detailed examination.

Medical Uses

Indications for Small Bowel Evaluation

Capsule endoscopy serves as the primary diagnostic tool for evaluating obscure (OGIB), defined as persistent or recurrent bleeding of unknown origin following negative bidirectional endoscopy ( and ). This includes both occult bleeding, characterized by without visible blood loss, and overt bleeding with visible hemorrhage. The procedure is recommended as the first-line imaging modality for hemodynamically stable patients with suspected small bowel sources, offering a diagnostic of approximately 60% for identifying clinically significant lesions such as angioectasias, ulcers, or erosions. Compared to standard upper and lower endoscopies, capsule endoscopy detects additional bleeding sources in 20-30% of cases, particularly vascular lesions that account for up to 40% of findings. In patients with , particularly , capsule endoscopy is indicated for assessing small bowel involvement when symptoms suggest ileal or proximal disease despite inconclusive ileocolonoscopy or cross-sectional imaging. It excels at visualizing mucosal abnormalities, including aphthous and deep ulcers, strictures that may cause obstructive symptoms, and fistulas connecting bowel segments or to adjacent organs. These findings aid in confirming the diagnosis, monitoring disease activity, and guiding therapeutic decisions, such as escalating anti-inflammatory therapy. The Lewis Score or Capsule Endoscopy Activity Index is often used to quantify severity, with scores >135 indicating significant inflammation. Guidelines recommend its use in suspected small bowel after negative conventional endoscopy, though patency assessment is advised in patients with known strictures to prevent retention. Capsule endoscopy also plays a role in screening for small bowel tumors in high-risk populations, such as those with familial polyposis syndromes or a history of , where it can detect polyps, gastrointestinal stromal tumors, or lymphomas that are often on other . In celiac disease, it is employed to evaluate complications in refractory cases, identifying ulcerative jejunoileitis, , or small bowel . These applications enhance early detection, with yields for neoplastic lesions ranging from approximately 3% to 9% in patients with suspected bleeding, though routine use is reserved for symptomatic or high-risk individuals due to cost and invasiveness considerations.

Applications in Other Areas

Capsule endoscopy has been adapted for esophageal evaluation through specialized devices, such as the PillCam ESO, which enable non-sedated imaging of the esophagus for screening conditions like and . In patients with , esophageal capsule endoscopy (ECE) demonstrates moderate diagnostic accuracy for detecting , with pooled sensitivity of 77% and specificity of 86%, though it is not recommended as a replacement for (EGD) due to limitations in capability. For in cirrhotic patients, ECE offers feasible screening with pooled sensitivity of 83% and specificity of 85%, providing a less invasive alternative when EGD is not tolerated, but it requires further validation for routine use. Colon capsule endoscopy (CCE), particularly the second-generation PillCam Colon 2, extends the technology to colorectal cancer screening by visualizing the entire colon after bowel preparation and prokinetics. In prep-optimized patients, CCE achieves high sensitivity for polyp detection, ranging from 80% to 90% for lesions ≥6 mm, making it a viable option for average-risk individuals who decline traditional colonoscopy. Studies confirm its effectiveness in identifying significant findings, with completion rates of 86% to 92% and comparable polyp miss rates to colonoscopy in back-to-back evaluations. Emerging applications include pediatric , where capsule endoscopy aids in diagnosing obscure , suspected , and other small bowel pathologies in children, with diagnostic yields up to 58% in cases with negative prior imaging. Safety profiles are favorable, though retention risks necessitate patency assessments, particularly in younger patients with strictures. In post-surgical settings, such as after resection for , capsule endoscopy effectively monitors recurrence by detecting active lesions in 78% of patients shortly after and tracking progression over months, offering a noninvasive baseline for longitudinal assessment without the need for repeated invasive procedures.

Manufacturers and Products

Leading Companies

, which acquired (the acquirer of ) in 2015, has established itself as the dominant force in the capsule endoscopy market with the PillCam system. originally received FDA approval for the PillCam in 2001, marking the introduction of the first commercial wireless capsule endoscope. Since then, PillCam has become the market leader, with over 4 million units utilized in procedures worldwide as of 2021, enabling widespread adoption for small bowel visualization. CapsoVision, a U.S.-based company founded in 2005 and headquartered in , has carved out a niche by prioritizing omnidirectional imaging technologies in capsule endoscopy. The company's innovations focus on comprehensive 360-degree views of the , distinguishing its contributions to non-invasive diagnostics. IntroMedic, a South Korean firm established in , has played a key role in advancing capsule endoscopy through its MiroCam system, which emphasizes high-resolution and real-time viewing capabilities. Since entering the market, IntroMedic has contributed to global accessibility by developing systems that support efficient, detailed examination of the small bowel. Other leading companies include , which offers the EndoCapsule system for small bowel imaging, and Chongqing Jinshan Science & Technology (Group) Co., Ltd., known for the MicroCam capsule, both contributing to market competition and innovation as of 2025.

Available Devices and Features

Several commercial capsule endoscopy devices are available for small bowel evaluation, each offering distinct features tailored to enhance visualization and procedural efficiency. The , developed by , features a single camera for imaging of the small bowel mucosa. It captures images at a resolution of 320 × 320 pixels with an adaptive of 2 to 6 frames per second (fps), depending on capsule movement, and provides a life of at least 8 hours to support extended transit through the . In contrast, the CapsoCam Plus from CapsoVision provides a panoramic 360° view using four laterally mounted cameras and 16 light-emitting diodes for illumination, eliminating blind spots in small bowel assessment. This device incorporates 1.5 GB of onboard storage, allowing image capture without an external data recorder and enabling patient mobility during the procedure; it operates at up to 20 total (approximately 5 fps per camera) and received FDA clearance in 2016 for adult use, with subsequent expansions for pediatric applications in patients aged two and older. The MiroCam system by IntroMedic includes options like the MC1600 model with a constant frame rate of 6 and a 170° per camera, supporting detailed small bowel imaging over an operational duration of up to 12 hours. A key differentiator is the optional magnetic navigation feature in the MiroCam variant, which uses an external handheld to guide capsule orientation and position, particularly useful for targeted views. Data from the capsule is transmitted via human body communication during recording and can be transferred post-procedure using for efficient workflow integration.
DeviceCamerasResolutionFrame RateBattery LifeUnique Features
PillCam SB3 (Medtronic)1320 × 320 pixels2–6 fps (adaptive)≥8 hoursWide-angle 156° view; external recorder required
CapsoCam Plus (CapsoVision)4 (panoramic)High-resolution (specific pixels not disclosed)Up to 20 fps totalUp to 15 hoursOnboard storage; no external hardware; FDA-cleared 2016
MiroCam (IntroMedic)1 or 2 (model-dependent)320 × 320 pixels6 fps (constant)Up to 12 hoursMagnetic navigation option; data transfer

Safety and Risks

Common Side Effects

Capsule endoscopy is generally a safe and well-tolerated procedure, with the majority of patients experiencing minimal or no adverse effects beyond those associated with bowel . Common side effects are typically mild and transient, arising from the physical presence and movement of the capsule through the . Abdominal discomfort or is frequently reported, often attributed to the capsule's stimulating bowel or causing temporary distension. These symptoms usually subside as the capsule progresses and do not require intervention. Nausea or mild cramping can occur shortly after capsule , particularly during the initial gastric retention phase or early small bowel transit, and generally resolves within a few hours without specific treatment. Such effects are linked to the body's natural response to the foreign object and are more common in patients sensitive to gastrointestinal stimulation. Patients are advised to rest and hydrate during this period to alleviate discomfort. Although uncommon, allergic reactions to the capsule's biocompatible materials, such as the outer shell or internal components, have been documented in isolated cases. These may manifest as mild skin irritation, , or, rarely, more systemic responses, and are managed through close and symptomatic care if they arise. Pre-procedure screening for material sensitivities is recommended to minimize this risk.

Retention and Management

Capsule retention represents the most significant complication associated with capsule endoscopy, defined as the failure of the device to pass through the and be excreted within two weeks of . The overall retention rate is approximately 1-2% as of recent meta-analyses, though it rises to 2-5% in patients with known and can reach 13-21% in cases of suspected small bowel strictures. Key risk factors for retention include known gastrointestinal obstructions, strictures, inflammatory conditions such as , and (NSAID)-induced ulcers leading to diaphragm disease. Absolute contraindications encompass swallowing disorders that impair safe ingestion of the capsule, suspected or confirmed small bowel obstructions, and due to limited safety data. Management strategies prioritize prevention and to minimize harm. Patency capsules, such as the PillCam patency system, are used for pre-procedure testing to assess patency by dissolving if a stricture is present, thereby reducing retention risk in high-risk patients. If retention occurs, retrieval options include endoscopic when feasible, particularly in the or proximal small bowel, or surgical for cases lodged in strictures or obstructions. Adverse events like retention must be reported to the U.S. (FDA) via the Manufacturer and User Facility Device Experience (MAUDE) database to ensure ongoing device safety monitoring.

Advancements and Future Directions

AI Integration and Recent Innovations

has significantly enhanced capsule endoscopy by enabling automated detection of lesions such as ulcers and polyps, improving diagnostic accuracy and efficiency. models, particularly convolutional neural networks (CNNs), have achieved sensitivities exceeding 90% for identifying ulcers and erosions in small bowel capsule endoscopy (SBCE), with one multicenter validation study reporting 90.2% and 84.4% specificity across devices like PillCam SB3 and Olympus EC-10. For lesions, meta-analyses of AI-assisted systems show pooled sensitivities of 94% and specificities of 97%, facilitating precise identification while minimizing inter-observer variability. These models also reduce review times substantially; for instance, AI-assisted reading shortened analysis from 57.2 minutes to 9.0 minutes per exam, representing an approximately 84% reduction, though typical efficiencies range from 50-70% in clinical workflows by prioritizing relevant frames. Such advancements streamline high-volume screening, with systems like NaviCam ProScan demonstrating 99.9% for lesions and cutting reading time to 3.8 minutes. Recent developments in 2024-2025 have focused on -based ecosystems for analysis in capsule , with several FDA clearances marking progress toward integrated . AnX Robotica's NaviCam ProScan, cleared by the FDA in January 2024, provides -assisted reading for suspected small bowel , enhancing detection across pediatric and adult patients while supporting integration for workflow efficiency. Similarly, CapsoVision's CapsoCam Plus , leveraging -based , received FDA clearance expansions in 2025 for pediatric use and enables remote procedure tracking, reducing on-site review burdens. These tools represent a shift toward scalable, diagnostics, with Olympus's broader OLYSENSE ecosystem—launched in 2025—incorporating for applications that complement capsule systems through connectivity, though primarily validated for detection. Innovations in therapeutic capsules include prototypes for , expanding capsule endoscopy beyond diagnostics. Robotic capsules equipped with magnetic actuation, such as those proposed by Yim et al., integrate permanent magnets for locomotion and precise drug release at sites like the , using external fields to trigger injection for conditions including . The SonoCAIT prototype, first developed in 2017, employs technology for site-specific release in the lower GI tract, addressing by minimizing systemic exposure. High-resolution techniques further support these applications; for example, shape-from-shading algorithms applied to single capsule images generate accurate colon models, improving lesion characterization with minimal distortion even from low-quality footage. Stereo camera-enabled capsules achieve detailed 3D mapping of subepithelial tumors, enhancing surgical planning.

Emerging Technologies and Clinical Trials

Recent advancements in capsule endoscopy have focused on enhancing navigational through magnetic and tethered systems, addressing limitations in passive capsule transit. Magnetic capsule endoscopy (MCE) employs external magnets to the device, improving visualization of the small bowel and by reducing gastric transit time and increasing completion rates. A prospective randomized demonstrated that magnetic significantly enhances small bowel capsule endoscopy completion rates by facilitating faster pyloric passage, with successful transpyloric achieved in 59% of cases and a pyloric transit time reduced compared to non-steered capsules. Similarly, a 2025 systematic and of six studies reported an of 1.29 for improved completion rates with magnetic guidance, defined as successful visualization of the before battery depletion, though the increase was not statistically significant across all cohorts. Tethered capsules, connected by a thin , enable real-time and retrieval, allowing for targeted imaging of the upper without full ingestion risks. In a clinical , tethered capsule endomicroscopy achieved a success rate of 89% in unsedated patients, providing high-resolution microstructural images of the and with minimal discomfort. Ongoing multicenter clinical trials from 2024 to 2025 are evaluating these technologies alongside diagnostic enhancements, including -assisted detection of small bowel lesions. A 2024 multicenter study validated an model for identifying and across various small bowel capsule endoscopy devices, achieving superior diagnostic performance with reduced reading times compared to standard-of-care methods, though specific device comparisons like CapsoCam versus PillCam were not detailed. For instance, integration with the PillCam Crohn’s in earlier foundational work showed 92.4% overall accuracy for and detection, with 97.1% sensitivity, informing recent trial designs focused on severity scoring. Additionally, trials are exploring cloud-based platforms for remote video analysis and home delivery models to streamline patient access. A 2024 on home-delivered colon capsule endoscopy using 5G-enabled smartboxes reported an 88% patient satisfaction rate and 96% completion rate, with real-time consultations facilitating follow-up without hospital visits, potentially reducing healthcare burdens. Emerging research compares capsule endoscopy to invasive alternatives like (DBE), particularly in pediatric and (IBD) populations, highlighting reduced procedural risks. In pediatric very early-onset IBD, a multi-institutional study of 82 children under 6 years old found small bowel capsule endoscopy feasible and safe, detecting abnormalities in 42% of cases with delivery via gastroscope, outperforming magnetic resonance enterography for mucosal detail while avoiding the and technical challenges of DBE, which requires a minimum weight of about 11 kg. These trials underscore capsule endoscopy's advantage in minimizing needs—often performed with intravenous or general in only 35.6% to 64.4% of pediatric cases—compared to DBE's routine deep requirements, enabling broader application in young patients with suspected small bowel involvement in IBD.

References

  1. [1]
    Capsule endoscopy camera - Mayo Clinic
    Capsule endoscopy is a procedure that uses a tiny wireless camera to take pictures of the organs in the body that food and liquids travel through.
  2. [2]
    Capsule endoscopy: MedlinePlus Medical Encyclopedia
    Aug 8, 2023 · It is about the size of a large vitamin pill. The person swallows the capsule, and it takes pictures all the way through the digestive ( ...
  3. [3]
    The history of time for capsule endoscopy - PMC - NIH
    Capsule endoscopy was conceived by inventive minds of good people. In the beginning there was a will to do something for medicine.
  4. [4]
    Capsule Endoscopy - Johns Hopkins Medicine
    Capsule endoscopy is a noninvasive diagnostic procedure to visualize the inside of your digestive tract. You swallow a capsule that contains a tiny camera, a ...
  5. [5]
    Pillcam's inventor regrets sale of 'Biblical' tech to foreign firm
    Apr 23, 2015 · It was really a matter of “shrinking” the missiles that Iddan was working with to the size of a small, ingestible pill with a camera attached – ...
  6. [6]
    M2A capsule endoscopy. A breakthrough diagnostic tool for small ...
    The M2A Capsule Endoscope received FDA clearance in August 2001 and is currently in use at major medical centers throughout the United States and Europe.
  7. [7]
    Capsule endoscopy in the evaluation of patients with ... - PubMed
    This pilot study demonstrates that capsule endoscopy provides excellent visualization of the small intestine, is well tolerated by patients, and is safe.
  8. [8]
    Outcome of patients with obscure gastrointestinal bleeding after ...
    Outcome of patients with obscure gastrointestinal bleeding after capsule endoscopy: report of 100 consecutive cases. Gastroenterology. 2004 Mar;126(3):643-53.
  9. [9]
    Covidien Completes Acquisition of Given Imaging - Medtronic
    Feb 27, 2014 · Covidien has acquired all outstanding Given shares for $30 per share or an aggregate consideration of approximately $860 million, net of cash and short-term ...
  10. [10]
    Covidien buys camera-in-a-pill maker Given Imaging for $860 mln
    Dec 8, 2013 · Medical device maker Covidien PLC will acquire Israeli camera-in-a-pill maker Given Imaging for $30.00 per share in cash, or a total of $860 ...
  11. [11]
    Endoscope Capsules: The Present Situation and Future Outlooks
    The concept of a capsule indicated for small bowel analysis was developed by two groups. In 1996, a gastroenterologist named Dr. Paul Swain demonstrated the ...<|control11|><|separator|>
  12. [12]
    PillCam™ SB 3 System | Medtronic (NZ)
    Physical Properties ; Length: 26.2 mm. Diameter: 11.4 mm · 3.0 g · Biocompatible plastic ...
  13. [13]
    Capsule Endoscopy - StatPearls - NCBI Bookshelf - NIH
    The capsules range in size but are approximately 1.1 by 1.6 cm in size. ... capsule endoscopy and is the most common complication. Capsule retention is ...
  14. [14]
    Capsule endoscopy: Present status and future expectation - PMC
    “Capsule” has components of camera including optical dome, light source, lens, a complementary metal oxide semiconductor imager (CMOS), battery, and a wireless ...
  15. [15]
    [PDF] Analysis of current and future technologies of capsule endoscopy
    Jul 17, 2020 · A typical capsule endoscope contains several primary components; the external casing, optical window, LED array, optical lens, CMOS image sensor ...
  16. [16]
    Capsule Endoscopy - Abdominal Key
    Jul 31, 2022 · The small bowel capsule (Figure 9.1) (PillCam SB3, Medtronic) is cylindrical and measures 11 mm in diameter and 26 mm in length. It has one ...Missing: dimensions | Show results with:dimensions
  17. [17]
    A randomized trial comparing wireless capsule endoscopy with ...
    The capsules contained a complimentary metal oxide silicon (CMOS) chip camera, transmitter, light-emitting diode (LED) illumination, and silver oxide batteries ...Missing: components | Show results with:components
  18. [18]
    Emerging Issues and Future Developments in Capsule Endoscopy
    The CE has a battery life of 15 hours and captures images at 20 fps for the first 2 hours of operation and 12 fps for the duration of the procedure (12).
  19. [19]
    Capsule designs with various camera configurations. We have...
    d) A panoramic design consisting of four cameras placed on the sides of capsule robot body with 90°rotational offset to each other, resulting in a 360°view.
  20. [20]
    PillCam™ SB 3 Capsule Endoscopy System - Medtronic
    PillCam™ SB 3 capsule is a camera inside a capsule that captures images at a frame rate of two to six frames per second (fps) based on capsule speed as it ...Missing: resolution | Show results with:resolution
  21. [21]
    Capsule Endoscopy: Current Trends, Technological Advancements ...
    Jun 4, 2025 · It has a 10-h battery life and is used for evaluating the stomach, duodenum, colon, and small intestine. The weight of this capsule is 4.5 g, ...
  22. [22]
    Product Review: Medtronic PillCam™ Capsule - TechSci Research
    The third-generation system, PillCam™ SB3 offers a 30% improvement in image resolution and has the capacity to increase number of photographic frames from 2 to ...Missing: specifications | Show results with:specifications
  23. [23]
    Color Enhancement in Endoscopic Images Using Adaptive Sigmoid ...
    Wireless capsule endoscopy (WCE) is an established methodology that offers ... It has been used in several researches related to image processing [25–27].
  24. [24]
    Consistency and Standardization of Color in Medical Imaging
    Jul 9, 2014 · Conditions impacting color quality include lighting, camera settings (e.g., focus, flash, and white balance), compression, and views. Image ...
  25. [25]
    Compression for Bayer CFA Images: Review and Performance ... - NIH
    These compression methods can be roughly divided into the compression-first-based (CF-based) scheme and the demosaicing-first-based (DF-based) scheme.
  26. [26]
    Robotic wireless capsule endoscopy: recent advances and ...
    May 30, 2024 · The RF signal wirelessly transmits distinct images from within the gastrointestinal tract to an array of body-attached sensors, and it ...
  27. [27]
    A Fluorescence-Based Wireless Capsule Endoscopy System ... - NIH
    Apr 6, 2020 · The capsule is integrated with an image sensor, illumination source, batteries, communication module, and other components [5,6]. Once the ...
  28. [28]
    Capsule endoscopy of the small bowel - PMC - NIH
    This should be performed at a speed comfortable to the viewer, probably at a maximum of 15 frames per second. Some prior experience of endoscopy appears to be ...Ce Reading Software · Figure 1 · The End Of Conventional...
  29. [29]
    Capsule Endoscopy | University of Michigan Health
    During the test: After swallowing the capsule, you may drink clear liquids and take your medications after 2 hours, and you may eat after 4 hours. Avoid MRI ...
  30. [30]
    Capsule endoscopy - PMC - NIH
    The capsule is excreted with the feces, usually within 24 to 48 h[16]. CE is usually performed as an outpatient procedure. The presence of intestinal contents ...
  31. [31]
    Capsule retention: prevention, diagnosis and management
    Since in clinical practice the majority of capsules are excreted within 3–7 days (5), in order to avoid unnecessary X-ray exposure, it seems reasonable to plan ...
  32. [32]
    How to read small bowel capsule endoscopy: a practical guide ... - NIH
    Sep 21, 2020 · Basics of review & important steps in SB capsule endoscopy reading. The first step in SBCE evaluation is a quick overview of the entire video. ...
  33. [33]
    [PDF] The role of endoscopy in the management of suspected small-bowel ...
    Obscure GI bleeding (OGIB) has been defined as overt or occult bleeding of unknown origin that persists or recurs after an initial negative bidirectional ...
  34. [34]
    yield of capsule endoscopy (ce) compared to other modalities in ...
    Results: Fourteen studies compared the yield of CE with push enteroscopy (PE) for OGIB. The yield for CE and PE was 66% and 34%, respectively (n = 375; IY = 33 ...
  35. [35]
    British Society of Gastroenterology guidelines on inflammatory ... - Gut
    We suggest performing small bowel capsule endoscopy when small bowel Crohn's disease is suspected despite normal or inconclusive investigations. ... ECCO ...<|control11|><|separator|>
  36. [36]
    ECCO-ESGAR Guideline for Diagnostic Assessment in IBD Part 1
    Aug 23, 2018 · There are currently two validated indexes available, namely the Capsule Endoscopy Crohn's Disease Activity Index [CECDAI] and the Lewis Score, ...Chapter 2: Monitoring known IBD · Video-capsule endoscopy · Capsule endoscopy
  37. [37]
    Capsule endoscopy in celiac disease - PMC - NIH
    Capsule endoscopy has been reported to be able to demonstrate intussusception, ulcerative jejunoileitis, lymphoma and adenocarcinoma in patients with celiac ...
  38. [38]
    The Role of Capsule Endoscopy in the Diagnosis and Management ...
    Jan 7, 2024 · The role of capsule endoscopy in the diagnosis and management of small bowel tumors: a narrative review.
  39. [39]
    Current and Future Use of Esophageal Capsule Endoscopy - PMC
    Jul 31, 2018 · CONCLUSIONS. Capsule endoscopy has been proven to be a feasible and safe diagnostic option for patients with esophageal diseases. Several ...
  40. [40]
    Clinical utility of colon capsule endoscopy: a moving target? - NIH
    Oct 9, 2023 · Regarding detecting significant neoplastic lesions, the sensitivity of CCE and CTC was 96.1% and 79.3%, respectively. The detection rate for ...
  41. [41]
    Capsule endoscopy in pediatric patients - PMC - NIH
    The present review provides data on the indications, diagnostic yield, adverse events and limitations of the WCE technique in children and tries to predict the ...
  42. [42]
    Prospective postsurgical capsule endoscopy in patients with Crohn's ...
    AIM: To clarify the usefulness of postsurgical capsule endoscopy (CE) in the diagnosis of recurrent small bowel lesions of Crohn's disease (CD).
  43. [43]
    Capsule endoscopy: now a viable alternative to traditional ...
    Aug 22, 2025 · Capsule endoscopy was first introduced following the US Food and Drug Administration's (FDA) approval of Given Imaging's PillCam in 2001.
  44. [44]
    FDA clears next-gen Medtronic PillCam solution suite - MassDevice
    May 17, 2024 · “For over 20 years, Medtronic PillCam has led the world in capsule endoscopy, touching over 4 million patient lives,” Thomas wrote. “Now, we ...
  45. [45]
    CapsoVision - Crunchbase Company Profile & Funding
    It was founded in 2005 and headquartered in Saratoga, California. Read More.
  46. [46]
    Capsule endoscopy in gastrointestinal disease: Evaluation ...
    Apr 1, 2022 · Current indications for capsule endoscopy of the small intestine in adults (Table 1) include diagnosis of obscure GI bleeding and chronic iron- ...
  47. [47]
    Product | IntroMedic
    The MiroCam® capsule captures images of the GI tract, which are sent via Human Body Communication to sensor pads which are affixed to the patient body.
  48. [48]
    MiroCam™ capsule for obscure gastrointestinal bleeding
    Authors concluded that MiroCam was safe and effective to explore the entire small bowel, with good image quality and real-time feasibility [9]. The aim of this ...
  49. [49]
    [PDF] August 27, 2021 Given Imaging Ltd. (Medtronic) Randy Prebula ...
    Aug 27, 2021 · The PillCam SB 3 capsule is intended for visualization of the small bowel mucosa. • It may be used in the visualization and monitoring of ...
  50. [50]
    PillCam™ SB 3 Capsule | Medtronic
    Units per box ... The PillCam™ patency capsule is a dissolvable capsule used to verify adequate patency of the gastrointestinal tract prior to capsule endoscopy.
  51. [51]
    [PDF] Video capsule endoscopy - ASGE
    CCE requires a more intensive bowel preparation than is used for colonos- copy, which may be potentially challenging for patients. ASGE guidelines state that ...
  52. [52]
    Product Specifications - Capsovision
    CapsoCam Plus was designed to provide efficient and accurate diagnostic images and one of its most important features is its Smart Motion Sense™ technology.
  53. [53]
    CapsoView - Capsovision
    Award-winning 360º panoramic view. 4 cameras and 16 LED lights. 1.5GB onboard storage. Up to 20 frames per second. 3-5 FPS for each of ...
  54. [54]
    [PDF] October 21, 2016 CapsoVision, Inc Azimun Jamal Director of QA/RA ...
    Oct 21, 2016 · The CapsoCam Plus (SV-3) Capsule Endoscope System performance testing was conducted per the appropriate FDA. Recognized Consensus Standards and ...
  55. [55]
  56. [56]
    State-of-the-art and future perspectives in ingestible remotely ...
    Dec 1, 2024 · The capsule, with dimensions of 35.0 mm in length and 12.5 mm in diameter, was equipped with an enteric coating that underwent pH-dependent ...1. Introduction · 2. State-Of-The-Art Smart... · 3.8. Balancing Design...
  57. [57]
    Capsule Endoscopy: Prep, Procedure, What It Can Diagnose
    Avoid eating and drinking, as your stomach needs to be empty for the procedure. Usually, you must stop eating at least 12 hours before a capsule endoscopy. Your ...
  58. [58]
    Patient Information
    ### Summary of Common Side Effects of Capsule Endoscopy
  59. [59]
    [PDF] Cautions for Capsule Endoscopy Patient Manual - Olympus Medical
    Reddening or irritation of the skin may sometimes be caused by the adhesive on the antenna lead cover. If this is observed, consult a physician immediately. ○ ...
  60. [60]
  61. [61]
    A Descriptive Analysis of Capsule Endoscopy Events in the FDA ...
    Capsule retention occurs when the capsule remains in the digestive tract for 2 weeks or more, requiring directed medical, endoscopic, or surgical intervention.
  62. [62]
    Retention during capsule endoscopy: Is it a real problem in routine ...
    The rate of capsule retention depends on the indication for CE: 1.4% for OGIB, 1.5% for suspected CD, 5–13% for known CD and 21% for suspected small bowel ...
  63. [63]
    Patency testing improves capsule retention rates but at what cost? A ...
    Aug 8, 2023 · Two patients (1.2%) reported abdominal pain and bloating during the patency test. Both were managed conservatively and had passed the PC on ...
  64. [64]
    Contraindications for video capsule endoscopy - PMC - NIH
    In conclusion, swallowing disorders with the inability to safely swallow the capsule are a contraindication for standard procedure. However, if endoscopic ...
  65. [65]
    Clinical Experience with the PillCam Patency Capsule prior to Video ...
    The PillCam patency capsule (PC) is given before a video capsule endoscopy (VCE) in order to minimize the risk of capsule retention (CR).
  66. [66]
    Capsule retention: prevention, diagnosis and management - PubMed
    Therefore, retrieval of the retained capsule is recommended if asymptomatic CR lasts for long time. A safe capsule retrieval is usually performed by endoscopy, ...
  67. [67]
  68. [68]
    Artificial intelligence-assisted capsule endoscopy for detecting ...
    Mar 31, 2025 · AI holds the potential to automatically detect various types of lesions, shorten the reading times of capsule endoscopy and reduce missed ...
  69. [69]
    Deep learning meets small-bowel capsule endoscopy
    Oct 16, 2025 · AI-assisted SBCE could streamline high-volume workflows, standardize readings across varying expertise levels, and improve triage by ...
  70. [70]
    Recent technological advances in video capsule endoscopy
    Sep 29, 2025 · CapsoCam Plus uses a unique 360° panoramic imaging system with four lateral cameras ... view of 156° and generates two to six images per second.
  71. [71]
    FDA approves AnX Robotica's AI endoscopy tool
    Jan 3, 2024 · The tool is designed to help reviewers interpret small bowel capsule endoscopy (SB-CE) images for suspected small bowel bleeding in adult patients.
  72. [72]
    CapsoVision's IPO Raises $27.5M Towards AI-Enabled Capsule ...
    Jul 17, 2025 · Its flagship product, CapsoCam Plus, which has been cleared by the FDA for small bowel visualization, features four cameras, offering a 360 ...
  73. [73]
    Olympus Launches OLYSENSE CAD/AI in the US and Europe : 2025
    Sep 30, 2025 · In the United States, the CADDIE computer-aided detection device received Food and Drug Administration (FDA) 510(k) clearance (July 2024) to ...Missing: capsule | Show results with:capsule
  74. [74]
    Capsule Endoscopy: Current Trends, Technological Advancements ...
    Conceived in 2000 by Gavriel Iddan, CE employs a diminutive, ingestible capsule ... History of capsule endoscopy. In Small Intestine Disease: A ...Missing: Technion | Show results with:Technion
  75. [75]
    Drug Delivery Strategies for the Lower GI Tract: A Review - De Munter
    Sep 7, 2025 · SonoCAIT is one of the most recent and innovative capsule endoscopy prototypes designed for targeted drug delivery using US technology, The ...
  76. [76]
    New 3D models of the colon can help detect disease more rapidly
    Jun 25, 2024 · It is possible to re-create a fairly accurate three-dimensional (3D) model of the colon of some patients based on a single image taken by a capsule endoscopy ...
  77. [77]
    Magnetic Steering of Capsule Endoscopy Improves Small Bowel ...
    Feb 6, 2019 · Conclusions: Magnetic steering of capsule endoscopy improves small bowel CECR by reducing GTT, adding further support to MCE as a practical tool ...
  78. [78]
    Effect of magnetically guided capsule endoscopy on gastrointestinal ...
    Aug 28, 2025 · This systematic review and meta-analysis compare the effectiveness of these strategies in improving transit times and completion rates. Methods.
  79. [79]
    Tethered Capsule Endomicroscopy for unsedated microscopic ... - NIH
    Jun 4, 2021 · TCE device swallowing rate was 34/38 (89%). The appearance of physiologic upper gastrointestinal wall, including its microscopic pathology was ...
  80. [80]
    AI-ASSISTED CAPSULE ENDOSCOPY FOR DETECTION OF ...
    Oct 9, 2025 · This multicenter study aimed to validate an AI model for ulcers and erosions across different SBCE devices. Methods: A multicenter, cross- ...Missing: clinical PillCam
  81. [81]
    Integration of Artificial Intelligence-Enhanced Capsule Endoscopy in ...
    Jun 9, 2025 · As AI continues to advance, its integration into endoscopic practice has the potential to optimize diagnostic accuracy, streamline workflow, and ...
  82. [82]
    Exploring the feasibility of home-delivered capsule endoscopy with ...
    Oct 31, 2024 · This study demonstrates that home-delivered CCE is a feasible and acceptable alternative to traditional hospital-based screening methods.Missing: cloud- | Show results with:cloud-
  83. [83]
    Feasibility and Safety of Small Bowel Capsule Endoscopy in Very ...
    Jul 8, 2025 · Small bowel capsule endoscopy is a valuable modality for evaluating small intestinal lesions in children with inflammatory bowel disease.