OD600

OD600, or optical density at 600 nanometers, is a spectrophotometric technique used in to estimate the concentration of microbial cells, such as and , in liquid suspensions by measuring the resulting from light scattering. This method provides a rapid, non-destructive way to assess cell density without direct counting, serving as a cornerstone for monitoring culture growth in laboratories worldwide. The principle underlying OD600 relies on the of passing through a suspension, where the scattered reduces proportionally to the number of s in dilute samples (typically up to OD600 ≈ 1). The 600 nm is specifically selected because bacterial cells, with sizes ranging from 0.5 to 5 μm, scatter most efficiently at this visible red range, while absorption by cellular components or common culture media like Luria-Bertani broth is minimal, thereby reducing background interference. Additionally, 600 nm avoids the cell-damaging effects of s, making it suitable for repeated measurements on live cultures. In practice, OD600 is instrumental for tracking microbial growth phases—lag, exponential (log), stationary, and decline—and determining optimal harvest points, often at OD600 values of 0.4 to 0.6 during late log phase to maximize viable cell yields. For instance, in Escherichia coli, an OD600 of 1.0 approximates 8 × 108 cells per milliliter, though exact conversions depend on strain, growth conditions, and require species-specific calibration against methods like colony-forming units or dry weight. Measurements are performed using a spectrophotometer with a 1 cm path length cuvette, blanked against the culture medium, and samples are typically vortexed for homogeneity. Despite its ubiquity, OD600 has limitations: the signal becomes non-linear at higher densities due to multiple events, necessitating sample dilution; results vary with cell morphology, aggregation, and instrument ; and it primarily reflects total rather than viability. To enhance accuracy, with standards like silica microspheres has been shown to improve precision, with over 82% of replicates achieving residuals within 1.1-fold of predicted values. These considerations underscore the method's role as a qualitative that benefits from complementary quantitative techniques in rigorous applications.

Fundamentals

Definition and Purpose

OD600, or optical density at 600 , is defined as the of a microbial suspension measured at a wavelength of 600 using a spectrophotometer with a standard 1 cm light path. This metric quantifies the caused by light and by cells in liquid culture, providing an indirect measure of concentration. The 600 wavelength is selected because it minimizes from cellular pigments while effectively detecting from bacterial cells, making it suitable for a wide range of microorganisms. The primary purpose of OD600 is to serve as a non-invasive for estimating concentration, enabling quick assessments without destructive sampling or direct enumeration techniques like colony counting. For instance, in cultures grown under standard conditions in Luria-Bertani medium, an OD600 of 1.0 approximates 8 × 108 cells per milliliter, though exact conversions depend on , growth conditions, and require species-specific . This correlation allows researchers to standardize experiments and track efficiently in . As a dimensionless value, OD600 represents the negative logarithm of the and is approximately linear for estimation in the range of 0.1 to 1.0, beyond which dilutions are often required to maintain accuracy. This scale ensures proportional readings within typical experimental densities, supporting consistent inter-laboratory comparisons when standardized protocols are followed.

Underlying Principles

The optical density at 600 nm (OD600) is fundamentally based on the interaction of with microbial s in suspension, where the primary mechanism is scattering due to rather than specific molecular absorption. Microbial s, typically ranging from 0.5 to 5 μm in size, scatter incident through processes like or Mie scattering, depending on the wavelength and dimensions relative to the ; at 600 nm, scattering predominates over absorption because bacterial cellular components exhibit minimal absorbance in this visible range. This effect attenuates the transmitted , providing a measure of concentration without requiring or . The relationship between OD600 and cell concentration is described by an adaptation of the Beer-Lambert law, which originally applies to but extends to -dominated systems under certain conditions. The law states that A (equivalent to OD) is proportional to the concentration c of the scattering particles, the path l (typically 1 in standard cuvettes), and the \epsilon, expressed as: A = \epsilon l c In microbial suspensions, \epsilon incorporates both and minor contributions, and the proportionality holds linearly at low densities (OD600 ≲ 0.2) where single scattering dominates; at higher densities, multiple scattering introduces nonlinearity, requiring for accurate concentration estimates. The selection of 600 nm as the measurement wavelength optimizes sensitivity to cell density while minimizing artifacts. This wavelength avoids strong absorption by common growth media components (e.g., nutrients in broth) and microbial pigments like or , which absorb more in UV or shorter visible ranges; it also provides sufficient efficiency for without the photodamage associated with UV . OD600 correlates with and approximates viable cell counts during the , where cells are relatively uniform in size and shape, but this relationship is influenced by variations in cell morphology—such as or clumping—which alter the cross-section and thus the effective \epsilon. For instance, larger or irregularly shaped cells scatter more light per unit , potentially overestimating if not calibrated against direct counts like colony-forming units.

Measurement Methods

Standard Protocol

The standard protocol for OD600 measurement begins with preparation of the microbial culture. Cultures are grown to the desired growth phase in an appropriate medium under controlled conditions, such as and , before sampling. To ensure homogeneity, the culture is gently vortexed or swirled to resuspend settled cells, while avoiding excessive that could damage cells or introduce bubbles. Sterile technique is employed throughout to prevent , including the use of sterile and work surfaces. If the anticipated optical density exceeds the upper limit of the linear range (typically up to OD600 ≈ 1.0)—samples are diluted serially, often at a 1:10 ratio using the same or a compatible to avoid osmotic shock. is avoided for dilutions to preserve integrity. The original OD600 value is then calculated by multiplying the measured diluted value by the dilution factor. This dilution strategy ensures readings fall within the proportional range governed by the Beer-Lambert law. For the measurement itself, a spectrophotometer is first zeroed using a blank consisting of uninoculated in a to account for background . A 1 cm path length —either for reusability or disposable plastic—is filled with 1-3 mL of the prepared sample, ensuring no air bubbles are present. The is inserted into the spectrophotometer, which is set to 600 nm , and the is recorded promptly. Measurements are performed in replicates (at least triplicates) to assess variability and improve reliability. Best practices include maintaining consistent temperature during measurement, typically between 25°C and 37°C to match growth conditions and minimize thermal effects on cell settling. Cuvettes must be clean and matched pairs used for blank and sample to reduce errors. Samples should be measured immediately after preparation to avoid evaporation or further growth, and the spectrophotometer should be warmed up as per manufacturer instructions for stable readings.

Instrumentation and Variations

The core instrumentation for OD600 measurements is a UV-Vis spectrophotometer equipped with a source and detector capable of precise readings at 600 nm, paired with s that provide a standard 1 cm path length to adhere to the Beer-Lambert law principles. Common types include macro-volume quartz s for high accuracy in traditional setups, micro-volume s for smaller sample sizes (typically 50–500 µL), and disposable s made from materials like or PMMA, which offer convenience and reduce cross-contamination risks while maintaining suitable transparency at 600 nm. Variations in instrumentation adapt OD600 measurements to specific scales and environments. Microplate readers enable high-throughput analysis in 96-well formats, where a typical 200 µL sample volume yields an effective path length of approximately 0.6 , necessitating software-based to normalize readings to the standard 1 path length. For continuous, monitoring in large-scale fermenters, immersion or dip-cell probes—such as in-line optical density sensors—allow real-time assessment at 600 without disrupting the process, supporting sanitary integration in bioreactors up to thousands of liters. of these instruments requires regular validation using standards like monodisperse bead suspensions (e.g., 0.5–1 µm diameter), which mimic microbial scattering properties and enable cross-instrument standardization, often with software for automated logging and to convert to cell density equivalents. Emerging technologies enhance precision and portability for OD600 applications. Nano-volume spectrometers, such as those using or systems, facilitate measurements with sample volumes under 2 µL by concentrating the liquid path, though they may require dilution for dense cultures to stay within linear ranges. Portable handheld devices, including battery-powered dilu-photometers and ultrafast spectrophotometers, support field or on-site use with minimal sample preparation, offering over 8 hours of operation and integration with mobile apps for data export.

Applications

Microbial Growth Monitoring

OD600 is a primary method for monitoring microbial growth dynamics in bacterial and cultures by quantifying changes in optical over time, which reflects increases in cell . During the , OD600 remains low as cells adapt to the medium without significant division. This transitions to the log or , where OD600 increases exponentially, doubling with each as cells actively replicate under optimal conditions. The stationary follows, marked by a plateau in OD600 as nutrient depletion and waste accumulation balance growth and death rates. Finally, the death shows a decline in OD600 due to ongoing cell and reduced viability. To track these phases, cultures are typically sampled every 30-60 minutes, with OD600 measured using a standard spectrophotometric at 600 nm, and values plotted against time to generate a curve. From the linear portion of the log phase on a , the specific rate μ can be calculated as μ = (ln OD₂ - ln OD₁) / (t₂ - t₁), where OD₁ and OD₂ are optical densities at times t₁ and t₂, respectively, providing a quantitative measure of replication in units of h⁻¹. The relationship between OD600 and actual cell density varies by strain due to differences in cell size and shape; for , an OD600 of 1 corresponds to approximately 8 × 10⁸ colony-forming units (CFU) per ml, while yeast such as exhibit lower cell counts per OD unit—around 1.5 × 10⁷ cells per ml at OD600 = 1—owing to their larger cell volume. In practice, for recombinant protein expression in E. coli, cultures are often inoculated at an initial OD600 of 0.05 from an overnight starter and monitored until reaching OD600 of 0.6-1.0 in mid-log phase for induction and harvest, optimizing yield during active growth.

Biotechnological Uses

In biotechnological processes, OD600 serves as a key metric for optimizing fermentation conditions in bioreactors, enabling real-time monitoring of microbial biomass to adjust nutrient feeding and sustain exponential growth phases critical for metabolite production. For instance, in fed-batch fermentations of Escherichia coli for poly(3-hydroxyalkanoates) (PHA) biosynthesis—a biodegradable polymer used in biofuels and plastics—OD600 measurements guide the transition from growth to production stages, with induction occurring when OD600 exceeds 30 to maximize yields up to 20 g/L PHA. Similarly, in recombinant protein production, such as reteplase, OD600 tracks culture density to fine-tune inducer additions, achieving high cell densities while preventing overflow metabolism. For recombinant protein expression, OD600 determines optimal induction and harvest timings to enhance soluble protein yields in expression hosts like E. coli BL21(DE3). is typically performed at mid-log phase (OD600 of 0.6–0.8) to balance and metabolic capacity, followed by expression at reduced temperatures (15–25°C) for 16–20 hours, with harvests at OD600 of 1.0–2.0 to minimize inclusion body formation. This approach is widely adopted in industrial biomanufacturing for therapeutics, where OD600-based protocols correlate with downstream purification efficiency. In antibiotic susceptibility testing, OD600 quantifies inhibition by measuring pre- and post-treatment in assays, facilitating determination of minimum inhibitory concentrations (s). According to CLSI guidelines, inocula are standardized to an OD600 of approximately 0.1 in cation-adjusted Mueller-Hinton to ensure reproducible results across aerobic , with MIC endpoints defined by the lowest concentration showing no visible growth (OD600 ≤ 0.05–0.1). This supports rapid screening in pharmaceutical development, correlating OD600 reductions with 90–100% inhibition accuracy for agents like . During industrial scale-up, OD600 is adapted via online optical sensors in large-volume bioreactors to enable non-invasive tracking and through correlations with weight (DCW). Probes like the AFGUARD® system, operating on scattered light at near-infrared wavelengths, provide OD600 equivalents up to 2.2 g/L DCW in continuous fermentations, such as production by coagulans, allowing automated feeding adjustments that maintain steady-state growth. These sensors mitigate sampling disruptions in volumes exceeding 100 L, ensuring scalability from to production while validating OD600-DC W linearity (R² > 0.95).

Limitations and Alternatives

Sources of Inaccuracy

One primary source of inaccuracy in OD600 measurements arises from the non-linearity of the optical density signal at higher cell densities. The underlying Beer-Lambert law assumes single events, which holds reliably only in the low-density regime (typically OD600 ≲ 0.2–1.0), where OD is approximately proportional to cell concentration. Beyond this range, multiple predominates, causing light to be redirected multiple times and resulting in an overestimation of cell density relative to actual counts. For example, undiluted samples exceeding OD600 > 1.0 often require to restore linearity, as the parabolic deviation can lead to errors exceeding 50% in density estimates without correction. Biological variability among cells introduces further inconsistencies in OD600 readings. Cell clumping or aggregation, common in certain strains or under suboptimal conditions, disrupts uniform light scattering and yields poor correlation with total cell numbers, with calibration residuals up to 28-fold higher when compared to direct counting methods like . Morphological changes, such as filamentation induced by stress factors like antibiotics or nutrient limitation, elongate cells and increase scattering per cell without a corresponding rise in cell count, thereby inflating OD600 values disproportionately. Species-specific differences exacerbate this issue; for instance, like typically exhibit an OD600 of 1 corresponding to approximately 8 × 10^8 cells/mL, whereas ratios for other species vary due to differences in cell size and wall thickness. Environmental factors can significantly interfere with OD600 accuracy by mimicking or masking cell-derived . Media from non-cellular components, such as extracellular or metabolic byproducts, adds extraneous light attenuation that is indistinguishable from cellular contributions, leading to overestimated densities in complex growth media. Bubbles, often introduced during sample handling or , cause erratic and can significantly elevate readings if not removed by gentle tapping or . Temperature fluctuations indirectly affect measurements by altering cell morphology or promoting evaporation in open systems like microplates, which concentrates samples and affects readings. Additionally, contribute to equivalent to viable ones, as OD600 quantifies total rather than metabolic activity, resulting in viability overestimations during stationary phase or exposure where rates can exceed 50%. Calibration deficiencies represent a systemic source of inaccuracy, stemming from the absence of standardized conversion factors across instruments and conditions. OD600-to-cell count relationships vary by spectrophotometer and path length, necessitating device-specific to achieve residuals below 10%; without them, inter-instrument discrepancies can reach 2-fold. Moreover, the correlation between OD600 and colony-forming units (CFU/mL) weakens markedly in the stationary phase, where morphological shifts and viability loss cause up to 3.3-fold variations in ratios.

Complementary Techniques

Direct counting methods provide absolute cell numbers, offering a more precise alternative to OD600 by enumerating individual cells without relying on light attenuation assumptions. Flow cytometry, for instance, uses laser-based detection to count bacterial cells at rates exceeding 10,000 per second, enabling high-throughput analysis and calibration against OD600 for improved accuracy in estimating cell density. Hemocytometers, traditional chamber-based counters, allow manual microscopic enumeration of cells in a known volume, correlating bacterial counts (e.g., 10^8 to 10^9 cells/mL at OD600 = 1) with optical density for validation in low-density cultures. To distinguish live from dead cells—addressing OD600's oversight of viability—stains like propidium iodide (PI) are integrated with flow cytometry; PI penetrates compromised membranes of non-viable bacteria, fluorescing red upon binding DNA, while live cells exclude it, achieving up to 95% accuracy in viability assessment. Viability assays complement OD600 by quantifying only metabolically active cells, crucial for applications where total biomass includes non-viable debris. Colony-forming unit (CFU) plating involves diluting and spreading samples on plates, incubating to count visible colonies, which represent viable s capable of division; this method correlates with OD600 but reveals discrepancies in stressed populations, with typical yields of 10^8 to 10^9 CFU/mL at OD600 = 1 for . ATP bioluminescence assays, such as those using , detect intracellular ATP as a proxy for metabolic activity, producing light proportional to viable cell numbers (e.g., 10^5 to 10^7 relative light units per 10^6 s), offering rapid (under 10 minutes) results insensitive to OD600's aggregation artifacts. Biomass proxies extend OD600's utility in complex media where cell morphology varies, providing total dry matter or protein content as stable metrics. Dry weight measurement entails filtering, washing, and oven-drying cell pellets to weigh , yielding correlations like 0.27 g/L dry weight per OD600 unit for Vibrio natriegens, ideal for filamentous or aggregated microbes where OD600 underestimates mass. Protein assays, notably the method using dye, quantify total cellular protein (e.g., 100-200 μg/mL at OD600 = 1), binding to basic residues for colorimetric detection at 595 nm; this approach is particularly effective in nutrient-rich media, normalizing for extracellular interferents that skew OD600 readings. Advanced optical techniques refine and , surpassing OD600's bulk by resolving subpopulations. Nephelometry measures forward at 90° to assess clarity, correlating scattered intensity with bacterial (e.g., linear response up to 10^9 particles/mL), useful for detecting early aggregation not evident in absorbance-based OD600. Imaging combines with flow, capturing high-resolution images of thousands of cells per second to quantify size (e.g., 1-5 μm diameters for typical ), enabling phenotypic heterogeneity that calibrates OD600 against variable morphologies in dynamic cultures.

References

  1. [1]
    Robust estimation of bacterial cell count from optical density - Nature
    Sep 17, 2020 · Measurements of OD via absorbance at 600 nm (OD600) were calibrated using three protocols and for each of these a model was devised for the ...<|control11|><|separator|>
  2. [2]
    Cell Density measurement by OD600 Method
    ### Summary of OD600 Method for Cell Density Measurement
  3. [3]
    OD600 Measurements: Tips for Implementation - Eppendorf Canada
    Feb 24, 2020 · The measurement of optical density at 600 nm is a simple and widely used method to monitor the growth of microorganisms in liquid culture ...
  4. [4]
    Concentration vs. Optical Density of ESKAPEE Bacteria - NIH
    Dec 21, 2024 · Bacteria scatter light most efficiently at wavelengths around 600 nm [6]. However, measurements between 600 nm and 700 nm can be used [7]. The ...
  5. [5]
    OD600 Measurement For Cell Density: Bacterial And Yeast Growth
    OD600 measurements are typically used to determine the stage of growth of a bacterial culture, these measurements help ensure that cells are harvested at an ...
  6. [6]
    A brief history of bacterial growth physiology - Frontiers
    Or, to determine the growth rate of a culture, optical density (mass) measurements were made at least 10 times in the course of each doubling of the culture.
  7. [7]
    [PDF] Relationship between Population Density and Specific Growth Rate ...
    Experience showed that optical density was a linear function of population density within values of optical density of 0.0-0-6. In practice, a sample of a ...
  8. [8]
    General calibration of microbial growth in microplate readers - Nature
    Dec 13, 2016 · The Beer-Lambert law (Supplementary Note 1) assumes that light is only absorbed to derive OD ~ C, which is true if the light received by the ...Results · Calibration Of Od Typically... · Methods
  9. [9]
    Concentration vs. Optical Density of ESKAPEE Bacteria: A Method to ...
    In bacterial studies, optical density at specific wavelengths (usually 600 nm, referred to as OD600) became a standard way of quantifying cultures, although ...Missing: seminal | Show results with:seminal
  10. [10]
    Measure microbial growth using the „OD600“ | BMG LABTECH
    Oct 11, 2019 · The most common way to assess microbial growth in solution is the measurement of the optical density at 600 nm or short OD 600.The principle behind... · Considerations for microbial...
  11. [11]
    None
    ### Standard Protocol for OD600 Cell Density Measurement
  12. [12]
    OD600 Measurements | Technical Note 168 - DeNovix
    Jan 14, 2020 · The purpose of this note is to highlight sources of the potential differences observed in OD 600 measurements between instruments and present performance data ...<|separator|>
  13. [13]
  14. [14]
    [PDF] NanoDrop One Microvolume UV-Vis Spectrophotometers
    Use cuvettes to measure dilute samples and optical density of bacterial cultures or to perform kinetics experiments. Includes cuvette temperature control and ...Missing: equipment | Show results with:equipment
  15. [15]
    OD600 (Cell Density, Bacterial Growth, Yeast Growth) - Implen
    OD600 measurements are typically used to determine the stage of growth of a bacterial culture, these measurements help ensure that cells are harvested at an ...Missing: core | Show results with:core
  16. [16]
  17. [17]
    Reconsidering Dogmas about the Growth of Bacterial Populations
    According to the literature, the characteristic bacterial growth curve pattern comprises four phases: lag, exponential growth, stationary phase, and death phase ...2.1. Cell Culture And Growth... · 2.8. Proteomics Analysis · 3. Results
  18. [18]
    [PDF] Bacterial Growth Curves: Analysis through OD₆₀₀ measurements
    By measuring the “absorbance” at long wavelengths. (600 nm is typically used by convention – OD₆₀₀), the growth of the bacteria can be monitored much more ...Missing: seminal | Show results with:seminal
  19. [19]
    Estimating microbial population data from optical density - PMC - NIH
    The spectrophotometer has been used for decades to measure the density of bacterial populations as the turbidity expressed as optical density–OD.Results · Decision Letter 0 · Author Response To Decision...
  20. [20]
    [PDF] DIRECT YEAST CELL COUNT AT OD600 - Tip Biosystems
    We found a yeast cell number of 1.5 x 10^7 cells per 1 OD600 unit, this is within the data found in the literature.
  21. [21]
    [PDF] Protein Expression and Purification Series - Bio-Rad
    An OD600 of 0.6–1.0 is the typical target for induction. At this point the cells are dividing rapidly, and production of the protein will be optimal. For ...
  22. [22]
    Optimizing a Fed-Batch High-Density Fermentation Process for ... - NIH
    Feb 26, 2021 · In this study, we developed and optimized a fed-batch fermentation process in a stirred tank reactor, beginning with the biosynthesis of poly(3- ...
  23. [23]
    Optimization of Fermentation Conditions for Reteplase Expression ...
    OD600 of samples was evaluated to measure the growth of culture and ... Expression of reteplase under optimum fermentation conditions in a 2-L bioreactor ...
  24. [24]
    Strategies to Optimize Protein Expression in E. coli - PMC
    After the overnight starter culture is used to inoculate the large‐scale culture (step 5), the cells are grown to mid‐log phase (OD600 of ∼0.6 to 0.9) in Luria ...Missing: harvest | Show results with:harvest
  25. [25]
    Engineering protein production by rationally choosing a carbon and ...
    Sep 4, 2019 · In this study, our aim was to find the most favourable conditions for carrying out recombinant protein production in E. coli BL21 using two different ...<|separator|>
  26. [26]
    Minimum Inhibitory Concentration (MIC) Determination - BMG Labtech
    For the determination of MIC, OD600 is used to distinguish between concentrations of an antibiotic that are sufficient to inhibit microbial growth and those ...
  27. [27]
    An Innovative Optical Sensor for the Online Monitoring and Control ...
    Mar 21, 2016 · Online monitoring of biomass during the batch fermentation of B. coagulans for the production of LA, using optical density (OD600) and cell dry ...
  28. [28]
    Contact-free infrared OD measurement for online monitoring of ...
    Dec 15, 2020 · Various online OD-measurement techniques for the monitoring of large scale bioreactors up to high cell densities are well established and used ...Regular Article · Introduction · Microorganisms And Media
  29. [29]
    Understanding OD600: Principle and Considerations
    Optical Density at 600 nm (OD600) is a widely used spectroscopic method for estimating the concentration of bacterial cells in a culture.Missing: definition | Show results with:definition
  30. [30]
    [PDF] The OD600DiluPhotometer™ User Manual - Implen
    It is recommended however that you check for air bubbles before the measurement. In case of an air bubble in the cell please remove it by gently tapping the ...
  31. [31]
    How to optimise OD600 measurements - BMG Labtech
    OD600 measurements were started after the plate was inserted into the microplate reader at RT (25°C). Evaporating liquid led to the formation of condensation on ...
  32. [32]
    [PDF] Bacterial growth curve by OD600 and SoloVPE - Repligen
    The aim of this study is to detect the bacterial growth curve of E. coli by two methods, such as standard spectrophotometer and SoloVPE method. The cell- ...
  33. [33]
    [PDF] Technical Insight: The Inconsistencies of OD600 - SBT Instruments
    Conversion from OD-units to CFU/mL is dependent on instrument configuration ... OD600-to-CFU ratios will shift during an E. coli fermentation. A ...<|separator|>
  34. [34]
    Robust estimation of bacterial cell count from optical density - PMC
    The most common method for estimating the number of cells in a liquid suspension is the use of optical density measurements (OD) at a wavelength of 600 nm ( ...Results · Experimental Data Collection · Fig. 1. Study DesignMissing: morphology filamentation
  35. [35]
    Growth Dynamics and Survival of Liberibacter crescens BT-1, an ...
    Oct 16, 2019 · The total number of bacterial cells ml−1 per OD600 was determined by dividing the total bacterial number ml−1 determined from hemocytometer ...
  36. [36]
    Assessment and Interpretation of Bacterial Viability by Using ... - NIH
    Its use in combination with flow cytometry brought up new questions about how to interpret LIVE/DEAD staining results.
  37. [37]
    Comparison of Two Optical-Density-Based Methods and a Plate ...
    For wells not reaching an OD600 of 0.2 within the time frame of the experiment, viability of bacteria was determined by plating the content of the well.Missing: assay | Show results with:assay
  38. [38]
    Real-time monitoring of extracellular ATP in bacterial cultures using ...
    Jan 22, 2021 · ATP-based hygiene and cell viability testing. The ATP bioluminescence assay is based on a two step-oxidation of D-luciferin catalysed by the ...
  39. [39]
    Metabolism of the fast-growing bacterium Vibrio natriegens ...
    Oct 16, 2017 · The OD600 values were converted to cell dry weight concentrations using the following relationship for V. natriegens: 1.0 OD600 = 0.27 gDW/L, ...
  40. [40]
    Reprogramming microbial populations using a programmed lysis ...
    Nov 25, 2021 · In this study, a programmed lysis system (PLS) is developed to reprogram microbial cooperation to enhance chemical production.
  41. [41]
    Application of an Online-Biomass Sensor in an Optical Multisensory ...
    Sep 17, 2014 · The biomass sensor signal could successfully be correlated and calibrated with well-known measurement methods like OD600, cell dry weight (CDW) ...
  42. [42]
    The Application of Imaging Flow Cytometry for Characterisation and ...
    In this study, we applied high-throughput Imaging Flow Cytometry (IFC) to characterise and quantify, at single-cell level and over time, the phenotypic ...Missing: complementary | Show results with:complementary