Resuscitator

A resuscitator, also known as a bag-valve- (BVM), is a portable, hand-operated designed to deliver positive pressure ventilation to patients experiencing or apnea, by manually inflating their lungs with oxygen-enriched air. It typically consists of a self-inflating bag, a one-way to prevent rebreathing of exhaled gases, and an such as a face or endotracheal , enabling providers to and ventilate individuals who cannot spontaneously. These devices are essential in settings like (CPR), care, and , where immediate respiratory support can be lifesaving. The modern resuscitator was invented in 1953 by Danish anesthesiologist Henning Ruben and engineer Holger Hesse amid a epidemic in , which created a in demand for ventilation tools during an oxygen shortage caused by a national truck drivers' strike. Launched commercially in as the Ambu bag by Testa-Laboratorium (later renamed Ambu), it was the first self-inflating resuscitator, featuring a simple design with a latex bulb reinforced by bicycle spokes (later replaced by foam rubber) that required no external power source or oxygen supply for basic function. This innovation addressed the limitations of earlier bellows-style devices from the early 20th century, such as the Dräger Pulmotor of 1907, which were bulky, oxygen-dependent, and prone to fatigue during prolonged use. By the , the Ambu bag had largely supplanted these predecessors, becoming a standard in due to its reliability, portability, and ease of sterilization. Resuscitators are classified primarily into self-inflating and flow-inflating types, with the former—exemplified by the Ambu bag—relying on to automatically refill the bag with ambient air after compression, making it ideal for prehospital and resource-limited environments. Flow-inflating models, in contrast, require a continuous oxygen source to inflate the bag and are often used in controlled clinical settings for more precise delivery. Available in adult and pediatric sizes, with the latter incorporating pop-off valves to limit pressure and prevent in infants and children, resuscitators are constructed from durable, latex-free materials to minimize infection risks and ensure single-patient use in disposable variants. In clinical practice, resuscitators facilitate positive pressure ventilation at rates of 10–12 breaths per minute for adults, delivering tidal volumes of 400–600 mL while monitoring for chest rise and avoiding , which can complicate CPR outcomes. They integrate with advanced airway devices like supraglottic airways or laryngoscopes and are often equipped with accessories such as (PEEP) valves to improve oxygenation in conditions like (ARDS). Despite their simplicity, effective use demands training to achieve proper mask seal and ventilation technique, underscoring their role as a foundational tool in global emergency response protocols.

Overview

Definition and Purpose

A resuscitator is a portable that delivers positive pressure to patients with inadequate or absent spontaneous , helping to prevent and maintain adequate oxygenation. These devices function by forcing oxygen-enriched air into the lungs via a , , or other , restoring respiratory function in critical situations where normal has ceased. The primary purposes of resuscitators encompass immediate during , acute , or , where rapid intervention is essential to sustain life. They also serve as a temporary bridge to more advanced systems in clinical settings and are employed by emergency medical responders in both pre-hospital and environments to support patients until definitive care is available. In global emergency protocols, such as those from the , bag-valve-mask resuscitators are the standard for initial ventilation during out-of-hospital , forming a core component of .

Basic Components

A standard resuscitator, also known as a bag-valve-mask (BVM) , comprises several core components that enable manual positive pressure ventilation to support breathing in unable to ventilate adequately. These include a self-inflating bag serving as the primary reservoir for air or oxygen, one-way valves to direct gas flow and prevent rebreathing, a for delivering gas to the airway, and an oxygen inlet port for supplemental oxygen delivery. The self-inflating bag is the central element, constructed to recoil automatically after compression, drawing in fresh gas from the or an oxygen to refill itself. It typically has a capacity of 500 for pediatric use and 1,500–1,600 for adults, allowing delivery of tidal volumes around 300–800 depending on squeeze force and patient size. This design ensures reliable inflation without external power, facilitating intermittent or continuous ventilatory support. One-way valves are integral to unidirectional gas flow: the intake valve, often featuring flap or disc membranes, admits or oxygen into the bag while blocking ; the patient valve, including non-rebreathing mechanisms like duckbill or single-shutter valves, directs exhaled gas away from the patient to prevent re-inhalation of carbon dioxide-enriched air. These valves maintain low resistance—typically 2–4 cmH₂O for inspiration and expiration—to minimize during operation. The patient interface consists of a face mask that seals over the and or a standard 15 mm connector for attachment to an endotracheal tube, ensuring targeted delivery of ventilatory gases directly to the airway. Masks are sized variably (e.g., sizes 0–5 for infants to adults) to achieve an airtight seal without excessive pressure on facial tissues. An oxygen inlet port, usually a nipple or tubing connector, allows attachment to an external oxygen supply, often with a bag (600 mL pediatric, 2,600 mL adult) to achieve FiO₂ levels up to 100% at flows of 10–15 L/min. Without oxygen, the device delivers room air (21% FiO₂); with a and high flow, oxygenation improves significantly for hypoxic patients. Resuscitators are typically fabricated from biocompatible, durable materials such as for reusable models, which can withstand autoclaving for sterilization, or (PVC) for disposable versions to ensure single-use sterility and cost-effectiveness. These materials provide flexibility for bag compression, chemical resistance, and latex-free composition to reduce risks. Bag capacities and valve dead spaces (around 7 mL) are standardized to support safe across age groups. In terms of flow dynamics, squeezing the bag manually generates , commonly 10–40 cmH₂O, to inflate the lungs with tidal volumes of 6–7 mL/kg ideal body weight, delivered over 1–2 seconds at rates under 12 breaths per minute. This creates a that drives gas through the valves and interface, with exhalation occurring passively as the bag refills and chest recoil expels CO₂ via the outlet valve. Excessive squeeze can exceed safe limits, risking . Variations in setup enhance safety in advanced models: a pop-off , often set at 35–60 cmH₂O, automatically vents excess to protect against lung overdistension, particularly in pediatric devices; some integrate a manometer to monitor real-time airway pressures, aiding precise control during use. These features are optional but recommended for high-risk scenarios to maintain pressures below 30–40 cmH₂O.

Types of Resuscitators

Manual Resuscitators

Manual resuscitators, commonly known as bag-valve- (BVM) devices, are hand-operated tools designed for providing positive pressure ventilation to patients who are not adequately. These devices consist of a self-inflating bag connected to a one-way valve system and a face , allowing a single operator or team to deliver controlled breaths without relying on external power sources. The Ambu bag, a widely recognized proprietary example, exemplifies this design and has become synonymous with manual resuscitation efforts in emergency settings. The core design features a compressible, self-inflating made from durable, medical-grade materials like SEBS, with an adult-sized bag typically holding a of 1.5 to 2 liters. Integrated non-rebreathing valves ensure unidirectional , preventing exhaled gases from re-entering the bag while allowing intake from ambient air or an attached oxygen . When squeezed, the bag delivers a of approximately 500-600 mL for adults, adjustable based on squeeze force and duration, which supports effective inflation without excessive . This configuration enables compatibility with various sizes for airtight seals over the and . In operation, the rescuer rhythmically squeezes the bag to force air or oxygen-enriched gas into the patient's lungs via , with the bag automatically refilling upon release through intake valves drawing from the environment or . For adult patients, ventilations are typically administered at a of 10-12 breaths per minute, each lasting 1 second to mimic natural and minimize risks like . The device's simplicity allows for immediate use in diverse scenarios, from prehospital care to operating rooms, bridging the gap until is available. Key advantages of manual resuscitators include their portability and reliability in resource-limited environments, as they require no electricity or batteries and can function solely with room air if oxygen is unavailable. Production costs are low, with disposable adult units often priced under $20, making them accessible for widespread deployment in emergency kits and ambulances. When used correctly, they achieve effective in the majority of cases, supporting oxygenation and until more definitive interventions, though operator skill is crucial to avoid complications like gastric . The invention of the manual resuscitator traces back to Danish anesthesiologist Henning Ruben, who invented the first self-inflating bag in 1953 amid an oxygen supply crisis in , with commercial launch in 1956, revolutionizing portable manual ventilation by eliminating the need for compressed gases or electrical power.

Expired Air Resuscitators

Expired air resuscitators are basic ventilation devices designed for mouth-to-mask or mouth-to-barrier delivery, utilizing the rescuer's exhaled breath to provide positive pressure ventilation to a non-breathing . These devices feature a transparent mask that seals over the patient's mouth and nose, incorporating a one-way and outflow valve to direct the rescuer's expired air into the patient's airway while preventing reverse flow of the patient's exhaled gases back to the rescuer. An inflow valve allows the rescuer to draw in fresh room air between breaths, and some early designs included a relief valve to limit pressure and avoid ; modern devices rely on proper rescuer technique for pressure control. Unlike powered or bag-based systems, these resuscitators require no external equipment or power source, relying entirely on the rescuer's manual exhalation and lung capacity to generate tidal volumes typically limited to 500-800 mL per breath, though effective delivery of approximately 500 mL (6-7 mL/kg for adults) is recommended to ensure adequate ventilation without excessive gastric insufflation. In operation, the rescuer places the mask securely on the patient's face and exhales steadily through the device at a controlled rate of 10-12 breaths per minute, observing chest rise to confirm effective delivery. The exhaled air provides an oxygen concentration of approximately 16-17%, sufficient for short-term oxygenation in emergencies, as it maintains alveolar oxygen tension around 80 mm Hg. One-way valves and optional bacterial/viral filters (with up to 99.8% efficiency) minimize cross-infection risks by blocking pathogens from the patient's . This method is particularly suited to resource-limited settings, such as layperson interventions during or . The primary advantages of expired air resuscitators include their immediate availability in standard CPR kits, portability, and zero risk of mechanical failure, making them ideal for bystander resuscitation as endorsed by (AHA) guidelines since the . They enable rapid initiation of without specialized training beyond CPR , historically validated for maintaining oxygenation in scenarios like or . However, prolonged use carries a risk of in the patient due to the 3.5-4.1% content in the rescuer's exhaled air, which can inhibit cardiac function and elevate blood CO2 levels over time. Additionally, in infectious disease outbreaks, these devices are unsuitable without robust barriers, as the close proximity increases transmission potential despite valves, with rare documented cases of pathogens like or crossing during mouth-to-mask contact.

Oxygen-Powered Resuscitators

Oxygen-powered resuscitators are medical devices that utilize compressed oxygen or air from a to automate the delivery of positive , making them ideal for scenarios requiring sustained or hands-free support in critical care settings. These systems typically employ a demand mechanism or continuous design connected directly to an oxygen source, such as an E- , enabling the provision of high-concentration oxygen—ranging from 40% to 100%—at adjustable rates up to 30 liters per minute (L/min). The lightweight and compact construction, often weighing under 0.5 kg and measuring around 140 x 63 x 73 mm, facilitates portability without reliance on batteries or electrical power. In operation, these resuscitators initiate through patient-triggered demand or fixed-time cycling, with modes selectable for override or delivery at rates of 10-20 breaths per minute (). Pressure is inherently limited to 40-60 cmH₂O to prevent , achieved via built-in relief valves that terminate flow upon reaching the threshold, accompanied by an audible alarm for high pressure or obstructions. Additional safety features include (PEEP) options of 2-4 cmH₂O and compatibility with masks, endotracheal tubes, or tracheostomies for patients over 10 kg. Low-pressure alarms may activate if the oxygen supply drops below 30 L/min, ensuring reliable performance during use. Key advantages of oxygen-powered resuscitators include significant reduction in rescuer fatigue by automating consistent delivery, precise control over inspiratory flows to minimize gastric , and their suitability for transport where hands-free operation allows focus on other interventions. These devices support minute volumes of 9-18 / in automatic mode, enhancing oxygenation efficiency in prehospital environments. is determined by the equation: \text{Tidal volume} = \text{Flow rate} \times \text{Inspiratory time} For instance, at a flow rate of 15 L/min (0.25 L/s) and an inspiratory time of 1 second, this yields approximately 250 mL, scalable to adult needs of 500 mL or more by adjusting parameters. Early models from the 1960s, such as those developed by companies like Bird and Bennett, began integrating oxygen-powered features with basic ventilatory systems to improve reliability in emergency settings.

Historical Development

Early Innovations

Early efforts in mechanical resuscitation predated the , with devices emerging as rudimentary tools for , particularly in response to accidents where asphyxiation from toxic gases was common. These , often paired with simple masks, were used to force air into victims' lungs, marking an initial shift from manual methods toward mechanical assistance in industrial emergencies. In the 1740s, English physician John Fothergill advanced these concepts by advocating the use of to distend the lungs of apparently drowned individuals, reporting William Tossach's successful 1744 case of mouth-to-mouth to restore breathing in a suffocated coal miner, while advocating mouth-to-mouth respiration as a preferred method for reviving apparently individuals over due to its availability and controlled force. Fothergill's work emphasized the physiological need for lung inflation to revive asphyxiated patients, influencing subsequent European resuscitation practices focused on and suffocation. The early 20th century saw the introduction of more sophisticated mechanical resuscitators, exemplified by the 1907 Pulmotor developed by Heinrich Dräger in . This portable device, powered by pressurized oxygen, alternated positive and negative airway pressures to simulate breathing for asphyxiation victims, and it was the first commercially produced ventilator of its kind. Mass production ramped up in the 1910s, with widespread deployment during for treating gas poisoning and battlefield injuries. By the , precursors to the , such as earlier negative-pressure chamber designs, began influencing the evolution toward more reliable portable resuscitators, though these early innovations suffered from significant limitations including bulkiness, which hindered mobility, and unreliability in consistent pressure delivery, often leading to inadequate or . Despite these challenges, the adoption of such devices in regions with high incidences of and industrial accidents contributed to substantial reductions in mortality rates for respiratory emergencies, transforming outcomes from near-certain fatality to viable recovery in many cases.

Mid-20th Century Advancements

In the , a pivotal advancement in resuscitator design emerged with the invention of the self-inflating bag by Danish anesthesiologist Henning Ruben and engineer Holger Hesse, introduced commercially in 1956 as the Ambu bag. This device represented the first portable, disposable resuscitator featuring non-rebreathing valves, including the innovative Ruben valve, which allowed for efficient one-way airflow without the need for external gas sources or electricity, enabling reliable manual ventilation in resource-limited settings. Unlike earlier bellows-based systems, the Ambu bag's lightweight rubber construction—initially weighing around 500 grams—facilitated immediate use by a single operator, marking a shift toward user-friendly, field-deployable tools that enhanced emergency response capabilities post-World War II. During the and , further refinements focused on improving safety, monitoring, and compatibility, including the integration of oxygen inlet ports to deliver enriched air mixtures up to 100% oxygen, which addressed limitations in hypoxic environments. A significant material transition occurred from durable but cumbersome metals and rubbers to lightweight plastics like (PVC), promoting disposability, easier sterilization, and reduced infection risk in clinical and prehospital settings. These changes drastically lowered device weight from over 5 in prior metal-framed models, such as early 20th-century Pulmotors, to under 1 , boosting portability for crews and . Key milestones in standardization came in the 1970s, when the () endorsed the inclusion of BVM resuscitators in () protocols as part of training, building on its 1963 formal adoption of techniques. This endorsement, alongside the 1979 National Conference on CPR and Emergency Cardiac Care, accelerated the integration of BVMs into global emergency guidelines, emphasizing their role in maintaining oxygenation during . By the late 1970s, these portable, reliable devices had achieved widespread adoption in , contributing to enhanced survival outcomes in out-of-hospital s through more effective ventilation support.

Clinical Applications

Indications for Use

Resuscitators, such as bag-valve-mask (BVM) devices, are primarily indicated for managing apnea during , where immediate ventilatory support is essential to maintain oxygenation and circulation. They are also deployed in cases of severe , such as peripheral oxygen saturation (SpO2) below 90% despite supplemental oxygen, to prevent further deterioration in perfusion. Additional primary indications include airway obstruction from foreign bodies or anatomical issues, and respiratory compromise induced by , such as chest injuries impairing . In specific clinical scenarios, resuscitators play a in out-of-hospital (OHCA), where BVM is a standard component of protocols to deliver oxygen during chest compression pauses. emergencies, including during induction or recovery, often necessitate their use to bridge until or recovery of spontaneous breathing. For , positive pressure via resuscitator is recommended if heart rate remains below 100 beats per minute after initial and , with particular considerations for preterm infants who may require adjusted oxygen concentrations. Evidence-based criteria for initiation include absent respirations (apnea) or severe in adults that compromises . Ventilation should continue until spontaneous breathing resumes, an advanced airway is secured, or is achieved, with monitoring for chest rise and improving SpO2 to confirm efficacy. Contraindications to resuscitator use include an intact gag reflex without prior , as this increases the risk of and during mask application. Resuscitators are not suitable for or long-term ventilation, as they are designed for acute, intermittent support rather than sustained mechanical assistance.

Operation Techniques

Operation of a resuscitator, such as a bag-valve-mask (BVM) device, begins with proper setup to ensure airway patency and effective delivery of oxygen-enriched air. Select the appropriate mask size based on the patient's age—adult masks for patients over 8 years or weighing more than 20 kg, pediatric for younger children, and masks for neonates and small infants—to achieve a secure seal without excessive pressure on the face. Attach the mask to the self-inflating bag and connect an oxygen source if available, delivering at least 15 L/min to achieve high inspired oxygen concentrations up to 100%. Position the patient and perform the head-tilt/chin-lift maneuver to open the airway, or use jaw thrust if cervical spine injury is suspected; insert an in unresponsive patients without a reflex or a if semi-conscious to prevent obstruction. The ventilation sequence involves delivering controlled breaths to mimic normal while avoiding . Squeeze the bag firmly but slowly over approximately 1 second to deliver a of 6-7 mL/kg of ideal body weight—typically 400–600 mL for adults, 4–8 mL/kg (e.g., 15–30 mL for term neonates) for newborns, and 6–7 mL/kg for older infants and children—observing for visible chest rise as the primary indicator of adequate delivery. Administer breaths at a rate of 10–12 per minute for adults (one every 5–6 seconds), 20–30 per minute for infants and children, and 30–60 per minute for neonates (as of 2025 guidelines). Allow full between breaths by releasing the bag completely, and avoid excessive force to prevent gastric . Continuous monitoring during operation is essential to adjust and prevent complications. Assess for symmetric chest rise and bilateral breath sounds after each breath to confirm inflation; watch for signs of gastric distension, such as abdominal movement or regurgitation, which may indicate improper seal or excessive pressure. Check for air leaks around the by listening for hissing sounds and readjust the seal as needed, using or if available to target above 94% and end-tidal CO2 between 35-45 mm Hg. If leaks persist, reposition the head or switch to a two-person . For advanced scenarios, employ a two-person to enhance seal efficacy and reduce operator fatigue, where one provider maintains seal using both hands (e.g., thumbs on the mask connector and fingers lifting the ) while the second squeezes the bag. This method delivers more consistent volumes compared to one-person operation. When integrating with (CPR), deliver two breaths after every 30 chest compressions in adults (30:2 ratio) until an advanced airway is placed, then transition to asynchronous ventilation at 10 breaths per minute without interrupting compressions. In pediatric CPR (infants and children), use a 15:2 ratio for two rescuers or 30:2 for one rescuer (as of 2025 guidelines).

Modern Innovations and Guidelines

Recent Technological Advances

In response to the , significant innovations in automated resuscitators emerged post-2010, focusing on low-cost, portable designs to bridge ventilator shortages in overwhelmed healthcare systems. These devices automate the of bag-valve masks (BVMs), providing consistent without requiring constant manual effort. Notable examples include the Emergency-Vent, a robotic gripper-based system developed in 2020 to deliver controlled breaths at a production cost of approximately $100–$200, and the RepRapable automated BVM, which uses off-the-shelf and 3D-printed components for rapid assembly in resource-limited environments. Other NIH-supported research highlighted designs like the ABCD , costing around $400–$800 in production, emphasizing portability and ease of assembly for use. Modern automated resuscitators incorporate smart features such as integrated pressure and flow sensors for real-time monitoring, achieving accuracies of ±8% or better in validated systems, which helps prevent over- or under-ventilation. Battery-powered models, like the AuRes automatic resuscitator, offer 2 hours of continuous operation on battery, suitable for short-term field or transport scenarios. While AI-driven rate adjustments remain more prevalent in full ventilators, resuscitator prototypes have adopted automated feedback loops to maintain respiratory rates based on sensor data, enhancing reliability during crises. Key innovations include hybrid manual-automatic modes, as seen in devices like the OP-Vent, which allow seamless switching between operator-squeezed and mechanized compression for adaptive clinical needs. Disposable single-use variants, such as the Ambu SPUR II and Dispo-Bag resuscitators, utilize medical-grade, latex-free materials to minimize risks, proving essential for pandemic protocols and single-patient applications. Additionally, 3D-printable prototypes, exemplified by the AIR (Automated Inflating Resuscitator), enable on-demand production in disaster zones using accessible printers and materials. These advances have facilitated greater adoption in low-resource settings, where manual ventilation often leads to inconsistent delivery. Field trials and bench evaluations, including a 2022 biological study on mechanical resuscitators, demonstrated stable volumes and effective oxygenation maintenance in porcine models, outperforming manual methods in consistency and reducing clinician fatigue. A 2024 implementation study in resource-constrained areas, such as on the Unisabana-HERONS low-cost mechanical ventilator in , further confirmed stable arterial blood gases (PaO2 ~69-80 mmHg, PaCO2 ~38-40 mmHg) and SpO2 levels (~94-96.8%) comparable to standard ventilators, supporting their role in improving outcomes during outbreaks and emergencies.

Current Standards and Training

Current international guidelines for resuscitator use emphasize standardized protocols to optimize outcomes in scenarios. The 2025 () guidelines recommend bag-valve-mask (BVM) ventilation as a primary strategy during out-of-hospital (OHCA), delivering sufficient (approximately 500–600 mL) to achieve visible chest rise while avoiding excessive ventilation that could lead to gastric insufflation or reduced . Similarly, the European Resuscitation Council (ERC) Guidelines 2025 endorse BVM as essential equipment for standard CPR in OHCA, prioritizing it alongside chest compressions in the initial response algorithm. In low-income settings, the (WHO) standards advocate for manual resuscitators, such as self-inflating bag-and-mask devices, as the cornerstone of basic neonatal and adult resuscitation due to their simplicity, reusability, and minimal resource requirements. Training protocols for resuscitator proficiency are integrated into (BLS) certification programs, which mandate hands-on sessions to build competency. AHA-accredited BLS courses typically include approximately 2 hours of practical training using simulation manikins to practice BVM techniques, , and scenario-based application, ensuring rescuers can maintain an airtight seal and deliver appropriate ventilation rates of 10 breaths per minute without interrupting compressions. These sessions emphasize retention through deliberate practice, with certification requiring demonstrated competence in adult, child, and infant scenarios. Recertification is required every 2 years to reinforce these skills, as proficiency can decline without regular refreshers. Regulatory frameworks ensure resuscitators meet safety and performance benchmarks. In the United States, the (FDA) classifies manual resuscitators as Class II devices, subjecting them to 510(k) premarket notification and general controls to mitigate moderate risks associated with ventilation delivery. Internationally, the ISO 80601-2-84:2023 standard specifies requirements for basic safety and essential performance of ventilators, including resuscitators, through rigorous testing of flow accuracy, pressure limits, and alarm functions to support reliable use in prehospital environments. Following the , virtual reality (VR) training modules for resuscitation skills have gained traction, with over 30% of programs incorporating them by 2025 to enhance realism and accessibility while reducing infection risks during hands-on practice.

Limitations and Safety Considerations

Potential Risks

The use of resuscitators, particularly bag-valve-mask (BVM) devices, carries risks of , defined as lung injury resulting from excessive airway pressure during ventilation. Clinical studies indicate that , including and , occurs in approximately 13% of cases involving positive pressure ventilation over several days, with manual BVM contributing to acute instances through vigorous squeezing or inadequate pressure monitoring. This risk is heightened when peak inspiratory pressures exceed safe thresholds, potentially leading to alveolar rupture. Gastric insufflation represents another common complication, where air enters the instead of the lungs, increasing the likelihood of regurgitation and . In scenarios involving BVM , gastric insufflation has been observed in up to 30% of cases, particularly with high volumes or short inspiratory times, and can result in rates of 2.5% during procedures. Untrained operators exacerbate this issue, as improper mask seal or excessive force promotes esophageal airflow. Device-specific risks include valve malfunctions in manual resuscitators, which can lead to rebreathing of exhaled and subsequent . Mis-assembly or defective non-rebreathing valves can cause inadequate ventilation or complete failure, resulting in and elevated levels. In powered resuscitators, prolonged delivery of high fractional inspired oxygen (FiO2 >60%) poses a risk of , manifesting as pulmonary damage from and impaired . This is particularly relevant in extended use, where can induce bronchitis-like symptoms and . Patient-specific factors amplify these risks; pediatric patients, with their smaller and more compliant airways, face higher susceptibility to and overdistension from standard adult devices. Additionally, expired air resuscitation methods historically carried transmission risks, but post-2020 guidelines incorporating have significantly mitigated this concern during scenarios. Complications during resuscitation are predominantly due to improper mask seals leading to hypoventilation or aspiration, as evidenced by observational data from emergency interventions.

Best Practices for Mitigation

To minimize risks associated with resuscitator use, preventive measures focus on device configuration and operational techniques. Pop-off valves on bag-valve-mask (BVM) resuscitators, particularly for pediatric use, should be set to limit peak inspiratory pressure to approximately 35 cmH₂O, preventing barotrauma while ensuring adequate tidal volumes (higher limits of 45-60 cmH₂O apply for adults). Implementing a two-rescuer technique, where one provider maintains a tight mask seal using both hands while the other squeezes the bag, reduces mask leaks by approximately 50% compared to single-rescuer methods, thereby improving ventilation efficacy. Regular device checks, including inspection for cracks, valve functionality, and mask integrity as per manufacturer guidelines (typically quarterly for reusable components and after each use for cleaning), ensure reliability and prevent equipment failure during emergencies. Infection control protocols have been strengthened following the , emphasizing disposable components to reduce cross-contamination. Single-use masks are recommended for BVM resuscitators in high-risk scenarios, as they eliminate the need for disinfection and minimize aerosol transmission of pathogens. For powered resuscitators, incorporating high-efficiency particulate air () filters on the exhalation port captures viral particles, aligning with guidelines for aerosol-generating procedures. Monitoring protocols integrate continuous waveform to verify effective , targeting an end-tidal CO₂ (ETCO₂) level greater than 10 mmHg, which confirms adequate pulmonary blood flow and chest compression quality during . If BVM is required for more than 10 minutes without achieving stable oxygenation or if complications arise, transitioning to an advanced airway device, such as a supraglottic airway or endotracheal tube, is advised to sustain long-term support. In disaster settings, where resource is common, manual resuscitators like self-inflating BVMs are prioritized over powered units to conserve oxygen supplies and avoid dependency on or compressed gas, which may be depleted or unavailable.

References

  1. [1]
    emergency resuscitator - Product Classification - FDA
    Device, emergency resuscitator. Definition, Emergency ventilator incorporating bag and valve, intended to provide emergency respiratory support by means of ...
  2. [2]
    Manual resuscitator - MEDEVIS
    A hand-operated device designed to provide or assist ventilation in patients who are apnoeic or exhibit inadequate respiration.
  3. [3]
    Understanding your bag-valve-mask resuscitator - PMC - NIH
    The self-inflating resuscitator or bag-valve-mask (BVM) is a basic but essential part of an anaesthetist's toolkit for resuscitation.
  4. [4]
    Ambu Bag: Celebrating 65 Years of Lifesaving Innovation - History
    Many bellows and resuscitators were introduced in the 1950s and 60s, but they “were considered clumsy and tiring in prolonged use and rendered largely obsolete” ...
  5. [5]
    Resuscitators - Virtual Museum
    Dräger produced their first resuscitator, the Pulmotor, in 1907. The Pulmotor remained in use until the 1940s. ... The Lung Motor, a resuscitator manufactured in ...Missing: types | Show results with:types
  6. [6]
    Chapter 16: Portable Ventilation Devices - AccessAnesthesiology
    There are two types of portable manual resuscitators: self-inflating and flow-inflating systems. +++. SELF-INFLATING SYSTEMS. ++. Self-inflating manual ...
  7. [7]
    Resuscitators Selection Guide: Types, Features, Applications
    Manual resuscitators come in two sizes: adult and pediatric. Pediatric resuscitators are for children. Pediatric resuscitators come with a pop-off valve. When ...
  8. [8]
    Understanding your bag-valve-mask resuscitator - BJA Education
    May 2, 2023 · Like the modern ventilator, the BVM resuscitator was developed during the polio epidemic in 1953. The combination of the mass demand for manual ...
  9. [9]
    Resuscitation Equipment & Their Uses - Emcare Services
    Jul 20, 2023 · A resuscitator is a medical device that uses positive pressure to inflate the lungs of a person, who's unconscious and has stopped breathing ...
  10. [10]
    RESUSCITATOR Definition & Meaning - Merriam-Webster
    Oct 14, 2025 · especially : a device used to restore or assist the respiration of a person who is not breathing normally. Medical Definition. resuscitator.
  11. [11]
    How To Do Bag-Valve-Mask (BVM) Ventilation - Critical Care Medicine
    Bag-valve-mask (BVM) ventilation is the standard method for rapidly providing rescue ventilation to patients with apnea or severe ventilatory failure.
  12. [12]
    Part 7: Adult Basic Life Support: 2025 American Heart Association ...
    Oct 22, 2025 · When ventilating adult patients in cardiac arrest, it is reasonable to give enough tidal volume to produce visible chest rise while avoiding ...
  13. [13]
    The Self-Inflating Resuscitator – Evolution of an Idea - Sage Journals
    In 177 AD, Galen used bellows to inflate the lungs of various animals. – “if when the animal is already dead, you blow air in through the larynx, you will of ...
  14. [14]
  15. [15]
    [PDF] Laerdal Silicone Resuscitators - NET
    The Laerdal Silicone Resuscitator is a self-inflating manual resuscitator that is intended for patients requiring total or intermittent ventilatory support.
  16. [16]
    Bag-Valve-Mask Ventilation - StatPearls - NCBI Bookshelf - NIH
    May 3, 2025 · The tidal volume needed for an adult is typically 500 to 600 mL ... [15] Additionally, using low-pressure, low-volume insufflation ...Continuing Education Activity · Introduction · Preparation · Technique or Treatment
  17. [17]
    [PDF] A HealthTech Report - PATH
    Type of bag material: In general, resuscitator bags are produced in silicone, PVC, or rubber. Silicone is often the chosen material for reusable resuscitators ...<|separator|>
  18. [18]
    Bag-Valve-Mask (BVM) Ventilation • LITFL • CCC Airway
    Oct 6, 2024 · self-inflating resuscitation device · oxygen inlet nipple · air intake valve · oxygen reservoir with two one way valves · non-rebreathing valve that ...
  19. [19]
    Manual bag valve mask ventilation performance among respiratory ...
    Nov 1, 2020 · The mean ventilation parameters among all participants include a tidal volume of 599.70 ml, pressure of 26.35 cmH2O, flow rate of 26.35 l/min, ...
  20. [20]
    It's In The Bag: Tidal Volumes in Adult and Pediatric Bag Valve Masks
    The adult BVM provided a mean tidal volume of 807.7mL versus the pediatric BVM providing 630.7mL, both of which exceeded the upper threshold of 560mL of tidal ...
  21. [21]
    BVM Air Volume Guide: Correct Ventilation Techniques |
    Sep 5, 2025 · What is the recommended tidal volume for adult BVM ventilations? The American Heart Association recommends delivering 500-600 mL of air ( ...
  22. [22]
    Bag-Valve-Mask Ventilation Technique - Medscape Reference
    Jan 29, 2024 · This basic airway management technique allows oxygenation and ventilation of patients until a more definitive airway can be established.
  23. [23]
  24. [24]
    Henning Ruben • LITFL • Medical Eponym Library
    Dec 5, 2021 · This prompted Ruben, to conceive the idea of the self inflating bag designed to provide manual artificial ventilation without compressed gases.
  25. [25]
    Ambu Bag - Wood Library-Museum of Anesthesiology (WLM)
    In 1954, a shortage of oxygen deliveries to Danish hospitals prompted him to invent the first self-inflating resuscitator, the Air Mask Bag Unit (AMBU).
  26. [26]
    MOUTH-TO-MASK RESUSCITATION - JAMA Network
    Mouth-to-mask resuscitation uses a mask, flutter valve, and outflow valve to prevent expired air flow, and an inflow valve for fresh air. It is safe, simple, ...
  27. [27]
    Part 6: Advanced Cardiovascular Life Support | Circulation
    Rescue breathing (ventilation using exhaled air) will deliver approximately 16% to 17% inspired oxygen concentration to the patient, ideally producing an ...
  28. [28]
    Mouth-to-mask ventilation: a superior method of rescue breathing
    Mouth-to-mask ventilation produced tidal volumes (1,093 ml and 1,200 ml) not significantly different from those seen with endotracheal intubation. If clinical ...
  29. [29]
    A Reappraisal of Mouth-to-Mouth Ventilation During Bystander ...
    Studies have identified reticence to perform mouth-to-mouth ventilation as a significant barrier to more frequent performance of bystander CPR.
  30. [30]
    History of CPR | American Heart Association CPR & First Aid
    It's as old as the healing arts, as new as a drone delivering an automated external defibrillator (AED). And from primitive methods (like whipping an ...
  31. [31]
    [PDF] CPR MEDICAL DEVICES INC. OPERATING MANUAL OXYLATOR ...
    The OXYLATOR EMX is an oxygen powered resuscitator/ inhalator which requires no power source (i.e. battery) other than the supply of compressed oxygen in an ...
  32. [32]
    [PDF] CAREvent® ALS - User Manual - O-Two Medical Technologies
    Jul 19, 2023 · The CAREvent® ALS Handheld Resuscitator is a pneumatically powered, time/volume cycled ventilatory resuscitator with the added feature of a ...
  33. [33]
    EMT Resuscitator - Allied Medical LLC
    Advanced pneumatic technology limits and regulates gas flow to 40 lpm during the manually-triggered breath, but will allow flow rates of up to 115 lpm during a ...
  34. [34]
    (PDF) The Effectiveness of Oxygen-Powered Inhalation Devices in ...
    Jun 29, 2020 · An effective method for ventilation in such scenarios might be a triggered oxygen-powered resuscitator. Methods: We report seven ...Missing: function | Show results with:function
  35. [35]
    Transport Ventilation | Respiratory Therapy
    Oxygen driven resuscitators, also known as demand valves, may be found on an ambulance. These devices provide RCPs with the ability to ventilate the patient ...
  36. [36]
    Early ICU Ventilators - Virtual Museum
    The Bennett PR-2 was used often used to continuously ventilate patients in early ICUs. ... The Bird ventilators, such as the Bird Mark 7 pictured here, were used ...Missing: resuscitator | Show results with:resuscitator
  37. [37]
    Breathing Apparatus for Mine Rescue in the UK, 1890s–1920s
    Inspired in some respects by diving gear, the apparatus was used in mine rescue work and operations to recover mines after an explosion. The chapter focuses on ...
  38. [38]
    [PDF] on the Recovery ofa Man Dead
    a Man Dead in Appearance, by d/- tending the Lungs with Air. Printed at. Edinburgh, 1744; by John Fothergill,. Licent. Coll. Med. Lond. Read Feb. z1 ...
  39. [39]
    ARTIFICIAL RESPIRATION, THE HISTORY OF AN IDEA
    Fothergill, probably the greatest worker of this resuscitation ... Thus far the techniques of artificial respiration have entered the fields of physiology,.
  40. [40]
    [PDF] It began with the Pulmotor The History of Mechanical Ventilation
    His Pulmotor created positive and negative airway pressure alternatively and was powered by oxygen under pressure. In 1907. Heinrich Dräger was awarded the.Missing: steam- | Show results with:steam-
  41. [41]
    Draeger Pulmotor - Wood Library-Museum of Anesthesiology (WLM)
    The Pulmotor was introduced by the Draeger company, of Lubeck, Germany, in 1907. This example was made by a Draeger branch in Pittsburgh, PA.Missing: steam- | Show results with:steam-
  42. [42]
    A practical mechanical respirator, 1929: the "iron lung" - PubMed
    Popularly named the iron lung, the Drinker respirator supported thousands of patients afflicted with respiratory paralysis during the polio era.
  43. [43]
    From Mouth‐to‐Mouth to Bag‐Valve‐Mask Ventilation: Evolution ...
    May 27, 2014 · This paper draws a historical view of manual ventilation's evolution ... self-inflating bags to be used as complete resuscitation devices.
  44. [44]
    Bellows and Self-Inflating Bags | Request PDF - ResearchGate
    Aug 7, 2025 · 238, 239 The first self-inflating resuscitator bag was designed by the Danish anesthesiologist Henning Ruben (1914Ruben ( to 2004)-the AMBU bag.Missing: invention | Show results with:invention
  45. [45]
    US7051596B1 - Manual resuscitators with integral manometer
    A manometer integrated into a manual resuscitator includes a housing with a communication path, and a pressure gauge integrally forming part of the housing.
  46. [46]
    Standards and Guidelines for Cardiopulmonary Resuscitation (CPR ...
    The National Conference on Cardiopulmonary Resuscitation (CPR) and Emergency Cardiac Care (ECC) held in September 1979 developed standards and guidelines.
  47. [47]
    Cardiopulmonary Resuscitation | Circulation
    Dec 19, 2006 · The 2005 American Heart Association guidelines endorse rapid defibrillation, especially for inpatients, in which the goal is delivery of a ...
  48. [48]
    Bag-Valve-Mask Ventilation and Survival From Out-of-Hospital ...
    Nov 12, 2023 · This novel multicenter study demonstrates that lung inflation occurs infrequently with BVM ventilation during 30:2 CPR for OHCAs. Ventilation ...
  49. [49]
    Part 5: Neonatal Resuscitation: 2020 American Heart Association ...
    Oct 21, 2020 · Most newborns who are apneic or have ineffective breathing at birth will respond to initial steps of newborn resuscitation (positioning to open ...
  50. [50]
    The Dos and Don'ts of Bag-Valve Mask Ventilation - JEMS
    Aug 8, 2014 · Recognize the need to ventilate a patient, and do so immediately. Hypoventilation occurs when the rate of spontaneous ventilations falls below 8 ...Missing: indications | Show results with:indications
  51. [51]
  52. [52]
    MIT volunteers created open-source, low-cost ventilator for Covid-19
    Apr 22, 2020 · Engineers at the U of I developed a prototype for a simple emergency ventilator, called Rapid Vent, that would only cost $100 to $200 to make.
  53. [53]
    Recent advances in low-cost, portable automated resuscitator ... - NIH
    Dec 2, 2021 · Automated AMBU bags or resuscitator devices aim to assist patient breathing via compressing and releasing the AMBU bags at specific frequency, ...
  54. [54]
    Introducing a novel respiratory function monitor for neonatal ...
    The monitor automatically calculates and displays expiratory tidal volume (VTE) passing through the sensor with an accuracy of ±0.5 mL or ±8% and ventilation ...Simulation And Education · Introduction · Discussion<|control11|><|separator|>
  55. [55]
    AuRes automatic resuscitator – Polbionica
    The advantage of the AuRes automatic resuscitator is its lightweight housing, minimalist design, and battery sufficient for 2 hours of continuous operation.
  56. [56]
    Ambu® SPUR®II Disposable Resuscitator
    A disposable manual resuscitator featuring a single-shutter valve for reliable functionality, and a thin self-inflating bag with excellent recoil.Missing: infection COVID- 19
  57. [57]
    Dispo-Bag™ Manual Resuscitator - GaleMed Corporation
    Disposable for Single-Patient Use · Reduces the risk of cross-contamination—ideal for infection control protocols. · Latex-Free and Medical-Grade Materials · Safe ...
  58. [58]
    Automated Inflating Resuscitator (AIR): Design and Development of ...
    In this report we describe the design and development of the automated inflating resuscitator (AIR), a 3D-printed ventilator training device.Missing: field 2020-2023
  59. [59]
    Biological evaluation of a mechanical ventilator that operates by ...
    Mar 3, 2022 · This study evaluates the physiological responses of healthy swine maintained under volume- or pressure-controlled mechanical ventilation by a mechanical ...Missing: impact | Show results with:impact
  60. [60]
    The development and implementation of a low-cost mechanical ...
    May 15, 2024 · The effectiveness and safety outcomes included maintaining arterial blood gases and pulse oximetry saturation (SpO2), respiratory pressures and ...
  61. [61]
    Part 3: Adult Basic and Advanced Life Support: 2020 American ...
    Oct 21, 2020 · These guidelines are designed primarily for North American healthcare providers who are looking for an up-to-date summary for BLS and ALS for ...
  62. [62]
    Neonatal resuscitation in low-resource settings: What, who, and how ...
    Evidence from several observational studies shows that facility-based basic neonatal resuscitation may avert 30% of intrapartum-related neonatal deaths.
  63. [63]
    Basic Life Support (BLS) Course Options
    The online portion of HeartCode BLS can be completed in approximately 1 to 2 hours. Time to complete the hands-on skills session varies from 60 ...Missing: protocol | Show results with:protocol
  64. [64]
    Resuscitator, Manual, Non Self-Inflating - Product Classification - FDA
    A non self-inflating manual resuscitator, also called a hyperinflation system, is a manual ventilator intended to ventilate a patient by forcing a volume of ...
  65. [65]
    ISO 80601-2-84:2023
    ### Summary of ISO 80601-2-84 for Resuscitators or Ventilators Performance Testing
  66. [66]
    Part 12: Resuscitation Education Science: 2025 American Heart ...
    Oct 22, 2025 · Virtual reality (VR) and augmented reality (AR) can increase realism during resuscitation training. However, while VR can improve knowledge ...
  67. [67]
    Barotrauma and Mechanical Ventilation Treatment & Management
    Feb 1, 2024 · In more than 900 patients with a cumulative incidence of barotrauma of 13% over the first 4 study days, no relation was detected between peak ...
  68. [68]
    Sudden severe barotrauma from self-inflating bag-valve devices
    These devices are generally safe, but minor deviations in their recommended use can expose patients to airway pressures greater than 135 cm H2O. We present a ...
  69. [69]
    and pressure-controlled ventilation in children with the i-gel - Medicine
    In several studies, the incidence of gastric insufflation varied from 0% to 30% depending on the PIP and the presence of a gastric tube. We hypothesized that ...Missing: untrained | Show results with:untrained
  70. [70]
    Bag-Mask Ventilation during Tracheal Intubation of Critically Ill Adults
    Feb 18, 2019 · Hypoxemia is the most common complication during tracheal intubation of critically ill adults and may increase the risk of cardiac arrest ...
  71. [71]
    Manual Resuscitators Flashcards | Quizlet
    Defective/malfunctioning non-rebreather valve. III. Faulty pressure release valves. PEEP. Positive End-Expiratory Pressure. Common mechanical ventilator ...
  72. [72]
    Oxygen Toxicity in Critically Ill Adults - PMC - NIH
    Damage from excessive oxygen can arise from both direct pulmonary toxicity from a high FiO2 and from systemic effects of high PaO2 in blood and tissues (5, 37) ...
  73. [73]
    Practice of Excessive FIO2 and Effect on Pulmonary Outcomes in ...
    Prolonged high FIO2 in patients requiring mechanical ventilation worsens gas exchange, decreases ciliary efficacy, and produces hyperoxic bronchitis and ...
  74. [74]
    2023 International Consensus on Cardiopulmonary Resuscitation ...
    In-water expired-air resuscitation may be considered by trained rescuers ... infection transmission and on consistency with current recommendations on using PPE ...
  75. [75]
    Positive-Pressure Ventilation | Textbook of Neonatal Resuscitation
    These valves are usually set to release at 30 to 40 cm H2O pressure, but they are not reliable and may not release until higher pressures are achieved. Some ...
  76. [76]
    Oxygen Unit Maintenance | DAN Southern Africa
    Jun 6, 2018 · Thoroughly inspect your unit at least once a month and after every use. Make sure the cylinder is not just full, but full of oxygen, and that the oxygen washer ...
  77. [77]
  78. [78]